DT-3015N Portable Xenon Stroboscope Instruction Manual
- June 16, 2024
- DT
Table of Contents
- DT-3015N Portable Xenon Stroboscope
- Safety Precautions
- 1. Overview of this product
- 2. Before use
- Specifications
- Product Usage Instructions
- Safety Precautions
- Q: Can I use the product in a flammable environment?
- Q: How should I clean the unit?
- Q: What should I do if the unit heats up during prolonged
DT-3015N Portable Xenon Stroboscope
INSTRUCTION MANUAL
Safety Precautions
Be sure to read the entire instruction manual thoroughly before initial set- up, operation and maintenance.
The instruction manual provides two grades of safety warnings: “Danger” and “Caution”.
Each of them is an important description related to safety. Be sure to observe.
DANGER : This indicates the possibility of fire, severe injury, and even death if a user disregards the instruction and operates the unit improperly.
CAUTION : This indicates the possibility of minor injury or property damage if a user operates the unit improperly. However, depending on the circumstances, there is still the possibility that severe injury may result. Be sure to observe.
We categorize the type of those precautions using the following symbols throughout the manual.
A prohibited action you must not do.
A forced action you must always do.
1. Overview of this product
A stroboscope tachometer is a measurement instrument to measure the speed (cycle) of rotating objects that rotate at a constant speed, or moving objects that repeatedly operate at a constant cycle. When the rotation (motion) cycle matches with the flash cycle while the strobe flash is periodically applied on a rotating or moving object, the rotating (moving) object image appears to stand still. This stroboscope tachometer is the non-contact type, and can be used to read the object frequency when such a still image appears. Also, a stroboscope can be used to make images of rotating or moving objects stand still or that slightly move in order to observe their appearance in slow -motion.
Main features
- Wide flash range – 30 to 35000 FPM(flashes per minute)
- Emission in synchronization with the external trigger pulse
- Phase Shift function.(PHA mode)
- Rechargeable
2. Before use
2.1 Checking the supplied items
Check that the five items below are supplied.
2.2 Charging
Be sure to charge the battery before initial use.
Before charging, be sure to check that the power is turned OFF.
Please check the state of the power switch OFF with the picture in “3.1 Main
unit” on page 6.
2.2.1 Charging method
Turn the power OFF, and connect the AC adapter’s connector to the main unit.
Next insert the AC adapter’s AC plug into an outlet. The battery lamp lights
up, and charging starts. Charging is complete in approximately 2 hours, and
then the charging lamp goes off.
The battery lamp( B-CHG ) | State |
---|---|
Off | AC adapter not connected, or Charging complete |
Flash | Charging error (Stop charging) |
Continuous light | Charging |
Be sure to use the supplied AC adapter.
Use at the ambient temperature 15°C or higher.
(If the ambient temperature is lower than 15°C, Charging may not be possible.)
*If the battery overheats, a charging error may occur.
In this case, let the internal temperature decrease and then reconnect the AC
adapter.
Note 1) When the power is turned ON during charging, charging is canceled and emission starts. When the unit uses the AC adapter as its power source, it is not charged. (Note that emission and charging cannot be performed simultaneously.)
Note 2) Though the unit can be charged while battery power still remains, charging will still end when fully charged.
It is recommended, however, that the unit be charged when the battery has run out (“LLLLLL” is indicated on the display) to avoid reducing the battery life itself.
Note 3) If the unit is charged immediately after a long emission time at a high ambient temperature,the temperature detection circuit will turn on, and the charging lamp may not light up. In that case, disconnect the AC adapter, and let the unit cool. When the product temperature has decreased, connect the AC adapter again and start charging.
Note 4) Using in a high temperature environment may shorten the battery life.
Specifications
-
Safety Precautions: Danger and Caution
-
Usage: Emission in synchronization with external trigger pulse,
Phase Shift function, Rechargeable -
Operating Conditions: 45-85% relative humidity
Product Usage Instructions
Safety Precautions
Before using the product, it is crucial to adhere to the safety
precautions outlined in the manual. Failure to follow these
precautions may result in injury or damage to the unit.
Danger Warnings
-
Never use in flammable environments as it may result in
fire. -
Avoid looking directly into the light source to prevent eye
injury. -
Do not subject the unit to strong impact or dropping to avoid
abnormal operation.
Caution Warnings
-
Avoid exposure to water, direct sunlight, condensation, dust,
dirt, salt, iron, oil, chemicals, corrosive or combustible
gases. -
Do not alter, modify, or dispose of the unit improperly to
prevent injury due to abnormal operation. -
Clean the unit with a soft dry cloth or diluted neutral
detergent. Avoid using volatile chemicals like benzine, thinner, or
alcohol.
Operating Conditions
Operate the unit within a relative humidity range of 45-85% to
prevent malfunction.
Charging
Before initial use, ensure the battery is fully charged. Follow
these steps for charging:
- Turn the power OFF.
- Connect the AC adapter’s connector to the main unit.
- Insert the AC adapter’s plug into an outlet. The battery lamp
will light up to indicate charging has started.
