EL-WiFi-21CFR-DULT
21CFR Dual Channel Ultra-Low-Temperature Cryogenic Vaccine WiFi Data Logger
- Can be installed as part of a 21CFR compliant system with additional controls including permission-based use, authority level sign-off, and full system audit
- Measurement range -100 to +100°C (-148 to +212°F)
- T-type probe on 1m cable
- Wirelessly stream and view data on the 21CFR Cloud
- Configurable high and low alarms with indicator
- Easy sensor set-up using free 21CFR WiFi Sensor Software
The EL-WiFi-21CFR-DULT is designed to monitor vaccines in cryogenic dry ice storage, in accordance with 21CFR part 11 data security standards. Data is streamed wirelessly over any compatible WiFi network to be viewed on the 21CFR Cloud. During configuration, the sensor will search for an existing wireless network whilst physically connected to the PC. It can then be placed anywhere within range of the network. If the sensor temporarily loses connectivity with the network, it will log readings until it is able to communicate again with the 21CFR Cloud (max 30 days at 10-second sample interval). The sensor is IEEE 802.11bgn* (2.4GHz) compliant, supports WEP, WPA/WPA2 encryption, and enterprise networks (PEAP, TTLS, FAST). The EL-WiFi-21CFR-DULT has a protection rating of IP40 and the probe tip IP67. The unit is freestanding, but it can be attached to a wall or surface using the bracket provided. The unit can be clipped in and out of the bracket as required.
SPECIFICATIONS
| Minimum| Typical| Maximum| Unit
---|---|---|---|---
Battery life| | > 6| | Months
Temperature measurement range| -100 (-148)| | +100 (+212)|
°C (°F)
USB supply voltage (@SOOmA)| 4.5| 5.0| 5.5| Vdc
Operating temperature range (logger)| -20 (-4)| | +60 (+140)|
°C (°F)
Logging period (user configurable)| 10 sec| 10 min| 12 hrs|
Transmission period (user configurable)| 1 min| 1 hr| 24 hrs|
Temperature measurement resolution| | 0.1 (0.2)| | °C (°F)
Temperature display resolution| | 0.1| |
Accuracy| | ±2.5 (±4.5)| | °C (°F)
IP Rating| Logger IP40, Probe Tip IP67 (Bung fully inserted, not
permanently powered, thermocouple
connector fitted and fully inserted, a device mounted vertically.)
Dimensions (Display)| 82 x 70 x 36mm (3.22 x 2.75 x 1.41″)
Probe dimensions**| 300 x 0 1.5mm, lm cable (11.8 x 0 0.05, 39.3″ cable)
- MAC Address starting 98:8B: AD:2……. only. **Excluding mounting bracket.
21CFR COMPLIANCE
Permission based use and access|
---|---
Authority based action sign-off|
Data records cannot be edited or deleted|
Complete system audit trail|
For a detailed compliance checklist, please visit
www.lascarelectronics.com/markets/21cfr/|
ACCESSORIES
PSU-5VDC-USB-USA | USB Mains Power Adapter for the USA |
---|---|
PSU USB-UK | USB Mains Power Adapter for the UK |
PSU USB-EU | USB Mains Power Adapter for the UK |
EL-P-TC-T-ULT | USB Mains Power Adapter for the UK |
INCLUDED IN THE BOX
EL-WIFI WALL BRACKET
|
Wall mounting bracket for EL-WiFi sensors
---|---
2 x EL-P-TC-T-ULT
|
Thermocouple probe
CABLE USB A-MICRO B
|
USB Type-A to Micro B
CALIBRATION CERTIFICATES ARE NOW AVAILABLE
Lascar now offers a Traceable Calibration Certificate Service on Temperature
Data Loggers. Using reference equipment that has been calibrated by a
UKAS/NIST accredited laboratory and using apparatus traceable to national or
international standards. For more information, please visit
www.lascarelectronics.com/calibration.
21CFR WIFI SENSOR SOFTWARE
EasyLog 21CFR WiFi software* is available as a free download from
download.filesthrutheair.com. Easy to install and use, EasyLog 21CFR WiFi
Sensor Software allows easy sensor set-up and connection of sensors to a WiFi
network and the 21CFR Cloud.
21CFR CLOUD
21CFR Cloud subscription delivers all of the flexibility of a remote
monitoring system: Interact with sensors via any internet-enabled device;
manage and monitor sensors;
access event logs and set up email alerts; assign authorization levels to user
accounts and view comprehensive audit trail records. Unlike the standard
EasyLog Cloud service, 21CFR Cloud includes other controls such as permission-
based use, authority sign-off, and full system audit to ensure data monitored
and collected is regulated to 21CFR Part 11 standard. For full pricing
information and to create an account, please visit
https://21cfr.wifisensorcloud.com/. Any queries or questions, please do not
hesitate to contact our team.
A Cloud account subscription is created during the WiFi sensor set-up process using EasyLog 21CFR WiFi Sensor Software. 21CFR WiFi Sensors are only compatible with the 21CFR Cloud 21cfr.wifisensorcloud.com. When setting up your logger, make sure you select the T-Type probe. Download the latest version of the software free of charge from download.filesthrutheair.com
Battery Life and Power Supply
The battery can be recharged (unit must be between 0 – 40 °C) via a PC, a
USB +5V wall adapter, or a portable USB battery pack using the USB lead
provided. It can also be permanently powered by a USB wall adapter or USB
battery pack. Readings may be affected while the internal battery is being
charged. However, once charged, the continued connection of the charger will
have no effect. Battery life is dependent on: transmission period, WiFi
encryption method, WiFi encryption key rotation frequency (determined
by the router/access point), signal strength between router/access point and
WiFi device, the present volume, and type of WiFi traffic from other devices,
sample rate, and operating temperature. Specifications liable to change
without prior warning *Requires Windows 7, 8.1, 10
www.lascarelectronics.com/data-loggers
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>