MASiMO Bridge Percutaneous Nerve Field Stimulator User Guide

June 5, 2024
MASiMO

MASiMO Bridge Percutaneous Nerve Field Stimulator

THE KIT CONTAINS

MASiMO-Bridge-Percutaneous-Nerve Field-Stimulator-
FIG-1

  • Bridge Stimulator with inserted but not activated batteries
  • Sterile Wire Harness for use with Bridge Stimulator
  • (8) Round Bandages to fasten the appliance leads
  • (2) Stimulator Adhesive
  • Compound Benzoin Tincture
  • (2) Alcohol Swab
  • Sureprep® Protective Wipe
  • Tegaderm™
  • Bio-Hazard Bag
  • Patient Information Card
  • Tweezers
  • Surgical Marker
  • Medical Light
  • Quick Reference Guide
  • Directions For Use

PUTTING ON BRIDGE

  1. Ask the patient which side they generally sleep on and if they have a preference of which ear they would like the device applied to.
  2. Discuss if the patient wears glasses, hearing aids or uses a CPAP machine, to ensure Bridge is placed around them appropriately.
  3. If the patient has long hair, pin the hair up and away from the ear.MASiMO-Bridge-Percutaneous-Nerve Field-Stimulator-FIG-3
  4. Use the alcohol pad to clean off the anterior (ventral) and posterior (dorsal) of the outer ear to remove all surface oils, and make up and disinfect the skin.
  5. Additionally use the Sureprep wipe on zones of ear where leads will be implanted.
  6. The correct location for implantation of the electrode needles is determined by visualizing the cranial and occipital neurovascular bundles through transillumination.
  7. To mark zones 1-4 for lead placement:
  8. Place the medical light tip against the skin on the opposite side of the zone identified and view the zone as light passes through the tissue. The neurovascular bundles will be visibly outlined.
  9. Mark a location within 0.5-2mm of a neurovascular bundle with provided surgical marker.

NOTE: to avoid bleeding and discomfort, do not place zone location on the neurovascular bundle.MASiMO-Bridge-Percutaneous-Nerve Field-Stimulator-
FIG-4

Zone 1 Great Auricular Nerve

  • Center of the ear lobe.

Zone 2 Auricular Branch of the Vagus Nerve

  • Anterior and superior to the tragus (lay the tweezers across the center of the ear horizontally to assist with marking the implantation site).

NOTE: Check for a pulse in this area and mark zone AWAY from the pulse.

Zone 3 Auriculotemporal Nerve

  • At the triangular fossa

Zone 4 Lesser Occipital Nerve

  • Dorsal side of the triangular fossa.
  1. Place the double-sided stimulator adhesive on the side of the stimulator contacting the patient’s ear.
  2. Insert the wire harness into the stimulator until firmly seated to start therapy.
  3. Hold the stimulator with the inserted wire harness posterior to the patient’s ear. Ensure that the stimulator is properly positioned extending the wires over the ear allowing enough length so the leads can be placed at the desired location.
  4. After the proper position of the device is confirmed apply the compound benzoin tincture to the skin where the stimulator is to be applied. Also, apply the tincture at the sites marked for lead placement.
  5. Gently press the stimulator with inserted wire harness onto the skin at the desired application site for 3-5 seconds.
  6. Start with the placement of the ground lead (white color) on zone 1. Secure the electrode needle and place a round adhesive bandage on the flat side of the lead. Implant the lead percutaneously with gentle pressure.

NOTE: Round adhesive can be applied to the electrode lead either before or after needle implantation.

  1. Repeat step 13 for the remaining zones with any of the grey leads.
  2. If bleeding and discomfort occur, reevaluate placement location with transillumination and reposition the leads.
  3. Wait 5-10 mins to ensure that patient is able to feel the stimulation.
  4. If the patient cannot feel or tolerate the stimulation after 5-10 min, , reposition the leads.MASiMO-Bridge-Percutaneous-Nerve Field-Stimulator-FIG-9
  5.  Gather any dangling wires and hold it close to the stimulator. Place the Tegaderm™ strip over the stimulator to hold it in place.
  6. If withdrawal symptoms begin to return within 5 days of therapy starting, direct the patient to perform a pinch test to make sure the leads are in good contact with the ear. For the pinch test, simply pinch down on the lead.

REMOVAL OF BRIDGE

Have the patient return to the office after five days for removal.

  • Gently remove each of the electrodes one at a time by slowly lifting off the round adhesive patch, then pulling the electrode needle out.
  • After the electrodes have been removed, carefully pull off the stimulator from behind the ear.
  • Once removal is complete, clean the areas of the ear where the electrodes were placed with an alcohol wipe.
  • Place the used device into the biohazard waste bag provided and dispose of the bag as medical waste, following your appropriate state regulations.