FAQ
Q: Can I use the product in a flammable environment?
A: No, using the product in a flammable environment may result
in a fire hazard. It is strictly prohibited.
Q: How should I clean the unit?
A: Clean the unit with a soft dry cloth or a cloth dampened with
diluted neutral detergent. Avoid using volatile chemicals like
benzine, thinner, or alcohol.
Q: What should I do if the unit heats up during prolonged
use?
A: To prevent overheating, fix the strobe using a tripod or
similar support. Avoid direct skin contact with the unit to prevent
burns and potential malfunctions.
Safety Precautions
Be sure to observe
Be sure to read the entire instruction manual thoroughly before initial set-
up, operation and maintenance.
The instruction manual provides two grades of safety warnings: “Danger” and
“Caution”. Each of them is an important description related to safety. Be sure
to observe.
This indicates the possibility of fire, severe injury, and even death if a
user disregards the instruction
Danger and operates the unit improperly.
This indicates the possibility of minor injury or property damage if a user
operates the unit improperly.
Caution However, depending on the circumstances, there is still the
possibility that severe injury may result.
Be sure to observe.
We categorize the type of those precautions using the following symbols
throughout the manual.
A prohibited action you must not do.
A forced action you must always do.
Never use in flammable environments.
May result in fire.
Danger
Never look directly into the light source.
May result in eye injury.
Do not apply strong impact to the unit, or drop it.
Failure to follow this could result in abnormal operation.
Caution
Avoid the followings.
Water, direct sunlight, condensation, dust, dirt, salt, iron, oil, chemicals,
corrosive and/or combustible gases.
Do not alter, modify or dispose improperly.
Failure to follow this could result in injury due to abnormal operation.
Wipe clean the unit with a soft dry cloth if it gets dirty. Or immerse a cloth
in water diluted neutral detergent, wring it, and wipe clean the unit with it.
Do not use any volatile chemicals, such as benzine, thinner, or alcohol.
Since continuously emitting light for long time causes the unit’s housing to
heat up, fix the strobe using a tripod, etc. (Avoid direct skin contact with
the unit, such as holding it by hand).
Failure to follow this could result in mild burns.
Operate within 0-35°C(32-95°F)
Failure to follow this could result in malfunction.
Operate within 45-85%RH
Failure to follow this could result in malfunction.
1
Contents
1 Overview of this product ………………………………………………………………. 3 2 Before
use…………………………………………………………………………. 3
2.1 Checking the supplied items …………………………………………………………… 3 2.2
Charging………………………………………………………………………….. 4
2.2.1 Charging method………………………………………………………………… 4 2.2.2 Indication of low
battery voltage ( LOWBAT ) ……………………………………………. 5 3 Part names and functions
……………………………………………………………. 5 3.1 Main unit………………………………………………………………………….. 5 3.2
Operation Panel …………………………………………………………………….. 6 3.3 Display
…………………………………………………………………………… 7 3.3.1 Part names and function instructions
…………………………………………………. 7 4 Function and operations………………………………………………………………. 8
4.1 Power ON / OFF …………………………………………………………………….. 8 4.1.1 LAMP ON / OFF
…………………………………………………………………. 9 4.1.2 Flash timer …………………………………………………………………….. 9
4.1.3 Heat Lamp Indicator ……………………………………………………………… 9 4.2 Emission mode and
settings ………………………………………………………….. 10 4.2.1 How to switch INT and EXT
……………………………………………………….. 10 4.3 Internal oscillation
emission…………………………………………………………… 11 4.3.1 FPM mode setting
(INT)…………………………………………………………… 11 4.3.2 PHA mode (INT) …………………………………………………………………
13 4.4 External synchronous emission ……………………………………………………….. 14 4.4.1 FPM mode
setting (EXT) ………………………………………………………….. 15 4.4.2 PHA mode setting (EXT)
………………………………………………………….. 16 4.5 Function mode…………………………………………………………………….. 17
4.5.1 The setting items ……………………………………………………………….. 17 4.5.2 Instructions about
function mode …………………………………………………… 18 4.6 Saving function
……………………………………………………………………. 24 4.6.1 Saving the setting values
…………………………………………………………. 24 4.6.2 How to save the setting
values……………………………………………………… 25 4.6.3 Initialize ……………………………………………………………………….
26 4.6.4 Initializing the setting values……………………………………………………….. 27 4.7 Reading
memory function …………………………………………………………… 28 4.7.1 Reading values
………………………………………………………………… 28 4.7.2 How to read the
memory………………………………………………………….. 28 4.8 External signal I/O connector
specifications and Pin assignment ……………………………….. 29 4.9 External pulse
input…………………………………………………………………. 30 4.10 External pulse output
………………………………………………………………. 30 4.11 Lamp replacement………………………………………………………………….
31 4.12 Battery replacement ……………………………………………………………….. 32 5
Specifications……………………………………………………………………… 33 5.1 Specifications
list…………………………………………………………………… 33 5.2 External dimensions
………………………………………………………………… 34 6 Troubleshooting……………………………………………………………………. 35
2
1 Overview of this product A stroboscope tachometer is a measurement
instrument to measure the speed (cycle) of rotating objects that rotate at a
constant speed, or moving objects that repeatedly operate at a constant cycle.
When the rotation (motion) cycle matches with the flash cycle while the strobe
flash is periodically applied on a rotating or moving object, the rotating
(moving) object image appears to stand still. This stroboscope tachometer is
the non-contact type, and can be used to read the object frequency when such a
still image appears. Also, a stroboscope can be used to make images of
rotating or moving objects stand still or that slightly move in order to
observe their appearance in slow -motion.
Main features
Wide flash range 30 to 35000 FPM(flashes per minute)
Emission in synchronization with the external trigger pulse
Phase Shift function.(PHA mode)
Rechargeable
2 Before use 2.1 Checking the supplied items
Check that the five items below are supplied.
One(1) DT-3015N
Main unit
One(1) External signal I/O connector (8 pins) RM15WTPZ-8S(71)
Accessories
Instruction Manual Warranty
One(1) Dedicated AC adapter (3m)
One(1) This document One(1) document
3
2.2 Charging
Be sure to charge the battery before initial use. Before charging, be sure to check that the power is turned OFF. Please check the state of the power switch OFF with the picture in “3.1 Main unit” on page 6.
2.2.1 Charging method
Turn the power OFF, and connect the AC adapter’s connector to the main unit.
Next insert the AC adapter’s AC plug into an outlet. The battery lamp lights up, and charging starts.
Charging is complete in approximately 2 hours, and then the charging lamp goes off.
The battery lamp( B-CHG )
State
Off Flash Continuous light
AC adapter not connected, or Charging complete Charging error (Stop charging) Charging
Be sure to use the supplied AC adapter. Use at the ambient temperature 15°C or higher.
(If the ambient temperature is lower than 15°C, Charging may not be possible.)
*If the battery overheats, a charging error may occur.
In this case, let the internal temperature decrease and then reconnect the AC adapter.
Remove the connector cover, and connect the AC adapter
Note 1) When the power is turned ON during charging, charging is canceled and
emission starts. When the unit uses the AC adapter as its power source, it is
not charged. (Note that emission and charging cannot be performed
simultaneously.)
Note 2) Though the unit can be charged while battery power still remains,
charging will still end when fully charged. It is recommended, however, that
the unit be charged when the battery has run out (“LLLLLL” is indicated on the
display) to avoid reducing the battery life itself.
Note 3) If the unit is charged immediately after a long emission time at a
high ambient temperature,the temperature detection circuit will turn on, and
the charging lamp may not light up. In that case, disconnect the AC adapter,
and let the unit cool. When the product temperature has decreased, connect the
AC adapter again and start charging.
Note 4) Using in a high temperature environment may shorten the battery life.
Never charge the unit under the following conditions.
Using another AC adapter than supplied Charging near flammable materials
Adapter of
another device
Oil
4
2.2.2 Indication of low battery voltage ( LOWBAT )
If the remaining battery decreases below the specified amount, the battery voltage is lowered and the number display is changed to indicate the current setting value and “LLLLLL” alternately. Emission can be performed even when the battery voltage is low. If the remaining battery decreases further, the number display indicates only “LLLLLL”, which provides the emission stop status. Press the power switch, and shut off the power immediately. In either case, charge the unit according to the procedure in “2.2.1 Charging method” .
Operation When the power is ON
Display Indicates alternately
Press the power switch, immediately
The normal display (this display differs depending on each mode)
Indication of low battery voltage The same LED as the left lights up
The remaining battery decrease further
3 Part names and functions 3.1 Main unit
Xenon Lamp
Indication of emission stop
Operation Panel
*This figure shows when the power is off Power ON
Power OFF Power Switch
Power Cable 5
3.2 Operation Panel
o. Key Power Switch Dial (LAMP POWER)
MODE LOAD(SAVE)
SIG
UNIT ×2 ÷2
Function Instructions
Turns the power ON / OFF
Turn this clockwise or counterclockwise to change the emission frequency and
some setting value. Press and hold the center of this to turn the lamp ON /
OFF.
Press this to switch the LED display as follows Flashes Per Minute mode (FPM
mode) Phase Shift mode (PHA mode DEG = delay angle ) Phase Shift mode ( PHA
mode ms = delay time ) FPM mode …
Press this to load the saved settings. Also, Press and hold this to save the
current display and mode settings.
Select Internal/ External/ Parameter mode. Internal oscillation emission
(referred to as “INT” )Flash by internal signal External synchronous emission
(referred to as “EXT”)Synchronized flash with external signal Each press of
this switches the delay type as following, delay angle(DEG) delay time(ms) DEG
ms
Press this while on INT to double the current set emission count (frequency).
(If the double count is over than the flash range, nothing will change on the
display.)
Press this while on INT to halve the current set emission count (frequency).
(If the half count is lower than the flash range, nothing will change on the
display.)
6
3.3 Display 3.3.1 Part names and function instructions
Battery lamp
SELECT number display
Numerical display
Delay type display
B-CHG
1
Heat lamp
DEG HEAT ms
UNIT
×2
2 3 4 INT EXT
5 FPM PHA
SIG LOAD SAVE
Mode display Emission setting display
÷2
MODE
o. Key Emission setting display
Mode display SELECT number display Numerical display Battery lamp
Display
INT EXT FPM PHA
1 2 3 4 5
B-CHG
Instructions Each press of “SIG” switches emission settings. INT Internal
oscillation emission EXTExternal synchronous emission
Each press of “MODE” switches flashing mode and delay type. FPMFlashes per
minutes mode / PHAPhase mode
This indicates the memories for saving. (called SELECT number) Each press of
“LOAD” switches SELECT number (1 to 5) Standard Operation : indicates flash
rate. PHA Mode : Angle and Time are displayed. Function Mode : Each setting
value is indicated.
Lights up while charging, and goes off when charging is complete. Flashes if
charging error.
Delay type display
DEG ms
Each press of “MODE” switches flashing mode and delay type. DEGdelay angle / msdelay time ( millisecond)
Heat lamp
HEAT
When the ambient temperature of the Xenon lamp rises above a certain level, this lamp indicator will flash.
7
4 Function and operations 4.1 Power ON / OFF
Press the power switch when the power is OFF to turn the power ON. When power is turned ON, the model is indicated, followed by internal oscillation emission or external synchronous emission.
Operation When the power is OFF
Display
Press the power switch
Displayed “dt” for 0.5 sec.
Displayed the model for a sec.
When turning the power on for the first time, “EXT FPM mode” is displayed.
From the second time on, the same as
when the power is previously turned OFF is displayed.
Press the power switch when the power is ON to turn the power OFF, then the indication goes off.
Operation
Display
When the power is ON
Press the power switch
Indicated letters go off
*When the power is turned ON, the following indication may be displayed.
The following indication is an error message when the reading previously used setting value fails.
Press the “MODE” to return to normal emission screen from the error message display.
Operation
Display
Press the “× 2”
Error message
Returns to the normal emission screen
8
4.1.1 LAMP ON / OFF
Press and hold the center of the dial to turn ON the LAMP when emission setting is displayed on the screen. Press and hold again while flashing to turn OFF the LAMP.
Operation
Display
Press and hold the dial to flash. Press and hold again to stop flashing.
Press and hold *When turning the power switch OFF while flashing, it will be
turned ON flashing at the next time.
4.1.2 Flash timer
When the Flash timer is set to a value other than 0 [minutes] in Function mode 3,
the flashing will automatically cease after set time from the start of the light emission.
When the flashing stops by this flash timer, the display indicates “OFF”. Press and hold the dial to return to the previous display.
Operation
Display
Press and hold
After set time from the start
Return to the previous display
4.1.3 Heat Lamp Indicator
When the ambient temperature of the Xenon lamp rises above a certain level,
the Heat Lamp Indicator will flash. If the temperature rises more, the Heat
Lamp Indicator will remain on and the light emission will cease. *The
temperature tends to rise more quickly the higher the emission frequency is
set.
The ambient temperature of the Xenon lamp Less than 65°C (149°F) Over 65°C (149°F) Over 70°C (158°F)
HEAT Off Flash On
Light emission Keep flashing Keep flashing Stop flashing
9
4.2 Emission mode and settings
Emission mode
Instructions
Internal oscillation emission ( “INT” )
Flash at the set frequency.
External synchronous emission ( ” EXT” ) Flash in synchronization with the external trigger pulse.
Function mode
Configure the settings.
4.2.1 How to switch INT and EXT Each press of “SIG” switches “INT” and “EXT”.
Power OFF Displays the model
Power ON
INT
FPM mode
Turn the dial
Increase / decrease the emission count (frequency)
EXT FPM mode Display external input pulse frequency
PHA mode
Turn the dial
Increase / decrease the delay angle or the delay time
PHA mode
Turn the dial
Increase / decrease the delay angle or the delay time
Confirm saving Return to the previous display
Read the saved value. (SELECT number)
Saving function
Press and hold
Confirm saving Return to the previous display
10
4.3 Internal oscillation emission
On “INT”, the Xenon lamp flashes at the displayed emission count. “INT” has the following 2 mode settings.
Mode settings Instructions
FPM mode
Set the emission count per minute [FPM] (flashes per minute)
PHA mode
Shift the timing of flash. The phase can be changed by a degree or a
millisecond. ( One cycle is 360°)
Display of INT
SELECT number
Mode display
Display “INT”
The emission count (frequency)
4.3.1 FPM mode setting (INT)
The emission count (frequency) can be set in FPM mode. Turn the dial in a CW direction to increase the emission count, and in a CCW direction to decrease it. (Turn the dial fast to change the setting value greatly, and slowly to change it slightly.)
Operation
Display CW
Increase / decrease The emission count
(frequency)
Decrease Minimum value is 30
CCW
Increase Maximum value is 35000
Press “SIG” to
switch to the “EXT”
INT
EXT
11
Press “×2″to double the current set the emission count
Press “÷2″to halve the current set the emission count
Press “MODE” to switch to PHA mode
FPM mode
If it is over than the flash range, nothing will change on the display.
If it is lower than the flash range, nothing will change on the display.
PHA mode
12
4.3.2 PHA mode (INT)
When the rotation (motion) cycle of a measured object matches with the strobe flash cycle, the measured object appears to stand still. Use the PHA mode in order to adjust the stop angle (position). The phase can be changed by 1° using dial within the range between 1° and 359° in the PHA mode. Press “UNIT” to switch to PHA mode (ms). (The phase can be changed by 0.01 [ms] within the range between 0.00 and 999.99 [ms] and by 0.1 [ms] within the range between 1000.0 and 1994.4 [ms].) A time longer than the light emission cycle cannot be set. (PHA mode ms) Press “MODE” in PHA mode to switch to FPM mode.
Operation
Display
Decrease delay angle
CW Increase delay angle
Increase / decrease Delay angle / time
PHA mode (DEG)
CCW
Each press of “UNIT” switches DEG to ms
Decrease delay time
CW Increase delay time
PHA mode (ms)
CCW
Press “MODE” to switch to FPM mode
PHA mode
FPM mode
13
4.4 External synchronous emission
External synchronous emission is the function to emit a strobe flash in synchronization with an external trigger pulse input. You can set which edge of the external trigger pulse triggers emission, the rising edge or falling edge. A timing (delay) from the external trigger pulse input with the strobe flash emission can be optionally set using time and angle. EXT has the following 2 mode settings.
Mode setting FPM mode
Instruction
Display the measured value of the emission count (frequency). The unit is FPM (flashes per minute)
PHA mode Shift the timing of flash. The phase can be changed by a degree or a millisecond. ( One cycle is 360°)
Measured value of the emission count (frequency)
SELECT number
Mode display
Display “EXT”
*If the external trigger pulse cycle is beyond the specifications range, the following letters are indicated. [Without the delay angle / time setting] Measurable range 27 to 35020 [fpm]
When the external input pulse frequency goes below 27 [fpm], underlines are displayed.
When the external input pulse frequency goes beyond 35020 [fpm], overlines are displayed.
[With the delay angle setting] Measurable range 27 to 35020 [fpm]
When the external input pulse frequency goes below 27 [fpm], When the external input pulse frequency goes beyond 35020 [fpm],
an underline on the first, left two digits is displayed.
an overline on the first, left two digits is displayed.
[With the delay time setting] Measurable range 27 to 35020 [fpm]
When the external input pulse frequency goes below 27 [fpm], an underline on the first, left two digits is displayed.
When the external input pulse frequency goes beyond 35020 [fpm], an overline on the first, left two digits is displayed.
14
4.4.1 FPM mode setting (EXT) Operation
Display
Press “MODE” to switch to PHA mode
Press “SIG” to switch to INT
EXT FPM mode
PHA mode Display in the previous unit
(DEG or ms)
INT FPM mode
15
4.4.2 PHA mode setting (EXT)
Delay emission can be set within the input signal range between 27 and 35020
[fpm]. In the PHA mode, the phase from the external trigger pulse entry to
strobe flash emission can be changed by 1° using the dial within the range
between 1° and 359°. Press “UNIT” to toggle between PHA mode (DEG) and PHA
mode (ms). The phase can be changed by 0.01 [ms] within the range between 0.00
and 999.99 [ms], by 0.1 [ms] within the range between 1000.0 and 2216.0 [ms].
A time longer than the light emission cycle cannot be set. (PHA mode ms) Press
“MODE” in PHA mode to switch to FPM mode.
Operation
Decrease delay angle
Display CW
Increase delay angle
Increase / decrease Delay angle / time
CCW
Press “UNIT” to toggle between DEG and ms
Decrease delay angle
CW Increase delay angle
CCW
Press “MODE” to switch to FPM mode
PHA mode
FPM mode
16
4.5 Function mode
Turn the power ON while pressing the “MODE” to enter the function mode. During the function mode, turn the dial (CW/CCW) to change the settings. And press “MODE” to save the setting and move to the next setting item. (F1F2F3F4F5F6 end ) If the power is turned OFF in the middle of the function mode, the changed value will not be saved in the memory.
4.5.1 The setting items
The setting items in the following table can be configured in the function mode.
F1
Set to add decimal point (for INT and EXT FPM mode)
Add decimal point
No decimal point
F2
Trigger edge setting (for EXT)
Flash in the rising edge.
Flash in the falling edge.
F3
Auto emission stop time setting
Emission is performed continuously
Stop emission automatically when no operation is performed for a certain period of time
and display “OFF”.
The available setting time is within the range between 0 [min] and 120 [min],
which can be changed by 1 [ms]
F4
Choose the input circuit
Open collector input
Voltage pulse input
F5
Measurement range setting (for INT)
The maximum value of the emission count can be set within the range between 30 and 35000 [fpm] (The value is limited to the maximum.)
