AIRPREP ACD 210 Cub Sampler Instruction Manual
- September 5, 2024
- AIRPREP
Table of Contents
AirPrep Cub Sampler™ (ACD 210 & ACD 220)
and Air Sampling System User Manual
PRODUCT OVERVIEW
AirPrep™ Cub Sampler ACD210 model has an internal battery
AirPrep™ Cub Sampler ACD220 model is plug-in only
1.1 PRODUCT DESCRIPTION
The AirPrep Cub Sampler is a dry filter air sampler that operates using
InnovaPrep’s unique, patented rapid filter elution products to provide samples
for a wide range of indoor and outdoor biomonitoring applications.
1.2 COLLECTION CAPABILITY
The AirPrep Cub Sampler is suited for the collection of bioaerosols (bacteria,
molds, and viruses) and other particulate matter, including sub-micron sized
particles, and airborne molecular contamination. The AirPrep Cub Sampler
efficiently captures micron- and submicron-sized particles, including viruses,
bacteria, pollen, molds and fungal spores, as well as non-biological particles
from 0.01 micron to 10+ microns.
- Users can select from three flowrate modes: 50 LPM, 100 LPM, or 200 LPM from the front panel as described in Section 4.
- Users can select from four timed modes: 30 minutes, 1 hour, 2 hours, or continuous from the front panel as described in Section 4.
- Expected continuous run times for ACD210 battery: 50LPM – 16hrs, 100LPM – 8hrs, 200LPM – 4hrs
- Since liquids are not used in the collector, the system can be operated in extreme temperatures (see specifications in Section 6).
1.3 MATERIALS PROVIDED:
The AirPrep Cub Sampler includes the power supply cord (Item #AC04076) and a
universal fitting installed in the base of the unit for common off-the-shelf
camera tripods.
*Upon receipt, always inspect components when unpacking. If damage is observed, do not use and contact InnovaPrep or local distributor immediately. Please use the original box and foam packaging in the event a return shipment is necessary.
1.4 MATERIALS REQUIRED BUT NOT INCLUDED- SEE AirPrep Consumables
Selection Guide
for more information.
The single-use AirPrep Filter & Elution Kit comes with everything you need to
recover a fluid sample (as shown in Figure 1).
Users have a choice between two kits with different elution buffer types:
- AirPrep Filter & Elution Kit with TRIS – Item AC8100T – Recommended for use with rapid molecular methods of analysis.
- AirPrep Filter & Elution Kit with PBS – Item AC8100P – Recommended for classical culture analysis.
The AirPrep Filter & Elution Kit includes:
- Elution Buffer Canister
- Elutor Adapter
- Filter
- Sample Cup
These kits and multipacks are available for purchase on the InnovaPrep website https://www.innovaprep.com/store .
METHOD OF OPERATION
The AirPrep Cub Sampler uses an electret filter to capture particles. To
recover the trapped particles, a novel Wet Foam Elution™ process is employed
to break the electrostatic charge and flush the particles off the filter into
a small liquid volume in seconds.
2.1 SAMPLE COLLECTION ON AN ELECTRET FILTER
The dry 52 mm electret filters are made with a combination of positively
charged and negatively charged fibers. This substantially increases the
collection efficiency of the filter and allows for the use of lower pressure
drop filters, which, in turn, allows for higher sampling rates for extended
periods using battery power.
2.2 INSTANT WET FOAM ELUTION™
The patented wet foam elution process uses a very specific high-quality foam
to elute the filter. The elution buffer is composed of water, a low
concentration surfactant (0.15% Tween 20), and a pH buffer (Tris or PBS). This
solution is carbonated, which dissolves significant amount of CO2 into the
fluid. During the elution process, the valve opens as the elution canister is
pushed into the elutor adapter that is fitted onto the filter, forcing the
elution buffer through the pressure orifice, through the filter and into the
sample cup. As the fluid passes from the high-pressure environment on one side
of the orifice, to the low-pressure environment on the other side, the
dissolved CO2 expands, comes out of the solution to form micro-bubbles. These
micro-bubbles increase the volume of the fluid six-fold or more as it passes
through the filter. The foam immediately collapses back into a liquid and is
ready for sample processing and analysis. InnovaPrep’s patented Wet Foam
Elution process ( https://www.innovaprep.com/patents ) is much more
efficient than liquid rinsing for the following reasons:
- Volume Expansion – When rinsing a filter with liquid, most of the liquid volume is used to fill the dead space inside the filter housing; only a small portion of the fluid covers the filter surface. This can be avoided to an extent by reducing the cross-sectional area of the fluid path across the filter, but a large portion of the liquid is still underutilized. Because our elution buffer is 80-90% gas, this fills the empty space without contributing to the final sample volume.
- Increased Viscosity – Liquid has a tendency toward “channeling” when flowing through a filter, creating an area of high flow in the center of the fluid path, while the portion of flow in contact with the filter surface is much slower. The higher viscosity of our elution buffer as a foam prevents channeling and allows for a more uniform flow through the interstitial spaces of the filter.
- Bubble Dynamics – The micro-bubbles in the foam behave as deformable solids. As they travel through the filter they move as a rigid body.
- Exfoliating Action – As the micro-bubbles in the foam impact against each other and burst, the turbulence and energy produced helps to lift particles that are adhering to the filter fibers.
- Buffered Elution – The buffered elution fluid is conductive and breaks down the electrostatic charges holding the particles to the electret filter, effectively releasing them in the liquid sample.
SAMPLER COMPONENTS
3.1 AIRPREP CUB SAMPLER COMPONENTS
- Omni-Directional Inlet Cap – protects from water ingress
- Inlet – screws off left to open for loading or recovering the filter cartridge. Screws on right to close
- Front Panel – Power on and Mode selections (see section 4 for instructions)
- Power Supply and Plug
- Universal Tripod Connection – for attaching to a common, off-the-shelf camera tripod is located on the bottom of the instrument (not shown)
PROCEDURES
This section includes some of the basic instructions for operation.
4.1 SETTING UP THE AIRPREP CUB SAMPLER FOR COLLECTION
Push the barrel plug of the power supply (Figure 2) into the power jack on the
bottom of the instrument. Plug the AC Power Cord into the power supply and
plug the other end into an appropriate wall outlet.
4.2 INSTALLING THE FILTER
Open the inlet cap by twisting it to the left to lift off. Insert the filter
into the top of the AirPrep Cub Sampler, as shown in figure 2. The filter
housing is designed to only fit in one way. Screw the inlet lid on (twisting
right) to close.
4.3 CONTROL PANEL INTRODUCTION
The control panel consists of four timer setting LED indicator lights, three
flowrate LED indicator lights, and one momentary pushbutton switch
(Hold/Mode).
Two variables constitute a “run mode”: flowrate and runtime.
To turn on and off:
- Press Hold/Mode for 2 seconds to Power on/power off
To change run modes
- press the Hold/Mode for 4 seconds
Reset Button
- A reset button is located on the bottom of each instrument, pressing this will result in an immediate system interrupt and reset of the system.
4.3.1 LED INDICATOR LIGHT
The Battery LED Indicator Light (Model 210 only) will flash in the following
situations:
- When a run mode is selected that exceeds the expected battery capacity
- After the battery is drained low enough to shut down the fan before the timer expired
- The battery LED indicator light will start to flash slowly as the charge approaches a critical level- increasing in frequency until it is steady on when the battery is at its preset critical charge.
The battery LED indicator will turn solid if the power supply is plugged in
and the system is charging. It will shut off when it is fully charged.
Battery Status Indicator
By holding the power button for 2 seconds, then release, the Power LED will
light up solid, followed by a combination of the five green LEDs on the left
side of the panel. Each green LED that is lit indicates 20% of the battery.
Example: If 3 of the 5 green LEDs are lit this would indicate approximately a
60% battery charge.
Flowrate LED Indicator Light
The flowrate LED indicator for the selected run mode will stay on while the
fan is running.
Timer LED Indicator Light
- The timer LED Indicator light will flash slowly as they countdown, as they approach the expiration they will increase in frequency until the unit times out and shuts down. The light will remain solid for the last 10% of the timer.
- If the timer is set to continuous, the LED indicator light will stay steadily on.
