ACCU-CHEK 07453701 Test Strips Instruction Manual
Important Information
NOTE: In this package insert the term “blood glucose” is used when referring to “blood sugar.”
Intended Use for the Accu-Chek Guide Meter
The Accu-Chek Guide Blood Glucose Monitoring System is comprised of the
Accu-Chek Guide meter and Accu-Chek Guide test strips.
The Accu-Chek Guide Blood Glucose Monitoring System is intended to
quantitatively measure glucose in fresh capillary whole blood from the
fingertip, palm, and upper arm as an aid in monitoring the effectiveness of
glucose control. The Accu-Chek Guide Blood Glucose Monitoring System is
intended for in vitro diagnostic single patient use by people with diabetes.
The Accu-Chek Guide Blood Glucose Monitoring System is intended to be used by
a single person and should not be shared. This system is not for use in
diagnosis or screening of diabetes mellitus, nor for neonatal use. Alternative
site testing should be done only during steady-state times (when glucose is
not changing rapidly).
Accu-Chek Guide control solutions are for use with the Accu-Chek Guide Blood
Glucose Monitoring System to check that the meters and test strips are working
together properly and that the test is performing correctly.
Intended Use for the Accu-Chek Guide Link Meter
The Accu-Chek Guide Link Blood Glucose Monitoring System is comprised of the
Accu-Chek Guide Link meter and Accu-Chek Guide test strips. The Accu-Chek
Guide Link Blood Glucose Monitoring System is intended to quantitatively
measure glucose in fresh capillary whole blood from the fingertip, palm, and
upper arm as an aid in monitoring the effectiveness of glucose control. The
Accu-Chek Guide Link Blood Glucose Monitoring System is intended for in vitro
diagnostic single patient use by people with diabetes.
The Accu-Chek Guide Link Blood Glucose Monitoring System is intended to be
used by a single person and should not be shared.
This system is not for use in diagnosis or screening of diabetes mellitus, nor
for neonatal use.
Alternative site testing should be done only during steady-state times (when
glucose is not changing rapidly).
Accu-Chek Guide control solutions are for use with the Accu-Chek Guide Link
Blood Glucose Monitoring System to check that the meters and test strips are
working together properly and that the test is performing correctly.
The Accu-Chek Guide Link Blood Glucose Monitoring System is intended to be
used to wirelessly transmit glucose values to the MiniMed™ 770G system with
Bluetooth® wireless technology through the use of Bluetooth low energy
communication.
Intended Use for the Accu-Chek Guide Me Meter
The Accu-Chek Guide Me Blood Glucose Monitoring System is comprised of the
Accu-Chek Guide Me meter and Accu-Chek Guide test strips.
The Accu-Chek Guide Me Blood Glucose Monitoring System is intended to
quantitatively measure glucose in fresh capillary whole blood from the
fingertip, palm, and upper arm as an aid in monitoring the effectiveness of
glucose control. The Accu-Chek Guide Me Blood Glucose Monitoring System is
intended for in vitro diagnostic single patient use by people with diabetes.
The Accu-Chek Guide Me Blood Glucose Monitoring System is intended to be used
by a single person and should not be shared.
This system is not for use in diagnosis or screening of diabetes mellitus, nor
for neonatal use.
Alternative site testing should be done only during steady-state times (when
glucose is not changing rapidly).
Accu-Chek Guide control solutions are for use with the Accu-Chek Guide Me
Blood Glucose Monitoring System to check that the meters and test strips are
working together properly and that the test is performing correctly.
Introduction
Testing your blood glucose regularly may help you better manage your diabetes,
which can prevent or slow the development of diabetes complications. Your
healthcare professional can tell you what blood glucose range is appropriate
for you. It is very important to stay in your target range. If your blood
glucose is too low, you may experience anxiety, shakiness, sweating, headache,
increased hunger, dizziness, pale skin color, sudden change in mood or
irritability, fatigue, difficulty concentrating, clumsiness, palpitations,
and/or confusion. If your blood glucose is too high, you may experience
increased thirst, frequent urination, blurred vision, drowsiness, and/or
unexplained weight loss.
Important Safety Information
-
During normal testing any blood glucose meter or lancing device may come in contact with blood. All parts of the kit are considered biohazardous and can potentially transmit infectious diseases from bloodborne pathogens, even after you have performed cleaning and disinfecting.1,2
-
Meters and lancing devices should never be used by more than one person. Do not share them with anyone, including other family members, due to the risk of infection from bloodborne pathogens.1,2 Do not use on multiple patients!
-
Clean and disinfect the meter and the lancing device before allowing anyone else to handle them. Do not allow anyone else to test with the meter or lancing device.
-
It is important to keep the meter and lancing device clean and disinfected. Read and follow the meter and lancing device cleaning and disinfecting instructions found in the meter User’s Manual.
-
Wash and dry hands thoroughly using soap and water before and after handling the meter, lancing device, and test strips.
Warning -
Do not eat the test strips. They are only for use outside the body (in vitro).
-
Choking hazard. Small parts. Keep away from children under the age of 3 years.
-
Use of this device on multiple patients may lead to transmission of Human Immunodeficiency Virus (HIV), Hepatitis C Virus (HCV), Hepatitis B Virus (HBV), or other bloodborne pathogens.
Limitations
- Do not use the meter at high hematocrit levels above 65 % or low hematocrit levels below 10 %.
- Not for use in diagnosis or screening of diabetes mellitus.
- Not for neonatal use.
- Abnormally high concentrations of ascorbic acid (vitamin C) resulting in blood concentrations in excess of 5 mg/dL may cause inaccurate results. If you are not sure please check with your doctor.
- Do not use the meter system to measure blood glucose in people who are experiencing cardiovascular collapse (severe shock) or decreased peripheral blood flow.
- Do not use this system during xylose absorption test.
- Not for use on critically ill patients, patients in shock, dehydrated patients, or hyperosmolar patients.
- This system has not been tested at altitudes higher than 10,150 feet.
Before You Start Testing
- Carefully read and follow all instructions in the package insert when testing your blood glucose.
- The Accu-Chek Guide test strips are for testing fresh capillary whole blood.
- Your hematocrit should be between 10–65 %. Ask your healthcare professional if you do not know your hematocrit.
- If you have poor circulation, testing your own blood glucose may not be right for you. Ask your healthcare professional.
Test Strip Storage and Handling
- Use the test strips at temperatures between 43–113 °F.
- Use the test strips between 10–90 % relative humidity. Humidity is the amount of dampness in the air.
- Do not expose the test strips to heat, moisture, or humidity. Temperatures outside the required range, as well as moisture and humidity, can damage the test strips and produce incorrect results.
- Use the test strip immediately after removing it from the container.
- Discard the test strips if they are past the Use By date printed on the test strip container. If the Use By date is missing or cannot be read, do not use the test strips.
- Store the test strip container at temperatures between 39–86 °F. Do not freeze. Do not store the test strip container in rooms where the air is humid such as the kitchen, laundry room, or bathroom.
- Store the test strips in their original container with the cap closed. Close the container tightly immediately after removing a test strip to protect the test strips from humidity.
- Store the test strip container between 10–90 % relative humidity.
STEP 1 Getting Ready to Test
- Get your supplies together. You need the meter, a test strip, a lancing device, and a lancet.
- Prepare the lancing device.
- Check the Use By date on the test strip container. Do not use test strips past the Use By date.
STEP 2 Performing a Control Test
What control solution to use
Use the Accu-Chek Guide control solutions.
How the control solution works
The control solution contains a known amount of glucose that acts like blood
when you apply it to the test strip. Performing a control test lets you know
that the meter and test strips are working properly.
When to perform a control test
- You open a new test strip box
- You left the test strip container open
- You think the test strips are damaged
- You want to check the meter and test strips
- The test strips were stored in extreme temperatures, humidity, or both
- You dropped the meter
- Your test result does not match how you feel
- You want to check if you are performing the test correctly
How to perform a control test
Refer to the control solution package insert or User’s Manual for
instructions. If you need control solutions, talk to your pharmacist or visit
accu-chek.com to
order online.
What the control results mean
Compare the control result to the range on the label of the test strip
container. If the control result is within the range, you know that the meter
and test strips are working properly. You can now test your blood glucose.
If the control result is not within the range…
- Were the test strips or control solutions past the Use By or discard date?
- Did you wipe the tip of the control solution bottle before use?
- Were the caps on the test strip container and the control solution bottle always closed tightly?
- Was the test strip used immediately after it was removed from the test strip container?
- Were the test strips and control solutions stored in a cool, dry place?
- Did you follow the directions?
- Did you choose the correct control solution level, either 1 or 2, when you
performed the control test?
For more information, refer to the control solution package insert or User’s Manual. For questions, contact the Accu-Chek Customer Care Service Center toll-free at 1-800-858-8072. We offer assistance in many languages.
STEP 3 Performing a Blood Glucose Test
You are now ready to test your blood glucose. Read the “Getting a good blood
drop” section of this package insert if you have problems or are new to
testing.
Procedure
-
Wash hands with warm soapy water and dry thoroughly.
-
Insert the metallic end of the test strip into the meter. The meter turns on.
On the Accu-Chek Guide and Accu-Chek Guide Link meters, Preparing to test appears on the display.
On the Accu-Chek Guide Me meter, a flashing drop appears on the display. -
Check the Use By date on the test strip container. Do not use test strips past the Use By date.
-
Use the lancing device to get a blood drop.
-
Touch the yellow edge of the test strip to the blood drop. Do not put blood on top of the test strip.
On the Accu-Chek Guide and Accu-Chek Guide Link meters, remove your finger from the test strip when Analyzing appears on the display.
On the Accu-Chek Guide Me meter, remove your finger from the test strip when the flashing hourglass appears on the display. Failure to move your finger away from the test strip could give inaccurate results. -
The test result appears on the display.
-
Take the test strip out of the meter. Put the lancet and the test strip in a puncture-proof container, such as a biohazard container.
-
Wash hands thoroughly with soap and water.
Getting a good blood drop
If you have trouble getting a good blood drop, here are some tips:
Fingertip testing
-
Wash your hands in warm, soapy water, then rinse and dry thoroughly.
(Warming your fingers can increase blood flow.) -
If you use an alcohol wipe, make sure the testing site is dry before getting a blood drop.
-
Let your hand hang by your side. (This increases blood flow.)
-
Squeeze your finger at the knuckle for 3 seconds, then let go. Repeat.
-
Prick your finger, and then squeeze at the knuckle to form a blood drop. Do not squeeze too hard.
Alternative site testing–Rub the skin prior to lancing to increase blood flow.
Your Blood Glucose Results
The normal fasting glucose level for a non-diabetic adult is below 100 mg/dL.
The normal glucose level for a non-diabetic adult 2 hours post 75 g Oral
Glucose Tolerance Test (OGTT) is less than 140 mg/dL.3,4 For people with
diabetes: Consult your healthcare professional for the blood glucose range
appropriate for you. You should treat your low or high blood glucose as
recommended by your healthcare professional.
These test strips deliver results that correspond to blood glucose
concentrations in plasma as per the recommendation of the International
Federation of Clinical Chemistry and Laboratory Medicine (IFCC). Therefore,
your meter displays blood glucose concentrations that refer to plasma although
you always apply whole blood to the test strip.
Symptoms of low or high blood glucose
Being aware of the symptoms of low or high blood glucose can help you
understand your blood glucose results and decide what to do if they seem
unusual.
Low blood glucose (hypoglycemia): Symptoms of hypoglycemia may include,
but are not limited to, anxiety, shakiness, sweating, headache, increased
hunger, dizziness, pale skin color, sudden change in mood or irritability,
fatigue, difficulty concentrating, clumsiness, palpitations, and/or confusion.
High blood glucose (hyperglycemia): Symptoms of hyperglycemia may
include, but are not limited to, increased thirst, frequent urination, blurred
vision, drowsiness, and/or unexplained weight loss.
If you are having any of these symptoms, test your blood glucose. If your
blood glucose result is displayed as LO or HI, and you have symptoms of low or
high blood glucose, follow your healthcare professional’s instructions or
contact your healthcare professional. For example, if your healthcare
professional has advised you to immediately treat a low blood glucose result
(such as by eating something), then do so.
Warning: Follow the advice of your healthcare professional before you
change your therapy!
If LO is displayed on the meter, blood glucose may be below 20 mg/dL.
If HI is displayed on the meter, blood glucose may be over 600 mg/dL.
If your blood glucose result does not match the way you feel, follow these
steps:
- Perform a control test.
- Review the test procedure and repeat the test with a new test strip.
- If your blood glucose result still does not match the way you feel, follow your healthcare professional’s instructions or contact your healthcare professional immediately.
Do not change your treatment because of just one result.
NEVER ignore symptoms of low or high blood glucose.
For detailed information on error codes, refer to the User’s Manual.
Testing from Alternative Sites with the Accu-Chek G uide, Accu-Chek
Guide Link, and Accu-Chek Guide Me Meters
Warning: Do not use alternative site testing to calibrate a continuous
glucose monitoring system or to make insulin dosing calculations. Warning:
Alternative site testing should be done only during steady-state times (when
blood glucose is not changing rapidly).
Alternative site testing is ONLY approved for the Accu-Chek Guide test strips
when used with the Accu-Chek Guide, Accu-Chek Guide Link, or Accu-Chek Guide
Me meters.
You have the option of obtaining a blood sample from other sites on your body
besides the fingertip. Alternative sites include the palm and upper arm. Blood
obtained from the fingertip can be used at any time to test blood glucose. If
blood from an alternative site is used, there are certain times when testing
is not appropriate (see below). This is because your blood glucose level
changes faster in your fingertip than in the alternative sites. These
differences may cause you to make the wrong therapeutic decision, producing
adverse health effects. Read the following section before you test from
alternative sites.
IMPORTANT
-
Talk to your healthcare professional before you test from alternative sites.
-
Only fingertip samples should be used to calibrate a continuous glucose
monitoring (CGM) device or to make insulin dosing calculations.
You may perform a test from the palm or upper arm: -
immediately before a meal.
-
while fasting.
You may NOT perform a test from the palm or upper arm: -
up to 2 hours following a meal, when blood glucose values can rise quickly.
-
after injecting bolus insulin, when blood glucose values can decrease rapidly.
-
after exercise.
-
if you are sick.
-
if you think your blood glucose is low (hypoglycemia).
-
if you sometimes do not notice when your blood glucose is low.
If your blood glucose does not match how you feel, perform a fingertip test to confirm your result. If the fingertip test result still does not match how you feel, contact your healthcare professional. For more information on testing from alternative sites, contact the Accu-Chek Customer Care Service Center at 1-800-858-8072.
Performance Characteristics
Test principle: The blood glucose monitoring systems are plasma
calibrated to
allow easy comparison of results with laboratory methods. Blood from your
fingertip reacts with the chemicals in the test strip to create a harmless
electrical current in the test strip. The meter reads the current and gives
you the blood glucose result.
Sample size: 0.6 μL
Test time: <4 seconds
System measuring range: 20–600 mg/dL
User performance study: In a study conducted at 2 sites with capillary
blood samples, the following results were obtained by untrained patients:
Results for glucose concentrations less than 75 mg/dL
within ±5 mg/dL| within ±10 mg/dL| within ±15 mg/dL
8/12 (66.7 %)| 12/12 (100 %)| 12/12 (100 %)
Results for glucose concentrations greater than or equal to 75 mg/dL
within ±5 %| within ±10 %| within ±15 %| within ±20 %
63/108 (58.3 %)| 103/108 (95.4 %)| 107/108 (99.1 %)| 108/108 (100 %)
User performance study: In a study conducted at 1 site with Palm AST samples, the following results were obtained by untrained patients:
Results for glucose concentrations less than 75 mg/dL
Alternative Site| within ±5 mg/dL| within ±10 mg/dL| within
±15 mg/dL
Palm| 8/9 (88.9 %)| 9/9 (100 %)| 9/9 (100 %)
Results for glucose concentrations greater than or equal to 75 mg/dL
Alternative Site| within
±5 %
| within
±10 %
| within
±15 %
| within
±20 %
Palm| 175/363 (48.2 %)| 308/363 (84.8 %)| 356/363 (98.1 %)| 362/363 (99.7 %)
User performance study: In a study conducted at 1 site with Upper Arm AST samples, the following results were obtained by untrained patients:
Results for glucose concentrations less than 75 mg/dL
Alternative Site| within ±5 mg/dL| within ±10 mg/dL| within
±15 mg/dL
Upper Arm| 7/10 (70 %)| 10/10 (100 %)| 10/10 (100 %)
Results for glucose concentrations greater than or equal to 75 mg/dL
Alternative Site| within
±5 %
| within
±10 %
| within
±15 %
| within
±20 %
Upper Arm| 156/355 (43.9 %)| 281/355 (79.2 %)| 337/355 (94.9 %)| 351/355 (98.9 %)
Precision
Precision studies using control solutions (day-to-day precision) and blood
(within-system precision) are shown below:
Control solutions | Low | Mid | High |
---|---|---|---|
N | 300 | 300 | 300 |
mean [mg/dL] | 44.9 | 116.6 | 297.4 |
SD [mg/dL] | 1.4 | 2.8 | 6.8 |
CV [%] | 3.1 | 2.4 | 2.3 |
Blood | 1 | 2 | 3 |
--- | --- | --- | --- |
N | 300 | 300 | 300 |
mean [mg/dL] | 40.5 | 81.7 | 132.1 |
SD [mg/dL] | 1.4 | 2.0 | 2.8 |
CV [%] | 3.5 | 2.4 | 2.1 |
Reagent composition†
Mediator | 6.6 % |
---|---|
FAD-GDH enzyme | 21.3 % |
Buffer | 22.6 % |
Stabilizer | 2.3 % |
Non-reactive ingredients | 47.2 % |
†Minimum at time of manufacture
Warranty
Roche warrants that your Accu-Chek Guide test strips will be free from defects
in materials and workmanship until the product expiration date printed on the
label if the test strips are used and stored in the manner described in this
package insert and in your Accu-Chek blood glucose meter User’s Manual. If,
prior to the expiration date of the test strips, there is a defect in
materials or workmanship, Roche will replace the test strips free of charge.
Your sole and exclusive remedy with respect to the test strips shall be
replacement. Any warranty claim should be directed to the Accu-Chek Customer
Care Service Center at 1-800-858-8072.
THE ABOVE WARRANTY IS EXCLUSIVE OF ALL OTHER WARRANTIES, AND ROCHE MAKES NO
OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION THE IMPLIED
WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT
SHALL ROCHE BE LIABLE TO THE PURCHASER OR ANY OTHER PERSON FOR ANY INCIDENTAL,
CONSEQUENTIAL, INDIRECT, SPECIAL OR PUNITIVE DAMAGES ARISING FROM OR IN ANY
WAY CONNECTED WITH THE PURCHASE OR USE OF THE TEST STRIPS. NO WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE, IF ANY IS IMPLIED FOR THE
SALE OF THE TEST STRIPS, SHALL EXTEND FOR A LONGER DURATION THAN THE
EXPIRATION DATE OF THE TEST STRIPS.
Some states do not allow limitations on how long an implied warranty will last
or the exclusion of incidental or consequential damages, so the above
limitation and exclusion may not apply to you. This warranty gives you
specific legal rights, which vary from state to state.
Additional Information
The User’s Manual contains more information.
For questions, contact the Accu-Chek Customer Care Service Center toll-free at
1-800-858-8072. We offer assistance in
many languages.
References
-
FDA Public Health Notification: Use of Finger stick Devices on More than One Person Poses Risk for Transmitting Blood borne Pathogens: Initial Communication, Update November 29, 2010. http://wayback.archive-it.org/7993/20161022010458/
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm224025.htm. Accessed April 22, 2019. -
CDC website on “Infection Prevention During Blood Glucose Monitoring and Insulin Administration,” http://www.cdc.gov/injectionsafety/blood-glucose-monitoring.html. Accessed July 11, 2019.
-
American Diabetes Association website; Diagnosing Diabetes and Learning about Prediabetes. http://www.diabetes.org/diabetes-basics/diagnosis/. Accessed April 22, 2019.
-
American Diabetes Association. 2. Classification and diagnosis of diabetes; Standards of Medical Care in Diabetes-2019. Diabetes Care 2019; 42 (Suppl. 1): S13-S28.
Explanation of Symbols
|
Global Trade Item Number
---|---
|
Serial number
Problems or Questions?
For questions, contact the Accu-Chek Customer Care Service Center toll-free at
1-800-858-8072. We offer assistance in
many languages.
Manufactured in the U.S.A. using U.S. and imported materials.
Roche Diabetes Care, Inc.
9115 Hague Road
Indianapolis, IN 46256
accu-chek.com
U.S. Pat.: http://www.roche-diagnostics.us/patents
ACCU-CHEK, ACCU-CHEK GUIDE, and FASTCLIX are trademarks of Roche. The
Bluetooth® word mark and logos are registered trademarks owned by Bluetooth
SIG, Inc. and any use of such marks by Roche is under license.
MiniMed is a registered trademark of Medtronic, Inc.
© 2020 Roche Diabetes Care