MASiMO LAB12321A0224 Mighty Sat Medical Pulse Oximeter
Read these instructions before use. If you have a serious problem, please contact your competent authority and Masimo.
MEDICAL ELECTRICAL EQUIPMENT
- WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY by ANSI/AAMI ES 60601-1:2005, CAN/CSA C22.2 No. 60601-1:2008, and applicable Particular, (ISO 80601-2-61:2011) and related Collateral (IEC 60601-1-11:2010) Standards for which the product has been found to comply by Intertek.
Patents: www.masimo.com/patents.htm.
- Masimo, , MightySat, MightySat OTC, Signal Extraction Technology, SET, and Signal I.Q. are federally registered trademarks of Masimo Corporation.
- All other trademarks and registered trademarks are the property of their respective owners.
- © 2024 by Masimo Corporation.
Glossary
- Apnea: Describes a condition where breathing has stopped.
- Hemoglobin: The protein in red blood cells that moves oxygen from the lungs throughout the body and carbon dioxide from the tissues back to the lungs.
- Necrosis: The death of most or all of the cells in a tissue due to disease, injury, or failure of the blood supply.
- Oxygen Saturation: The percentage of hemoglobin in blood that is bound to oxygen.
- Parameter: A parameter is an element of a system that is critical in evaluating the system’s condition.
- Perfusion: The bodily process of delivering blood to the capillary beds.
- Pulse Oximeter: A medical device that provides oxygen saturation and pulse rate based on how light passes through the body.
- Quality of Service (QoS): Description or measurement of the overall performance of the Bluetooth communication between MightySat and the mobile device as seen by the user.
Device Description
Description of the MightySat
The MightySat® Medical Pulse Oximeter is for self-monitoring of SpO2 and pulse
rate by individuals 18 years and older. Key features available:
- Masimo SET® technology for SpO2 and pulse rate in no motion and low perfusion.
- Bluetooth® LE for the wireless connection to smart devices.
Understanding Health Data Parameters
The following information describes the measurement feature data collected.
Feature | Explanation |
---|---|
Oxygen Level (SpO 2 ) | Percentage of usable oxygen in your blood. |
Low oxygen (< 90%) can be a sign of a problem that may be related to your
breathing or blood circulation.
Pulse Rate (PR)| A measure of the number of times your heart beats per
minute. The average pulse rate is typically 65 to 100 beats per minute.
Perfusion Index (Pi)| Feature that reflects the relative strength of
the signal used for the SpO2 and Pulse Rate. The Pi will increase with better
blood circulation. Your SpO2 values may be more reliable when your Pi value is
higher. Pi values of 0.25 or below are typically considered low. If your Pi is
low, try warming your hands or using a different finger.
Who is MightySat for?
Feature (Use) | Who is it for? |
---|---|
General | Individuals 18 years and older |
SpO 2 | Individuals 18 years and older |
Pulse Rate | Individuals 18 years and older |
What does the MightySat do and Who is MightySat for (Intended Use and Indications for Use)
- The Masimo MightySat® Medical Pulse Oximeter is intended for the spot-checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR).
- The Masimo MightySat® Medical Pulse Oximeter is indicated for use with individuals 18 years and older who are well or poorly perfused under no motion conditions.
- The Masimo MightySat® Medical Pulse Oximeter is not intended for diagnosis or screening of lung disease and treatment decisions using the device should only be under the advice of a healthcare provider.
Self-Monitoring Do’s and Do Not’s
- Do self-monitor your oxygen and pulse rate as recommended by your healthcare provider. This does not mean diagnosing and treating.
- Do check your finger can cover the bottom pad of the MightySat and close it comfortably.
- Do wait a few seconds for your oxygen and pulse rate values to become brightly displayed and steady for the most reliable values.
- Do familiarize yourself with your normal (baseline) SpO2 levels. Focus on changes from your baseline over time and not solely on a single measurement at any given moment.
- Do seek medical attention if you are not feeling well even if your readings are normal.
- Do review the information on the limitations and ways to improve the accuracy of the reading provided in this manual.
- Do Not self-diagnose and treat yourself. Your physician uses the measurements along with other symptoms and your medical history in their treatment decision. This cannot be replaced by a pulse oximeter.
- Do Not change your treatment plan on your own. Never alter oxygen delivery settings or medical treatment prescribed by your physician.
- Do Not rely only on the readings. Reading can provide a false sense of security to a lung or health condition that has not yet affected your blood oxygen.
- Do Not use as an apnea monitor. Oxygen changes may be delayed from when your breathing stops.
- Do Not use for continuous monitoring or for long periods. No alarms are provided and prolonged pressure on the finger can affect blood circulation.
General Warnings and Precautions
Before use, read the following carefully.
Safety Warnings
- WARNING: Do not use if it appears damaged or if you think it is damaged. Exposed electrical circuits can lead to harm.
- WARNING: Do not adjust, repair, open, disassemble, or modify MightySat. Such changes may lead to injury and/or false readings.
- WARNING: Do not use in areas filled with flammable gases: such as anesthetics, oxygen, oxygen-enriched environments, or nitrous oxide to prevent the risk of fire.
- WARNING: Do not use MightySat during magnetic resonance imaging (MRI) or in an MRI environment.
- The device contains materials that can be attracted by the MR magnet core which can make it a risk of projectile injury.
- Metal components can heat up during MR scanning which can present thermal injury and burns.
- Artifacts can be created in the MR image.
- Strong magnetic fields may prevent the device from operating properly.
- WARNING: Do not attempt to secure it to the finger using any external means (e.g., tapes or bands). Too much pressure to a finger can cause skin injury.
- WARNING: Avoid using it for long periods. Skin injury or placement changes may result in inaccurate readings.
- WARNING: Keep small parts away from small children. Small items such as the battery door, battery, and lanyard may become a choking hazard.
- WARNING: Do not use the lanyard if it can get wrapped around the neck. Strangulation may occur.
- WARNING: The following can cause damage that can create a safety or performance issue with your MightySat:
- Do not set it on a wet surface.
- Do not immerse it in liquid.
- Do not attempt to sanitize or sterilize using heat, steam, boiling or similar means.
- Clean only with the solutions listed in this manual.
- Do not clean while in use.
- CAUTION: Do not use devices that are sensitive to magnets (e.g., pacemakers). The magnet provided with the MightySat could hinder the proper operation of the device.
- Note: Use and store as directed in the Specifications section in this manual.
Performance Warnings
-
WARNING: Do not self-diagnose or self-medicate based on the measurements. Always consult your doctor.
-
WARNING: See the Risk and Benefits section for information on limitations and ways to get the most accurate readings.
-
WARNING: MightySat is not intended to be used as an apnea monitor.
-
WARNING: If any measurement seems questionable, check your vital signs by alternate means and then check that the MightySat is functioning properly. Contact your doctor if you are still unsure about your measurements.
-
WARNING: MightySat is not designed to be used during medical procedures such as surgeries, Magnetic Resonance Imaging (MRI) sessions, and
cardioversion/defibrillation procedures. -
WARNING: MightySat is not intended for use during electrocautery.
-
CAUTION: The displayed parameter data may require additional time to stabilize to get the most accurate readings.
-
CAUTION: When using a smart device, keep both devices within the recommended range of each other. See the Specifications section for details. Moving outside of this range may cause a loss in connection.
-
CAUTION: Move the devices away from sources that may interfere with Bluetooth connection. The presence of other devices that may create radio frequency interference
-
(RFI). This may result in a loss of Quality of Service. Examples of devices that cause RFI are other smart devices, devices with remote controls, and baby monitors.
-
Note: The display may be difficult to view when exposed to direct sunlight or bright lights.
-
Note: The display will shut off automatically if there are no readings.
Pulse Oximetry Performance
- WARNING: If SpO2 values indicate hypoxemia, do not self-diagnose. Confirmation by a medical professional is required.
- WARNING: SpO2 is empirically calibrated in healthy adult volunteers with normal levels of carboxyhemoglobin (COHb) and methemoglobin (MetHb).
- CAUTION: Optical, pleth-based measurements (e.g., SpO2 and PR) can be affected by the following:
- Incorrect placement or securement (e.g., too loose, too tight).
- Blood pressure cuff inflated or constricting the same arm.
- Intravascular dyes such as indocyanine green or methylene blue.
- Arterial catheter
- Venous congestion
- Abnormal venous pulsations (e.g., tricuspid valve regurgitation, Trendelenburg position).
- Abnormal pulse rhythms due to physiological conditions or induced through external factors (e.g., cardiac arrhythmias, intra-aortic balloon, etc.).
- Moisture, birthmarks, skin discolouration, tattoos, henna, or foreign objects in the light path.
- Elevated levels of bilirubin
- Physiological conditions that can significantly shift the oxygen disassociation curve.
- A physiological condition that may affect vasomotor tone or changes in vasomotor tone.
- Loss of pulsatile signal
CAUTION: No or inaccurate SpO2 readings may additionally be caused by:
- Elevated levels of COHb and/or MetHb. Note: High levels of COHb or MetHb may occur with a seemingly normal SpO2..
- Severe anemia.
- Very low arterial perfusion.
- Hypocapnic or Hypercapnic conditions.
- Excessive motion.
- Vasospastic disease such as Raynaud’s.
- Hemoglobinopathies and synthesis disorders (e.g., thalassemias, Hb s, Hb c, sickle cell).
- Peripheral vascular disease.
- EMI radiation interference.
Cleaning and Service Warnings
WARNING: To prevent damage:
- Remove batteries when not in use for more than 30 days
- Replace both batteries at the same time to avoid mixing fully and partially charged batteries.
- Use only AAA alkaline batteries.
WARNING: Dispose of the battery according to local laws.
FCC STATEMENT
Compliance Warnings
WARNING: Any changes or modifications not expressly approved by Masimo
shall void the warranty for this equipment and could void the user’s authority
to operate the equipment.
WARNING: The frequency bands of this device (2.4 GHz and 5.15 to 5.25
GHz) are only for indoor use, under international telecommunication
requirements.
WARNING: Per international telecommunication requirements, the frequency
band of 2.4 GHz is only for indoor usage to reduce the potential for harmful
interference to co-channel mobile satellite systems.
CAUTION: Comply with local laws and regulations when disposing of
MightySat, its accessories, and its battery.
Note: When using with a device with wireless features, consideration
should be taken to local government frequency allocations and technical
parameters to minimize the possibility of interference to/from other wireless
devices.
Note: This device complies with Part 15 of the FCC Rules. Operation is
subject to the following two conditions:
- This device may not cause harmful interference, and
- this device must accept any interference received, including interference that may cause undesired operation.
Note: This equipment has been tested and found to comply with the limits for a Class B digital device, according to part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used under the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and the receiver.
- Consult the dealer or an experienced radio/TV technician for help.
Note: This equipment has been tested and found to comply with the Class B
limits for medical devices according to the IEC 60601-1-2. These limits are
designed to provide reasonable protection against harmful interference in all
establishments, including domestic establishments.
Note: This Class B digital apparatus complies with Canadian ICES-003.
Risks and Benefits
See the Benefits and Risks of the MightySat below.
Your Body and Oxygen
When we breathe, oxygen moves in our blood around our body. Pulse oximeters
like the MightySat help to identify the oxygen level in the blood. A normal
resting SpO2 levels are typically 95% or greater. However, normal values can
be lower for individuals with lung disease, advanced age, or those living at
high altitudes. A SpO2 below 90% is typically considered low and outside the
normal range, but it is important to not self-diagnose. Consult your
healthcare provider.
- Low SpO2 Less than 90% of SpO2
- Additionally, if your oxygen level is trending lower, or you are not feeling well, seek the advice of your physician.
Risks of MightySat
- Inappropriate use or misuse of the device. You should follow the information provided in this manual and not self-diagnosis or treat yourself. Always seek the advice of your physician if you are not feeling well.
- With electrical devices, there may be a risk of electrical hazards. These risks are reduced by the design and testing of MightySat.
- As with all devices with small parts, there is a risk of a child swallowing and choking. Keep small parts away from small children to prevent such hazards.
- As with all items applied directly to the skin, there is a risk of skin irritation and discomfort.
Benefits of MightySat
- MightySat is a portable medical device able to provide clinically validated oxygen saturation (SpO2) and pulse rate (PR).
- MightySat wirelessly connects to a smart device for storage and review of data.
Use of Pulse Oximeters and their Limitations
Pulse oximeters provide blood oxygen levels, SpO2, based on how light shines
through the finger. The accuracy of your SpO2 depends on how well the light
through the finger is measured.
Limitations
Things that interfere with the light can affect your SpO2 accuracy. Some
things can be controlled, while others need awareness. Controllable things:
- Choose a finger that completely covers the sensor.
- Not wearing nail polish.
- Keeping fingers dry and free of foreign objects.
- Not blocking the light through your finger.
- Avoiding direct exposure to bright lights, including direct sunlight.
- Keeping your finger still.
- Warming your hands to improve circulation.
- Avoid use on the same arm with the blood pressure cuff inflated.
- Keep away from other electrical equipment that may cause readings to be affected (e.g., microwaves, strong radio transmitters).
Other things that require awareness:
- Skin pigment or colour.
- Skin or finger thickness.
- Health conditions affecting how oxygen is carried in your blood (e.g., sickle cell, severe anemia).
- Poor blood circulation.
- Presence of blood components not able to carry oxygen (e.g., elevated carbon monoxide levels in your blood, recent tobacco use).
- Age and medical history.
Description
Parts List
These items should be included with your MightySat:
Item | Description |
---|---|
1 | MightySat |
2 | Batteries |
3 | Lanyard |
4 | Carrying Case |
MightySat Features
Item | Description | Function |
---|---|---|
1 | Enclosure Clip | Attach a lanyard to the clip. |
2 | Bluetooth Indicator (Optional) | Displays when Bluetooth is enabled on |
the device.
3| Battery Status Indicator| Displays remaining battery life.
4| Main Screen| Display for measurements and indicators.
Note: Numbers are dim when confidence in the measurement is low.
5| Waveform and SIQ or Pulse Bar| When the waveform option is turned on, the Waveform and SIQ line display. The SIQ line will be taller with greater confidence in the readings displayed. If your SIQ is low, try keeping your finger still.
When the waveform option is turned off, the Pulse Bar displays the pulse rate. The taller the Pulse Bar the greater the confidence in the readings displayed.
For more reliable SpO2 values, use a SpO2 value with a taller SIQ line and a higher Pi value that is above 0.25.
6| Touchpad| User interface for settings (see Using the Touchpad on page 13 in this manual).
Basic Setup and Use
- The following steps include basic setup and use for operating the MightySat device.
The MightySat requires two alkaline AAA batteries to operate. To install batteries, follow the steps below:
Note: MightySat will not work if the batteries are inserted backwards.
- Place the MightySat with the display screen facing down.
- Push lightly on the battery button (identified by the arrow) to release and remove the battery cover.
- Insert two new AAA alkaline batteries to match the labels (+ and -) inside of the battery case.
- Snap the battery door back onto the device.
CAUTION: Ensure that the battery door is installed before use.
- MightySat turns on when the device is opened to place on a finger.
-
To open the MightySat, squeeze the back of the device (1).
-
Insert the ring finger with the display facing up (2). For a right-handed person, this would be the left-hand ring finger. For a left-handed person, this would be the right-hand ring finger.
-
Once the device is correctly placed on the finger, release the device to close. The display should be facing up (3).
-
The tip of the finger should touch the finger stop as shown in the image below.
Note: Ensure the finger is correctly positioned. -
Check the fit on the finger by making sure the finger can cover the entire inside bottom pad. If your finger is too small, choose a different finger. If no finger is big enough, then the MightySat should not be used.
Note: The top half of MightySat should rest comfortably on the top of the
finger. If MightySat feels tight and your finger feels uncomfortable, try a
smaller finger as your finger may be too thick.
CAUTION: Do not press the top of the device while on the finger.
WARNING: Do not attempt to secure the MightySat to the finger using
additional pressure.
- After correctly placing on the finger, readings will display.
- If you are recording your SpO2 and Pulse Rate values, wait a few seconds for the SpO2 and Pulse Rate values to be brightly lit and steady before recording.
Note: The SpO2 and Pulse Rate values will appear dim until reliable SpO2 and Pulse Rate are available. If readings are not displayed, see Troubleshooting.
Your Readings
-
Normal resting SpO2 levels are typically 95% or greater. However, normal values can be lower for individuals with lung disease, advanced age, or those living in high altitudes. Please consult your healthcare provider to define your normal values.
-
Changes or trends in SpO2 may be more meaningful than a single SpO2 value.
-
Pay attention to other signs or symptoms of low oxygen level, such as:
-
Bluish colouring in the face, lips, or nails
-
Shortness of breath, difficulty breathing, or a cough that gets worse
-
Restlessness and discomfort
-
Chest pain or tightness
-
Fast or racing pulse rate
-
The MightySat turns off after removing it from the finger.
Advanced Setup and Use
Using the Touchpad
The multi-function Touchpad on MightySat is located below the display screen.
Note: The display is not a touch screen.
Desired Function | Required Action | Description |
---|---|---|
Rotate the Main Screen for a better view while on the finger.* | Tap and | |
release the Touchpad to rotate once in a clockwise direction. | The Main Screen |
can be rotated for the best viewing angle.
|
Enter the Menu Screen.| Press and hold the Touchpad.| The menu allows changes
to MightySat settings. See Main Menu Options
Navigate the Menu Screen.| Swipe left or right on the Touchpad.| Switches
between menu items on the Menu Screen.
Select a menu item.| Touch and release the Touchpad.| Select an item on the
Menu Screen to switch between options or enable/disable that option.
See Main Menu Options
Exit the Menu Screen.| Swipe right on the Touchpad to display the arrow, then touch and release.| Returns to the Main Screen.
See Main Menu Options
- This function does not rotate the Menu Screen.
Main Menu Options
Use the Touchpad to navigate the Menu Options. See Using the Touchpad The Menu options are:
Menu Icon| Menu Option| Description| Default Setting|
Options
---|---|---|---|---
|
Back
|
Return to Main Screen.
|
N/A
|
N/A
|
Waveform
|
Allows the user to choose to display the waveform on the screen.
|
On
|
On or Off *
|
Brightness
|
Change the brightness of the display.
|
100%
|
25%, 50%, 75%, and
100%
|
Bluetooth (Optional)
|
For connection with a smart device. Enables or disables Bluetooth LE.
|
On
|
On or Off *
|
About
|
Hardware and software information about the device. Serial number; software version; Bluetooth LE Mac Address
|
N/A
|
N/A
When On, the icon is white, when Off, the icon is grey (dimmed).
Connecting to a Smart Device via Bluetooth (Optional)
Note: Bluetooth LE is a feature available for use with compatible smart
devices. For a full list of compatible smart devices, see www.masimo.com.
Bluetooth Connection
The MightySat provides a Bluetooth LE connection to a smart device. The
Bluetooth connection is only available with the Masimo App. Bluetooth icon
will appear when connected. MightySat connects to a single device at a time to
reduce risk of unauthorized access.
Note: The MightySat requires the use of the Masimo App to connect to a
smart device.
Pair MightySat to Smart Device
-
Ensure Bluetooth is enabled on the smart device through the device settings.
-
From your compatible smart device, perform one of the following:
- For AndroidTM-powered devices, go to the Google PlayTM store.
- For Apple® devices, go to the App StoreSM.
-
Search and download the “Masimo” app.
-
Launch the Masimo App.
-
Turn the MightySat Bluetooth On. See Main Menu Options on page 13 of this manual for further instructions.
-
Follow the Masimo App on-screen instructions to pair a device.
-
When the Masimo App identifies the MightySat, press/select it to pair.
-
Once MightySat is connected to a smart device, the Masimo App returns to the Main screen.
Note: A connection icon will appear on the MightySat device when a Bluetooth connection has been established. -
Place MightySat on the patient’s finger. Confirm that readings on MightySat and readings displayed on the Masimo App are the same without a delay greater than 10 seconds.
Note: If the delay is greater than 10 seconds, move the MightySat closer
to the smart device and repeat the connection process.
Note: Turn off the Bluetooth on the MightySat to prevent unauthorized
connections when not used.
CAUTION: When using Bluetooth, keep both devices within range of each
other. See Specifications for details. Moving out of range may cause a loss in
connection with the smart device.
CAUTION: When using Bluetooth, move the devices away from sources that
may interfere with the connection. Interference may result in loss of Quality
of Service. See Specifications on page 15 for details.
Verify Paired MightySat
-
On the smart device, access the Masimo Personal Health app Options.
-
Locate Paired Device.
Note: Sensor Mode is always Bluetooth Sensor when using MightySat with a smart device. -
Compare the Paired Device information to the BT SN (Bluetooth Serial Number) displayed on the MightySat About screen, see Main Menu Options on page 13 for information on accessing the About screen.
Disconnect Paired MightySat
- On the smart device, access the Masimo Personal Health app Options.
- Press/select the Paired Device information.
- Select Forget this Device. MightySat will be disconnected from the smart device. MightySat will need to be paired if it is to be used with this smart device again.
Troubleshooting
Error or Error Message| Possible Causes| Recommended
Solutions
---|---|---
A red battery symbol displays on Main Screen| Low battery| Replace low
batteries as soon as possible. (see Installing the AAA Batteries in this
manual)
Device does not display readings| · Incorrect finger placement
· Incorrect battery orientation
· No battery
· Low battery
· Environmental influences
| · Wait for measurement.
· Reposition finger (see Using MightySat in this manual)
· Re-orient batteries
· Replace with new batteries
· Relocate device
· Contact Masimo Technical Services
Device display does not turn on| · No battery
· Incorrect battery orientation
· Device damaged
| · Replace with new batteries
· Re-orient batteries
· Contact Masimo Technical Services
Numbers appear dim| · Brightness set low
· Exposed to bright lights or sunlight
· Incorrect finger placement
· Measurement site may be poorly perfused
| · Check brightness setting in menu
· Relocate device so that it is not directly under bright lights
· Reposition finger (See Using MightySat in this manual)
· Contact Masimo Technical Services
Device keeps turning off while on the finger| · Incorrect finger placement
· Environmental influences
· Device damaged
| · Reposition finger (See Using MightySat in this manual)
· Relocate device
· Replace with new batteries
· Contact Masimo Technical Services
Measurement does not display on the smart device using optional Bluetooth LE| · Bluetooth LE not connected
· Compatible app not installed on smart device
· Device damaged
· Smart device damaged
| · Confirm Bluetooth LE is on for the MightySat and the smart device
· Confirm a compatible app is installed on the smart device
· Close and re-launch the compatible app on the smart device
· Check that MightySat is connected to the correct smart device
· Contact Masimo Technical Services
Specifications
Display Ranges
Parameter | Display Ranges |
---|---|
SpO2 (Oxygen Saturation) | 0% to 100% |
PR (Pulse Rate) | 25 bpm to 240 bpm |
Pi (Perfusion Index) | 0.02% to 20% |
The MightySat includes LEDs that emit wavelengths range from 600 nm to 1000 nm
with a peak optical power less than 15 mW that are used to estimate your
oxygen saturation. This information about the wavelength range may be more
useful for your healthcare professional.
Performance Specifications
- Accuracy root mean squared, or ARMS, is used to show how well a pulse oximeter provides SpO2 compared to a blood sample, SaO2. The ARMS is the potential error that can be expected between the SpO2 and SaO2. The lower the ARMS value the better the accuracy.
- The ARMS is based on clinical studies on healthy volunteers as their oxygen is lowered. The ARMS is the value where at least 2/3 (67%) of the differences in SaO2 to SpO2 fell.
- For the MightySat, the ARMS testing was done under 2 different conditions; no motion and low perfusion.
Condition | Description |
---|---|
No Motion | Volunteers keep their fingers still during measurements. |
Low Perfusion | Simulator used to simulate very low blood circulation during |
measurements and different skin tones.
Below is the ARMS for the MightySat.
SpO 2 Accuracy
Condition| Range| Population| A RMS *****
No Motion [1]| 70% to 100%| Adults, Pediatrics| 1.5%
Low perfusion [2]| 70% to 100%| Adults, Pediatrics| 2%
See the SpO2 Performance Study Summary on page 16 for additional SpO2 accuracy information.
Pulse Rate (PR)
Condition| Range| Population| A RMS *****
No Motion [3]| 25 bpm to 240 bpm| Adults, Pediatrics| 3 bpm
Low perfusion [3]| 25 bpm to 240 bpm| Adults, Pediatrics| 3 bpm
- ARMS accuracy is a statistical calculation of the difference between device measurements and reference measurements. Approximately two-thirds of the device measurements fell within +/- ARMS of the reference measurements in a controlled study.
SpO2 Performance Study Summary
Table below shows the results from the SpO2 clinical study. The study used 39 healthy volunteers (22 dark skin and 17 light skin). Data was collected on each volunteer while at different oxygen levels when their finger was still. The overall measured ARMS value was within1.5% under no motion.
Overall Measured A RMS Value
Range| Bias| Precision| A RMS| LOA
70% – 100%| -0.02| 1.26| 1.26| [-2.50 2.45]
- The following are the results for each 10% grouping.
Measured A RMS Values
Range| A RMS| Npairs
90% – 100%| 1.16 %| 894
80% – 90%| 1.19 %| 646
70% – 80%| 1.52 %| 490
- The Bland-Altman plot below shows the difference between SpO2 and SaO2 during the study. Between the two dotted lines is where 95% of the differences were found, or the limits of agreement. This is different from the ARMS value which is set at about 67%. This generally means about 67% of SpO2 values were within 1.26% of the blood sample values and about 95% of the SpO2 values were within 2.48 to -2.60% of the blood sample values. For example, if the SpO2 value reads 95%, then the true oxygen saturation in the blood can be generally thought to be between 92% to 98% based upon the limits of agreement and 93% to 97% based upon the ARMS.
- The dotted lines on the Bland-Altman plot also help to identify data points that are outside of the limits of agreement, or potential outliers. The upper dotted line at 2.48% and lower dotted line at -2.60% shows that during the study the SpO2 value was about equally above and below the true oxygen saturation in the blood. In other words, there is an equal chance that the true oxygen saturation can be higher or lower than the SpO2 value you see. The plot is provided to help you understand that just because you see one value on the MightySat that it will always match the value from a blood test.
Specific SaO2 Range of 85-95%
Bias, Precision, ARMS, and LoA results were also computed for 85-95% SaO2. The
performance in this SpO2 range may help understand how well a pulse oximeter
can work when the blood oxygen is in the transition between normal to low
oxygen levels.
Sat Level| Bias| Precision| A RMS| LOA|
Npairs
---|---|---|---|---|---
[85-95]| -0.35| 1.17| 1.22| [-2.63 1.94]| 738
Subgroup Analysis: By Pigment and Gender
The data from the study was also analyzed by gender and skin pigment. The skin
pigments were grouped using the color-based Massey-Martin scale (1-10). The
subjects with a score of 1-3 were grouped as “Light”, and subjects with a
score of 4-10 were categorized as “Dark”. The performance by gender and skin
pigment may help understand how well a pulse oximeter can work on different
people. Comparing the ARMS and LoA shows that MightySat performed similarly on
all groups.
Group| Bias| Precision| A RMS| LOA|
Npairs
---|---|---|---|---|---
Light Male| -0.13| 1.26| 1.27| [-2.61 2.35]| 384
Dark Male| 0.11| 1.44| 1.44| [-2.70 2.93]| 612
Light Female| -0.06| 1.11| 1.11| [-2.11 2.23]| 258
Dark Female| -0.18| 1.30| 1.32| [-2.73 2.38]| 216
Figure 2 provides a scatter plot, which shows the relationship between the
SpO2 to the SaO2 blood reference for different subgroups (i.e., Light Male,
Dark Male, Light Female and Dark Female). The best result would be a diagonal
line that passes through the same value for SpO2 and SaO2. For example, a line
that passes through 90% SpO2 and SaO2, 80% SpO2 and SaO2, and 70% SpO2 and
SaO2.
While the Bland Altman plot (Figure 1) shows the difference between the SpO2
and SaO2 values, the scatter plot (Figure 2) shows the relationship between
the SpO2 and SaO2 values. Using both plots can be helpful to understand the
SpO2 performance on different people. For example, from the Bland Altman plot
(Figure 1), you see that your SpO2 value can be about +/- 2.5% different from
the SaO2 value based upon the limits of agreement. While the scatter plots
help to see if your SpO2 is more likely to be higher or lower than the SaO2
based upon your gender and skin pigment group and your SpO2 value. For
example, if you are a male with dark skin pigment (Massey Martin score 4 or
greater) and your SpO2 value is 95% you might expect your SaO2 value to be
between 92-98%. This is because the scatterplot shows at 95% SpO2, SaO2 values
are between 98-92% as part of the study. However, at a SpO2 value of 75%, you
might expect your SaO2 value to be around 70-75%. This is because the
scatterplot at 75% SpO2 shows more SaO2 values between 75-70%. Fortunately,
the performance of the MightySat is similar for all subgroups.
Battery Life
Item | Description |
---|---|
Operating | 1.5 Volt AAA Battery (2) |
Battery Life | ≥15 hours (screen brightness at 50%) |
Expected Service (Useful) Life
Item | Description |
---|---|
Expected Service (Useful) Life | 2 Years |
Environment
Item | Description |
---|---|
Operating Temperature | 5°C to 40°C (41°F to 104°F) |
Storage Temperature | -40°C to 70°C (-40°F to 158°F) |
Operating Humidity | 10% to 95% non-condensing |
Storage Humidity | 10% to 95% non-condensing |
Atmospheric pressure | 540 mBar to 1060 mBar |
Physical Characteristics
Item | Description |
---|---|
Dimensions | 3” x 1.7” x 1.3” |
(7.6 cm x 4.3 cm x 3.3 cm)
Weight without Battery| 0.2 lbs. (90g)
Compliance
Safety Compliance
ANSI/AAMI ES60601-1 CSA C22.2 No. 60601-1 IEC/EN 60601-1
IEC 60601-1-6
IEC 60601-1-11
ISO 80601-2-61
EMC Compliance
IEC 60601-1-2, Class B
ISO 80601-2-61: Clause 202, 20 V/m radiated immunity
Equipment Classifications per IEC 60601-1
Degree of Protection against electric shock| Type BF applied part
Mode of Operation| Continuous Operation
Degree of Protection from Liquid Ingress| IP23, Protection from ingress of
particulates > than 12.5 mm and ingress from spraying water.
Environment| Not for use in the presence of flammable anesthetics
Bluetooth LE Wireless Technology Information
Bluetooth LE Wireless Technology Information
Modulation Type| GFSK
Max. Output Power| -1 dBm
Frequency Range| 2402 MHz – 2480 MHz
Antenna Peak Gain| -7 dBi
Recommended Range| ~10 feet (~3 meters) line-of-sight
Quality of Service (QoS)| Delay <10 seconds
Security| Proprietary binary protocol
Radio Compliance
Radio Modes| Bluetooth LE
USA| FCC ID: VKF-MSAT01A
FCC parts 15.207 and 15.247
Canada| IC-7362A-MSAT01A RSS-210
Europe| EN 300 328
EN 301 489-17
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
Guidance and Manufacturer’s Declarations – Electromagnetic Emissions
The ME Equipment is intended for use in the electromagnetic environment
specified below. The customer or the user of the ME Equipment should assure
that it is used in such an environment.
Emission Test| Compliance| Electromagnetic Environment –
Guidance
RF Emissions CISPR 11| Group 1| ME Equipment uses RF energy only for its
internal function. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF Emissions CISPR 11| Class B| This device is not intended to be operated in
the home environment. This device has not been evaluated for use in aircrafts.
Guidance and Manufacturer’s Declarations – Electromagnetic Emissions
Harmonic emissions IEC 61000-3-2| N/A|
Voltage fluctuations/flicker emissions
IEC 61000-3-3
|
N/A
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
Recommended Separation Distances
Recommended Separation Distance Between Portable and Mobile RF Communication Equipment and the ME Equipment
The ME Equipment is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the
ME Equipment can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment
(transmitters) and the ME Equipment as recommended below, according to the
maximum output power of the communication equipment.
Rated maximum output power of transmitter (W)| Separation Distance According
to Frequency of Transmitter (m)
150 kHz to 80 MHz
| 80 MHz to 800 MHz
| 800 MHz to 2.5GHz
0.01| 0.12| 0.018| 0.035
0.1| 0.37| 0.057| 0.11
1| 1.17| 0.18| 0.35
10| 3.7| 0.57| 1.1
100| 11.7| 1.8| 3.5
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1 : At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 : These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Symbols
The following symbols may be found on the MightySat, or packaging and are
defined below.
Citations
- The MightySat® Medical Pulse Oximeter has been validated for no motion accuracy in human blood studies on healthy adult male and female volunteers with light to dark pigmented skin in induced hypoxia studies in the range of 70% to 100% SpO2 against a laboratory co-oximeter.
- The MightySat® Medical Pulse Oximeter has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2 simulator and Masimo’s simulator with signal strengths of greater than 0.02% and transmission of greater than 5% for saturations ranging from 70% to 100%.
- The MightySat® Medical Pulse Oximeter has been validated for pulse rate accuracy for the range of 25 bpm to 240 bpm in bench top testing against a Biotek Index 2 simulator and Masimo’s simulator with signal strengths of greater than 0.02% and transmission of greater than 5% for saturations ranging from 70% to 100%.
Cleaning, Disinfecting, and Service
Cleaning and Disinfecting
- WARNING: Before cleaning, read Cleaning and Service Warnings on page 6.
- WARNING: Before cleaning, make sure the device is off and is not applied to a finger.
- CAUTION: Thoroughly clean and low level disinfect before applying it to on a new person.
- Note: Before cleaning, remove the batteries and make sure the battery cover is re-attached correctly.
- As good practice, you should clean the device if you see visible debris and disinfect before use each day.
- To clean, follow the instructions below:
- Wipe the outer surfaces using a dampened soft cloth twice with one of the recommended cleaning solutions listed below, or until the surfaces are free of any visible residue.
- Note: Pay particular attention to cracks, crevices, and hard to reach areas of the device.
- Repeat the above cleaning step using a fresh wipe.
- Allow to dry thoroughly before using again.
- To conduct low level surface disinfection of the MightySat, follow the instructions below:
- Note: Follow cleaning instructions prior to disinfecting the device.
- Visibly wet the sensor pads and outer surfaces using a soft cloth dampened with a 10% (1:10) chlorine bleach to water solution.
- Allow the solution to sit for 10 minutes on the sensor pads before wiping them with a dry soft cloth.
- Allow the MightySat to dry thoroughly before using again.
- The surfaces of the MightySat may be cleaned with the following:
- 70% Isopropyl Alcohol
- Cidex Plus (3.4% glutaraldehyde)
- 10% (1:10) chlorine bleach to water solution
- If you see a color change to the finger pads, the device may be at the end of its reuse life. Performance may start to become affected. If this occurs, you should stop using the device.
- CAUTION: To avoid permanent damage to the MightySat, do not use undiluted bleach (5% – 5.25% sodium hypochlorite) or any other cleaning solution not recommended.
Service
WARNING: Do not attempt to repair the MightySat as this may cause damage
to the device and prevent it from operating properly.
If the device does not appear to be operating correctly, see Troubleshooting
on page 15 section in this manual.
Note: To maintain the proper functionality of the battery compartment and avoid possible damage from alkaline batteries that may leak, remove batteries from the device when not in use for long periods of time.
Product Support
For additional help, contact Masimo Technical Services at
949-297-7498. Local contact information
can be found at http://service.masimo.com.
Limited Warranty
- Masimo warrants to the original end-user purchaser the Masimo-branded hardware product MightySat® Medical Pulse Oximeter Pulse Oximeter and any software media contained in the original packaging against defects in material and workmanship when used in accordance with Masimo’s user manuals, technical specifications, and other Masimo published guidelines for a period of 24 months from the original date the Product was obtained by the end-user purchaser.
- Masimo’s sole obligation under this warranty is the repair or replacement, at its option, of any defective Product or software media that is covered under the warranty.
- To request a replacement under warranty, Purchaser must contact Masimo and obtain a returned goods authorization number so that Masimo can track the Product. If Masimo determines that a Product must be replaced under warranty, it will be replaced and the cost of shipment covered. All other shipping costs must be paid by purchaser.
- The above described warranty is in addition to any statutory rights provided to Purchaser under applicable laws and regulations of the region in which the product was sold to the extent that those rights cannot be disclaimed and are superseded by the above described warranty to the extent permitted under applicable laws and regulations of the region in which the product was sold.
Exclusions
The warranty does not apply to any non-Masimo branded product or any software,
even if packaged with the Product, or any Product that was: (a) not new or in
its original packaging when supplied to purchaser; (b) modified without
Masimo’s written permission; (c) supplies, devices, or systems external to the
Product; (d) disassembled, reassembled, or repaired by anyone other than a
person authorized by Masimo; (e) used with other products, like new sensors,
reprocessed sensors, or other accessories, not intended by Masimo to be used
with the Product; (f) not used or maintained as provided in the operator’s
manual or as otherwise provided in its labeling; (g) reprocessed,
reconditioned, or recycled; and (h) damaged by accident, abuse, misuse, liquid
contact, fire, earthquake or other external cause.
No warranty applies to any Product provided to Purchaser for which Masimo, or
its authorized distributor, is not paid; and these Products are provided AS-IS
without warranty.
Limitation of Warranty
Except as otherwise required by law or altered by the purchase agreement, the
above warranty is the exclusive warranty that applies to the Product and
software media, and Masimo does not make any other promises, conditions, or
warranties regarding the Product. No other warranty applies, express or
implied, including without limitation, any implied warranty of
merchantability, fitness for a particular purpose, satisfactory quality, or as
to the use of reasonable skill and care. In so far as the above warranties
cannot be disclaimed, Masimo limits the duration and remedies of the
warranties to the duration and to the remedies set forth above and as
permitted by law. See the licensing terms for the terms and conditions that
apply to and Software accompanying the Product. Additionally, Masimo will not
be liable for any incidental, indirect, special, or consequential loss,
damage, or expense arising from the use or loss of use of any Products or
Software. In no event shall Masimo’s liability arising from any Product or
Software (under contract, warranty, tort, strict liability, or otherwise)
exceed the amount paid by purchaser for the Product or Software. The above
limitations do not preclude any liability that cannot legally be disclaimed by
contract.
Sales & End-User License Agreement
This document is a legal agreement between you (“purchaser”) and Masimo
Corporation (“Masimo”) for the purchase of this Product (“Product”) and a
license in the included or embedded Software (“Software”) except as otherwise
expressly agreed in a separate contract for the acquisition of this Product,
the following terms are the entire agreement between the parties regarding
your purchase of this Product. If you do not agree to the terms of this
agreement, promptly return the entire Product, including all accessories, in
their original packages, with your sales receipt to Masimo for a full refund.
Restrictions
- Copyright Restrictions: The Software and the accompanying written materials are copyrighted. Unauthorized copying of the Software, including Software that has been modified, merged, or included with other software, or the written materials is expressly forbidden. Purchaser may be held legally responsible for any copyright infringement that is caused or incurred by Purchaser’s failure to abide by the terms of this Agreement. Nothing in this License provides any rights beyond those provided by 17 U.S.C. §117.
- Use Restrictions: Purchaser may physically transfer the Product from one location to another provided that the Software is not copied. Purchaser may not electronically transfer the Software from the Product to any other instrument. Purchaser may not disclose, publish, translate, release, distribute copies of, modify, adapt, translate, reverse engineer, decompile, disassemble, or create derivative works based on the Software or the written materials.
- Transfer Restrictions: In no event may Purchaser transfer, assign, rent, lease, sell, or otherwise dispose of the Product or the Software on a temporary basis. Purchaser shall not assign or transfer this License, in whole or in part, by operation of law or otherwise without Masimo’s prior written consent; except that the Software and all of Purchaser’s rights hereunder shall transfer automatically to any party that legally acquires title to the Product with which this Software is included. Any attempt to assign any rights, duties or obligations arising hereunder other than as set forth in this paragraph shall be void.
- U.S. Government Rights: If Purchaser is acquiring Software (including the related documentation) on behalf of any part of the United State Government, the following provisions apply: the Software and documentation are deemed to be “commercial software” and “commercial computer software documentation,” respectively pursuant to DFAR Section 227.7202 FAR 12.212, as applicable. Any use, modification, reproduction, release, performance, display or disclosure of the Software (including the related documentation) by the U.S. Government or any of its agencies shall be governed solely by the terms of this Agreement and shall be prohibited except to the extent expressly permitted by the terms of this Agreement.