HACH DR300 Pocket Colorimeter User Manual
- June 6, 2024
- HACH
Table of Contents
- Section 1 Specifications
- Section 2 General information
- 2.1 Safety information
- 2.1.1 Use of hazard information
- 2.1.2 Precautionary labels
- 2.1.3 Certification Canadian Radio Interference-Causing Equipment
- FCC Part 15, Class “B” Limits
- Operation is subject to the following conditions:
- 2.2 Product overview
- 2.3 Product components
- Section 3 Install the batteries
- Figure 2 Install the batteries
- Section 4 User interface and navigation
- 4.1 Keypad description
- Section 5 Set the time
- 3 Shows when the Hach Communication Dongle is installed.
- Refer to Download a test procedure on page 14.
- Figure 5 Sample cell orientation
- Figure 6 Install the instrument cap over the cell holder
- Section 7 Show measurements
- Section 8 Calibration
- 8.1 Standard calibration adjust
- 8.1.1 Do a standard calibration adjust
- 8.1.2 Set the standard calibration adjust to off
- 8.2 User-entered calibration curve
- 8.2.1 Enter a calibration curve with standards
- 8.2.2 Enter a calibration curve with the keypad
- 8.2.3 Set to the factory default calibration
- Section 9 Maintenance CAUTION
- 9.1 Clean the instrument
- 9.3 Replace the batteries
- Section 10 Troubleshooting
- 8 The value that flashes will be 10% over the upper test range limit.
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
DR300 Pocket Colorimeter
DOC022.97.90639
DR300
08/2021, Edition 4
User Manual
Section 1 Specifications
Specifications are subject to change without notice.
Specification
|
Details
---|---
|
Dimensions (W x H x D) 6.9 x 15.7 x 3.4 cm (2.7 x 6.2 x 1.3 in.)
Enclosure
|
IP67, waterproof at 1 m (3.3 ft) for 30 minutes when battery compartment is closed and locked.
Light source
|
Light emitting diode (LED)
Detector
|
Silicon photodiode
Display
|
LCD with backlight
Weight
|
0.25 kg (0.55 lb)
Power requirements
|
4 AAA batteries; approximate life of 5000 tests (use of backlight decreases this number)
Rechargeable batteries are not recommended.
Operating environment
|
0 to 50 °C (32 to 122 °F), 0 to 90% relative humidity non-condensing
Storage temperature
|
–20 to 55 °C (–4 to 131 °F), 0 to 80% relative humidity non condensing
Wavelength
|
Fixed wavelength ±2 nm, different for each model
Filter bandwidth
|
15 nm
Absorbance range
|
0 to 2.5 Abs
Sample cell
|
25 mm (10 mL) and 1 cm (10 mL)
Data storage
|
Last 50 measurements
Specification
|
Details
---|---
Bluetooth® 1
|
Bluetooth® is on when the optional Hach Communication Dongle is installed.
Certifications
|
CE
Warranty
|
1 year (EU: 2 years)
Section 2 General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential damages resulting from any defect or omission in this manual. The manufacturer reserves the right to make changes in this manual and the products it describes at any time, without notice or obligation. Revised editions are found on the manufacturer’s website.
2.1 Safety information
The manufacturer is not responsible for any damages due to misapplication or
misuse of this product including, without limitation, direct, incidental and
consequential damages, and disclaims such damages to the full extent permitted
under applicable law. The user is soley responsible to identify critical
application risks and install appropriate mechanisms to protect processes
during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this
equipment. Pay attention to all danger and caution statements. Failure to do
so could result in serious injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do
not use or install this equipment in any manner other than that specified in
this manual.
1 The Bluetooth® word mark and logos are registered trademarks owned by the Bluetooth SIG, Inc. and any use of such marks by HACH is under license.
2.1.1 Use of hazard information
DANGER
Indicates a potentially or imminently hazardous situation which, if not
avoided, will result in death or serious injury.
