HAMILTON MEDICAL G5 Mechanical Ventilator User Manual

June 6, 2024
HAMILTON MEDICAL

G5 Mechanical Ventilator

.
HAMILTON-G5

Operator’s Manual
REF 159001, 159002, 159003
Software version 2.8x 624074/13 | 2020-08-14
0 1 9 7

Operator’s Manual
HAMILTON-G5
2020-08-14 624074/13
Be sure to read the Addendum/Errata for the Operator’s Manual, included at the end of this document. The Addendum/Errata is
also available at www.hamilton-medical.com, in MyHamilton.

© 2020 Hamilton Medical AG. All rights reserved. Printed in Switzerland. No part of this publication may be reproduced, stored in a database or retrieval system, or transmitted in any form or by any means, electronic, mechanical, or by photocopying, recording, or otherwise, without prior written permission of Hamilton Medical AG. This document may be revised, replaced, or made obsolete by other documents by Hamilton Medical AG at any time and without notice. Ensure that you have the most current applicable version of this document; if in doubt, contact the technical support department of Hamilton Medical AG, Switzerland. While the information set forth herein is believed to be accurate, it is not a substitute for the exercise of professional judgment. Nothing in this document shall limit or restrict in any way Hamilton Medical AG’s right to revise or otherwise change or modify the equipment (including its software) described herein, without notice. In the absence of an express, written agreement to the contrary, Hamilton Medical AG has no obligation to furnish any such revisions, changes, or modifications to the owner or user of the equipment (including software) described herein. The equipment must be operated, serviced, or upgraded only by trained professionals. Hamilton Medical AG’s sole responsibility with respect to the equipment and its use is as stated in the limited warranty provided in the device Operator’s Manual. Hamilton Medical AG shall not be liable for any loss, cost, expense, inconvenience, or damage that may arise out of misuse of the product, or if non-Hamilton Medical AG parts were used when replacing parts, or if serial numbers were amended, deleted, or removed.

If returning parts to Hamilton Medical AG, be sure to use the standard Hamilton Medical returned goods authorization (RGA) procedure. Disposal of parts shall follow all local, state, and federal regulation with respect to environmental protection.
Hamilton Medical AG will make available, on request, circuit diagrams, component parts lists, descriptions, calibration instructions, or other information that will assist appropriately trained personnel to repair those parts of the equipment designated by Hamilton Medical AG to be repairable.
For all proprietary as well as third-party trademarks used by Hamilton Medical AG, see www.hamilton-medical.com/trademarks. Product and/or company names marked with a § symbol may be the trademarks and/or registered trademarks of their respective owners, including but not limited to Aerogen§, Nihon Kohden§, Masimo§, Masimo SET§, Masimo rainbow SET§, Respironics§.
Manufacturer
Hamilton Medical AG Via Crusch 8, CH-7402 Bonaduz, Switzerland Phone: (+41) 58 610 10 20 Fax: (+41) 58 610 00 20 info@hamilton-medical.com www.hamilton- medical.com
Distributor in USA
Hamilton Medical, Inc. 4990 Energy Way, P.O. Box 30008 Reno, NV 89520 Phone: 775-858-3200 Toll-free: 800-426-6331 Fax: 775-856-5621 marketing@hamilton- medical.net

4

English | 624074/13

Table of Contents

Table of Contents
Preface ……………………………………………………………… 17 Chapter 1 Safety information ………………………………………………. 19
1.1 Overview……………………………………………………………………………. 20
1.2 Electromagnetic susceptibility ………………………………………………… 20
1.3 Fire and other hazards………………………………………………………….. 21
1.4 General operation and setup…………………………………………………. 21 1.4.1 General operation and setup………………………………………….. 21 1.4.2 Electrical: power and batteries………………………………………… 22 1.4.3 Gas supply ………………………………………………………………….. 23 1.4.4 CompactFlash port……………………………………………………….. 24
1.5 Setting up for ventilation………………………………………………………. 24 1.5.1 Patient breathing circuits, components, and accessories ……… 24 1.5.2 Preoperational check and tests ……………………………………….. 25 1.5.3 Humidifier…………………………………………………………………… 26 1.5.4 IntelliCuff……………………………………………………………………. 26 1.5.5 CO2 sensor setup and operation …………………………………….. 27 1.5.6 Nebulization………………………………………………………………… 29
1.6 Ventilating the patient …………………………………………………………. 29 1.6.1 Specifying patient settings……………………………………………… 29 1.6.2 Neonatal ventilation ……………………………………………………… 30 1.6.3 Apnea backup……………………………………………………………… 30 1.6.4 TRC settings………………………………………………………………… 30 1.6.5 Noninvasive ventilation………………………………………………….. 31 1.6.6 P/V Tool Pro ………………………………………………………………… 31 1.6.7 Using high flow oxygen therapy ……………………………………… 31
1.7 Monitoring and alarms…………………………………………………………. 31
1.8 Using the trolley………………………………………………………………….. 32
1.9 Maintenance………………………………………………………………………. 32 1.9.1 General maintenance, cleaning, and disinfection……………….. 32

Hamilton Medical | HAMILTON-G5 Operator’s Manual

5

Table of Contents

1.9.2 Preventive maintenance ………………………………………………… 34 1.9.3 O2 sensor …………………………………………………………………… 34
1.10 Service and testing ………………………………………………………………. 34
Chapter 2 System overview …………………………………………………. 35
2.1 Overview……………………………………………………………………………. 36 2.1.1 Standard features and options ……………………………………….. 36
2.2 Physical descriptions…………………………………………………………….. 39 2.2.1 About the ventilator……………………………………………………… 39 2.2.2 About the main display …………………………………………………. 44 2.2.3 About the patient breathing circuits ………………………………… 46 2.2.4 About the trolley and mounting variations ……………………….. 51 2.2.5 Setting up the monitor………………………………………………….. 52
2.3 Navigating the windows and controls……………………………………… 53 2.3.1 Accessing windows ………………………………………………………. 53 2.3.2 Adjusting controls ………………………………………………………… 54 2.3.3 Selecting list items………………………………………………………… 54
Chapter 3 Preparing the ventilator………………………………………… 55
3.1 Overview……………………………………………………………………………. 56
3.2 Connecting to a power source ………………………………………………. 56 3.2.1 Using battery power……………………………………………………… 56
3.3 Connecting the oxygen supply ………………………………………………. 57 3.3.1 Working with heliox as a gas source………………………………… 57 3.3.2 Selecting the gas source type …………………………………………. 58
3.4 Setting up the patient breathing circuit …………………………………… 58 3.4.1 Breathing circuit connections on the ventilator ………………….. 59 3.4.2 Working with the expiratory valve set………………………………. 59 3.4.3 Selecting the breathing circuit components ………………………. 60 3.4.4 Assembling the patient breathing circuit ………………………….. 61 3.4.5 Positioning the breathing circuit ……………………………………… 61

6

English | 624074/13

Table of Contents

3.5 Setting up esophageal/transpulmonary pressure monitoring ……….. 62 3.6 Turning the ventilator on and off …………………………………………… 62
Chapter 4 Setting up external devices and sensors…………………… 63
4.1 Overview……………………………………………………………………………. 64 4.2 Installing a module………………………………………………………………. 64 4.3 Setting up a humidifier…………………………………………………………. 64 4.4 Setting up the IntelliCuff cuff pressure controller………………………. 65
4.4.1 About the IntelliCuff tubing …………………………………………… 66 4.4.2 Setting up IntelliCuff …………………………………………………….. 66 4.5 Setting up CO2 monitoring …………………………………………………… 66 4.5.1 Mainstream CO2 measurement………………………………………. 67 4.5.2 Sidestream CO2 measurement ……………………………………….. 69 4.6 Setting up SpO2 monitoring………………………………………………….. 70 4.7 Enabling sensors …………………………………………………………………. 70 4.8 Setting up nebulization ………………………………………………………… 71 4.8.1 Setting up a pneumatic nebulizer ……………………………………. 71 4.8.2 Setting up an Aerogen nebulizer …………………………………….. 72 4.9 Connecting to an external patient monitor or other device …………. 72
Chapter 5 Specifying ventilation settings ……………………………….. 73
5.1 Process overview …………………………………………………………………. 74 5.2 Selecting the patient group …………………………………………………… 74
5.2.1 About system defaults: pre-configured settings …………………. 75 5.3 Entering patient data …………………………………………………………… 75 5.4 Performing the preoperational check, tests, and calibrations ………. 76
5.4.1 Performing the preoperational check……………………………….. 77 5.4.2 Performing the breathing circuit tightness test ………………….. 78 5.4.3 Calibrating the adult/pediatric flow sensor ……………………….. 79 5.4.4 Calibrating the O2 sensor………………………………………………. 80 5.4.5 Performing a zero calibration on the CO2 sensor/adapter……. 81

Hamilton Medical | HAMILTON-G5 Operator’s Manual

7

Table of Contents

5.4.6 Testing the alarms………………………………………………………… 82 5.5 Selecting the ventilation mode ………………………………………………. 84
5.5.1 Reviewing and adjusting ventilation settings……………………… 85 5.5.2 About the trigger types …………………………………………………. 86 5.5.3 About apnea backup ventilation……………………………………… 88 5.5.4 About tube resistance compensation (TRC) ………………………. 89 5.6 Setting alarm limits ……………………………………………………………… 91 5.7 Starting ventilation………………………………………………………………. 93 5.8 Stopping ventilation …………………………………………………………….. 93 5.9 About the control parameters ……………………………………………….. 93
Chapter 6 Specifying neonatal settings …………………………………..101
6.1 Setting up for neonatal ventilation …………………………………………. 102 6.1.1 Setting the patient group and weight………………………………. 102 6.1.2 Setting up the patient breathing circuit ……………………………. 103
6.2 Performing the preoperational check, tests, and calibrations ………. 105 6.2.1 Calibrating the neonatal flow sensor ……………………………….. 106
6.3 Selecting the ventilation mode ………………………………………………. 107 6.4 Setting the patient weight for ventilation ………………………………… 107 6.5 Alarms for neonatal ventilation ……………………………………………… 108 6.6 O2 enrichment for neonates …………………………………………………. 108 6.7 Specifying volume limitation for neonates ……………………………….. 108
Chapter 7 Ventilation modes ………………………………………………..109
7.1 Overview……………………………………………………………………………. 110 7.1.1 Breath types and timing options……………………………………… 110 7.1.2 Ventilation modes ………………………………………………………… 110 7.1.3 Ventilation controls and settings …………………………………….. 112
7.2 Volume-controlled modes, flow control…………………………………… 114 7.2.1 (S)CMV mode ……………………………………………………………… 114 7.2.2 SIMV mode …………………………………………………………………. 115

8

English | 624074/13

Table of Contents
7.3 Volume-controlled modes, flow cycled ……………………………………. 116 7.3.1 Volume Support (VS)…………………………………………………….. 116
7.4 Volume-targeted modes, adaptive pressure control …………………… 118 7.4.1 APVcmv mode …………………………………………………………….. 118 7.4.2 APVsimv mode…………………………………………………………….. 120
7.5 Pressure-controlled modes ……………………………………………………. 121 7.5.1 P-CMV mode ………………………………………………………………. 121 7.5.2 P-SIMV mode ………………………………………………………………. 122 7.5.3 DuoPAP mode……………………………………………………………… 123 7.5.4 APRV mode…………………………………………………………………. 123 7.5.5 SPONT mode……………………………………………………………….. 125
7.6 Intelligent Ventilation …………………………………………………………… 126 7.6.1 ASV mode…………………………………………………………………… 126 7.6.2 INTELLiVENT-ASV mode ………………………………………………… 128
7.7 Noninvasive modes ……………………………………………………………… 128 7.7.1 NIV mode……………………………………………………………………. 129 7.7.2 NIV-ST mode……………………………………………………………….. 130 7.7.3 nCPAP-PS mode…………………………………………………………… 131 7.7.4 High flow oxygen therapy ……………………………………………… 132
7.8 Ambient state …………………………………………………………………….. 133 7.9 Working with noninvasive modes…………………………………………… 134
7.9.1 Required conditions for use……………………………………………. 134 7.9.2 Contraindications…………………………………………………………. 134 7.9.3 Potential adverse reactions …………………………………………….. 135 7.9.4 Control settings in noninvasive ventilation………………………… 135 7.9.5 Alarms in noninvasive ventilation…………………………………….. 135 7.9.6 Monitored parameters in noninvasive ventilation……………….. 136 7.9.7 Additional notes about using noninvasive ventilation………….. 136

Hamilton Medical | HAMILTON-G5 Operator’s Manual

9

Table of Contents
7.10 Working with ASV ………………………………………………………………. 137 7.10.1 Clinical workflow with ASV ……………………………………………. 137 7.10.2 Maintaining adequate ventilation ……………………………………. 139 7.10.3 Reviewing alarm settings……………………………………………….. 139 7.10.4 Monitoring ASV …………………………………………………………… 140 7.10.5 Weaning …………………………………………………………………….. 141 7.10.6 Functional overview ……………………………………………………… 142
Chapter 8 Monitoring ventilation ………………………………………….149
8.1 Overview……………………………………………………………………………. 150 8.2 Viewing numeric patient data ……………………………………………….. 150
8.2.1 About the main monitoring parameters (MMP) …………………. 150 8.2.2 About the secondary monitoring parameters (SMPs) ………….. 151 8.2.3 Viewing patient data in the Monitoring window ……………….. 152 8.3 Viewing graphical patient data………………………………………………. 152 8.3.1 Selecting a display layout ………………………………………………. 153 8.3.2 Selecting display options ……………………………………………….. 154 8.3.3 Working with waveforms ………………………………………………. 155 8.3.4 Working with Trend graphs……………………………………………. 159 8.3.5 Working with loops………………………………………………………. 160 8.4 Working with Intelligent panels……………………………………………… 161 8.4.1 Dynamic Lung panel: real-time ventilation status ……………….. 161 8.4.2 Vent Status panel: real-time ventilator dependence status …… 164 8.4.3 ASV Graph panel: real-time patient condition and targets …… 166 8.4.4 ASV Monitoring panel: real-time values……………………………. 167 8.5 Monitoring transpulmonary/esophageal pressure ……………………… 167 8.6 About the monitored parameters …………………………………………… 168 8.7 Viewing patient ventilation time …………………………………………….. 178 8.8 Viewing device-specific information ……………………………………….. 178

10

English | 624074/13

Table of Contents

Chapter 9 Responding to alarms …………………………………………..179
9.1 Overview……………………………………………………………………………. 180 9.1.1 Alarm limit indicators ……………………………………………………. 182 9.1.2 Responding to an alarm ………………………………………………… 182 9.1.3 Temporarily silencing an alarm ……………………………………….. 183
9.2 About the alarm buffer ………………………………………………………… 183 9.3 Adjusting alarm loudness (volume) …………………………………………. 185 9.4 Troubleshooting alarms………………………………………………………… 185 9.5 Working with a distributed alarm system (DAS) ………………………… 201
9.5.1 Enabling global AUDIO OFF……………………………………………. 202 9.5.2 About DAS-related alarms……………………………………………… 202
Chapter 10 Ventilation settings and functions …………………………..205
10.1 Overview……………………………………………………………………………. 206 10.2 Accessing settings during ventilation ………………………………………. 206
10.2.1 Accessing patient data during ventilation …………………………. 206 10.2.2 Accessing settings during ventilation ……………………………….. 207 10.3 Entering/exiting Standby ………………………………………………………. 208 10.4 Oxygen enrichment……………………………………………………………… 209 10.4.1 Adjusting the oxygenation level for O2 enrichment ……………. 209 10.4.2 Suctioning maneuver ……………………………………………………. 210 10.5 Manual breath ……………………………………………………………………. 210 10.6 Inspiratory and expiratory hold ………………………………………………. 211 10.6.1 Inspiratory hold ……………………………………………………………. 211 10.6.2 Expiratory hold…………………………………………………………….. 211 10.7 Working with a nebulizer ……………………………………………………… 212 10.7.1 Specifying duration and synchronization settings……………….. 212 10.7.2 Working with a pneumatic nebulizer……………………………….. 213 10.7.3 Working with an Aerogen nebulizer………………………………… 214 10.8 Locking and unlocking the touch screen………………………………….. 214

Hamilton Medical | HAMILTON-G5 Operator’s Manual

11

Table of Contents
10.9 Capturing a screenshot ………………………………………………………… 215 10.10 About the Event log …………………………………………………………….. 215 10.11 Setting display options …………………………………………………………. 216
10.11.1 Setting date and time……………………………………………………. 216 10.11.2 Day and night display brightness …………………………………….. 217
Chapter 11 Working with P/V Tool ………………………………………….219
11.1 Overview……………………………………………………………………………. 220 11.1.1 Conditions for use………………………………………………………… 220 11.1.2 Indications for use………………………………………………………… 221 11.1.3 Contraindications for use ………………………………………………. 221
11.2 Using the P/V Tool……………………………………………………………….. 221 11.3 Opening the P/V Tool …………………………………………………………… 222 11.4 Adjusting the control settings………………………………………………… 223 11.5 Performing a P/V Tool maneuver ……………………………………………. 224
11.5.1 Viewing data……………………………………………………………….. 224 11.5.2 Choosing the data to display………………………………………….. 225 11.6 Analyzing the data ………………………………………………………………. 227 11.7 Using reference curves …………………………………………………………. 227 11.8 Performing a recruitment maneuver ……………………………………….. 228
Chapter 12 Working with external devices………………………………..229
12.1 Working with the HAMILTON-H900 humidifier ………………………… 230 12.1.1 Accessing humidifier controls on the ventilator………………….. 231 12.1.2 About the humidifier operating modes…………………………….. 232 12.1.3 Changing humidity using temperature controls …………………. 234 12.1.4 Entering Standby………………………………………………………….. 235 12.1.5 Turning the humidifier on/off …………………………………………. 235 12.1.6 About humidifier-related alarms……………………………………… 236 12.1.7 About humidifier-related parameters……………………………….. 239

12

English | 624074/13

Table of Contents

12.2 Working with IntelliCuff……………………………………………………….. 239 12.2.1 Accessing IntelliCuff controls on the ventilator ………………….. 240 12.2.2 Turning IntelliCuff on and off…………………………………………. 241 12.2.3 About IntelliCuff modes ………………………………………………… 242 12.2.4 Setting the cuff pressure ……………………………………………….. 243 12.2.5 Deflating the cuff…………………………………………………………. 244 12.2.6 About IntelliCuff-related alarms ……………………………………… 244 12.2.7 About IntelliCuff- related parameters ……………………………….. 246 12.2.8 Last Patient settings with IntelliCuff…………………………………. 246
Chapter 13 Maintenance……………………………………………………….247
13.1 Overview……………………………………………………………………………. 248
13.2 Cleaning, disinfection, and sterilization …………………………………… 248 13.2.1 General guidelines for cleaning ………………………………………. 248 13.2.2 General guidelines for disinfection ………………………………….. 249
13.3 Preventive maintenance ……………………………………………………….. 250
13.4 Performing maintenance tasks ………………………………………………. 252 13.4.1 Maintaining the filters …………………………………………………… 252 13.4.2 Replacing the galvanic O2 sensor ……………………………………. 253 13.4.3 Charging and storing batteries ……………………………………….. 254 13.4.4 Replacing batteries……………………………………………………….. 254
13.5 Repacking and shipping ……………………………………………………….. 255
Chapter 14 Configuration ……………………………………………………..257
14.1 Overview……………………………………………………………………………. 258
14.2 Accessing Configuration mode………………………………………………. 258
14.3 Configuring general settings …………………………………………………. 258 14.3.1 Selecting the language………………………………………………….. 258 14.3.2 Selecting the breath timing philosophy…………………………….. 258 14.3.3 Selecting the units of measure ……………………………………….. 258 14.3.4 Configuring adjustable alarms………………………………………… 259

Hamilton Medical | HAMILTON-G5 Operator’s Manual

13

Table of Contents

14.3.5 Setting the minimum alarm loudness (volume) ………………….. 259 14.3.6 Enabling the Check flow sensor for water alarm………………… 260 14.4 Configuring MMPs………………………………………………………………. 260 14.5 Configuring Vent Status settings ……………………………………………. 260 14.6 Configuring communication options ………………………………………. 261 14.6.1 Configuring I:E timing …………………………………………………… 261 14.6.2 Selecting a communication protocol………………………………… 261 14.6.3 Configuring HAMILTON-H900 humidifier communication …… 262 14.6.4 Configuring distributed alarm system (DAS) communication… 262 14.7 Configuring nebulization options …………………………………………… 262 14.8 Activating IntelliCuff ……………………………………………………………. 262 14.9 Activating SpO2 and CO2 measurement …………………………………. 262 14.10 Defining system default settings …………………………………………….. 263 14.10.1 Exporting or importing default settings ……………………………. 264 14.10.2 Choosing the ASV version ……………………………………………… 264 14.10.3 Enabling the display of resistance- and compliance-related
parameters………………………………………………………………….. 264 14.11 Configuring software and hardware options ……………………………. 265
14.11.1 Reviewing installed options ……………………………………………. 265 14.11.2 Adding software options……………………………………………….. 265 14.11.3 Activating hardware options ………………………………………….. 265 14.12 Copying configuration settings………………………………………………. 266
Chapter 15 Parts and accessories…………………………………………….267
15.1 Overview……………………………………………………………………………. 268
Chapter 16 Specifications ………………………………………………………277
16.1 Physical characteristics………………………………………………………….. 278 16.2 Environmental requirements………………………………………………….. 279 16.3 Pneumatic specifications ………………………………………………………. 280 16.4 Electrical specifications …………………………………………………………. 281

14

English | 624074/13

Table of Contents
16.5 Control settings ………………………………………………………………….. 283 16.6 Monitored parameters …………………………………………………………. 288 16.7 Alarms ………………………………………………………………………………. 295 16.8 Configuration …………………………………………………………………….. 297 16.9 ASV technical data ………………………………………………………………. 300 16.10 Ventilator breathing system specifications ……………………………….. 302 16.11 Technical performance data ………………………………………………….. 303
16.11.1 Accuracy testing…………………………………………………………… 309 16.11.2 Essential performance …………………………………………………… 309 16.12 Functional description of ventilator system ………………………………. 309 16.12.1 Gas supply and delivery…………………………………………………. 310 16.12.2 Gas monitoring with the flow sensor……………………………….. 311 16.12.3 Pneumatic diagram ………………………………………………………. 312 16.13 Symbols used on device labels and packaging ………………………….. 313 16.13.1 Symbols used on the trolley……………………………………………. 315 16.14 Standards and approvals ………………………………………………………. 315 16.15 Disposal and year of manufacture ………………………………………….. 316 16.16 Warranty……………………………………………………………………………. 317
Glossary ……………………………………………………………..319 Index …………………………………………………………………327 Addendum HiFlow ………………………………………….. Add-1 Addendum HAMILTON-G5/S1……………………………. Add-5

Hamilton Medical | HAMILTON-G5 Operator’s Manual

15

Table of Contents

16

English | 624074/13

Preface

HAMILTON-G5 Documentation This guide is part of a documentation suite that includes, among others, the following documents:
Table 1. HAMILTON-G5 documentation suite

Document title

Description

Operator’s Manual (this guide)

Provides detailed information about the setup and use of the HAMILTON-G5 ventilator.

INTELLiVENT-ASV Operator’s Manual

Provides setup and use information for the INTELLiVENT-ASV ventilation mode.

Pulse Oximetry Instructions for Use

Provides setup and use information for using SpO2 and related sensors with the ventilator.

Volumetric Capnography User Guide Provides reference information for CO2 capnography.

HAMILTON-H900 Instructions for Use Provides specifications, and setup and use information for the HAMILTON-H900 humidifier.

IntelliCuff Instructions for Use

Provides specifications, and setup and use information for the IntelliCuff cuff pressure controller.

Aerogen Solo/Aerogen Pro Instructions Provides specifications, and setup and use information

for Use

for the Aerogen Solo and Aerogen Pro nebulizers.

Communication Interface User Guide

Provides an overview of the communication interface, including how to connect the ventilator to external devices for data communication and support for nurse call remote alarms.

Service Manual

Provides information about installing and setting up the medical equipment, as well as additional technical and servicing information for the ventilator.

EMC Declarations Guide

Provides emissions and EMC-related safety and use information.

Be sure to read the documentation before using the device or accessories.
To download the latest version of this manual or other documents, free of charge, visit the MyHamilton website. To register, go to: https://www.hamiltonmedical.com/MyHamilton

Hamilton Medical offers the Hamilton Medical College, which provides a variety of learning modules free of charge. To register, go to: http://college.hamiltonmedical.com

Hamilton Medical | HAMILTON-G5 Operator’s Manual

17

Preface

Conventions used in this guide In this manual: · Button and tab names are shown in a
bold font. · The notation XX > XX shows the
sequence of buttons/tabs to touch to open the associated window. For example, the text Open the System > Settings window means touch the System button, then touch the Settings tab. · Software version: The software version for the ventilator is displayed in the System > Info window and should match the version on the title page of this manual. · Units of measure: Pressure is indicated in cmH2O, length in cm, and temperature in degrees Celsius (°C). The unit of measure for length is configurable. · The graphics shown in this manual may not exactly match what you see in your environment. Safety messages are displayed as follows:
WARNING A WARNING alerts the user to the possibility of injury, death, or other serious adverse reactions associated with the use or misuse of the device.
CAUTION A CAUTION alerts the user to the possibility of a problem with the device associated with its use or misuse, such as device malfunction, device failure, damage to the device, or damage to other property.

NOTICE A NOTICE emphasizes information of particular importance. In tables, safety messages are indicated as follows:
WARNING!
CAUTION!
NOTICE!
Intended use The HAMILTON-G5 ventilator is designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients. The device is intended for use in the hospital and institutional environment where health care professionals provide patient care. The HAMILTON-G5 ventilator is intended for use by properly trained personnel under the direct supervision of a licensed physician. The HAMILTON-G5 ventilator may be used for transport within a hospital or hospital type facility provided compressed gas is supplied. The device is not to be used in the presence of flammable anesthetic agents or other ignition sources. The ventilator is not to be used in an environment with magnetic resonance imaging (MRI) equipment. The device is not intended for transportation outside the hospital or for use in the home environment.

18

English | 624074/13

1
Safety information
1.1 Overview…………………………………………………………………………….. 20 1.2 Electromagnetic susceptibility …………………………………………………. 20 1.3 Fire and other hazards…………………………………………………………… 21 1.4 General operation and setup………………………………………………….. 21 1.5 Setting up for ventilation……………………………………………………….. 24 1.6 Ventilating the patient ………………………………………………………….. 29 1.7 Monitoring and alarms………………………………………………………….. 31 1.8 Using the trolley…………………………………………………………………… 32 1.9 Maintenance……………………………………………………………………….. 32 1.10 Service and testing ……………………………………………………………….. 34
19

1 Safety information

1.1 Overview
This chapter provides safety information related to setting up and operating the ventilator and trolley, as well as providing service.
Be sure to review this Operator’s Manual before using the ventilator and any accessories.
Be sure to read the Instructions for Use provided with any devices and accessories used with the ventilator before use.
Carefully review all sections of this safety chapter before setting up the ventilator and accessories, and ventilating the patient.
If you have questions about any of the information in this manual, contact your Hamilton Medical representative or technical service personnel.
1.2 Electromagnetic susceptibility
WARNING · MR UNSAFE. Keep away from mag-
netic resonance imaging (MRI) equipment. The HAMILTON-G5 poses unacceptable risks to the patient, medical staff, or other persons within the MR environment. · Correct function of the device may be impaired by the operation of high-frequency surgical equipment, microwaves, shortwaves, or strong magnetic fields in close proximity. · Follow precautions for electrostatic discharge (ESD) and electromagnetic interference (EMI) to and from the ventilator and any connected devices and accessories.

· Use of accessories, transducers, and cables other than those specified or provided by the manufacturer of this equipment can result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment, and may result in improper operation.
· Portable RF communications equipment, including peripherals such as antenna cables and external antennas, should be placed no closer than 30 cm (12 in) to any part of the HAMILTON-G5 ventilator, including any cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment can occur.
· The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11, class A). If it is used in a residential environment (for which CISPR 11, class B is normally required) this equipment might not offer adequate protection to radio-frequency communication services. The user might need to take mitigation measures, such as relocating or reorienting the equipment.
The HAMILTON-G5 complies with the IEC 60601-1-2 EMC (Electromagnetic Compatibility) Collateral Standard. The ventilator requires special precautions regarding electromagnetic compatibility (EMC). It must be installed and put into service according to the EMC information provided in the ventilator EMC Declarations (PN 624896). Portable and mobile RF communications equipment can affect the ventilator and all medical electrical equipment.

20

English | 624074/13

1 Fire and other hazards

1.3 Fire and other hazards

1.4.1 General operation and setup

WARNING · It is not permitted to use any of the
equipment with flammable gases or anesthetic agents, or in insufficiently ventilated areas. Danger of fire! · Do not use the ventilator with any equipment or high-pressure gas hoses that are worn or contaminated with oil or grease. · Highly compressed oxygen together with flammable sources could lead to spontaneous explosions. · In case of fire, immediately secure the patient’s ventilatory needs, turn off the ventilator, and disconnect it from its gas and electrical sources. · Do not use if primary power source cables are damaged. · The HAMILTON-G5 can be used in an oxygen-enriched environment. To reduce the risk of fire, use only breathing circuits intended for use in oxygen-enriched environments. Do not use antistatic or electrically conductive tubing.
1.4 General operation and setup
This section provides the following safety information: · General operation and setup · Electrical: power and batteries · Gas supply · USB ports

WARNING
· Modifications to the device and any accessories are not permitted.
· An O2 sensor must be installed. · In case of ventilator failure, the lack of
immediate access to appropriate alternative means of ventilation can result in patient death. · If a fault is detected in the ventilator or its life-support functions are in doubt, disconnect the ventilator from the patient and immediately start ventilation with an alternate device (for example, a resuscitation bag), using PEEP and/or increased oxygen concentration when appropriate. The ventilator must be removed from clinical use and serviced by a Hamilton Medical authorized service engineer. · Use only parts and accessories specified in Chapter 15 and in the product e-catalog, or that are specified as being compatible with this ventilator. Doing so ensures proper ventilation operation, avoids degraded performance, and keeps your warranty in force. · The use of this equipment is restricted to one patient at a time. · Only use the ventilator and its components and accessories according to the intended use and as described in the associated Instructions for Use. · Do not connect any component or device to the exhaust port of the expiratory valve unless authorized by Hamilton Medical. · The ventilator must not be used in a hyperbaric chamber.

Hamilton Medical | HAMILTON-G5 Operator’s Manual

21

1 Safety information

· If there is damage to any part of the ventilator, do not use the device. Technical service is required.
· Do not simultaneously touch conductive components (for example, the USB port) or conductive parts of the ventilator enclosure and the patient.
· Additional equipment connected to medical electrical equipment must comply with the respective IEC or ISO standards. All configurations must comply with the requirements for medical electrical systems, IEC 60601-1, clause 16.
· Anybody connecting additional equipment to medical electrical equipment configures a medical system and is responsible for ensuring that the system complies with the requirements for medical electrical systems. Local laws take priority over the above-specified requirements.
· Do not block the holes between the HAMILTON-G5’s To patient and From patient ports. These holes are vents for the overpressure and ambient valves.
CAUTION To prevent possible patient injury and ventilator overheating, do NOT block the cooling fan vents.

NOTICE · The barometric pressure is only mea-
sured and compensated during ventilator installation and setup, and with every service. There is no automatic calibration for barometric compensation. · Due to the ventilator’s base flow, the exhaust gas output is larger than the patient’s actual exhaled volume.
1.4.2 Electrical: power and batteries
WARNING · Ventilation stops if the battery or bat-
teries are discharged and no external power supply is connected. · To minimize the risk of electrical shock, plug the ventilator power cord into an appropriate grounded power receptacle. It is the hospital’s responsibility to ensure that the receptacle is properly grounded (earth). · The HAMILTON-G5 requires protective earth grounding, because it is a class I device, as classified according to IEC 60601-1. · Power sockets that can lead to a failure of ventilation must have a locking device. · It is the responsibility of the operator to ensure that the power system of any device connected to the ventilator power outlet complies with the requirements for medical electrical systems as well as local regulations. · Periodically check or replace the battery.

22

English | 624074/13

1 Gas supply

· Check the battery charge level before ventilating a patient and before unplugging the ventilator for transport or other purposes.
· Connect only the HAMILTON-H900 to the power strip.
CAUTION To electrically isolate the ventilator electrical circuits from all poles of the primary power supply simultaneously, disconnect the power plug.
NOTICE · Set up the ventilator in a location
where the primary power supply is accessible. · Only authorized service personnel may replace the power cable. · Battery life indications are approximate. The actual battery life depends on ventilator settings, battery age, and level of battery charge. To ensure maximum battery life, maintain a full charge and minimize the number of complete discharges. · After power has been interrupted, the device stores the last settings, including any specified alarm limits. Upon reconnection with the power supply, the device resumes ventilation with these stored settings.

1.4.3 Gas supply
CAUTION Always check the status of the oxygen and air cylinders or other supply before using the ventilator during transport.
NOTICE · To prevent damage to the ventilator,
connect only clean, dry medical grade oxygen and air. · When the ventilator is not in use, disconnect all gases.
1.4.3.1 Working with Heliox
WARNING To prevent possible hypoxia or death, connect a heliox gas supply that contains a minimum of 20% oxygen. The ventilator supports the following gas mixtures (HE% / O2%): 78/22, 79/21, and 80/20.
CAUTION To prevent heliox from entering the wall gas supply, connect compressed air with a minimum pressure of 2.8 bar.

Hamilton Medical | HAMILTON-G5 Operator’s Manual

23

1 Safety information

NOTICE
· When Heliox is in use: ­ The alarm lamp is illuminated in blue (when an alarm is generated, the lamp alternates blue with yellow or red, depending on the alarm priority) ­ O2 monitoring cannot be disabled
· Heliox is disabled when any of the following are selected or active: ­ Nebulization ­ INTELLiVENT-ASV mode ­ Tube resistance compensation (TRC)
· In the System > Gas source window, ensure that the selected gas source type matches the gas source connected to the ventilator. A mismatch can result in inaccurate gas delivery and volume monitoring.
· Calibrate the flow sensor after: ­ Switching between air and heliox connections ­ Significant changes in O2 concentration during heliox ventilation

1.5 Setting up for ventilation
This section provides the following safety information: · Patient breathing circuits, components,
and accessories · Performing preoperational checks and
testing · Humidifier · IntelliCuff · CO2 monitoring setup and operation · Nebulization · SpO2 monitoring setup and operation
See the Pulse Oximetry Instructions for use.
1.5.1 Patient breathing circuits, components, and accessories In addition to the information provided in this section, carefully review the information in Sections 1.3 and 1.4.

1.4.4 CompactFlash port
NOTICE The CompactFlash port is for data export and program update only (screenshots and log files). A Hamilton Medical CompactFlash card is recommended.

WARNING
· To prevent patient or ventilator contamination, always use a bacteria filter or HMEF between the patient and the inspiratory port. If no bacteria filter is used, the exhaled gas can contaminate the ventilator.
· Ensure that all of the components of the breathing circuit set, including but not limited to flow sensor, humidifier, and other accessories, match the associated intended use for the target patient group.

24

English | 624074/13

1 Preoperational check and tests

· Adding attachments or other components/assemblies to a breathing system can change the pressure gradient across the ventilator, which can adversely affect ventilator performance.
· For each new patient, always use a new or reprocessed breathing circuit to avoid cross contamination.
· During ventilation, regularly check the breathing circuit filter for increased resistance and blockage.
NOTICE · Any bacteria filter, HMEF, or addi-
tional accessories in the expiratory limb may substantially increase flow resistance and impair ventilation. · When adding components to the Hamilton Medical breathing circuit configurations, do not exceed the inspiratory and expiratory resistance values of the ventilator breathing system as specified in Section 16.10, as required by ISO 80601-2-12. · Pressure and volume measurement accuracy may be affected by using a breathing circuit with high resistance. Accuracy was tested with Hamilton Medical devices using the breathing circuits PN 260039 for adults, PN 260189 for pediatrics, and PN 151969 for neonates.

1.5.2 Preoperational check and tests
CAUTION · To prevent possible patient injury, dis-
connect the patient from the ventilator before running the preoperational tests, and use another source of ventilatory support. · To ensure the ventilator’s safe operation, always run the preoperational check before using the ventilator on a patient. · Do NOT use the ventilator until necessary repairs are completed and all preoperational tests have passed.
NOTICE · To ensure that all breathing circuit
connections are leak-tight, perform the tightness test every time you connect a circuit or change a circuit part. · If there is a mismatch between the selected patient group and the type of flow sensor connected, the calibration fails. Ensure you are using the correct flow sensor for the patient. · Wait 2 minutes before calibrating the flow sensor following a switch between air and heliox, or a significant change in the Oxygen setting. This allows the mixture to stabilize.

Hamilton Medical | HAMILTON-G5 Operator’s Manual

25

1 Safety information

1.5.3 Humidifier

1.5.4 IntelliCuff

WARNING · Before using a humidifier, review the
Instructions for Use as well as the Instructions for Use provided with its accessories. · To prevent possible patient injury and equipment damage, do not turn the humidifier on until the gas flow has started and is regulated. Turn the humidifier off before stopping gas flow. · Adding attachments or other components/assemblies to a connected humidifier can change the pressure gradient across the ventilator, which can adversely affect ventilator performance. · Regularly check the water traps and the breathing circuit limbs for water accumulation. Empty as required.
NOTICE The humidifier is not powered by the ventilator when operating on the backup battery.

WARNING · Never connect the tubing to any other
device or connector other than to the IntelliCuff port on the ventilator and to the inflating tube on the tracheal tube or tracheostomy tube. · Disconnect the IntelliCuff tubing from the tracheal or tracheostomy tube when IntelliCuff is turned off. · When the IntelliCuff tubing is connected to the ventilator, IntelliCuff starts applying the last-set or default pressure as soon as a pressure above 0 is detected in the tubing, even if IntelliCuff is disabled and the ventilator is in Standby.
CAUTION · Use only Hamilton Medical disposable
tubing with a filter and safety valve. Use of any other tubing will result in the immediate loss of cuff pressure if disconnected at the ventilator. Use of any other tubing without a filter may result in the device being contaminated. · Check tubing regularly. Bent or kinked tubes can provide incorrect monitoring information.

26

English | 624074/13

1 CO2 sensor setup and operation

1.5.5 CO2 sensor setup and operation
WARNING · Monitor the CO2 waveform (capno-
gram) on the ventilator display. If it appears abnormal, check the patient, settings, and the breathing circuit components, including the CO2 sensor sampling line. Adjust and replace components as appropriate. · If the capnogram appears abnormal, inspect the CO2 airway adapter and replace if needed. · Elevated baseline can be caused by sensor problems or an issue with the patient. · Do not use any CO2 sensor/adapter if they appear to be damaged or if they fail to operate properly. Refer servicing to Hamilton Medical authorized personnel. · Do not use the CO2 components when they are wet or have exterior condensation. · In NIV and neonatal ventilation with uncuffed tubes, leaks may influence the volumetric capnogram and the measured values. · Always connect all components securely and check for leaks according to standard clinical procedures. Displacement of the nasal or combined nasal-oral cannulas can cause lowerthan-actual CO2 readings.

· Positioning of tubes and cables: ­ Do not position the cables or tubing in any manner that may cause patient entanglement or strangulation. ­ Support the tubing to avoid stress on the ET tube. ­ Do not apply excessive tension to any cable or tubing.
· During use, a system leak, such as that caused by an uncuffed ET tube or damaged CO2 sensor may significantly affect sensor readings, including flow, volume, pressure, and other respiratory parameters.
· Leakages in the breathing or sampling system may cause the displayed CO2 values to be significantly underreported (too low).
· Keep all cleaning agents away from the CO2 sensor electrical connections.
· For the CO2 sensor/adapter, use only cleaning and disinfection agents that are recommended in the Approved cleaning agents for CO2 components, available on MyHamilton.
· Periodically check the sensor and tubing for excessive moisture or secretion build-up, and replace if needed. Excessive moisture can affect measurements.
· LoFlo sidestream CO2 sensor. Do not use with patients who cannot tolerate the removal of 50 ml ±10 ml/ min from their total minute volume. In adaptive modes (such as ASV, APVcmv, and APVsimv), the removal is fully compensated.

Hamilton Medical | HAMILTON-G5 Operator’s Manual

27

1 Safety information

· LoFlo sidestream CO2 sensor. Use of devices containing PVC plasticized with DEHP should be limited to the amount of time treatment is medically necessary, especially for neonates and pregnant or nursing mothers.
CAUTION · All devices are NOT protected against
reanimation with a defibrillator. Disconnect the CO2 sensor before using a defibrillator on the patient. · Always use the correct CO2 adapter. In adult patients, smaller geometrics increase airway resistance and induce low tidal volumes and intrinsic PEEP. In neonatal patients, large geometrics impede effective CO2 removal. · Do NOT place the CO2 sensor on the patient. It can burn the skin as the sensor may reach a temperature of 46°C (115°F). · Use during nebulization may influence the CO2 measurements. In addition, the medication can contaminate the sensor windows, causing the sensor to fail prematurely. · LoFlo sidestream CO2 sensor. Remove the sampling kit sample cell from the module when not in use. · LoFlo sidestream CO2 sensor. Do NOT stick finger into the sample cell receptacle.

NOTICE
· Position airway adapters with windows in a vertical, not a horizontal, position. This helps keep patient secretions from pooling on the windows. If pooling occurs, remove the adapter, rinse with water, and reconnect.
· Do not combine the neonatal CO2 airway adapter and the adult flow sensor. Doing so can increase resistance, create artifact, or lead to hypoventilation, intrinsic PEEP, or overinflation.
· Do not place the CO2 sensor/adapter between the ET tube and the elbow, as this may allow patient secretions to enter the tubing and block the adapter windows.
· The CO2 sensors and accessories that have contact with the patient are not made with natural rubber latex.
· Nitrous oxide, elevated levels of oxygen, helium, and halogenated hydrocarbons can influence the CO2 measurement.

28

English | 624074/13

1 Nebulization

1.5.6 Nebulization For additional safety information related to Aerogen§ nebulizers, see the Aerogen Solo/Aerogen Pro Instructions for Use.
WARNING · Nebulization of drugs can cause an
occlusion and increased resistance of a connected expiratory filter. Check the filter frequently for increased resistance or blockage. · Connect the nebulizer in the inspiratory limb per your institution’s policy and procedures. Connecting the nebulizer between the flow sensor and the endotracheal tube increases dead space and causes incorrect volume measurements. · Pneumatic nebulization affects the delivered oxygen concentration. · Nebulization can affect the accuracy of CO2 measurements.
CAUTION To prevent the expiratory valve from sticking due to nebulized medications, regularly check and clean or replace the expiratory valve membrane and/or the expiratory filter.
NOTICE · Pneumatic nebulization is disabled:
­ During neonatal ventilation (if needed, use an Aerogen nebulizer1) ­ When using Hi Flow O2 ­ When using heliox · Only use approved piezo nebulizers with the HAMILTON-G5.

1.6 Ventilating the patient
This section provides the following safety information: · Specifying patient settings · Neonatal ventilation · Apnea backup · TRC settings · P/V Tool Pro · Noninvasive ventilation · Using high flow oxygen therapy
1.6.1 Specifying patient settings
WARNING · It is the clinician’s responsibility to
ensure that all ventilator settings are appropriate, even when “automatic” features, such as ASV, or default settings are used. · To prevent possible patient injury: ­ Make sure the ventilator is set up for the appropriate patient group with the appropriate breathing circuit components. ­ For each patient group, make sure you select the correct patient sex and height (Adult/Pediatric) or weight (Neonatal). Correct entries help prevent hyper- or hypo-ventilation. · The ventilator is a high-flow device that can operate with flows above 60 l/min and with a high oxygen concentration.

1 Not available in all markets.

Hamilton Medical | HAMILTON-G5 Operator’s Manual

29

1 Safety information

1.6.2 Neonatal ventilation In addition to the information provided in this section, carefully review the information in Sections 1.5 and 1.6.
WARNING Prolonged exposure to high oxygen concentrations may cause irreversible blindness and pulmonary fibrosis in preterm neonates. Be especially careful when performing oxygen enrichment.
CAUTION · To prevent increased CO2, do NOT
use an adult airway adapter for neonates as it will increase dead space. · To determine appropriate tidal and minute volumes for neonatal patients, you must consider (anatomic) dead space. Artificial airways (Y-piece, flow sensor, ET tube, CO2 airway adapter) may increase the dead space. · When using active humidification, prevent water accumulation in the flow sensor by ensuring that the flow sensor is positioned at a 45° angle relative to the floor. Excess water can affect the flow sensor measurements and lead to inaccurate volume delivery, potentially resulting in hypoventilation.

NOTICE When switching between the Adult, Pediatric, and Neonatal patient groups, you must calibrate the flow sensor and perform the tightness test.
1.6.3 Apnea backup
CAUTION We recommend you enable apnea backup ventilation whenever a mode that allows spontaneous breathing is selected. Apnea backup is enabled by default.
1.6.4 TRC settings
WARNING To ensure patient safety, check that the Pressure alarm limit is set appropriately when using TRC, as real pressure may be higher than the set pressure.
CAUTION To prevent patient injury, be especially careful when defining TRC settings, as using the incorrect tube type or size setting can endanger the patient.

30

English | 624074/13

1 Noninvasive ventilation

1.6.5 Noninvasive ventilation
NOTICE · As a precaution, you must be pre-
pared to intubate the patient and start invasive ventilation at any time while noninvasive ventilation is in use. · The use of a mask can increase dead space. Always comply with the mask manufacturer’s instructions when using noninvasive ventilation.
1.6.6 P/V Tool Pro
WARNING Do not attempt to use the P/V Tool on an active patient as it can cause patient discomfort and erroneous readings.
NOTICE · During a maneuver and for 30
seconds following the end of the maneuver, all patient alarms are silenced. · Apnea time begins after the end of the maneuver. · Use of the P/V Tool provides information that, in conjunction with hemodynamic data and other clinical information, may be used to optimize PEEP and other ventilator settings. · During the maneuver, the high Pressure alarm is automatically set to Ptop + 5 cmH2O. When the maneuver is finished, the high Pressure alarm limit returns to the previous setting. If IntelliCuff is connected, Pcuff may also be affected. For details, see Section 12.2.4.1.

· A calibrated flow sensor and a tight circuit produce the best results.
1.6.7 Using high flow oxygen therapy
WARNING · Use only interfaces intended for high
flow oxygen therapy that allow the patient to exhale, such as a nonocclusive high-flow nasal cannula, tracheal adapter, or tracheal mask. This is important because exhalation through the expiratory valve is not possible when using high flow oxygen therapy. · Ensure the ventilator’s gas pipeline system does not exceed the pipeline design flow capacity. If the system exceeds the flow capacity, it can interfere with the operation of other equipment using the same gas source. · Always use active humidification during high flow oxygen therapy.
1.7 Monitoring and alarms
CAUTION · To prevent possible patient injury,
make sure the alarm limits are appropriately set before you place the patient on the ventilator. · The HAMILTON-G5 oxygen monitoring function can be disabled, except when Heliox is in use. Ensure that an alternative means of oxygen monitoring is always available and enabled.

Hamilton Medical | HAMILTON-G5 Operator’s Manual

31

1 Safety information

· To ensure that oxygen monitoring is always fully functional, replace an exhausted or missing O2 sensor as soon as possible or use an external monitor that complies with ISO 80601-2-55.
NOTICE · The HAMILTON-G5 is not intended to
be a comprehensive vital sign monitor for patients on life-support equipment. Patients on life-support equipment should be appropriately monitored by qualified medical personnel and suitable monitoring devices. · The use of an alarm monitoring system does not give absolute assurance of warning for every type of issue that may arise with the ventilator. Alarm messages may not pinpoint a problem exactly; the exercise of clinical judgment is necessary. · It is recommended that additional independent monitoring devices, including pulse oximeters measuring SpO2, be used during mechanical ventilation. The operator of the ventilator must still maintain full responsibility for proper ventilation and patient safety in all situations. · Do not pause the audible alarm when leaving the patient unattended. · The Auto function is not available during neonatal ventilation.

1.8 Using the trolley
WARNING · To prevent possible personal injury
and equipment damage, including tipping: ­ Lock the trolley’s wheels when parking the ventilator. ­ Take care when crossing thresholds. · To prevent accidental extubation, check the patient tubing support arm joints and secure as necessary.
1.9 Maintenance
This section provides the following safety information: · Maintenance, cleaning, and disinfection · Preventive maintenance
1.9.1 General maintenance, cleaning, and disinfection
WARNING · Reprocessing of Hamilton Medical sin-
gle-use products can affect the product properties and may cause injury to the patient. For example, a change to the surface structure during reprocessing may lead to a change in the tear strength or cause actual cracking. Furthermore, an altered surface structure may result in a microbial aggregation of spores, allergens and pyrogens, for example, or cause an increase in the number of particles released as a result of chemical changes in the material properties.

32

English | 624074/13

1 General maintenance, cleaning, and disinfection

· To reduce the risk of cross-contamination, regularly clean and replace the fan filter. For details, see Table 13-3 and Section 13.4.1.
· To prevent patient exposure to sterilizing agents and to prevent premature deterioration of parts, sterilize parts using only the techniques recommended in Chapter 13 and in any associated Reprocessing Guide or Instructions for Use provided with each part.
· Hamilton Medical does not assume any liability for the proper functioning of single-use items if they are reprocessed and reused by the user.
· Always use caution when handling bacteria filters to minimize the risk of bacterial contamination or physical damage. Dispose of used filters immediately after use. Follow your hospital procedures for disposal.
· Follow the cleaning, disinfection, and sterilization procedures for each component as described in this guide and in the cleaning agent manufacturer’s Instructions for Use.
· Always disconnect the device and any accessories, including CO2 sensor/ adapter, from electrical power before cleaning and disinfection to reduce the risk of electric shock.
CAUTION · Do NOT sterilize or immerse the CO2
sensor in liquids. · Do NOT attempt to sterilize the inte-
rior components of the ventilator. · Do NOT attempt to sterilize the entire
device with ETO gas. · Incorrect concentrations or residence
times of sterilization agents may lead to bacterial resistance.
Hamilton Medical | HAMILTON-G5 Operator’s Manual

· To prevent premature deterioration of parts, make sure the disinfecting chemical is compatible with the part material. Use only registered/ approved cleaning and disinfection solutions, as approved by your institution’s protocol, after each patient use, according to the cleaning agent manufacturer’s recommendations.
· Intrusion of fluids, or immersing parts in fluids, will damage the device.
· Do NOT pour fluids onto the device surfaces.
· Do NOT use abrasives materials (for example, steel wool or silver polish), hard brushes, pointed instruments, or rough materials on surfaces.
· Thoroughly rinse all patient or airway contact component to ensure removal of residual cleaning/disinfection agents.
· Cleaning and disinfection agent residues can cause blemishes or fine cracks, especially on parts exposed to elevated temperatures during sterilization.
NOTICE · Because sanitation practices vary
among institutions, Hamilton Medical cannot specify specific practices that will meet all needs or be responsible for the effectiveness of these practices. · This Operator’s Manual only provides general guidelines for cleaning, disinfecting, and sterilizing. It is the operator’s responsibility to ensure the validity and effectiveness of the actual methods used.
33

1 Safety information

· For specific information on cleaning, disinfecting, and sterilizing autoclavable (reusable) accessories and components, refer to the appropriate Reprocessing Guide and Instructions for Use provided with each part.
1.9.2 Preventive maintenance
NOTICE · Dispose of all parts removed from the
device according to your institution’s protocols. Comply with all local, state, and federal regulations with respect to environmental protection, especially when disposing of the electronic device or parts of it (for example, O2 sensor). · We recommend that you document all maintenance procedures. · It is not allowed to perform service or maintenance on the device while a patient is connected. · If no bacteria (inspiratory) filter is used, the device must be considered contaminated and must be serviced.
1.9.3 O2 sensor
NOTICE · To prevent leakage within the ventila-
tor, make sure an O2 sensor is installed at all times, even if you use an external monitor or disable oxygen monitoring. · The paramagnetic O2 sensor must only be replaced if it fails. In this case, have the ventilator serviced.

1.10 Service and testing
· To ensure proper servicing and to prevent possible physical injury, only Hamilton Medical authorized service personnel may service the ventilator using information provided in the Service Manual. In addition, all accessories and devices must only be serviced by Hamilton Medical authorized service personnel.
· The manufacturer can only be responsible for the safety, reliability, and performance of the ventilator if all of the following requirements are met: ­ Appropriately trained personnel carry out assembly operations, extensions, readjustments, modifications, maintenance, or repairs. ­ The electrical installation of the relevant room complies with the appropriate requirements. ­ The ventilator system is used in accordance with the operator’s manual. ­ Do not attempt service procedures other than those specified in the service manual.
· Any attempt to modify the ventilator hardware or software without the express written approval of Hamilton Medical automatically voids all warranties and liabilities.

34

English | 624074/13

2
System overview
2.1 Overview…………………………………………………………………………….. 36 2.2 Physical descriptions……………………………………………………………… 39 2.3 Navigating the windows and controls………………………………………. 53
35

2 System overview

2.1 Overview
The HAMILTON-G5 ventilator system comprises the following main components: · Detachable monitor with integrated
alarm lamp and touch screen display · Ventilation unit for gas mixing and
control, and patient breathing circuit for gas delivery and exchange · Oxygen monitoring using a galvanic or optional paramagnetic sensor · Optional connections to a humidifier, IntelliCuff cuff pressure controller, SpO2 and CO2 sensors, and external data interfaces · Trolley, shelf, or pendant mount
The ventilator system offers the following main features: · Monitoring: Real- time waveforms,
numerical monitoring, trends, loops, and Intelligent panels showing the patient’s real-time breathing status, ventilator dependence, and targets, CO2 and SpO2 measurements (when enabled) · Alarms: Adjustable and non-adjustable to ensure patient safety · Configurable startup settings for each patient group · Monitoring and control of the HAMILTON-H900 humidifier from the ventilator · Monitoring and control of the IntelliCuff cuff pressure controller from the ventilator · Transpulmonary pressure measurement · Support for pneumatic or Aerogen nebulization

2.1.1 Standard features and options The ventilator offers a robust set of standard equipment and features, as well as optional modes and features for the supported patient groups. Table 2-1 lists the standard software configuration and options. Table 2-2 lists the standard equipment (hardware) and options.

36

English | 624074/13

2 Standard features and options

Table 2-1. Standard software configuration and options

Function

Patient group

Adult/Pediatric

Neonatal

Standard: X Option: O Not applicable: —

Patient groups

X

O

Modes

Intelligent ventilation modes

ASV

X

INTELLiVENT-ASV

O

Volume-targeted, pressure-controlled modes

APVcmv

X

X

APVsimv

X

X

Volume-controlled, flow-controlled modes

(S)CMV

X

SIMV

X

Volume-controlled, flow-cycled mode

Volume Support (VS)

X

X

Pressure-controlled modes

DuoPAP, APRV

X

X

P-CMV

X

X

P-SIMV

X

X

SPONT

X

X

Noninvasive modes

Hi Flow O2

O

O

NIV, NIV-ST

X

nCPAP-PS

O

Hamilton Medical | HAMILTON-G5 Operator’s Manual

37

2 System overview

Function
Other functions P/V Tool, P/V Tool Pro Intellisync+ Flow and pressure triggers TRC Suctioning tool Trends/Loops

Patient group

Adult/Pediatric

Neonatal

O

O

O

X

X

X

X

X

X

X

X

Table 2-2. Standard equipment (hardware) configuration and options

Functions Standard: X Option: O
Trolley, shelf mount, or pendant mount solution (selected when ordering) External battery Modules for external sensors/devices: CO2, SpO2, Nebulizer, Humidifier Heliox ventilation Extended communication ports: CompactFlash, USB, DVI, COM (RS-232), Special interface Communication protocols (for use with the COM ports): HAMILTON-G5 / Polling, HAMILTON-G5 / Block, HAMILTON-G5 / Block (ACK), Galileo / Polling, DraegerTestProtocol, Humidifier Paramagnetic O2 sensor Paux port HAMILTON-H900 humidifier integration IntelliCuff cuff pressure controller integration

HAMILTON-G5
X O O
O X
O
O X O O

38

English | 624074/13

2 Physical descriptions

2.2 Physical descriptions
This section provides an overview of the ventilator, breathing circuit sets, and trolley. Figure 2-1. HAMILTON-G5 with accessories

2.2.1 About the ventilator Figures 2-2 through 2-5 provide an overview of the device.

1 Support arm and infusion arm
2 Display and controls
3 Breathing circuit connections

4 Breathing circuit 5 Humidifier 6 Trolley

Hamilton Medical | HAMILTON-G5 Operator’s Manual

39

2 System overview
Figure 2-2. Front view, ventilator monitor

1

Alarm lamp*

6

Screen lock/unlock

2

Touch screen display (Figure 2-6)

7

Nebulizer key

3

Manual breath key

8

Standby key

4

O2 enrichment key

9

Print screen key

5

Audio Pause key

10 Press-and-Turn (P&T) knob

  • The alarm lamp illuminates in blue when heliox is in use.

40

English | 624074/13

Figure 2-3. Front view, ventilator body

2 About the ventilator

1

Paux port

7

2

Nebulizer port

8

3

Flow sensor connection

9

ports

4

IntelliCuff port

10

5

Expiratory valve set

11

6

From patient expiratory port

Hamilton Medical | HAMILTON-G5 Operator’s Manual

Exhaust port To patient inspiratory port Inspiratory filter
CO2/SpO2/Aerogen/ Humidifier option module ports Status indicator panel (Section 2.2.1.1)
41

2 System overview
Figure 2-4. Rear view, ventilator body

1

Fan filter

8

AC power socket

2

Serial number label

9

Monitor cable

3

Power button

10 Fuse compartment

4

High-pressure air DISS or NIST inlet

fitting

11 Oxygen sensor with cover

5

High-pressure oxygen DISS or NIST inlet 12 Potential equalization conductor

fitting (for heliox, see Section 3.3)

6

High-pressure gas water trap with filter 13 RS-232 COM1, COM2 ports

7

Reservoir pressure relief valve exhaust

14 Special interface

42

English | 624074/13

Figure 2-5. Rear view, ventilator monitor

2 About the ventilator

1

Monitor cable storage

5

CompactFlash port

2

Tilt-release lever

6

USB port

3

Monitor cable

7

DVI-I connection port

4

Mounting post with swivel lock/

release latch

CAUTION! For training purposes only. Not for use with a connected patient.

Hamilton Medical | HAMILTON-G5 Operator’s Manual

43

2 System overview

2.2.1.1 About the status indicators on the ventilator Indicator lights on the front of the ventilator unit show important ventilation status information. Table 2-3. Status indicator panel

Symbol

Description
Alarm indicator. Solid red when an alarm is active.
For alarm related information, see Chapter 9.
Primary power indicator. Solid blue when the ventilator is plugged in and connected to primary (AC) power.
Power indicator. Solid green when the ventilator is turned on.

2.2.2 About the main display Directly access all the windows for mode, controls, alarms, and monitoring from the main display during normal ventilation (Figure 2-6).

44

English | 624074/13

Figure 2-6. Main display

2 About the main display

1

Message bar (color coded)

8

Window buttons: Alarms, Controls,

Monitoring, Graphics, Tools, Events,

System

2

Configurable graphic display

9

Power source

3

Window buttons: Patient, Additions,

Modes

10 Humidifier quick access icon

4

INTELLiVENT-ASV button

11 Main monitoring parameters (MMP)

5

Active mode and selected patient group 12 Secondary monitoring parameters (SMP)

6

IntelliCuff quick access icon and/or

Heliox icons (when installed and

selected)

13 Audio Pause indicator and countdown timer

7

Main controls for the active mode

Hamilton Medical | HAMILTON-G5 Operator’s Manual

45

2 System overview

2.2.3 About the patient breathing circuits For details about connecting and setting up the breathing circuit, see Section 3.4. When setting up the patient breathing circuit, keep in mind the following important points: · To prevent patient or ventilator con-
tamination, be sure to connect a bacteria (inspiratory) filter or HMEF between the patient and the inspiratory port.
Figure 2-7. Adult/pediatric breathing circuits Adult/Ped: Dual limb with humidifier

· Connect the CO2 sensor in front of or behind the flow sensor, according to your institution’s protocol.
· During neonatal ventilation with active humidification, prevent water accumulation in the flow sensor by ensuring that the flow sensor is positioned at a 45° angle relative to the floor.
Adult/Ped: Coaxial with HMEF

1

To patient inspiratory port

9

Y-piece

2

From patient expiratory port

10 CO2 sensor/adapter

3

Expiratory valve set

11 Flow sensor

4

Flow sensor connection ports

12 Humidifier

5

Bacteria filter

13 Coaxial inspiratory/expiratory limb

6

Inspiratory limb to humidifier

14 Expiratory limb extension

7

Heated inspiratory limb with tempera- 15 HMEF

ture sensor, to patient

8

Heated expiratory limb

46

English | 624074/13

2 About the patient breathing circuits

Figure 2-8. Adult/pediatric breathing circuit: high flow oxygen therapy

Adult/Ped: Dual limb, high flow oxygen therapy

Adult/Ped: Single limb, high flow oxygen therapy

1

To patient inspiratory port

7

Heated expiratory limb

2

From patient expiratory port

8

Y-piece

3

Expiratory valve set

9

Adapters (various)

4

Bacteria filter

10 Nasal cannula

5

Inspiratory limb to humidifier

11 Attachment strap

6

Heated inspiratory limb with tempera- 12 Humidifier

ture sensor, to patient

Hamilton Medical | HAMILTON-G5 Operator’s Manual

47

2 System overview
Figure 2-9. Neonatal breathing circuits Neonatal/pediatric: Dual limb with humidifier

Neonatal/pediatric: Dual limb with HMEF

1

To patient inspiratory port

9

Heated expiratory limb

2

From patient expiratory port

10 Y-piece

3

Expiratory valve set

11 CO2 sensor/adapter

4

Flow sensor connection ports

12 Flow sensor

5

Bacteria filter

13 Humidifier

6

Inspiratory limb to humidifier

14 Inspiratory limb

7

Heated inspiratory limb with tempera- 15 Expiratory limb

ture sensor, to patient

8

Unheated inspiratory limb extension,

for use in incubator

16 HMEF

48

English | 624074/13

2 About the patient breathing circuits

Figure 2-10. Neonatal breathing circuits: high flow oxygen therapy

Neonatal/pediatric: Dual limb, high flow oxygen therapy

Neonatal/pediatric: Single limb, high flow oxygen therapy

1

To patient inspiratory port

7

Unheated inspiratory limb extension,

for use in incubator

2

From patient expiratory port

8

Heated expiratory limb

3

Expiratory valve set

9

Y-piece

4

Bacteria filter

10 Connection to patient interface (options not shown)

5

Inspiratory limb to humidifier

11 Humidifier

6

Heated inspiratory limb with tempera-

ture sensor, to patient

Hamilton Medical | HAMILTON-G5 Operator’s Manual

49

2 System overview
Figure 2-11. Neonatal breathing circuit: nCPAP-PS Neonatal: nCPAP-PS

1

To patient inspiratory port

2

From patient expiratory port

3

Expiratory valve set

4

Flow sensor connection ports

5

Bacteria filter

6

Inspiratory limb to humidifier

7

Heated inspiratory limb with tempera-

ture sensor, to patient

8

Unheated inspiratory limb extension,

for use in incubator

9

Heated expiratory limb

10 Flow sensor (connected to expiratory port)

11 Y-piece

12 Humidifier

50

English | 624074/13

2 About the trolley and mounting variations

2.2.4 About the trolley and mounting variations The HAMILTON-G5 can optionally be ordered with a standard trolley, pendant mount, or a shelf mount solution. The trolley has space for oxygen cylinders.
2.2.4.1 Preparing the trolley for intrahospital transport Before proceeding, review the safety information in Chapter 1.
WARNING · Only the components listed in this
section are approved for intrahospital transport. · Use of additional items, such as a patient support arm, can result in the trolley tipping over.

If using a HAMILTON-G5 trolley, the ventilator and its components, as well as the trolley, must be configured and positioned as follows during transport within the hospital:
· The oxygen cylinders must be securely attached to the trolley.
· Only the following components are allowed to be connected during transport: ­ Breathing circuit ­ Flow sensor ­ CO2 sensor (mainstream or sidestream) ­ SpO2 sensor, including Masimo adapter ­ Infusion arm (water bottle holder)

CAUTION To prevent possible equipment damage, avoid overloading the HAMILTON-G5’s basket and tray, or placing objects on the HAMILTON-G5 that might compromise its stability.
NOTICE The O2 cylinder can only be mounted on the Universal trolley.

Hamilton Medical | HAMILTON-G5 Operator’s Manual

51

2 System overview

2.2.5 Setting up the monitor The HAMILTON-G5 offers multiple mounting options for the monitor. You can also adjust the tilt and view angle of the monitor.
2.2.5.1 Mounting the monitor The following mounting options are available for the ventilator monitor: · Top of the trolley (1) · Trolley rail (2) · Standard hospital rail (3) · Shelf (4) · Pendant system (4) Figure 2-12. Mounting options

2.2.5.2 Adjusting the monitor You can adjust the monitor’s position and set it to the desired orientation and angle by turning and tilting it, as needed.
To tilt the monitor up and down 1. Pull the tilt handle toward you (1), and
adjust the angle of the monitor (2). 2. Release the handle to lock the moni-
tor’s position. Figure 2-13. Tilting the monitor up and down

Contact your Hamilton Medical representative for more information.

To turn the monitor side to side 1. Press the bottom of the monitor post
latch to unlock it (1), and turn the monitor to the desired angle (2). 2. Press the top of the latch to lock the monitor’s position.

52

English | 624074/13

2 Navigating the windows and controls

Figure 2-14. Turning the monitor side to side
2.3 Navigating the windows and controls
Use the touch screen and the Press-andturn knob (referred to as the P&T knob) to access data and specify settings. You interact with the HAMILTON-G5 user interface as follows: · Touch elements on the display to open
windows and make selections. · Use the P&T knob to select, specify,
and confirm selections. A selected item is highlighted in yellow. This section describes how to navigate the interface.

2.3.1 Accessing windows
To open a window
4 Do any of the following to open a window: Touch the button and any needed tabs. Turn the P&T knob to move the cursor to the button or tab, then press the P&T knob.
To close a window
4 Do any of the following to close a window: Touch the window button again. Touch the X button. Turn the P&T knob to move the cursor to the X button, then press the P&T knob.

Hamilton Medical | HAMILTON-G5 Operator’s Manual

53

2 System overview

2.3.2 Adjusting controls
Specifying settings involves activating a control, setting a value, and confirming the setting.
To adjust a control setting 1. Activate the control by doing any of
the following: Touch the control to select and activate it. Turn the P&T knob to move the cursor to the control, then press the P&T knob to activate it. The activated control is orange.
Figure 2-15. Selected control (yellow outline), activated control (orange)

2.3.3 Selecting list items Some selections are presented in a scrollable list.
To select a list item 1. In a list, touch the scroll bar to select
and activate it. 2. Turn the P&T knob to scroll through
the list, and when the desired selection is highlighted, press the knob to select it.

2. Adjust the value by turning the P&T knob to increase or decrease the value.
3. Confirm the setting by doing any of the following: Touch the control again. Press the P&T knob.
The new setting is immediately applied.

54

English | 624074/13

3
Preparing the ventilator
3.1 Overview…………………………………………………………………………….. 56 3.2 Connecting to a power source ……………………………………………….. 56 3.3 Connecting the oxygen supply ……………………………………………….. 57 3.4 Setting up the patient breathing circuit ……………………………………. 58 3.5 Setting up esophageal/transpulmonary pressure monitoring ………… 62 3.6 Turning the ventilator on and off ……………………………………………. 62
55

3 Preparing the ventilator

3.1 Overview
Preparing the ventilator for use comprises the following steps:

To …
Connect to a power source.
Connect the oxygen supply.
Set up the patient breathing circuit, including performing the preoperational check.
Connect external devices and sensors.
Turn on the ventilator.
Select the patient group, mode, and alarm limits, and enter patient data.

See … Section 3.2 Section 3.3 Section 3.4
Chapter 4
Section 3.6 Chapter 5

3.2 Connecting to a power source
Before proceeding, review the safety information in Chapter 1. Always check the reliability of the primary power outlet before plugging in the ventilator. When connected to primary power, the AC power symbol in the bottom right corner of the display shows a frame around it. In addition, the primary power symbol on the status indicator panel is lit.

To connect the ventilator to a primary power supply 1. Connect the ventilator to an outlet
that supplies AC power. Make sure the power cord is well seated into the ventilator socket and secured with the power cord retaining clip to prevent unintentional disconnection. 2. Connect one end of a grounding cable to the equipotential grounding post on the ventilator (Figure 2-4) and the other to a properly grounded outlet.
3.2.1 Using battery power A mandatory backup battery protects the ventilator from low power or failure of the primary power source. The backup battery is labeled INT on the ventilator. When the primary power source fails, the ventilator automatically switches to operation on backup battery with no interruption in ventilation. An alarm sounds to signal the switch-over. Silence the alarm to confirm notification of the power system change and reset the alarm. If battery power is completely lost, a buzzer sounds continuously for at least two minutes. Batteries are charged whenever the ventilator is connected to the primary power supply, whether or not it is turned on. The battery and power source symbols in the bottom right corner of the display show the power source in use. See Table 3-1. A frame around a power symbol indicates the current ventilator power source. An optional second battery is available. It is labeled EXT on the display, and is only shown when installed.

56

English | 624074/13

3 Connecting the oxygen supply

Figure 3-1. Power source indicators on display

Table 3-1. Battery/power state
Power icon Battery/power state on display
Device is plugged into primary power and the battery is charging.
Device is running on battery power.
Battery is fully charged.

Battery is partially charged.

Power icon on Status indicator panel

Battery has less than 10% charge left.
Battery is either defective or not installed.
See Section 2.2.1.1.

If a battery is not fully charged, recharge it by connecting the ventilator to the primary power source. For details, see Section 16.4. Chapter 13 describes how to replace the optional battery.

3.3 Connecting the oxygen supply
Before proceeding, review the safety information in Chapter 1. High-pressure oxygen, provided by a central gas supply or a gas cylinder, is supplied through DISS or NIST male gas fittings. The ventilator uses high-pressure oxygen, air, and heliox from wall supplies, cylinders, or the VENTILAIR II medical air compressor. With the optional cylinder holder, you can mount oxygen cylinders to the trolley. If you use gases from cylinders, secure the cylinders to the trolley with the accompanying straps. To connect the gas supply to the ventilator 4 Connect the gas hose to the ventila-
tor’s oxygen inlet fitting (Figure 2-4).
3.3.1 Working with heliox as a gas source Before proceeding, review the safety information in Chapter 1. Heliox is a mixture of helium and oxygen, and can be indicated for patients in cases of acute and life-threatening upper airway obstruction. This action is taken as a temporary measure to provide a decrease in the patient’s work of breathing while the cause of the obstruction is treated. Administering heliox can make it easier to ventilate, because its lower density can allow a patient to produce inspiratory and expiratory flows with less turbulence.

Hamilton Medical | HAMILTON-G5 Operator’s Manual

57

3 Preparing the ventilator

3.3.2 Selecting the gas source type
Before starting ventilation, be sure to select the appropriate gas source.
You set the source in Standby mode.
To select the gas source 1. In Standby mode, open the System >
Gas Source window. 2. Touch the appropriate button for the
desired gas source. Select Air or Heliox as appropriate. When Heliox is selected, the alarm lamp on top of the display is lit blue. 3. Close the window. 4. Calibrate the flow sensor.
Figure 3-2. Gas source window

3.4 Setting up the patient breathing circuit
Before proceeding, review the safety information in Chapter 1. Connecting the breathing circuit comprises the following steps. For neonatal ventilation, see Chapter 6.

To …
Install the expiratory valve.
Select the appropriate breathing circuit and components.
Assemble the breathing circuit.
Adjust the position of the breathing circuit.
Connect external devices and sensors.
Perform any required tests, calibrations, and the preoperational check.

See … Section 3.4.2 Section 3.4.3
Section 3.4.4 Section 3.4.5 Chapter 4 Chapter 5

1 System 2 Gas Source

3 Air, Heliox2

2 If the option is installed and activated.
58

English | 624074/13

3 Breathing circuit connections on the ventilator

3.4.1 Breathing circuit connections on the ventilator Figure 3-3 illustrates the key ports on the ventilator for connecting the breathing circuit set. For breathing circuit diagrams, see Section 2.2.3.
Figure 3-3. Key connection ports, front of ventilator

3.4.2 Working with the expiratory valve set This section describes how to assemble, install, and disassemble the expiratory valve set.
To assemble/install the expiratory valve set 4 Position the expiratory valve set
(Figure 3-4) in the expiratory port (1) and twist clockwise until it locks into place (2). Figure 3-4. Installing the expiratory valve set

To disassemble the expiratory valve set 4 Remove the expiratory valve set from
the expiratory valve port on the ventilator.

1 To patient inspira- 5 Nebulizer port tory port

2 From patient expiratory port

6 Flow sensor connection ports

3 Expiratory valve exhaust

7 IntelliCuff tubing port

4 Paux port

8 CO2, SpO2, Aerogen, and Humidifier module ports, if installed

Hamilton Medical | HAMILTON-G5 Operator’s Manual

59

3 Preparing the ventilator

3.4.3 Selecting the breathing circuit components

Select the correct breathing circuit parts for your patient.

For neonatal ventilation, see Chapter 6.

Table 3-2. Breathing circuit component specifications

Patient data/ Component
Patient height (cm)

Adult > 130

Pediatric 30 to 150

IBW (kg)

30

Tracheal tube ID > 4 (mm)

Breathing circuit 15 to 22 limb ID (mm)3

Flow sensor

Adult/Ped

3 to 48 > 4
10 to 22
Adult/Ped

CO2 airway adapter

Adult/Ped4 Adult/Ped4

3.4.3.1 Using a filter in the breathing circuit Before proceeding, review the safety information in Chapter 1. Inspiratory bacteria filter To prevent patient or ventilator contamination, be sure to connect a bacteria (inspiratory) filter or HMEF between the patient and the inspiratory port. For neonatal patients, use a neonatalpediatric bacteria (inspiratory) filter or HMEF.

If no inspiratory filter is used, the exhaled gas can contaminate the ventilator. If you are not using an inspiratory filter, and an exhalation obstructed alarm is generated, the ventilator may be contaminated. Have the ventilator serviced.
Expiratory bacteria filter Before using an expiratory filter with nebulization, review the safety information in Section 1.5.6. An expiratory filter is not required on the HAMILTON-G5, but you may use one according to your institution’s protocol. It is not required because the expiratory valve design prevents internal ventilator components from coming into contact with the patient’s exhaled gas. If you use an expiratory filter, place it on the patient side of the expiratory valve cover. Monitor closely for increased expiratory circuit resistance. An Exhalation obstructed alarm may also indicate excessive expiratory circuit resistance. If the Exhalation obstructed alarm occurs repeatedly, remove the expiratory filter immediately. If you otherwise suspect increased expiratory circuit resistance, remove the expiratory filter or replace the filter to eliminate it as a potential cause.
Heat and moisture exchanging filter (HMEF) The HMEF is a passive humidification component used together with a bacteria filter. Use an HMEF when ventilating with a coaxial breathing system.

3 When using coaxial breathing sets, follow the manufacturer’s recommendations for each patient group. 4 When tracheal tube ID > 4 mm.
60

English | 624074/13

3 Assembling the patient breathing circuit

3.4.4 Assembling the patient breathing circuit
Assemble the appropriate breathing circuit for your patient. Commonly used standard breathing circuit configurations are illustrated in Section 2.2.3. For neonatal ventilation, see Chapter 6.
3.4.4.1 Connecting the flow sensor
NOTICE To prevent inaccurate flow sensor readings, make sure the flow sensor is correctly connected.
Before proceeding, review the safety information in Chapter 1.
To connect a flow sensor to the breathing circuit
1. Insert a flow sensor into the breathing circuit in front of the patient connection. See also the breathing circuit diagrams in Section 2.2.3.
Flow sensor connection dual limb circuit, Y-piece

The blue tube attaches to the blue connection port. The clear tube attaches to the silver connection port. 3. Calibrate the flow sensor and perform the Tightness test. See Section 5.4.
3.4.5 Positioning the breathing circuit
NOTICE · To prevent water accumulation in the
flow sensor and tubing, position the flow sensor tubing on top of the flow sensor. · Ensure there is no undue stress placed on any tubing or cables.
After assembly, position the breathing circuit so that the hoses will not be pushed, pulled, or kinked as a result of a patient’s movement, transport, or other activities, including scanner bed operation and nebulization. The next step is to perform all required tests, calibrations, and the preoperational check. See Chapter 5.

Flow sensor connection coaxial circuit

2. Attach the blue and clear tubes to the flow sensor connection ports on the ventilator (Figure 3-3).

Hamilton Medical | HAMILTON-G5 Operator’s Manual

61

3 Preparing the ventilator

3.5 Setting up esophageal/ transpulmonary pressure monitoring
The Paux port allows you to use pressure readings other than airway pressure (Paw), for example, from an esophageal balloon catheter, for monitoring purposes. Transpulmonary pressure is also calculated using a combination of the Paw and Paux pressures.
To display Paux-related parameters 1. Connect an esophageal catheter to
the Paux port on the front of the ventilator (Figure 2-3). 2. Open the Monitoring > Paw/Paux window. 3. Touch the Pes (Paux) button to activate Paux as the standard pressure input. To revert to using airway pressure, touch the Paw button. The associated pressure-related parameters are available in the Monitoring window. For details, see Section 8.5.
3.6 Turning the ventilator on and off
To turn on the ventilator 4 Press the Power button on the back of
the ventilator. The ventilator runs a self-test. After a short time, the Standby window is displayed. Proceed with setting up the ventilator and patient, as appropriate.

Figure 3-5. Power button
To turn off the ventilator Note that while using IntelliCuff, you must first deflate the cuff and turn off the device before turning off the ventilator. 1. From active ventilation, press the
Standby key to open the Activate Standby window. 2. Touch the Activate standby button to confirm. 3. Press the power button on the back of the ventilator. The ventilator turns off. In the event of a technical fault or the device will not turn off 4 Press and hold the Power button (Figure 2-4) on the back of the device for about 10 seconds to turn off the ventilator.

62

English | 624074/13

4
Setting up external devices and sensors
4.1 Overview…………………………………………………………………………….. 64 4.2 Installing a module……………………………………………………………….. 64 4.3 Setting up a humidifier………………………………………………………….. 64 4.4 Setting up the IntelliCuff cuff pressure controller……………………….. 65 4.5 Setting up CO2 monitoring ……………………………………………………. 66 4.6 Setting up SpO2 monitoring…………………………………………………… 70 4.7 Enabling sensors ………………………………………………………………….. 70 4.8 Setting up nebulization …………………………………………………………. 71 4.9 Connecting to an external patient monitor or other device ………….. 72
63

4 Setting up external devices and sensors

4.1 Overview
The HAMILTON-G5 supports a variety of external devices and sensors for ventilation, including: · Humidifier · IntelliCuff cuff pressure controller · CO2 monitoring sensors · Pulse oximetry (SpO2 monitoring)
sensors · Nebulizers
This chapter describes how to set them up for ventilation.
4.2 Installing a module
For SpO2 and CO2 sensors, and Aerogen nebulizer use, the associated option module must be installed. An additional HAMILTON-H900 humidifier module is also available.
To install a module 1. If present, remove the cover plate
from the module slots. 2. Slide in the module until it clicks into
place.

Figure 4-1. Sensor, nebulizer, and humidifier connection modules

1 Release button

2 Connection modules

To remove a module 1. Press the release button on top of the
module, and pull the module out. 2. If desired, replace the module slot
cover.
4.3 Setting up a humidifier
Before proceeding, review the safety information in Chapter 1. When used with the HAMILTON-H900 humidifier, the ventilator supports integration of humidifier operation and data monitoring directly from the ventilator display5. Other humidifiers are supported, without the integration. To connect a non-Hamilton Medial humidifier, refer to the manufacturer’s instructions for use.

5 Not available in all markets.
64

English | 624074/13

4 Setting up the IntelliCuff cuff pressure controller

To connect the HAMILTON-H900 humidifier to the ventilator 1. First, enable the Humidifier option on
the ventilator, and ensure either a COM port is configured for the humidifier or the Humidifier module is installed. For details, see Sections 14.11.3 and 14.6.3. 2. Connect the HAMILTON-H900 humidifier power cable to the dedicated power socket on the ventilator (Figure 2-4). 3. Connect a potential equalization cable to the humidifier and to a grounding socket at your facility. 4. Connect the communication cable to the bottom of the humidifier, and to the ventilator. On the ventilator, you can either connect the cable to the configured RS-232 COM port on the back of the ventilator (option 1 below) or to the Humidifier module on the front (option 2 below), whichever is available. Figure 4-2. Connecting the humidifier communication cable to COM port (1) or to module port (2)
If data export is configured, humidifier data is also transmitted from the ventilator to an external monitoring system.

