MINIWARE DT71 Mini Digital Tweezers User Manual

June 8, 2024
Miniware

User Manual

MINIWARE DT71 Mini Digital Tweezers

MINIWARE DT71 Mini Digital Tweezers

Thank you for purchasing DT71 Mini Digital Tweezers. Please read this manual before using the device. This user manual is based on APP1.05.

Safety Statement

Read carefully all the following safety precautions to avoid personal injuries and prevent damage to the device or any products connected to it. Failure to follow these safety instructions could result in personal injuries or risk of fire.

  1. Please use only the power cable dedicated for this product or certified by your country/region.
  2. Before connecting and disconnecting DT71’s tips to device under test, please power off the circuit to be tested, and after connecting DT71’s tips correctly, power on and measure the circuit.
  3. It is recommended to unplug the controller when DT71 is not in use.
  4. To avoid fire or electric shock, please observe all terminal ratings and marking instructions to avoid damage to the device. Before connecting DT71, please consult the product manual or product label for information about the rated values;
  5. After the power is turned on, do not touch the exposed connectors and components. Do not use it when you suspect that the product is malfunctioning. Please contact after-sales service for testing, maintenance, adjustment or parts replacement;
  6. Static electricity can cause damage to DT71, and measure should be made in anti-static areas if possible. Before connecting DT71 to the device under test, the inner and outer conductors should be grounded briefly to discharge static electricity.
  7. Please keep DT71 surface clean and dry; do not operate in humid, flammable and explosive environment.

Warnings:
Please do not disassemble DT71 controller or test arms. Once disassembled, it cannot be repaired!

Liability Statement:
Any damage of the product or losses related to the product damage, if it’s man-caused, or assumed to be man-caused, the liability will belong to the user. The user is responsible for any damage or loss caused by unauthorized disassembly or modification of the product.

Product Introduction

DT71 Mini Digital Tweezers is a multi-function measure tool with a fully differential input and built-in signal generator. DT71 has a unique ternary structure, which can be split into a controller, test arms  and tweezer tips, flexible in replacement and combination. DT71 uses thickened gold-plated tweezer tips, which can be replaced according to the application scenario. It can measure various devices such as resistor, capacitor, inductor, voltage, frequency, diode, etc. to help users quickly identify components. The built-in micro signal generator of DT71 can output a variety of signals, providing a perfect solution for the debugging and maintenance of complex electronic systems and the classification and detection of discrete chip components.

Appearance

FIG 1 Appearance

Introduction

DT71 Mini Digital Tweezers is a new concept portable LCR tester, which can automatically identify the type of electronic components and actively select the appropriate range for measure; the built-in signal generator can output a variety of required signals for electronic system debugging and maintenance.

FIG 2 Introduction

Parameters

FIG 3 Parameters

Installation And Charging

Installation

  1. Fully insert the 3.Smm audio plug of DT71 controller into the socket of the test arms. After the correct insertion, the screen will display the bootup icon and firmware version, and then enter user interface;                                                             FIG 4 Installation
  2. Tap the touch button to select measure mode for use.

FIG 5 Installation

Button and menu

DT71 has no physical buttons, only a hidden touch button on the top of the controller. Users can switch functions and modes by tapping the touch button.

FIG 6 Button and menu

FIG 7 Menu

FIG 8 Menu

How to charge

Insert the 3.Smm male plug of the dedicated data cable into DT71’s test arms, and then connect a DCSV power supply to charge the test arms. The charging indicator light is always on when charging, and off when fully charged.

FIG 9 How to charge

Automatic sleep and wake up

DT71 has automatic sleep function, when DT71 remains static for 60 seconds (factory setting), it will automatically enter sleep mode; When you need to use it again, close (short-circuit) the tweezer tips to wake up DT71 to enter user interface.

Measure

Preparation before measuring

Before measuring, please connect DT71 controller to test arms, or short- circuit the tweezer tips to wake up DT71. When the screen displays, select the correct measure mode to start measuring.

Measure parameters and accuracy:

FIG 10 Measure parameters and accuracy

The test data comes from laboratory environment and is for reference only.
True error range = ±(readingaccuracy + corresponding resolutionvalue)

Note: Maximum absolute input voltage: -5V – SOV; Input resistance: 1 Mn.

Manual Measure Mode

FIG 11 Manual Measure Mode

FIG 12 Manual Measure Mode

Automatic Identification Mode

FIG 13 Automatic Identification Mode

  1. Long press DT71’s touch button to switch to automatic identification mode (Identify); DT71 can automatically identify inductor.capacitor. resistor. diode;
  2. The screen will display the primary parameters and secondary parameters of the measured object.

Signal Output

FIG 14 Signal Output

Calibration

Zero Calibration

  1. Long press DT71’s touch button to switch to calibration mode (Calibration);
  2. Keep the tweezer tips closed and short circuit until the screen prompts the next step;
  3. Open the tweezer tips and keep it open until the screen prompts the next step;
  4. After the open calibration is successful, the screen will prompt whether to save the calibration data, tap the touch button to confirm and save.
  • If you mistakenly enter calibration mode, long press the touch button to exit.

Accurate Calibration
For accurate calibration, please visit www.miniware.com.cn and download the DT71 calibration accessories, and operate according to the instructions.

Config File

Insert DT71 controller into the Data Cable’s 3.5mm female socket, connect the Data Cable to your PC via USB Type-C cable; a a-character string removable hard disk will appear on your PC. Open the CAL.INI configuration file in the removable disk and set the parameters.

Parameter Setting

FIG 16 Parameter Setting

Restore Factory Setting
Open the CAL.INI config file, delete all data, enter “load default” and save it to restore the factory settings.

User-Defined Wave Setting
USER_WAVEFORM = {
Ox7FF, Ox87F, Ox8FF, Ox97E, Ox9FC, OxA77, 0xAFO, OxB66, 0xBD9, 0xC48. OxCB2, 0xD18, 0xD78, 0xDD3, 0xE29, 0xE77, 0xECO, OxF01, 0xF3C, OxF6F, OxF9A, OxFBE, OxFDA, OxFEE, OxFFA, OxFFE, OxFFA, OxFEE, OxFDA, OxFBE, OxF9A, OxF6F, OxF3C, OxF01, 0xECO, OxE77, 0xE29, 0xDD3, 0xD78, 0xD18, OxCB2, 0xC48, 0xBD9, 0xB66, 0xAFO, OxA77, 0x9FC, Ox97E, Ox8FF, Ox87F, Ox7FE, Ox77E, Ox6FE, Ox67F, Ox601, 0x586, 0x50D, Ox496, 0x424, 0x3B5, Ox34B, Ox2E5, 0x285, 0x22A, Ox1D4, 0x186, 0x13D, OxOFC, OxOC1, 0x08E, Ox063, 0x03F, Ox023, 0xOOF, Ox003, 0xOOO, Ox003, 0xOOF, Ox023, 0x03F. Ox063, 0x08E, OxOC1, 0xOFC, Ox13D, Ox186, 0x1D5, 0x22A, Ox285, 0x2E5, Ox34B, Ox3B5, 0x424, 0x497, 0x50D, Ox586, 0x601, 0x67F, Ox6FE, Ox77E,
OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, OxOOO, }

Note:

  1. Use hexadecimal number OxOOO-OxFFF to represent 0-3V signal waveform;
  2. The output waveform only loads the first 100 points of valid data; red font modification is invalid.

Firmware Upgrade

  1. Visit www.miniware.com.cn to download the latest DT71 firmware to your PC.
  2. Insert DT71 controller into the Data Cable’s 3.Smm female socket, connect the Data Cable to your PC via USB Type-C cable; a a-character string removable hard disk will appear on your PC, entering DFU setting mode.
  3. Copy the .hex firmware to the root directory of that disk. After the extension of the firmware changes from 11 .hex” to 11.rdy11 , restart DT71, thus the firmware is upgraded.

FIG 17 Firmware Upgrade

Standard Service

One year of free warranty will be provided, if the damage was not caused by false manipulation
by the user. Please contact your seller for warranty details.Tweezer tips are consumables products, once it’s used, no replacement will be provided.

Legal Statements

Disposal
Do not dispose this product with domestic waste.

  • This device complies with the WEEE Directive (this additional product label indicates that this electronic product must not be disposed of in household waste).
  • Handling and recycle: Disposal of the product shall be manipulated according to laws and regulations in your area.

Statement of fulfilling FCC standard

This device fulfills part 15 of the FCC regulations. Device must fulfill below 2 conditions:

  1. Device must not generate interference;
  2. Device must be able to resist any interferences on it, including interferences that could cause dangerous manipulation.

Statement of fulfilling CE standard
This product with CE logo on it fulfills related Euro Union laws and regulations.

Read More About This Manual & Download PDF:

MINIWARE DT71 Mini Digital Tweezers User Manual – Download [optimized]
MINIWARE DT71 Mini Digital Tweezers User Manual – Download

Questions about your Manual? Post in the comments!

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Miniware User Manuals

Related Manuals