Laserliner 082.046A CondenseSpot Plus Infrared Thermometer Instruction
Manual
Laser liner
Completely read through the operating instructions, the „Warranty and
Additional Information“booklet as well as the latest information under the
internet link at the end of these instructions.
Follow the instructions they contain. This document must be kept in a safe
place and if the laser device is passed on, this document must be passed on
with it.
Function/Application
The Condense Spot Plus is an infrared temperature measuring device with an integrated hygrometer which permits non-contact temperature measurement of surfaces and calculation of dew point temperature. The instrument measures how much electromagnetic energy is emitted in the infrared wavelength range and uses this information to calculate surface temperature. With the integrated sensors, the device detects thermal bridges as well as condensation moisture.
General safety instructions
– The device must only be used in accordance with its intended purpose and
within the scope of the specifications.
– The measuring tools and accessories are not toys. Keep out of reach of
children.
– Modifications or changes to the device are not permitted, this will
otherwise invalidate the approval and safety specifications.
– Do not expose the device to mechanical stress, extreme temperatures,
moisture or significant vibration.
– The device must no longer be used if one or more of its functions fail or
the battery charge is weak.
– Please ensure compliance with the safety regulations set out by local and
national authorities with regard to the correct and proper use of the device.
Safety instructions
Using class 2 lasers
– Attention: Do not look into the direct or reflected beam.
– Do not point the laser beam towards persons.
– If a person‘s eyes are exposed to class 2 laser radiation, they should shut
their eyes and immediately move away from the beam.
– Tampering with (making changes to) the laser device is not permitted.
– Under no circumstances should optical instruments (magnifying glass,
microscope, binoculars)
Laser outlet
Safety instructions
Dealing with electromagnetic radiation
– The measuring device complies with electromagnetic compatibility regulations
and limits in accordance with the EMC Directive 2014/30/EU.
– Local operating restrictions – for example, in hospitals, aircraft, petrol
stations or in the vicinity of people with pacemakers – may apply. Electronic
devices can potentially cause hazards or interference or be subject to hazards
or interference.
– The measuring accuracy may be affected when working close to high voltages
or high electromagnetic alternating fields.
- LC display
- Set emissivity factor
- Mode setting: dp /
- HEAT Humidity / ambient temperature sensor
- Infrared sensor
- 8-point laser circle outlet
- SET button / dp / rH / T-A selection
- ON / Trigger
- Battery compartment
a Hold function
b Bar graph, condensation moisture indicator
c Dew point temperature in °C
d Battery charge
e Measurement °C
f Measured value display
g Quick emissivity display
h Dew point mode (dp) with relative humidity (rH) and ambient temperature
(T-A) display
i Laser beam switched on, temperature measurement (infrared)
j Thermal bridge mode (HEAT)
k Active measurement indicator
I LOW, CHK, HI display in thermal bridge mode.
Laserliner
1. Inserting batteries
Open the battery compartment and insert batteries according to the symbols. Be
sure to pay attention to polarity.
2.ON / OFF
Auto switch-off after 30 seconds.
Make sure that the humidity/ambient temperature sensor (4) is folded in during
transport
3.Continuous measurement / Hold
For continuous measurement activate the laser (see figure) and keep the button pressed.
Release the button as soon as the target laser pinpoints the measurement
location.
The measured value is held.
4 Setting the emissivity
An integrated sensor head detects the material/surface-specific infrared
radiation emitted by all objects.
The level of these emissions is determined by the specific emissivity of the
material (0.01 to 1.00). After switching on for the first time, the device is
preset to an emissivity factor of 0.95, which is suitable for
most organic materials and non-metals (plastics, ceramics, wood, rubber,
paints, varnishes and stone). For materials with different emissivity factors,
refer to the table under Point 9.
Wherever possible, paint or matt black stickers can be applied on non-coated
metals and metal oxides, which due to their low, temperature-unstable
emissivity are suitable for IR measurement only for a limited extent, as well
as on surfaces with unknown emissivity in order to set the emissivity factor
to 0.95. If this s not possible, use a contact thermometer.
The emissivity last selected is set after the device is switched on. Check the
emissivity setting ! before each measurement.
The device features a quick selection function for the stored emissivity factors (0.95, 0.85, 0.75, 0.65, 0.55) and accurate setting between 0.01 – 1.00.
Quick emissivity factor selection
Accurate emissivity factor setting
5. Mode selection
Laserliner
6 Dew point mode/condensation moisture indicator
The dew point temperature is the point the temperature must drop below for the air to precipitate its water vapour in the form of droplets, mist or dew. Condensation water therefore occurs for example when the temperature of an inside wall or window reveal is lower than the dew point temperature of the room. These areas are then damp and are a breeding ground for mildew/mould with possible material damage.
The CondenseSpot Plus calculates the dew point with the integrated sensors for
ambient temperature and relative humidity. At the same time, the surface
temperature of objects is determined with the aid of
infrared temperature measurement. By comparing these temperature it is
possible to identify areas that are exposed to the risk of condensation
moisture. The condensation moisture indicator (b) displays the
result in the form of a bar graph, supported by visual and acoustic signals
indicating the high probability of condensation moisture occurring.
No risk of condensation moisture
Slight risk of condensation moisture „dP“ symbol flashing
The condensation moisture indicator (b) is shown in all modes.
The device therefore constantly shows the risk level of condensation moisture.
The measured values for relative humidity and ambient temperature can be
called up:
7. Thermal bridge mode
In relation to buildings, the term thermal bridge refers to an area, e.g. an inside wall, at which heat is transferred outwards faster than at the rest of the inside wall. When viewed from the inside, the temperature of this area is lower than the surrounding areas and when viewed from outside higher than the surrounding areas. This often indicates faulty or inadequate insulation.
For this purpose, the Condense Spot Plus compares the ambient temperature with the surface temperature. If there are large differences in both temperatures, the device will trigger warnings in 2 stages: „CHK“ is displayed in borderline cases or the display lighting changes to „blue“ or „red“ to indicate the differences are very large.
Ambient temperature:
20°C No thermal bridge
Ambient temperature:
20°C Possible thermal bridge, continue checking area
Ambient temperature:
20°C Thermal bridge, display lights blue with acoustic signal
Ambient temperature:
12°C Thermal bridge, display lights red with acoustic signal
8. Menu settings
9. Emission coefficient table
Information on maintenance and care
Clean all components with a damp cloth and do not use cleaning agents, scouring agents and solvents. Remove the battery(ies) before storing for longer periods. Store the device in a clean and dry place.
Calibration
The measuring device should be calibrated and tested on a regular basis to ensure it is accurate and working properly. We recommend the measuring device is calibrated every 1 – 2 years. If necessary, contact your distributor or the UMAREX-LASERLINER service department.
Technische Daten (Technische Änderungen vorbehalten. 21W28)
Messgröße
| Infrarottemperatur Luftfeuchte Umgebungstemperatur
Modus| Taupunkt Wärmebrücke
Messbereich Umgebungstemperatur| -20°C … 65°C
Genauigkeit Umgebungstemperatur| 0°C … 50°C (± 1°C)
<0°C und >50°C (± 2,5°C)
Auflösung Umgebungstemperatur| 0,1°C
Messbereich Infrarottemperatur| -40°C … 365°C
Genauigkeit Infrarottemperatur
| -40°C … 0°C (± (1°C + 0,1°C/1°C))
0°C … 30°C (± 1°C oder ± 1%, je nach größerem Wert)
30°C (± 2°C oder ± 2%, je nach größerem Wert)
Auflösung Infrarottemperatur| 0,1°C
Messbereich Luftfeuchte| 1% … 99%
Genauigkeit (absolut) Luftfeuchte (relativ)| 20% … 80% (± 3%)
<20% und >80% (± 5%)
Auflösung Luftfeuchte (relativ)| 0,1%
Messbereich Taupunkttemperatur| -50°C … 50°C
Genauigkeit Taupunkttemperatur
| 20% rH … 30% rH (± 2,5°C)
31% rH … 40% rH (± 2°C)
41% rH … 95% rH (± 1,5°C)
Auflösung Taupunkttemperatur| 0,1°C
Laserwellenlänge| 650 nm
Laserklasse| 2 / < 1 mW (EN 60825-1:2014/AC:2017)
Automatische Abschaltung| nach 30 Sekunden
Emissionsgrad| einstellbar, 0,01 … 1,00
Optik| 12:1 (12 m Messentfernung : 1 m Messfläche)
Stromversorgung| 2 x 1,5V LR6 (AA)
Betriebsdauer| ca. 20 Std.
Arbeitsbedingungen| 0°C … 50°C, Luftfeuchtigkeit max. 80% rH, nicht
kondensierend, Arbeitshöhe max. 2000 m über NN (Normalnull)
Lagerbedingungen| -10°C … 60°C, Luftfeuchtigkeit max. 80% rH, nicht
kondensierend
Abmessungen (B x H x T)| 60 x 205 x 150 mm
Gewicht| 384 g (inkl. Batterien)
EU directives and disposal
This device complies with all necessary standards for the free movement of
goods within the EU.
This product is an electric device and must be collected separately for
disposal according to the European Directive on waste electrical and
electronic equipment.
Further safety and supplementary notices at:
http://laserliner.com/info?an=AFV
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>