4.3.2 POWER ON
When powering-on the sampler, the previous run mode will be recalled and
started by default.
4.3.3 POWERING OFF MANUALLY BEFORE TIME EXPIRES
To power-off the Air Sampler before the time expires, hold the Hold/Mode
button for 2 seconds, the unit will power-off.
**4.3.4 POWER OFF DUE TO LOW BATTERY***
4.3.5 SELECTING THE FLOW RATE
-
Press Hold/Mode button for 4 seconds, if fan was already running it will shut off after the first two seconds.
a. All 3 Flowrate LED Indicator lights will flash -
Let up and momentarily press the Hold/Mode button
a. The 50LPM LED indicator light will flash. If desired, stop to select or -
Momentarily press the Hold/Mode button again
a. The 100LPM LED indicator light will flash. If desired, stop to select.
Additional momentary presses will continue to cycle through the 50, 100 and 200 LPM LEDs -
Press the Hold/Mode button for 2 seconds to save. Selected flowrate mode is saved, and LED indicator light illuminates steadily; all timer LED indicator lights start flashing
4.3.6 HOW TO SELECT A TIMED RUN MODE
-
Once the flow rate has been set. Press the Hold/Mode button momentarily.
a. All four timed mode LED indicator lights will flash -
Let up and momentarily press the Hold/Mode button
a. The 30 Min LED indicator light will flash. If desired, stop to select or -
momentarily press the Hold/Mode button again
a. The 1-hour LED indicator light will flash. If desired, stop to select.
Additional momentary presses will continue to cycle through the remaining timer modes. -
Press Hold/Mode button for 2 seconds to save selection. Selected flowrate mode is saved and LED indicator light illuminates.
a. Fan will start.
If a time & flowrate combination requires more battery than is available, the
battery LED will flash to alert you
4.4 RECOVERY OF THE SAMPLE
- When the run is completed, recover the filter from the sampler with gloved hands by twisting the inlet to the left to open.
- Remove the filter cassette, invert it top-side-down, and press onto the sample cup firmly until it clicks all the way around the cup. Perform this on a stable surface.
- Cap the filter with the elutor adapter as shown in Figure 6 (make sure the filter and elutor adapter are fully sealed).
- Place the filter assembly in a zip-type bag for transporting dry, or alternatively, elute the filter as described in the following section.
4.5 ELUTING THE SAMPLE FROM THE FILTER
To extract the captured particles from the filter:
- Place the sample cup, filter, and elutor adapter, as shown in Figure 6, on a stable surface.
- Firmly press the elution buffer canister into the elutor adapter. The elution buffer will be released as a foam from the canister through the filter when the canister valve is pressed down. Hold the canister down until all the foam is released. The foam quickly collapses into a liquid in the sample cup and available for sample processing and analysis.
4.6 CLEANING THE SAMPLER
Common off-the-shelf alcohol or bleach wipes can be used for cleaning and
disinfection of the entire sampler. If applicable, disconnect the AC power
plug prior to cleaning.
AIRPREP CUB SAMPLER- MODEL ACD210 BATTERY SAFETY INFORMATION
14.4V 3350mAh
Operating temperature:
- Charge: 10 – 45°C
- Discharge: -20 – 6O°C
Storage temperature:
Recommended 25+/-5°C at 50% state of charge
- Do not immerse the AirPrep Cub Sampler in liquid. Exposure to liquid may damage the battery or the circuits. As a result, the battery may generate heat, smoke, catch fire, or explode.
- Do not use or place the AirPrep Cub Sampler near an open flame, heater, or high temperature (above 80°C).
- Do not use unauthorized chargers. Use of an unauthorized charger could cause the battery to generate heat, smoke, catch fire, or explode.
