ISO 9847 Pyranometer
User Guide
Hukseflux
Thermal Sensors
ISO 9847 Pyranometer
ISO 9847 & ASTM G207 CALIBRATION
The world is moving towards indoor calibration of pyranometers.
The leading pyranometer manufacturers and many others, calibrate pyranometers
indoors at normal incidence, according to the ISO 9847 and equivalent ASTM G
207 standards. Compared to outdoor calibration, the method has the fundamental
advantage that directional response and temperature response are perfectly
reproducible. It works under generally accepted reference conditions: 0 °
zenith angle is the reference for directional error and 20 °C is the reference
instrument temperature.
Figure 1 SRC02 indoor pyranometer calibration system
Figure 2 Application example: sensor being prepared for calibration in an
indoor calibration system
Introduction
The tradition of outdoor calibration of pyranometers was born out of necessity. The directional- and temperature response of the first pyranometers were unreliable and could not accurately be determined. Local calibration at realistic solar angles and temperatures was the way to include the unknown directional error and temperature response into the calibration. Modern instruments are produced within known and narrow performance limits. For these sensors, indoor calibration is best. Calibration reference sensors are still calibrated outdoor under the natural sun to minimise spectral errors.
ISO 9847, IIc indoor calibration
The indoor calibration method of ISO 9847, type IIc, works by transferring the sensitivity of a calibration reference sensor to an identical test sensor under a lamp. The reference sensor is calibrated outdoors under the spectrum of the natural sun. The IIc procedure involves an unshaded and a shaded measurement and exchange of the instrument positions in the potentially inhomogeneous beam of the lamp. The procedure also includes a beam-stability verification. The method is a “transfer” between identical instruments and does not rely on intensity or the spectral composition of the lamp.
Requirements
The IIc method requires use of a reference sensor with identical sensor and optics as the test sensor. Sensors of the same model have identical spectral sensitivity and identical non-linearity. Using a different model / brand or higher class sensor leads to errors; if the spectral response or linearity of the calibration reference differs from the sensor under test, the use of a low intensity, low-colour temperature lamp instead of the natural sun, introduces linearity- and spectral errors.
Standards
ISO 9847: Solar Energy Calibration of field pyranometers by comparison to a reference pyranometer. ASTM G207 – 11 Standard test method for indoor transfer of calibration from reference to field pyranometers.
Calibration hierarchy
The calibration reference sensor is traceable to WRR. Typically, the
calibration reference used in indoor calibration is calibrated following ISO
9846, type I outdoor calibration. However, you may also add another step in
the hierarchy using reference sensors calibrated indoors.
Calibration reference conditions
If the temperature response, non-linearity and directional response of the
calibration reference are known, you may apply corrections from outdoor
calibration conditions to the typical reference conditions, or add a general
“transfer error” to cover possible differences. Reference conditions are not
standardised, but the main manufacturers use:
- irradiance level 1000 W/m2
- normal incidence irradiance
- instrument temperature 20 °C
- horizontal instrument position
- spectrum: solar irradiance on a clear day
The calibration lamp only serves as a means to compare the calibration
reference to an identical sensor under test. The above reference conditions
therefore also apply to the indoor calibration result. The calibration remains
traceable to the calibration reference conditions (including the spectrum and
irradiance level) that were valid for the calibration reference sensor during
its outdoor calibration. The lamp typically has a colour temperature of around
3000 K.
Benefits
- calibration at normal incidence, which is the reference condition for directional response
- calibration at 20 ° C which is the reference condition for instrument temperature
- change of sensitivity is directly traceable to sensor / coating degradation
- reference condition comply with IEC Standard Test Conditions for solar energy testing (STC), as applied in Photovoltaic (PV) module and system testing
- fast, independent of weather
Reference users of the same method
- Hukseflux: The Netherlands, USA, India, China, Japan, Singapore, Australia, Brazil
- ISOCAL North America Inc., USA
- GeoSUN Africa, South Africa
- TÜV Rheinland, Germany
- DWD Deutscher Wetterdienst, Germany
- KNMI The Netherlands Meteorological Institute, The Netherlands
- EKO Instruments, Japan
- Kipp & Zonen B. V., The Netherlands, UK, Germany, France, Singapore, USA
- Campbell Scientific Canada, Canada
- Meatech, India
Further endorsements
In the new IEC 61724-1:2017 Photovoltaic
System Performance Monitoring – Guidelines for Measurement, Data Exchange and
Analysis allows calibration according to ISO 9847.
Attainable uncertainty at Hukseflux
Typical contributions to the uncertainty budget are the uncertainty of the
calibration reference, 0.75 % uncertainty of the transfer to reference
conditions and 0.5 % uncertainty of the method.
The calibration reference uncertainty is higher for lower class sensors.
The expanded calibration uncertainties (k = 2) we attain at Hukseflux for
different pyranometer classes:
- < 1.2 % for Class A
- < 1.5 % for Class B
- < 1.8 % for Class C
Avoiding uncertainty in directional response, these uncertainties are not far from those of high-quality outdoor experiments.
About Hukseflux
Hukseflux is the leading expert in measurement of energy transfer. We design
and manufacture sensors and measuring systems that support the energy
transition. We are market leaders in solar radiation- and heat flux
measurement. Customers are served through the main office in the Netherlands,
and locally owned representations in the USA, Brazil, India, China, South East
Asia and Japan.