Watchdata WD-C7808 Facial Recognition Thermometer User Manual

June 6, 2024
Watchdata

Watchdata WD-C7808 Facial Recognition Thermometer User Manual

v 1.2

Product Application Mode

Stranger Temperature Measurement – Standalone Version

Once power on, WD-C7808 Facial Recognition Thermometer (referred to as terminal) will automatically enters into stranger temperature measurement application mode. When someone walks into the detection zone, the terminal will automatically capture the human face, the temperature will be displayed on the screen and broadcasted by audio. When the temperature is above 37.3℃, there will be automatic light and audio alarm.

Applicable scenarios include public places with low flow of people, such as tax bureaus, industrial and commercial bureaus and other government service halls, parks, banks, library, exhibition hall, etc.

Fixed Personnel Temperature Measurement – Standalone Version

In this application mode, the whitelist of personnel should be loaded into terminal in advance. The terminal takes photos while measuring the temperature. The photos are compared with those in the whitelist to determine whether they match, along with the audio broadcast. When the temperature is above 37.3℃, there will be automatic light and audio alarm.

Whitelist import and history record export are supported by USB interface.

Online Temperature Measurement

Online temperature measurement application mode requires terminal goes online. When working at this application mode, the terminal will not only export comparison record of whitelist, but also do real-time uploading of stranger recognition records to server.

Online temperature measurement application mode requires a separate private deployment of the management platform developed by Watchdata.

Operating Instructions

Power On

Insert charger in charging port at back side of terminal, the temperature measurement software will be started automatically. It takes about 10 seconds for the boot process, please wait patiently.

Power On

After power on of terminal, calibration with ambient temperature is required. It is recommended to wait 10 minutes after power on before use terminal for temperature measurement.

Main Interface

After power on of terminal, the main interface is the facial recognition and temperature measurement interface. There are 3 ICONS in the lower left corner of the display area, with the following descriptions with their representations and functions.

Main Interface

Sound control – click this icon the volume control bar will pop up, which can be used to adjust system volume.

Server connection status – if the dot in the icon is red, it indicates that server is not connected; otherwise, if the dot in the icon is green, it indicates that server is connected.

Organization code modification – click to modify organization code, organization code equals to company brand.

Terminal-related Settings

For fixed personnel temperature measurement mode and online temperature measurement mode, terminal carries out the personnel list import and identify record export.

Enter System Management Interface

In the main interface of terminal, click the button in the upper right corner of the main interface for 5 times in a row. There shows a window to enter password.

Enter System Management Interface

Enter password “6116” (default password) in the input box, click “OK” button, it should enter the system management interface, which is displayed as below.

Click the icon in the upper right corner to quickly enter the android settings interface for related operations.

Enter System Management Interface Continued

Server Address Setting

After entering system management interface, select the “communication settings” as shown in the following figure.

Server Address Setting

URL address – management platform address, the default is Watchdata management platform address: 101.201.181.195:12080

Reg Code – the organization code is used to identify the company to which the device belongs. The default value is wd-test.

Device serial number – device serial number cannot be changed by the user.

Note: the Reg Code here is consistent with the Organization code modification in the main interface.

Device Name Setting

After entering system management interface, select the “basic parameter settings” as shown in the following figure.

Device Name Setting

Device name – user-defined device name, if user need to have multiple devices connect to the server, device name is used for distinction.

Device gate – the default value is “Entrance”, it can be set according to the actual installation environment as Entrance, Exit or Entrance /Exit.

Installation address – equipment installation address defined by user.

Temperature Compensate Setting

Terminal has function of temperature compensation by certain manual calibration using a more accurate temperature measuring device, such as a mercury thermometer. After entering system management interface, select the “temperature settings” as shown in the following figure.

Temperature Compensate Setting

Calibrate the temperature value by sliding up and down with finger for temperature compensation.

List of People

After entering system management interface, select the “List of people” to view the list of people in the terminal, including white list, guest and blacklist.

Historical Record Review

After entering system management interface, select the “History” to view the history recognition data in the terminal, which is shown as follows.

The history comparison record is displayed on the interface. Double-click to view the details of the record.

Note: only the identification records of people on the whitelist are recorded. The identification records of strangers are not kept.

Open/Close navigation bar

After entering system management interface, select the “Basic parameter settings” interface, then “Swipe up and down to display the navigation bar” to choose whether enable this function.

Navigation bar

Networking Settings

Swipe up on the lower side of the screen to display the navigation bar and enter the android desktop.

Use android desktop Settings button to go to WIFI Settings or Ethernet Settings. This setting is the same as for a universal android device.

Terminal Mounting Methods – Desktop

Terminal of WD-C7808 has three ways of placing, providing a great convenience for customer of variety scenarios.

Figure 3.1

With additional base in the package, terminal of WD-C7808 can be placed on table, desk and similar positions. The additional base is suitable to place a contactless reader inside.

The following 3 steps show how to install terminal with desktop method.

Step 1: Drop the main engine on the base and move it back from the front
Step 1

Step 2: Put the host on the surface of the base and push it until host and base completely stuck, and cannot move back, as shown in the figure.
Step 2

Step 3: As the hexagon screws at the bottom of the host can be seen from the through hole on the back of the base, use the hexagon tool with the equipment to screw out the screws 3 to 4 times counterclockwise to fix the base and the host in front and back positions
Step 3