For additional details about: · Connecting the humidifier to the
breathing circuit, see Section 2.2.3. · Working with the humidifier, see the
HAMILTON-H900 Instructions for use. · Controlling the humidifier from the
ventilator, see Chapter 12.
4.4 Setting up the IntelliCuff cuff pressure controller
The ventilator supports the use of an optional IntelliCuff cuff pressure controller, and offers integrated operation and monitoring of the device. For details on using IntelliCuff during ventilation, see Section 12.2. The following options are available: Integrated and standalone Integrated IntelliCuff The IntelliCuff port on the front of the ventilator connects inside the ventilator to an integrated automatic cuff pressure controller module. The integrated cuff controller comprises a small pump and pressure monitoring device with two independent pressure sensors. When in use, the cuff controller increases the cuff pressure as needed, compensates for leaks, and reduces any excess pressure, if required. To aid with intubation and extubation, the cuff controller generates a small vacuum to completely deflate the cuff. For setup details, see Section 4.4.2.

Hamilton Medical | HAMILTON-G5 Operator’s Manual

65

4 Setting up external devices and sensors

Standalone IntelliCuff IntelliCuff is connected as a standalone device, and all controls and operations are available on the device itself. For details on using IntelliCuff as a standalone device, see the IntelliCuff Instructions for use.

Figure 4-3. Connect IntelliCuff tubing to IntelliCuff port on ventilator (1)

4.4.1 About the IntelliCuff tubing The IntelliCuff connector allows connection only from the ventilator end (with the shut-off valve) of the Hamilton Medical cuff pressure tubing. The ventilator end of the tubing has a built-in shut-off valve, which prevents loss of cuff pressure in the event of a disconnection from the ventilator. The patient end of the tubing fits the connector (pilot balloon) for cuff pressure measurement on the ET tube or the tracheotomy cannula.
4.4.2 Setting up IntelliCuff For each patient, you connect the cuff and tubing to the patient and to the ventilator, and specify the desired settings.
To connect the cuff tubing 1. Connect the cuff tubing to the patient
as described in the IntelliCuff Instructions for use. 2. Connect the other end of the cuff tubing to the IntelliCuff port on the front of the ventilator (Figure 2-3).

To enable the IntelliCuff option on the ventilator, see Section 14.11.3. For operation details, see Section 12.2 and the IntelliCuff Instructions for use.
4.5 Setting up CO2 monitoring
Before proceeding, review the safety information in Chapter 1. CO2 monitoring data is helpful for the assessment of the patient’s airway integrity or ensuring proper endotracheal tube placement, among other applications. Two CO2 measurement options are available: mainstream and sidestream. Which option you use depends on the clinical setting.6 Enabling CO2 measurement on the ventilator requires enabling the CO2 hardware (in Configuration) and enabling the sensor. In addition, the CO2 module must be installed.

6 The volumetric capnogram is only available when using a mainstream CO2 sensor.
66

English | 624074/13

4 Mainstream CO2 measurement

Table 4-1. CO2 measurement overview

For details about … Mainstream CO2 measurement, connection, and use Sidestream CO2 measurement, connection, and use Enabling the CO2 hardware
Installing a module Enabling the CO2 sensor

See … Section 4.5.1
Section 4.5.2
Section 14.11.3 Section 4.2 Section 4.7

Figure 4-4. Mainstream CO2 monitoring components and assembly

4.5.1 Mainstream CO2 measurement The CO2 monitoring option comprises the following components (shown in Figure 4-4): communication module, airway adapter, and CO2 sensor. The sensor generates infrared light and beams it through the airway adapter to a detector on the opposite side. CO2 from the patient, flowing through the mainstream airway adapter, absorbs some of this infrared energy. The system determines the CO2 concentration in the breathing gases by measuring the amount of light absorbed. The ventilator displays CO2 measurements as numeric values, waveforms, trends, and loops.

1 Communications module with CO2 connection port

3 CO2 sensor

2 Airway adapter

Hamilton Medical | HAMILTON-G5 Operator’s Manual

67

4 Setting up external devices and sensors

4.5.1.1 Connecting the mainstream CO2 sensor
CAUTION When using active humidification, prevent water accumulation in the CO2 adapter/sensor by ensuring that they are positioned at a 45° angle relative to the floor. Excess water can affect the sensor measurements.

Do not place the airway adapter between the ET tube and the elbow, as this may allow patient secretions to accumulate in the adapter.7 The sensor cable should face away from the patient. 5. Secure the sampling line safely out of the way.
Figure 4-5. Connecting CO2 sensor/adapter (1) to breathing circuit (Adult/Ped shown)

NOTICE You must use an appropriate adapter to connect the mainstream CO2 sensor to a neonatal flow sensor.
Before proceeding, review the safety information in Chapter 1. Ensure the CO2 sensor and adapter are clean and dry before connection.
To set up mainstream CO2 monitoring 1. Connect the sensor cable to the CO2
connection port (1) on the ventilator (see Figure 4-4). 2. Attach the CO2 sensor (3) to the airway adapter (2), aligning the arrows on both components. Press the components together until they click. 3. When connecting a CO2 sensor for the first time, perform the zero calibration of the sensor/adapter, if needed, as described in Section 5.4.5. 4. Connect the sensor/adapter to the breathing circuit proximal to the patient, in a vertical position. See Figure 4-5.

To verify the quality of the connection 4 Check the capnogram (CO2 wave-
form) on the ventilator display. If CO2 levels are higher than expected, check the patient condition. If you determine that the patient’s condition is not contributing, calibrate the sensor (Section 5.4.5).
To remove the sensor cable 4 Pull back on the connector sheath and
disengage from the connection port on the ventilator.

7 You can connect the CO2 sensor in front of or behind the flow sensor according to your institution’s protocol.

68

English | 624074/13

4 Sidestream CO2 measurement

4.5.2 Sidestream CO2 measurement The LoFlo CO2 module is a sidestream CO2 monitoring system comprising the following components: communication module, airway sampling adapter, and CO2 module. See Figure 4-6. The module generates infrared light and beams it through the sample cell to a detector on the opposite side. CO2 from the patient that is aspirated into the sample cell absorbs some of this energy. The system uses a sampling rate of 50 ml/min. The system determines CO2 concentration in the breathing gases by measuring the amount of light absorbed by these gases. The ventilator displays CO2 measurements as numeric values, waveforms, trends, and loops.
Figure 4-6. Sidestream CO2 monitoring components and assembly

4.5.2.1 Connecting the sidestream CO2 sensor
Before proceeding, review the safety information in Chapter 1.
To set up CO2 sidestream monitoring 1. Connect the CO2 module cable to the
CO2 connection port (1) on the ventilator (see Figure 4-6). 2. Insert the sample cell (4) into the CO2 module (2) as shown in Figure 4-6. It clicks into place. Inserting the sample cell into the module automatically starts the sampling pump. Removing the cell turns the pump off. 3. Perform the zero calibration of the adapter, if necessary, as described in Section 5.4.5 before connecting it to the breathing circuit. 4. Connect the adapter between the inspiratory limb and the flow sensor (or between the inspiratory limb and HMEF, if used). See Figure 4-7. The sampling line should face away from the patient. 5. Secure the sampling line safely out of the way.
Figure 4-7. Connecting CO2 adapter (1) to the breathing circuit

1 Communications module with CO2 connection port
2 CO2 module
3 Airway adapter

4 Sampling cell
5 Connecting sampling cell to module

Hamilton Medical | HAMILTON-G5 Operator’s Manual

69

4 Setting up external devices and sensors

To remove the sample cell 1. Remove the airway adapter from the
breathing circuit. 2. Press down on the locking tab and
remove the sample cell from the CO2 module.

4.6 Setting up SpO2 monitoring
The HAMILTON-G5 supports input of SpO2 and related pulse oximetry data, and provides integrated monitoring and data display. Enabling SpO2 measurement on the ventilator requires enabling the SpO2 hardware (in Configuration) and enabling the sensor(s). Table 4-2. SpO2 measurement overview

For details about …

See …

Activating the SpO2 hard- Section 14.11.3 ware

Installing a module

Section 4.2

Enabling the SpO2 sensor(s)

Section 4.7

Working with SpO2 data

Pulse Oximetry Instructions for Use

4.7 Enabling sensors
Before proceeding, review the safety information in Chapter 1. In addition to hardware activation for CO2 and SpO2 measurement (Section 14.11.3, the O2, CO2, and/or SpO2 sensors must be individually enabled for monitoring data to be available.
To enable sensor monitoring 1. Open the System > Sensors on/off
window. 2. Select the appropriate checkboxes
(O2, CO2, SpO2 left, SpO2 right) to enable/disable the monitoring functions, as desired. The ventilator always enables O2 monitoring upon restart. Figure 4-8. System > Sensors on/off window

1 System 2 Sensors on/off

3 O2, CO28, SpO2 left8, SpO2 right8

8 If the option is installed and activated.
70

English | 624074/13

4 Setting up nebulization

4.8 Setting up nebulization
The HAMILTON-G5 supports the following nebulizer types: · Pneumatic · Aerogen§, 9, 10
This section describes how to connect and set up the nebulizer for use. Nebulizer and operation details are provided in Section 10.7.

To connect a pneumatic nebulizer to the breathing circuit set 1. Connect the nebulizer as shown in
Figure 4-9. 2. Connect the nebulizer tubing to the
ventilator Nebulizer port (Figure 2-3).
Figure 4-9. Connecting a pneumatic nebulizer

4.8.1 Setting up a pneumatic nebulizer Setting up and using a pneumatic nebulizer comprises the following steps:
Table 4-3. Nebulizer setup and use overview

To …

See …

Enable or disable volume compensation in Configuration. By default, enabled.

Section 14.7

Connect the nebulizer to the breathing circuit and ventilator, and set it up for use.

This section

Configure duration and breath cycle synchronization settings, and start nebulization.
Information about supported nebulizers and their operation is also provided.

Section 10.7

1 Breathing circuit (coaxial shown)
2 Nebulizer

3 Nebulizer tubing to ventilator
4 Flow sensor

For additional details, refer to the manufacturer’s instructions for use.

9 Not available in all markets. 10 If the option is installed and activated.

Hamilton Medical | HAMILTON-G5 Operator’s Manual

71

4 Setting up external devices and sensors

4.8.2 Setting up an Aerogen nebulizer Before proceeding, review the safety information in Chapter 1. The HAMILTON-G5 supports the use of an Aerogen nebulization system11. The system comprises the Aerogen module and connection port on the ventilator (Figure 2-3), and the Aerogen Solo or Aerogen Pro nebulizer. Setting up and using an Aerogen nebulizer comprises the following steps:

To …

See …

If not installed, install the Aerogen module.

Section 4.2

In Configuration, enable the Aerogen option.

Section 14.7

Connect Aerogen to the breathing circuit and the ventilator, and set it up for use.

Aerogen Solo/ Aerogen Pro Instructions for Use

Configure duration and breath cycle synchronization settings, and start nebulization.
Information about supported nebulizers and their operation is also provided.

Section 10.7

4.9 Connecting to an external patient monitor or other device
You can connect the ventilator to a patient monitor, PDMS, computer, or distributed alarm system using the communication ports on the ventilator. For details, see the Communication Interface User Guide, available on MyHamilton. By connecting the ventilator to a distributed alarm system, you can activate global AUDIO OFF for most alarms for an unlimited period of time. For details, see Section 9.5.

11 If the option is activated.
72

English | 624074/13

5
Specifying ventilation settings
5.1 Process overview ………………………………………………………………….. 74 5.2 Selecting the patient group ……………………………………………………. 74 5.3 Entering patient data ……………………………………………………………. 75 5.4 Performing the preoperational check, tests, and calibrations ……….. 76 5.5 Selecting the ventilation mode ……………………………………………….. 84 5.6 Setting alarm limits ………………………………………………………………. 91 5.7 Starting ventilation……………………………………………………………….. 93 5.8 Stopping ventilation ……………………………………………………………… 93 5.9 About the control parameters ………………………………………………… 93
73

5 Specifying ventilation settings

5.1 Process overview
This section explains how to set up the HAMILTON-G5 for ventilation on an individual patient. Setting up ventilation generally comprises the following steps, each of which is described in this chapter: · Selecting the patient group · Specifying patient data · Performing the preoperational check,
including: ­ Performing a breathing circuit tightness test ­ Calibrating the flow sensor, O2 sensor, and zero calibration of the CO2 sensor · Testing alarms · Selecting the ventilation mode · Reviewing and adjusting control settings · Reviewing and adjusting alarm limits

5.2 Selecting the patient group
Before proceeding, review the safety information in Chapter 1. The HAMILTON-G5 supports the following patient groups: Adult, Pediatric, and Neonatal. Table 5-1. Patient groups

Adult
Sex: M, F
Height: 130 to 250 cm
IBW: 30 to 139 kg
Minimum delivered tidal volume: 100 ml

Pediatric

Neonatal

Sex: M, F
Height: 30 to 150 cm
IBW: 3 to 42 kg
Minimum delivered tidal volume: 20 ml

Weight: 0.2 to 30 kg
Minimum delivered tidal volume: 2 ml

To select the patient group and initial settings
4 For a new patient, touch the desired patient group tab in the Standby window (Figure 5-1): ­ Adult ­ Pediatric ­ Neonatal Touch Last patient to reuse the last active ventilator parameters. The selected patient group appears under the mode name (Figure 2-6).
The settings saved with the selected patient group are loaded and displayed (Section 5.2.1), in addition to the default patient sex/height/IBW (Adult/Pediatric) or weight (Neonatal).

