Blue-White F-550 Flow Meter Industries Instruction Manual
- June 23, 2024
- Blue-White
Table of Contents
Blue-White F-550 Flow Meter Industries Instruction Manual
SAFETY INFORMATION
Please read this manual completely before unpacking, installing, and operating this equipment. In particular, pay attention to all dangers, warnings, and precautions, otherwise, it may cause serious injury to the operator or damage to the equipment.
Symbol | Description |
---|---|
Warning (Risk of electric shock) | |
Caution (Refer to the user’s guide) |
| Ground, Protective Conductor Terminal
Note: When in doubt regarding your electrical installation, contact a licensed electrician.
Introduction
Congratulations on purchasing the F-550 Flow Meter!
Your Flow Meter is calibrated and ready for use. The unit is designed to give
many years of service. For more information or questions on this product or
any of our Industrial products, please visit us at www.blue-white.com
What’s in the box
-
Flow meter (1)
(Flow meter may contain plastic tubing for protection during shipping. Remove before installation.) -
Quick Start Guide
Check all packaging for completeness and inspect for damage before installation.
System Features
- Dual Scale (GPM / LPM), screen printed with white backing on opposite side for easy reading
- Tested and calibrated for clean water applications
- Acceptable for outdoor installations – One piece machined acrylic meter body
- Threaded adapters and panel mount option make for simple install on piping
- Unique O-ring design prevents leaks
- Upward flow, vertical installation only
Application Notes
Although meters may be suitable for other chemicals, Blue-White cannot guarantee their suitability. It is the responsibility of the user to determine the suitability of the flow meter in their application. Flow meters are tested and calibrated for water only.
Ordering Information / Model Number Matrix
Engineering Specifications
Flow Meter Specifications
Maximum Working Pressure | 250 psi (17.2 bar) @ 70 ˚F (21 ˚C) |
---|---|
Maximum Fluid Temperature | 200˚F (93 ˚C) @ 0 PSI |
Full Scale Accuracy | +/- 5% |
Calibration Fluid | Water, specific gravity 1.0 |
Scale Length | 4” (100mm) |
Environment | Acceptable for direct sunlight exposure |
Maximum Pressure Drop | 3 PSI |
Approximate Shipping Weight | 2 lb (.91 kg) |
Materials of Construction
Wetted Components:|
---|---
Meter Body| Cast acrylic rod, Annealed
Adapters| Polypropylene
Guide Rod Holder| Polysulfone
O-ring Seals| Viton® (optional EP)
Float| 316SS or PTFE (varies by model)
Guide Rod| 316SS
Non-Wetted Components
Bulkhead Nuts| Polypropylene
Temperature vs. Pressure Graph
Flow Specifications
| Feed Rate| | Max Pressure| Max Temperature|
---|---|---|---|---|---
Model| | | | | Float Material
| GPM| LPM| PSI (bar)| °F (°C)|
F-55250L| .025–.250| 0.1–1.0| 250 (17.2)| 70 (21)| 316ss
F-55375L| 0.1–1.0| 0.4–4.0| 250 (17.2)| 70 (21)| 316ss
F-55376L| 0.2–2.0| 0.75–7.5| 250 (17.2)| 70 (21)| 316ss
F-55500L| 0.5–5.0| 2.0–20.0| 250 (17.2)| 70 (21)| 316ss
F-55750L| 1.0–10.0| 4.0–40.0| 250 (17.2)| 70 (21)| 316ss
F-55005L| 1.0–5.0| 3.0–18.0| 250 (17.2)| 70 (21)| PTFE
F-55010L| 1.0–10.0| 4.0–40.0| 250 (17.2)| 70 (21)| 316ss
F-55015L| 1.0–15.0| 5.0–60.0| 250 (17.2)| 70 (21)| 316ss
F-55200L| 2.0–20.0| 7.5–75.0| 250 (17.2)| 70 (21)| 316ss
Dimensions
Non-Adjustable Panel Mount
Adjustable Panel Mount