WARNING
Indicates a potentially or imminently hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or
moderate injury.
NOTICE
Indicates a situation which, if not avoided, may cause damage to the
instrument. Information that requires special emphasis.
2.1.2 Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the instrument could occur if not observed. A symbol on the instrument is referenced in the manual with a precautionary statement.
|
This symbol, if noted on the instrument, references the instruction manual for operation and/or safety information.
---|---
|
Electrical equipment marked with this symbol may not be disposed of in European domestic or public disposal systems. Return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.
2.1.3 Certification Canadian Radio Interference-Causing Equipment
Regulation, ICES-003, Class B:
Supporting test records reside with the manufacturer. This Class B digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
FCC Part 15, Class “B” Limits
Supporting test records reside with the manufacturer. The device complies with Part 15 of the FCC Rules.
Operation is subject to the following conditions:
- The equipment may not cause harmful interference.
- The equipment must accept any interference received, including interference that may cause undesired operation.
Changes or modifications to this equipment not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at their expense. The following techniques can be used to reduce interference problems:
- Move the equipment away from the device receiving the interference.
- Reposition the receiving antenna for the device receiving the interference.
- Try combinations of the above.
2.2 Product overview
This instrument is a portable filter photometer used for testing water.
Note: This instrument has not been evaluated to measure chlorine and
chloramines in medical applications in the United States.
2.3 Product components
Make sure that all components have been received. Refer to Figure 1. If any items are missing or damaged, contact the manufacturer or a sales representative immediately. Figure 1 is an example and shows the parts supplied with LPV445.99.00110. Other instruments come with different components.
1 DR300
|
5 Storage case
---|---
2 AAA alkaline batteries
|
6 Reagents
3 Sample cells, 25 mm (10 mL), glass
|
7 Hach Communication Dongle (optional, supplied separately)
4 Sample cells, 1 cm (10 mL), plastic
Section 3 Install the batteries
WARNING
Explosion hazard. Incorrect battery installation can cause the release of
explosive gases. Be sure that the batteries are of the same approved chemical
type and are inserted in the correct orientation. Do not mix new and used
batteries.
Refer to Figure 2 to install the batteries. Then, push to set the instrument to on.
Figure 2 Install the batteries
1 Coin
2 Battery cover
3 Plastic insert for dongle2
Section 4 User interface and navigation
4.1 Keypad description
Figure 3 shows the keypad and gives the key functions.
2 Only remove the plastic insert to install the Hach Communication Dongle.
Refer to the installation instructions supplied with the dongle.
8 English
-
Range key: Selects the measurement range (e.g., LR or HR). Push and hold for 3 seconds to enter or exit menu mode.
In menu mode, scrolls up or increases the value of the selected digit. -
Zero key: Sets the zero value before a measurement. In menu mode, goes back one menu level or moves the cursor to the previous digit.
-
Power key: Sets the power to on and off. Push and hold for 5 seconds to reset the instrument. The calibration is not deleted.
-
Backlight key: Sets the backlight to on and off. In menu mode, scrolls down or decreases the value of the selected digit.
-
Read key: Starts a sample measurement. In menu mode, selects the menu option shown or moves the cursor to the next digit.
4.2 Display description Figure 4 shows the values and icons shown on the
display.
1 Numeric display: Measured value or menu options
2 Range icon: Points to the selected measurement range
3 Measurement ranges or parameters
4 Bluetooth® icon: Bluetooth® is on3.
5 Battery icon: Battery power level. Flashes when the battery power level is
low.
6 Parameter and measurement ranges
7 Calibration adjusted icon: The factory default calibration was adjusted or a
user-entered calibration curve was entered.
Section 5 Set the time
Set the time (24-hour format).
- Push and hold for 3 seconds to enter menu mode. The time shows (or 00:00).
- Push to set the time.
