testo 106 Digital Food Waterproof Thermometer Instruction Manual
- June 6, 2024
- testo
Table of Contents
testo 106
Food Thermometer
General Information
Instruction Manual
Please read this document carefully and familiarise yourself with the
operation of the product before putting it to use.
Keep this documentation to hand so that you can refer to it when necessary.
Product Description
Safety Information
Avoid electrical hazards:
Do not measure on or near live parts!
Preserving product safety/warranty claims:
Operate the instrument properly and according to its intended purpose and within the parameters specified. Do not use force.
Do not store together with solvents (e.g. acetone).
Open the instrument only when this is expressly described in the documentation for maintenance purposes.
Ensure correct disposal:
Dispose of defective rechargeable batteries and spent batteries at the collection points provided.
Send the instrument directly to us at the end of its life cycle. We will ensure that it is disposed of in an environmentally friendly manner.
Protect the environment:
The button cell used in the instrument contains 1,2Dimethoxyethane (CAS
110714). See EC Regulation No. 1907/2006 (REACH) Art. 33.
Intended Use
Test 106 is a robust food thermometer.
The product is designed for the following tasks/areas:
· Food sector: Production, food service, spot check measurement
· Measurement of liquids, pastes and semisolid materials
The following components of the product are designed for continuous contact
with foodstuffs in accordance with the regulation
(EG) 1935/2004:
The measurement probe is up to 1 cm before the probe handle or the plastic
housing. If provided, the information about penetration depths in the
instruction manual or the mark(s) on the measurement probes should be noted.
The product may not be used in the following areas:
· In areas where there is a danger of explosion
· For diagnostic measurement in medicine
Technical Data
Feature | Values |
---|---|
Measurement range | -50 to +275 °C |
Resolution | 0.1 °C |
Accuracy | ±0.5 °C (-30.0 to +100.0°C) |
±1.0 °C (-50.0 to -30.1°C)
±1 % (+100.1 to +275.0°C)
Response time t99| 10s (Measured in moving liquid)
Measuring rate| 2 measurements per second
Working temperature| -20 to +50°C
Transport/ Storage temperature| -40 to +70°C
Power supply| 2 x round cell type CR2032
Battery life| 350h (typically)
Housing| ABS
Protection class| IP67 (with TopSafe)
Dimensions| 220 x 35 x 20 mm
EC guideline| 2014/30/EC
Information on standards
This product fulfills with TopSafe guidelines in accordance with the EN 13485
standard.
Suitability: S, T (storage, transport)
Environment: E (transportable thermometer)
Accuracy class: 0.5
Measurement range: 50 to +275 °C
According to EN 13485, the measuring instruments should be checked and
calibrated regularly under the terms of EN 13486 (Recommended: Yearly).
Contact us for more information.
Initial Operation
- Open the battery compartment.
- Insert battery (2 x Type CR2032). Watch out for +/.
- Close battery compartment.
Operation
If the signal sound is switched on, there will be a sound every time is
pressed and alarm limits are exceeded.
If the alarm LED is switched on, it flashes when upper or lower alarm values
are exceeded.
If Auto-Off is switched on, the instrument switches off automatically after 10
minutes if no button has been activated.
If Auto Hold is switched on, AUTO HOLD flashes. Once the reading is stable
within a fixed time duration, it will be frozen.
AUTO HOLD lights up.
Restart measurement: .
7.1 Switching On/Off
Switch on an instrument: .
Switch off instrument: Keep pressed.
7.2 Measuring
Observe required immersion/penetration depth for correct measurement results:
15 mm.
Instrument is switched on.
Immerse / penetrate probe into the object being measured.
The current reading is shown.
Holding reading manually
Hold reading: .
- Reading is frozen and HOLD lights up.
Restart measurement: .
Exceeding the alarm value
- ALARM and (upper alarm value exceeded) or (lower alarm value exceeded) light up.
7.3 Setting alarm limits
Instrument is switched on.
-
Show set upper alarm value ( ): .
or
Show set lower alarm value ( ): . -
Set alarm value: or . Keep button pressed to increase the set speed.
-
Return to measurement: .
If necessary repeat the procedure for the upper and lower alarm value.
Setting instrument
The instrument is switched off.
-
Open configuration mode: Keep pressing.
-
Select temperature unit (°C or °F): .
Confirm selection: . -
Switch Auto Hold on or off : .
Confirm selection: .
If Auto Hold has been switched on:Set time duration (5, 10, 15, or 20s) at which a reading has to be stable (change lower than 0.2°C) for it to be frozen: . Confirm selection: .
-
Switch sound on or off : .
Confirm selection: . -
Switch Alarm LED on or off : .
Confirm selection: . -
Switch Auto-Off on or off : .
Confirm selection: .
– Configuration is complete. The instrument changes to the measurement mode.
Service and Maintenance
9.1 Changing the battery
- Open the battery compartment.
- Insert battery (2 x Type CR2032). Watch out for +/.
- Close battery compartment.
9.2 Cleaning the instrument
Do not use any aggressive cleaning agents or solutions.
Clean the housing and probe with a damp cloth (soap water).
Questions and Answers
Question
| Possible causes|
Possible solution
---|---|---
lights up.| Batteries empty| Change batteries.
– – – lights up.| Measurement range exceeded.| Measurements can only be
carried out in the range specified.
Instrument cannot be switched on.| Batteries empty.| Change batteries.
The instrument switches itself off.
| The auto-Off function is switched on.| Switch off Auto-Off function (See Switching on the instrument).
If we have not answered your question, please contact your local dealer or
Testo´s Customer Service.
For contact details, please visit
www.testo.com/servicecontact
Accessories
Name | Item no. |
---|---|
TopSafe | 0516 8265 |
Frozen food drill | 0554 0826 |
Wall bracket and protecting cap | 0554 0825 |