Scigiene HMC_BLW FLM PLUS Marijuana Moisture Meter User Manual

July 12, 2024
Scigiene

Scigiene HMC_BLW FLM PLUS Marijuana Moisture Meter

Scigiene-HMC-BLW-FLM-PLUS-Marijuana-Moisture-Meter-
fig-1

Specifications

  • Product Name: Humimeter FLM PLUS Marijuana Moisture Meter
  • Sensor: Cone (loose bud) sensor
  • Measuring Unit Range: 0 – 100%
  • Calibration Curves: Marijuana Curve, Digit Curve, Free calibration curves 1-3
  • Power Source: 4 x 1.5V Alkaline AA batteries
  • Startup Time: Approx. 3 seconds

Product Usage Instructions

Start-up:

  1. Remove the rubber protection cover and insert 4 new 1.5V Alkaline AA batteries following the correct polarity.
  2. Press the power button for approximately 3 seconds to switch on the device.
  3. After startup, check the displayed information for serial number, software version, battery status, and memory allocation.
  4. To switch off the device, press the button for about 3 seconds or leave it idle for 10 minutes for automatic shutdown.

Changing the Sensor:
To change the sensor, screw the desired sensor onto the device. The instrument will automatically recognize the sensor and display corresponding calibration curves.

Measuring Procedure:

  1. Ensure that the device temperature matches the material to be measured by letting it adjust to the surrounding temperature for at least half an hour.
  2. Avoid direct sunlight exposure during measurement.

FAQ

  • What is the purpose of the Marijuana Curve in calibration?
    The Marijuana Curve shows the current water content in % of marijuana loose bud using the cone (bud) sensor.

  • How do I determine the material reference moisture using Humimeter FLM?
    The humimeter FLM calculates the moisture referred to the total mass using the formula %F = (Mn/Mt) * 100, where Mn is the mass of the sample with average moisture content, Mt is the mass of the dried sample, and %F is the calculated absolute moisture content.

SPECIFICATION

Calibration curve| Explanation| Compressed density| Unit| Measuring range| Sensor
---|---|---|---|---|---
Marijuana Curve| Marijuana loose

bud

| 100 to 160 kg/m³| % wc| 6 – 40%| 13736
Digit| special products| –| | 0 – 100| 13736
Empty 1| special products| –| | | 13736
Empty 2| special products| –| | | 13736
Empty 3 RT| special products| –| | | 13736
Test block| Only for device check, not for measuring!| 13736

Explanation of calibration curves

Marijuana Curve:
Shows the current water content in % of marijuana loose bud using the cone (bud) sensor!

Digit:

  • The digit curve is a unit less calibration curve with a range from 0 up to 100, which corresponds to the entire measurement range of the device. With this curve, special products can be measured.
  • The higher the value the wetter the material. By means of a comparative measurement by a reference method, a table with comparison values can be created.
    Very dry: 0 very wet: 100

Free calibration curves 1-3:

  • There are four free calibration curves which can be used for measuring special products (temperature compensated).
  • On request Scigiene / Schaller GmbH can develop customer specific calibration curves for your product.

Test block:
This calibration curve is only used for checking the instrument with the optional test block accessories; part# HM-13888-M (Calibration Test block for FLM Humimeter cone/bud sensor)

Determination of the material reference moisture

The humimeter FLM determines the water content, which means that it calculates the moisture referred to the total mass:

Scigiene-HMC-BLW-FLM-PLUS-Marijuana-Moisture-Meter-
fig-2

Humimeter FLM With cone (loose bud) sensor

Scigiene-HMC-BLW-FLM-PLUS-Marijuana-Moisture-Meter-
fig-3

Start-up

  • Insert the four delivered 1.5V Alkaline AA batteries as described below:
    1. At first remove the rubber protection cover. For that, hold the rubber housing at the upper side and pull it over. In case of an optional USB interface you must remove the protection cap before.
    2. Press with your finger onto the arrow of the battery cap and pull it back.
    3. Put four new batteries 1.5 Volt AA Alkaline batteries in the device. Make sure that the position of the battery poles is correct.
    4. Press down the batteries and close the cap.Scigiene-HMC-BLW-FLM-PLUS-Marijuana-Moisture-Meter-fig-4
  • For switching on the instrument, press the button for approx. 3 seconds. Now the LCD display lights up. After switching on, the serial number, the software version, the battery status and the memory allocation are displayed. Then the instrument is ready for use. For switching off the instrument, press the button again for approx. 3 seconds. If no button is pressed for 10 minutes, the instrument switches off automatically.

Changing the sensor
For changing the sensor, just screw the desired sensor on the humimeter device. The instrument automatically recognises the sensor and shows the corresponding calibration curves.

Measuring procedure

  1. For a correct measurement please ensure that the device has the same temperature than the material to measure. For that reason, let your device adjust to the surrounding temperature of the material for at least half an hour before measuring (protect from direct sunlight!)
  2. Switch on the device: Press the key for 3 seconds.
  3. Connect the desired sensor (if no sensor is connected, the display shows “no sensor”).
  4. Now select the desired calibration curve by pressing the buttons . (Pressing the key in the measuring window for at least 3 seconds, a list with all available sorts will appear. Curves displayed in grey are not available for the currently connected sensor.)
    The currently set calibration curve is indicated in the upper part of the display.
    For the explanation of the different curves, please see page 2 of this manual .Scigiene-HMC-BLW-FLM-PLUS-Marijuana-Moisture-Meter-
fig-5

Marijuana cone (loose bud) sensor:
To ensure correct measuring results, the measuring chamber has to be full to the brim with material, but must not be precompressed (see illustration picture). Put the cap (compressing fixture) onto the measuring chamber and tighten it to the stop. As soon as the measuring chamber is completely closed, the display shows the measuring value.

Scigiene-HMC-BLW-FLM-PLUS-Marijuana-Moisture-Meter-
fig-7

Marijuana cone/ground samples gravimetric procedure
To ensure correct measuring results, Using the provided scale and beaker wight out 3.0gm of cannabis products (whole buds or ground products) and add this to the Umbel chamber Put the cap (compressing fixture) onto the measuring chamber and tighten it to the stop. As soon as the measuring chamber is completely closed, the display shows the measuring value.

Scigiene-HMC-BLW-FLM-PLUS-Marijuana-Moisture-Meter-
fig-1

Scigiene-HMC-BLW-FLM-PLUS-Marijuana-Moisture-Meter-
fig-8

Keypad symbols

Scigiene-HMC-BLW-FLM-PLUS-Marijuana-Moisture-Meter-
fig-9

Operating the instrument

  • Switching on:
    Press the button for 3 seconds.

  • Setting date and time:
    2 times > Options -> Date/Time

  • Saving measuring data:
    The shown value can be saved by pressing the button . Name the saved value by pressing the button.

  • Hold:
    Select the menu item “Datalog time“ in the menu “Options“. Activate “Hold” there and change to the measuring window again. Pressing the button below the symbol , the measuring value will remain on the display until another button is pressed.

  • Display lighting:
    Press the key briefly; the display lighting switches off automatically after approx. 30 seconds. Pressing any key activates the display lighting again.

  • Switching off:
    Press the key for 3 seconds. The instrument switches off after releasing the key. The instrument switches off automatically after approx. 10 minutes.

  • Measuring range:
    If the measuring value is blinking in grey, the valid measuring range is exceeded. In this case the accuracy will be decreasing.

Activation of the “super user” function

2 times – Options – Unlock
Enter the 4-digit password by using the button (standard is the 4-digit serial number) and confirm by pressing the button.

Changing the user level

Changing from advanced user to single user:

  • Make sure that you have activated the “super user” functions as per the instructions above. Afterwards change to the menu and choose “Options”.
  • In the submenu please select “o Userlevel” (2 times – Options – o Userlevel)
  • Confirm by pressing the button. Now the single user is activated.

Changing from single user to advanced user:

  • Keep both the buttons pressed directly after switching on the device. Your humimeter automatically starts the main menu. Activate the “super user” functions as per the instructions above.
  • Navigate to “Options – o Userlevel” and confirm by pressing the button.

Changing the batteries

If the battery symbol appears in the measuring window or if a critical charge of battery is shown in the status (!), the batteries must be changed IMMEDIATELY. If you do not use your humimeter device for a longer period, remove the batteries. For eventual resulting damages, we cannot provide any warranty.

Exemption from liability

  • For misreadings and wrong measurements and of this resulting damage we refuse any liability.
  • This is a device for quick determination of moisture. The moisture depends on multiple conditions and multiple materials. Therefore, we recommend a plausibility check of the measuring results.
  • Each device includes a serial number and the guarantee stamp. If those are broken, no claims for guarantee can be made. In case of a faulty device, please contact Scigiene (www.scigiene.com).

Transfer saved data to the PC

(only with optional USB data interface module)

  • To send your saved logs to the PC, connect the humimeter device to your PC using the USB cable that was provided with your device. Carefully loosen the protection cap on your humimeter and plug in the USB mini B connector.
  • The bigger connector should be connected to a USB slot on your PC. Start the LogMemorizer software on your PC and switch on your humimeter. The data transfer can be started on your humimeter or on the software.

Starting the data transfer on the humimeter:
Press the key until you reach the menu (see image on the right). Then choose “Send Logs” and confirm by pressing the key. Now choose ”Manual Logs” and confirm with again. All saved logs will be sent to your PC.

Starting the data transfer on your PC:

  • Press the field “communication” in the LogMemorizer software. The menu window shown on the right opens.
  • For transferring the data you can select “Import last manual log” (the last saved measuring series is transferred) or “Import all manual logs” (all saved logs are transferred). If you click on one of these menu items, the transfer starts immediately.
  • For the initial configuration of the software, please press the F1 key of your PC and read the help file.

Print saved data

(only with optional USB data interface module and Schaller thermo printer)

  • To print your saved data, connect the device to the printer using the printer cable that was delivered with your device. Carefully loose the protection cap on the humimeter. At first plug in the side of the connector with the close plastic casing at the humimeter.
  • Then switch on the device.
  • Not till then the other side of the cable has to be plugged in at the printer. Switch on the printer by pressing . Now the green LED is blinking. If it does not blink, please change the batteries and try again.
  • Press the button at your humimeter until you reach the menu (see image on the right). Choose “Print Logs” and confirm by pressing .
  • Now you can select if you want to print the last saved measuring series or all saved measuring series (logs).
  • Confirm by pressing again. The selected logs are printed out now.
  • To save paper, please think of clearing the data storage regularly.Scigiene-HMC-BLW-FLM-PLUS-Marijuana-Moisture-Meter-fig-22

Device maintenance instructions

  • To ensure your device lasts for a long time, please do not expose it to strong mechanical loads or heat e.g. dropping it or direct sunlight exposure. Clean your device using a dry cloth. Any kind of wet cleaning damages the device.
  • The instrument is not rainproof. Keep it in dry areas. When the device isn’t used for a longer period (2 months) or when the batteries are empty, they should be removed to prevent a leakage of the battery acid.

Application range air humidity sensors

Within the normal application range (normal range) the accuracy of the device is as indicated. A long-term application beyond the normal application range (max. range), particularly at an air humidity of more than 80%, can lead to higher measuring errors (+3% after 60 hours). Back in the normal application range, the sensor will return to the indicated accuracy automatically.

Scigiene-HMC-BLW-FLM-PLUS-Marijuana-Moisture-Meter-
fig-23

Packaging

If the device isn´t used for a long duration of time, store it in the cardboard cylinder (or optionally the wooden or plastic case) that the device was delivered with. Do not discard the packaging! In case of returning the instrument for warranty or repair, use the original packaging. For damages in transport due to inadequate packaging; we refuse any liability.

Technical data

  • Resolution of display
    • 0.1% material moisture (Marijuana)
    • 0.5°C temperature
  • Operating temperature 0°C to 40°C / 32 to 104°F
  • Temperature range -15°C to 85°C / 5 to 185°F
  • Storage temperature -20°C to 60°C / -4 to 140°F
  • Temperature compensation automatically
  • Power supply 4 pcs. of 1.5Volt AA Alkaline batteries
  • Auto Switch Off after approx. 10 minutes
  • Current consumption 55 mA (with lighting)
  • Display 128 x 64 matrix display, lighted
  • Dimensions 145 x 65 x 27 mm
  • Weight approx. 250g (with batteries and rubber protection cover)
  • Degree of protection IP 40
  • Scope of supply Humimeter FLM and cones (marijuana bud) sensor with rubber protection cover, 4 of 1.5Volt AA Alkaline batteries
  • Optional accessories External sensors, test block, software etc.

External sensors

Measurement: measuring range / resolution / accuracy

Marijuana cones (loose bud) moisture sensor (13736):

  • water content: 6 to 40% / 0.1%
  • temperature °C: -15 to +85°C / 0.5°C / ±0.5°C (at 25°C)
  • temperature °F: 5 to 185°F / 0.9°F / ±0.5°F (at 77°F)

Most common reasons for misreadings

  • Product temperature out of application range
    Material below 0°C or above +40°C may cause faulty measurements. The storage of cold material in a warm storage area usually creates condensed water which may lead to major measuring errors.

  • Too short conditioning time

  • Discrepancy in temperature between device and material
    Please ensure that the device and the material under test are being stored at the same temperature before measuring. Protect your measuring device from direct sunlight for a reasonable time period before taking a measurement. A high temperature difference has a negative effect on the stability of the measurement results.

  • Wrong calibration curve for warmed-up insertion probe

  • Wrong calibration curve
    Double check the correct selection of the calibration curve before measuring.

  • Wet or mouldy material

  • Frozen measuring material or material containing snow This leads to a major decrease in accuracy.

  • Plug-in direction (for bales)
    The plug-in direction has a great influence on the accuracy. Necessarily follow the instructions according to page 13 of this manual!

  • Compressed density outside the application range If the compressed density differs from that specified, there may be deviations.

  • Moving the measuring head after plugging in leads to misreading!

  • Water film at the measuring head
    After measuring wet material a water film can arise on the sensor head. This may cause a too high result in the following measurements. After measuring wet material, clean both black plastic parts of the measuring head accurately with a dry cloth.