Bonvoisin SF-400D Electronic Analytical Balance User Manual

June 6, 2024
Bonvoisin

SF-400D Electronic Analytical Balance
User Manual

Bonvoisin SF 400D Electronic Analytical Balance

Any Questions, contact us!
Email: [email protected]
Whatsapp/Wechat: +8615958157037

Specifications

Capacity: 600g
Accuracy: 0.01g
Units: g/oz/ct/dwt/T/gn
Auto Power off 270 seconds without operation
Power Mode: AC adapter/2pcs 1.5V AA batteries
Operation Temperature: 5-35℃
Display: LCD 5 digits

Couting Function ( PCS )

  1. Put the relevant amount of goods that you are going to count as the sample ( 25, 50, 75, 100 ).
  2. Press and hold the pcs key for two seconds, then you will see25. Press the MODE key and you will see ” 25, 50, 75, 100 “. Please select the number that you are willing to put as the sample number. Then press the ” PCS ” key to confirm it.

Now you can put the goods that you want to count on the center of the platform and the LCD will show the actual number of the goods on the platform.

Units Conversion

Press Mode key to converting units from g/oz/ct/dwt/T/gn

Tare Function

  1. Turn on scale
  2. Place the container on the platform
  3. Press ” TARE ” and wait until ” 0.00 ” is displayed.

Calibration

  1. Turn on the scale
  2. press and hold the CAL key until LCD displays 600 then 200 or 500.
  3. Press the TARE button to choose 200g or 500g and then press the MODE button.
  4. Put the calibration weight accordingly and it will display PASS.

Notice
Avoiding any exposure to extreme heat or cold. Dale works better when placed and started at normal room temperature. Always allow the scale to acclimate to normal room temperature for at least one hour before use.
Keep your scale in a clean environment. Dust, dirt moisturevibration, air currents, and a close distance to other electronic equipment can all cause an adverse effect on the reliability and accuracy of your scale.
Handle with care. Gently apply all items to be weighed onto the tray top.
Although this scale is designed to be quite durable, try to avoid rough treatment as this may permanently damage the internal sensor.
Avoid shaking, dropping or other shocking the scale.
This is a precision instrument and must be handled with extreme care.