Notice! Please remove the guide before giving the device to the patient. Only medical professionals could adjust pressure settings.

Provider Setup Navigation

  1. Press “ Menu ” to any Menu (See Operation Flow on the right) while in the Standby screen.
  2.  Hold down both the “ UP”   & ” Down ” buttons, and then simultaneously press the “ Menu” for 1 second. When the screen starts blinking, the Provider Mode would be unlocked.
  3. Press “UP ” and ” Down ” to adjust the settings.
  4.  Press “ Menu ” to confirm the settings.
  5.  Press the “ Start/Standby ” to go back to the standby screen.

Provider Mode Quick Guide

APEX XT Auto Automatic Positive Airway Pressure - Provider
Mode

Re-set the Compliance Meter

  1. In the standby screen, press “ Menu” to select [ CM xxxx.x hr ] menu.
  2.  Accessing the Provider Mode first ( Press ), then the screen starts blinking.
  3.  Press “ UP ” or ” Down ” , then the screen will show [Clear].
  4. Press “ Menu ” to confirm the settings, and the Compliance Meter record will be erased. The screen will show [Clear OK] when finishing the process.

APEX XT Auto Automatic Positive Airway Pressure - Compliance
Meter

Engineering Mode Quick Guide

Accessing the Engineering Mode
1. Plug the power supply into the iCH device.
2. Before it shows [ Standby ] screen, hold down both the “ Starts/Standby ” and “ Menu” buttons for 1 second, the
Engineering Mode would be unlocked.

APEX XT Auto Automatic Positive Airway Pressure - Total
Meter
Re-set the Total Meter

  1. Accessing the Engineering Mode first, then the screen will show [ TM xxxx.x h r] menu.
  2. Hold down both the “ UP ”   & ” Down ” buttons, and simultaneously press the “ Menu ” , and then the screen starts blinking.
  3. Press “ UP ” or ” Down ” , then the screen will show [ Clear ].
  4. Press “ Menu ”    to confirm the settings, and the screen will show [ TM CLr Clear ],and the Total Meter record will be erased.
  5.  The screen will show [ TM End Clear OK ] when finishing the process.