Non-adjustable Panel Mount
Pipe Size | A | B | C | D | E | Mount Hole |
---|---|---|---|---|---|---|
F-55250L (1/4”) | 7-1/4” | 2-1/2” | 1-1/16” | 5-5/8” | 1-1/4” | 9/16” |
F-55375L (3/8”) | 8-3/8” | 2-3/4” | 1-11/32” | 6-1/2” | 1-1/2” | 11/16” |
F-55376L (3/8”) | 8-3/8” | 2-3/4” | 1-11/32” | 6-1/2” | 1-1/2” | 11/16” |
F-55500L (1/2”) | 8-3/8” | 2-3/4” | 1-11/32” | 6-1/2” | 1-1/2” | 7/8” |
F-55750L (3/4”) | 9” | 3-3/4” | 1-1/2” | 6-1/2” | 1-3/4” | 1-1/16” |
F-55005L (1”) | 10-5/8” | 4” | 1-3/4” | 8” | 2” | 1-21/64” |
F-55010L (1”) | 10-5/8” | 4” | 1-3/4” | 8” | 2” | 1-21/64” |
F-55015L (1”) | 10-5/8” | 4” | 1-3/4” | 8” | 2” | 1-21/64” |
F-55200L (1”) | 10-5/8” | 4” | 1-3/4” | 8” | 2” | 1-21/64” |
Adjustable Panel Mount
Pipe Size| A| B| C| D| E| F| Mount
Hole
---|---|---|---|---|---|---|---
F-55250LA (1/4”)| 7-1/4”| 3-7/16”| 1”| 5-5/8”| 2-7/16”| 1-11/32”| 9/16”
F-55375LA (3/8”)| 8-3/8”| 4-1/2”| 1-11/32”| 6-1/2”| 3-1/4”| 1-3/4”|
11/16”
F-55376LA (3/8”)| 8-3/8”| 4-1/2”| 1-11/32”| 6-1/2”| 3-1/4”| 1-3/4”|
11/16”
F-55500LA (1/2”)| 8-3/8”| 4-1/2”| 1-11/32”| 6-1/2”| 3-1/4”| 1-3/4”| 7/8”
F-55750LA (3/4”)| 9”| 5-1/2”| 1-1/2”| 6-1/2”| 3-1/4”| 15/16”| 1-1/16”
F-55010LA (1”)| 10-5/8”| 5-1/4”| 1-3/4”| 8”| 3-1/2”| 17/16”| 1-21/64”
F-55015LA (1”)| 10-5/8”| 5-1/4”| 1-3/4”| 8”| 3-1/2”| 17/16”| 1-21/64”
F-55200LA (1”)| 10-5/8”| 5-1/4”| 1-3/4”| 8”| 3-1/2”| 17/16”| 1-21/64”
Installation
Installation Requirements
Before beginning installation, review your system flow and pressure to ensure your F-550 is proper for your system parameters.
Pipe Requirements:
- The flow meter is designed to work best when installed in a perfectly vertical pipe system.
- Ensure piping design prevents any stress on the meter fittings, minimizes vibration, and ensures location with direct visibility and minimal possibility of damage from forcible contact.
- Pipe should be clean and free of debris.
- Connecting pipe or panels should be uniform and not warped.
- Never allow the flow meter to support the weight of related pipe or tubing.
- On/Off solenoid valves, quick-opening valves, and pumps installed before the flow meter must be accompanied by flow regulating/throttling valves, or similar device, to prevent sudden and forcible flows through the flow meter which can cause damage to flow meter components.
- Do not exceed maximum working pressure (refer to Flow vs. Pressure chart)
Tools Required
- PTFE Tape
- Level
Note: Keep flow meter away from pipe glue fumes. Ensure all pipe glue is dry before installation flow meter. Pipe glue solvents will damage acrylic.
Example of acrylic flow meter exposed to glue fumes / solvents. Crazing of
acrylic
Warning Always wear protective clothing, face shield, safety glasses and gloves when working on your equipment. Use caution when lifting and mounting equipment. When in doubt, contact factory for assistance.
Installation Instructions
I nstallation Instructions
- **** The flow meter must be installed in a vertical plane to ensure accuracy. Prepare piping and components per standard piping installation.
- Use only PTFE tape on the threaded adapters. Pipe dope and glue will damage the meter body and fittings. Or install the flow meter after all glued fittings have dried and fumes have been purged from the system.
- Mounts and supports must be sturdy enough to support the plumbing and prevent vibration. Vibration and heavy loads will damage the meter.
