IVI Foundation SC6540 Multiplexer IVI Driver Getting Started User Guide

September 7, 2024
IVI Foundation

SC6540 Multiplexer IVI Driver Getting Started

Product Information

Specifications

  • Product: Multiplexer SC6540
  • Manufacturer: IKONIX Group
  • IVI Driver Version: 1.1
  • Location of Help Document: IVI
    FoundationIVIDriversSC6540SC6540.chm

Product Usage Instructions

1. IVI Driver Setup

Follow these steps to set up the IVI Driver:

  1. Download the IVI Driver from the manufacturer’s website.

  2. Run the self-extracting setup file and follow the installation
    wizard instructions.

  3. If prompted to download IVI Shared Components, visit the IVI
    Foundation Website and download the appropriate version (32-bit or
    64-bit).

  4. Install the shared components and continue with the IVI Driver
    installation.

  5. Optionally install the source code of the IVI Driver if
    needed.

  6. The IVI driver will be installed under the path IVI
    FoundationIVI, with *.dll files in the Bin folder and help
    documents in ..DriversSC6540.

2. Getting Started with C

To begin using the IVI-COM driver with C#, follow these
steps:

  1. Ensure you have the SC6540 IVI Driver, IVI Shared Components,
    VISA driver, and Microsoft Visual Studio or another IDE
    installed.

  2. Prepare a Main Scanner with a HV Module and a GB Module.

  3. Use the IVI-COM driver to import and control the device through
    a C# program.

  4. We recommend using an IVI-COM interop for easier program
    development.

Frequently Asked Questions (FAQ)

Q: Where can I find the IVI Driver for SC6540?

A: You can download the SC6540 IVI Driver from the
manufacturer’s website or refer to the help document located at
IVI FoundationIVIDriversSC6540SC6540.chm.

Q: What are the prerequisites for using the IVI Driver with

C#?

A: The prerequisites include having the SC6540 IVI Driver, IVI
Shared Components, VISA driver, and an IDE like Microsoft Visual
Studio installed. Additionally, you need a Main Scanner with a HV
Module and a GB Module.

Document Version 1.1
IVI Driver Getting Started Guide
For Multiplexer SC6540,
Overview
This application note will describe the installing instructions and several programming examples for IVI Instrument Driver of Scanner. To understand more about the IVI drivers, please refer to the website of IVI Foundation. For more detail of the SC6540 IVI driver, please check the help document, SC6540.chm, located at the path of IVI FoundationIVIDriversSC6540.
1. IVI Driver Setup
Instructions on downloading and Installing IVI Instrument Drivers from website. Download and install Shared Components from IVI Foundation Website.
2. Getting Started with C#
A tutorial using IVI driver establishes communication with the instrument by C# programming.
3. Getting Started with C++
A tutorial using IVI driver establishes communication with the instrument by C++ programming.
4. Getting Started with Python
A tutorial using IVI driver establishes communication with the instrument by Python programming.
5. Getting Started with LabVIEW
A tutorial using IVI driver establishes communication with the instrument by LabVIEW programming.
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

1. IVI Driver Setup
After downloading the IVI Driver, run the self-extracting setup file and you will see the installation wizard to start setup. Please follow the below instruction to complete the installation.
The setup will detect if IVI Shared Components are installed. If prompted with the following screen, click on Download, The IVI Foundation Website will be opened.
Please download the latest IVI Shared Components either 32-bit or 64-bit version. After downloading, install the shared components and continue the installation.
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

After the IVI Shared Components are installed, please follow the steps to complete installation.
There are options for installing the source code of the IVI Driver, if it is necessary.
The IVI driver would be installed under the path of “IVI FoundationIVI”. For the files with “*.dll” extension name would be located in the “Bin” folder. And the necessary help documents will be in the folder of “..DriversSC6540”.
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

2. Getting Started with C#
Introduction
This chapter describes the procedures of using the IVI-COM driver of IKONIX Group by C# programming language. In this exercise, the programmer could import the driver and complete a short program controlling the device step-by step.
The C# could call IVI-C driver, either. However, we suggest that an IVI-COM interop would be easier for you to develop the program.
Requirements
SC6540 IVI Driver IVI Shared Components, https://www.ivifoundation.org/shared_components/Default.aspx VISA (Virtual Instrument Software Architecture) driver,
https://www.ni.com/en/support/downloads/drivers/download.ni-visa.html Microsoft Visual Studio or other IDEs A Main Scanner with a HV Module and a GB Module
Download the Drivers
Please go to the website of the Associated Research to download the latest version of IVI drivers or contact the vendors. Follow the steps and instructions in Chapter 1 to complete the installation.
References
On the website of IVI Foundation, there are documentations you might be interested while implementing controlling the devices. You could find the resources of developing with an IVI driver, https://www.ivifoundation.org/resources/default.aspx. The IVI Shared Components could be download from https://www.ivifoundation.org/shared_components/Default.aspx. There are several documents on the website for understanding the IVI.
In the installed directory, there are several documents for your reference understanding the SC6540 IVI Driver. A help file, SC6540.chm, would be located at the path of IVI FoundationIVIDriversSC6540. In this help file, you could find all of the provided functions and their hierarchy.
There are four types of sample code for your reference which are located at the path of IVI FoundationIVIDriversSC6540Examples, including C#, C++, Python and LabVIEW as well.
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

Development
1 Create a C# project 1.1 Open Visual Studio IDE and create a new C# console project.
2 Import Libraries 2.1 Right-click on the reference and select Add Reference in the solution explorer 2.2 Click on the Browse button and go to the path of “IVI FoundationIVIBinPrimary Interop Assemblies” and choose AssociatedResearch.SC6540.Interop.dll and Ivi.Driver.Interop.dll.
2.3 Declare to use the name spaces for the interop assemblies that are specified to reference in the previous section.
using AssociatedResearch.SC6540.Interop;
3 Start programming 3.1 Create an object of the driver and use the initialize method to build up the connection.
//Initialize // var driver = new SC6540(); string resourceName = “ASRL4::INSTR”; string optionString = “Cache=false, InterchangeCheck=false, QueryInstrStatus=true, RangeCheck=false, RecordCoercions=false, Simulate=false”; driver.Initialize(resourceName, true, false, optionString); // Disable all channels driver.Execution.DisableAllChannels();
For more detail for the parameters of the Initialize() method, please refer to the help document, SC6540.chm, which is located at “IVI FoundationIVIDriversSC6540″. The first parameter resourceName is a string type and indicates the interfaces type and address of the connection. The resource name,”ASRL4::INSTR”, represents a serial port with
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

address 4. For example, a GPIB connection could be “GPIB0::8::INSTR”. For TCP/IP connection, it will be in the format of “TCPIP0::192.168.0.1::10001::SOCKET”. The 10001 is the TCP/IP connection port of SC6540. The address depends on the configurations. The optionString, “Cache=false, InterchangeCheck=false, QueryInstrStatus=true, RangeCheck=false, RecordCoercions=false, Simulate=false”, configures the options for the IVI driver with enabled QueryInstrStatus which will check the status at the end of each user operation. Please refer to the SC6540.chm for more details. After initialization, it is suggested to disable all of the channels by DisableAllChannels() method at the first step.
3.2 Configure Withstand Channels
// Withstand test (ACW, DCW) // Enable Withstand (High) channels int[] hvChannels = { 1, 2, 3 }; driver.Execution.ConfigureWithstandChannels(hvChannels);
// Enable Return (Low) channels int[] rtnChannel = { 4, 5, 6 }; driver.Execution.ConfigureReturnChannels(rtnChannel);
// After the multiplexer was configured, the safety tester could start output for withstand test on those connections. Thread.Sleep(1000);
// Disable all channels driver.Execution.DisableAllChannels();
For some units, we may need a multiplexer, SC6540, to extend the capacity on a HIPOT tester, such as Omnia2, HypotUltra, Hypot4 of Associated Research. Before a withstand test operated, the multiplexer needs to be configured to the test point. Both the ConfigureWithstandChannels() and ConfigureReturnChannels() method needs an integer array as input. After, the HV and Return channels are switched, we could add a procedure performing a withstand output. And disable the relays by DisableAllChannels().
3.3 Configure Continuity Channels
// Continuity (CONT) // Enable Continuity (High) channels int[] contChannels = { 4, 5 }; driver.Execution.ConfigureContinuityChannels(contChannels);
// Enable Return (Low) channels int[] contRtnChannel = { 7 }; driver.Execution.ConfigureReturnChannels(contRtnChannel);
// After the multiplexer was configured, the safety tester could start output for continuity test on those connections. Thread.Sleep(1000);
The ConfigureContinuityChannels() method takes an integer array as input. For the connection on the return circuits, we could invoke the method of ConfigureReturnChannels() which is shared with the withstand channels.
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