F6
Set the pulse width of the synchronous output signal to 400 [µs] or 150 [µs]
The pulse width is 400 [µs]
The pulse width is 150 [µs]
17
4.5.2 Instructions about function mode
How to move to function mode. Turn the power ON while pressing the “MODE” to enter the function mode.
Operation
Display
The power switch ON while pressing “MODE”
Turn the power ON
Displays “Func” for two seconds
while pressing “MODE” And move to function mode 1
In function mode 1, you can set weather to add decimal point. (Not reflected on EXT)
Operation
Display Indicates alternately
Change the setting value
Turn the dial and it will stop indicating alternately.
CCW
*Previous setting value The default setting is “0.0
CW
CCW0 ( add decimal point)
Go to function mode 2
CW0.0 (no decimal point) Go to function mode 2
18
In function mode 2, in which the edge of the external pulse triggers emission, the rising edge or falling edge, can be set (for EXT).
Operation
Display Indicates alternately
Change the setting value
Turn the dial and it will stop indicating alternately.
CCW
*Previous setting value The default setting is “L-H”
CW
CCW: L-H (Rising edge)
CW: H-L (Falling edge)
Go to function mode 3
*This setting changes the timing of emission as shown in the following figure. The timing of emission
External trigger pulse Flash at the rising edge Flash at the falling edge
Go to function mode 3
19
In function mode 3, auto emission stop time can be changed.
Operation
Display Indicates alternately
Change the setting value
Turn the dial and it will stop indicating alternately.
CCW
*Previous setting vale The default setting is “0”
CW
CW+1 CCW-1 The timer can be set between 0 [min] and 120 [min]
Go to function mode 4
Go to function mode 4
20
In function mode 4, the input circuit setting can be changed. Operation
Display Indicates alternately
Change the setting value
Turn the dial and it will stop indicating alternately.
CCW
*Previous setting vale The default setting is “OFF”
CW
CCWOFF Voltage pulse input (Pull down the external input circuit)
CWOn Open collector input (Pull up the external input circuit)
Go to function mode 5
Go to function mode 5
21
In function mode 5, the measurement range on INT can be set.
Operation
Display Indicates alternately
Change the setting value
Turn the dial and it will stop indicating alternately.
CCW
*Previous setting vale The default setting is “35000”
CW
Go to function mode 6
Increase / decrease by 1 [fpm] Available setting range: 30 to 35000 [fpm] Go to function mode 6
22
In function mode 6, the pulse width of the synchronous output signal can be set to 400 sec or 150 sec.
Operation
Display Indicates alternately
Change the setting value
Turn the dial and it will stop indicating alternately.
CCW
*Previous setting vale The default setting is “400”
CW
150 sec
400 sec
Save the changes and go to the home screen
Go to the home screen
23
4.6 Saving function
This unit can save the setting values in two ways. Power OFF saveWhile using
the unit while on INT and EXT, turn the power OFF to save the setting value to
that which it was before turning the power OFF. When the power is turned ON
again, operation starts from the previous setting value.
Memory saveThe setting values on the screen can be saved as the SELECT number.
Since the SELECT number have five memories, the setting can be saved as five
patterns.
SELECT number display
4.6.1 Saving the setting values The setting values are saved as the following figure.
save ×cannot save
Emission setting(INT / EXT) Mode setting(FPM / PHA)
The emission count(FPM) INT Delay angle(PHA)
Delay time(PHA) Delay angle(PHA) EXT Delay time(PHA)
Memory save
INT
EXT
×
×
×
×
×
Power OFF save
Power OFF during INT
×
×
Power OFF during EXT
× × ×
1
2
1 When you press “ZERO” in Delay angle mode, the value (delay angle) on the display is different from the actual value. In that case, the value on the display will be saved.
2 When you turn the power OFF to save, the calculated delay time based on delay angle will be saved.
24
4.6.2 How to save the setting values
The setting values can be saved by the operations as followings.
Operation
Display
Press and hold “LOAD”
Press and hold “LOAD”
Press “×2” to cancel the saving function.
“SAvE”and SELECT number (the default is “1”) display alternately every 0.5
seconds. Turn the dial and change SELECT number to save.
CW +1 CCW-1
When”1″displays on the screen, turn the dial CW to display “5”. When”5″displays on the screen, turn the dial CCW to display “1”.
Press “LOAD” and confirm the saving function. After the saving, go to the previous screen.
Press “×2” and cancel the saving function. Go to the previous screen without saving.
25
4.6.3 Initialize The memory can be erased by initializing the saving function. Operation
Display
The power switch ON while pressing “÷2” and “MODE”
The power switch ON while pressing “÷2” and “MODE”
Press “×2” and Confirm initialization
Displays for a second
Press “×2” and confirm initialization
Displays for a second And go to the home screen
26
4.6.4 Initializing the setting values
The current settings will be erased and replaced as follows. (Including the function mode.)
The initial emission setting
The initial mode
Default SELECT number
FPM
INT
PHASE(DEG)
PHASE(ms)
FPM
EXT
PHASE(DEG)
PHASE(ms)
The initial setting EXT FPM 0 1 1500 0 0 0 0 0
display *2
F1
0.0
F2
L-H
F3
0
F4
OFF
F5
35000
F6
400
1 Memory (SELECT number) is not loaed 2 On EXT, underlines are displayed until the external signal inputs occur.
27
4.7 Reading memory function 4.7.1 Reading values
Press “LOAD” and read the saving values as the figure on 4.6.1
4.7.2 How to read the memory
Operation
OFF (SELECT0)
Display
SELECT1
Each press of “LOAD” switches SELECT numbers.
OFF12345
SELECT2 SELECT3
Set the setting values which have been saved in SELECT number on display.
SELECT4 SELECT5
28
4.8 External signal I/O connector specifications and Pin assignment
RM15WTRZB-8P(71) (Hirose Electric Group)
Pin number Signal name Remarks
1
+19V
Power input +
2
-19V
Power input –
3
12V
Power supply+ (DC12V) for sensor
4
OUT
External pulse output
5
IN
External pulse input
6
GND
Power supply- (DC12V) for sensor External pulse output common External pulse input common
7
NC
–
8
FG
Earth
1 2
8
7
6
3 5
4
Pin assignment
29
4.9 External pulse input Connect the unit to external devices (sensors, etc.) to allow the strobe to emit light using the pulse signal from the devices on external synchronous emission.
Available input frequency
Available input signal
Available input pulse width Input impedance
Available measurement range Available delay emission range
Hi 2.5 to 12 V Lo 0 to 0.5 V 50 s or more (edge trigger) 10 k ( at voltage
input )
27 to 35020 fpm0.45 to 583.7 Hz 27 to 35020 fpm0.45 to 583.7 Hz
[Input circuit]
Input/output Signal Jack
5pin 6pin
12V
12V
4.7k 10k
10k
Switch this in function mode 4 On open collector input OFF voltage pulse input
Go to the inputcircuit Insulated with phototransistor
4.10 External pulse output
Outputs pulses to external devices simultaneously with light emission.
Output circuit spec 12 V output Output pulse width about 400 s
[Output circuit] Input/output Signal Jack
4pin
4.7k
*Output square wave as following figure.
Battery type is not grounded
6.8k Come from the output circuit Insultaed with phototransister Battery type
is not grounded
Flash timing Synchronous output
About 400 sec
Pulse width can be changed in function mode 6 (150 sec)
30
4.11 Lamp replacement
The life of a xenon lamp is about 1200 hours when it is emitted at 1500 fpm
each time. Although rotation speed is displayed, no flash is emitted. When the
flash is intermittently emitted, this indicates the lamp must be replaced. Be
sure to replace the lamp with the following steps. Be sure to use the
specified lamp. Please contact us or the retailer where you purchased this
product if you need it.
Be sure to turn the power OFF. May result in electric shock
The lamp is hot and may cause burns. After emission stops, let the stroboscope sit for 30 minutes or longer. Be sure stroboscope is cool to the touch before replacing the lamp.
Remove the protection window by loosening the 4 screws. Insert a thin screw
driver into a hole of the protection window and pull out.
Remove the reflector and Insert the lamp insertion holder as far as it will
go. Then turn it CW two or three times and pull out straight with lamp.
Be sure to use a lamp insertion holder. Pulling out lamp by hand may cause
injury.
Insert new lamp into the lamp insertion holder with seeig the figure below.
And press it to the socket in the proper direction.
Turn the lamp insertion holder CCW and pull out. And put carefully the
reflector on the case properly.
Put the protection window back with 4 screws. *In order to maintain
protection, be sure to fix the protective window. (Tightening torque: 0.3N-m
or less)
4-Mounting screws for protection window (Tightening torque: 0.3 N-m or less)
Insert the screw part of the holder and turn it CW
Lamp Reflector
Protection window
Insert this part into socket to fit the groove
31
Lamp
The lamp insertion holder
(One side is threaded) *Sold Separately
4.12 Battery replacement If the built-in battery-life duration becomes short,
replace the battery with one that is specificated. Be sure to use the
specified battery. Please contact us or the retailer where you purchased this
product if you need it.
Be sure to turn the power OFF. May result in electric shock
Do not replace the battery near flammable materials.
Remove the six screws on the bottom of the DT-3015N main unit, and take out
the built-in battery. Remove the connector and battery fixing screws. Cut the
insulok that is holding the battery onnector with nippers,etcs.
Please take care not to scratch the connector cable and the battery with nippers. Remove the four battery fixing screws and the top cover. Take out the battery from the battery case, and replace with a new battery.
Please be careful that the harness faces the outside. Insert the Insulock attached to the battery through the Insulock through hole on the top cover. Secure the connector cable with the Insulock so that the battery connector contacts the top cover.
(Refer to the Detail drawing) *Please tighten it strongly. Attach the top
cover to the battery case and fix it with the fixing screws (4 places-
tightening torque 0.3 Nm). Connect the battery connector to the main body
connector. Install the battery in the main unit and fix it with the battery
case fixing screws (6 places-tightening torque 0.6 N·m).
Connector cable
Battery connector
Insulok
Body connector
Insulok through hole Four-battery fixing screws
Detail drawing Top cover
Battery case
Six-battery case fixing screws 32
5 Specifications 5.1 Specifications list
Emission count
Resolution
Internal oscillation emission
Phase change function Jump function Limit function Integer function
Delayed emission function
Input signal
External synchronous
emission
Frequency measurement range Phase change function Resolution
Delayed emission function
Display
Setting devices
Lamp
Emission source input
Emission pulse width
Emission timer
Saving function Power supply for sensor
Power
Operating temperature Operating humidity Protection structure Weight
30 to 35000 fpm 30.0 to 5000.0 0.1 fpm
to 8000.0 0.2 fpm to 10000.0 0.5 fpm to 35000.0 1.0 fpm
Available (PHA mode) Available (press “×2” and “÷2″ while on INT) Available
(configurable in function mode 5) Available (configurable in function mode 1)
Measurement range30 to 35000 fpm Within the range (angle)0 to 359°, available
to set by1° Within the range (time) 0.00 to 999.99 ms, available to set by
0.01 ms
1000.0 to 1994.4 ms ,available to set by 0.1 ms H level 2.5 to 12 V L level 0
to 0.5 V Pulse width : 50 µs or more *Input impedance : 10 k or more 27 to
35020 fpm Available (PHA mode)
1 fpm Measurement range27 to 35020 fpm Within the range (angle)0 to 359°,
available to set by1° Within the range (time) 0.00 to 999.99 ms, available to
set by 0.01 ms
1000.0 to 2216.0 ms ,available to set by 0.1 ms 6-digit 7 segment red LED
Multi-turn encoder, tact switch Xenon Lamp 15 W 20 s Emission continuously or
automatically stop in 1 to 120 [min] (adjustable in function mode 3) Power OFF
saving Memory saving ( SELECT number) DC12V 40mA
Built-in NiMH battery
·Charging time : Approx. 2 hours ·Continuous emission time Approx. 3 hours
(when the emission count is 1500FPM)
Supplied AC adapter
·Input: AC 100 V to 240 V 50/60Hz 0 to 40°C (32 to 104°F)
45 to 85% (Non- condensing) None Approx. 1.9 kg ( main unit only) (4.2lb)
Notes 1 The lamp may flicker in high frequency, but it is not failure. It can be measured by instruments so please continue to use it. 2 If noise causes signal integrity issues from the external input sensors, please take measures such as connecting a ferrite core to the signal line.
33
5.2 External dimensions 34
6 Troubleshooting
Troubleshooting
Symptoms
Emission occurs inconsistently.
Emission sometimes stops.
Display does not change by turning the dial. Information appears on the
digital display,but no emission occurs. Emission occurs but does not match the
display. Nothing is indicated on the display and no emission occurs. At
measurement, the object does not stand still completely When the power is
turned ON,”LLLLLL” is indicated, and no emission occurs. Though the battery is
fully charged, “LLLLLL”is soon indicated on the display. The charging lamp
does not light up when the AC adapter is connected. Emission occuers when the
AC adapter is plugged in, but stops when unplugged. Though the battery is
fully charged, the charging lamp remains lit.
Though the charging lamp lights up,charging does not start.
Factors The Xenon lamp failure. The Xenon lamp failure. Internal circuit malfunction. The Xenon lamp failure. Internal circuit malfunction.
Causes
The Xenon lamp life. Internal circuit failure. The Xenon lamp life. Internal
circuit failure.
Internal circuit failure.
Short battery life Internal circuit failure.
Internal circuit failure.
Internal circuit malfunction. The rotartion speed cannnot be set accurately.
The battery voltage is low.
The battery capacity is low. Internal circuit malfunction. The power does not
reach the main unit. The battery is exhausted.
Internal circuit malfunction.
The temperature detection circuit is working.
Internal circuit failure.
Due to resolution.
The battery is not charged. Short battery life Battery malfunction Internal
circuit failure. AC adapter is not connected. AC adapter malfunction
The battery voltage is too low.
Short battery life Charging circuit failure
Not charged within specified temperature.
Solution
Replace the lamp,If the problem has not been solved, ask for repair. Replace
the lamp,If the problem has not been solved, ask for repair.
Ask for repair.
Replace the lamp,If the problem has not been solved, ask for repair.
Ask for repair.
Ask for repair.
It cannot be set below the second decimal place.
Charge the battery.
Replace the battery.
Ask for repair. Check the connection. Ask for repair. Replace the battery.If
the problem has not been solved, ask for repair. Replace the battery.If the
problem has not been solved, ask for repair. Charge in a room temperature
environment. If the problem has not been solved ,ask for repair.
Q A
Questions Is it possible to see the object 2m away? Can I shoot the video? Can
I take a picture? Is this washable? Do you have explosion-proof models? Is it
possible to use at 200V
Answer It depends on the surrounding environment. No.
No.
It is not washable.
Note
Please check it by using demo.
It does not have the signal for the video. It does not have the signal for
pictures. Do not wash.
No.
Impossible at present.
Yes. It can be used at 100-240V.
The included ac adapter is a free power supply up to 240V.
35