Scigiene SP-840006 Light Meter Installation Guide
- June 10, 2024
- Scigiene
Table of Contents
Light Meter
Part#: SP-840006
Instruction Manual
INTRODUCTION
Compact and portable, your new meter can be used to monitor light levels or check the LUX level of a particular light source. The sensor’s exclusive photo diode and color correction filter meets the C.I.E. photopic spectrum with a cosine correction factor meets the standard. Scigiene light meters are cosine corrected for a quick and accurate response. Large easily legible LCD. The meter has an external zero adjustment, data hold, low-battery indicator, and a moisture resistant front panel. Comes with soft carrying case, detachable light sensor, photo-sensor cover, instructions and a 9-Volt battery.
PANEL DESCRIPTION
1. DISPLAY
2. ON button
3. OFF button
4. HOLD button| 5. X1 (2,000 RANGE) button
6. X10 (20,000 RANGE) button
7. ZERO SETTING terminal
8. SENSOR INPUT socket| 9. STAND
10. BATTERY
11. PHOTO SENSOR
12. SENSOR plug
---|---|---
MEASUREMENT PROCEDURES
- Insert the SENSOR plug into the SENSOR INPUT socket.
- Turn the meter on using the ON button.
- Remove the lens cap from the PHOTO SENSOR.
- Press the 1X button for measurements of 0 to 1999 LUX, or select the 10X button for measurements of 2,000 to 19,990 LUX. If you are uncertain about which range to use, select the highest range. In 10X mode, “ “appears in the right bottom LCD.
- Range Overload is indicated when “1” appears on the left side of the display. If this occurs, switch to the highest range or discontinue use.
- Point the PHOTO SENSOR toward the source and read the results on the display. To convert LUX results to FC, multiply the meter reading by the factor of .0929
- Push the HOLD button to freeze the reading in the display. The word “HOLD” is displayed. Press the HOLD button again to exit this function.
- Press the OFF button to turn the unit off.
NOTES:
-
Keep the lens cap on the PHOTO SENSOR at all times except when actually taking a reading.
-
Fluctuations in the reading are generally due to shadows or fluctuations in the line voltage.
Ambient temperature, drafts and ventilation also affect the luminous flux output. -
Avoid range overload and do not store in areas of high temperature and/or humidity. Remove the battery for long-term storage.
ZERO ADJUSTMENT & CALIBRATION
WARNING: Only use the terminal marked “ZERO” for adjustment. The unmarked terminals are for professional laboratory calibration only and tampering with these terminals will result in inaccurate readings and require professional calibration of the meter.
- With the sensor cover securely in place and the meter turned on, the LCD display should indicate “0000”. A slight variation will not affect readings; however, if a precise reading is required you may use the ZERO SETTING terminal to reset the unit.
- To make an adjustment, cover the PHOTO SENSOR and press the 1X button.
- Use a small screwdriver to turn the ZERO SETTING terminal until “0000” is displayed.
- To maintain accuracy, annual laboratory calibration is recommended.
BATTERY REPLACEMENT
Replace the battery when the low battery icon is displayed in the left corner of LCD. In-spec measurements may be made for several hours after the low battery indicator appears. Remove the battery compartment screw and slide the cover away from the meter, remove the battery and replace with a 9V battery (alkaline or heavy-duty type). Replace the cover.
SPECIFICATIONS
Display | Maximum reading 19,990 |
---|---|
Overload Indicator | 1 is displayed on the left of the LCD |
Range* | 2,000 Range = 0 to 1,999 LUX |
20,000 Range = 2,000 to 19,990 LUX
Resolution| 2,000 Range = 1 LUX
20,000 Range = 10 LUX
Accuracy| ±5% + 4 digits
Operating Environment| 32 to 122°F (0 to 50°C), less than 80% RH
Weight| 9.9 oz (280g)
Dimensions| Main Unit: 7.9 x 2.7 x 1.2″ (200 x 68 x 30mm)
Photo Sensor: 3.2 x 2.2 x 0.3” (82 x 55 x 7mm)
Photo Sensor Lead: Extends to appx. 3½ ft
Optional Accessories| SP-840090 – Water Resistant Instrument Pouch
SP-840093 Field Tripod
*Electronic Specifications (23 ± 5°C): Tested under the environment RF Field Strength less than 3 V/M, and frequency less than the 30 MHz only.
LIGHT SPECTRUM CHART
WARRANTY
Scigiene warrants this product against defects in materials and workmanship for a period of five (5) years from the date of purchase, and agrees to repair or replace any defective unit without charge. If your model has since been discontinued, an equivalent Scigiene product will be substituted if available. This warranty does not cover probes, batteries, battery leakage, or damage resulting from accident, tampering, misuse, or abuse of the product. Opening the meter to expose its electronics will void the warranty. To obtain warranty, please contact Scigiene.
https://www.scigiene.com/
1295 Morningside Avenue, Unit 16-18
Scarborough, ON M1B 424 Canada
Phone: 416-261-4865 Fax:
416-261-7879
www.scigiene.com
Read User Manual Online (PDF format)