Please fill in water full before turning on the machine:( Distill water/normal clean water. 3L) exchange water once in 2 weeks. connect the handle, push straight tighten and lock the side tips.

Steps of filling in water

open “over flow”. insert water funnel to “injection”. fill in approximately 3L water. when “Overflow” comes out water.it is full. Press the injection blue hole frame to take out the water funnel.

Turning on machine: press the green button and it will light on.

Machine Installation

**Left and Right:

**

Front and back:
Machine Installation

General Structure

Cry lipolysis + 360°Vacuum cup body and Double chin
General Structure

Big size vacuum cup with silicone frame. for abdomen. Selection of different size handle for abdomen, waist, leg, butt, arm and chin.

Key Technology

Minus -15℃ freezing vacuum sculptor/Vacuum suction fat in cup

Main Function

  1. Body shaping, fat dissolving, repel the cellulite.

  2. Lymphatic drainage, promoting metabolism & blood circulation.

  3. Decrease the excessive water and prevent the fat from
    accumulation.

  4. Fastest effect fat removal.

  5. Complementing the collagen, skin rejuvenation.

User Interface
User Interface

Select language
Russian, Spanish, English, French, Italian. Click to enter in Software.

  • The main menu interface for Cry lipolysis + vacuum and hand piece control display
       control display
    360°body handle screen Double chin handle screen
    control display  control
display
    control display

  • Cry lipolysis Treatment settings

    1. -5 degree freeze treatment time 20-30 minutes for each part,-10 degree treatment time 15-20 mins.-15 degree no more than 15 mins.to avoid frostbite. antifreeze membrane disposable for each client. Setting for vacuum 3S. release 1S.Machine body host screen 1 snow is -3 degree. Press handle screen “work” to start vacuum and handle.

    2. Multi-sizes handles for different area treatment. each area treatment time 20-30 mins. move the handle step by step.

    3. Take a another cooling procedure at the same position. you needs to wait at least 10 days.
      Wait for first procedure fat remove out of body and skin recover from Metabolism.

    4. Cry lipolysis cool-sculpturing treatment is designed for those physically healthy people who have local fat deposits. It can target at the fat at a specific part, such as waist, abdomen, back, etc.

Points for attention

  1. Read the user manual and machine body yellow stickers first before operation.
  2. Use antifreeze pads on skin for treatment.
  3. The patient cannot receive the treatment if he/she has following medical electronic equipment:
    • Implanted cardiac pacemakers or other implanted electronic devices;
    • Life-support artificial heart-lung machine or other equipment;
    • Portable ECG measurement device;

Contraindication

Please avoid the treatment from these people :

  1. Pregnant women or women who may be pregnant. 2.Scarring, infection of the treatment area
  2. History of keloid scarring
  3. Use of medication known to induce photosensitivity
  4. Fragile skin
  5. Patients undergoing the surgical procedure without consulting with a surgeon
  6. Patients with tumors
  7. Patients with serious disease
  8. Patients with low/high blood pressure

Warranty

  1. Manufacturer warrants defects in materials and workmanship for a period of 1 year from the date of original purchase.
  2. Manufacturer won’t bear any responsibility:
    • Damage caused by accident, abuse, misuse or misapplication, it will be repaired with repair fee.
    • Damage caused by service performed by anyone who is not technician or qualified by manufacturer.
    • A product or part that has been modified without any permission of manufacturer.
    • machine is been opened without notice to manufacturer. disposable screw sticker is gone.

Technical Parameters

  1. Host display: 8 “TFT color touch screen
  2. Voltage : 220V/50Hz or 110V/50Hz
  3. Power:800W
  4. Cooling liquid: distill water/clean water
  5. Vacuum: 650mmHg
  6. Pump flow rate: 60L/min
  7. Vacuum pressure: 0-86Kpa
  8. Freezing temperature: -1℃ to -15℃
  9. (3 kinds of handles for selection)
    360 ° big handle with 4 cups size
    208.5cm
    21
    9cm
    2210cm
    23
    1.5cm
    360 ° medium handle with 4 cups size
    136cm
    14
    7cm
    168cm
    17
    9cm
    *small chin handle 84cm
    small chin handle**

Packing list:

  • Machine Body with power wire 1 pc
  • Handle support 1 pc
  • Cooling handle 1 pcs/2 pcs/3pcs(selective)
  • Water funnel 1 pc
  • Vacuum filter 1 pc
  • User brochure 1 pc