WinAPEX ET8138 Touch Sensitive Multimeter with Colour Display Instruction Manual
- June 6, 2024
- WinAPEX
Table of Contents
6000counts
Operation Manual of Pocket Type Intelligent Scanning Digital
Multimeter
Overview
6000counts is a pocket type 3 5 / 6-digit pocket type real effective value
intelligent scanning digital instrument; the series is triple display: main
display, secondary display, analog bar. Except for the display frequency of AC
voltage measurement sub-display, another function sub-display is normal
temperature. In the process of using, users can operate continuously. There is
no need to turn the dial to select the function. The meter will automatically
recognize and measure based on the input AC or DC current/resistance. It has
stable performance, high precision, high reliability, clear reading, and
overload protection. Driven by AAA 1.5V battery, the instrument adopts a
large-screen LCD display and adopts a boost power supply. Even at the edge of
a 2.3V low battery, it can ensure the high brightness of the backlight and
flashlight. The meter is easy to carry and is very popular among users.
This series of instruments can manually switch and measure DC voltage and AC
voltage, capacitance, NDV, diode, continuity test, live wire judgment,
frequency measurement, and true RMS. It is a superior tool and an ideal tool
for laboratories, factories, radio enthusiasts, and families.
Safety Precautions
This series of instruments is designed in accordance with IEC1010 (Safety Standards promulgated by the International Electrotechnical Commission), please read the safety precautions before use.
- When measuring voltage, please do not input the limit voltage exceeding the effective value of DC 700V or AC 500V;
- The voltage below 36V of the current file is a safe voltage;
- When changing functions and ranges, the test pen should leave the test point;
- Select the proper function and range, beware of the wrong operation, although the series of instruments with full-range protection, for safety reasons, you still pay more attention;
- Safety Symbol Description: “ ”Exist dangerous voltage,“ ” Ground,“ ” Double insulation, “ ”Operator must refer to the instruction manual,“ ”Low voltage symbol.
Characteristics
1. General Characteristics
1-1. Display mode: triple display LCD; main display, sub-display, analog bar;
1-2. Maximum display: 5999 (3 5/6) bit automatic polarity display;
1-3. Measurement method: double integral A/D conversion;
1-4. Sampling rate: about 3 times per second;
1-5. Over range display: The highest position shows “OL”;
1-6.Continuous power operation function;
1-7. Low voltage display:“ ”symbol appears;
1-8. Working environment:(0~40)℃, relative humidity<80%;
1-9. Power: AAA 2* 1.5V battery;
1-10. Volume (size): 139×68×18mm (L×W×H);
1-11. Weight: about 175g(including 1.5V battery);
1-12. Attachment: One instruction manual, one certificate, one outer packaging
box, one pair of test leads, two AAA1.5V batteries.
2. Technical characteristics
2-1. Accuracy: ±(a% of reading + least significant digit), to ensure accuracy
Ambient temperature: (23 ±5)℃, relative humidity <75%, calibration guarantee
period from the date of manufacture for one year.
2-2. Performance (Note “▲” indicates that the meter has this function)
Function|
---|---
DC voltage DCV| ▲
AC voltage ACV| ▲
Resistor / Diode / On-Off Test / Capacitor| ▲
Frequency F| ▲
Touch key| ▲
Continuous power operation| ▲
NCV| ▲
Zero line / Fireline test| ▲
Full unit symbol| ▲
Backlight manual/Auto-off| ▲
True RMS measurement| ▲
Temperature(℃/℉)| ▲
Flashlight lighting| ▲
Operation panel instructions
-
Simulation bar;
-
LCD display;
-
Live wire simulation indication;
-
Normal temperature display status;
-
Quick selection of capacitor/flashlight key;
-
Data lock key;
-
Power on and off for more than 3 seconds; short trigger for AC voltage/
DC voltage / diode / buzzer / capacitor switching, so as to cycle; -
Electric field measurement (NCV) / neutral wire measurement (LIVE)/more than 3s℃/℉
-
The red probe is the input port of voltage, resistance, diode, capacitor, frequency, and buzzer; Black is the negative end of the input.
Note:
-
The backlight is always on, LCD is highlighted, and the user is constantly on In use, the instrument will not turn off the power supply; when it stops using, it will automatically Power off; truly intelligent operation without power off, refer
to the power off instructions for details. -
The table adopts a touch design for keys, which is more safe and reliable. When using touch, Every time a key is triggered, it is better to have an interval of more than 1 second to avoid false triggering
-
When touching any touch key, the touch key and panel are on, and you can switch the function key.
After 5 seconds, the touch key and panel light are turned off and locked.
Technical Specifications
1. DC voltage / AC voltage automatic scanning test (DCV / ACV)
Accuracy/ Range | 6000counts | Resolution |
---|---|---|
DC/AC6V | ±(0.8%+5) | 0.001V |
DC/AC60V | 0.01V | |
DC/AC600V | 0.1V | |
DC1000/AC750V | ±(0.8%+10) | 1V |
Input impedance: 10MΩ; overload protection: 1000V DC or 750V AC feature
Validity Measurement: frequency response 50Hz-1kHz;
The specific operations are as follows:
1-1. Long press POWER is greater than 2S, the boot display is automatically
scanned state “AUTO”.
1-2. The black test lead is negative, the red test lead is positive, and the
test lead reliably touches the test point.
1-3. When the voltage measured at the negative end of the black test lead
input port and the positive end of the red test lead is greater than 0.8V,
regardless of the AC voltage or the DC voltage, the meter will compare the DC
component and the AC component, take the larger component signal, and then
according to the measured value The size is automatically switched between
6V/60V/600V/1000V, and then the measured value is displayed on the LCD.
Note:
- The input voltage must not exceed DC1000V or AC750V. If it exceeds, there is a danger of damage to the instrument circuit. When high voltage circuit, pay special attention to avoid electric shock;
- After completing all measurement operations, disconnect the test leads from
the circuit under test.
2. Resistance (Ω)
Range | 6000counts | Resolution |
---|---|---|
600Ω | ±(0.8%+5) | 0.1Ω |
6kΩ | ±(0.8%+3) | 1Ω |
60kΩ | 10Ω | |
600kΩ | 100Ω | |
6MΩ | 1kΩ | |
60MΩ | ±(2.5%+3) | 10kΩ |
Input impedance: 10MΩ; overload protection: 1000V DC or 750V AC peak.
The specific operations are as follows:
2-1. The boot display is in the automatic scan status “AUTO”.
2-2. The black test lead is negative, the red test lead is positive, and the
test lead reliably touches the test point.
2-3. If the measurement resistance at both ends of the test lead is less than
50Ω, the buzzer will emit a continuous beep, and fast buzzer measurement is
required. Cycle the power button to enter the buzzer for rapid measurement.
There will be a “ ” symbol displayed on the LCD, fast buzzer/diode
measurement, this function can be automatically switched according to the
needs of the measurement.
2-4. If you are measuring the resistance of a closed-loop, you must discharge
the resistance at both ends of the resistance to be measured. Otherwise, if
the voltage in the loop is greater than 0.8V, the meter will mistake it for
voltage measurement and enter the electricity measurement mode.
2-5. When measuring the value at the negative end of the black test lead input
port and the positive end of the red test lead, the meter will automatically
switch between 600Ω/6k Ω/60kΩ 600kΩ/6M Ω/60MΩ according to the resistance
measurement value, and then the measured value will be displayed on the LCD.
Note:
- When measuring low resistance, the test leads will bring internal
resistance. In order to obtain accurate readings, you can record the short
circuit
value of the test leads first, and subtract the value when the test leads are short-circuited from the measurement readings;
Fast continuity test / diode / capacitor
Range
| display value|
test condition
---|---|---
“ AUTO| Diode forward voltage drop| Forward DC current is about 1mA, open
circuit voltage is about 3V
The buzzer sounds long and the resistance of the test is less than (50±20) Ω.|
Open circuit voltage is about 0.4V, press “power” for two-speed function
switching
Capacitance(C)
Accuracy/
Range| 6000counts|
Resolution
---|---|---
10nF| ±(3.5%+20)| 10pF
100nF| 100pF
1uF| 1nF
10uF| 10nF
100uF| 100nF
1mF| 1uF
10mF| 10uF
60mF| ±(5%+3)| 100uF
Overload protection: 1000V DC, 750V AC peak.
4-1. The boot display is in the automatic scan status “AUTO”.
4-2. The black test lead is negative, the red test lead is positive, and the
test lead reliably touches the test point.
4-3. If you need fast continuity test/diode/capacitance measurement,
constantly trigger the “power” switch, enter the fast continuity
test/diode/capacitance measurement in one cycle, and select the corresponding
function measurement according to the measurement requirements.
When measuring the capacitance, the measured capacitance The size will
automatically select different ranges, and the measured value will be
displayed on the LCD. The capacitance measurement range is 10nF/100nF
/1uF/10uF/100uF/1mF/10mF/60mF.
Note:
- When measuring capacitance with a 10nF file, there may be residual reading on the screen display value. This number is the distributed capacitance of the test pen. It is an accurate reading and can be subtracted after measurement.
- When the large capacitance file measures severe leakage or breakdown capacitance, some values will be displayed and unstable; when measuring large capacitance, the reading takes several seconds to stabilize, which is normal when measuring large capacitance;
- Please fully discharge the capacitor before testing the capacitor capacity, otherwise, it will enter the voltage measurement mode.
- Unit: 1F=1000mF, 1mF=1000uF, 1uF=1000nF, 1nF=1000pF
5. NCV measurement;
The operation is as follows:
5-1. Press and hold the “NCV/LIVE” button; enter the EF measurement.
5-2. The front end of the meter has NCV test points. As long as the point is close to the AC voltage, the buzzer will emit different continuations according to the different strength of the signal, and the LCD will also display according to the strength of the signal. A different number of segments.
5-3. Cycle to trigger “NCV/LIVE” key, enter live live live live wire measurement, display live character on LCD, close to live wire test point with a red meter, and make reliable contact with the point, the black meter end is suspended, and does not contact any test point; if the test point is a live wire, the LCD will display OL, and the buzzer will send continuous sound.
Frequency measurement
Range | 6000counts | Resolution |
---|---|---|
10Hz | ±(0.1%+3) | 0.01Hz |
100Hz | 0.1Hz | |
1kHz | 1Hz | |
10kHZ | 10Hz | |
100kHz | 100Hz | |
1MHz | 1kHz | |
10MHz | 10kHz |
Input sensitivity: 1V RMS; overload protection: 550V DC or AC peak (no more
than 10 seconds)
Frequency measurement
- trigger power key; switch to frequency measurement function;
- The black test lead is negative, the red test lead is positive, and the test lead reliably touches the test point;
- Measurement values will be displayed on the LCD.
Note:
- When the input exceeds 10Vrms, it can be read, but the error may be larger;
- In a noisy environment, it is best to use a shielded cable when measuring small signals;
- When measuring high voltage circuits, special care should be taken to avoid electric shock;
- It is forbidden to input voltage values exceeding 250V DC or AC peak to avoid damage to the meter.
Automatic switch machine
When the instrument is out of service for about 5 minutes, the instrument will
automatically power off; to restart the power supply, long-press the “power”
key for more than 3 seconds, and the LCD will display as “auto scan” AUTO “,
In order to make it more convenient for the user, the design of the machine
adds the design of continuous operation and power supply, that is, the user
will not automatically shut down the machine during the use process, and can
only enter the shutdown state after 5 minutes of stop using, be careful:
- In the manual nonmeasurement state, no matter the AC voltage or DC voltage level because the instrument probe is completely welded at the input end, there will be certain data on the LCD. In the design, we define the manual DC voltage level, the input end has 10 words, the AC voltage level has 100 words, and the capacitance has 100 words. In this state, we default it to no measurement state, 5 minutes It will turn off automatically.
Troubleshooting
If your instrument does not work, the following method can help you solve the general problem, if the fault still can not be excluded, please contact the service center or dealer.
Failure phenomenon | Inspection site and method |
---|---|
Not shown | Battery not connected |
Replace the battery
Low battery symbol| Replace the battery
Current is not input| Replace fuse
Resistance display error| The test pen is not in contact
This manual is subject to change without notice;
The contents of this manual are considered correct. If the user finds any
errors, omissions, etc., please contact the manufacturer;
The company does not bear the accidents and hazards caused by the user’s wrong
operation;
The functions described in this manual are not intended for the reasons of the
product for special purposes.
60000-0128C200813