OMEGA FLSW3400A Optical Sensor Switch User Guide
- October 27, 2023
- Omega
Table of Contents
- UNPACKING THE Optical Sensor Switch
- DESCRIPTION
- INSTALLATION
- Electrical Connections
- SPECIFICATIONS
- CE Compliance
- OPERATING INSTRUCTIONS
- TROUBLESHOOTING
- APPENDIX 1 COMPONENTS DIAGRAM
- APPENDIX 2 DIMENSIONAL DRAWINGS
- WARRANTY/DISCLAIMER
- RETURN REQUESTS/INQUIRIES
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
**FLSW3400A Optical Sensor Switch
User Guide
Shop online at omega.com
e-mail: info@omega.com
For latest product manuals:
www.omegamanuatinfo
**
UNPACKING THE Optical Sensor Switch
1.1 Inspect Package for External Damage
Your Optical Sensor Switch was carefully packed in a sturdy cardboard carton,
with anti-static cushioning materials to withstand shipping shock. Upon
receipt, inspect the package for possible external damage. In case of external
damage to the package contact the shipping company immediately.
1.2 Unpack the Optical Sensor Switch
Open the carton carefully from the top and inspect for any sign of concealed
ship- ping damage. Please keep all packing and notify Omega Customer Service
of any damage.
When unpacking the instrument please make sure that you have all the items
indicated on the Packing List. Please report any shortages promptly.
1.3 Packing List
Desired Flow Meter with Hi-Lo
Sensors Installed
OSSM
Power Supply
Splitter
User’s Guide| Quantity
1
1
Optional
2
1
---|---
1.4 Returning Merchandise for Repair
Please contact the customer service representative at 1-800-826-6342 ext. 2208
for a Return Authorization Number (AR).
It is mandatory that any equipment returned for servicing be purged and
neutralized of any dangerous contents including but not limited to toxic,
bacterially infectious, corrosive or radioactive substances.
DESCRIPTION
The 150 mm Flow Meter with Hi-Lo Alarm Optical Sensor Switche is a noninvasive
means for detection of a HI or LOW flow.
This sensor is ideal for signaling an alarm, cutoff valve, or other device
when the float passes the detector.
The Hi-Lo Optical Sensor Switch consists of the OSSM module and two self-
contained mini-slim photoelectric sensors (Thrubeam type). Every sensor has a
transmitter and receiver. Two sets of sensors are mounted on two solid
carriers on opposite sides of the flow tube. The float inside the flow tube is
detected as it passes across the beam of light. The sensors can be used to
detect the float passage beyond the set point of the sensor and can also be
set to mo111
Pagenitor the float position at a specific level, signaling when the float is
outside of the range of the sensor light beam.
INSTALLATION
Electrical Connections
The Optical Sensor Switch gets a power through the OSSM module.
The module could be powered up either by a +12+15 VDC power supply with a
minimum current rating of 250 mA, or by the 9-pin D-connector (DB9M) (Fig. 1)
for momentary or latch operations.
SUPPLYING DC POWER TO THE POWERJACK AND THE “D” CONNECTOR AT THE SAME TIME
WILL DAMAGE THE METER. DC POWER JACK POLARITY IS CENTER POSITIVE.
WARNING: DO NOT CONNECT POWER SUPPLY WITH MORE THAN 15VDC VOLTAGE
3.1. Insert the LOW flow sensor’s plugs into the splitter, then the splitter’s
plug into the module’s phone jack #1 (Fig. 1).
3.2. Insert the HIGH flow sensor’s plugs into the splitter, then the
splitter’s plug into the module’s phone jack #2 (Fig. 1).
3.3 Connect the power supply through the power jack (Fig. 1). Or
Attach the 9-pin D-connector to the module (fig. 1 & 2).
During the initial power up of the Optical Sensor Switch with no flow
conditions the red and green LED on the module will be “On”. This is an
indication that power is on and the sensors are ready for operation.
Both sensors will have 2 green lights “On” from each side of the flow meter
Important notes:
In general, “D” Connector numbering patterns are standardized. There are, how-
ever, some connectors with nonconforming patterns and the numbering sequence
on your mating connector may or may not coincide with the numbering sequence
shown in our pin configuration table above. It is imperative that you match
the appropriate wires in accordance with the correct sequence regardless of
the particular numbers displayed on your mating connector.
Make sure power is OFF when connecting or disconnecting any cables in the sys-
tem. The power input is protected by a 600 mA (medium time-lag) resettable
fuse.
If a shorting condition or polarity reversal occurs, the fuse will cut power
to the valve circuit. Disconnect the power to the unit; remove the faulty
condition, and reconnect the power.
The fuse will reset once the faulty condition has been removed.
Use of the Optical Sensor Switch in a manner other than that specified in this
manual may impair the protection provided by the equipment.
SPECIFICATIONS
MATERIALS OF | CONSTRUCTION: |
---|---|
END BLOCKS: | Aluminum or 316 Stainless Steel. |
ELASTOMERS: | Buna & Viton (Aluminum), Viton (316 SS). |
FLOW TUBE: | Borosilicate Glass. |
POWER INPUT: | + 12+15 VDC. Minimum current rating 250 mA peak to peak max |
100mV.
POWER CONSUMPTION:| Less than 200 mA.
ACCURACY:| +/- 2% of full scale.
REPEATABILITY:| 0.5% of full scale.
AMBIENT TEMPERATURE:| 20-55 deg. C.
RESPONSE TIME:| 0.5 milliseconds.
LIGHT IMMUNITY:| 4 Element, point light source, red LED 650 nm
DRY CONTACT CLOSURES:| 2 normally open, and normally closed relays (1A, 30VDC
max.)
ALARM:| 70 dB audible buzzer and/or visual LED
ALARM OPTIONS:| High, Low, or High/Low.
BUZZER:| User configurable, momentary or latch.
RESET:| Reset button or remote through “D”-connector to disable relay or
buzzer (remote Reset is TTL compatible active Low).
Environmental (per IEC 664):| Installation Level II; Pollution Degree
CE Compliance
Any model Optical Sensor Switch bearing a CE marking on it, is in compliance
with the below stated test standards currently accepted.
EMC Compliance with 89/336/EEC as amended;
Emission Standard: EN 55011:1991, Group 1, Class B
Immunity Standard: EN 55082 1:1992
OPERATING INSTRUCTIONS
6.1. Sensor Alignments
Two sensors’ carriers can be set in any position along the flow meter tube.
WARNING : Do not install the sensors closer than 1/2” (10 mm) from one to
another. Doing so may trigger a false alarm condition.
To set up the switches position:
-Loosen the 2 red thumbscrews;
-Slide the switch carrier to the desired position;
-Tighten the thumbscrews.
Stable operation indicator (green LED) turns on with a stable incoming beam
and with a stable blocked light. Output indicator (orange LED) turns on when
the beam from emitter is blocked by the float.
Each Optical Sensor Switch is shipped from the factory with the sensors
aligned and does not require additional adjustment of the focus distance even
if the flow tube was replaced or reinstalled (see p. 6.2).
6.2. DIP switch configuration
The DIP switch, located near the 9-pin D-connector can be used to set up
custom’s settings separately for High and Low alarm.
In order to activate the sensor’s alarm the corresponding switch has to be
installed in the “On” position on the Configuration DIP switch (see Fig. 3).
Each sensor (High or Low) can be separately set for Momentary or Latch
operation. In addition the Relay and Buzzer can be configured individually.
As an example, if the SW1 is set to “On”: the High flow alarming (latching
mode) is active, then the green LED (on left side of the D-connector) and
relay will be energized in the LATCH mode. They will stay in the energized
state without any conditions (even if the flow rate changes and the sensor
have become inactive) until the RESET button is pressed or external RESET TTL
signal is applied. If momentary mode is set (DIP switch SW3 is set to “On”)
the relay and green LED will be energized only when the High flow sensor is
active (Alarm Condition). The same is true for the Buzzer settings.
All Switches “Off” position–Alarm is disabled (LED, Relays and Buzzer).
ALARM DIP SWITCH CONFIGURATION “On”- pressed down “Off’-raised up
HIGH ALARM| LOW ALARM| ACTION WHEN SWITCH IS “ON”
SW1| SW5| LED and Relay are On (LATCH)
SW2| SW6| BUZZER is On (LATCH)
SW3| SW7| LED and Relay are On momentary (FOLLOW SENSOR)
SW4| SW8| BUZZER is On momentary (FOLLOW SENSOR)
Figure 3: Optical Sensor DIP Switch Configuration.
TROUBLESHOOTING
7.1. Common Conditions
Your Optical Sensor Switch was thoroughly checked at numerous quality control
points during and after manufacturing and assembly operations. It was
carefully packed to prevent damage during shipment. Should you feel that the
instrument is not functioning properly please check for the following common
conditions first:
Are all cables connected correctly?
Are there any leaks in the installation?
Is the power supply correctly selected according to requirements? When several
devices are used a power supply with appropriate current rating should be
selected.
Were the connector pinouts matched properly? When interchanging with other
manufacturers’ equipment, cables and connectors must be carefully wired for
correct pin configurations.
7.2. Troubleshooting Guide
N.o | INDICATION | LIKELY REASON | SOLUTION |
---|---|---|---|
1 | No green LED | ||
lights on sensors. | Power supply off. | Check connection of the |
power supply.
No connections between
sensors and splitter; no
connections between
splitter and OSSM| Check connection
between sensors,
splitters & the OSSM.
|
2| Buzzer or Relay does not work.| Wrong configuration of the DIP switch.|
Make correct configuration of the DIP switch according to Figure 3.
Pc board defect.| Return to factory for replacement.
3| The sensor does not react on
passing float (no orange LED)| Sensor’s carrier is not locked into the
position| Tighten the red thumbscrews
7.3. Technical Assistance
Repair personnel. Please call our Technical Assistance at 800 872 9436 ext.
2298. Please have your Serial Number and Model Number ready when you call.
APPENDIX 1 COMPONENTS DIAGRAM
APPENDIX 2 DIMENSIONAL DRAWINGS
WARRANTY/DISCLAIMER
OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials
and workmanship for a period of 13 months from date of purchase. OMEGA’s
WARRANTY adds an additional one (1) month grace period to the normal one (1)
year product warranty to cover handling and shipping time. This ensures that
OMEGA’s customers receive maximum coverage on each product.
If the unit malfunctions, it must be returned to the factory for evaluation.
OMEGA’s Customer Service Department will issue an Authorized Return (AR)
number mmediately upon phone or written request. Upon examination by OMEGA,
if the unit is found to be defective, it will be repaired or replaced at no
charge. OMEGA’s WARRANTY does not apply to defects resulting from any
action of the purchaser, including but not limited to mishandling, improper
interfacing, operation outside of design limits, improper repair, or
unauthorized modification. This WARRANTY is VOID if the unit shows evidence of
having been tampered with or shows evidence of having been damaged as a result
of excessive corrosion; or current, heat, moisture or vibration; improper
specification; misapplication; misuse or other operating conditions
outside of OMEGA’s control. Components in which wear is not warranted, include
but are not limited to contact points, fuses, and triacs.
OMEGA is pleased to offer suggestions on the use of its various products.
However, OMEGA neither assumes responsibility for any omissions or errors nor
assumes liability for any damages that result from the use of its products in
accordance with information provided by OMEGA, either verbal or written. OMEGA
warrants only that the parts manufactured by the company will be as specified
and free of defects. OMEGA MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY
KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED
WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The
remedies of purchaser set forth herein are exclusive, and the total liability
of OMEGA with respect to this order, whether based on contract, warranty,
negligence, indemnification, strict liability or otherwise, shall not exceed
the purchase price of the component upon which liability is based. In no event
shall OMEGA be liable for consequential, incidental or special damages.
CONDITIONS : Equipment sold by OMEGA is not intended to be used, nor
shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or
with any nuclear installation or activity; or (2) in medical applications or
used on humans. Should any Product(s) be used in or with any nuclear
installation or activity, medical application, used on humans, or misused in
any way, OMEGA assumes no responsibility as set forth in our basic
WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify
OMEGA and hold OMEGA harmless from any liability or damage whatsoever arising
out of the use of the Product(s) in such a manner.
RETURN REQUESTS/INQUIRIES
Direct all warranty and repair requests/inquiries to the OMEGA Customer
Service Department.
BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED
RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID
PROCESSING DELAYS). The assigned AR number should then be marked on the
outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and
proper packaging to prevent breakage in transit.
FOR WARRANTY RETURNS, please have thefollowing information available BEFORE
contacting OMEGA:
1. Purchase Order number under which the product was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the product.| FOR
NON-WARRANTY REPAIRS, consult OMEGA for current repair charges. Have the
following information available BEFORE contacting OMEGA:
1. Purchase Order number to cover the COST of the repair,
2. Model and serial number of the product, and
3. Repair instructions and/or specific problems relative to the product.
---|---
OMEGA’s policy is to make running changes, not model changes, whenever an
improvement is possible.
This affords our customers the latest in technology and engineering.
OMEGA is a registered trademark of OMEGA ENGINEERING, INC.
© Copyright 2009 OMEGA ENGINEERING, INC. All rights reserved. This document
may not be copied, photocopied, repro- duced, translated, or reduced to any
electronic medium or machine-readable form, in whole or in part, without the
prior writ- ten consent of OMEGA ENGINEERING, INC.
Where Do I Find Everything I Need for Process Measurement and Control?
OMEGA…Of Course! Shop online atomega.comSM
TEMPERATURE
- Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies
- Wire: Thermocouple, RTD & Thermistor
- Calibrators & Ice Point References
- Recorders, Controllers & Process Monitors
- Infrared Pyrometers
PRESSURE, STRAIN AND FORCE
- Transducers & Strain Gages
- Load Cells & Pressure Gages
- Displacement Transducers
- Instrumentation & Accessories
FLOW/LEVEL
- Rotameters, Gas Mass Flowmeters & Flow Computers
- Air Velocity Indicators
- Turbine/Paddlewheel Systems
- Totalizers & Batch Controllers
pH/CONDUCTIVITY
- pH Electrodes, Testers & Accessories
- Benchtop/Laboratory Meters
- Controllers, Calibrators, Simulators & Pumps
- Industrial pH & Conductivity Equipment
DATA ACQUISITION
- Data Acquisition & Engineering Software
- Communications-Based Acquisition Systems
- Plug-in Cards for Apple, IBM & Compatibles
- Datalogging Systems
- Recorders, Printers & Plotters
HEATERS
- Heating Cable
- Cartridge & Strip Heaters
- Immersion & Band Heaters
- Flexible Heaters
- Laboratory Heaters
ENVIRONMENTAL MONITORING AND CONTROL
- Metering & Control Instrumentation
- Refractometers
- Pumps & Tubing
- Air, Soil & Water Monitors
- Industrial Water & Wastewater Treatment
- pH, Conductivity & Dissolved Oxygen Instruments
OMEGAnet® OnlineServiceomega.com |
Internet e-mail info@omega.com
---|---
Servicing North America:
U.S.A.: ISO 9001 Certified
Servicing North America: Omega Engineering, Inc., One Omega Drive, P.O. Box
4047 Stamford, CT 06907-0047
Toll-Free: 1-800-826-6342
FAX: (203) 359-7700| Tel: (203) 359-1660
e-mail:info@omega.com
---|---