Aerosol Sampling & Collection System Specifications
| ACD200 Bobcat| ACD210 & ACD220 Cub
---|---|---
Particle Range| Optimized for 0.01 μm to 10.0 μm, minimizes capture of
particles >11 μm
Inlet| Patented built-in omnidirectional super-isokinetic
Decontamination| Hydrogen peroxide or alcohol wipes
Environmental| Indoor or Outdoor Use in non-condensing environments
Technology| Dry electret filter media, built-in mass flow sensor for
consistent sampling rates| Dry electret filter media with calibrated sampling
mode
Temp/ Humidity Range| Operating temp. -2°C to 60°C Storage temp. -2°C to 40°C|
Operating temp. -2°C to 60°C
Storage temp. -2°C to 30°C
Flow| 100 or 200 L/min| 50, 100 or 200 L/min
Noise| 71.3 dBa at 200L/Min| 78.6 dBa at 200L/Min (without QuietDock)
68.3 dBa at 200L/Min with Quiet Dock (Sold Separately)
Physical Size| Filter: 4.5 cm diameter active filter in 6.0cm diameter
Case: 17cm x 16cm x 30cm (W/L/H)| Filter: 52mm diameter active filter
Case: 20.32cm x 12.7cm x 12.7cm (W/L/H)
Weight| 3.5kg with battery| 1.3kg w/ battery (ACD210), 1.1kg w/o battery
(ACD220)
Power Source| Universal wall supply: 82-265V (47-63 Hz)
Option A: Rechargeable 7.5 Ah Lithium Ion 28.8V
Option B: 4.8 Ah Nickel Metal Hydride 24V (2 each 12V sections)| Universal
wall supply: 82-265V (47-63 Hz) Lithium Ion 14.8V, 3350 mAH (ACD210 only)
Rated Input Current| 3.0A| 5.0A
Rated Input Voltage| 12 VDC| 24 VDC
Expected Run Times| Up to 14 hrs on continuous mode, or 6 days on extended
run.
Battery type dependent| 50 LPM- 16hrs
100 LPM- 8 hrs
200 LPM- up to 4 hrs
Battery type dependent
System Controls| Active control for consistent sampling rates, 5 Status lights
with light discipline for military applications, 5 user-selectable run modes
and system status check button| Intuitive controls with user selected flow
rate, timer controls and status lights
Mounting| Built-in tripod extends to 40″ at inlet| Built-in tripod mount. 72″
tripod sold separately
Hard Shell Rolling Travel Case| Customizable rugged case suitable for
transport and field use. Large enough to accommodate the instrument and
consumables.
Dimensions: Exterior: 25″ x 20″ x 12″ Interior: 21″ x 17″ x 10.5″
Sample Recovery
Filter Recovery Method | Wet Foam Elution |
---|---|
Final Sample Volume | ~ 6-7mL |
Recovery Efficiency | 86-95%* will vary based upon particle type and size |
Recovery Time | 11 ~ 20 Seconds |
Sample Recovery Power Requirement | None – single use consumables |
Sample-to-Sample Carryover | None – System uses single-use filters and an easy |
to clean flow path
Recovery Consumables: Single Use Kits| AC08100T- Filter and Elution Kit -Tris
AC08100P- Filter and Elution Kit -PBS
Click here for AirPrep Consumable Selection Guide
Recommended Storage Temperature:| 2-25 °C (for AirPrep Elution Buffer)
Please contact InnovaPrep at
customerservice@innovaprep.com ,
or an official InnovaPrep distributor for technical assistance and
troubleshooting.
InnovaPrep
www.innovaprep.com
132 E. Main Street, Drexel, MO 64742
Phone: +1 833-833-3763
General Inquiries: info@innovaprep.com
Sales: sales@innovaprep.com
Customer Service:
customerservice@innovaprep.com
Caution: Any changes, or modifications not expressly approved by
InnovaPrep could void the user’s authority to operate the equipment.
Note: This equipment has been tested and found to comply with the limits
for a Class A digital device, pursuant to part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environment. This
equipment generates, uses, and can radiate radio frequency energy and, if not
installed and used in accordance with the instruction manual, may cause
harmful interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.
FOR RESEARCH USE ONLY
© 2021-2023 InnovaPrep LLC. All rights reserved.
InnovaPrep, AirPrep, FluidPrep, CP Select, Concentrating Pipette, Wet Foam
Elution are trademarks or registered trademarks of InnovaPrep LLC. All other
brands and names contained herein are the property of their respective owners.
InnovaPrep products are patent protected
https://www.innovaprep.com/patents