Synergy Air Elite Low Air Loss Alternating Therapy System
Synergy® Air Elite
Low Airloss Alternating
Therapy System rom Hill-Rom
© 2005 by Hill-Rom Services, Inc. ALL RIGHTS RESERVED.
Manufactured for Hill-Rom by: Caremed Supply, Inc.
Distributed by:
HILL-ROM COMPANY, INC.
4349 CORPORATE ROAD
CHARLESTON, SC 29405
No part of this text shall be reproduced or transmitted in any form or by any
means, electronic or mechanical, including photocopying,
recording, or by any information or retrieval system without written
permission from Hill-Rom Services, Inc. (Hill-Rom).
The information in this manual is confidential and may not be disclosed to
third parties without the prior written consent of Hill-Rom.
First Edition
First Printing 2005
Printed in the USA
CareSelections™ is a trademark of Hill-Rom Services, Inc.
Hill-Rom® is a registered trademark of Hill-Rom Services, Inc.
Synergy® is a registered trademark of Mediq/PRN Life Support Services, Inc.
Underwriters Laboratories, Inc.® is a registered trademark of Underwriters Laboratories, Inc.
The information contained in this manual is subject to change without notice. Hill-Rom makes no commitment to update or keep current, the information contained in this manual.
Hill-Rom reserves the right to make changes without notice in design, specifications, and models. The only warranty Hill-Rom makes is the express written warranty extended on the sale or rental of its products.
To order additional copies of this manual (139122), refer to the back cover for contact information. For countries not listed on the back cover, contact your distributor.
NOTE:
The back cover is a comprehensive list of Technical Support contact
information for Hill-Rom. The product discussed in this manual may not be
available in all of the countries listed.
Document Symbols
This manual contains different typefaces and symbols designed to improve readability and increase understanding of its content:
- Standard text—used for regular data.
- Boldface text—emphasizes a word or phrase.
- NOTE:—sets apart special data or important instruction clarification.
- WARNING or CAUTION
Intended Use
WARNING:
Contraindication: The Synergy® Air Elite Low Airloss Alternating Therapy
System surface is not for use with patients who have an unstabilized spinal
cord injury. Patient injury could occur.
WARNING:
The Synergy® Air Elite Low Airloss Alternating Therapy System surface is not
for pediatric applications. Patient injury could occur.
Features
Controls
- Comfort Control Soft/Firm: Press the Soft or Firm control to adjust the Comfort Control setting shown on the eight increment scale.
- Cycle Timer: Press the Cycle Timer control to adjust the Alternating Therapy cycle time in 5-minute increments between minutes and 95 minutes.
- Alternate: Press the Alternate control to start Alternating Therapy.
- Static: Press the Static control to stop the Alternating therapy and supply low airloss therapy only.
- A uto Firm: Press the Auto Firm control to quickly inflate the air cells and supply uniform firmness to help the caregiver do routine procedures.
- Standby/Operate: Press the Standby/Operate control to start or stop the control unit.
- Power Failure: For additional safety, the Power Failure alarm and indicator come on when the power to the control unit fails or the main Power switch is turned off.
- Alarm Reset: Press to turn the Power Failure alarm off.
Specifications
Safety Tips
To help prevent the risk of hospital bed fires, make sure facility personnel
follow the safety tips in the FDA Public Health Notification:
Safety Tips for Preventing Hospital Bed Fires. (US only)
WARNING:
Evaluate patients for entrapment risk according to facility protocol, and
monitor patients appropriately.
WARNING:
Evaluate patients for entrapment risk according to facility protocol, and
monitor patients appropriately. Ensure that all siderails are fully latched
when in the raised position. Failure to do either of these could result in
serious injury or death.
NOTE:
Siderails are intended to be a reminder to the patient of the unit’s edges,
not a patient-restraining device.When appropriate, Hill-Rom recommends that
medical personnel determine the proper methods necessary to ensure a patient
remains safely in bed.
SHOCK HAZARD:
The potential for electrical shock exists with electrical equipment.
Failure to follow facility protocols may result in death or serious personal
injury.
WARNING:
-
Contraindication: The Synergy® Air Elite Low Airloss Alternating Therapy System surface is not for use with patients who have an unstabilized spinal cord injury. Patient injury could occur.
-
The Synergy® Air Elite Low Airloss Alternating Therapy System surface is not for pediatric applications. Patient injury could occur.
-
Before you plug the control unit into an applicable power source, carefully read the instructions. Failure to do so could cause patient injury, personal injury, or equipment damage.
-
Do not use the unit in the presence of any flammable gases such as anesthetic agents. Patient injury, personal injury, or equipment damage could occur.
-
Keep the control unit and mattress away from flammable liquids.
Failure to do so could cause patient injury, personal injury, or equipment damage. -
Keep the mattress away from sharp objects. Failure to do so could cause patient injury or equipment damage.
-
Do not put any heating device close to the mattress. Patient injury, personal injury, or equipment damage could occur.
-
Use the unit under the order of a physician only. Failure to do so could cause patient injury or equipment damage.
-
When you use the unit, it may be necessary to change the patient’s position from time to time. Failure to do so could cause patient injury or equipment damage.
-
Only Hill-Rom-authorized persons should service the Synergy® Air Elite Low Airloss Alternating Therapy System. Service by unauthorized persons could cause personal injury or equipment damage.
CAUTION:
- Do not drop the control unit. Equipment damage could occur.
- Do not store the Synergy® Air Elite Low Airloss Alternating Therapy System in direct sunlight or extreme cold conditions. Equipment damage could occur.
Installation
-
Remove the existing mattress from the bed frame.
-
Turn the Synergy® Air Elite Low Airloss Alternating Therapy System mattress so its air tubes are at the foot end of the bed.
-
Put the mattress on the bed frame.
-
Use the elastic straps to attach the mattress to the bed frame.
-
Hang the control unit on the footboard of the bed frame.
-
Attach the air tube connectors and exhaust connector to the lefthand side of the control unit.
-
Examine the air hoses. Make sure the air hoses are not kinked under the mattress.
-
Plug the unit into an applicable power source.
-
Turn the Power switch On. The Standby indicator comes on.
-
Press the Standby/Operate control. The Operate indicator comes on, and the control unit starts.
-
Press the Auto Firm control. The Auto Firm indicator comes on, and the mattress starts to inflate.
-
Let the mattress fully inflate.
NOTE: In Auto Firm mode, the mattress should fully inflate within 2 to 3 minutes. -
After the mattress is fully inflated, move the patient on to the mattress in accordance with facility protocol.
-
After the patient is in position on the mattress, press the Auto Firm control to stop the inflation of the mattress.
Instructions for Use
CPR Mode
To quickly deflate the mattress for CPR procedures, pull the CPR release on
the side of the control unit to disconnect the hose connector.
Alternating Therapy Mode
Alternating Therapy mode inflates and deflates the 18 air cells in this order:
- Every third air cell fully inflates
- Air cells 2, 5, 8, 11, 14, 17 deflate
- Every next third air cell fully inflates
- Air cells 3, 6, 9, 12, 15, 18 deflate
- Every final third air cell fully inflates
To start the Alternating Therapy mode, do as follows:
- Press the Alternate control. The Alternate indicator comes on.
- To set the necessary cycle time interval of the therapy, press the Cycle Timer control to adjust the
Cycle Timer setting in 5 minute increments between 5 minutes and 95 minutes.
NOTE:
To estimate the total cycle time, multiply the Cycle Timer setting by 4.
Static Mode
Static mode supplies low airloss therapy without alternating therapy.
1. Press the Static control. The Static indicator comes on.
NOTE: When Static mode is on, the Cycle Timer display is off.
2. If necessary, adjust the Comfort Control setting:
• Decrease pressure—press the Soft (-) control.
• Increase pressure—press the Firm (+) control.
3. Put your hand between the air cells and the foam pad in the mattress
assembly. Make sure at least 2″ (5 cm) exists between the patient and the top
of the foam pad.
Auto Firm Mode
Auto Firm mode quickly inflates the air cells to their maximum pressures to supply uniform firmness to help he caregiver do routine procedures. To turn the Auto Firm mode on or off, press the Auto Firm control. When the Auto Firm mode is on, the Auto Firm indicator comes on.
Power Failure Alarm
If power to the control unit is lost, or the Power switch is turned off, the
Power Failure indicator and alarm come on to show that power to the unit has
failed. To turn the alarm off, press the Alarm Reset control.
Cleaning
WARNING:
Follow the product manufacturer’s instructions. Failure to do so could result
in personal injury or equipment damage.
SHOCK HAZARD:
The potential for electrical shock exists with electrical equipment. Failure
to follow facility protocols may result in death or serious personal injury.
SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.
SHOCK HAZARD:
Do not expose the unit to excessive moisture. Personal injury or equipment
damage could occur.
CAUTION:
Do not use harsh cleansers, solvents, or detergents. Equipment damage could
occur.
General Cleaning
We recommend that you clean the unit with detergent and warm water.
Do not use excessive liquid, harsh cleansers, or phenol-based cleaning
solutions.
Control Unit
- Before you clean and examine the control unit, turn the Power switch Off, and unplug the unit from its power source.
- Use a damp soft cloth with mild detergent and warm water to clean the control unit.
- If soil is present, wipe down the control unit with a sodium hypochlorite solution diluted to 1000 ppm or an EPA registered (US only), tuberculocidal, disinfectant.
NOTE:
The control unit case is made from acrylonitrile butadiene styrene (ABS)
plastic.
Mattress
To clean the mattress on the bed, use a damp soft cloth with mild detergent
and warm water. Wipe down the air cells with a solution of sodium hypochlorite
(1000 ppm) or any other non-phonetic germicidal solution.
If the top coverlet or bottom encasement is soiled, either discard it, or do as follows to machine wash and thermally disinfect it:
- Put on applicable protective personal equipment (PPE).
Discard
a. Remove and discard the top coverlet or bottom encasement in accordance with the facility’s procedures for contaminated waste.
b. Replace with a clean top coverlet or bottom encasement.
Wash
Industrial laundry—
a. Break wash cycle in cold water for 10 minutes.
b. Main wash cycle I at 140°F (60°C) for 6 minutes.
c. Main wash cycle II at 158°F (70°C) for 10 minutes.
d. Cold rinse for three standard cycles
e. Extraction cycle for 5 minutes.
Normal laundry—
a. Pre-wash in cold water for a standard cycle.
b. Main wash cycle at 158°F (70°C) for 10 minutes.
c. Extraction cycle for 2 minutes.
d. Cold rinse for a standard cycle.
e. Extraction cycle for 5 minutes.
2. Let the top coverlet or bottom encasement air dry, or machine dry at 110°F (43°C) for one hour.
Steam Cleaning
Do not use any steam cleaning device on the unit. Excessive moisture can
damage mechanisms in this unit.
Cleaning Hard to Clean Spots
To remove difficult spots or stains, we recommend that you use standard
household cleansers and a soft bristle brush. To loosen heavy, dried-on soil,
you may first need to saturate the spot.
Disinfecting
When there is visible soil and also between patient use, we recommend that you
disinfect the unit with an EPA registered (US only), tuberculocidal,
disinfectant.
Dilute the disinfectant as specified on the manufacturer’s label.
Maintenance
WARNING:
Only Hill-Rom-authorized persons should service the Synergy® Air Elite Low
Airloss Alternating Therapy System. Service by unauthorized persons could
cause personal injury or equipment damage.
No daily maintenance is necessary. Only Hill-Rom-authorized persons should service the Synergy® Air Elite Low Airloss Alternating Therapy System. For minor problems, refer to “Troubleshooting” on page 13.
Service Calls
When you call Hill-Rom about your unit, be prepared to give the serial number
from the product identification label. When you give the serial number, the
Hill-Rom representative can identify your unit and give you the information
you need more quickly.
Troubleshooting
WARNING:
Only Hill-Rom-authorized persons should service the Synergy® Air Elite Low
Airloss Alternating Therapy System. Service by unauthorized persons could
cause personal injury or equipment damage.
Control Unit Does Not Operate
- Make sure the power cord is plugged into the control unit and an applicable power source that supplies the correct voltage.
- It may be necessary to replace the fuse in the control unit. Call Hill-Rom Technical Support for assistance.
Some Air Cells Have Lower Air Pressure than Others with Normal Air Pressure
- Put the unit in Static mode. Unzip the top coverlet from the bottom encasement, and examine the air tubes between the air cells and the main air supply line for kinks. Adjust the air tubes as necessary.
- Examine the air cells for leaks. If a leak is found, call Hill-Rom Technical Support for assistance.
Control Unit Supplies Air but Mattress Does Not Inflate
- Make sure the unit is plugged into an applicable power source that supplies the correct voltage.
- Examine the air tubes for kinks. Adjust the air tubes as necessary.
- Make sure the air tubes are correctly connected. Connect the air tubes as necessary.
- Unzip the top coverlet from the bottom encasement, and examine the air cells and air tubes for leaks. If a leak is found, call Hill-Rom Technical Support for assistance.
Storage and Handling
CAUTION:
Do not store the Synergy® Air Elite Low Airloss Alternating Therapy System in
direct sunlight or extreme cold conditions.
Equipment damage could occur.