RCP ADF-B06 Pulse Oximeter User Guide
- June 6, 2024
- RCP
Table of Contents
- RCP ADF-B06 Pulse Oximeter
- INTRODUCTION
- NOTES ON SAFETY
- PRODUCT OVERVIEW
- SYMBOLS
- POWER-ON KEY/FUNCTIONAL KEY PERATIONS
- ALERT SOUND SETTINGS
- BRIGHTNESS SETTING
- POWER SAVING FUNCTION
- ALERT RANGE SETTING
- TAKING A MEASUREMENT
- CLEANING
- DISINFECTING
- MAINTENANCE
- TECHNICAL SPECIFICATION
- STORAGE AND TRANSPORTATION
- CONTACT INFORMATION AND NOTICES
- FCC Compliance Statement
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
RCP ADF-B06 Pulse Oximeter
Model # ADF-B06
INTRODUCTION
This Pulse Oximeter is provided to you by Remote Care Partners (RCP), for use
exclusively with the health monitoring program from your provider.
Please read this manual carefully and keep it handy for future reference.
NOTES ON SAFETY
Please read this section carefully before using the unit.
Do not attempt to open the electrical compartment of the Oximeter.
Periodically change the contact position between the Oximeter probe and the
finger for a measurement that lasts a long time. Adjust the position of the
probe before the measurement lasts two hours, and check the integrity of the
skin, the blood circulation condition of the finger as well as the position of
the finger.
Do not use on newborn babies.
Consult your provider if the result is outside the normal range. A normal
%SpO2 is 95% or greater.A normal PR is between 60-100.
Do not directly expose your eyes to light-emitting components of the Oximeter,
as that could cause harm to your eyes.
For details about clinical limitations and contraindications, please carefully
consult the relevant medical literature.
Warning: Do not use the Oximeter in an environment with any inflammable
gases, inflammable anesthetic, or other inflammable substances.
Warning: Dispose of exhausted batteries in accordance with environment
protection regulations.
Warning: Do not use the Oximeter in an MRI or CT environment.
Warning: Do not operate the Oximeter when it is damp, wet, or has
condensation. Avoid moving the Oximeter from an excessively cold environment
to a high-temperature moist environment.
Handling batteries properly:
As soon as old batteries run out, replace them with new batteries.
- Do not use old and new batteries together.
- Align the polarities of batteries correctly.
- When the unit will not be used for more than 3 months, remove the batteries. Otherwise, batteries may leak and cause damage to the unit.
PRODUCT OVERVIEW
- Symbol of oxygen saturation
- Measured value of oxygen saturation
- Bar graph
- Plethysmogram
- Measured value of Pulse rate
- PI
- Symbol of pulse rate
- Battery power indication/Bluetooth symbol (alternately displayed)
SYMBOLS
POWER-ON KEY/FUNCTIONAL KEY PERATIONS
After powering on the Oximeter, hold the power-on key/functional key for about one second. The Oximeter shows a parameter setting interface. Press or hold the power-on/functional key to perform corresponding operations. Hold it to set an item or press it to switch an option or switch the display mode. Press means no more than 0.5 seconds, while Hold means more than 0.5 seconds.
ALERT SOUND SETTINGS
Hold the power-on key/functional key while the Oximeter is in powered-on
state.
Parameter setting interface 1 is displayed.
Move “” to the corresponding option by pressing the function key. Hold the
function key to set Alm to on and set Beep to off.
When Alm is set to on and the measured values of the blood oxygen saturation
and pulse rate go beyond the upper limit or lower limit, the Oximeter gives
off an alert sound.
When Alm is set to off and the measured values go beyond the limit, the
Oximeter will not give any alert sound.
When Beep is set to on, a tick will be heard along with pulse beats during
pulse rate measurement.
When Beep is set to off, no sound will be output along with pulse beats during
pulse rate measurement.
While the “” symbol stays on the Restore option, hold the function key to
restore factory settings.
BRIGHTNESS SETTING
On parameter interface 1, press the functional key to select the Brightness option and then hold the functional key to set the brightness to a value ranging from 1 to 5. The greater the value, the greater the brightness of the screen.
POWER SAVING FUNCTION
The ADF-B06 has a power-saving feature. It can save power by reducing screen brightness automatically. If you leave your finger in the device for more than 150 seconds, the brightness will reduce to level 1. To make it return to the original brightness level, press the function key.
ALERT RANGE SETTING
On parameter interface 2, press the functional key to switch between options.
On this interface, you can set the upper limit and lower limit of SpO2 Alm and
PR Alm.
While the “” symbol stays on the +/- option, hold the function key to set the
option to + or -.
In + mode, select the corresponding option and hold the functional key to
increment the upper or lower limit; in – mode, hold the functional key to
decrement the upper or lower limit.
Move “” to the Exit option and hold the function key to return to the
monitoring interface.
Do not allow other people to use your pulse oximeter. The pulse oximeter is specifically assigned to you and you only. Allowing other people to use your pulse oximeter can mislead your provider.
TAKING A MEASUREMENT
- Squeeze the closed end of the device to open the other end. Stick one finger completely into the rubber pad cavity of the Oximeter,
- Keep the fingernail surface upward and release the closed end so the device fits snugly on your finger.
- Press the power key to power on the Oximeter. If you do not completely insert your finger into the cavity, the measurement may be inaccurate.
- The reading will automatically transfer to the RCP SmartHub and will then be sent to your provider.
Notes:
- Do not move your finger during measurement.
- Keep your body still during the measurement.
- After the readings become stable, read the measured values of oxygen saturation and the pulse rate on the screen.
- The Oximeter will automatically turn off 10 seconds after you remove your finger.
About the display
Continuously press the functional key in the monitoring process. The monitored
data and the display mode will be cyclically displayed on the OLED screen in
two different ways (large fonts and plethysmogram) and four directions, as
shown in the following figure.
Replace the batteries when the battery capacity is insufficient and the symbol flickers on the screen.
CLEANING
Power off the instrument and remove the batteries before cleaning. Ensure that the appearance of the instrument is neat, dust-free, and dirt-free. Clean the outer surface of the instrument (including the OLED screen) using 75% medical alcohol and a piece of dry soft cloth. Caution: Avoid running the Oximeter underwater or submerging it in water or any other liquid during cleaning.
DISINFECTING
Before measurement with the instrument, wipe the rubber finger pad using a piece of dry soft cloth moistened with 75% medical alcohol. Clean the finger to be measured using medical alcohol for disinfection purposes before and after use.
MAINTENANCE
- Remove the batteries from the battery slot and properly store them if you do not plan to use the Oximeter for a long period of time.
- Store the Oximeter between 14 and 122°F (–10 to +50°C) and at humidity levels no greater than 93%.
- Periodically check the Oximeter for damage.
- Avoid using the Oximeter in an environment with inflammable gases or using it in an environment where the temperature or humidity is excessively high or low.
- Check the accuracy of the oxygen saturation and pulse rate readings by using an appropriate calibration apparatus.
TECHNICAL SPECIFICATION
Dimensions
Width| 62.0 mm
Depth| 37.0 mm
Height| 32.0 mm
Weight (including batteries)| 42.5 g
Peak wavelength range of the light emitted from the probe
Red Light| 660 nm ± 3
Infrared Light| 905 nm ± 5
Maximum optical output power of the probe
Infrared Light| 1.2 mW
Bluetooth module
| 4.0 (single mode)
Normal working condition
Working Temperature| 5°C to 40°C (41°F to
104°F)
Relative Humidity| 15% to 80%, non- condensing
Atmospheric Pressure| 70 kPa to 106 kPa
Rated Voltage| DC 3.0 V
Default values and conditions of alert
Parameter| Value
Oxygen saturation| Upper limit: 100
Lower limit: 94
Pulse rate| Upper limit: 130
Lower limit: 50
Alert condition| When the alert switch is on and the actual measured value goes beyond the preset alert parameter range, the Oximeter
gives an alert sound
Technical parameters
Parameter| Value
Display range| O2 saturation| 35% to 100%
Pulse rate| 25 bpm to 250 bpm
Resolution
| Oxygen saturation| 1%
Pulse rate| 1 bpm
Measurement precision| Oxygen saturation| ±2% (70% to 100%)
No requirement (≤ 69%)
Pulse rate| ±2 bpm
Alert range
| Oxygen saturation| Upper limit: 50% to 100%
Lower limit: 50% to 100%
Pulse rate
| Upper limit: 25 bpm to 250 bpm Lower limit: 25 bpm
to 250 bpm
Alert error
| Oxygen saturation| ± 1% of the preset value
Pulse rate| The greater of ±10% of the preset value and ±5 bpm
PI| Weak PI| Min. 0.2%
STORAGE AND TRANSPORTATION
Packaged products should be stored in well-ventilated rooms without corrosive gas and with an ambient temperature of –10 to +50°C, a relative humidity lower than 93% (without condensation), and an atmospheric pressure of 50–106 kPa.
CONTACT INFORMATION AND NOTICES
Remote Care Partners
855-477-7000
www.remotecarepartners.com
AndseFit LTD
Unit 513 Lakeside 1,
No. 8 Science Park West Ave
Hong Kong Science Park
Hong Kong
FCC Compliance Statement
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
- This device may not cause harmful interference, and
- This device must accept any interference received, including interference that may cause undesired operation.