74

English | 624074/13

5 About system defaults: pre-configured settings

Figure 5-1. Standby window

The Default setups are defined in Configuration (Chapter 14).
5.3 Entering patient data

1 New patient, Last patient tabs
2 Patient groups
3 Preop check

4 Gender/height/ IBW (or Weight for Neonatal) for selected default
5 Start (When Hi Flow O2 is selected: Start therapy)
6 INTELLiVENTASV12

5.2.1 About system defaults: pre-configured settings For each of the patient groups, a different default configuration can be defined. During patient setup, you can then quickly pre-configure the ventilator according to your standard protocols, and modify settings as needed. Each Default setup defines a ventilation mode, mode control settings, graphic display selection, and O2 enrichment and nebulizer settings.
12 Not available in all markets.
Hamilton Medical | HAMILTON-G5 Operator’s Manual

CAUTION Entering the correct patient data ensures safe ventilation settings for start up and apnea backup. Before proceeding, review the safety information in Chapter 1. Specifying the correct patient data is particularly important, as the ventilator uses this data as a basis for some calculations and initial mode control settings. · For the Adult and Pediatric patient
groups, the ventilator uses sex and patient height to calculate the ideal body weight (IBW). · For Neonatal patients, the ventilator uses the patient body weight. To enter patient data 4 In the Standby window: Adult/Pediatric. Specify the patient sex and height. The device calculates the patient IBW. Neonatal. Specify the patient weight.
75

5 Specifying ventilation settings

5.4 Performing the preoperational check, tests, and calibrations
The tests and calibrations described in this section help verify the safety and reliability of the ventilator. If a test fails, troubleshoot the ventilator as indicated or have the ventilator serviced. Make sure the tests pass before you return the ventilator to clinical use. The test results are stored in memory, including when the ventilator is turned off. This allows the ventilator to be checked and kept in storage, ready for use. The audible alarm is paused during calibration. Table 5-2. When to perform tests and calibrations

Test or calibration
CO2 sensor/ adapter zero calibration (mainstream/ sidestream)
Alarm tests

When to perform
Required after connecting a CO2 sensor or when a related alarm occurs. Recommended after switching between different airway adapter types.
As desired

To access tests and calibration functions 1. Do either of the following:
­ Open the System > Tests & calib window. ­ From the Standby window, touch the Preop check button. 2. Touch the button for the desired operation.
Figure 5-2. System > Tests & calib window

Test or calibration
Preoperational check
Flow sensor calibration and tightness test
O2 sensor calibration, if needed

When to perform
Before connecting a new patient to the ventilator.
After connecting a new breathing circuit or component (including a flow sensor).
After installing a new O2 sensor or when a related alarm occurs. Not required with a paramagnetic O2 sensor.

1 System
2 Tests & calib 3 Flow Sensor

4 Tightness (shown uncalibrated)
5 O2 sensor
6 CO2 sensor (shown disabled)

A checkmark indicates the component is calibrated and ready. A red X indicates the calibration was unsuccessful. A box with

76

English | 624074/13

5 Performing the preoperational check

no marks indicates the test/calibration has not been performed. A grayed-out box indicates the CO2 sensor is not enabled.

5.4.1 Performing the preoperational check Before proceeding, review the safety information in Chapter 1.
When to perform Before connecting a new patient to the ventilator. To perform the preoperational check 1. Use a setup as described in Table 5-3. 2. Perform all of the steps in Table 5-4. To ensure that the ventilator functions according to specifications on your patient, we recommend that your test circuit be equivalent to the circuit used for ventilation. Table 5-3. Test breathing circuit setup

Component Breathing circuit Flow sensor
Test lung

Specification Adult/pediatric, ID10 to ID22
Adult/pediatric, with calibration adapter Demonstration lung, 2 liter, with adult ET tube between flow sensor and lung

If using heliox, follow the preoperational steps described in Table 5-5.

Table 5-4. Preoperational check

Do or observe…

Verify …

1 Connect ventilator to primary power and an oxygen supply.

2 Assemble the

The breathing

patient breath- circuit is assembled

ing circuit.

correctly.

3 Turn on the ventilator.

During the self test, the alarm lamp is lit red and the buzzer sounds briefly.

4 With the ventila- The System > Tests

tor in Standby, & calib window

touch Preop

opens.

check in the

Standby

window.

5 Perform the tightness test.

The test passes. See Section 5.4.2.

6 Calibrate the flow sensor.

The calibration is successful. See Section 5.4.3.

7 If necessary, run The calibration is

the O2 sensor successful. See

calibration.

Section 5.4.4.

8 If necessary, run The zero calibration the CO2 sensor is successful. See zero calibration. Section 5.4.5.

9 Generate test alarms.

The corresponding alarm message is displayed in the message bar. See Section 5.4.6.
Note that patient alarms are suppressed in Standby.

Hamilton Medical | HAMILTON-G5 Operator’s Manual

77

5 Specifying ventilation settings

Table 5-5. Preoperational check with Heliox

Do or observe…

Verify …

1 Connect ventilator to primary power, Heliox, compressed air, and oxygen supplies.

2 Assemble the

Breathing circuit is

patient breath- assembled correctly.

ing circuit.

3 Turn on the ventilator.

During the self test, the alarm lamp is lit red and the buzzer sounds briefly.

4 Select Air as the See Section 5.4.6.1. gas source, disconnect the air supply, and generate an Air
supply failed alarm.

5 Select Heliox as See Section 5.4.6.1. the gas source, disconnect the Heliox supply, and generate a
Heliox supply failed alarm.

6 Select the gas source to use for ventilation.

7 Perform the tightness test.

The test passes. See Section 5.4.2.

8 Calibrate the flow sensor.

The calibration is successful. See Section 5.4.3.

9 If necessary, cal- The calibration is

ibrate the O2

successful. See

sensor.

Section 5.4.4.

10 If necessary, run The zero calibration the CO2 sensor is successful. See zero calibration. Section 5.4.5.

78

Corrective action A checkmark indicates the component is calibrated and ready. A red X indicates the calibration was unsuccessful. If the ventilator does not pass the preoperational check, have it serviced.
5.4.2 Performing the breathing circuit tightness test Before proceeding, review the safety information in Chapter 1. This test checks for leakage in the patient breathing circuit.
When to perform After installing a new or decontaminated breathing circuit or component (including a flow sensor). To perform the tightness test 1. Set up the ventilator for ventilation,
complete with breathing circuit and flow sensor. 2. In the System > Tests & calib window, touch Tightness. The text Disconnect patient is now displayed. 3. Disconnect the breathing circuit at the patient side of the flow sensor. Do not block the open end of the flow sensor. The text Tighten patient system is now displayed.
English | 624074/13

5 Calibrating the adult/pediatric flow sensor

4. Block the opening (wearing a glove is recommended).

If the problem still persists, have the ventilator serviced.

The text Patient system tight is now displayed. 5. Connect the patient. 6. When the test is complete, verify that there is a checkmark in the Tightness checkbox.
To cancel the test while it is in progress 4 Touch Tightness again.
In case of test failure If the test fails, a red X is displayed in the Tightness checkbox. Perform the following checks, repeating the tightness test after each one, until the test is successful: · Check the breathing circuit for a dis-
connection between the ventilator and the flow sensor, or for other large leaks (for example, breathing circuit, humidifier). · Check that the flow sensor and expiratory valve set are properly seated. · If the calibration still fails, replace the flow sensor. · If the calibration still fails, replace the expiratory valve membrane. · If the calibration still fails, replace the expiratory valve set. · If the calibration still fails, replace the breathing circuit.

5.4.3 Calibrating the adult/pediatric flow sensor This calibration checks and resets the calibration points specific to the flow sensor in use, and measures circuit resistance compensation. Ensure you are using the correct flow sensor for the selected patient group. If there is a mismatch, calibration fails. For details about calibrating a neonatal flow sensor, see Section 6.2.1.
When to perform After connecting a breathing circuit or component. To calibrate an adult/pediatric flow sensor 1. Set up the ventilator for ventilation,
complete with breathing circuit and flow sensor. 2. In the System > Tests & calib window, touch Flow Sensor. If you have not already disconnected the patient, the message line displays Disconnect patient. 3. Disconnect the patient now.
4. When prompted, attach the calibration adapter to the flow sensor and flip them 180° so the adapter is

Hamilton Medical | HAMILTON-G5 Operator’s Manual

79

5 Specifying ventilation settings

directly connected to the limb (as shown below). 5. When prompted, flip the flow sensor/ adapter 180° again, so the flow sensor is directly connected to the limb, and remove the calibration adapter.

· Check the breathing circuit for a disconnection between the ventilator and the flow sensor, or for other large leaks (for example, breathing circuit, humidifier).
· Check that the flow sensor and expiratory valve set are properly seated.
· If the calibration still fails, replace the flow sensor.
· If the calibration still fails, replace the expiratory valve membrane.
· If the calibration still fails, replace the expiratory valve set.
If the problem persists, have the ventilator serviced.

6. When calibration is complete, verify that there is a checkmark in the Flow Sensor checkbox.
7. When successful, continue with other tests or ventilation.
To cancel an ongoing calibration 4 Touch Flow Sensor again.
In case of calibration failure If the calibration fails, a red X is displayed in the Flow Sensor checkbox. Perform the following checks, repeating the calibration after each one, until calibration is successful: · Ensure that the flow sensor is appropri-
ate for the selected patient group.

5.4.4 Calibrating the O2 sensor The device tests the O2 sensor and resets the calibration points specific to the sensor in use. The galvanic O2 sensor requires approximately 30 minutes startup time to reach stable values. O2 monitoring during this time period may be more variable. We recommend waiting 30 minutes prior to calibrating the O2 sensor. The paramagnetic O2 sensor does not require startup time, and is only calibrated once, upon installation.
To perform O2 sensor calibration 1. Ensure the appropriate gas supplies
are connected to the ventilator. 2. In the System > Tests & calib window,
touch O2 sensor. 3. When calibration is complete, the
message O2 sensor calibration OK is displayed. Verify that there is a checkmark in the O2 sensor checkbox.

80

English | 624074/13

5 Performing a zero calibration on the CO2 sensor/adapter

In case of calibration failure If the calibration fails, a red X is displayed in the O2 sensor checkbox. Perform the following checks, repeating the calibration after each one, until calibration is successful: · Ensure a Hamilton Medical O2 sensor is
installed. · If the second calibration attempt fails,
and you are using a galvanic O2 sensor, replace the sensor. If the problem persists, have the ventilator serviced.
5.4.5 Performing a zero calibration on the CO2 sensor/adapter
CAUTION · Always perform zero calibration with
the CO2 sensor (mainstream) or CO2 module (sidestream) connected to the airway adapter. · Be sure NOT to cover both ends of the airway adapter with your fingers.
Before proceeding, review the safety information in Chapter 1. The CO2 adapter zero calibration compensates for optical differences between airway adapters and for sensor drift. Note that the CO2 sensors are calibrated at the factory; you only need to zero the adapters as described below.

Zero calibration requirements for mainstream CO2 sensors Perform a zero calibration in the following cases:
· With the first use of the sensor
· When changing between airway adapter types (for example, from single use to reusable)
· When the CO2 sensor calibration needed alarm is generated
Zero calibration requirements for sidestream CO2 sensors
You only need to perform a zero calibration with sidestream CO2 sensors when the CO2 sensor calibration needed alarm is generated.
Wait at least 20 seconds — and for best results, 2 minutes — to perform the zero calibration after removing the adapter from the patient’s airway. This time allows any CO2 remaining in the adapter to dissipate.
To perform the zero calibration of the CO2 sensor/adapter (mainstream) and sensor/ module (sidestream)
For reference, see Figure 4-4 for the mainstream CO2 assembly and Figure 4-6 for the sidestream assembly. 1. Connect the CO2 sensor (mainstream)
or the CO2 module (sidestream) to the ventilator, and ensure CO2 monitoring is enabled. Once connected, wait approximately 90 seconds for the device to warm up. 2. Disconnect the CO2 adapter from the breathing circuit. 3. Attach the CO2 adapter to the sensor (mainstream) or plug the sample cell into the CO2 module (sidestream).

Hamilton Medical | HAMILTON-G5 Operator’s Manual

81

5 Specifying ventilation settings

Place these components away from all sources of CO2 (including the patient’s and your own exhaled breath) and the exhaust port of the expiratory valve. 4. In the System > Tests & calib window, touch CO2 sensor. Do not move the components during calibration. 5. When the zero calibration is complete, verify that there is a checkmark in the CO2 sensor checkbox.
In case of zero calibration failure
If the zero calibration fails, a red X is displayed in the CO2 sensor checkbox. Perform the following checks, repeating the zero calibration after each one, until it is successful: · Check the airway adapter and clean if
necessary. · If the zero calibration still fails, ensure
there is no source of CO2 near the airway adapter. · If the zero calibration still fails, connect a new adapter. · If the zero calibration still fails, connect a new CO2 sensor (mainstream) or CO2 module (sidestream).
If the problem persists, have the ventilator serviced.

5.4.6 Testing the alarms The HAMILTON-G5 performs a self-check that verifies proper alarm function during startup. We recommend testing the alarms as part of the preoperational check. For all of the tests, use a demonstration lung assembly as described in Section 5.4.1.
5.4.6.1 Testing the air and Heliox supply failure alarms
To test the Air supply failure alarm 1. In Standby, open the System > Gas
source window, and touch the Air button. 2. Dis

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals