Hill-Rom 242S Likorall Overhead Patient Lifts Installation Guide
- June 17, 2024
- Hill-Rom
Table of Contents
- Hill-Rom 242S Likorall Overhead Patient Lifts
- Product Information
- Product Description
- Safety Instructions
- Safety Instructions
- Definitions
- Technical Data
- Measurements
- Assembly
- Operation
- Charging the Batteries
- Charging via hand control
- MultiStation
- Transferring from room to room
- Maximum Load
- Recommended Lifting Accessories
- Quick-Release Hook
- Simple Troubleshooting
- Inspection and Maintenance
- FAQ
- Q: Can Likorall be used for gait training?
- Q: Are there different sizes and designs available for the slings?
- Q: Can Likorall be used to move a patient between separate rooms without
- Q: Is the lift suitable for use in areas with flammable mixtures?
- Q: What should I do if I have any uncertainties or questions?
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Hill-Rom 242S Likorall Overhead Patient Lifts
Installation Guide
7EN120115-03
2011-06-30
Applies to the following models:
Likorall 242 S Prod. no. 3122009-3122010
Likorall 242 ES Prod. no. 3122005-3122006
Likorall 242 S R2R Prod. no. 3122011-3122012
Likorall 242 ES R2R Prod. no. 3122007-3122008
Likorall 243 ES Prod. no.3123001-3123002
Likorall 250 ES Prod. no. 3122501-3122502
Likorall 250 S, IRC Prod. no. 3124050
Product Information
Specifications
-
Models: Likorall 242 S, Likorall 242 ES, Likorall 242 S R2R,
Likorall 242 ES R2R, Likorall 243 ES, Likorall 250 ES, Likorall 250 S, IRC -
Product Numbers:
- 3122009-3122010 (Likorall 242)
- 3122005-3122006 (Likorall 242 R2R)
- 3122011-3122012 (Likorall 243, 250)
- 3122007-3122008 (Likorall 242 ES)
- 3123001-3123002 (Likorall 243 ES)
- 3122501-3122502 (Likorall 250 ES)
- 3124050 (IRC)
Product Description
Likorall is one of the world’s most popular overhead lift motors. It is designed and developed for Liko’s complete range of fixed installed and freestanding lift systems. A Likolift system is always close at hand and easy to use.
All common lifts and transfers can be performed using Likorall, such as between bed/wheelchair, to/from floor, toilet visits, gait training, and together with stretchers. Likorall R2R (room to room) enables the patient to be moved between two rail systems in separate rooms without connecting rails and without making holes over doors.
A large and complete range of accessories are available for Likorall, including many different sling models in several sizes and designs.
Safety Instructions
Installation of Likorall to rail carriages must be made by personnel authorized by Liko and in accordance with the current installation instructions.
Before lifting, always ensure that:
-
The patient is never left unattended during a lifting situation.
-
Likorall has been tested by accredited testing institutes and complies with the requirements for MDD Class 1 products (MDD 93/42/EEC).
-
Likorall complies with the requirements in EN ISO 10535:2006, IEC 60601-1, IEC 60601-1-2, ANSI/AAMI ES60601-1, and CAN/CSA C22.2 no 60601-1.
-
The lift should not be modified under any circumstances.
Contact Liko/Hill-Rom for any questions. -
Particular care must be observed when using strong sources of electromagnetic disturbance, such as diathermy, to ensure diathermy cables are not positioned on or near the lift. Consult the responsible assistive-device technician or supplier for assistance.
-
The lift should not be used in areas where flammable mixtures may occur, such as areas where flammable goods are stored.
IMPORTANT!
Read the instruction guide for both the patient lift and lifting accessories before use. Lifting and transferring a person always involves a certain level of risk. It is important to completely understand the contents of the instruction guides. The equipment should be used only by trained personnel. Please contact your Liko/Hill-Rom representative in the event of any uncertainties or questions.
NOTE!
This instruction guide contains important information for users of the product. All those who use the product should review and fully understand the contents of the instruction guide. Remember to keep the instruction guide in a place where it is always available to those using the product.
Safety Instructions
Installation of Likorall to rail carriages must to be made by personnel authorized by Liko and in accordance with the current installation instructions.
Before using the first time make sure that:
- The lifting accessories are properly attached to the lift
- The batteries have been charged for at least 8 hours • you have read the instruction guide for the lift and lifting accessories
- Personnel using the lift are informed of the correct use of the lift and lifting accessories
- The lifting accessory is selected appropriately, in terms of type, size, material and design with regard to the patient’s needs.
Before lifting, always ensure that:
- The lift strap is not twisted or worn and can move in and out of the lift freely
- The lifting accessories are not damaged
- The lifting accessory is correctly and securely applied to the patient in order to prevent injuries
- The lifting accessories are properly attached to the lift
- The lifting accessories hang vertically and can move freely
- The sling bar latches are intact; missing or damaged latches must always be replaced
- The sling’s strap loops are correctly connected to the sling bar hooks when the sling straps are extended, but before the patient is lifted from the underlying surface.
Never leave a patient unattended during a lifting situation!
Likorall have been tested by accredited testing institutes and comply with the
requirements for MDD Class 1 products (MDD 93/42/EEC).
Likorall complies with the requirements in EN ISO 10535:2006, IEC 60601-1, IEC
60601-1-2,
ANSI/AAMI ES60601-1 and CAN/CSA C22.2 no 60601-1.
Under no circumstances must the lift be modified. If you have any questions, please contact Liko/Hill-Rom.
Particular care must be observed when using strong sources of electromagnetic disturbance, such as diathermy, etc, so that diathermy cables are not positioned on or near the lift. If you have questions, please consult the responsible assistive-device technician or the supplier.
The lift may not be used in areas where flammable mixtures may occur, for example in areas where flammable goods are stored
Likorall 242, 243, 250 • 7EN120115-03
Definitions
Technical Data
Likorall is equipped with a SFS (Single Fault Safety) safety drum. This patented safety design provides protection against uncontrolled lowering. The lift strap has a tenfold safety.
Measurements
- Minimum distance from ceiling to CSP at maximum lifting height.
** Built-in dimension: the distance between the attachment point for the lift unit on the carriage and the CSP at maximum lifting height.
*** Lifting interval: the distance between maximum lifting height and minimum lifting height measured in CSP.
Assembly
Handle reinforcement (exclusively: Likorall 242 S, ES)
Press the reinforcement (included in the bag containing the instruction guide)
labelled ”Emergency lowering” onto the emergency lowering handle.
Operation
Hand Control Likorall
Likorall is operated with a light push on the buttons of the hand control. The arrow directions should be the same as the direction shown in the illustration. The movement ceases when the button is released. For the Likorall ES, there is a number of hand controls to choose from, depending on how the lift and rail system are equipped, and an IR hand control for wireless operation.
If needed, the lifting motion can also be controlled without hand control via the buttons and respectively, on the end cover of the lift, see page 6.
Emergency stop
For emergency stop: pull the red emergency stop cord.
To reset: turn the button in the direction of the arrows.
The red cord on the lift motor´s end cover is intended for use if an emergency
situation occurs.
Lifting accessory with quick-release hook (Likorall: 242 R2R, 243 ES, 250
ES)
Push down the red catch and connect the quick-release hook to the Q-link.
Release and check that the catch locks in order to prevent involuntary
unhooking from the Q-link. Read more about Liko´s Quick-release Hook system on
page 9.
Mechanical emergency lowering (Likorall 242 S/ ES)
- Move the emergency lowering handle up and down until the patient has been lowered and the lift strap is completely slack. Always ensure that the emergency lowering is made to a bed, wheelchair or other suitable location.
- After mechanical emergency lowering has been performed,
resetting/adjustment of the lifting height is required:
– Lower the sling bar so that the lift strap is completely slack.
– Hold down the emergency lowering handle halfway.
Simultaneously, tighten the lift strap by turning the black wheel counter
clockwise with the other hand.
Repeat until the desired height has been achieved.
Installation of Latches
After installation, check that the latch locks and runs freely in the sling bar hook.
Electrical emergency lowering
Emergency lower by pushing in the button on the end cover of the lift. Always
ensure that the emergency lowering is made to a bed, wheelchair or other
suitable location.
Carriages with adjustable friction brake
The amount of drag along the rail can be adjusted on carriages equipped with a
friction brake. Turn the brake clockwise for increased resistance and counter
clockwise to reduce resistance.
The following carriages have a friction brake: prod. no. 3126011 and 3126015.
Hang-up HandControl Hanger
When not in use, the hand control can be hung on the Hang-up Hand control hanger.
SSP Limit Switch
The lifting motion is stopped electrically with a light touch on the SSP Limit
Switch from, for instance, the strap stops, Q-link or if the lift strap is
subjected to harmful strain, for example, if it is pulled sideways or folded
over during the lifting motion. If the SSP Limit Switch is activated so that
the lifting motion stops, the lift can once again be operated after the lift
strap is straight again (a short delay in the lifting motion is normal in
these cases).
The SSP Limit Switch protects the lift motor from mechanical strain and it
also prevents squeeze injurie.
Ensure that the lift strap is kept straight and stretched when it runs in and
out of the lift motor.
Indication: charging
Likorall indicates in two ways that the battery needs charging:
• Buzzer: sounds when lifting
• LED,(A) : flashes (red) when lifting
When either of these signals sounds or illuminates, the lift should be charged
as soon as possible.
See Charging the batteries, page 7.
Never move the lift by pulling the hand control!
Charging the Batteries
In order to ensure maximum battery life, it is important to charge the batteries regularly. We recommend that you charge the batteries after use or every night. Full charge is achieved after max. 8 hours. Fully charged batteries will last for approximately 60 lift cycles.
The lift cannot be charged with the emergency stop engaged.
NOTE! If the lift will not be used for a longer period of time, a charging system should be activated to allow the battery to charge or the emergency stop button should be pressed in order to prevent the battery from discharging.
Likorall has several charging systems, these systems must never be activated parallelly.
Charging via hand control
1. Check that the emergency stop button is not pressed.
2. Place the hand control in the charger’s socket.
3. Connect the charger to a 100-240 V AC electric socket.
4. A green LED illuminates to indicate that the charger is connected to the
electrical grid.
5. The charging starts automatically, and a yellow LED indicates that the
batteries are charging.
6. When the battery is fully charged, the charger shuts off automatically and
the yellow LED turns off.
Alternative charging procedure
Charging the Batteries
In-Rail Charging
Liko´s In-Rail Charging system is an easy to use charging system. When using
the In-Rail Charging system the lift will charge continously. During active
lifting the In-Rail Charging system will “pause” and re-activate charging
automatically after completed operation.
Indications from IRC:
– A green LED-light on the lift indicate that the lift is ready to use.
– A yellow LED-light on the lift indicate that the lift has low battery
capacity, contact Liko if the problem persist.
For more information, please contact your Liko reprecentative or visit our website www.liko.com.
MultiStation
As an alternative to charging via the hand control,
the batteries can be charged with a MultiStation installed on the rail system.
In this case, Likorall must be equipped with a contact rail or a transfer
motor. The Likorall batteries are then charged by parking the lift in the
charging position under the MultiStation (see figure).
This charging procedure is also suitable when Likorall is operated by HandControl IR.
Likorall comply with the Directive 2002/96/EC on waste electrical and
electronic equipment.
Old batteries are to be deposited at the nearest recycling facility in
accordance with local regulations or given to personnel authorized by Liko
/Hill-Rom.
Transferring from room to room
Liko’s R2R Room-to-Room system is an effective solution for safe and easy transfer of patients between two or more rooms. The R2R system is mounted without making openings in walls over doors and full isolation is therefore retained between the rooms supported by the system.
The transfer is performed in a safe and easy manner, with the aid of separate rail systems for each room. Liko’s R2R system enables linking together two Likorall when transferring from room to room. The actual transfer operation between two rooms is performed with a comfortable transition for the patient from one Likorall to another.
Mounting Universal SlingBar R2R (Room-to-Room)
The Universal SlingBar R2R has a double hook with Quick-release. The double
hook has protection against unintentional detachment.
1. Move Lift 1 with the patient as close to the door way as possible. Lower the lift as far as possible, bearing in mind the patient´s comfort.
2. Move Lift 2 as close to the door opening as possible.
Lower the lift strap from Lift 2 a sufficient length and connect the Q-link to
the Universal SlingBar R2R.
Check that the safety latches on the R2R connector function properly
3. Raise Lift 2. The patient is successively moved to the next room and finally suspended in Lift 2 only. When the pressure is relieved from the lift strap for Lift 1, disconnect the lift strap from the Universal SlingBar R2R and the transfer can be carried out in the next room.
NOTE! To free the Q-link from the Universal SlingBar R2R, it may be necessary to let out additional strap from Lift 1.
Maximum Load
Different maximum loads may apply to different products on the assembled lift system: lift, sling bar, sling and any other accessories used. For the assembled lift system, the maximum load is always the lowest maximum load rating for any of the components. For example, a Likorall which is approved for 200 kg (440 lbs) can be equipped with a sling bar which is approved for 300 kg (660 lbs). In this case, the maximum load of 200 kg (440 lbs) applies to the assembled lift system. Study the markings on the lift and lifting accessories or contact your Liko/Hill-Rom representative if you have any questions.
Recommended Lifting Accessories
Using lifting accessories other than those approved can entail a risk.
The Liko assortment contains many sling bars, slings, stretchers, scales and
other accessories to solve most lifting needs. Below is a list of lifting
accessories suitable for Likorall.
To select suitable slings and other lifting accessories, please see the
“Lifting accessories” brochure. For additional guidance in selecting a sling,
study the instruction guide for the respective sling models. Here you will
also find guidance for combining Liko’s sling bars with Liko’s slings.
Contact yourLiko/Hill-Rom representative or visit www.liko.com for advice and
information on Liko’s product range
SlingBar Mini 220 Prod. no. 3156005
Max. load 205 kg (450 lbs)
Universal SlingBar 350 with Quick-release Hook™ Prod. no. 3156084
Fixed connection, prod. no. 3156074
Max. load 300 kg (660 lbs)
Universal SlingBar 450 with Quick-release Hook™ Prod. no. 3156085
Fixed connection, prod. no. 3156075
Max. load 300 kg (660 lbs)
Universal SlingBar 600 with Quick-release Hook™ Prod. no. 3156086
Fixed connection, prod. no. 3156076
Max. load 300 kg (660 lbs)
Universal SlingBar 670 Twin with Quick-release Hook™ Prod. no. 3156087
Fixed connection, prod. no. 3156077
Max. load 300 kg (660 lbs)
Universal SideBars 450 Prod. no. 3156079
including bag
Max. load 300 kg (660 lbs)
Sling Cross-bar 450 with Quick-release Hook™ Prod. no. 3156022
Fixed connection, prod. no. 3156021
Max. load 300 kg (660 lbs)
Sling Cross-bar 670 with Quick-release Hook™ Prod. no. 3156019
Fixed connection, prod. no. 3156018
Max. load 300 kg (660 lbs)
Sling bars with fixed connection can be equipped with Quick-release Hook
Quick-Release Hook
Liko’s Quick-release Hooks are a system for quick change of lifting accessories on Liko’s mobile and stationary lifts equipped with the Q-link. The Quick-release Hook can be installed on lifting accessories with a fixed connection.
The Quick-release Hook Universal fits the Universal SlingBar 350, 450 and 600 (prod. no. 3156074 – 3156076). Quick-release Hook TDM fits the SlingBar Mini 220 (prod. no. 3156005), Sling Cross-bar 450 and 670 (prod. no. 3156021 and 3156018) and Universal TwinBar 670 (prod. no. 3156077).
See the “Guide to Liko’s Quick-release Hook System”, which can be downloaded from our website www.liko.com, or contact Liko/Hill-Rom for more information about the advantages and uses of the Quick-release Hook system.
Stretcher
All stretchers in Liko’s product range can be used with Likorall.
Contact Liko/Hill-Rom for further information.
Scale
For weighing persons in combination with the Likorall,
we recommend LikoScale 350. This can easily be installed with the LikoScale
Adapter Kit.
Please contact Liko/Hill-Rom for more information
Liko Support Springs for bounce and gait training
Is used in order to produce a softer, springy motion during gait training. The
springs are available in two sizes: short prod. no. 3156512 and long prod. no.
3156511.
These springs are intended for patient weights up to 70 kg (154 lbs).
For patient weights up to 100 kg (220 lbs): prod. no. 3156513 (short).
Also see the instruction guide for Liko MasterVest, models 60 and 64 or
Liko LiftPants model 92, for more information.
Liko Bathing Chair Prod. no. 3156025
Contact Liko/Hill-Rom for further information.
SlingBar Cover Paddy 30 Prod. no. 3607001
Fits Universal SlingBars 350, 450 and 600, as well as SlingBar Slim 350
Hang-up HandControl Hanger Prod. no. 3156100
Sold in set of 10 pcs.
Multi-Connector Prod. no. 3126111
Intended for installation on Likorall for control of switches and/or if the
rail system is equipped with a MultiStation for charging via the rail.
Transfer motor Likorall ES Prod. no. 3126044
Fits Likorall with ES designation
Max. 250 kg (550 lbs)
HandControl Remote IR Prod. no. 3126060
Likorall with ES designation can be equipped with a wireless hand control
(IR). The hand control operates normally within a range of 0-5 meters (0-196
inch.) from the lift.
Quick-Release Carriage LR Prod. no. 3126008
Fits the Quick-release Adapter LR prod. no. 3126009
Quick-Release Adapter LR Prod. no. 3126009
Fits Quick-Release Carriage LR prod. no. 3126008
Easy Switch
Developed for users/patients who have reduced strength or movement of the
hand. Easy to install on the hand control.
2-button Prod. no. 3107010
4-button Prod. no. 3107011
Fits Likorall hand controls prod. no. 3126034, 3126035 and 3126036.
Parking Panel 600, LR/MR Prod. no. 3126075
Parking Panel 1500, LR/MR Prod. no. 3126080
Can be completed with the following accessories:
Hook for SlingBar Prod. no. 3126070
Hook for Accessories Prod. no. 3126071
Bracket for Charger Prod. no. 3126100
Quick Reference Guide (see respective product)
Simple Troubleshooting
The lift doesn’t work
1. Check to ensure that the emergency stop button is not pressed in.
2. Check that the hand control is correctly connected.
3. Charge the battery.
4. If the lift still does not work properly, please contact Liko/Hill-Rom.
A repeated signal can be heard from the lift
1. Charge the battery immediately.
2. If the lift still does not work properly, please contact Liko/Hill-Rom.
The lift stops in the high position
1. Check to ensure that the emergency stop button is not pressed in.
2. Check that the hand control is correctly connected.
3. Use the selected mechanical or electrical emergency lowering device to
lower the patient onto a firm surface.
4. Charge the battery.
5. If the lift still does not work properly, please contact Liko/Hill-Rom.
The lift does not achieve maximum lifting capacity
1. Charge the battery.
2. If the lift still does not work properly, please contact Liko/Hill-Rom.
In event of unusual sounds or any leakage from the lift
Inspection and Maintenance
Care and Maintenance
For trouble-free operation, certain details should be checked each day the lift is used:
- Inspect the lift and check to make sure that there is no external damage.
- Check the sling bar attachment.
- Check the lift strap for wear and to ensure the strap is not twisted.
- Check the functionality of the latches.
- Check the operation of the lift movement.
- Check to make sure that the emergency lowering functions correctly.
- Check that the mechanical emergency lowering functions and that the lifting height is set correctly.
- Charge the batteries each day the lift is used and check to ensure that the charger works.
When necessary, clean the lift with a moist cloth, using warm water or disinfectant.
NOTE! Do not use cleaning agents that contain phenol or chlorine, since these can damage aluminium and plastic materials.
The lift should not be exposed to running water.
Service
Likorall should be periodically inspected at least once a year.
Periodic inspection, repair and maintenance should be performed only in
accordance with the Liko
Service Manual and by personnel authorized by Liko and using original Liko
spare parts.
Service Agreement
Liko offers the opportunity to enter into service contracts for the
maintenance and periodic inspection of your Liko products.
Expected Life Time
The product has an expected life time of 10 years when correctly handled,
serviced and periodically inspected in accordance with Liko’s instructions.
Transport and Storage
During transportation, or when the lift is not to be used for a long time, the
emergency stop should be engaged.
The environment where the lift is transported and stored should have a
temperature of 10–40 °C (50–104 °F) and a relative humidity of 30 to 75 %. The
air pressure should be 700–1060 hPa.
Recycling
For instructions on how to recycle your Liko product, please visit our
website: www.liko.com.
Product Changes
Liko’s products undergo continuous development, which is why we reserve the
right to make product changes
without prior notice. Contact your Liko/Hill-Rom representative for advice and
information about product upgrades.
Design and Quality by Liko in Sweden
Liko is quality certified in accordance with ISO 9001 and its equivalent for
the medical device industry, ISO 13485.
Liko is also certified in accordance with environmental standard ISO 14001.
FAQ
Q: Can Likorall be used for gait training?
A: Yes, Likorall can be used for gait training along with other common lifts and transfers.
Q: Are there different sizes and designs available for the slings?
A: Yes, Likorall offers a large and complete range of accessories, including many different sling models in several sizes and designs.
Q: Can Likorall be used to move a patient between separate rooms without
connecting rails?
A: Yes, Likorall R2R (room to room) enables the patient to be moved between two rail systems in separate rooms without the need for connecting rails or making holes over doors.
Q: Is the lift suitable for use in areas with flammable mixtures?
A: No, the lift should not be used in areas where flammable mixtures may occur, such as areas where flammable goods are stored.
Q: What should I do if I have any uncertainties or questions?
A: Please contact your Liko/Hill-Rom representative in the event of any uncertainties or questions.