GIMA M28243 Super Vega Battery Suction Unit User Manual

June 6, 2024
GIMA

GIMA logo

PROFESSIONAL MEDICAL PRODUCTS

ASPIRATORE SUPER VEGA BATTERY SUPER VEGA BATTERY SUCTION UNIT ASPIRATEUR SUPER VEGA BATTERY SAUGER SUPER VEGA BATTERY ASPIRADOR SUPER VEGA BATTERY ASPIRADOR SUPER VEGA BATTERY SSAK AKUMULATOROWY SUPER VEGA BATTERY ASPIRATOR SUPER VEGA BATTERY ΣΥΣΚΕΥΗ ΑΝΑΡΡΟΦΗΣΗΣ SUPER VEGA BATTERY

28243

GIMA M28243 Super Vega Battery Suction Unit - Gima
S.p.A.Gima S.p.A.
Via Marconi, 1 – 20060 Gessate (MI) Italy
gima@gimaitaly.comexport@gimaitaly.com
www.gimaitaly.com
Made in ItalyGIMA M28243 Super Vega Battery Suction Unit -
0476

SUPER VEGA BATTERY SUCTION UNIT is a suction pump particularly suited for use in hospital wards,
in patients with a tracheotomy, in small surgical applications, and in post- operative treatments at home.
A device that can be used for nasal, oral, and tracheal aspirations in adults or for body liquids in children (for example mucus, phlegm, and blood). A device designed to offer ease-of-transport and almost continuous use thanks to the adoption of an electronic system to manage the power supply. The luminous LED positioned on the front panel indicates its activation and the charge status of the device. Its body is made from plastic with high thermal and electrical insulation in compliance with recently introduced European safety regulations. Supplied with sterilizable polycarbonate jug with overflow valve. Features a suction regulator and vacuum gauge located on the front panel.

GENERAL WARNING

Read the instruction manual carefully before use. The device is for use by qualified personnel (surgeon / professional nurse/assistant) The use of the device at home is restricted to an adult in full possession of mental faculties and/or home carers
The instrument must not be disassembled. For technical service always contact Gima S.p.A.

IMPORTANT SAFETY RULES

  1. Check the condition of the unit before each use. The surface of the unit should be carefully be inspected for visual damage. Check the mains cable and do not connect to power if the damage is apparent;

  2. Before connecting the appliance always check that the electric data indicated on the data label and the type of plug used, correspond to those of the mains electricity to witch it’s to be connected;

  3. Respect the safety regulations indicated for electrical appliances and particularly:
    – Use original components and accessories provided by the manufacturer to guarantee the highest efficiency and safety of the device;
    -The device can be used only with the bacteriological filter;
    – Never immerse the appliance into the water;
    – Do not place or store the aspirator in places where it may fall or be pulled into the bathtub or washbasin.
    In the event it is accidentally dropped, do not attempt to remove the device from the water whilst the plug is still connected: disconnect the mains switch, remove the plug from the power supply and contact the
    GIMA technical service department. Do not attempt to make the device work before it has been thoroughly checked by qualified personnel and/or the
    GIMA technical service department.
    -Position the device on stable and flat surfaces in a way that the air inlets on the back aren’t obstructed;
    – To avoid incidents, do not place the aspirator on unstable surfaces, which may cause it to accidentally fall and lead to a malfunction and/or breakage. Should there be signs of damage to the plastic parts, which may expose the inner parts of the energized device, do not connect the plug to the electrical socket? Do not attempt to make the device work before it has been thoroughly checked by qualified personnel and/or the GIMA technical service department.
    – Don’t use in the presence of inflammable substances such as anesthetic, oxygen, or nitrous oxide;
    – Don’t touch the device with wet hands and always prevent the appliance from coming into contact with liquids;
    – Don’t leave the appliance connected to the power supply socket when not in use;
    – Don’t pull the power supply cable to disconnect the plug remove the plug from the mains socket correctly;
    – Store and use the device in places protected against the weather and far from any sources of heat. After each use, it is recommended to store the device in its own box away from dust and sunlight.
    – In general, it is inadvisable to use single or multiple adapters and/or extensions. Should their use be necessary, you must use ones that are in compliance with safety regulations, however, take care not to exceed the maximum power supply tolerated, which is indicated on the adapters and extensions.

  4. For repairs, exclusively contact technical service and request the use of original spare parts. Failure to comply with the above can jeopardize the safety of the device;

  5. Use only for the purpose intended. Don’t use for anything other than the user defined by the manufacturer. The manufacturer will not be responsible for damage due to improper use or connection to an electrical system not complying with current regulations.

  6. The medical device requires special precautions regarding electromagnetic compatibility and must be in- stalled and used in accordance with the information provided with the accompanying documents: the SU-
    PER VEGA BATTERY SUCTION UNIT device must be installed and used away from mobile and portable
    RF communication devices (mobile phones, transceivers, etc.) that may interfere with the said device.

  7. Instrument and accessory discharging must be done according to current regulations in the country of use.

  8. WARNING : Do not change this equipment without the permission of the manufacturer GIMA S.p.A. None of the electric or mechanical parts have been designed to be repaired by customers or end-users. Don’t open the device, do not mishandle the electric/mechanical parts. Always contact technical assistance

  9. Using the device in environmental conditions different than those indicated in this manual may harm seriously the safety and the technical characteristics of the

  10. The medical device is in contact with the patient by means of a disposable probe (not supplied with the
    device). Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO 10993-1 standard on material biocompatibility.

  11. The product and its parts are biocompatible in accordance with the requirements of regulation EN 606011.

  12. Operation of the device is very simple and therefore no further explanations are required other than those indicated in the following user manual.

  13. The lead battery integrated into the device is not to be considered as ordinary domestic waste. Such a component must be disposed of in a specific collection center in order to be recycled.

  14. Use in Home-Care: Keep all accessories of the device out of reach of children under  36 months of age since they contain mall parts that may be swallowed.

  15. Do not leave the device unattended in places accessible to children and/or persons not in full possession of mental faculties as they may strangle themselves with the patient’s tube and/or the power cable.

The manufacturer cannot be held liable for accidental or indirect damages should the device be modified, repaired without authorization or should any of its components be damaged due to accident or misuse. Any minimal modification/repair on the device voids the warranty and does not guarantee compliance with the technical requirements provided by the MDD 93/42/EEC (and subsequent changes) and it’s normative.

CONTRAINDICATIONS

– Before using the ASPIRATOR, consult the instructions for use: failure to read all the instructions in this manual can be harmful or the patient.
– The device cannot be used to drain chest fluids;
– The device must not be used for suction of explosive, corrosive, or easily flammable liquids.
– ASPIRATOR is not suitable for MRI. Do not introduce the device in MRI environments.

TECHNICAL CHARACTERISTICS

Model SUPERGA BATTERY 230/12V
Typology (MDD 93/42/EEC) Class Ila Medical device
UNI EN ISO 10079-1 Classification HIGH VACUUM / LOW FLOW
Power Feeding 14V — — — 4A with AC/DC adapter (input: 100-240V— –

 50/60Hz – 100VA) or Internally powered equipment (Pb Battery 121.1= 4A) or with cigarette lighter adapter (12V = 4A)
Maximum Suction Pressure (without jar)| -75kPa (- 0.75 bar)
Minimum Suction Pressure (without jar)| Less -25kPa (-0.25 bar)
Maximum Suction Flow (without jar)| 161 /min
Weight| 3.50 Kg
Insulation Class (when used with the AC/DC adapter)| Class II
Insulation Class (when used with an internal battery)| Internally Powered Equipment
---|---
Insulation Class (when used with a car cigarette lighter cable)| Class II
Size| 350x 210x 180 mm
Battery Holding Time| 80 minutes
Battery Time Charge| 240 minutes
Accuracy of Vacuum Indicator| ±5%
Working Condition| Room temperature: 5 + 35°C
Room humidity percentage: 10 + 93 % RH Atmospheric pressure: 800 + 1060 hPa
Conservation condition and Transport| Room temperature: – 25 +70°C
Room humidity percentage: 0 + 93% RH Atmospheric pressure: 500 + 1060 hPa

The technical specifications may change without notice

CLEANING OF THE DEVICE

Use a soft dry cloth with not – abrasive and not – solvent detergents. To clean the device’s external parts always use a cotton cloth dampened with detergent. Don’t use abrasive or solvent detergents. Before carrying out any cleaning and/or maintenance operation, disconnect the appliance from the power supply, unplug it or turn off the switch on the device.

Particular care should be taken to ensure that the internal parts of the equipment do not get in touch with liquids. Never clean the equipment underwater.

During all clearing operations use protective gloves and apron (if need be, also wear a face mask and glasses) to avoid getting in contact with contaminated substances (after each utilization cycle of the machine).

ACCESSORIES SUPPLIES

  • Complete aspiration jar 1000ml
  • Conical fitting
  • Tubes set 6 mm x 10 mm
  • Hydrophobic and antibacterial filter
  • AC/DC adapter
  • Power supply cord for ac/dc adapter
  • Cigarette lighter cable

Antibacterial and hydrophobic filter: designed for a single patient with the aim of protecting patient and machine against infections cross. It blocks the passage of the liquids that come into contact with it. Always replace it if yes suspected to be contaminated and/or wetted or discolored. If the aspirator is used on patients in pathological situations not known and where it is not possible to assess any indirect contamination, replace the filter after each use. The filter is not designed to be contaminated,   disassembled, and/or sterilized. In case the pathology of the patient and/or is known where there is no danger of indirect contamination, we recommend replacing the filter after each work shift or otherwise every month even if the device is not used.On request, versions with a complete 2000ml jar are also available

Suction catheter: Single-use device to be used on a single patient. Do not wash or re-sterilize after use. Reuse may cause cross infections. Don’t use after the lapse of the sell-by date

WARNING : Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO 10993-1 standard on material biocompatibility.

Aspiration jar : The mechanical resistance of the component is guaranteed up to 30 cycles of cleaning and sterilization. Beyond his limit, the physical- chemical characteristics of the plastic material may show signs of decay. Therefore, we recommend that you change it.

Silicone tubes: the number of cycles of sterilization and/or cleaning is strictly linked to the employment of the said tube. Therefore, after each cleaning cycle, it is up to the final user to verify whether the tube is suitable for reuse. The component must be replaced if there are visible signs of decay of the material constituting the said component.

Conical fitting: the number of cycles of sterilization and the number of cleaning cycles is strictly linked to the employment of the said component. Therefore, after each cleaning cycle, it is up to the final user to verify whether the fitting is suitable for reuse. The component must be replaced if there are visible signs of decay in the material constituting the said component.
The service life of the device : More than 1000 hours of operation (or 3 years) in accordance with the standard conditions of testing and operation. Shelf life: maximum 5 years from the date of manufacture.

CLEANING OF ACCESSORIES

Before using the device, the manufacturer advises you to clean and/or sterilize the accessories. Washing and/or cleaning the  autoclavable jar as to be carried out as follows:

  • Wear protective gloves and apron (glasses and face mask if necessary) to avoid contact with contaminating
    substances;

  • Disconnect the tank from the device and remove the said container from the support of the device.

  • Separate all the parts of the cover (overflow device, washer).

  • Disconnect all tubes from the jar and the protection filter

  • Wash each part of the container from secretions under cold running water and then clean every single part
    in hot water (temperature not exceeding 60°C)

  • Once again, carefully wash every single part using, if necessary, a non-abrasive brush to remove any deposits.
    Rinse with hot running water and dry all parts with a soft cloth (non- abrasive). It is possible to wash with commercial disinfectants by carefully following the instructions and dilution values supplied by the manufacturer.

After cleaning, leave the parts to dry in an open, clean environment.

  • Dispose of the aspiration catheter according to that provided by local laws and regulations.
    The silicone aspiration tubes and the conical fitting may be carefully washed in hot water (temperature must not exceed 60°C). After cleaning, leave the parts to dry in an open, clean environment.

When cleaning is complete, reassemble the container for liquid aspirations according to the following procedure:

  • Place the overflow valve into its seat in the cover (under VACUUM connector)
  • Insert floating valve keeping the o-ring towards the opening of the cage
  • Place the o-ring into its seat around the cover
  • After completing assembling operations always make sure that covers seals perfectly to avoid vacuum leak-
    ages or liquid exit

The jar and the cover can be autoclaved by placing the parts into the autoclave and running one sterilization stem cycle at 121°C (1 bar relative pressure – 15 min) making sure that the jar is positioned upsidedown.
Mechanical resistance of the jar is guaranteed up to 30 cycles of sterilization and cleaning at the indicated conditions (EN ISO 10079-1). Beyond this limit, the physical-mechanical characteristics of the plastic may decrease and replacement of the part is therefore recommended.
After sterilization and cooling at an environmental temperature of the parts make sure that these are not damaged.
The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative pressure – 15 min). The conical connector can be sterilized on autoclave using a sterilization cycle at 121°C  (1 bar relative pressure – 15  in).

DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER

PERIODICAL MAINTENANCE CHECKS

PERIODICAL MAINTENANCE CHECKS The SUPER VEGA BATTERY SUCTION UNIT suction equipment does not need maintenance or lubrication. It is, however, to inspect the unit before each use. With regard to training, given the information contained in the user manual and since it is easy to understand the said device, it doesn’t appear to be necessary. Unpack the instrument and always check the integrity of plastic parts and  AC/DC  switching adapter,  feeding cable,  they might have been damaged during previous use.  Connect the cable to the electrical network and turn the switch on. Close the aspirator outlet with your finger and with the suction regulator in maximum vacuum position check that the vacuum indicators reach -75kPa (-0.75   ar) minimum  (internal battery). Rotate the knob from right to left and check the aspiration regulating control.  The vacuum indicator should go down  -25kPa (-0.25  bar). Verify that loud noises are not present,  these can indicate wrong functioning.  The device is protected by a  safety fuse  (F  10A L  250V) situated in the cigarette lighter cable.  When replacing, always check the type and value as indicated.
Internally, the device is protected  (see electrical specifications)  by two fuses F1,  F2  (T  15A  L  125V) that cannot be reached from the outside. Therefore,  contact the manufacturer to request the assistance of an authorized and qualified technician when they need to be replaced.  If it’s replaced make sure that its replacement is always the same type and value, as indicated.   The device is made up of a  lead battery that cannot be accessed by outside. In order to replace it, consult the technical staff authorized by the manufacturer.

USE ONLY THE RECOMMENDED BATTERIES FROM GIMA. THE USE OF OTHER BATTERIES
ARE NOT RECOMMENDED AND INVOLVES THE CANCELLATION OF THE WARRANTY

In the event that the service personnel has to replace the internal battery, pay special attention to the polarity
of the same component. The + / – polarities are indicated directly on the battery.

If the overfill security system it’s activated, don’t proceed with the liquid aspiration. If the overfill security system doesn’t work  here are two cases:

Fault type

| Cause|

Solution

---|---|---
1. Red light on| Battery run down| Hook up the power cord to the electricity mains, positioning the equipment power switch on 0.
2. No light| Defective AC/DC adapter or technical internal problem| Contact the technical service.
3. No aspiration| Jar Cap badly screwed down| Unscrewed the cap then re-screw it correctly
4. No aspiration| The lid seal is not in its seat| Unscrew the cap and insert the seal properly in it seat
5. The float doesn’t close| if the cap has been washed, ensure that the float is not partially detached| Insert the float into its place
6. The float doesn’t close| The float it’s covered by dirty material| Unscrewed the cap, leave the and put in on au­toclave
7. Low suction| The foam inside the jar| Fill the jar to 1/3 full of ordinary water
8.   No aspiration due to flow leakage of mucus| Filter blocked| Replace filter
9.The Vacuum power on the patient side is either very low or absent| · Vacuum regulator set to minimum
· Protection filter blocked or damaged
· Connection tubes blocked, kinked or disconnected
·Shut-off valve blocked or damaged
·Pump motor damaged| · Turn the vacuum regulator clockwise and
check the value of the vacuum on the gauge
·Replace the filter
·Replace or reconnect the tubes, check the jar connections
·Empty the jar, or disconnect the tube from the jar and unblock the shut-off valve. The unit will only work in the upright position
· Contact the technical service
10. Noisy| Technical internal problem| Contact the technical service
Faults 1 -2-3- 4-5-6-7 -8-9-10

| None of the remedies has achieved the desired results| Contact the seller or GIMA After-sales Assistance Service

INSTRUCTION FOR USE **

**

GIMA S.p.A. will provide upon request electric diagrams, components list, descriptions, setting instructions, and any other information that can help the technical assistance staff for product repair. 1st case – If the overfill security system doesn’t work the aspiration will be stopped by the bacteriological filter
who avoid the liquid penetration inside the device.
2nd case – If both the security system and the bacteriological filter do not work, there is the possibility that liquid has leaked inside the device, in this case, return the device to GIMA technical service.

  • The device must be checked before each use in order to identify any operating faults and/or damage due to transport and/or storage.
  • The work surface must be flat and stable in order that the control panel can be reached and so that the vacuum gauge, vase, and antibacterial filter are clearly visible.
  • It is recommended not to hold the device in your hand and/or avoid prolonged contact with the body of the device.

WARNING: For correct use, position the suction unit on a flat and stable surface, thus maximizing the available volume of the vase and improving the efficiency of the overflow device.

Operation with AC/DC power pack:

  • Connect the short silicone tube with the antibacterial filter to the suction port. The other tube, connected to
    the filter on one end, must be connected to the spout on the vase lid with the float assembled inside (overflow
    device). The overflow device is triggered (the float closes off the internal lid fitting) when the maximum volume (90% of the effective vase volume) is reached, thus preventing the liquid from penetrating the inside of the machine. The device must be used n a flat, horizontal surface.

  • Connect the long silicone tube to the free spout on the lid; the free end of the tube must be connected to the conical fitting for the probe coupling, to which the suction probe must then be connected.

  • Connect the universal power pack to the device using the dedicated connector and insert the power cord plug into the socket. To start the treatment, press the switch into position I to turn the device on

  • Set the desired vacuum level (Bar / kPa) through the vacuum regulator. Turn the knob in a clockwise direction to increase the vacuum level: these values can be read on the “vacuum gauge”.

  • To suspend and/or end the treatment, press the switch again and pull the plug out of the socket

  • To mitigate the formation of foam inside the vase, unscrew and remove the lid from the vase, and fill the
    latter with 1/3 water (to facilitate cleaning operations and speed up depressurization during operation), then screw the lid back into the vase.

  • Remove the accessories and proceed with cleaning operations.

  • At the end of each use place the device back in the box, protected against dust

WARNING: The power cord plug is the element of separation from the electrical mains, even if the device is equipped with an on/off button. Once the device is in use, the power plug must remain accessible to allow another method of disconnection from the electrical mains.

Filter assembling

GIMA M28243 Super Vega Battery Suction Unit - Filter
assembling

Make sure the filter is assembled with the arrows on the side of the patient.
WARNING : The inside of the medical device must be regularly checked for the presence of liquids or other visible contamination secretions). In the presence of liquids or other visible contamination, immediately replace the medical device due o the risk of an insufficient vacuum flow rate. These products have been designed, tested, and manufactured exclusively for single patient use and for a period no longer than 24 hours.

Operation using cigarette lighter DC 12V

  • Connect the device’s external plug 12V to the lighter plug with the cigarette lighter cable. Check the battery power status of the vehicle before the cigarette lighter cable. Press the switch to start suction
  • Press the switch to the I position to turn it on.
    WARNING : Only use the originally supplied or recommended replacement cigarette lighter cables (view the chapter “Important Safety Rules”)

Operation with Internal Battery

  • Press the switch in the position I to turn the device on (the external power pack doesn’t need to be connected)
  • The fully charged battery life is about 60 minutes with continuous operation.

WARNING: Before using the device, check the battery power status. Before each use proceed with charging the battery.  o, maintain the device in good conditions, recharge the battery every 3 months (when not in use).

Recharging operations: to be able o charge the internal battery it is necessary to connect the universal switching adapter to the electric network for approx. 240  minutes with the main switch to position 0.

TAB. I – INDICATOR LIGHTS DURING OPERATIONS
When an external power supply is connected (regardless of the state of the battery charger) and when the
the device is working (after having turned it on), the LED stays in a FIXED GREEN position.

LED Signal Phase Problem / Cause Solution
Flashing Green Led During recharge Battery recharge running Wait
Steady Greed Led During recharge Recharging cycle complete Remove power

supply
Steady Red Led| During battery operation| Flat battery| Start recharging cycle

WARNING: During this signal, you will hear a long, continuous beep (duration of sound 0.8 sec/sound frequency: every 8.5 sec), which notifies the user regarding the battery dis­charge.

Flashing Red Led| The device automatically turns off when the battery is flat| Battery completely flat| When the device is restarted the LED will flash red: begin the battery recharge cycle immedi­ately
Steady Orange Led| During battery operation| Intermediate status| Guaranteed battery function / Recharge when the red LED signal comes on.

GIMA M28243 Super Vega Battery Suction Unit - -1

NEVER USE THE DEVICE WITHOUT A JAR AND / OR PROTECTION FILTER

RISK OF ELECTROMAGNETIC INTERFERENCE AND POSSIBLE REMEDIES
 This section contains information regarding the conformity of the compliance with the EN 60601-1-2 Standard (2015). The SUPER VEGA BATTERY SUCTION  NIT surgical aspirator is an electro-medical device that requires particular precautions regarding electromagnetic compatibility and which must be installed and commissioned according to the electromagnetic compatibility information supplied. Portable and mobile radio communication devices (mobile phones, transceivers, etc.) may interfere with the medical device and should not be used in close proximity with, adjacent to, or on top of the medical device. If such use is necessary and unavoidable, special precautions should be taken so that the electro-medical device functions properly in its intended operating configuration (for example, constantly and visually checking for the absence of anomalies or malfunctions). The use of accessories, transducers, and cables differing from those specified, with the exception of transducers and cables sold by the appliance and system manufacturer as spare parts,  can lead to an increase in emissions or in a decrease of the immunity of the device or system. The following tables supply information regarding the EMC (Electromagnetic compatibility) characteristics of the electro-medical device.

Guidance and manufacturer’s declaration — electromagnetic Emissions

The SUPER VEGA BATTERY SUCTION UNIT is intended for use in the electromagnetic environment specified below. The customers or the user of the SUPER VEGA BATTERY SUCTION UNIT should make sure that it’s used in such an environment.
Emissions test| Compliance| Electromagnetic environment – guidance
Irradiated / Conducted emissions CISPR11| Group 1| The SUPER VEGA BATTERY SUCTION UNIT only used RF energy only for its internal functioning. Therefore, its RF emissions are very low and are not cause interference in the proximity of any Electronic ap­pliances.
Irradiated / Conducted emissions CISPR11| Class [8]| The SUPER VEGA BATTERY SUCTION UNIT can be used in all environments, including domestic and those connected directly to the public mains distribution that supplies power to environments used for domestic scopes.
Harmonic emissions EN 61000-3-2| Class [A]
Voltage fluctuations / flicker emissions EN 61000-3-3| Complies
Guidance and manufacturer’s declaration – Immunity Emissions

The SUPER VEGA BATTERY SUCTION UNIT is intended for use in the electromagnetic environment specified below. The customers or the user of the SUPER VEGA BATTERY SUCTION UNIT should make sure that it’s used in such an environment.
Immunity Test| Level Indicated by the EN 60601-1-2| Compliance Level| Electromagnetic environment guidance
Electrostatic

discharge (ESD) EN 61000-4-2

| +1-8kV on contact +1-15kV in air| The device doesn’t change its state| Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient / burst EN 61000-4-4| ± 2kV power supply lines

t 1kV for input 1

output lines

| The device doesn’t change its state| The main power quality should be that of a typical commercial environment or hospital.
Surge

EN 61000-4-5

| t 1kV differential mode +1-2 kV ordinary mode| The device doesn’t change its state| The main power quality should be that of a typical commercial environment or hospital.
Loss of voltage, brief voltage interruptions and variations

EN 61000-4-11

| 6% Ur (>95% dip UT) for 0,5 cycle

40 % Ur (60% dip Ur) for 5 cycle

70 % UT (30% dip UT) for 25 cycle  6 % Ur (>95%6p Ur) for 5 sec

| – –| Mains power quality should be that of a typi­cal commercial environment or hospital If the user of the SUPER VEGA BATTERY SUC­TION UNIT request that the appliance oper­ates continuously, the use of a continuity unit is recommended.
A magnetic field with network frequency (50/60 Hz)

EN 61000-4-8

| 30A/m| The device doesn’t change its state| The power frequency magnetic field should be measured in the intended installation location to make sure that its sufficiently low.
Note UT is the value o the power supply voltage
Guidance and manufacturer’s declaration – Immunity Emissions

The SUPER VEGA BATTERY SUCTION UNIT is intended for use in the electromagnetic environment specified below. The customers or the user of the SUPER VEGA BATTERY SUCTION UNIT should make sure that it’s used in such an environment.
Immunity Test| Level indicated by the EN 6C601-1 -2| Compliance Level| Electromagnetic environment-guidance
Conducted Immunity 61000-44
Radiated
Immunity 61000-4-3| 3Vrms ISOldiz to 80Mhz (for non-He- supporting devices) 10V/m 80MHz to 2.7GHz (for non-kle- supporting devices)| V1 = 3 V RMS
El =10Vim| The portable and mobile RF communication devices. including cables. must not be used closer to the SU­ PER VEGA BATTERY SUCTION UNIT device. then the separation distance calculated by the equation applica­ble to the transmitter frequency.
Recommended separation distance d= 1 t Vii
d = pipTI  from 80 MHz to 800MHz
Et
d =[ 21,ff, from 800 MHz to 2.7 G Hz
E. Where P is the maximum nominal output voltage of the the transmitter in Watt (W) depending on the manufactur­ er of the transmitter and the recommended separation distance in meters (m). The intensity of the field from the feed RF transmitters. as determined by an elec. tro- magnetic study d the site , . could be lower than tying evel c4 conformity of each frequency interval ‘. It is possible to check for interference in property to devices identified by the fc4- 4 1 0 liming symbols: 4

icon

Note 1: At 80 MHZ and 800 MHz the interval with the Nghest frequency is &OW

Note 2: These guidelines may not be applicable in all situations. The Sofa-magnetic propagation is influenced by the absorption and by reflection from buildings. objects and people.

a) The field intensity for fixed transmitters such as the base stations for radiotelephones (mobile and excess) and terres­ trial mobile radio. amateur lad, devices. radio AM and FM transmitters and N transmitters can not be theoretically and accurately foreseen. To establish a demo-magnetic environment generated by fixed RF transmitters. an electromagnet. lc study of Me site should be considered. If the field intensity measured in the place where the device will be used surpass­es the above- mentioned applicable level of conformity. the normal functioning of the device should be monitored. if abnor­ mal performance arises. additional measures such as changing the device’s direction or positioning may be necessary.

b) The field intensity on an interval frequency of t SO kHz to 80 MHZ should be less than 10 Vim.

The recommended separation distance between portable and mobile radio- communication devices and the monitor

The SUPER VEGA BATTERY SUCTION UNIT surgical aspirator is intended to operate in an electro-magnetic envi­ronment where RF irradiated interferences are under control. The client or operator of the SUPER VEGA BATTERY SUCTION UNIT device can help prevent electromagnetic interference by keeping a minimum distance between the portable and mobile RF communication devices (transmitters) and the SUPER VEGA BATTERY SUCTION UNIT device, as recommended below, in relation to the radio-communication maximum output power.
The maximum nominal output power of the Transmitter W| The separation distance from the frequency transmitter (m)
150KHz to

d.1 -5 :11/73
Lir ,

| 80MHz| 80MHz to 800MHz

d.[121w,

E’

| 800MHz to

_dItIVT.

LE’ J

| 2,7GHz
0,01| 0,12| 0,12| 0,23
0,1| 0,38| 0,38| 0,73
1| 1,2| 1,2| 2,3
10| 3,8| 3,8| 7,3
100| 12| 12| 23
For transmitters with a maximum nominal output power not shown above, the recommended separation distance in meters (m) can be calculated using the equation applicable to the transmitter frequency, where P is the maximum nominal output power of the transmitter in Watt (W) depending on the transmitter’s manufacturer.

Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied

Note 2: These guidelines may not be applicable in all situations. The electromagnetic propagation is influenced by the absorption and by the reflection from buildings, objects, and people

SYMBOLS

| Caution: read instructions (warnings) carefully| icon
12| Follow instructions for use
---|---|---|---
| Keep in a cool, dry place| | Keep away from sunlight
| Manufacturer| | Date of manufacture
| Product code| | Lot number
| Medical Device complies
with Directive 93/42/EEC| | Type BF applied part
| WEEE disposal| | Class II applied
| Serial number| | Temperature limit
| Battery| | Atmospheric Pressure limit
| Direct current| | Alternating current
| Covering Protection rate| | Mains frequency
| ON / OFF| icon 22| Humidity limit

Disposal: The product must not be disposed of along with other domestic waste. The users must
dispose of this equipment by bringing it to a specific recycling point for electric and electronic equipment.

DISPOSAL OF WASTE BATTERIES – (Directive 2006/66/EC) This symbol on the battery or on the packaging indicates that the battery provided with this product shall not be treated as household waste. By ensuring these batteries are disposed of correctly, you will help prevent potentially negative consequences for the environment and human health which could otherwise be caused by inappropriate waste handling of the battery. The recycling of the materials will help to conserve natural resources. At the end of their life hand the batteries over to the applicable collection points for the recycling of waste batteries. For more detailed information about recycling this product or battery, please contact your local Civic Office, your household
waste disposal service, or the shop where you purchased the product.

GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals