kinetik WELLBEING DT-K111D Rapid Flexible tip Digital Thermometer User Manual
- June 6, 2024
- kinetik WELLBEING
Table of Contents
Rapid Flexible tip Digital Thermometer
Model: DT-K111D
0197
Congratulations on your purchase of this product. Please read the instructions carefully before using the thermometer for the first time, and keep these in a safe place. This product is intended for the measurement of human body temperature. this product is for home and hospital use, the operator shall be at least 11 years old and the patient can be the operator.
Operating Instructions
Before using, please disinfect the probe at first. To switch on, press the ON/OFF button next to the display; a short beep will sound, indicating that the thermometer is operational. At the same time, the thermometer runs a self- check test, during which all the digital segments appear on the LCD. When the letters “L” and an ashing “°C” or “°F” display, the thermometer is now ready for use. If the ambient temperature is below 32°C or 89.6°F, then “L°C” or “L°F” will appear on the LCD and if it is more than 42°C or 107.6°F, then “H°C” or “H°F” will appear on the LCD. During the reading, the current temperature is displayed continuously, and the “°C” or “°F” symbol flashes. The measurement is completed when a constant temperature value has been reached. The temperature value is considered constant when the temperature rises less than 0.1°C within 4 seconds. As soon as the constant temperature value is reached, a beep will sound 12 times, and the “°C” or “°F” symbol will stop flashing. The highest temperature measured appears on the LCD. However, please note that this thermometer is a maximum thermometer, i.e. the displayed temperature can increase slightly if measurement continues after the beep, this is particularly the case with axillary measurements, the temperature value should be recorded which approximates the core body temperature. In this instance please note the description under “Methods of measuring temperature”. When the measurement is completed, please switch the thermometer off by pressing the ON/OFF button. After the temperature has been displayed, the thermometer will shut off automatically after 4 minutes and a beep sound.
Memory function
Switch the thermometer on, a short beep will sound. At the same time, the thermometer runs a self-check test, during which all the digital segments appear on the LCD. After that, The last reading value with “°C” or “°F” will appear automatically on the LCD for about 2 seconds. The reading is only over- written when a new temperature value is recorded.
Methods of measuring temperature
It is important to remember that the body temperature reading depends on the
site where it is measured. For this reason, the measurement site must always
be specified in order to ensure that a correct temperature reading is
recorded.
In the rectum(rectal)
This is the most accurate method from a medical point of view because it comes
closest to the core body temperature. Lubricating the thermometer tip with
ungluing agents before use. Do not use vaseline. The thermometer tip is
inserted carefully into the rectum for a maximum of 13 mm. The usual measuring
time is approximately 10 seconds.
Under the arm(axillary)
Placing the thermometer in the armpit provides a measurement of surface
temperature that can fluctuate by around 0.5°C to 1.5°C from rectal
temperature readings in adults. The usual measuring time for this method is
approximately 23 to 29 seconds. It should be noted that an exact reading won’t
be obtained if, for example, the armpit temperature is lower than normal (the
armpit temperature has been cooled down). If in this case, we recommend
extending the measuring time by around 5 minutes in order to obtain the most
precise possible reading that corresponds as closely as possible to the core
body temperature.
In the mouth(oral)
There are different heat zones in the mouth. As a general rule, the oral
temperature is 0.3°C to 0.8°C lower than the rectal temperature. To ensure
that the reading is as accurate as possible, place the thermometer tip to the
left or right of the root of the tongue. The thermometer tip must have
constant contact with the tissue during the reading and be placed under the
tongue in one of the two heat pockets at the back, keep the mouth closed
during the reading and breathe evenly through the nose. Do not eat or drink
anything before the measurement. The usual measuring time is approximately 25
seconds.
Note: We strongly recommend the rectal method as the most accurate method
for identifying the basal temperature, and advise you to extend the measuring
time by 1 Minute after the beep.
How to change the measuring unit
Under the power-off state, press the ON/OFF button for 2 seconds, the display will show “b” and the unit in use. Press the ON/OFF button to switch the unit, Then the thermometer will be operational after 3 seconds
Cleaning and disinfection
The best way to clean the thermometer tip is by applying a disinfectant (e.g. 70% medical alcohol) with a damp cloth. It shall be disinfected before each use. The Waterproof thermometers can be immersed in liquid or lukewarm water for thorough cleaning and disinfection.
Safety precautions
- The incorrect methods of measuring temperature may lead to inaccurate measurement results.
- Do not expose to high temperatures or direct sunlight.
- Do not drop the thermometer. It is neither shock-proof nor impact-resistant.
- Do not modify this device without the authorization of the manufacturer.
- Do not bend or open the device (except the battery compartment).
- Do not clean with thinners, petrol, or benzene. Only clean with water or disinfectant.
- Do not immerse the Waterproof thermometers underwater 15cm for longer than 30 minutes.