3.4 Configure Ground Bond Channel
//Ground Bond Test (GND) // Enable Ground Bond test channel driver.Execution.ConfigureGndChannel(2); // After the multiplexer was configured, the safety or ground bond tester could start output for ground bond test on those connections. Thread.Sleep(1000);
Only one ground bond channel could be enabled at one time. Therefore, the ConfigureGndChannel() takes a integer as input parameter at once. When a ground bond channel was set, the other switches would be disabled automatically.
3.5 Configure SC6540 for Dual Check
// DualCheck (Withstand test + Ground bond test) // Enable Grond Bond channel and Withstand Channel at the same time. driver.Execution.ConfigureGndAndWithstandChannel(5, “HLOOOOL”); // After the multiplexer was configured, the safety tester could start dual check on those connections. Thread.Sleep(1000);
The DualCheck is the function allows the instrument to simultaneously run a hipot and AC Ground Bond test. By ConfigureGndAndWithstandChannel(), we could configure both ground bond channel and withstand channels at the same time. This function takes one integer input as ground bond channel and one string as the raw configuration for HV module. The configurations will be H (High), L (Low) or O (Open). For example, the syntax for this command would be “HLOOOOL” which indicates the channel 1 is connected to HV and channel 2 and 7 would be connected to Return.
3.6 Close the session
driver.Execution.DisableAllChannels(); driver.Close(); Console.WriteLine(“Done – Press Enter to Exit”); Console.ReadLine();
Close() would close the I/O session to the instrument.
4 Completed example The completed sample code could be found at the path of “IVI FoundationIVIDriversSC6540Examples”.
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

3. Getting Started with C++
Introduction
This chapter describes the procedures of using the IVI-COM driver of IKONIX Group by C++ programming language. In this exercise, the programmer could import the driver and complete a short program controlling the device step-by step.
Requirements
SC6540 IVI Driver IVI Shared Components, https://www.ivifoundation.org/shared_components/Default.aspx VISA (Virtual Instrument Software Architecture) driver,
https://www.ni.com/en/support/downloads/drivers/download.ni-visa.html Microsoft Visual Studio or other IDEs A Main Scanner with a HV Module and a GB Module
Download the Drivers
Please go to the website of the Associated Research to download the latest version of IVI drivers or contact the vendors. Follow the steps and instructions in Chapter 1 to complete the installation.
References
On the website of IVI Foundation, there are documentations you might be interested while implementing controlling the devices. You could find the resources of developing with an IVI driver, https://www.ivifoundation.org/resources/default.aspx. The IVI Shared Components could be download from https://www.ivifoundation.org/shared_components/Default.aspx. There are several documents on the website for understanding the IVI.
In the installed directory, there are several documents for your reference understanding the SC6540 IVI Driver. A help file, SC6540.chm, would be located at the path of IVI FoundationIVIDriversSC6540. In this help file, you could find all of the provided functions and their hierarchy.
There are three types of sample code for your reference which are located at the path of IVI FoundationIVIDriversSC6540Examples, including C#, C++ and Python as well.
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

Development
1 Create a C++ project 1.1 Open Visual Studio or any other IDEs and create a new C++ console project.
2 Include Directories 2.1 Right-click on the project and select properties. 2.2 Expand the Configuration Properties and select VC++ Directories on the left menu. 2.3 Click on the drop-down column of the Include Directories and select <Edit..> to open the edit window.
2.4 Select the New Line button to add an include directories. There will be two necessary paths need to be added. IVI FoundationIVIBin $(VXIPNPPATH)VisaCom
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

2.5 Click OK to complete including the directories. 2.6 Use the #import operator to import the necessary DLLs

include “stdafx.h” #include #include #import

no_namespace #import no_namespace #import no_namespace 3 Start programming 3.1 Create an instance of the driver by pointer and use the initialize method to build up the connection. HRESULT hr = ::CoInitialize(NULL); ISC6540Ptr driver(__uuidof(SC6540)); _bstr_t resourceName = “ASRL4::INSTR”; _bstr_t optionString = “Cache=false, InterchangeCheck=false, QueryInstrStatus=true, RangeCheck=false, RecordCoercions=false, Simulate=false”; driver->Initialize(resourceName, true, false, optionString); // Disable all channels driver->Execution->DisableAllChannels(); For more detail for the parameters of the Initialize() method, please refer to the help document, SC6540.chm located at “IVI FoundationIVIDriversSC6540″. The first parameter resourceName is a string type and indicates the interfaces type and address of the connection. The resource name,”ASRL4::INSTR”, represents a serial port with address 4. For example, a GPIB connection could be “GPIB0::8::INSTR”. For TCP/IP connection, it will be in the format of “TCPIP0::192.168.0.1::10001::SOCKET”. The 10001 is the TCP/IP connection port of SC6540. To know more about the options of the Initialize() method, please refer to the SC6540.chm document. After initialization, it is suggested to disable all of the channels by DisableAllChannels() method at the first step. 4.1 Configure Withstand Channels // Withstand test (ACW, DCW) // Enable Withstand (High) channels int hvChannels[3] = { 1,2,3 }; SAFEARRAY* hvChannelsSafeArray = ConvertToSafeArray(hvChannels, sizeof(hvChannels)/sizeof(*hvChannels)); driver->Execution->ConfigureWithstandChannels(&hvChannelsSafeArray); SafeArrayDestroy(hvChannelsSafeArray); //// Enable Return (Low) channels int rtnChannels[3] = { 4,5,6 }; SAFEARRAY* rtnChannelsSafeArray = ConvertToSafeArray(rtnChannels, sizeof(rtnChannels) / sizeof(*rtnChannels)); driver->Execution->ConfigureReturnChannels(&rtnChannelsSafeArray); SafeArrayDestroy(rtnChannelsSafeArray); // After the multiplexer was configured, the safety tester could start output for withstand test on those connections-> Sleep(1000); // Disable all channels 28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-[847-367-4077](https://phone.gd/phone/847-367-4077) | Fax: 1-[847-367-4080](https://phone.gd/phone/847-367-4080) | www.arisafety.com

driver->Execution->DisableAllChannels();
For some units, we may need a scanner, SC6540, to extend the capacity on a HIPOT tester, such as Omnia2, HypotUltra, Hypot4 of Associated Research. Before a withstand test operated, the multiplexer needs to be configured to the test point. Both the ConfigureWithstandChannels() and ConfigureReturnChannels() method needs an integer array as input. After, the HV and Return channels are switched, we could add a procedure performing a withstand output. And disable the relays by DisableAllChannels(). There is a C++ function, ConvertToSafeArray(), in the above sample which is to convert the integer array to a SAFEARRAY type.
//ConvertToSafeArray is the function to create a SAFEARRAY type from a std::int[] // SAFEARRAY ConvertToSafeArray(int channels[], int size) {
CComSafeArray csaData(size); for (int i = 0; i < size; i++) {
csaData.SetAt(i, channels[i]); } return csaData.Detach(); }
For the COM components, it take the SAFEARRAY type as the array parameters. Therefore, we need to create it with assigned size. Also, it is suggested to release the SAFEARRAY by SafeArrayDestroy() after it is no longer been used.
4.2 Configure Continuity Channels
// Continuity (CONT) // Enable Continuity (High) channels int contChannels[2] = { 4, 5 }; SAFEARRAY
contChannelsSafeArray = ConvertToSafeArray(contChannels, sizeof(contChannels) / sizeof(contChannels)); driver->Execution->ConfigureContinuityChannels(&contChannelsSafeArray); SafeArrayDestroy(contChannelsSafeArray);
// Enable Return (Low) channels int contRtnChannel[1] = { 7 }; SAFEARRAY
contRtnChannelsSafeArray = ConvertToSafeArray(contRtnChannel, sizeof(contRtnChannel) / sizeof(*contRtnChannel)); driver->Execution->ConfigureReturnChannels(&contRtnChannelsSafeArray); SafeArrayDestroy(contRtnChannelsSafeArray);
// After the multiplexer was configured, the safety tester could start output for continuity test on those connections Sleep(1000);
The ConfigureContinuityChannels() method takes an SAFEARRAY as input which is converted by ConvertToSafeArray(). For the connection on the return circuits, we could invoke the method of ConfigureReturnChannels() which is the same with the withstand channels.
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