- Push the or to change the number that flashes. Push to the next digit. Push to go to the previous digit. to go
3 Shows when the Hach Communication Dongle is installed.
Section 6 Do a test DANGER
Chemical or biological hazards. If this instrument is used to monitor a
treatment process and/or chemical feed system for which there are regulatory
limits and monitoring requirements related to public health, public safety,
food or beverage manufacture or processing, it is the responsibility of the
user of this instrument to know and abide by any applicable regulation and to
have sufficient and appropriate mechanisms in place for compliance with
applicable regulations in the event of malfunction of the instrument.
DANGER
Chemical exposure hazard. Obey laboratory safety procedures and wear all of
the personal protective equipment appropriate to the chemicals that are
handled. Refer to the current safety data sheets (MSDS/SDS) for safety
protocols.
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with
local, regional and national regulations.
The generic steps to do a test follow. To do a test for a specific parameter
(e.g., chlorine), download the test procedure from the manufacturer’s website.
Refer to Download a test procedure on page 14.
1. Push to select the applicable measurement range (e.g., LR or HR).
2. Prepare the blank. Refer to the test procedure.
3. Clean the sample cell with a no-lint cloth.
4. Insert the blank sample cell into the cell holder. Make sure to install
the blank sample cell in the correct and consistent orientation so that the
results are more repeatable and precise. Refer to Figure 5.
5. Install the instrument cap over the cell holder. Refer to Figure 6.
6. Push to set the instrument zero.
7. Remove the blank sample cell.
8. Prepare the sample. Refer to the test procedure.
9. Clean the sample cell with a no-lint cloth.
10. Insert the sample cell into the cell holder. Make sure to install the
sample cell in the correct and consistent orientation so that the results are
more repeatable and precise. Refer to Figure 5.
11. Install the instrument cap over the cell holder. Refer to Figure 6.
12. Push . The display shows the results in concentration units or
absorbance. Note: The result flashes if the result is less or more than the
instrument range.
13. Remove the sample cell from the cell holder.
14. Immediately empty and rinse the sample cell. Rinse the sample cell and
cap three times with deionized water (or distilled water). Note: As an
alternative, use tap water to rinse the sample cell if the samples measured
have a higher concentration than the tap water.
Figure 5 Sample cell orientation
1 Orientation mark4
2 Sample cell, 25-mm (10 mL), glass5
3 Sample cell, 1-cm (10 mL), plastic6
4 Some variants of the instrument have sample cells without an orientation
mark.
5 Use the glass sample cell for low-range chlorine tests. 6 Use the plastic
sample cell for high-range chlorine tests.
Figure 6 Install the instrument cap over the cell holder
6.1 Download a test procedure
- Go to http://www.hach.com.
- Enter “DR300” in the Search box.
- Select the “Downloads” option on the left side in the “Search Type” box.
- Scroll down to “Methods/Procedures”. 5. Click the link for the applicable test procedure to download it. 14 English
Section 7 Show measurements
Note: The instrument saves a maximum of 50 measurements. After 50 measurements are done, new measurements replace the oldest measurements.
- Push and hold for 3 seconds.
- Push until “rCL” (recall) shows, then push . “ 01 ” shows. Measurement 01 is the last measurement done.
- Push to scroll forward. The measurement number is followed by the measurement value and then the time.
- To go to a measurement number, push number shows, then push or . Note: Measurements cannot be deleted.until a measurement
- Push and hold for 3 seconds to go back to measurement mode.
Section 8 Calibration
This instrument is calibrated at the factory. No user calibration is necessary.
8.1 Standard calibration adjust
Use the standard calibration adjust (SCA) option when a calibration must be
adjusted to meet regulatory requirements. The factory calibration is adjusted
slightly with the standard calibration adjust (SCA) option so that the
instrument shows the expected value of the standard solution. The adjusted
calibration is then used for all test results. This adjustment can increase
the test accuracy when there are slight variations in the reagents or
instruments.
Note: For instruments with factory-calibrated ranges or methods, the
standard calibration adjust (SCA) feature is disabled when a user-entered
calibration is entered into the instrument. To set SCA back to on, set the
instrument to the factory default calibration. Refer to Set to the factory
default calibration on page 20.
8.1.1 Do a standard calibration adjust
-
Complete the test procedure for the range to calibrate. For the sample, use the standard solution concentration given in the test procedure documentation. Note: If a standard solution concentration is not given in the test procedure documentation, a different known standard can be used.
-
When the test procedure is completed, push and hold for 3 seconds.
-
Push until “SCA” shows, then push . The display shows the standard calibration adjust value.
-
If a different known standard is used, enter the value of the standard:
a. Push until “Edit” shows, then push . b. Enter the value of the standard.
Push the or to change the number that flashes. Push to go to the next digit. Push to go to the previous digit. -
Push to add the standard calibration adjust value to the factory calibration curve.
8.1.2 Set the standard calibration adjust to off
To use the factory default calibration again, set standard calibration adjust (SCA) to off.
- Push and hold for 3 seconds to enter menu mode.
- Push until “SCA” shows, then push.
- Push until “OFF” shows, then push.
Note: To set the SCA function to on again, do a standard calibration adjust.
8.2 User-entered calibration curve
This instrument accepts a user-prepared calibration curve. The calibration curve can be from 0 to 2.5 absorbance. Make sure that the calibration curve includes standard values that are less and more than the range of interest. The instrument range will be the same as the calibration range. For example, when the standards that are used are 1.00, 2.00 and 4.00. The instrument range is 1.00 to 4.00.
There are two options to enter a user calibration curve: · Enter a calibration
curve with standards–The standard solution
values are entered with the keypad and the absorbance values are measured. ·
Enter a calibration curve with the keypad–The standard solution values and
absorbance values are entered with the keypad. Note: If the instrument is set
to off or the instrument power is removed before a user-entered calibration
curve is completed, the calibration curve is not saved. The instrument
automatically switches off in user-entered calibration entry mode after 60
minutes of no activity. User-entered calibrations are completed when the user
goes out of calibration (cal) mode or edit mode.
8.2.1 Enter a calibration curve with standards
WARNING
Chemical exposure hazard. Obey laboratory safety procedures and wear all of
the personal protective equipment appropriate to the chemicals that are
handled. Refer to the current safety data sheets (MSDS/SDS) for safety
protocols.
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with
local, regional and national regulations.
Note: As an alternative, deionized water can be used for the blank unless the
sample is significantly more turbid or has more color than deionized water.
-
Push to set the instrument to the range to calibrate (e.g., LR or HR).
-
Prepare the blank. Refer to the test procedure.
-
Clean the sample cell with a no-lint cloth.
-
Set the instrument to zero.
a. Insert the blank sample cell in the cell holder.
b. Install the instrument cap over the cell holder.
c. Push . The display shows “- – – -“, then “0.00”. -
Push and hold for 3 seconds to enter menu mode.
-
Push until “USEr” shows, then push .
-
Push until “CAL” shows, then push .
-
When “S0” shows on the display, push .
-
Enter 00.00 (or 000.0) for the blank value. Push the or to change the number that flashes. Push to go to the next digit. Push to go to the previous digit.
-
When “A0” shows on the display, push to measure the absorbance of the blank. The display shows the absorbance value for “S0”.
-
Remove the sample cell from the cell holder.
-
Prepare the sample. Refer to the test procedure. For the sample, use the standard solution concentration given in the test procedure documentation.
-
Clean the sample cell with a no-lint cloth.
-
Push to show “S1” (or “Add”), then push .
-
Enter the concentration value of the first calibration standard, then push .
-
When “A1” shows on the display, do the steps that follow to measure the absorbance:
a. Insert the reacted standard sample cell in the cell holder. b. Install the instrument cap over the cell holder. c. Push . The display shows the absorbance value for “S1”. -
The calibration is completed with two calibration points. If additional standards are necessary for calibration: Do steps 11 16 again to measure more calibration standards.
-
Remove the sample cell from the cell holder.
-
Immediately empty and rinse the sample cell. Rinse the sample cell and cap three times with deionized water (or distilled water). Note: As an alternative, tap water can be used to rinse the sample cell if the concentration of the parameter in the tap water is less than the samples measured.
-
Push and hold for 3 seconds to go back to measurement mode.
8.2.2 Enter a calibration curve with the keypad
At least two data pairs are necessary to enter a user-prepared calibration curve. A concentration value and the absorbance value for the given concentration is necessary for each data pair. A maximum of 10 data pairs can be entered.
-
Push to set the instrument to the range to calibrate (e.g., LR or HR).
-
Push and hold for 3 seconds to enter menu mode.
-
Push until “USEr” shows, then push .
-
Push until “Edit” shows, then push .
-
When “S0” shows on the display, push .
-
Enter the first data pair.
The first data pair is S0 (concentration value) and A0 (absorbance value).
Push or to change the number that flashes. · Push to go to the next digit. · Push to go to the previous digit. -
Do steps 5 and 6 again to enter the second data pair (S1 and A1).
-
The calibration is completed with two data pairs. If additional data pairs are necessary for calibration: a. When “Add” shows, push . b. Do steps 5 and 6 again to enter more data pairs.
-
Push and hold for 3 seconds to go back to measurement mode.
8.2.3 Set to the factory default calibration
To remove a user-entered calibration curve from the instrument and use the factory calibration, do the steps that follow:
1. Push and hold for 3 seconds to enter menu mode.
2. Push until “USEr” shows, then push .
3. Push until “dFL” (default) shows, then push .
Section 9 Maintenance CAUTION
Multiple hazards. Only qualified personnel must conduct the tasks described in
this section of the document.
NOTICE
Do not disassemble the instrument for maintenance. If the internal components
must be cleaned or repaired, contact the manufacturer.
9.1 Clean the instrument
Clean the exterior of the instrument with a moist cloth and a mild soap
solution and then wipe the instrument dry as necessary. 9.2 Clean the sample
cells
CAUTION
Chemical exposure hazard. Obey laboratory safety procedures and wear all of
the personal protective equipment appropriate to the chemicals that are
handled. Refer to the current safety data sheets (MSDS/SDS) for safety
protocols.
CAUTION
Chemical exposure hazard. Dispose of chemicals and wastes in accordance with
local, regional and national regulations.
Most laboratory detergents are used at recommended concentrations. Neutral
detergents, such as Liquinox, are safer to use when regular cleaning is
necessary. To decrease the cleaning times, increase the temperature or use an
ultrasonic bath. To complete the cleaning, rinse a few times with deionized
water and then let the sample cell air dry. Sample cells may also be cleaned
with acid, followed by a thorough rinse with deionized water. Note: Always use
acid to clean sample cells that were used for low-level metal tests.
Special cleaning methods are necessary for individual procedures. When a brush
is used to clean sample cells, take extra care to avoid scratches on the
interior surfaces of the sample cells.
9.3 Replace the batteries
Replace the batteries when the battery power level is low. Refer to Install the batteries on page 7.
Section 10 Troubleshooting
Error
|
Description
|
Solution
---|---|---
E-00
|
No Zero
|
In user calibration mode, a standard solution was measured before the instrument zero was set. Measure a blank solution to set the instrument to zero.
E-01
|
Ambient light
error7
|
There is ambient light in the cell holder. Make sure that the instrument cap is fully installed on the cell holder. Refer to Do a test on page 11.
E-02
|
LED error7
|
The LED (light source) is out of
regulation. Replace the batteries. Make sure that the LED in the cell holder comes on when or is pushed.
7 When an E-01 or E-02 error occurs on a measurement, the display shows “_.__”. The decimal place depends on the chemistry. If the E-01 or E-02 error occurs while the instrument is set to zero, set the instrument to zero again.
English 21
Error
|
Description
|
Solution
---|---|---
E-03
|
Standard adjust error
|
• The measured value of the standard solution is more than the adjustment limits. Prepare a fresh standard.
• The standard solution is not within the concentration range that can be used for standard calibration adjust. Prepare a standard with a value at or near the recommended concentrations given in the procedure.
• Make sure that the concentration of the standard solution is entered correctly.
Reading flashes followed by
E-04
|
The reading is more or less than the instrument
range.8
|
If the reading is less than the instrument range, make sure that the instrument cap is fully installed on the cell holder. Measure a blank. If the blank reading is not zero, set the instrument to zero again.
If the reading is more than the instrument range, identify if there is a light blockage in the cell holder. Dilute the sample. Do the test again.
E-06
|
Absorbance error
|
The absorbance value is not correct or the user-entered calibration curve has fewer than two points. Enter or measure the absorbance value again.
E-07
|
Standard value error
|
The standard solution concentration is equal to another standard solution concentration that is already entered in the user-entered calibration curve. Enter the correct standard concentration.
E-09
|
Flash error
|
The instrument is not able to save data. Push and hold for 5 seconds to reset the instrument.
E-10
|
Environment
temperature too high or too low
|
The ambient temperature is out of range. Use the instrument only in the specified operating conditions. Refer to
Specifications on page 3.
8 The value that flashes will be 10% over the upper test range limit.
Error
|
Description
|
Solution
---|---|---
E-12
|
Low battery
power
|
Battery power is too low. Replace the batteries. Refer to Install the batteries on page 7.
E-13
|
Parameter load failure
|
The memory of the instrument is defective. Contact technical support.
E-14 followed by “_.__” or “0” if no zero was present
|
Zero
measurement
invalid
|
The zero measurement is too low. Use a sample cell filled with water and try again. If the error continues, contact technical support.
E-15 followed by “_.__”
|
Absorbance too high
|
Identify if there is a light blockage in the cell holder. Clean the cell holder. Dilute the sample. Do the test again.
Note: This instrument can not read absorbance values higher than 3.5 Abs.
E-20
|
Signal
measurement out of range
|
There is too much light on the light detector. Make sure that the instrument cap is fully installed on the cell holder. Do the test again. If the error continues, contact technical support.
E-21
|
Signal
measurement
unstable
|
There is an unstable signal on the light detector. There is too much or unstable ambient light. Make sure that the
instrument cap is fully installed on the cell holder. Do the test again. If the error continues, contact technical support.
E-22
|
Hardware error
|
The electronic system is defective. Contact technical support.
The following errors can occur immediately after an instrument update.
Error
|
Description
|
Solution
---|---|---
E-30
|
No application
|
There was an error during the application update. A valid application was not found on the instrument. Update the instrument again.
E31
|
Bootloader update failed
|
There was an error during the transmission of the bootloader update. Update the bootloader again.
Error
|
Description
|
Solution
---|---|---
E-32
|
Application update failed
|
There was an error during the transmission of the application update. Update the instrument again.
E-66
|
Update failed
|
The instrument is defective. Contact technical support.
Section 11 Replacement parts and accessories
W A R N I N G
|
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the instrument or equipment malfunction. The replacement parts in this section are approved by the manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to the company website for contact information.
Replacement parts
Description Quantity Item no.
AAA batteries, alkaline 4/pkg 4674300 Instrument cap 1 LPZ445.99.00006 Battery cover 1 LPZ445.99.00007 Sample cell, 25 mm (10 mL), glass 6/pkg 2427606 Sample cell, 1 cm (10 mL), plastic 2/pkg 4864302
Accessories
Description Quantity Item no.
Hach Communication Dongle 1 LPV446.99.00012 Soft-sided case/holster 1 5953100
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>