- The thermometer contains small parts (battery, battery this reason, do not leave the thermometer unattended in the hands of children
- Avoid bending the thermometer tip by more than45 degrees.
- If the ambient temperature is over 35°C or 95°F, dip the thermometer tip in cold water for approx. 5 to 10 seconds prior to measuring the temperature.
- Persistent fever, in particular in children, has to be treated by a doctor, please get in touch with your doctor!
- Do not use near strong electromagnetic fields, i.e. keep it away from any radio systems and mobile phones.
ELECTROMAGNETIC COMPATIBILITY INFORMATION
This device is suitable for home healthcare environments and professional
healthcare facility environments. WARNING: Use of this equipment adjacent to
or stacked with other equipment should be avoided because it could result in
improper operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating normally.
The essential performance is the digital thermometer can offer the temperature
measurement. Do not use mobile (cellular) telephones and other devices, which
generate strong electrical or electromagnetic fields, near the medical device.
This may result in incorrect operation of the unit and create a potentially
unsafe situation. The recommendation is to keep a minimum distance of 7 m.
Verify the correct operation of the device in case the distance is shorter.
Guidance and manufacturer’s declaration – electromagnetic emissions
The device is suitable for use in the specified electromagnetic environment
and it has met the following standard’s emission requirements.
Phenomenon| Profession healthcare
facility environment| Home healthcare environment
Home healthcare
environment| CISPR 11, Group 1,
Class A or B| CISPR 11, Group 1, Class B
Harmonic distortion| IEC 61000-3-2,
Class A or not applicable| NA
Voltage fluctuations
and flicker| IEC 61000-3-3
or not applicable| NA
Guidance and manufacturer’s declaration – electromagnetic immunity
The device is suitable for use in the specified electromagnetic environment
and it meets the following immunity test levels. Higher immunity levels may
cause the device’s essential performance lost or degraded.
Phenomenon| Basic EMC standard
or test method| Professional healthcare
facility environment| Home healthcare
facility environment
Electrostatic
discharge| IEC 61000-4-2| 4/- 8 kV contact
+/- 2 kV, +/- 4 kV, +/ 8 kV, +/- 15 kV air
Radiated RF
EM fields| IEC 61000-4-3| 3V/m
80MHz-2.7GHz
80%AM at 1 kHz or 2Hz| 10V/m
80MHz-2.7GHz
80%AM at 1kHz or 2Hz
1kHz or 2Hz can be specified by
the manufacturer
Proximity fields from wireless communications equipment| IEC 61000-4-3| See
the RF wireless communication equipment table in recommended minimum
separation distances”.
Rated power frequency
magnetic fields| IEC 61000-4-8| 30A/m; 50 Hz or 60Hz
Electric fast
transients bursts| IEC 61000-4-4| NA
For input a.c. power port
d.c. power lines or signal input/output lines whose length exceeding 3m
Surges| IEC 61000-4-5| NA
For 1. input a.c. power port;
2.all d.c. power ports connected permanently to cables >3m
3. output signal output lines connected directly to outdoor cables
Conducted
disturbances
induced by
RF fields| IEC 81000-4-8| NA
---|---|---
For 1, input ac. power pat:
2.all d.c. power ports connected permanently to cables >3m
3.all patient-coupled cables
4.SIP/SOP whose maximum cable length z 3m
Voltage dips| IEC 61000-4-11| NA
Voltage
interruptions| IEC 81000-4-11| NA
UT: rated voltage(s): Eg. 25/30 cycles means 25 cycles at 50Hz or 30 cycles at
60Hz
Recommended mini num separation &stances
Nowadays. many RF wifeless equipment is being used in various healthcare
locations where medical equipment and/or systems are used. When they are used
in close proximity to medical equipment and/or systems. the medical equipment
and/or systems’ basic safety and essential performance may be affected. This
device has been tested with the immunity test level in the below table and
meets the related requirements of IEC 60601-1-2:2014. The customer and/or user
should help keep a minimum distance between RF wireless communications
equipment and this device as recommended below.
Test frequency
(MHz/| Band
(MHz)| Service| Modulation| Maximum| Power (W/
Distance
On)| Immunity
test level
pi ing
385| 380-390| TETRA 400| Pau/
moderation
18Hz| 2.| 0.3| 27
450| 430-470| GMRS 460
FRS 480| FM k 5 kHz
deviation
1 kHz sine| 2| 0.3| 28
710| 704-787| LTE Band
13. 17| Rise
modulation
217Hz| 0.2| 0.3| 9
745
780
810| 800-960| GSM 800900. TETRA800. OEN 820. COMA 850. LTE Band 5| Pulse
modjation
12Hz| 2| 0.3| 28
870
930
1720| 1700-1990| GSM 1800
COMA 1900:
GSM 1900.
