dynamic BIOSENSORS AS-2-Ga-lfs v5.1 Adapter Strand User Manual

September 6, 2024
dynamic BIOSENSORS

AS-2-Ga-lfs v5.1 Adapter Strand

Product Information

Specifications

  • Product Name: heliX+
  • Model: ADAPTER STRAND 2 with green dye Ga
  • Prehybridized with ligand strand
  • Manufacturer: Dynamic Biosensors GmbH & Inc.
  • Order Number: AS-2-Ga-lfs v5.1
  • Key Features: 2 spots with 2 different anchor sequences for
    DNA-encoded addressing

Product Description

The heliX+ ADAPTER STRAND 2 is a product prehybridized with a
ligand strand and features 2 spots with 2 different anchor
sequences for DNA-encoded addressing. It comes with green dye Ga
for easy visualization.

Contents and Storage Information:

  • Material: Adapter strand 2 – Ga – lfs

  • Cap Color: White

  • For research use only. Limited shelf life, check expiry date on
    label

  • To avoid freeze-thaw cycles, please aliquot the nanolever

  • Buffer: TE40

Product Usage Instructions

Preparation | MIX&RUN

  1. Mix Adapter strand 1 – Ga (400 nM) and conjugated Ligand strand
    (500 nM) at 1:1 ratio (v/v).

  2. Incubate the solution from step 1 at room temperature at 600
    rpm for 30 minutes for complete hybridization.

  3. Mix the solution from step 2 with Adapter strand 2 – Ga – lfs
    (200 nM) at 1:1 ratio (v/v).

  4. The solution is now ready for use in biochip
    functionalization.

Note: The stability of the solution is
dependent on the ligand molecule’s stability.

FAQ

Q: Can the heliX+ ADAPTER STRAND 2 be used for multiple

experiments?

A: The product is designed for single-use
experiments due to its limited shelf life and stability
concerns.

Q: How should I store the heliX+ ADAPTER STRAND 2?

A: Store the product according to the storage
information provided in the manual, and avoid exposing it to
frequent freeze-thaw cycles.

heliX+
User Manual
ADAPTER STRAND 2
with green dye Ga and prehybridized with ligand strand
Dynamic Biosensors GmbH & Inc. AS-2-Ga-lfs v5.1

Key Features
· Adapter strand 2 for functionalization of heliX® Adapter Chip Spot 2. · Compatible with heliX® Adapter Chip. · Includes Adapter strands for 50 regenerations. · Ideal for MIX&RUN sample preparation. · Adapter strand 2 carries a hydrophilic green dye (Ga) with a single negative net charge.
heliX® Adapter Chip Overview
2 spots with 2 different anchor sequences for DNA-encoded addressing.

Product Description

Order Number: AS-2-Ga-lfs

Table 1. Contents and Storage Information Material Adapter strand 2 – Ga – lfs

Cap White

Concentration Amount 200/250 nM 5 x 200 µL

For research use only. This product has a limited shelf life, please see expiry date on label.

To avoid many freeze thaw cycles please aliquot the nanolever.

Buffer TE40 [1]

Storage -20°C

AS-2-Ga-lfs v5.1

www.dynamic-biosensors.com

Page 1 of 3

Preparation | MIX&RUN
In-solution hybridization of adapter and ligand strands:
1. Mix Adapter strand 1 – Ga (400 nM) and conjugated Ligand strand (500 nM) at 1:1 ratio (v/v).
2. Incubate the solution of step 1 at RT at 600 rpm for 30 min to ensure complete hybridization.
3. Mix solution of step 2 and Adapter strand 2 – Ga – lfs (200 nM) at 1:1 ratio (v/v).
Solution is ready to use for biochip functionalization.
Stability of the solution is related to the stability of the ligand molecule.

Table 2. Additional material for functionalization of spot 1 and reference spot 2.

Material
Adapter strand 1 – Ga
Ligand strand carrying the conjugated ligand

Concentration 400 nM 500 nM

Buffer TE40 [1] PE40 [
2]

Related Product Name Adapter strand 1 with green dye Ga heliX® Amine Coupling Kit 1

Order No AS-1-Ga HK-NHS-1

Example

Required volume for 3 functionalizations: 100 µL with a final concentration of 100 nM.

Vial 1
Adapter strand 1 – Ga (400 nM)
25 µL

Conjugated Ligand strand (500 nM)
25 µL

Vial 2
Adapter strand 2 – Ga – lfs (200/250 nM)
50 µL

After incubation time, mix vial 1 and vial 2 to obtain 100 µL of ready-to-use DNA solution.

AS-2-Ga-lfs v5.1

www.dynamic-biosensors.com

Page 2 of 3

Contact

Dynamic Biosensors GmbH Perchtinger Str. 8/10 81379 Munich Germany

Dynamic Biosensors, Inc. 300 Trade Center, Suite 1400 Woburn, MA 01801 USA

Order Information order@dynamic-biosensors.com Technical Support support @dynamic-biosensors.com

www.dynamic-biosensors.com
Instruments and chips are engineered and manufactured in Germany. ©2024 Dynamic Biosensors GmbH | Dynamic Biosensors, Inc. All rights reserved.

[1] TE40: 10 mM Tris, 40 mM NaCl, 0.05 % Tween20, 50 M EDTA, 50 M EGTA [2] If the protein is not stable in PE40 (TE40, HE40), please check buffer compatibility with the switchSENSE® compatibility sheet.

AS-2-Ga-lfs v5.1

www.dynamic-biosensors.com

Page 3 of 3

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

dynamic BIOSENSORS User Manuals

Related Manuals