truflo UFM-H2O Series Ultrasonic Flow Meter Instruction Manual

August 9, 2024
truflo

truflo UFM-H2O Series Ultrasonic Flow Meter

truflo-UFM-H2O-Series-Ultrasonic-Flow-Meter-Product

Product Information

Specifications:

  • General: Ultrasonic Flow Meter
  • Display: 9 Digital Accumulate Flow Displays, 6 Digital Instantaneous Flow Displays, Various Condition Prompt symbols and units
  • Data Storage: Ferroelectric Storage Parameters usage, recording accumulate flow of 31 months and 31 days
  • Flow Measurement Cycle: Measurement Condition: 1 Time/Second (settable); Authentication Condition: 4 Time/Second
  • Output: RS485 / USART / Infrared interface; M-BUS Alternatives, 4G module | NB-lot Wireless Transmission
  • Communication Interface: OCT Alternatives
  • Output Signal Input Power: 24VDC Lithium Batteries 8-24VDC Standard Condition
  • Power Consumption: < 30uA, each battery can work for over 15 years constantly
  • Materials: Nylon + Glass Fiber PEEK ABS
  • Measuring Pipe Sensor Protection Cover: U5 D3 C Class E2 Class IP68

Installation Instructions:
When installing the meter, the upstream straight pipe line should be 5D, the downstream straight pipe line should be 3D, 20D from the pump (D is the pipe diameter), and ensure full pipe.

How to Use the UFM-H2O Series Ultrasonic Flow Meter:

  1. Correct Install Position:
    Ensure that the meter is installed with the upstream straight pipeline at 5 times the pipe diameter and the downstream straight pipeline at 3 times the pipe diameter. The distance from the pump should be 20 times the pipe diameter.

  2. Dimensions:
    The dimensions of the flow meter are not provided in the manual. Please refer to the product specifications for detailed dimensions.

Frequently Asked Questions (FAQ):

  • Can the Ultrasonic Flow Meter be used for hot water measurements?
    The manual does not specify the fluid temperature range for measurement. It is recommended to contact customer support for specific details on measuring hot water.

  • How often should the flow meter be calibrated?
    The manual provides a Calibration Method section (Page 14) that details how to calibrate the flow meter. It is recommended to follow the calibration instructions periodically as per the manufacturer’s guidelines.

Safety Information

  • Warning | Caution | Danger
    Indicates a potential hazard. Failure to follow all warnings may lead to equipment damage, injury, or death

  • Note | Technical Notes
    Highlights additional information or detailed procedure.

  • IMPORTANT!
    indicates situations or cases which, if not avoided, could result in damage or failure of the product’s equipment.

  • Do Not Use Tools
    Use of tool(s) may damage the product beyond repair and potentially void the product warranty.

Intended Use

  • The flow meter UltraFlo® should only be used for measuring the flow of pure, homogeneous liquids.
  • The UltraFlo® is not intended for use in medical applications.
  • The volume flow meter UltraFlo® is built by industry-standard EN 61010 regulations (corresponds to VDE 0411 “Safety specifications for electrical measurement, control, and laboratory devices”).
  • The manufacturer is not liable for any injury, damage, or harm due to inappropriate or unintended use or modifications of the flow meter. Conversions and/or changes to the flow meter may only be made if they are expressly performed by the operating instructions in this operating manual.

Personnel for Installation, Commissioning, and Operation

  • Assembly, electrical installation, commissioning, and maintenance of the flow meter must be carried out by qualified, trained personnel. The qualified personnel must have read and understood the operating instructions in this operating manual and must follow the operating instructions in this manual.
  • The installer has to ensure that the flow meter is correctly connected according to the electrical connection diagrams in this operating manual.
  • Serious injury or death from electric shock may occur if wiring, installation, disassembly, or removal of wires is performed while electrical power is energized

Technological Progress

The manufacturer reserves the right to revise, alter, or modify the flow meter to the most current technology without special prior notice. Further information about the latest updates and potential additions to these operating instructions is available at www.iconprocon.com.

Introduction

  • The UltraFlo® H2O series ultrasonic water meter is a new type of water meter specially designed for agricultural irrigation, garden management, and water resources monitoring.
  • The water meter has some advantages, such as low cost, high measurement accuracy, low power consumption, stable and reliable operation, and so on. The thickness of the water meter is only 2 inches(50.8mm), which saves installation space, and the IP68 scheme is adopted for all parts so that the water meter can be used in various harsh environments. The unique structure of the water meter makes the measurement free from the influence of water quality and pressure loss, which is a breakthrough in agricultural irrigation water measurement.
  • The UltraFlo® H2O Series Flow Meter utilizes ultrasonic technology to maximize water efficiency. It has no moving parts and is lightweight making it an excellent choice for plastic piping systems. The UltraFlo® H2O all-plastic design provides excellent protection against corrosive environments and flooded wet conditions

Technical Specifications

General

Executive Standard| ISO4064-2014, GBT778-2018
Accuracy Rating| 2 Class
Measurable Fluid| Water, Sewage, Seawater (Other Liquid needs be Customized), Liquid full of Pipeline
Fluid Temperature| 0.1-30 °C
Work Environment| Temperature: -30-45°C, Ghumidity: 100% ( RH )
Pressure| 200Psi
Pressure Loss| No pressure Loss
Display

Spot Display

| 9 Digital Accumulate Flow Display, 6 Digital Instantaneous Flow Display, Arious Condition Prompt symbols and units.

Data Storage

| Ferroelectric Storage Parameters usage, recording accumulate flow of 31months and 31days
Flow Measurement Cycle| Measurement Condition: 1Time/Second (settable);

Authentication Condition: 4Time/Second

Output

Communication Interface

| RS485 / USART / Infrared interface;

M-BUS Alternatives, 4G module | NB-lot Wireless Transmission

Output Signal| OCT Alternatives
Input Power
Built-in Batteries Power| 24VDC Lithium Batteries
External Power| 8-24VDC
Power Consumption| Standard Condition < 30uA, each battery can work for over 15 years constantly
Materials
Measuring Pipe| Nylon + Glass Fiber
Sensor| PEEK
Protection Cover| ABS
Sensitivity Grade
Sensitivity Grade of Upstream Flow Field| U5
Sensitivity Grade of Downstream Flow Field| D3
Climate & Mechanical Environment Safety Grade|

C Class

Electromagnetism Compatibility Grade| E2 Class
Protection Grade| IP68

Installation Instructions

When installing the meter, the upstream straight pipeline should be ≥5D, the downstream straight pipe line should be ≥3D, 20D from the pump (D is the pipe diameter), and ensure full pipe.

truflo-UFM-H2O-Series-Ultrasonic-Flow-Meter-Fig- \(1\)

Correct Install Position

truflo-UFM-H2O-Series-Ultrasonic-Flow-Meter-Fig- \(2\)

Dimensions

Flanged Connection

truflo-UFM-H2O-Series-Ultrasonic-Flow-Meter-Fig- \(3\)


Nominal Diameter (mm | in)

| Dimensions (mm)| Flange Dimensions (mm)| ****

Pressure Class

---|---|---|---


L

| ****

H1

| ****

H2

| ****

W1

| ****

W2

| ****

D

| ****

D1

| ****

D2

| *Bolt Hole Dia Qty| Flange Thickness,C*
DN50 | 2″| 200| 82.5| 180| 82.5| 108| 165| 125| 102| 18
4| 19| 16
DN80 | 3″| 225| 100| 197| 100| 108| 200| 160| 138| 188| 20| 16
DN100 | 4″| 250| 110| 207| 110| 110| 220| 180| 158| 18
8| 22| 16
DN125 | 5″| 275| 125| 220| 125| 125| 250| 210| 188| 188| 22| 16
DN150 | 6″| 300| 142.5| 233| 142.5| 142.5| 285| 240| 212| 22
8| 24| 16
DN200 | 8″| 350| 170| 257| 170| 170| 340| 295| 268| 2212| 26| 16
DN250 | 10″| 450| 200.5| 284.5| 200.5| 200.5| 405| 355| 320| 26
12| 29| 16
DN300 | 12″| 500| 230| 310| 230| 230| 460| 410| 378| 2612| 32| 16
DN350 | 14″| 550| 260| 350| 260| 260| 520| 470| 438| 26
16| 35| 16
DN400 | 16″| 600| 290| 380| 290| 290| 580| 525| 490| 3016| 38| 16
DN450 | 18″| 700| 320| 410| 320| 320| 640| 585| 550| 30
20| 46| 16
DN500 | 20″| 800| 357.5| 447.5| 357.5| 357.5| 715| 650| 610| 33*20| 46| 16

Threaded Connection


Nominal Diameter (mm | in)

| Dimensions (mm)| ****

Pressure Grade (Psi)

---|---|---


L

| ****

H1

| ****

H2

| ****

W1

| ****

W2

DN15 | ½”| 165| 14| 123| 57| 130| 232.06
DN15 | ½”| 195| 18| 125| 57| 130| 232.06
DN25 | 1″| 160| 22| 127.5| 57| 130| 232.06
DN32 | 1 ¼”| 180| 25| 130| 57| 130| 232.06
DN40 | 1 ½”| 200| 33.5| 134| 57| 130| 232.06

Flow Ranges

Nominal Diameter (mm | in)| Turndown Ratio (R)| Flow (GPM)
---|---|---
Starting Flowrate| Minimum Flow Q1| Demarcation Flow Q2| Common Flow Q3| Overload Flow Q4
Mono Industrial Grade Small Diameter Water Meter (R=200)
DN15 | ½”| 200| 0.0136| 0.0550| 0.0880| 11.007| 13.758
DN20 | ¾”| 200| 0.0220| 0.0880| 0.1408| 17.611| 22.014
DN25 | 1″| 200| 0.0347| 0.1386| 0.2219| 27.738| 34.6725
DN32 | 1 ¼”| 200| 0.0880| 0.3522| 0.5635| 70.445| 88.057
DN40 | 1 ½”| 200| 0.1378| 0.5503| 0.8805| 110.071| 137.589
Two Channel Large Diameter Water Meter (R=100)
DN50 | 2″| 100| 0.440| 1.761| 2.817| 176.114| 220.143
DN80 | 3″| 100| 1.100| 4.402| 7.044| 440.286| 550.358
DN100 | 4″| 100| 1.761| 7.044| 11.271| 704.458| 880.573
DN150 | 6″| 100| 4.402| 17.6114| 28.178| 1761.147| 2201.433
DN200 | 8″| 100| 6.934| 27.738| 44.380| 2773.806| 3467.258
DN250 | 10″| 100| 11.007| 44.028| 70.445| 4402.867| 5503.584
DN300 | 12″| 100| 17.611| 70.445| 112.713| 7044.588| 8805.735
DN400 | 16″| 100| 27.517| 110.071| 176.114| 11007.168| 13758.961
DN450 | 18″| 100| 27.517| 110.071| 176.114| 11007.168| 13758.961
DN500 | 20″| 100| 44.028| 176.114| 281.783| 17611.470| 22014.337
Four Channel Large Diameter Water Meter (R=200)
DN80 | 3″| 200| 0.550| 2.201| 3.522| 440.286| 550.358
DN100 | 4″| 200| 0.880| 3.352| 5.635| 704.458| 880.573
DN150 | 6″| 200| 2.201| 8.805| 14.089| 1761.147| 2201.433
DN200 | 8″| 200| 3.469| 13.869| 22.190| 2773.806| 3467.258
DN250 | 10″| 200| 5.503| 22.014| 35.222| 4402.867| 5503.584
DN300 | 12″| 200| 8.805| 35.222| 56.356| 7044.588| 8805.735
DN350 | 14″| 200| 8.805| 35.222| 56.356| 7044.588| 8805.735
DN400 | 16″| 200| 13.758| 55.035| 88.057| 11007.168| 13758.961
DN450 | 18″| 200| 13.758| 55.035| 88.057| 11007.168| 13758.961
DN500 | 20″| 200| 22.014| 88.057| 140.891| 17611.470| 22014.337

Performance Error Curve

Display Instructions

truflo-UFM-H2O-Series-Ultrasonic-Flow-Meter-Fig- \(6\)

Key Description

There are two capacitive touch keys on the display, which are indicated as and .

| Scroll UP; Change Number; Active Keys and Display (press for 5 seconds).
| Scroll DOWN; Move Modify Cursor.
| Slide from to , Enter to next step menu; confirm the operation.
| Slide from to , Quit the current menu.

  • There are 4 main menu options for the water meter
  • In the menu M-0, press the modifying key will enter the sub-menu M-01, Press will display sub-menu M-02. Press will return to the main menu M-0.

M-0: Measured value display and working status, referred to as DISP

M00| Display instantaneous flow and positive cumulative flow
M01| Display instantaneous heat and positive/negative cumulative heat (heat meter)
M02| Display water pressure and water temperature (multi-function water meter)
M03| Display supply water temperature TI and return water temperature T2 (heat meter)
M04| Displays battery voltage total signal strength and signal quality
M05| Displays total propagation time (microseconds) and propagation time difference (nanoseconds)
M06| Display date time week
M07| Auto Verification Window
M08| Manual Verification Window
M09| Display the software version number and instrument serial number
M0A| Show Full Screen
M-1: Pipeline parameters and historical query, referred to as CHEC
M10| Enter the password to enter the submenu
M11| Display the number of sound channels, inner diameter, and distance between probes
M12| Display sample time, damping coefficient, angle coefficient
M13| Show negative accumulation
M14| Query the daily net cumulative amount
M15| Query the monthly net cumulative amount
M17| Display total working time and failure working time
M18| Display production date (calibration writing), calibration job number (increase calibration level)
M-2: Communication settings, referred to as COM
M21| Set date time week
M23| Set RS485 communication protocol
M24| Set RS485 , communication address
M25| RS485/MBUS communication baud rate verification
M26| Communication data display for debugging
M-3: Flow Rate Setting and Modification, for short CORR
M32| Display current zero value, low flow excision value
M33| Set instrument coefficient
M34| Set low flow excision value
M35| Reset accumulator (modify cumulant)
M36| Display instantaneous flow rate and static state zero setting
M39| Set the scale of the meter’s positive/negative cumulative flow retention
M3B| Manually set the traffic type for the default display window M00

Detailed Menu

truflo-UFM-H2O-Series-Ultrasonic-Flow-Meter-Fig- \(9\)

truflo-UFM-H2O-Series-Ultrasonic-Flow-Meter-Fig-
\(10\)truflo-UFM-H2O-Series-Ultrasonic-Flow-
Meter-Fig- \(11\)truflo-UFM-H2O-Series-
Ultrasonic-Flow-Meter-Fig- \(12\)

truflo-UFM-H2O-Series-Ultrasonic-Flow-Meter-Fig-
\(13\)truflo-UFM-H2O-Series-Ultrasonic-Flow-
Meter-Fig- \(14\)truflo-UFM-H2O-Series-
Ultrasonic-Flow-Meter-Fig- \(15\)

truflo-UFM-H2O-Series-Ultrasonic-Flow-Meter-Fig-
\(16\)truflo-UFM-H2O-Series-Ultrasonic-Flow-
Meter-Fig- \(17\)truflo-UFM-H2O-Series-
Ultrasonic-Flow-Meter-Fig- \(18\)

Calibration Method

  • The water meter is calibrated via the constant-flow method.
  • The constant-current method is to make the calibration device (standard water meter) and tested water meter into a stable flow state at a set flow point and calibrate the water meter by measuring accumulative flow at the same time.
  • Different calibration methods may cause errors. If use start-stop method to calibrate the water meter, it may cause errors. Try to extend the calibrating time when you use the start-stop method to calibrate the water meter, and make sure the duration of each turn-on time should be more than 60 seconds at least.

Communication

This series of ultrasonic water meters/heat meters can support multi-protocol communication at the same time. For example, in the same application, the above-mentioned multiple protocols can be used together or mixed. Without doing any firmware or parameter changes.

Communication Interface

  • The ultrasonic water/heat meter is equipped with 3 physical serial communication interfaces, one is an IR infrared communication interface, one is a TTL/USART level interface, and the other is an RS485/MBUS Parallel dual interface.
  • USART is a logic-level serial interface, suitable for direct connection to other external MCUs, such as communication modules such as NB-iot, Lora, etc., and can also be directly connected to pressure and temperature modules such as SPI interface, etc., with power consumption Very low features.
  • The IR infrared communication interface is the standard interface of this series of ultrasonic water meters, which meets the standards stipulated in CJ188.
  • The IR infrared communication interface will automatically enter the low power consumption state within 300 seconds without communication. Touch the button to wake up the infrared interface; the above three physical serial communication interfaces are independent of each other, allowing the three serial ports to be accessed at the same time without interfering with each other.

Serial Communication Interface Default Settings

  • RS485/MBUS
  • 9600, N,8,1IR 9600, N,8,1
  • TTL/USART 9600, N,8,1

Change serial port default settings
Serial port parameter setting can be done through a PC -based software named V00_SETUP. The current serial port parameters are displayed in the M26 menu.

Supported Communication Protocols

  • MODBUS
  • M-BUS
  • CJ188

The above protocols can be supported at the same time. Some protocols can be cross-supported at the same time.

Warranty, Returns & Limitations

Warranty
Icon Process Controls Ltd warrants to the original purchaser of its products that such products will be free from defects in material and workmanship under normal use and service in accordance with instructions furnished by Icon Process Controls Ltd for a period of one year from the date of sale of such products. Icon Process Controls Ltd obligation under this warranty is solely and exclusively limited to the repair or replacement, at Icon Process Controls Ltd option, of the products or components, which Icon Process Controls Ltd examination determines to its satisfaction to be defective in material or workmanship within the warranty period. Icon Process Controls Ltd must be notified pursuant to the instructions below of any claim under this warranty within thirty (30) days of any claimed lack of conformity of the product. Any product repaired under this warranty will be warranted only for the remainder of the original warranty period. Any product provided as a replacement under this warranty will be warranted for the one year from the date of replacement.

Returns
Products cannot be returned to Icon Process Controls without Icon’s prior authorization. To return a product that is thought to be defective please submit a customer return (MRA) request form and follow the instructions therein. All warranty and non-warranty product returns to Icon Process Controls must be shipped prepaid and insured. Icon will not be responsible for any products lost or damaged in shipment.

Limitations

This warranty does not apply to products which:

  1. Are beyond the warranty period or are products for which the original purchaser does not follow the warranty procedures outlined above;
  2. Have been subjected to electrical, mechanical, or chemical damage due to improper, accidental, or negligent use;
  3. Have been modified or altered;
  4. anyone other than service personnel authorized by Icon Process Controls Ltd have attempted to repair;
  5. have been involved in accidents or natural disasters;
  6. Are damaged during return shipment to Icon Process Controls.

Icon Process Controls reserves the right to unilaterally waive this warranty and dispose of any product returned to Icon where:

  1. There is evidence of a potentially hazardous material present with the product;
  2. The product has remained unclaimed at Levelpro for more than 30 days after Icon Process Controls dutifully requested disposition.

This warranty contains the sole express warranty made by Icon Process Controls Ltd in connection with its products. ALL IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, THE WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSLY DISCLAIMED. The remedies of repair or replacement as stated above are the exclusive remedies for the breach of this warranty. IN NO EVENT SHALL Icon Process Controls Ltd BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY KIND INCLUDING PERSONAL OR REAL PROPERTY OR FOR INJURY TO ANY PERSON. THIS WARRANTY CONSTITUTES THE FINAL, COMPLETE, AND EXCLUSIVE STATEMENT OF WARRANTY TERMS AND NO PERSON IS AUTHORIZED TO MAKE ANY OTHER WARRANTIES OR REPRESENTATIONS ON BEHALF OF Icon Process Controls Ltd. This warranty will be interpreted under the laws of the province of Ontario, Canada.

  • If any portion of this warranty is held to be invalid or unenforceable for any reason, such finding will not invalidate any other provision of this warranty
  • For additional product documentation and technical support visit www.iconprocon.com | e-mail: sales@iconprocon.com support@iconprocon.com | Ph: 905.469.9283.

www.iconprocon.com | sales@iconprocon.com | 905.469.9283.