Table of Contents
OperateManual
AEOLUS-370 implant
Safety Summaries
1.1 Notice Item
Note: This product is connected fine electrical components with wires.
When the equipment is failure, please do not try to disassemble and repair it
by yourself. The maintenance must be solved by the manufacturer or
professional personnel with relevant qualifications authorized .
1.2 Safety measures
Please observe the following basic safety precautions when use this product.
- Please cut off the power after use this equipment every time
- Please don’t tilt the equipment when using it.
- Please don’t touch the equipmet’s shell when the environment is wet
- Please don’t place the equipment in flammable, explosive and corrosive gas environment。
- Please don’t reuse disposable collection bags and disposable spitting tubes.
- Please put the foot switch in place after using the equipment
- Please don’t step on the foot switch again。
- Please don’t use additional parts except recommended by the manufacturer, avoid to reduce the performance of the machine
- Please don’t disassemble the machine. It should be repaired by qualified maintenance personnel.
- please do not use the equipment when the power cord or plug is damaged, the machine can’t work normally, the machine falls or is damaged, or the machine falls into the water.
- Contact the manufacturer immediately to negotiate and provide solutions.
Product Feature
2.1 Produt Use
- Product use range: It is used for vacuum suction during medical operation.
- This product’s consumables is disposable item, instant change and use, easy to operate, time-saving and worry-saving
- This product adopts the movable suction device made of oil-free lubrication negative pressure pump
2.2、Structural features and working principle
2.2.1 Product Structure 2.2.2 Working
Principle
- The product structure is composed of piston pump, vacuum meter, negative pressure regulating valve, check valve, waste liquid collection bottle, suction pipe and foot switch.
- Adopt oil-free lubrication piston pump, without oil mist pollution.
- Acrylic panel, high-end atmosphere, fashionable shape.
- Low noise, high suction, foot switch control, convenient operation.
- Equipped with check valve to prevent liquid from flowing into the pump.
- The negative pressure regulating valve can adjust the any pressure according to the demand.
Schematic diagram :
Main Parameter of the Product
Modle|
---|---
Voltage| 220V±10%
Power| 100W
Frequency| 50Hz
Rate Current| 0.48 A
Max Vacuum| 0.09Mpa
Air Flow| Noise| Bottle Capacit Y| Use life| N.W| Product size
---|---|---|---|---|---
201-/min| 55dB| 2L| 5 Year| 20KG| 3448.5 95cm
Working Conditions
Ambient temperature range: +5 °C ~ +40 °C
relative humidity range: 30% – 80%
Atmospheric pressure range: 860hpa ~ 1060hpa
Note: if the storage temperature of the equipment is lower than 5 °C, the
equipment should be placed in the normal working temperature environment for
more than 4 hours before use
Product Acceptance
5.1 Before installing and debugging the product, first users should be check
whether the product appearance is good, whether the type and quantity of
accessories are consistent with the accessories list in the manual. If there
is any shortage or damage, please contact the supplier or manufacturer in
time.
5.2 Identification check
After unpacking, please carefully check whether the nameplate, label and
identification of the product are complete.
5.3 Acessories List
Item | Standard | Quantity |
---|---|---|
Suction machine host | host | 1 pcs |
Transparent bottle | 2 L | 1 pcs |
Disposable collection bag | 2L | 50pcs |
Disposable suction tube | 8*6 | 50pcs |
Disposable suction tip | / | 50pcs |
Check valve | 1pcs | |
Operate manual | / | 1pcs |
Warranty card | / | 1pcs |
5.4 Connection Pipeline 5.5
Connecting Powe r
Take out the power cable , turn on the the power switch, power indicator is
on(power is on).
Note:The power socket must have three core socket to ensure reliable
grounding. It is forbidden to bend the ground wire insert and insert it into
the two core socket.
5.6 Calibration pipeline
Turn the negative pressure regulating valve with clockwise to the maximum ,
block the suction tube / tip with your fingers
Turn on the equipment, the equipment start working without abnormal noise, the
pointer on the vacuum meter will quickly rise to the limit negative pressure
value.
Open the suction tube / tip, the meter needle will return to 0mpa, indicate
the pipeline connection is correct.
5.7 Negative pressure regulation
- Block the suction inlet, turn on the power switch, adjust the negative pressure regulating valve, the negative pressure should be change within the range of 0~0.09mpa in the vacuum meter.
- In clinical use, the negative pressure regulating valve is used to control the negative pressure value
- When the equipment is overloaded, do not shut down immediately. Turn off the power supply after the negative pressure value is close to 0.
5.8 Operate Method
This product adopts the power switch and foot switch to control on & off, the
on & offmode can be switched any one.
The specific operations as follows
- Turn on the power, press the start button on the machine. At this time, the machine start working.
- When the machine work, gently step on the foot switch, the machine stop working.
- When the machine stop work, gently step the foot switch again , the machine start working again.
Replace Consumable Material
- Note: the waste liquid collection bag, suction tube and suction head in this product are all disposable. Do not reuse it after cleaning, avoid to affect health and safety.
- Please turn off the power when replace the disposable consumable material.
- Please replace the consumable material according to below operation:
Operate step:
- Turn off the power.
- Pull off the suction tube / tip from the disposable collection bag.
- Take out the disposable collection bag.
Electrial Principle
Electrical maintenance
should be operated by professional person
Troubleshooting
phenomenon | analysis | Solution |
---|---|---|
Unable to start | Whether the power plug is connected | Connect the Power Plut |
Whether the power switch is turned on | Turn on the power switch | |
open the control switch failure | Please inspect or replace | |
Suction efficiency reduction | Entrance is blocked | Inspect the Suction |
Pipeline
Suction pipeline leak the air| Inspect the suction Pipeline & connector if
loosen
Pump overloading| Whether poor voltage| Inspect the power voltage
Equipment repair and maintenance
- Proper maintenance can effectively improve the efficiency of the equipment, reduce failure rate, and help to extend the service life of the equipment.
- Please check if the pipe is blocked every week. If the pipe is blocked, the suction efficiency will be reduced. when serious cases, the motor maybe damaged due to overload operation for a long time.
- Please check if the transparent bottle is broken every week. Because the crack of the tank will seriously affect the suction power of the equipment, once the crack is found, please contact the manufacturer immediately to negotiate a solution.
- Check the sealing ring at the transparent bottle mouth every month for wear and displacement.
- please check the circuit and electronic components of the equipment every year ,to ensure the wires are not aging and the electrical components can operate normally.
After Serive
The company have an responsible for the return, replacement and repair of non- human quality problems of this product within one week from the sale date ; Under normal use and storage, if the product has quality problems within one year from the sale date , the company will provide free maintenance; If the product has quality problems within one year after.