DECT:
LIE Band 1, 3
4, 25: MATS| Pulse
modulation 217Hz| 2| 0.3| 28
1845
1970
2450| 2400-2570| Bluetooth,
WLAN. 802.11 WO. RRD 24.%. LTE Band 7| Pulse
moddation
217Hz| 2| 0.3| 28
5240| 5100-5800| WLAN 802.11
ern| Pulse
moddation
217Hz| 0.2| 0.3| 9
5500
5785
Battery replacement
The battery is low voltage and needs to replace when the” ” battery symbol
appears on the right of The LCD Remove the battery cover. Pull out the inner
frame gently, then take out the battery with a toothpick or tweezers, replace
the battery of the same type. The electrode of the battery: the “+” sign up
and “-” sign down. We advise you to remove the batteries if the device is not
going to be used for a longer period of time. The backlight of different
thermometers will have differences, it may be related to the battery power or
the versions of the thermometer, attribute normal phenomenon.
Product disposal
Please ensure environmental protection. Batteries do not belong to domestic
waste. Please hand them in at the collection point or the municipal recycle
material center as special waste. The lithium battery or fuel cell may lead to
excessive temperatures, fire, or explosion.
This symbol on products and/or accompanying documents means that consumed
electronic products must not be mixed with conventional domestic waste. Take
these products to the corresponding collection points for correct treatment
and recycling, where they will be accepted free of charge. For more
information on the closest collection point, Please enquire with your local
authorities.
Technical data
Type: maximum thermometer
Measurement range: (32.0~42.0)°C/(89.6~107.6)°F
Measurement accuracy:
+/- 0.1°C/0.2°F (35.5°C~42.0°C/95.9°F~107.6°F)
+/- 0.2°C/0.4°F (32.0°C~35.5°C/89.6°F~95.9°F)
Storage/transportation temperature: (-25~55)°C, ≤95%RH
Ambient temperature during use: (5~35)°C, ≤Min Scale: 0.1°C/0.1°F
Battery type:
Lithium battery, type CR 1632, 3V, service life minimum
100 hours under continuous operation.
Weight: Approx. 19g
Display: Illuminated display
Atmospheric pressure: 700~1060hPa
Explanation of symbols
or Battery is empty
Manufactured by
Type BF equipment
Manufacture date
Product disposal instructions for electronic devices
Read the instructions (actual symbol colors are white on a blue background)
Classification according to the degree of protection against ingress of water
as detailed in IEC 60529
The battery in this product complies with the requirements stated in European
Directives 2006/66/EEC
CE conformity marking
Stand by
L°C or L°F: temperature under 32°C or 89.6°F
H°C or H°F: temperature over 42°C or 107.6°F
Legal requirements and guidelines
This product complies with the European Directive for Medical Devices
93/42/EEC and carries the CE mark. The device also complies with the
specifications of the European Standard EN 12470-3 Clinical thermometers –
Part 3: Performance of compact electrical thermometers (non-predictive and
predictive) with the maximum device. The CE marking confirms that this is a
medical device with a measuring function in the sense of the Medical Devices
Act which has undergone a conformity assessment procedure. A Notified Body
confirms that this product fulls all the appropriate statutory regulations.
Calibration check
This thermometer is initially calibrated at the time of manufacture. If this
thermometer is used according to the operation instruction, periodic re-
adjustment is not required. The calibration check has to be carried out
immediately if there are indications that the product does not keep the
defined error limits or the calibration properties could have been affected by
intervention or by any other means. Please also observe any national statutory
regulations. The calibration check can be carried out by competent authorities
or by authorized service providers. A test instruction for calibration check
can be provided to the relevant authorities and authorized services providers
on request.
Support
Our manual should provide you with all the information you need to set up and
use this product.
If you have a question, have a look at our Troubleshooting page!
For further assistance, why not contact our Customer Care team directly? We’re
here to help!
Our Customer Care team is available from 9 am-5 pm, Monday to Friday
(excluding bank holidays).
We promise to respond to all queries and will ensure to resolve any issue you
may be having.
You can reach us by…
Phone (UK):
+44 1483 937967
Post (UK):
Kinetik Medical Devices Limited
Unit 3, Perrywood Business Park, Honeycrock Lane,
Salfords, Redhill. RH1 5DZ| Phone (NZ):
(09) 801 0627
Post (NZ):
Kinetik Wellbeing NZ
44 John Brian Drive, 0794 Redvale, Albany,
Auckland, NewZealand
---|---
Live Chat:
Simply visit www.kinetikwellbeing.com and
send us a message.
Email: customercare@kinetikwellbeing.com
Hangzhou Hua’an Medical & HealthInstruments Co., Ltd
Building 2, 1# Fuzhu Nan RD, Wuchang Town,
Yuhang District, Hangzhou, Zhejiang 310023 China
Shanghai International Holding Corp. GmbH (Europe)
Eiffestrasse 80, 20537 Hamburg Germany
E-Mail: shholding@hotmail.com
Tel.: 049-40-2513175 Fax.: 049-40-255726