4.3 Configure Ground Bond Channel
//Ground Bond Test (GND) // Enable Ground Bond test channel driver->Execution->ConfigureGndChannel(2); // After the multiplexer was configured, the safety or ground bond tester could start output for ground bond test on those connections Sleep(1000);
Only one ground bond channel could be enabled at one time. Therefore, the ConfigureGndChannel() takes a integer as input parameter at once. When a ground bond channel was set, the other switches would be disabled automatically.
4.4 Configure SC6540 for Dual Check
// DualCheck (Withstand test + Ground bond test) // Enable Grond Bond channel and Withstand Channel at the same time driver->Execution->ConfigureGndAndWithstandChannel(5, “HHHLLLOO”); // After the multiplexer was configured, the safety tester could start dual check on those connections Sleep(1000);
The DualCheck is the function allows the instrument to simultaneously run a hipot and AC Ground Bond test. By ConfigureGndAndWithstandChannel(), we could configure both ground bond channel and withstand channels at the same time. This function takes one integer input as ground bond channel and one string as the raw configuration for HV module. The configurations will be H (High), L (Low) or O (Open). For example, the syntax for this command would be “HLOOOOL” which indicates the channel 1 is connected to HV and channel 2 and 7 would be connected to Return.
3.2 Close the session
driver->Execution->DisableAllChannels(); driver->Close(); std::cout << “Done – Press Enter to Exit” <<std::endl; std::cin.get();
Close() would close the I/O session to the instrument.
4 Completed example The completed sample code could be found at the path of “IVI FoundationIVIDriversSC6540Examples”.
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

4. Getting Started with Python
Introduction
This chapter describes the procedures of using the IVI-COM driver of Ikonix Group by Python programming language. In this exercise, the programmer could import the driver and complete a short program controlling the device step-by step.
Requirements
SC6540 IVI Driver IVI Shared Components, https://www.ivifoundation.org/shared_components/Default.aspx VISA (Virtual Instrument Software Architecture) driver,
https://www.ni.com/en/support/downloads/drivers/download.ni-visa.html Python IDE Comtypes Library ( pip install comtypes) A Main Scanner with a HV Module and a GB Module
Download the Drivers
Please go to the website of the Associated Research to download the latest version of IVI drivers or contact the vendors. Follow the steps and instructions in Chapter 1 to complete the installation.
References
On the website of IVI Foundation, there are documentations you might be interested in while implementing controlling the devices. You could find the resources of developing with an IVI driver, https://www.ivifoundation.org/resources/default.aspx. The IVI Shared Components could be download from https://www.ivifoundation.org/shared_components/Default.aspx. There are several documents on the website for understanding the IVI.
In the installed directory, there are several documents for your reference understanding the SC6540 IVI Driver. A help file, SC6540.chm, would be located at the path of IVI FoundationIVIDriversSC6540. In this help file, you could find all of the provided functions and their hierarchy.
There are three types of sample code for your reference which are located at the path of IVI FoundationIVIDriversSC6540Examples, including C#, C++ and Python as well.
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

Development
1 Install the Comtypes library pip install cometypes
In order to call an external com DLL in Python, you will need comtypes library installed.
2 Create a Python file 2.1 Open any IDE of Python and create a new Python file.
3 Import Libraries 3.1 Import the cometypes library and SC6540_64.dll
import time import comtypes import comtypes.client as cc # Import the IVI-COM DLL cc.GetModule(‘SC6540.dll’) from comtypes.gen import SC6540Lib
4 Start programming 4.1 Create an object of the driver and use the initialize method to build up the connection.
driver = cc.CreateObject(‘SC6540.SC6540’, interface=SC6540Lib.ISC6540) # Initialize Driver and make connection optionString = ‘Cache=false, InterchangeCheck=false, QueryInstrStatus=true, RangeCheck=false, RecordCoercions=false, Simulate=false’ driver.Initialize(‘ASRL4::INSTR’, True, False, optionString) # Disable all channels driver.Execution.DisableAllChannels();
For more detail for the parameters of the Initialize() method, please refer to the help document, SC6540.chm located at “IVI FoundationIVIDriversSC6540″. The first parameter ResourceName is a string type and indicates the interfaces type and address of the connection. The resource name,”ASRL4::INSTR”, represents a serial port with address 4. For example, a GPIB connection could be “GPIB0::8::INSTR”. For TCP/IP connection, it will be in the format of “TCPIP0::192.168.0.1::10001::SOCKET”. The 10001 is the TCP/IP connection port of SC6540. There are other parameters for the options of the Initialize() method, please refer to the SC6540.chm for more detail. For example, “QueryInstrStatus=true” makes the session automatically query the error status for each command was sent.
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

4.2 Configure Withstand Channels

Withstand test (ACW, DCW) # Enable Withstand (High) channels hvChannels = {

1, 2, 3 }; driver.Execution.ConfigureWithstandChannels(hvChannels);

Enable Return (Low) channels rtnChannel = { 4, 5, 6 };

driver.Execution.ConfigureReturnChannels(rtnChannel);

After the multiplexer was configured, the safety tester could start output

for withstand test on those connections. time.sleep(1)

Disable all channels driver.Execution.DisableAllChannels();

For some units, we may need a multiplexer, SC6540, to extend the capacity on a HIPOT tester, such as Omnia2, HypotUltra, Hypot4 of Associated Research. Before a withstand test operated, the multiplexer needs to be configured to the test point. Both the ConfigureWithstandChannels() and ConfigureReturnChannels() method needs an integer array as input. After, the HV and Return channels are switched, we could add a procedure performing a withstand output. And disable the relays by DisableAllChannels().
4.3 Configure Continuity Channels

Continuity (CONT) # Enable Continuity (High) channels contChannels = { 4, 5

}; driver.Execution.ConfigureContinuityChannels(contChannels);

Enable Return (Low) channels contRtnChannel = { 7 };

driver.Execution.ConfigureReturnChannels(contRtnChannel);

After the multiplexer was configured, the safety tester could start output

for continuity test on those connections. time.sleep(1)
The ConfigureContinuityChannels() method takes an integer array as input. For the connection on the return circuits, we could invoke the method of ConfigureReturnChannels() which is shared with the withstand channels.
4.4 Configure Ground Bond Channel

Ground Bond Test (GND) # Enable Ground Bond test channel

driver.Execution.ConfigureGndChannel(2);

After the multiplexer was configured, the safety or ground bond tester could

start output for ground bond test on those connections. time.sleep(1)
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

Only one ground bond channel could be enabled at one time. Therefore, the ConfigureGndChannel() takes a integer as input parameter at once. When a ground bond channel was set, the other switches would be disabled automatically.
4.5 Configure SC6540 for Dual Check

DualCheck (Withstand test + Ground bond test) # Enable Grond Bond channel

and Withstand Channel at the same time. driver.Execution.ConfigureGndAndWithstandChannel(5, “HHHLLLOO”); # After the multiplexer was configured, the safety tester could start dual check on those connections. time.sleep(1)
The DualCheck is the function allows the instrument to simultaneously run a hipot and AC Ground Bond test. By ConfigureGndAndWithstandChannel(), we could configure both ground bond channel and withstand channels at the same time. This function takes one integer input as ground bond channel and one string as the raw configuration for HV module. The configurations will be H (High), L (Low) or O (Open). For example, the syntax for this command would be “HLOOOOL” which indicates the channel 1 is connected to HV and channel 2 and 7 would be connected to Return.
4.6 Close the session
driver.Execution.DisableAllChannels(); driver.Close(); print(“Done”)
Close() would close the I/O session to the instrument.
5 Completed example The completed sample code could be find at the path of “IVI FoundationIVIDriversSC6540Examples”.
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

5. Getting Started with LabVIEW
Introduction
This chapter describes the procedures of using the IVI-COM driver of Ikonix Group by LabVIEW programming language. In this exercise, the programmer could learn how to import the driver and complete a short program controlling the device step-by step.
Even though the programmers could control the device by IVI Driver. For the LabVIEW programmer, we suggest that using LabVIEW plug & play driver would be easier for your programming and debugging. The LabVIEW driver from Ikonix Group are all made up with commands directly, so you could clearly check how the commands were sent to instruments.
Requirements
SC6540 IVI Driver IVI Shared Components, https://www.ivifoundation.org/shared_components/Default.aspx VISA (Virtual Instrument Software Architecture) driver,
https://www.ni.com/en/support/downloads/drivers/download.ni-visa.html National Instruments LabVIEW (This example was written in LabVIEW 2014) A Main Scanner with a HV Module and a GB Module
Download the Drivers
Please go to the website of the Associated Research to download the latest version of IVI drivers or contact the vendors. Follow the steps and instructions in Chapter 1 to complete the installation.
References
On the website of IVI Foundation, there are documentations you might be interested while implementing controlling the devices. You could find the resources of developing with an IVI driver, https://www.ivifoundation.org/resources/default.aspx. The IVI Shared Components could be download from https://www.ivifoundation.org/shared_components/Default.aspx. There are several documents on the website for understanding the IVI.
In the installed directory, there are several documents for your reference understanding the SC6540 IVI Driver. A help file, SC6540.chm, would be located at the path of IVI FoundationIVIDriversSC6540. In this help file, you could find all of the provided functions and their hierarchy.
There are three types of sample code for your reference which are located at the path of IVI FoundationIVIDriversSC6540Examples, including C#, C++ and Python as well.
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

1 Open a new vi. 2 Import the DLL component.

Development

Open the Function Palette by right-clicking on the block diagram. Then select Connectivity -> ActiveX. Select or drop the Automation Open function on the block diagram. 3 Right-clicking on the Automation Open and select Select ActiveX Class -> Browse will open a window for choosing the DLL. 4 Select the Browse button and select the file SC6540_64.dll located at IVI FoundationIVIBin. The IVI SC6540 Type Library would be added into the Type Libraries drop down menu. 5 Select ISC6540 and then click OK to complete creating an object of SC6540 driver instance.
The Labview will automatically generate an Automation refnum of SC6540Lib.ISC6540 control and connect to the Automation Open function.
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

6 Start programming 6.1 Create an Invoke Node function and connect the reference to the output of Automation Refnum and then click on the Method and select Initialize to initialize the connection with device.
For more detail for the parameters of the Initialize() method, please refer to the help document, SC6540.chm located at “IVI FoundationIVIDriversSC6540″. The first parameter ResourceName is a string type and indicates the interfaces type and address of the connection. The resource name,”ASRL4::INSTR”, represents a serial port with address 4. For example, a GPIB connection could be “GPIB0::8::INSTR”. For TCP/IP connection, it will be in the format of “TCPIP0::192.168.0.1::10001::SOCKET”. The 10001 is the TCP/IP connection port of SC6540. The OptionString, “Cache=false, InterchangeCheck=false, QueryInstrStatus=true, RangeCheck=false, RecordCoercions=false, Simulate=false”, configures the options for the IVI driver with enabled QueryInstrStatus which will check the status at the end of each user operation. Please refer to the SC6540.chm for more details.
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

6.2 Configure Withstand Channels
For some units, we may need a multiplexer, SC6540, to extend the capacity on a HIPOT tester, such as Omnia2, HypotUltra, Hypot4 of Associated Research. Before a withstand test operated, the multiplexer needs to be configured to the test point. Both the ConfigureWithstandChannels() and ConfigureReturnChannels() method needs an integer array as input. After, the HV and Return channels are switched, we could add a procedure performing a withstand output. And disable the relays by DisableAllChannels(). 6.3 Configure Continuity Channels
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

The ConfigureContinuityChannels() method takes an integer array as input. For the connection on the return circuits, we could invoke the method of ConfigureReturnChannels() which is shared with the withstand channels. 6.4 Configure Ground Bond Channel
Only one ground bond channel could be enabled at one time. Therefore, the ConfigureGndChannel() takes a integer as input parameter at once. When a ground bond channel was set, the other switches would be disabled automatically. 6.5 Configure SC6540 for Dual Check
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

The DualCheck is the function allows the instrument to simultaneously run a hipot and AC Ground Bond test. By ConfigureGndAndWithstandChannel(), we could configure both ground bond channel and withstand channels at the same time. This function takes one integer input as ground bond channel and one string as the raw configuration for HV module. The configurations will be H (High), L (Low) or O (Open). For example, the syntax for this command would be “HLOOOOL” which indicates the channel 1 is connected to HV and channel 2 and 7 would be connected to Return. 6.6 Close the session
Close() method in ISC6540 class would close the I/O session to the instrument. Also, all of the references should be closed using the Close Reference function. 7 Completed example
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

The completed example with the state machine design pattern for your reference which are located at the path of IVI FoundationIVIDriversSC6540Examples, including C#, C++ and Python as well. However, we suggest that using LabVIEW plug & play driver would be easier for LabVIEW developers. If you need a LabVIEW driver, please download it from the website of Associated Research or contact the vendor.
28105 N. Keith Drive Lake Forest, IL 60045 USA Toll Free: 1-800-858-TEST (8378) US/Canada Phone: 1-847-367-4077 | Fax: 1-847-367-4080 | www.arisafety.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals