TFA Dostmann 30.3046.01 Funk Thermometer Cool Home Instruction Manual

August 16, 2024
TFA Dostmann

Instruction manual Kat. Nr. 30.3046.01

30.3046.01 Funk Thermometer Cool Home

www.tfa-dostmann.de/en/service/downloads/instruction-manuals

Wireless thermometer
Thank you for choosing this product from TFA.

Before you use this device

  •   Please make sure you read the instruction manual carefully.
  • The operating instructions are enclosed with the device or can be downloaded at www.tfa-dostmann.de/en/service/downloads/instruction-manuals
  • This product should only be used as described within these instructions.
  • Following and respecting the instructions in your manual will prevent damage to your instrument and loss of your statutory rights arising from defects due to incorrect use. We shall not be liable for any damage occurring as a result of non-following of these instructions.
  • Please pay particular attention to the safety notices!”
  • Please keep this instruction manual safe for future reference.

Delivery contents

  • Wireless thermometer (base station)
  • Transmitter (cat.-no. 30.3250.02)
  • Instruction manual

Range of application and all the benefits of your new instrument at a

glance

  • Simultaneous control of indoor and outdoor temperature
  • Ventilation recommendation if opening or closing the windows lowers the room temperature
  • Outdoor temperature wireless, distance range up to 100 m
  • Indication of maximum and minimum values showing time and date of recording
  • Two-level trend display
  • Adjustable ventilation alarm with two volume levels (Lo / Hi / OFF)
  • Radio-controlled clock with date
  • With fold-out stand, wall mounting hanger and two magnets

For your safety

  • This product is exclusively intended for the range of application described above.
  • This product should only be used as described within these instructions

WARNING

  • Keep the devices and the batteries out of reach of children.

  • Small parts can be swallowed by children (under three years old).

  • Batteries contain harmful acids and may be hazardous if swallowed.
    If a battery is swallowed, this can lead to serious internal burns and death within two hours. If you suspect a battery could have been swallowed or otherwise caught in the body, seek medical help immediately.

  • Batteries must not be thrown into a fire, short-circuited, taken apart or recharged. Risk of explosion!

  • Low batteries should be changed as soon as possible to prevent damage caused by leaking. Make sure the polarities are correct. Never use a combination of old and new batteries together, nor batteries of different types. Remove the batteries if the device will not be used for an extended period of time. Avoid contact with skin, eyes and mucous membranes when handling leaking batteries. In case of contact, immediately rinse the affected areas with water and consult a doctor.

CAUTION

  • Unauthorised repairs, alterations or changes to the product are prohibited.
  • Do not expose the device to extreme temperatures, vibrations or shocks.
  • Clean the device with a soft damp cloth. Do not use solvents or scouring agents.
  • The base station is suitable for indoor use only. Protect it from moisture!
  • The transmitter is splashproof, but not watertight.

Elements

Base station (Receiver)
A: Display (Fig. 1):

A 1: Trend arrow indoor temperature
A 2: Indoor temperature
A 3: Battery symbol base station
A 4: MAX/MIN values
A 5: Ventilation recommendation window OPEN (AUF)/CLOSED (ZU)
A 6: Reception symbol transmitter
A 7: Outdoor temperature| A 8: Trend arrow for outdoor temperature
A 9: Battery symbol transmitter
A 10: Date
A 11: Alarm symbol
A 12: Time zone
A 13: Time
A 14: DCF reception symbol
---|---

B: Buttons (Fig. 2):

B 1: SET button
B 2: –/MIN button| B 3: MAX/+ button
B 4: ALARM button
---|---

C: Housing (Fig. 2-3):

C 1: Stand (fold out)
C 2: Wall mounting hanger| C 3: Battery compartment
C 4: 2 magnets
---|---

Transmitter
D: Display (Fig. 4):

D 1: Channel 1/2/3
D 2: Transmission signal| D 3: Battery symbol for the transmitter
D 4: Temperature
---|---

E: Housing (Fig. 4):
E 1: Battery compartment
E 2: Support for wall mounting or table standing
E 3: Indentations for the holder for wall mounting

F: Buttons in the battery compartment (Fig. 4):
F 1: TX button
F 2: 1 2 3 switch for channel selection

Getting started

6.1 Insert the batteries

  • Place the instruments on a table at a distance of about 1.5 meters from each other. Avoid being close to possible sources of interference such as electronic devices and radio equipment.
  • Open the battery compartment of the transmitter and insert two new AA 1.5 V batteries, polarity as illustrated. All LCD segments will be displayed for a short moment. The indoor temperature appears on the display. The switch is set to channel 1.
  • Close the battery compartment.
  • Open the battery compartment of the base station.
  • Insert two new AAA 1.5 V batteries, polarity as illu-strated. The device will alert you with a beep and all LCD segments will be displayed for a short moment.
  • Close the battery compartment.

6.2 Reception of the transmitter

  • The base station will scan the outdoor temperature and the reception symbol flashes.

  • If the reception is successful, the outdoor temperature and the reception symbol appear steadily on the base station display.

  • If the reception of the temperature fails, “- -. -” remains on the display. Check the batteries and try it again.

  • You can also activate the reception manually. Hold the MAX/+ button on the base station for 5 seconds.
    The reception symbol for the transmitter will be flashing.

  • Press the TX button in the transmitter’s battery compartment. The transmission of the outdoor data takes place immediately and if the reception is successful, the outdoor temperature appears on the base station display.

  • If the transmitter signal is lost, the device automatically starts a transmitter search after 30 minutes.

6.3 Reception of the DCF frequency signal

  • After the reception of the transmitter, the clock will now scan the DCF frequency signal and the DCF reception symbol flashes on the display.
  • When the time code is successfully received after 310 minutes, the radio-controlled time, the date and the DCF symbol will be shown steadily on the display.
  • You can also activate the DCF reception manually. Press and hold the –/MIN button for 5 seconds. The DCF reception symbol flashes.
  • The DCF reception always takes place at 2:00 and 3:00 o’clock in the morning. If the signal is not successfully received, further attempts will be taken at 4:00 and 5:00 o’clock.
  • If the clock cannot detect the DCF-signal (e.g. due to interference, transmitting distance, etc.), the time can be set manually.
  • The clock will then work as a normal quartz clock (see: “Manual settings”).

Note on radio-controlled time

  • The time is transmitted from an atomic clock near Frankfurt am Main by a DCF-77 (77.5 kHz) frequency signal with a range of about 1,500 km. Your radiocontrolled clock receives the signal, converts it and always shows the exact time. The adjustment of Daylight Saving Time and Standard Time is also automatic.
  • Avoid being close to possible sources of interference that may affect radio reception (see “Location and mounting”).
  • During night-time, the atmospheric interference is usually less severe. A single daily reception is adequate to keep the accuracy deviation under 1 second.

Operation

7.1 Manual settings

  • Press and hold the SET button for three seconds to enter the setting mode.

  • The first adjustable value flashes on the display.

  • Use the MAX/+ or –/MIN button to make the desired setting.

  • Confirm with the SET button and go to the next setting.

  • The sequence is shown as follows:
    – Ventilation alarm volume (On Lo / On Hi / OFF) (default: On Lo)
    – DCF reception On/OFF (default: dCF On)
    – Time zone -12/+12 (default: TZ 00)
    – Hour, minutes
    – Year, month, date

  • Press and hold the MAX/+ or –/MIN button in the respective setting mode for quick setting. The device will automatically quit the setting mode if no button is pressed for 20 seconds.

7.1.1 Volume setting for the ventilation alarm

  • In the setting mode, you can set the volume of the ventilation alarm in 2 levels (Lo / Hi) or switch it off (OFF) (see “Ventilation recommendation and alarm”).

7.1.2 DCF reception

  • By default, the DCF reception is activated (dCF On) and after successful reception of the DCF signal no manual time setting is necessary.
  • Once the DCF time reception is deactivated (dCF OFF), the clock and date must be manually set. No time zone will appear.
  • If the DCF reception is activated, the manually set time will be overwritten by the DCF time when the signal is received successfully.

7.1.3 Time zone setting

  • In the setting mode you can make the time zone correction
  • The time zone correction is needed for countries where the DCF signal can be received but the time zone is different from the DCF time (e.g. +1=one hour plus).

Ventilation recommendation and alarm

  • The device checks the room temperature and the outside temperature simultaneously, and recommends to open/close windows to lower the room temperature.
    – Outdoor temperature is lower than indoor temperature: Window OPEN symbol
    – Outdoor temperature is higher than indoor temperature: Window CLOSED symbol

  • In addition, an alarm is triggered as soon as the display (window OPEN/CLOSED) changes and the current indoor and outdoor temperatures overlap. The display flashes and an alarm tone sounds if desired. This alarm function is configured in the setting mode (see “Volume setting for the ventilation alarm”).

8.1 Activate and deactivate the ventilation alarm

  • To ensure that the symbol (window OPEN/CLOSED) appears on the display and an alarm is triggered as soon as the display changes, the alarm must be active.

  • Press and hold the ALARM button for 3 seconds to activate the alarm function (default OFF).

  • The alarm symbol appears on the display and “On” is flashing, if the alarm is activated. Wait some seconds until the display returns to normal mode to confirm the setting. The alarm symbol remains on the display.

  • There are three different alarm symbols depending on the alarm volume setting:
    – Symbol with one sound wave: Alarm volume low (in setting mode On Lo).
    – Symbol with two sound waves: Alarm volume high (in setting mode On Hi).
    – Symbol without sound wave: no alarm sound (in setting mode OFF).

  • If you do not wish to be notified about the ventilation recommendations (e.g. outside the warm season), you can deactivate this function completely. Press and hold the ALARM button for 3 seconds. “OFF” is flashing.
    Wait some seconds until the display returns to normal mode to confirm the setting. The alarm symbol and the ventilation recommendation disappear.

8.2 Alarm event

  • After the ventilation alarm has been activated, an alarm is triggered when the current indoor and outdoor temperatures overlap for the first time. This is to draw attention to the change in ventilation needs.
  • The window display changes (OPEN → CLOSE or CLOSE → OPEN), the alarm symbol flashes and an alarm signal sounds three times in quick succession (provided that in setting mode the volume is activated). After that, the alarm signal sounds again  every minute
  • Follow the ventilation recommendation and open or close the windows. Press briefly the ALARM button to turn off the current alarm. The alarm function remains active and is triggered next time the ventilation recommendation changes.

Maximum/Minimum function

  • Press the –/MIN button in normal mode to show the minimum (MIN) values of indoor and outdoor temperature since the last reset with time and date of recording.
    Press the –/MIN button again to return to normal mode.

  • Repeat the process by pressing the MAX/+ button to see the maximum values.

  • The device will automatically quit the MAX/MIN mode if no button is pressed for 20 seconds.

  • Press and hold the SET button for 3 seconds while the highest or lowest values are displayed. The values will be deleted individually and reset to the current state.

Trend arrows

| Temperature increases by 0.2°C or more in the last 5 minutes
---|---
| Temperature increases by 1°C or more in the last 15 minutes
| Temperature decreases by 0.2°C or more in the last 5 minutes
| Temperature decreases by 1°C or more in the last 15 minutes

No arrow
Change within the last 5 minutes less than 0.2°C

Location and mounting

  • When placed outdoors, choose a shady and dry place for the transmitter. Direct sunlight may trigger incorrect measurement and continuous humidity damages the electronic components needlessly.
  • Place the base station in any room of the house.
  • Avoid being close to possible sources of interference that may affect radio reception. It is recommended to keep a distance of at least 2 metres to possibly interfering sources (TV, computer, microwave, wireless phones, baby monitors) and large metal objects (e.g. refrigerator). Within solid walls, especially ones with metal parts, the transmission range can be reduced considerably. In extreme cases, please place the unit close to a window to improve reception.
  • Check whether the transmission from the transmitter to the base station is possible and whether the DCF radio clock signal can be received.
  • If necessary, choose, another position for the outdoor transmitter and/or receiver.
  • If the transmission is successful, you can place or wall mount the base station and the outdoor transmitter with the suspension device.
  • To wall-mount the transmitter, screw the holder to the wall and click the outdoor transmitter into the holder.
  • Do not use the base station in the vicinity of radiators, other sources of heat or in direct sunlight.

Battery replacement

  • As soon as the corresponding battery symbol appears in the display, please change the batteries of the base station or transmitter
  • Please note: When the batteries are changed, the contact between the transmitters and the base station must be restored – so always restart both devices or start a manual transmitter search.

Troubleshooting

Problem Solution

No display
on the base station or the transmitter
➜ Ensure the batteries’ polarities are correct
➜ Change the batteries

No transmitter reception
Display “- -.-”
➜ No transmitter installed
➜ Start manual search for the outdoor transmitter
➜ Check the transmitter’s batteries (only use batteries/rechargeable batteries with 1.5V voltage!)
➜ Restart the transmitter and the base station according to the manual
➜ Choose another place for the transmitter and/or the base station
➜ Reduce the distance between the transmitter and the base station
➜ Check if there is any source of interference

No DCF reception
➜ Activate (On) the DCF reception in setting mode
➜ Wait for an attempt reception during the night
➜ Change the location (interference!)
➜ Restart the device according to the manual
➜ Set the clock manually

No ventilation tips
➜ Activate alarm

No alarm sound
➜ Set volume in setting mode

Incorrect indication
➜ Change the batteries

If your device fails to work despite these measures, please contact the retailer where you purchased the product from for advice.

Waste disposal

This product and its packaging have been manufactured using high-grade materials and components which can be recycled and reused. This reduces waste and protects the environment. Dispose of the packaging in an environmentally friendly manner using the
collection systems that have been set up.

Disposal of the electrical device
Remove non-permanently installed batteries and rechargeable batteries from the device and dispose of them separately.
This product is labelled in accordance with the EU Waste Electrical and Electronic Equipment Directive (WEEE).
This product must not be disposed of in ordinary household waste. As a consumer, you are required to take end-of-life devices to a designated collection point for the disposal of electrical and electronic equipment, in order to ensure environmentally-compatible disposal.

The return service is free of charge. Observe the current regulations in place!

Disposal of the batteries
Never dispose of empty batteries and rechargeable batteries with ordinary household waste. They contain pollutants which, if improperly disposed of, can harm the environment and human health. As a consumer, you are required by law to take them to your retail store or to an appropriate collection site depending on national or local regulations in order to protect the environment. The return service is free of charge.
The symbols for the contained heavy metals are: Cd=cadmium, Hg=mercury, Pb=lead.

Specifications

Temperature

Measuring range indoor 0°C … 50°C
Measuring range outdoor -40°C … 60°C
Resolution 0.1°C
Accuracy ±1 °C @ 0… 50 °C,

otherwise ±1.5°C

Base station

Power consumption 2 x 1,5 V AAA batteries (not included)
Housing dimension 118 x 20 (43) x 75 (73) mm
Weight 89 g (device only)

Transmitter

Power consumption| Batteries 2 x AA 1,5 V
(not included)
---|---

We recommend Alkaline batteries

Range max. 100 m (open field)
Transmission frequency 433 MHz
Maximum radio-frequency power < 10mW
--- ---
Housing dimension 48 (54) x 26 (44) x 129 (135) mm
Weight 62 g (device only)

No part of this manual may be reproduced without written consent of TFA Dostmann. The technical data are correct at the time of going to print and may change without prior notice. The latest technical data and information about this product can be found in our homepage by simply entering the product number in the search box.

EU Declaration of conformity
Hereby, TFA Dostmann declares that the radio equipment type 30.3046 is in compliance with Directive 2014/53/EU. The full text of the EU declaration of conformity is available at the following internet address: www.tfa- dostmann.de/service/downloads/ce

www.tfa-dostmann.de
E-Mail: info@tfa-dostmann.de
TFA Dostmann GmbH & Co.KG,
Zum Ottersberg 12, 97877 Wertheim, Germany

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals