testo 104-IR BT
Combined infrared and penetration thermometer
Instruction manual
General information
Please read this document carefully and familiarize yourself with the operation of the product before putting it to use. Keep this document on hand so that you can refer to it when necessary.
Safety information
Avoid electrical hazards:
- Do not conduct measurements on or near live parts!
Adhere to the product safety/warranty requirements:
- Always operate the instrument properly and according to its intended purpose and within the parameters specified. Do not use force.
- Do not store with solvents (e.g. acetone).
- Only open the instrument if this is expressly described in the documentation for the purpose of maintenance or repair work.
Ensure correct disposal:
- Dispose of defective rechargeable batteries and spent batteries at the collection points provided.
- Send the instrument directly to us at the end of its life cycle.
We will ensure that it is disposed of in an environmentally friendly manner.
Intended use
The testo 104-IR BT is a robust food thermometer.
The product is designed for the following tasks/areas:
-
Food sector: production, food service, spot check
measurement, incoming goods. -
Measuring liquids, pastes and semi-solid materials
The following product components are designed for continuous contact with foodstuffs in accordance with Regulation (EC) 1935/2004:
The immersion/penetration probe from the tip up to 2 cm before the probe handle or the plastic housing. If provided, information about penetration depths in the instruction manual or the mark(s) on the immersion/penetration probe should be observed.
The product should not be used in the following areas:
- Potentially explosive areas
- For diagnostic measurements in the medical sector
Technical data
Bluetooth® Module
The Bluetooth® option may only be operated in countries in which it is type
approved.
Feature | Values |
---|---|
Bluetooth® | Range >20 m (free field) |
Bluetooth® type | LSD Science & Technology Co., Ltd L Series BLE Module (08 Mai |
- based on TI CC254X chip
Qualified Design ID| B016552
Bluetooth® radio class| Class 3
Bluetooth® company| 10274
Certification
Belgium (BE), Bulgaria (BG), Denmark (DK), Germany (DE), Estonia (EE), Finland (FI), France (FR), Greece (GR), Ireland (IE), Italy (IT), Latvia (LV), Lithuania (LT), Luxembourg (LU), Malta (MT), Netherlands (NL), Austria (AT), Poland (PL), Portugal (PT), Romania (RO), Sweden (SE), Slovakia (SK), Slovenia (SI), Spain (ES), Czech Republic (CZ), Hungary (HU), United Kingdom (GB), Republic of Cyprus (CY).
EFTA countries
Norway, Switzerland, Lichtenstein und Iceland.
Other countries
USA
Information from the FCC (Federal Communications Commission)
This device complies with part 15 of the FCC Rules.
Commissioning is subject to the two following conditions: (1) This instrument
must not cause any harmful interference and (2) this instrument must be able
to cope with interference, even if this has undesirable effects on the
operation.
Changes
The FCC demands that the user be informed that any changes or modifications to
the instrument that are not explicitly approved by Testo AG may void the
user’s right to use this instrument.
General Technical data
Feature | Values |
---|---|
Penetration probe | |
Sensor type | NTC |
Measuring range | -50 to +250 °C / -58 to +482 °F |
Resolution | 0.1 °C/°F/°R |
Accuracy (±1 digit) | ±1.0 °C / ±1.8 °F (-50,0 to-30,1 °C /-58.0 to -22.1 °F) |
±0.5 °C / ±0.9 °F (-30,0 to +99,9°C /-22.0 to +211.9 °F)
±1 % of the measuring range (+100.0 to +250.0 °C / +212. to +482.0 °F)
Response time t99| 10 s (measured in moving liquid)
Measuring rate| 0.5 s
Infrared measurement|
Lens| 10:1 + opening diameter of the sensor (12 mm / 0.47″)
Spectral range| 8 to 14pm
Laser type| 2-point laser
Output / wavelength| < 1 rnW / 650nrn
Class / standard| 2 / DIN EN 60825-1:2007
Measuring range| -30 – +250 °C / -22 – +482 °F
Resolution| 0.1 °C/°F/°R
Accuracy (at 23°C, ±1 digit)| ±2.5 °C / ±4.5 °F (-30.0 to -20.1 °C / -22.0 to
-4.2 °F) ±2.0 °C / ±3.6 °F (-20.0 to -0.1 °C / -4.1 to 31.9 °F)
±1.5 °C / ±2.7 °F or ±1.5 % of the measuring value (0.0 to +250.0 °C / +32.0
to +482.0 °F)
Measuring rate| 0.5 s
General|
Operating temperature| -20 to +50 °C / -4 to +122 °F
Transportation/ storage temperature| – 30…+70 °C / -22 to +158 °F (without
batteries)
Power supply| 2 x AAA batteries
Housing| ABS/TPE/PC and die-cast zinc/stainless steel
Protection class| IP65
Dimensions| 281 x 48 x 21 mm / 11.06 x 1.89 x 0.83″ (immersion/ penetration
probe folded out)
178 x 48 x 21 mm / 7.01 x 1.89 x 0.83″ (immersion/ penetration probe folded
up)
Weight| 207 g / 0.4331bs (incl. batteries)
Standards| EN 13485
EC Directive| 2014/30/EC
Warranty| 2 years, warranty terms: see
www.testo.com/warranty
Information on standards
This product complies with the EN 13485 standard for penetration measurement.
Suitability: S, T (storage, transportation)
Environment: E (transportable thermometer)
Accuracy class: 0.5
Measuring range: -50…+250 °C
According to EN 13485, the measuring instrument should be checked and
calibrated regularly under the terms of EN 13486 (recommended frequency:
yearly).
Contact us for more information.
Product description
-
Infrared sensor
-
2-point laser
-
Display
-
Control keys:
· [ON]: switches the instrument on
· [OFF]: switches the instrument off (hold the button down)
· [ ]: switches to IR measurement, carry out IR measurement (hold button down)
· [ ]: switches to contact measurement
· [ /HOLD/MIN/MAX]: Hold reading, show minimum / maximum value, sends readings to the mobile Testo terminal (Bluetooth® ) -
Battery compartment (at the back)
-
Fold-out immersion/penetration probe, folding out the probe switches the instrument on
Commissioning
Inserting batteries
- Use a slotted screwdriver to undo the screw on the battery compartment.
- Open the battery compartment.
- Insert batteries (2x type AAA). Observe the polarity!
- Close the battery compartment.
- Tighten the screw.
Operation
Switching on/off
Switching on via fold-out probe
Fold-out the probe.
- All display segments light up briefly. Contact measurement is enabled ( lights up).
Switching on/off via control keys
Switch the instrument on press [ON].
– All display segments light up briefly. IR measurement is enabled ( lights
up).
Switch off the instrument: press and hold down [OFF] until the display goes
off.
The instrument switches off automatically if no key is pressed: for 10
minutes, or for 1 minute when the probe is folded up, and Bluetooth® Mode is
disabled.
Changing the measuring mode
Contact measurement —> IR measurement: press [ ].
IR measurement —> contact measurement: press [ ].
Measuring
Observe the information on IR measurement/contact measurement (see chapter
below).
Press [ ] to transfer the displayed reading in Bluetooth® mode to the mobile
Testo terminal.
IR measurement
-
– Instrument is switched on, IR measurement is enabled, Bluetooth® mode is enabled.
-
Start measurement: press and hold down [ ].
-
Lock in on measurement object using the laser points: laser points mark the edges of the measuring range.
-
– The current measuring value is displayed.
-
End measurement: release the key.
-
– Hold lights up. The last measuring value and min./max. value are saved until the next measurement, or until the instrument is switched off.
-
Switch between min., max. and recorded value: press [ ].
-
The min./max. values can be reset:
· press [ ] or switch the instrument off. -
Restart measurement: press and hold down [ ].
-
Setting the emission level:
• When IR measurement is enabled, press and hold down[ ] and [ ] at the same time ( lights up).
• Emission level is displayed.
• Use [ ] or [ ] to change the value and wait for 3 s.
Contact measurement
- – Instrument is switched on, contact measurement is enabled ( lights up), Bluetooth® mode is enabled.
- 1 Position the contact thermometer in the measurement object and initiate the measurement: press [ ].
- 2 End measurement: press [ ].
- – Hold lights up. The last measuring value and min./max. values are saved until the next measurement, or until the instrument is switched off.
- AutoHold function: if this function is enabled, the measurement is ended automatically as soon as the measuring value is stable, AutoHold lights up.
- Switch between min., max. and recorded value: press [ ].
- The min./max. values can be reset:
- · switch the instrument off, switch to IR measurement or, while the held measuring value is displayed (Hold lights up), press and hold down [ ] until Clr lights up.
- Restart measurement: press [ ].
Settings
-
– The instrument must be switched off to initiate settings mode.
-
If no button is pressed for 3 s in settings mode, the instrument switches to the next view.
-
1 With the instrument switched off, press and hold down [ ] and [ ] until AutoHold or Hold flashes.
-
2 Switch the AutoHold function on (AutoHold) or off (Hold): press [ ] or [ ].
– °C, °F or °R flashes. -
3 Set measurement unit to degrees Celsius (°C), degrees Fahrenheit (°F) or degrees Réaumur (°R): press [ ] or [ ].
– (on) or (oFF) flashes. -
4 Switch the laser on (on) or off (oFF): press [ ] or [ ].
-
5 Switch Bluetooth® on (on) or off (oFF): press [ ] or [ ].
-
– The instrument switches to IR measurement.
-
– Bluetooth® is enabled and can be detected by mobile Testo terminals with a Bluetooth® interface. When the connection is established, a beep is emitted and the Bluetooth® icon is shown in the display.
Service and maintenance
Changing the batteries
- Use a slotted screwdriver to undo the screw on the battery compartment.
- Open the battery compartment.
- Insert batteries (2x type AAA). Observe the polarity!
- Close the battery compartment.
- Tighten the screw.
Cleaning the instrument
Only use weak, commercially available neutral/household cleaning agents (e.g. washing-up liquid) to clean the instrument. Do not use aggressive cleaning agents or solvent! The housing and probe can be disinfected using an alcohol- based spray. In doing so, always follow the manufacturer’s instructions.
- Clean the housing and probe under running water and rub dry with a towel.
- Clean the lens carefully with a cotton swab dipped in water or medical alcohol.
Questions and answers
Question | Possible causes | Possible solution |
---|---|---|
lights up | Low battery | ►Change batteries. |
IR measurement:
– – – lights up.| Measurement values
outside measuring range| ►Keep to permissible measuring range.
Contact measurement: – – – lights.| Measurement values outside the permissible
range| ►Keep to measuring range.
The instrument cannot be switched on| Batteries dead.| ►Change batteries.
The instrument switches itself off.| Instrument switches off automatically
after 10 min in contact measurement mode and after 1 min after switching on in
IR measurement mode.| ►Switch the instrument on again
If we have not been able to answer your question, please contact your local dealer or Testo Customer Service. For contact details, please visit www.testo.com/service-contact.
Information on infrared (IR) measurement
Measuring method
IR measurement is a visual measurement
- Keep lens clean.
- Do not carry out measurement with a foggy lens.
- Keep the measuring range (the range between the instrument and the measurement object) free of obstacles. There must be no particles of dust or dirt, no humidity (rain, steam) and no gases.
IR measurement is a surface measurement
If there is dirt, dust, frost, etc. on the surface, only the outermost layer
is measured, i.e. the dirt.
For vacuum-packed food, do not measure at air pockets.
Where the values are critical, always measure separately with a contact
thermometer. Particularly in the food sector: measure core temperature with a
penetration/immersion thermometer.
Adjustment time
If the ambient temperature changes (change of location, e.g. measurement
indoors/outdoors), the instrument must be allowed to equalize for 15 minutes
for infrared measurement.
Emissivity
Materials have different emission levels. This means they emit various amounts of electromagnetic radiation. The emission level of the instrument has a default setting of 0.95. This is ideal for the measurement of food, non- metals (paper, ceramic, gypsum, wood, paints and varnishes) and plastics.
Measuring range, distance
Depending on the distance of the measuring instrument from the measurement
object, a specific measuring range is recorded.
Measuring lens (ratio of distance: measuring range)
In italics = laser
Not in italics = measuring range
- Optimized measuring distance
Information on contact measurement
Observe the minimum penetration depth for immersion/penetration probes:10x
probe diameter
Avoid using in aggressive acids or alkalis.
Bluetooth® pairing
This chapter gives an overview of the public Bluetooth® Low Energy (BLE) interface of the test 104-IR BT and shows how to test functionalities and readout values with a generic BLE scanner app. In addition, a list of available services and characteristics with information about the transferred data is included.
For testing the Bluetooth® GATT profile a Bluetooth® scanner app from the Google Play Store can be used, for example, the BLE Scanner app from Bluepixel Technologies LLP Tools (https://play.google.com/store/apps/details?id=com.macdom.ble.blescanner). This app scans the services and characteristics from the BLE device.
Scan Devices and Connect
After starting the BLE scanner, all devices in range are detected. The testo 104-IR BT is labeled with the device name and the respective serial number. For connecting the device, the “connect” button must be pressed.
Services
After connecting, all services available are displayed. To get an overview of
the characteristics and their properties and values, click on the service to
expand the list.
Read / Notify Characteristics
Device Information
To read the characteristic’s value, press the blue “R” button, to load the value.
Custom Service
Each characteristic Descriptor (1) (UUID: 0x2901) provides a string to describe what kind of data will be transferred on this characteristic. With the read button (2) the characteristic value will be read out and shown. To activate the notification, click on the “N” button (3). If notifications are enabled the button shows up in green.
Write Emissivity
To write a new emissivity value, press the “W” button on the characteristic, after that a new window pops up. Clicking “OK” sends the value to the device. Emissivity is valid in a range of 0.1 – 1.0 with either one or two after- point decimals.
BLE Interface Description
Service| Characteristic| Properties| j Description|
Range
---|---|---|---|---
Device Information (0x180A)
| Model Number String (00002a24-0000-1000-8000-00805f9b34fb)| read| String
(104IR BT)|
| Serial Number String (00002a25-0000-1000-8000-00805f9b34fb)| rear| String
(xxxxxxxx)|
| Firmware Revision String (00002a26-0000-1000-8000-00805f9b34fb)| read|
String (xxx.xxx.xxxx)|
Generic Access (0x1800)
| Device Name (00002a00-0000-1000-8000-00805f9b34fb)| read| I String (1041R
BT SN:xxxxxxxx)|
Custom Service (00005a90-3354-11e8-b467-0ed5f89f718b)
| Battery Level (00005a91-3354-11e8-b467-0ed5f89f718b)| read/ notify| String
(Battery Level + Unit)| 0% — 100%
| Surface Temperature (00005a92-3354-11e8-b467-0ed5f89f718b)| notify| String
(Temperature + Unit)| -22° to 482 *F / -30 to +250 °C
„ Overrange”, “Underrange”, “Invalid”
| Plunge Temperature (00005a93-3354-11e8-b467-0ed5f89f718b)| notify| String
(Temperature + Unit)| -58° to 482 °F / -50 to +250 °C
„ Overrange”, “Underrange”, “Invalid”
| Temperature Unit (00005a94-3354-11e8-b467-0ed5f89f718b)| read/ notify| Char
(Temperature Unit)| ar f “C”, “R”
| Emissivity| (00005a95-3354-11e8-b467-0ed5f89f718b)| | read/ write| String|
(Emissivity)| 0.1 ^a 1.0
For each characteristic in the custom service (00005a903354-11e8-b467-0ed5f89f718b), there is a descriptor (UUID: 0x2901) which provides a string with information about the data in this characteristic.
Declaration of Conformity
Declaration No. 0005 / 2015
Wir messen es:
EC declaration of conformity
We confirm that the following products:
testo 104-IR BT Best. Nr.: / Order No.: 0560 1045
corresponds with the main protection requirements and, if used according to
their intended purpose, comply with the essential requirements of the
directive:
directive:
R &TTE 199/5/EG (bis/until 13.06.2016)
RED 2014/53/EU (ab/from 14.06.2016)
standards
EN 61326-1:2013
EN 301 489-1 V1.9.2: 2011
EN 301 489-17 V2.2.1: 2012
EN 300 328 V1.8.1: 2012
LE EN 62479:2010
EN 60950-1:2006+A11:2009+A1:2010+Al2:2011+A2:2013
This declaration is given in responsibility for.
Testo AG
P.O. Box 1140
79849 Lenzkirch / Germany
www.testo.com
abgegeben durch I buy.
Dr. Roif Merte
(Name/ name)| Wolfgang Schworer
(Name/ name)
---|---
Head of Research & Development
(steuvngim Betrieb des Herstelters) (Position in the company of the
manufacturer)| Head of Firmware & Electronics
(Stellung im Betrieb des Hersteliers)
(Position in the company of the manufacturer
Lenzkirch 30.1 1.2015
ppa
( Legally valid nature)
| iv
( Legally signature )
testo Solutions USA, Inc.
2 West Market Street, Suite 500
West Chester, PA 19382
Telephone: 800-227-0729 ext. 200
E-Mail: solutions@testo.com
Internet: www.testo.com/solutions
www.testo.com
0970 1043 en 01