- Avoid imposing a sudden burst of flow to the meter. Solenoid valves will damage the meter. Such a burst will cause the float to impact the top float stop with destructive force.
- The Maximum fluid temperature is dependent on the working pressure. High pressure and temperature will damage the meter. (see temperature VS pressure chart
Reading your F-430 Flow Meter
Avoid imposing a sudden burst of flow to the meter. Solenoid valves may cause
the float to impact the top float stop with destructive force. Always read
float at top.
Maintenance and Troubleshooting
- The F-550 Flow Meter requires very little maintenance. However, the flow meter may become clogged due to excess debris in the water. If this happens, simply shut of the process pump or valve and remove any pressure from the piping. If isolation valves are installed, shut these to minimize the amount of water that may leak. Remove the flow meter and clean any debris in the flow meter. Use mild soap and a soft bristle brush.
- Reinstall the flow meter and test. If the float was removed during the cleaning process, ensure the float was reinstalled properly. The float is wider at the top.
- If flow meter requires NIST Calibration Certification, please contact the factory for this service
Spare Parts
** I tem| Product Name| F-55250L| F-5537-5L F-5537-6L|
F-55500L| F-55750L| F-55005L F-55010L F-55015LF-55200L| Qty.
Re q.
---|---|---|---|---|---|---|---
1| Cap| 76001-080| 91001-032| 91001-032| 91001-033| 91001-160| 1
2| O-ring (cap)| 90003-072| 90003-081| 90003-081| 90003-082| 90003-079|
1
3| Guide wire| 90007-634| 90007-534| 90007-534| 90007-535| 90007-606| 1
4| Float| –| –| –| –| –| 1
5| Float stopbottom| –| –| –| –| –| 1
6*| Body| –| –| –| –| –| 1
7| O-ring(adapter)| 90003-072| 90003-034| 90003-034| 90003-081|
90003-110| 2
8| Adapter| 76001-079| 91001-141| 91001-142| 91001-143| 91001-144| 2
9| Lock nut| 91001-153| 91001-153| 91001-154| 91001-171| 91001-172| 2
Contact factory
Item| Product Name| F-55250LA| F-55375LA F-55376LA| F-55500LA| F-55750LA|
F-55005LA F-55010LA F-55015LAF-55200LA| Qty. Req.
---|---|---|---|---|---|---|---
1| Cap| 76001-080| 91001-032| 91001-032| 91001-033| 91001-160| 1
2| O-ring (cap)| 90003-072| 90003-081| 90003-081| 90003-082| 90003-079|
1
3| Guide wire| 76001-108| 76001-109| 76001-109| 76001-110| 76001-111| 1
4| Float| –| –| –| –| –| 1
5| Float stopbottom| –| –| –| –| –| 1
6| Body*| –| –| –| –| –| 1
7| O-ring(adapter)| 90003-072| 90003-034| 90003-034| 90003-081|
90003-110| 1
8| Adapter| 76001-079| 91001-141| 91001-142| 91001-143| 91001-144| 1
9| Lock nut| 91001-153| 91001-153| 91001-154| 91001-171| 91001-172| 2
10| Inlet Adapter| 76001-153| 91001-173| 91001-174| 91001-175|
91001-176| 1
11| O-ring,Inlet adapter| 90003-034| 90003-045| 90003-045| 90003-110|
90003-082| 1
12| O-ring,Valve stem| 90003-004| 90003-010| 90003-010| 90003-072|
90003-113| 1
13| Valve stem| 76001-082| 76001-088| 76001-088| 76001-094| 76001-104| 1
14| O-ring,Valve body| 90003-014| 90003-081| 90003-081| 90003-017|
90003-021| 1
15| Valve body| 76001-081| 76001-086| 76001-086| 76001-093| 76001-103| 1
16| Retainer cap| 76001-084| 76001-090| 76001-090| 76001-095| 76001-106|
1
17| Knob valve| 76001-083| 76001-089| 76001-089| 76001-097| 76001-105| 1
18| Set screw| 90011-056| 90011-117| 90011-117| 90011-117| 90011-117| 1
F-550
WARRANTY
LIMITED WARRANTY
Your new Flow Meter is a quality product and is warrantied for 12 months from
date of purchase (proof of purchase is required). The unit will be repaired or
replaced at our discretion. Failure must have occurred due to defect in
material or workmanship and not as a result of operation of the product other
than in normal operation as defined in the manual. Warranty status is
determined by the serial label and the sales invoice or receipt. The serial
label must be on the unit and legible. The warranty status of the unit will be
verified by Blue-White or a factory authorized service center.
