Instructions For Use
Spectra S1 Plus / S2 Plus
**PLEASE READ ALL INSTRUCTIONS BEFORE USING THIS PRODUCT MODEL
:__ SERIAL#:___**
DO NOT THROW AWAY THIS MANUAL. PLEASE KEEP IT FOR FUTURE REFERENCE.
INTENDED USE The Spectra S1 Plus and Spectra S2 Plus are powered breast pumps to be used by lactating women to express and collect milk from their breasts.
IMPORTANT SAFEGUARDS
This breast pump is designed to be a single-user product. Use of this
breast pump by anyone other than the original owner presents health risks,
affects performance integrity, and voids the product warranty.
If you are a mother who is infected with Hepatitis B, Hepatitis C or Human
Immunodeficiency Virus (HIV), pumping breast milk will not reduce or remove
the risk of transmitting the virus to your baby through your breast milk.
Before each use, visually inspect the individual components for cracks, chips,
tears, discoloration or deterioration. If damage to any of the individual
components is observed, discontinue use of the pump and contact the
manufacturer or distributor for replacement parts.
READ THIS ENTIRE MANUAL BEFORE USING IT.
When using electrical products, basic safety precautions should always be
followed.
TO REDUCE THE RISK OF ELECTROCUTION:
- Do not use while bathing.
- Do not place or store products where it can fall or be pulled.
- Do not place or drop into water or other liquids.
- Do not reach for a product that has fallen into water or other liquids; unplug the product immediately.
- Do not sanitize the kit in the microwave; never immerse the main body in water
TO REDUCE THE RISK OF BURNS, ELECTROCUTION, FIRE, OR INJURY TO PERSONS:
- This product should never be left unattended when plugged into an electrical outlet.
- Do not leave this product unattended around infants or children.
- Use this product only for its intended use as described in this manual; do not use any accessories or attachments not expressly manufactured by Uzinmedicare for the Spectra S1 Plus / Spectra S2 Plus breast pump.
- Never operate this product if it has a damaged cord or plug if it is not working properly, if it has been dropped, damaged or dropped into water, or if any parts are broken or missing.
- Keep the cord away from heated surfaces.
- Never use the pump while driving, sleeping or drowsy.
- Do not use outdoors or operate in environments where aerosol spray products are being used or where oxygen is being administered.
- Do not disassemble or modify the product.
- Do not use adjacent to or stacked on other electrical equipment.
- Allergic reactions may occur.
- Lint, dust, light (including sunlight), etc. can adversely affect the safety and performance of this product.
- When not in use, unplug the power cord from the electrical outlet.
NOTE: If you feel pain in your breast or nipple, turn the pump off, pull the adapter plug out and remove the breast shield from your breast (the suction can be broken by shutting the pump off and inserting your fingers between your breast and the breast shield). Consult a healthcare professional or lactation advisor, if necessary.
Important Symbols
Indications for Use
The Spectra S1 Plus / S2 Plus is a powered, breast pump to be used by lactating women to express and collect milk from their breasts. The Spectra S1 Plus / S2 Plus is intended for a single user.
Components and Assembly
Before using the breast pump for the first time, sanitize the breast shield in
boiling water for 5 minutes.
NOTE: Do not boil for more than 5 minutes, as the breast shield may become
distorted.
How to Assemble the Back-Flow Protector:
- Insert the white silicone membrane into the lower case.
- Insert the upper case into assembled silicone membrane.
How to Assemble the Pump Ki t:
|
---|---
Push valve into breast shield| Insert back-flow protector into breast shield
|
Fasten the bottle onto the breast shield| Insert tubing into back-flow
protector
Connecting the Tubing to the Pump
Pumping Guide
- Relax your body.
- Stimulate your let-down reflex before pumping. Some women have found that gently massaging their breasts or applying warm compresses from the back of the breast toward the nipple can stimulate their let-down reflex.
- Express milk from each breast for 10-15 minutes.
NOTE: For optimal milk expression, we recommend that you pump frequently but in short intervals. (10-15 minutes per breast, every 2-3 hours, 7-8 times a day).
Getting Started
NOTE: The Spectra S1 Plus and Spectra S2 Plus are designed to turn off automatically after 30 minutes and remember your current settings for your next pumping session. If you want to start pumping on settings different from those used during your previous session, adjust the settings before you start expressing.
Even when you are using the breast pump, we encourage you to breastfeed your baby whenever you can.
- SETTING THE VACUUM LEVEL: Adjust the vacuum level by pressing the buttons.
The vacuum level is measured in mmHg on your breast pump and refers to how
strongly the pump will `suck’. Please be cautious when setting the vacuum
level on the Spectra S1 Plus / S2 Plus. Some women find that milk expression
occurs faster on higher vacuum levels. However, it is important not to set the
vacuum level so high that it makes you uncomfortable, as the resulting pain
will hinder your milk flow.
We, therefore, suggest that you start off with the vacuum level on low and
then gradually increase the vacuum level to a comfortable level. You may find
that the vacuum level that you are comfortable with varies significantly, from
day to day, during the day and even during the same expression session.
- SETTING THE CYCLE SPEED: Adjust the cycle speed by pressing the button.
The cycle speed is measured in CPM (cycles per minute) on your breast pump and
refers to how many times the pump `sucks’ and releases per minute. Your
Spectra S1 Plus/S2 Plus has been designed with variable cycle settings so that
you can find the setting that suits your body best.
Changing the cycle speed can make a dramatic difference in milk expression for
some women.
Every woman is different, so adjust the cycle speed setting to find the speed
that works best for you.
- MASSAGE AND EXPRESSION MODES
– Massage Mode: You will adjust the vacuum level in massage mode to find what
works best for you.
– Expression Mode: After the letdown begins, and once the milk is flowing,
your baby will start sucking more slowly and deeply, creating a stronger
suction to draw down the milk.
This mode has a slower cycle speed and higher vacuum level.
Please adjust the cycle speed and vacuum level in expression mode to find what
works best for you.
LCD Display
Button Controls
Cleaning and Disinfection
-
First Time Use
Before using the Spectra S1 Plus or S2 Plus breast pump for the first time, disassemble and wash in warm, soapy water all parts that come into contact with the breast and breastmilk. -
After Every Use
1. Disassemble and wash all parts of your pump kit that come into contact with the breast and breastmilk (breast shield kit, valve head, membrane nipple, nipple cover, disk, and cap) in warm water with mild soap. All parts in the pump kit are dishwasher safe. 2. 2. 2. Rinse in clean water and air- dry.
3. When parts are completely dry, cover until next use.
NOTE: TO PREVENT CONDENSATION FROM COLLECTING IN THE AIR TUBE, WASH AND DRY IT IMMEDIATELY. IF CONDENSATION REMAINS IN THE AIR TUBE, REPLACE THE TUBING. TO CLEAN THE MAIN BODY, WE RECOMMEND WIPING WITH A CLEAN, DAMP (NOT WET) CLOTH (NO SOAP).
How long can I store breast milk?
At Room Temperature:
Temperature | Storage Time |
---|---|
15°C (60°F) | 24 hours |
26°C (79°F) | 4 to 6 hours |
30°C (86°F) | 4 hours |
In the Refrigerator:
Temperature | Storage Time |
---|---|
0°C – 4°C (32°F – 39°F) | Up to 8 days |
In the Freezer:
Type of Freezer | Description | Storage Time |
---|---|---|
Compartmentalized | part of a refrigerator | Up to 2 weeks |
Self-Contained | bottom, top mount, side-by-side, or stand-alone freezer | 3 to |
4 months
Deep| constant -17°C (0°F) temperature| 6 to 12 months
Source: Breastfeeding Made Easier at Home and Work, womenshealth.gov1
How do I thaw and use stored breast milk?
- Always make sure to check the date on the milk container before using. Do not use updated milk or milk that has been stored too long.
- Thaw frozen milk in the refrigerator or under cool, running water.
- Heat the milk under warm, running water and gently swirl the container to mix the milk.
- If warm water is not available, heat a pan of water on the stove. Once the water is warm (not boiling) remove the pan from the stove and place the milk container in the pan. Never warm the milk container directly on the stove.
- Never microwave breast milk because microwaves can cause dangerous hot spots that could burn you or a baby
- Always test the temperature of the milk on the inside of your wrist before feeding it to a baby. The milk should feel warm, not hot. · After thawing, milk should be stored in the refrigerator for no more than 24 hours.
- Never refreeze thawed milk. Throw away previously frozen milk that is not used within 24 hours.
Battery Information (for Spectra S1 Plus Only)
The battery used in this device may present a risk of fire or chemical burns
if not used properly.
Do not disassemble, heat, incinerate or dispose of battery in fire.
Dispose of the battery in accordance with local, state, and federal
regulations.
Rechargeable Battery: 11.1V, 2000mAh, Lithium-Ion
Battery indicator displayed with 4 bars| Operating time remaining:
approximately 3 hours
Battery indicator displayed with 1 bar| Operation time remaining:
approximately 1 hour
Battery indicator flashing| Operation time remaining: approximately 20 minutes
Night Light
The innovative built-in LED light allows you to select between two lighting levels, so you always have a light source available for expressing milk.
Trouble Shooting
Condition | Possible Causes | Check | Solution |
---|---|---|---|
No Power | Adaptor | Is the power cord plugged in? Is the power cord damaged? |
Plug into the outlet. If it is damaged, contact the manufacturer or the vendor
from whom you purchased the pump.
Dead battery
(Spectra SI Plus Only)| Is the battery fully charged?| If it doesn’t charge,
contact the
manufacturer or the vendor from
whom you purchased the pump.
No Suction| Air tube| Is the tubing connected to the pump?| Connect the air
tube to the pump.
Valve| Are the valve and tube connected to the breast shield properly?|
Disassemble and reassemble the parts properly.
Air cap| Is the air cap closed?| Close the air cap.
Low Suction| Vacuum Level| Is the vacuum level set too low? Is the valve
intact?
Is the pump making an odd noise?| Contact the manufacturer or seller of the
pump.
Defective assembly| Are kit parts assembled properly?| Disassemble and
reassemble the kit parts properly.
Breast shield| Is the breast shield damaged?| Contact manufacturer to replace
the breast shield.
Valve| Is the valve damaged?| Contact manufacturer to replace the valve.
Pain| Too high suction| Is the cycle speed set too high?| Set to the lowest
suction level and see your lactation consultant
or health care professional for advice.
Use and Storage Conditions
Technical Specifications
Protection Type : Class II
Weight : 1.6 kg (approximately 3.5 lbs)
Adapter : Input 100V 240AC, 50/60Hz, 600mA
Output : 12V, 2A
Battery : 11.1V 2000mAh Li-Polymer
- Type of protection against electric shock: Class II and/or internally powered equipment Degree of protection against electric shock: Type BF applied parts.
- Classification according to the degree of protection against ingress of water: IPX0, ordinary equipment
- This equipment is not suitable for use in the presence of flammable anesthetics or oxygen.
- Mode of operation: Continuous.
Warranty
This product is warranted by Uzinmedicare Co., to the original retail
purchaser to be free from defects in material and workmanship for a maximum
period of two (2) years from the date of purchase. In the event of a defect or
failure to conform to this warranty, Uzinmedicare Co. will, at Uzinmedicare
Co.’s option, repair or replace this product without charge for such
replacement, or parts or labor, at the Uzinmedicare Co. office listed below.
The purchaser shall bear all responsibility and expense for returning this
product, including the risk of loss prior to receipt by Uzinmedicare Co., as
well as shipping, packing, and insurance costs.
ANY AND ALL IMPLIED WARRANTIES ARE LIMITED TO THE DURATION OF TWO (2) YEARS
FROM THE DATE OF PURCHASE. SOME STATES DO NOT ALLOW LIMITATIONS ON HOW LONG AN
IMPLIED WARRANTY LASTS, SO THE ABOVE LIMITATIONS MAY NOT APPLY TO YOU.
THE LIABILITY OF UZINMEDICARE CO., FOR BREACH OF ANY WRITTEN OR IMPLIED
WARRANTY, IS LIMITED TO REPAIR OR REPLACEMENT OF THIS PRODUCT. UZINMEDICARE
CO. WILL HAVE NO LIABILITY UNDER ANY CIRCUMSTANCES FOR ANY INCIDENTAL OR
CONSEQUENTIAL DAMAGES. SOME STATES DO NOT ALLOW THE EXCLUSION OR LIMITATION OF
INCIDENTAL OR CONSEQUENTIAL DAMAGES, SO THE ABOVE LIMITATIONS MAY NOT APPLY TO
YOU.
THIS WARRANTY GIVES YOU SPECIFIC LEGAL RIGHTS. YOU MAY ALSO HAVE OTHER RIGHTS
WHICH VARY FROM STATE TO STATE.
Uzinmedicare Co. will have no obligation under this warranty to repair any
malfunction or damage caused by accident or arising from misuse, abuse,
improper maintenance, unauthorized modification, or connection to an improper
power supply. Without limiting the generality of the foregoing, bending or
dropping of this product or its components and visible cracking of the
equipment housing will be presumed to be defects resulting from misuse or
abuse. Accessories that are not expressly manufactured by Uzinmedicare Co. and
intended for use with the Spectra S1 Plus or Spectra S2 Plus breast pump will
immediately void this warranty. A charge will be made for the repair or
replacement of damaged parts.
If you wish to make a claim under this warranty, you must return this product
to Uzinmedicare Co. at the address below, unless directed otherwise by
Uzinmedicare Co. Please contact Customer Service at Uzinmedicare Co. and
obtain a Return to Merchant Authorization (RMA) Number and reference the
number on your return package. The original invoice number or Purchase Order
Number must be referenced on the RMA. You must call first for an RMA. Returns
will not be accepted without an accompanying RMA.
Uzinmedicare US Customer Service:
Spectra Baby USA
6851 SW 21st Court, Suite 1
Davie, FL 33317
Website: http://www.spectrababyusa.com
Email:
customerservice@spectrababyusa.com
Telephone: 855-446-6622
WARRANTY CARD
Please complete the information below and mail the warranty card to the following address:
Spectra Baby the USA
6851 SW 21st Court, Suite 1
Davie, FL 33317
Website: http://www.spectrababyusa.com
Email:
customerservice@spectrababyusa.com
Telephone : 855-446-6622
NOTE: To facilitate the processing of your warranty claim, please include proof of purchase from an authorized retailer.
| Spectra S1 Plus / S2 Plus Electric Breast pump
---|---
Model No.|
Serial No.|
Name|
Email|
Home Phone|
Mobile Phone|
Street Address|
City and State|
Zip Code|
Country|
Electromagnetic Compatibility (EMC)
This equipment needs special precautions regarding Electromagnetic
Compatibility (EMC) and needs to be installed and put into service in
accordance with the EMC information provided in this Instructions for Use
Manual.
Wireless communications equipment such as wireless home network devices,
mobile phones, cordless telephones and their base stations, walkie-talkies can
affect the Spectra S1 Plus and S2 Plus and should be kept at least 1.0 meters
away from the equipment.
This device complies with part 15 of the FCC Rules. Operation is subject to
the following two conditions:
A. This device may not cause harmful interference.
B. This device must accept any interference received, including interference
that may cause undesired operation.
This equipment has been tested and found to comply with the limits for a Class
B digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
typical installation. This equipment generates, uses and can radiate radio
frequency energy, and if not installed and used in accordance with these
instructions, may cause harmful interference to radio communications. However,
there is no guarantee that harmful interference will not occur in a particular
installation. If this equipment causes harmful interference to radio or
television, which can be determined by turning the equipment off and on, the
user is encouraged to try to correct the interference using one or more of the
following measures:
– Reorient or relocate the receiving device.
– Increase the separation between the equipment.
– Connect the equipment into an outlet on a circuit different from that to
which the other device(s) are connected.
– Consult the manufacturer or a field service technician for help.
The Spectra S1 Plus & S2 Plus breast pumps should not be used adjacent to or stacked with other electrical equipment. If adjacent or stacked use is necessary, it is the responsibility of the user to verify that the pump operates normally in such a configuration.
This equipment has no essential performance. It has been tested for immunity to electromagnetic disturbances and passed testing based on certain criteria, such as component failures, changes in programmable parameters or operating mode, cessation or interruption of any intended operation, or the initiation of any unintended operation.
TABLE 1: Electromagnetic Emissions (IEC 60601-1-2:2007)
This equipment is intended for use in the electromagnetic environment specified below. It is the responsibility of the user of the Spectra S1 Plus or Spectra S2 Plus breast pump to ensure that it is used in such an environment.
Emission test | Compliance | Electromagnetic environment-Guidance |
---|
RF emissions
CISPR 11| Group I| This equipment uses RF energy only for internal function.
Its RF emissions thus are very low and arc not likely to cause interference
with nearby electronic equipment.
RF emissions
CISPR 11| Class B| This equipment is suitable for use in all establishments,
including domestic establishments and those directly connected to public low-
voltage power supply networks that supply buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2| Class A
Voltage fluctuations / Flicker emissions
IEC 61000-3-3| Complies
TABLE 2: Electromagnetic Immunity (IEC 60601-1-2:2007)
This equipment is intended for use in the electromagnetic environment
specified below. It is the responsibility of the user of the Spectra S1 Plus
or Spectra S2 Plus breast pump to ensure that it is used in such an
environment.
Immunity test| IEC 60601 test level| Compliance level| Electromagnetic
environment – Guidance
---|---|---|---
Electrostatic
discharge
IEC 610004-2| ±6 kV contact
±8 kV air| ±6 kV contact
*3 kV air| Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at
At least 30%.
Electric
fast transient /
burst
IEC 6100044| ±2 kV for power supply lines
±1 kV for input/output lines| ±2 kV for power supply lines
±I kV for input/output lines| Mains power quality should be typical of a
commercial or hospital environment.
Surge
IEC 61000-4-5| ±1 kV line(s) to link(s)
1 2 kV link(s) to earth| ±1 kV differential mode
±2 IN common mode| Mains power quality should be typical of a commercial or
hospital environment.
Power frequency
and magnetic
field
IEC 61000-4-8| 3 Alm at 50 and 60 Hz| 3 Mm at 50 and 60 Hz| Power frequency
magnetic fields should be at levels characteristic of a typical location in a
typical commercial or hospital
environment.
Voltage dips and
interruptions
IEC 61000-4-11| <5% UT (>95% dip in UT) for 0.5 cycle 40% UT (60% dip in UT)
for 5 cycles 70% UT (30% dip in UT) for 25 cycles <5% UT (>95% dip in UT) for
5 secs UT = the a.c. mains voltage before
application of the test level.| <5% UT (>95% dip in UT) for 0.5 cycle 40% UT
(60% dip in UT) for 5 cycles 70% UT (30% dip in UT) for 25 cycles <5% UT (>95%
dip in UT) for 5 secs UT = the a.c. mains voltage before application of the
test level.| Mains power quality should be typical of a commercial or hospital
environment. During a power outage, the equipment should be powered from a
battery or other constant, appropriate power supply.
Conducted RF
1EC 610004-6| 3 Vrms
150 kHz — 80 MHz| 3 Vrms
150 kHz — 80 MHz| Portable and mobile RF communications equipment should be
used no closer to the
Spectra SI Plus or Spectra S2 Plus breast pump (including cables) than the
recommended separation distance (d), based on the following equations:
d = 1.24P for 150 kHz — 80 MHz
d = 1.24P for 80 MHz — 800 MHz
d = 2.31P for 800 MHz — 2.5GHz
Radiated RF
1EC 610004-3| 3 V/m
80 MHz — 2500 MHz| 3 V/m
SO MHz – 2500 MHz
| | | P is the maximum output power rating of the transmitter in watts.
d is the recommended separation distance in meters.
Field strength from fixed RF transmitters, as determined by an electromagnetic
site survey, should be less than the compliance level in each
frequency range. b
Interference may occur in the vicinity of equipment marked with the following
symbol.
---|---|---|---
a Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast, and TV broadcast cannot be predicted accurately. If you
wish to use the equipment in the vicinity of fixed RF transmitters and you are
uncertain about the electromagnetic environment, please consider obtaining an
electromagnetic site survey. If the measured field strength in the location
where the equipment is being used exceeds the applicable RF compliance level
above, it is the user’s responsibility to verify that the equipment is
operating normally. If abnormal performance is observed, additional measures,
such as re-orienting or relocating the equipment, may be necessary.
b Over the frequency range 150 kHz to 80MHz, field strengths should be less
than 3 V/m.
TABLE 3: Recommended Separation Distances (IEC 60601-1-2:2007)
This equipment is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. Observing the recommended separation
distances between this equipment and other portable and mobile RF
communications equipment (transmitters) can reduce the likelihood of
electromagnetic interference.
Rated Maximum
Output Power of
Transmitter (W)| Separation Distance (m) Based on Frequency of
Transmitter
---|---
150 kHz – 80 MHz
d= 1.2#| 80 MHz – 800 MHz
d =1.24P| 800 MHz – 2.5GHz
d = 2.3,./P
0.01| 0.12| 0.12| 0.23
0.1| 0.37| 0.37| 0.74
1| 1.| 1.| 2.
10| 4.| 4.| 7.
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the
recommendation separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects,
and people.
Spectra S1 Plus / S2 Plus Breast Pump Instructions for Use
{00169689 2}
Distributed in the US Exclusively by:
Spectra Baby the USA
6851 SW 21St Court, Suite 1 Davie, FL 33317
Website: http://www.spectrababyusa.com
Email:
customerservice@spectrababyusa.com
Telephone :855-446-6622
References