Multicarein Multiparameter Diagnostic Device
“`html
Product Information
Technical Specifications
-
Measuring range:
- Glucose electro-chemical: 10-600 mg/dL – 0.6-33.3 mmol/L
- Cholesterol
- Triglycerides
-
Measuring Blood sample time: 5 seconds
-
Memory: 500 test results with date and time
-
Display: LCD
-
Batteries: 2 lithium batteries Cr 2032 3 Volts (32212) –
suitable for 1 year or 1,000 measurements -
Size – weight: 97x49x20.5 mm – 65 g (battery included)
Product Usage Instructions
Device Overview
The Multiparameter Diagnostic Device is designed for the
self-monitoring of cholesterol, triglycerides, and glucose levels.
It provides quick and accurate results using Amperometric
Technology for glucose and Reflectometric Technology for
cholesterol and triglycerides.
Testing Procedure
-
Prepare the device by inserting the required test strip
(glucose, cholesterol, or triglycerides). -
Collect a blood sample using the safety lancet provided.
-
Apply the blood sample to the corresponding test strip as
instructed in the user manual. -
Wait for the specified time (5 seconds for glucose) for the
results to display on the LCD screen.
Data Management
The device can store up to 500 test results with dates and
times. Use the provided manual to learn how to review and manage
the stored data.
Frequently Asked Questions (FAQ)
Q: How often should I perform self-monitoring tests with the
device?
A: The frequency of testing should be determined by your
healthcare provider based on your individual health needs. Follow
the recommended testing schedule provided by your healthcare
professional.
Q: Are the test strips and lancets reusable?
A: No, the test strips and lancets are designed for single-use
only. Discard them properly after each test to ensure accurate
results and reduce the risk of contamination.
Q: Can I switch between different measurement units (mg/dL –
mmol/L) on the device?
A: Yes, the device allows you to switch between mg/dL and mmol/L
measurement units. Refer to the user manual for instructions on
changing the unit settings.
“`
DIAGNOSTIC TESTS – LABORATORY
MULTICAREIN MULTIPARAMETER DIAGNOSTIC DEVICE
· 24147 MULTICAREIN – Italian
· 24148 MULTICAREIN – English
*Supplied with nylon bag, 5 cholesterol strips, 5 safety lancets, rapid guide (GB, FR, IT, ES, PT, GR, PL, HU) and IT or GB manual.
· 24149 MULTICAREIN STARTER KIT – GB, IT
· 24153 MULTICAREIN STARTER KIT BLUETOOTH
**Supplied in cardboard box with a nylon bag, 10 glucose, 5 cholesterol, 5
triglycerides strips, 25 safety lancets, user manual (GB, IT), rapid guide and
QR code with complete manual in 8 languages (GB, IT, FR, ES, PT, GR, PL, HU).
Multiparameter diagnostic device for the periodical self-monitoring of risk
factors: cholesterol, triglycerides and glucose. Portable device for easy test
procedure and quick results. Accurate with Amperometric Technology for glucose
and Reflectometric Technology for cholesterol and tryglicerides. Switchable
mg/dL – mmol/L Multilingual box: GB, FR, IT, ES, GR, PT, PL, HU, RU and
selectable internal software: GB, IT.
· 23982 SOFTWARE+CABLE FOR MULTICAREIN – GB, IT, – optional Operating system:
Windows XP, 7, 8, 10, Vista, 16 Mb Ram.
The first and only multi-parameter device with double technology
AMPEROMETRIC SUCTION STRIP – glucose
Amperometric
– Electro-chemical technology for a practical use
– Requires very little blood samples (0.5 µl)
– Short time for the results (5 sec.)
REFLECTOMETRIC SUCTION STRIP – cholesterol and tryglicerides
– It allows a more functional self-monitoring of triglycerides and
cholesterol.
24149
SELF CONTROL
0344
ISO 15197:2015
3 IN 1: CHOLESTEROL, TRIGLYCERIDES,
GLUCOSE
24147, 24148
Triglycerides
Cholesterol
Tube Tube Tube
STRIPS FOR MULTICAREIN***
of 5 of 25 of 50
9 LANGUAGE BOX – spare
– 24156 24157 Glucose strips biosensor + 1 chip
24158 24159 – Cholesterol strips + 1 chip
24160 24161 – Triglycerides strips + 1 chip
Strips are sold in a 9 languages box: GB, IT, FR, ES, PT, PL, GR, HU, Arabic.
GIMA code
CONTROL SOLUTION FOR MULTICAREIN***
23978 Glucose control solution
23979 Cholesterol control solution
23980 Triglycerides control solution
LANCETS FOR MULTICAREIN***
23917 Lancets 28G – sterile – bag of 25 pcs
23916 Lancets 28G – sterile – box of 100
***All spares are compatible with old Multicare versions
Cholesterol
Glucose
Triglycerides
TECHNICAL SPECIFICATIONS
Measuring range
Measuring Blood
time sample
Glucose electro-chemical: 10-600 mg/dL – 0.6-33.3 mmol/L 5″
0.5 µl
Cholesterol
130-400 mg/dL – 3-10.2 mmol/L 30″-60″ 10 µl
Triglycerides
50-500 mg dL – 0.56 – 5.6 mmol/L 30″-60″ 10 µl
Memory:
500 test results with date and time
Display:
LCD
Batteries:
2 lithium batteries Cr 2032 3 Volts (32212)
suitable for 1 year or 1,000 measurements
Size – weight:
97x49x20.5 mm – 65 g (battery included)
MISSION® CHOLESTEROL METER + LIPID PANEL TEST
· 23932 MISSION® CHOLESTEROL METER
Provides high quality and reliable results for complete Lipid
Panel testing in 2 min.
3 in 1 complete Lipid Panel device
tests for Total Cholesterol (CHOL),
High Density Lipoprotein (HDL)
and Triglycerides (TRIG)
Also includes calculated tests:
Low Density Lipoprotein (LDL),
CHOL/HDL ratio and Cardiac Risk
Assessment
Reliable
– can test whole blood,
0123
serum or plasma
– wide hematocrit range of 0-50%
– small printer available for
immediate printing of results
Easy to use
– large LCD display for easy reading
– colour coded test devices with
engraved test name for easy
identification
Supplied with 2 control devices,
4 batteries, carrying case and manual (GB, IT).
23932
Available FR,ES, DE, PT manual in www.gimaitaly.com
Multilingual box: GB, FR, IT, ES, DE, PT, PL, SE, FI, GR, NL, RO,
BG, RU, SA.
5
TECHNICAL SPECIFICATIONS
Detection principle
Reflectance Photometry
Time to results
2 minutes
Memory with date/time 200 results
Type of test
Quantitative
Tests
Calculated tests Specimen volume
CHOL, HDL, TRIG
LDL, CHOL/HDL ratio, Cardiac Risk Assessment 3-1 Lipid Panel Device: 35 µL
Measurement units Measurement range
CHOL Individual Test Device: 10 µL mg/dL or mmol/L
CHOL: 100-500 mg/dL (2.59-12.93 mmol/L)
HDL: 15-100 mg/dL (0.39-2.59 mmol/L)
TRIG: 45-650 mg/dL (0.51-7.34 mmol/L)
Pc interface for data transfer Mini USB Port
Calibration
Code Chip
Languages on the screen English abbreviations
Power source
4 AAA or AC Adaptor (Mini USB, 5V dc, 50 mA)
Battery life
1,000 tests
Automatic shut off
5 minutes
Meter dimensions – weight 137×79×26 mm (5.4″ x 3.1″ x 1.0″) – 145 g (5.11 oz)
Display dimensions
50×50 mm (2.0″ x 2.0″)
GIMA code
ACCESSORIES AND SPARES
Minimum order
23935 3 in 1 Lipid panel tests + capil. tubes 35 ul + lancets* box of 5
23933 3 in 1 Lipid panel tests + capillary tubes 35 ul box of 25
23934 Total Cholesterol tests + capillary tubes 10 ul box of 25
24024 Capillary transfer tubes 10 ul – plastic
box of 50
23917 Gima lancets 28G
box of 25
23916 Gima lancets 28G
box of 100
23919 Auto lancet device – AST method
–
23929 Mission printer**
–
33250 Thermal paper roll 57 mm x 25 m
box of 5
*Safety lancets
**Includes 1 paper roll, cable, adaptor
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>