GIMA Biological Microscope User Manual

June 6, 2024
GIMA

GIMA LOGO

GIMA Biological Microscope User Manual

GIMA Biological Microscope

Model: N-126

Manufacturer

NINGBO YONGXIN OPTICS CO., LTD.
No. 169 Mujin Road, Hi-tech Park Ningbo, China
Made in China

Imported by

Gima S.p.A.
Via Marconi, 1 – 20060 Gessate (MI) Italy
gima@gimaitaly.comexport@gimaitaly.com
www.gimaitaly.com

FIG 1 Product compliance

Safety Note

  1. Open the box carefully to avoid the accessories, like lens, dropping to ground or being damaged.

  2. Do keep the instrument out of direct sunlight, high temperature or humidity, dusty and easy shaking environment. Make sure the stage is flat, horizontal and firm enough.

  3. When moving the microscope, carefully carry it with the handle and the base.

  4. Make sure the instrument is earthed, to avoid lighting strike.

  5. For safety, be sure the main switch is in “O” (off) state and cut off the power supply before replacing the bulb or the fuse. If you replace the bulb during use or right after use, allow the lamp bulb and the lamp house to cool completely before touching.
    (Designated bulb: 1W S-LED)

  6. Check the input voltage: be sure the input voltage which is signed in the back of the microscope is consistent with the power supply voltage, or it will bring a serious damage to the instrument.

  7. Always use the power cord provided by Gima.

  8. The electrical equipment of the microscope should be discard as electronic waste.

Maintenance and Care

  1. All the lenses have been adjusted properly; do not dismount them by yourself please.
  2. The nosepiece and coarse and fine focusing parts are so delicate that it is forbidden to disassemble them carelessly by yourself.
  3. Keep the instrument clean, and do not pollute the optical element when wiping away the dust on the instrument.
  4. The contaminations on the prism, like fingerprints and oil smudges, could be gently wiped with a piece of soft cloth or tissue paper, gauze which has been immersed in pure alcohol or ether. (Note that the alcohol and ether are highly flammable, do keep them away from the fire or potential sources of electrical sparks, and use them in a drafty room as possible as you can.)
  5. Do not attempt to use organic solvents to clean the microscope components other than the glass components. To clean them, use a lint-free, soft cloth slightly moistened with a diluted neutral detergent.
  6. When using, if the microscope is splashed by liquid, cut off the power at once, and wipe away the splash.
  7. Do not disassemble any parts of the microscope, as this will affect the function or reduce the performance of the microscope.
  8. Place the instrument in a cool, dry position. When not using the microscope, keep it covered with a dust cover. Make sure the lamp socket is cool before covering the microscope.

1. Components Name

FIG 2 Components Name

2. Assembly

2.1 Assembly Diagram
The following figure shows the installation sequence of the components. The number in the figure shows the assembly steps.

  • Before installing, be sure every components is clean, do not score any parts or glass surface.
  • Keep well with hexagon wrench provided. When changing the components, you will need it again.

FIG 3 Assembly Diagram

2.2 Assembly procedure
Installing binocular viewing head (fig. 1-2)
Insert the binocular viewing head into the microscope head and turn it to a proper position, then tight it.

FIG 4 Assembly procedure

Installing the eyepiece (fig. 3-4)
Insert the eyepiece into the eyepiece tube until they are against each other (fig. 4).

FIG 5 Assembly procedure

Installing the objective (fig. 5-6)

  1. Adjusting the coarse focus knob until the support device of the mechanical stage reaches its low limit position.
  2. Screw the lowest magnification objective into the nosepiece from the left or the right side, then revolve the nosepiece clockwise and mount other objectives by the sequence of low to high magnification. Installing objective this way will make the change of magnification to be easier during using.
  • Clean the objective regularly, for lens is susceptible to dust.
  • When operating, use 10× magnification objective to search and focus specimen firstly, then replace with higher magnification objective if necessary.
  • When replacing the objective, slowly turn the nosepiece until you hear “clicked”, that means the objective is in place.

FIG 6 Installing the objective

Connecting the Power Cord (fig. 7-9)
The cable and cords are vulnerable when bent or twisted, never subject the power cord to excessive force.

  1. Set the main switch 1 to “0” (off) state before connecting the power cord.
  2. Insert the plugs 2 into the power jack 3 of the microscope safely.
  3. Plug the power cord 4 into the power supply receptacle 5. Make sure the connection is well.
  • Do use the supplied power cord all the time. If lost or damaged, select the same standard cord, please.
  • A wide range of voltage, like 100V~240V, is acceptable for this microscope.

FIG 7 Connecting the Power Cord

FIG 8 Connecting the Power Cord

Replacing the fuse (fig. 8-9)
Do remember to set the main switch 1 to the state “0” (OFF) and unplug the power cord before replacing the fuse. Rotate the fuse out of the holder 3, replace with a new fuse, then rotate it back to the holder again.

3. Adjustment and Operation

3.1 Adjustment set diagram

FIG 9 Adjustment and Operation

fig. 10

FIG 10 Adjustment and Operation

fig. 11

3.2 Adjustment and Operation
Brightness Adjustment (fig. 12-13)

  1. Connect the power cord and set the main switch 1 to “ – ”state (ON).
  2. Turning the brightness adjustment knob 2.

FIG 11 Adjustment and Operation

Placing the Specimen (fig. 14)

  1. Place the specimen the center of the mechanical stage and use the stage clips to clamp it.
  2. Turn the portrait and lateral adjustment knob 1 of the mechanical ruler, move the specimen to the required position.
  • Be careful when changing the objective. If you finish the observation with the short working distance objective, and want to change another one, becareful of not letting the objective touch the specimen.

FIG 12 Adjustment and Operation

Fig. 14 & 15

Focusing the Specimen (fig. 15)

  1. Focus the specimen with 10x objective. To avoid the objective touching the specimen during focusing, you should raise the mechanical stage to let the specimen close to the objective at first, then slowly separate them to bring the specimen to focus.
    Turn the coarse focus knob 1 conversely to lower the specimen and search images in the 10×ocular simultaneously, and then use the fine knob 2 to focus. After that, you can replace with other magnification objectives safely, and focus without the risk of damaging the specimen.
    To make the observation more convenient, you can use the locking set 3 to fix the stage in a vertical direction.

FIG 13 Adjustment and Operation

Condenser Adjustment (fig. 16)
Turn the condenser focus knob 1 to move the condenser up and down. Raise the condenser when using the high magnification objective, and descend it when using the low magnification one.

  1. Focus the specimen with 10x objective.
  2. Adjust the condenser focus knob 1 to get a clear image of the field iris diaphragm.
  3. Turn the condenser centering knobs 2 to center the image of field iris diaphragm in the field of view.
  • The condenser and the objective are coaxial. It has been adjusted before leaving factory, so the user needn’t to adjust them by self
  • The highest position of the condenser has been adjusted too. It also needn’t any user’s operation. (The top surface of the condenser is 0.03 mm – 0.4 mm lower than the stage top surface.)

FIG 14 Adjustment and Operation

Aperture Iris Diaphragm Adjustment (fig. 17)
Turn the aperture iris diaphragm stick 1 to adjust the aperture iris diaphragm.

  • The aperture iris diaphragm is designed forthe adjustment of the numerical aperture, not for the brightness.
    Generally, setting the aperture iris diaphragm to 70- 80% of the N.A. of the objective in use will provide an image with good contrast. If you want to observe the image of the aperture iris diaphragm, remove one eyepiece and look through the tube. You will see a dark circle encroaching on the bottom of the tube.

FIG 15 Adjustment and Operation

Adjusting the Interpupillary Distance (fig. 18)
The interpupillary distance range: 47mm-78mm While looking through the eyepieces, move both eyepieces round until the left and right fields of view coincide completely.

FIG 16 Adjustment and Operation

Adjusting the Diopter (fig. 19)

  1.  Turn the eyepiece 1 to adjust the diopter while looking through it.
  • The diopter range of the eyepiece is ±5 diopter. The number aligned to the line on the viewing head is the diopter in use.

FIG 17 Adjustment and Operation

Adjusting the tension adjustment collar (fig. 20)
Turn the tension adjustment collar 1 with your fingers. When the collar is turned in the direction of the arrow, the tension of the coarse adjustment knob 2 increases. Turning the collar in the opposite direction decreases the tension.
If the nosepiece descends on its own or if the specimen gets out of focus quickly even when it is brought into focus using the fine adjustment knob 3, it means the tension of the coarse adjustment knob is too low. Turn the collar in the direction of the arrow to increase the tension.

4. Technical Specifications

4.1 Main Specifications

FIG 18 Technical Specifications

4.2 Eyepiece and objectives

Achromatic Objective

FIG 19 Technical Specifications

Infinite Semi-plan Objective

FIG 20 Technical Specifications

Eyepiece

FIG 21 Technical Specifications

Total Magnification

FIG 22 Technical Specifications

FIG 23 Technical Specifications

5. Troubleshooting Guide

5.1 Optical System

FIG 24 Troubleshooting Guide

FIG 25 Troubleshooting Guide

FIG 26 Troubleshooting Guide

5.2 Mechanical System

FIG 27 Troubleshooting Guide

5.3 Electrical System

FIG 28 Troubleshooting Guide

FIG 29 Troubleshooting Guide

Disposal: The product must not be disposed of along with other domestic waste. The users must dispose of this equipment by bringing it to a specific recycling point for electric and electronic equipment.

GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals