kinetik WELLBEING ET-100A Inner Ear Thermometer User Manual
- June 7, 2024
- kinetik WELLBEING
Table of Contents
- Introduction
- Specification
- Specification
- Parts
- Warning and safety notices
- How to use (°C displays for example)
- Changing from Celsius to Fahrenheit
- Explanation of symbols
- Product disposal
- Cleaning and Storage
- Replacing the batteries
- Guarantee
- EMC Table
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
kinetik WELLBEING ET-100A Inner Ear Thermomet er
Introduction
Please read the instructions carefully before using. The ear thermometer is a device capable of achieving infra-red temperature measurement when placed in the auditory canal of a subject. It is a safe means of measuring human body temperature through the ear.
Specification
The standard we adopted is EN12470-5:2003 Clinical thermometers — Part 5: Performance of infrared ear thermometers (with maximum device). This device complies with the requirements of IEC 60601-1,IEC 60601-1-11 and IEC 60601-1-2.
Range of displayed temperature:
34°C – 44°C (93.2°F – 111.2F) Display L°C (L°F) when the temperature is under
34°C (93.2°F) Display H°C (H°F) when the temperature is over 44°C (111.2°F)
Operating ambient: Temperature:
16°C – 35°C (60.8°F – 95.0°F) Humidity: 80%rh
Storage and transport condition:
Temperature: -10°C to 55°C (14°F to 131°F) Humidity: 30%rh to 90%rh
Display resolution temperature range:
0.1°C (0.1°F)
Accuracy:
±0.2°C (from 35.5°C to 42.0°C) ±0.4°F (from 96.0°F to 107.6°F)
Display:
liquid crystal display, 4 digits temperature value: display the maximum
temperature in measuring process
Specification
temperature unit: centigrade or Fahrenheit display of memory: last ten
memories low battery warning: the LCD display and then a beep sound is heard
Power consumption: 30 mW
Battery: two 1.5V alkalescence batteries (AAA)
Battery life: 4000 takes or 6 months with 3-min. usage per day
Dimension: 134mm by 37mm by 67mm
Net weight: 74g
Beeper sign: on/off, measuring finish and low voltage warning, etc.
Self-testing sequence: Press the ‘on/recall’ button to turn on the
thermometer and all of symbol (See Fig A) should be displayed on the LCD
in one second.
Calibkinetik WELLBEING ET-100A Inner Ear Thermometer ration: This
thermometer is calibrated at the time of manufacture. If the thermometer is
used according to the instructions, periodic re-calibration is not required.
Manufacturing Batch code: see the label within the battery compartment.
Parts
Warning and safety notices
Cautions
- Condition the thermometer at room temperature for 30 minutes before taking a reading.
- Use a new probe cover with each and every measurement to avoid inaccurate readings. Without a probe cover, results will be at least 1°C higher.
- This thermometer MUST be used with a probe cover. Using the product without a probe cover could = result in inaccurate readings.
- The ‘ ‘ symbol on the display indicates the device is ready for use.
- Do not scratch the probe membrane as this will affect the accuracy of measurements.
- If the LCD does not change or respond when you press any of the buttons, remove the batteries and then replace to reset.
- Please don’t use the thermometer if your ear canal has become inflamed. Do not force the thermometer into the ear canal. If pain occurs, stop using immediately.
- Do not expose this thermometer to electric shock.
- Do not expose the thermometer to sunlight or to water.
- Do not use near strong electromagnetic fields, i.e. Keep it away from any radio systems and mobile phones
- Do not modify this device without authorization of the manufacturer.
- This device must always be kept in a clean, dry area
- : Degree of protection against electric shock is Type BF applied part.
- : Battery disposal should be completed in accordance with local regulations.
Note:
- In order to avoid the spreading of germs, use a new probe cover for each measurement.
- Check to ensure probe cover is fitted on firmly before use (please see the diagrams below). If the probe cover is broken, discard and apply a new one immediately.
How to use (°C displays for example)
Measuring human body temperature in the Ear Canal
-
Press the ‘on/recall’ button to turn on the thermometer. After the beep the thermometer will be ready to read when the LCD display matches Fig B.
-
Straighten the ear canal by pulling the outer ear up and back to give a clear view of the eardrum.
● For children under 1 year, Pull the ear straight back.
●For children ages 1 year to adult, Pull the ear up and back.
Newborn-1 year 1 year – adult (Caution: Many pediatricians suggest use of ear thermometers for children older than 6 months.)
-
While pulling the outer ear, insert the probe snugly into the ear canal (insert as deep as is comfortable, do not force, the probe into the ear canal). Press and immediately release the ‘SCAN’ button. Do not remove the thermometer from the ear until you hear the beep sound — this beep means that the measurement is complete. Remove the thermometer from the ear. The display shows the measured temperature
-
Repeat measurements: wait at least 1 minute between measurements. Ensure the Ear symbol is visible in the LCD. How to use (°C displays for example)
-
Memory search: Press the ‘on/recall’ button, the last ten memories (N0.9-N0.0) will take turns to display on the LCD as Fig C and Fig D.
-
The LCD will displays ‘Er 1’ as Fig E and cannot displays temperature\ when the ambient temperature is out of the range of 16°C to 35°C.
Note: Before measuring, the thermometer shall be stabilized at the operating ambient condition for a minimum of 30 min.
-
Shut off: The thermometer shall automatically shut off without manipulating in one minute.
Changing from Celsius to Fahrenheit
To switch the display between °C and °F, turn the unit on. Press and hold the
on/recall button for 3 seconds, the display will show as Fig. F , the unit
will prepare for a measurement.
Explanation of symbols
| battery is empty
---|---
| Product disposal instructions for electronic devices
| The battery in this product complies with the requirements stated in
European Directives 2006/66/EEC
| Model Reference
| Memory
| Type BF equipment
| Symbol for “COMPLIES WITH MDD 93/42/EEC REQUIREMENTS”
| subject temperature is
under 34°C
| subject temperature is
above 44°C
| Read IFU carefully
| Authorised representative in the european community
| Symbol for “MANUFACTURER
| Classification according to the degree of protection against ingress of
water as detailed in IEC 529
| Lot Number
Product disposal
Please ensure environmental protection. Batteries do not belong in the domestic waste. Please hand them in at collection point or the municipal recycle material centre as special waste. This symbol on products and/or accompanying documents means that consumed electronic products must not be mixed with conventional domestic waste. Take these products to the corresponding collection points for correct treatment and recycling, where they will be accepted free of charge. For more information on the close-set collection point, Please enquire with your local authorities.
Cleaning and Storage
-
Store thermometer and the Disposable probe covers in a dry location free from dust and contamination and away from direct sunlight. The ambient temperature at the storage location should remain fairly constant and within the range of -10°C to 55°C.
-
Use an alcohol swab or cotton swab moistened with alcohol (70% lsopropy) to clean the thermometer casing and the measuring probe before each use. Ensure that no liquid enters the interior of the thermometer.
-
Never use abrasive cleaning agents, thinners or gasoline for cleaning and never immerse the instrument in water or other cleaning liquids. Take care not to scratch the surface of the probe membrane or display.
Replacing the batteries
- When voltage of the batteries are low, the LCD will display ‘‘ symbol, please replace two new batteries in unit. The thermometer cannot work accurately under the condition of low voltage.
- The thermometer is supplied with two 1.5V ALKALESCENCE BATTERIES (AAA). Insert new batteries when the low battery symbol appears on the LCD.
- Remove the battery cover and take out the old batteries.
- Place two new batteries according to the ‘+’ or ‘-
- Please take out the batteries to avoid battery leaking if unit not used for over six months.
- The disposal of the probe cover, battery and device shall comply with the local environment requirements. The lithium battery or fuel cell may lead to excessive temperatures, fire or explosion.
Guarantee
This product is guaranteed for a period of one year from the date of purchase against mechanical and electrical manufacturing defects. There are no serviceable parts inside this device. Any attempted repair by unauthorised persons invalidates the warranty. In the unlikely event that you experience a problem, please return it to the retailer where you made the purchase, along with your receipt. This does not affect your statutory rights.
EMC Table
Guidance and manufacturer’s declaration – electromagnetic immunity
The ET-100A Device is intended for use in the electromagnetic environment
specified below.
The customer or the user of the ET-100A Device should assure that it is used
in such an environment.
Emissions test| Compliance| Electromagnetic environment – guidance
RF emissions CISPR 11| Group 1| The ET-100A uses RF energy only for its
internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions CISPR 11| Class B| The ET-100A is suitable for use in all
establishments, including domestic establishments and those directly connected
to the public low-voltage power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions IEC 61000-3-2| Not applicable
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
| Not applicable
The ET-100A Device is intended for use in the electromagnetic environment
specified below.
The customer or the user of the ET-100A Device should assure that it is used
in such an environment.
IMMUNITY test| IEC 60601
test level
| Compliance level| Electromagnetic environment – guidance
Electrostatic discharge (ESD) IEC 61000-4-2| 6 kV contact
8 kV air
| 6 kV contact
8 kV air
| Floors should be wood, concrete or ceramic tile. If floors are covered with
synthetic material, the relative humidity should be at least 30 %.
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8|
3 A/m
|
3 A/m
| Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
Radiated RF IEC 61000-4-3| 3 V/m
80 MHz to
2.5 Ghz| 3 A/m| Portable and mobile RF communications equipment should be used
no closer to any part of the ET-100A Device, including cables, than the
recommended separation distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance :
d 1.2 P 80 MHz to 800 MHz
d 2.3 P 800 MHz to 2,5 GHz
Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m).Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey, a should be less than the
compliance level in each frequency range. b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects
and people.
A. Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted the poetically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the ET-100A Device is used
exceeds the applicable RF compliance level above, the ET-100A Device should be
observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the
ET-100A Device.
B. Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m.
Recommended separation distances between
portable and mobile RF communications equipment and the ET-100A Device
The ET-100A Device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the ET-100A Device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the ET-100A Device as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
W
|
Separation distance according to frequency of transmitter
150 kHz to 80 Mhz
d 1.2 P
| 80 MHz to 800 Mhz
d 1.2 P| 800 MHz to 2,5 Ghz
d 2.3 P
0.01| 0.12| 0.12|
0.23
0.1
| 0.38| 0.38| 0.73
1| 1.2| 1.2|
2.3
10
| 3.8| 3.8| 7.3
100| 12| 12|
23
Hangzhou Hua’an Medical & Health Instruments Co., Ltd
Building 2, 1# Fuzhu Nan Rd, Wuchang Town,
Yuhang District, Hangzhou, Zhejiang 310023 China
Shanghai International
Holding Corp. GmbH (Europe)
Eiffestrasse 80, 20537 Hamburg Germany