WHAT IS NOT COVERED
- Wear items.
- Unit removal, or re-installation, and any related labor charge.
- Freight to the factory, or service center.
- Units that have been tampered with, or in pieces.
- Damage to the unit that results from misuse, carelessness such as chemical spills on the enclosure, abuse, lack of maintenance, or alteration which is out of our control.
- Units damaged by faulty wiring, power surges or acts of nature.
PROCEDURE FOR IN WARRANTY REPAIR
Contact the factory to obtain a RMA (Return Material Authorization) number.
Carefully pack the unit to be repaired. Please enclose a brief description of
the problem as well as the original invoice or sales receipt, or copy showing
the date of purchase. Prepay all shipping costs. COD shipments will not be
accepted. Warranty service must be performed by the factory or an authorized
service center. Damage caused by improper packaging is the responsibility of
the sender. When In-Warranty repair or replacement is completed, the factory
pays for return shipping to the dealer or customer.
PRODUCT USE WARNING
Blue-White products are manufactured to meet the highest quality standards in
the industry. Each product
instruction manual includes a description of the associated product warranty
and provides the user with important safety information. Purchasers,
installers, and operators of Blue-White products should take the time to
inform themselves about the safe operation of these products. In addition,
Customers are expected to do their own due diligence regarding which products
and materials are best suited for their intended applications. Blue-White is
pleased to assist in this effort but does not guarantee the suitability of any
particular product for any specific application as Blue-White does not have
the same degree of familiarity with the application that the customer/end user
has. While Blue-White will honor all of its product warranties according to
their terms and conditions, Blue-White shall only be obligated to repair or
replace its defective parts or products in accordance with the associated
product warranties. BLUE-WHITE SHALL NOT BE LIABLE EITHER IN TORT OR
IN CONTRACT FOR ANY LOSS OR DAMAGE WHETHER DIRECT, INDIRECT, INCIDENTAL, OR
CONSEQUENTIAL, ARISING OUT OF OR RELATED TO THE FAILURE OF ANY OF ITS PARTS OR
PRODUCTS OR OF THEIR NONSUITABILITY FOR A GIVEN PURPOSE OR APPLICATION.
CHEMICAL RESISTANCE WARNING
Blue-White offers a wide variety of wetted parts. Purchasers, installers, and
operators of Blue-White products must be well informed and aware of the
precautions to be taken when injecting or measuring various chemicals,
especially those considered to be irritants, contaminants or hazardous.
Customers are expected to do their own due diligence regarding which products
and materials are best suited for their applications, particularly as it may
relate to the potential effects of certain chemicals on Blue-White products
and the potential for adverse chemical interactions. Blue-White tests its
products with water only. The chemical resistance information included in this
instruction manual was supplied to Blue-White by reputable sources, but Blue-
White is not able to vouch for the accuracy or completeness thereof. While
Blue-White will honor all of its product warranties according to their terms
and conditions, Blue-White shall only be obligated to repair or replace its
defective parts or products in accordance with the associated product
warranties. BLUE-WHITE SHALL NOT BE LIABLE EITHER IN TORT OR IN CONTRACT FOR
ANY LOSS OR DAMAGE, WHETHER DIRECT, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL,
ARISING OUT OF OR RELATED TO THE USE OF CHEMICALS IN CONNECTION WITH ANY BLUE-
WHITE PRODUCTS.
Users of electrical and electronic equipment (EEE) with the WEEE marking per Annex IV of the WEEE Directive must not dispose of end of life EEE as unsorted municipal waste, but use the collection framework available to them for the retum, recycle, recovery of WEEE and minimize any potential effects of EEE on the environment and human health due to the presence of hazardous substances. The WEEE marking applies only to countries within the European Union (EU) and Norway. Appliances are labeled in accordance with European Directive 2002/96/EC Contact your local waste recovery agency for a Designated Collection Facility in your area.
+1 (714) 893 – 8529
customerservice@blue-white.com
5300 Business Drive
Huntington Beach, CA 92649
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>