ifm Optical Distance Sensor Instruction Manual

June 7, 2024
ifm

Table of Contents

ifm Optical Distance Sensor Instruction Manual

1 Preliminary note

1.1 Symbols used

► Instruction

Reaction, result
[…] Designation of pushbuttons, buttons or indications
→ Cross-reference
Important note
Non-compliance can result in malfunctions or interference. Information
Supplementary note.

1.2 Warning signs used

WARNING

Warning of serious personal injury.
Death or serious irreversible injuries may result.

2 Safety instructions

  • Please read this document prior to set-up of the unit. Ensure that the product is suitable for your application without any restrictions.
  • Improper or non-intended use may lead to malfunctions of the unit or to unwanted effects in your application. That is why installation, electrical connection, set-up, operation and maintenance of the unit must only be carried out by qualified personnel authorised by the machine operator.
  • In case of malfunction of the unit please contact the manufacturer. If the unit is tampered with and/or modified, any liability and warranty is excluded.

WARNING

Visible laser light; laser protection class 2.
Use of controls or adjustments other than those specified herein may result in hazardous radiation exposure. Damage to the retina is possible.
►► Do not stare into the laser beam!
►► Apply the enclosed labels (laser warning) in the immediate vicinity of the unit.
►► Adhere to the caution and warning notes on the product label.
►► Use the enclosed label for the power supply cable.
►► EN/IEC 60825-1 : 2007 and EN/IEC 60825-1 : 2014 complies with 21 CFR 1040
except for deviations pursuant to Laser Notice No. 50, dated June 2007

Label for supply cable

Product label

3 Functions and features

The unit is used as an optical distance sensor.

3.1 Applications
  • The optical distance sensor measures distances between 0.2 and 10 m.
  • It has a background suppression at > 10…19 m.
  • The measured value is shown in a 10-segment display.
  • According to the set output functions 2 output signals can be generated.
  • O1D100 / O1D120: Certification 21 CFR PART 1040

The distance between the sensor and the background must be limited to max. 19 m by the customer. Otherwise the measured value may be
ambiguous. → 5.1 Installation conditions As an alternative article O1D105 can be used (the distance between sensor and background can be max. 100 m).

4 Functions

4.1 Output function hysteresis

The hysteresis keeps the switching state of the output stable if the measured value varies about the sensing range. Both outputs (OUT1 and OUT2) can be set
as hysteresis function. → 10.2.4 Hysteresis function

4.2 Output function window

The window function enables the monitoring of a defined acceptable range. Both outputs (OUT1 and OUT2) can be set as window function. → 10.2.6 Window function

4.3 Analogue output function

The sensor can also output an analogue signal on output 2 (OUT2) which is proportional to the distance. → 10.2.11 Scaling of the measuring range (analogue output)

4.4 Switch off the laser

For safety and maintenance purposes the laser of the unit can be temporarily switched off via the input on pin 5.

Input signal at pin 5 Laser
Low / not connected On
High Off

5 Installation

5.1 Installation conditions

►►Install the unit so that the object to be detected is within a measuring range of 0.2…10 m.
The unambiguity range of the sensor is fixed to 19.2 m. Objects within a range

10…19.2 m are suppressed.
Reflecting objects in the direct beam path of the sensor – also in the range > 19.2 m – are to be avoided by the customer. Otherwise the measured values may be ambiguous.

5.2 Mounting accessories

The unit is supplied without mounting accessories.

Examples of mounting accessories Art. no.
Protective cover O1D E21133
Mounting set E2D101 + E20938 + E20951 E21079
Mounting set O1D (for rod mounting Ø 12 mm) E2D101
Mounting rod straight Ø 12 mm / M10 E20938
Mounting set O1D (for rod mounting Ø 14 mm) E2D111
Mounting rod straight Ø 14 mm / M12 E20939

Fixture for mounting and fine adjustment of O1D laser units (rod or free- standing; depending on the clamp)| E1D100

Example mounting:

  1. Mounting set for Ø 12 mm rod art. no. E2D101
  2. Mounting rod straight Ø 12 mm / M10 art. no. E20938

6 Electrical connection

The unit must be connected by a qualified electrician.

►►The national and international regulations for the installation of electrical equipment must be adhered to.
►►Ensure voltage supply according to EN 50178, SELV, PELV. O1D100 / O1D120: cULus, Supply Class 2
►►Disconnect power.
►►Connect the unit as follows:

O1D100 / O1D120 PNP

Core colours of ifm sockets:
1 = BN (brown), 2 = WH (white), 3 = BU (blue), 4 = BK (black), 5 = GR (grey).

6.1 Operation with IO-Link master

The unit is compatible with IO-Link master port class A (type A).
For operation with IO-Link master port class B (type B) observe the following: As a standard, the unit is not compatible with master port class B (type B). Pin 2 (OU2) and pin 5 (IN1) are used for manufacturer-specific functions. That means that the main supply voltage of the unit and the additional voltage supply (master port class B on pins 2/5) are not electrically isolated. With the following configurations the unit can be used with master port class B:

  • Connect unit and IO-Link master via 3 wires: Connect pins 1, 3 and 4 of the unit with the IO-Link master (do not connect pins 2 and 5).
  • Connect unit and IO-Link master via 4 wires: Deactivate pin 2 (OU2) via IOLink (setting OU2 = “off”) and connect pins 1, 2, 3 and 4 of the unit with the
    IO-Link master (do not connect pin 5).

7 Operating and display elements

1:| 4 x LED green| Lighting LED = power and set display unit (mm, m, inch)
2:| 4 x LED yellow (two not connected)| Indication of the switching status; lights, if the corresponding output is switched
3:| 4-digit alphanumeric display| Indication of the measured distance, the parameters and parameter values
4:| Programming button [SET]| Setting of the parameter values (scrolling by holding pressed; incrementally by pressing once)
5:| Programming button [MODE/ENTER]| Selection of the parameters and acknowledgement of the parameter values

8 Menu

8.1 Menu structure

8.2 Explanation of the menu

For the factory settings please refer to the end of these instructions (→ 14 Factory setting).

| Configuration for output 1

4 switching functions can be selected:

[Hno], [Hnc], [Fno], [Fnc] → 10.2.3 Configuration of OUT1
---|---
**| Switch point for hysteresis function OUT1**

Limit value at which the output with selected hysteresis function changes its switching state (object nearer/farther than distance set)

[SP1] is only active if [OU1] = [Hno] or [Hnc]

→ 10.2.5 Setting of the switch point for hysteresis function OUT1


| Switch points for window function OUT1

Limit values at which the output with selected window function changes its switching state (object present / not present between the distance “near” and the distance “far”)

[nSP1] = switch point “near” / [FSP1] = switch point “far” [nSP1] / [FSP1] are only active if [OU1] = [Fno] or [Fnc]

→ 10.2.7 Setting of the switch points for window function OUT1

| Configuration for output 2

4 switching functions and 2 analogue signals can be selected: [Hno], [Hnc], [Fno], [Fnc], [I], [U] → 10.2.8 Configuration of OUT1

| Switch point for hysteresis function OUT2

Limit value at which the output with selected hysteresis function changes its switching state (object nearer/farther than distance set)

[SP2] is only active if [OU2] = [Hno] or [Hnc]

→ 10.2.9 Setting of the switch point for hysteresis function OUT2



| Switch points for window function OUT2

Limit values at which the output with selected window function changes its switching state (object present / not present between the distance “near” and the distance “far”)

[nSP2] = switch point “near” / [FSP2] = switch point “far” [nSP2] / [FSP2] are only active if [OU2] = [Fno] or [Fnc]

→ 10.2.10 Setting of the switch points for window function OUT2

| Analogue start point

Measured value at which 4 mA / 0 V are provided [ASP] is only active if [OU2] = [I] or [U]

→ 10.2.11 Scaling of the measuring range (analogue output)


| Analogue end point

Measured value at which 20 mA / 10 V are provided [AEP] is only active if [OU2] = [I] or [U]

→ 10.2.11 Scaling of the measuring range (analogue output)

---|---
**| Teach mode**

Selection “sampling rate” or “repeatability”

→ 10.3 Teach mode

**| Extended functions**

Press [SET] to open the submenu “Extended functions”

→ 10.4 Extended functions

---|---





| Delay for the switching outputs [dSx] = switch-on delay; [drx] = switch-off delay�

The output does not immediately change its switching status when the switching condition is met but only after the delay has elapsed� If the switching condition is no longer met after the delay has elapsed, the switching status of the output does not change�

[dS2] and [dr2] are not effective if [OU2] = [I] or [U]�

→ 10.4.1 Setting of the time delay for switching outputs

**| Setting of the fault suppression time for switchting outputs/analogue output**

This function allows to suppress short-time saturation of the measuring element (such saturation can result from direct reflection or strong fluctuations in brightness)�

During the delay set, the latest valid value measured is displayed, the output signals remain unchanged�

→ 10.4.2 Setting of the fault suppression time for switching outputs /

analogue output


| Display setting

7 settings can be selected:

[d1], [d2], [d3], [rd1], [rd2], [rd3], [OFF]

→ 10.2.2 Setting of the display

| Setting of the display unit

Selection of the unit of measurement for [SP1], [SP2], [ASP], [AEP] Selection options: [mm] [m] [inch]

→ 10.2.1 Selection of the display unit


| Restore factory setting

→ 10.4.3 Reset of all parameters to factory setting

**| Display of the software version number**

→ 10.4.4 Display of the software version number

9 Operating modes

9.1 Run mode

The run mode is the normal operating mode. After power on the unit is in the Run mode. It carries out its monitoring function and generates output signals according to the set parameters.
The display indicates the current distance, the yellow LEDs signal the switching status of the outputs.

9.2 Display mode

Indication of the parameters and the set parameter values.
►►Press [MODE/ENTER] briefly. >> Unit goes to the Display mode. Internally it remains in the operating mode. The set parameter values can be read.
►►To scroll through the parameters, press [MODE/ENTER] briefly.
►►To display the respective parameter value, press [SET] briefly. >> After 15 s the unit returns to the Run mode.

9.3 Object reflectivity

Display of the object reflectivity (formerly: align mode).
►►Press [SET] in the Run mode.

The unit displays an orientation value for the object reflectivity (e.g. +100 corresponds to a white object, +020 corresponds to a grey object).

9.4 Programming mode

Setting the parameter values → 10.1 Parameter setting in general

10 Parameter setting

During parameter setting the unit remains internally in the operating mode. It continues its monitoring function with the existing parameters until the change has been finished.

10.1 Parameter setting in general
10.1.1 Setting of the parameter value

Select the display unit [Uni] before you define the values for the parameters. In case of subsequent changes of the display unit rounding errors during internal conversion to other units may falsify the set values. → 10.2.1 Selection of the display unit.

10.1.3 Electronic lock

The unit can be locked electronically to prevent unintentional settings. On delivery the unit is not locked.

10.1.3 Electronic lock

The unit can be locked electronically to prevent unintentional settings. On delivery the unit is not locked.

Timeout
If no button is pressed for 15 s during the setting procedure, the unit returns to the Run mode with unchanged values.

10.2 Configuration of the basic settings

10.2.1 Selection of the display unit

Set [Uni] before the values for the parameters [SPx], [nSPx], [FSPx],
[ASP], [AEP] are defined.
In case of subsequent changes of the display unit rounding errors during
internal conversion to other units may falsify the set values.
►► Change to [EF].
►► Select [Uni] and set the unit of measurement. Selection of the unit of measurement: [mm], [m], [inch] ►► Confirm with [MODE/ENTER].

The selected unit is indicated by a green LED on the display.

**10.2.2 Setting of the display


10.2.3 Configuration of OUT1

**

10.2.4 Hysteresis function

The hysteresis keeps the switching state of the output stable if the measured value varies about the sensing range. In either case set and reset points are
symmetrically arranged around the selected switch point [SPx]. The hysteresis is the distance between set and reset points; it is calculated on the basis of the repeatability with a safety factor of 1.5. The calculation is based on the repeatability for max. 40 klux. → 10.3.3 Table repeatability and accuracy Example Hno

  1. For the output function [Hno] the output switches when the object approaches and when the switch point (A) is reached.
  2. When the object is removed again, the output does not switch back before the reset point (B) is exceeded. The reset point (B) is greater than the set point (A).

When the output function [Hnc] was selected, set and reset point are reversed.
The output is switched off when the object approaches. When the object is removed, the output switches on.

Switching status of the outputs
Output function Object distance (D) Output status
[Hno] D < [SPx] Closed
D > [SPx] Open
[Hnc] D < [SPx] Open
D > [SPx] Closed
Example of output function [Hno]

Sampling rate 50 Hz, distance to the object 1200 mm, grey value (18 % remission):
Hysteresis = ± 10 mm (repeatability → 10.3.3 table) x factor 1.5 = 15 mm
–Reset point 1200 mm + (15 mm) = 1215 mm
–Set point 1200 mm – (15 mm) = 1185 mm

10.2.5 Setting of the switch point for hysteresis function OUT1

► Select the output function [Hno] or [Hnc] at [OU1].
► Confirm with [MODE/ENTER].
► Select [SP1] and set the switch point.
► Confirm with [MODE/ENTER].

10.2.6 Window function

It is possible to define a window for the object recognition for each of the two outputs (OUT1 / OUT2). Switches off when the object is detected.

[nSPx] = switch point “near”; [FSPx] = switch point “far”; FE = window If the measured value is between the switch point “near” [nSPx] and the switch
point “far” [FSPx], the output is open (when [OUx] = [Fnc]).

Switches when the object is detected

[nSPx] = switch point “near”; [FSPx] = switch point “far”; FE = window If the measured value is between the switch point “near” [nSPx] and the switch
point “far” [FSPx], the output is closed (when [OUx] = [Fno]). Switching status of the outputs Output function| Object distance (D)| Output status
---|---|---
[Hno]| D < [SPx]| Closed
D > [SPx]| Open
[Hnc]| D < [SPx]| Open
D > [SPx]| Closed

10.2.7 Setting of the switch points for window function OUT1

► Select the output function [Fno] or [Fnc] at [OU1].
► Confirm with [MODE/ENTER].
► Select [nSP1] and set the switch point “near”.
► Confirm with [MODE/ENTER].
► Select [FSP1] and set the switch point “far”.
► Confirm with [MODE/ENTER].


Current output 4 … 20 mA

MEW = final value of the measuring range
In the set measuring range the output signal is between 4 and 20 mA. Faults are also displayed:
Too much light or object too near: 3.5 mA for a rising edge ([ASP] < [AEP]), 20.5 mA for a falling edge ([ASP] > [AEP]).
Object too far or no object present: 20.5 mA for a rising edge; 3.5 mA for a falling edge.
Voltage output 0 … 10 V

MEW = final value of the measuring range
In the set measuring range the output signal is between 0 and 10V.

10.3 Teach mode

10.3.1 Setting of the sampling rate


Distance in [mm]

| Repeatability| Accuracy
---|---|---
white

90 % remission

| grey

18 % remission

| white

90 % remission

| grey

18 % remission

2001000| ± 50 mm| ± 75 mm| ± 150 mm| ± 180 mm
10002000| ± 55 mm| ± 100 mm| ± 150 mm| ± 200 mm
2000000| ± 175 mm| ± 225 mm| ± 250 mm| ± 320 mm
40006000| ± 275 mm| ± 400 mm| ± 350 mm| ± 500 mm
600010000| ± 600 mm| —| ± 700 mm| —

Values for sampling rate 50 Hz, extraneous light of 40…100 klx

*Range referred to black (6 % remission) ≤ 4000 mm. The values apply at:

  • constant ambient conditions (23°C / 960 hPa)
  • only after unit powered up for 10 minutes

10.4 Extended functions
10.4.1 Setting of the time delay for switching outputs

11.1 General information

This unit has an IO-Link communication interface which requires an IO- Linkcapable module (IO-Link master) for operation.
The IO-Link interface enables direct access to the sensor values and parameters and provides the possibility to set the parameters of the unit during operation.
In addition communication is possible via a point-to-point connection with a USB adapter cable.
You will find more detailed information about IO-Link at www.ifm.com.

11.2 Device-specific information

You will find the IODDs necessary for the configuration of the IO-Link device and detailed information about sensor values, diagnostic information and parameters
in the overview table at www.ifm.com.

11.3 Parameter setting tools

You will find all necessary information about the required IO-Link hardware and
software at www.ifm.com.

11.4 Extended functions

With IO-Link extended functions and measured data are available.

11.4.1 Teach function

The following teach functions are available:

  • Teach on background
  • Two-point teaching
    The teach functions are described in detail in the IODD.

11.4.2 Object reflectivity
The object reflectivity is provided via the display and as process data value (PDV).
The object reflectivity can, for example, be used to detect sensor soiling.

12 Set-up / operation

►►After mounting, wiring and programming check whether the unit operates correctly.

If the unit has been correctly set up, the distance to the object is indicated.

Lifetime of a laser diode: 50000 hours

12.1 Fault indication

Display| Possible cause| Switching output| Current output / voltage output| IO-Link process value distance 3)| IO-Link process value object reflectivity 3)
---|---|---|---|---|---
[Hno]| [Hnc]| [Fno]| [Fnc]| [ASP] < [AEP]| [ASP] > [AEP]


[++]

| too much light,

eg reflective surface

| ****

ON

| ****

OFF

| ****

OFF

| ****

ON

| ****

3,5 mA /

0 V

| ****

20,5 mA /

10 V

| ****

NoData

| ****

OL

[- -]| too little light, no object| OFF| ON| OFF| ON| 20,5 mA /

10 V

| 3,5 mA /

0 V

| NoData| UL



[nEAr]

| object to be measured outside the measuring range < 02 m| ****


ON

| ****


OFF

| ****


OFF

| ****


ON

| ****

3,5 mA /

0 V

| ****

20,5 mA /

10 V

| ****


UL

| ****


NoData



[FAr]

| object to be measured outside the measuring range > 10 m| ****


OFF

| ****


ON

| ****


OFF

| ****


ON

| ****

20,5 mA /

10 V

| ****

3,5 mA /

0 V

| ****


OL

| ****


NoData


[Errp]

| plausibility (eg object too fast)| ****

X1)

| ****

X1)

| ****

X1)

| ****

X1)

| ****

X1)

| ****

X1)

| ****

X1)

| ****

X1)

[LoFF]| laser switched

off

| OFF| ON| OFF| ON| 20,5 mA /

10 V

| 3,5 mA /

0 V

| NoData| NoData
[SC1]| short circuit in switching output 1| 2)| 2)| na| na
[SC2]| short circuit in switching output 2| 2)| 2)| X1)| X1)
[SC]| short circuit in all switching outputs|  |  | na| na

  1. unchanged
  2. [SC1] or [SC] only active, if output 2 is configured as switching output.
  3. See IODD of the device
13 Maintenance, repair and disposal

Faulty sensors must only be repaired by the manufacturer.
►►Keep the front lens of the sensor free from soiling.
►►After use dispose of the unit in an environmentally friendly way in accordance with the applicable national regulations.

14 Factory setting

Technical data and further information at www.ifm.com

Read More About This Manual & Download PDF:

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals