Smilecare UFR-106 Infrared Forehead Thermometer Instruction Manual

June 5, 2024
Smilecare

Smilecare UFR-106 Infrared Forehead Thermometer Instruction Manual

Smilecare UFR-106 Infrared Forehead Thermometer Instruction
Manual

Introduction

▲Your new Infrared forehead thermometer uses advanced infrared (IR) technology to measure temperature instantly and accurately on the forehead or object.
▲Easy to use and less measurement time This thermometer does not need to contact body or object to ensure the safety and hygiene. Its ergonomic design makes this thermometer be simple and very easy to check the temperature. It only takes 1 second to take measurement and reading.
▲Body and object modes
This thermometer supports to measure temperature of body and object.
Measuring range of object mode is 0.0 to 100.0°C (32.0°F~199.9°F ). That means except to take body temperature, it also supports to take the surface temperature of object as below,
Surface temperature of milk in the bottle;
Surface temperature of baby bath;
Environment temperature.
▲Color visible indication of alarm and alarm sounds
When body temperature is over 37.5°C, color indicator will show red color and an alarm sounds for 10 times.
▲Memory function
It stores up to 20 sets recent measurement data.
▲Auto power off
Automatically power off if left idle for Approx. 60±20s.
▲Measurement
Measurement time interval in 1 second and the measurement distance is within 1~3cm.
▲Intended use :
The Infrared Thermometer is intended for the intermittent measurement and monitoring of human body temperature from forehead or object. The device is indicated for use by people of all ages at homecare and in hospital.
▲Contraindication(s):None.
▲The patient is an intended operator. Object mode is for reference only.

Safety Information

To assure the correct use of the product, basic safety measures should always be followed including the warning and the caution listed in the instruction manual :

Symbol descriptions
The following symbols may appear in this manual, on the label, on the device, or on it’s accessories. Some of the symbols represent standards and compliances associated with the device and its use.

Smilecare UFR-106 Infrared Forehead Thermometer Instruction Manual -
Symbol Smilecare
UFR-106 Infrared Forehead Thermometer Instruction Manual -
Symbol

WARNING:

Do not dispose of electrical appliances as unsorted municipal waste, use separate collection facilities. Contact you local government for information regarding the collection systems available. If electrical appliances are disposed of in landfills or dumps, hazardous substances can leak into the groundwater and get into the food chain, damaging your health and well-being.

Classification

  1. Internally powered equipment;
  2. Type BF applied part; .
  3. Protection against ingress of water or Particulate matter: IP21;
  4.  Not category AP/APG equipment;
  5.  Mode of operation: Continuous operation.

The user must check that the equipment functions safely and see that it is in proper working condition before being used.

Product Structure

– Display

  1. Body temperature indicator
  2. Object temperature indicator
  3. Memory mode
  4. Temperature reading
  5. Temperature unit
  6. Low battery indicator

– Body

  1. Auto-sense Probe
  2. 2-Color LED
  3. Display Screen
  4. SET Button
  5. MEM Button
  6. START/STOP Button
  7. Battery Cover

Battery Installation

Remove the battery cover from the battery compartment, insert the battery.

  1. Remove the battery cover from .the battery compartment as the arrow direction accordingly.
  2. Insert 2 AAA powerful batteries into the compartment and ensure each battery is in the proper direction, Positive (+) and Negative (-) are displayed on the back of battery cover.
  3. Close the battery cover.

Low battery and replacement
When power on, the low battery symbol = will display once the unit start to work, and you must replace with new batteries, otherwise the unit can’ I work.

Battery type and replacement
Please use 2pcs AAA identical 1. 5V alkaline batteries.
Do not use the batteries beyond their expiry date.
Please remove the batteries if you do not need to use for long time.

WARNING
Dispose of the battery in accordance with all federal, state and local laws. To avoid fire and explosion hazard, do not burn or incinerate the battery.

Setting Mode

How To Set

1. Mode setting:

Smilecare UFR-106 Infrared Forehead Thermometer Instruction Manual - Mode
setting

2. Unit setting:
Long press SET Button when power on after you hear 1 beep, the screen will display t or °F , Mode will change between t and °F .when the SET Button is pressed. It will confirm the Celsius Unit automatically when you switch to t , and confirm the Fahrenheit Unit automatically when you switch to °F . Then press START/STOP Button make the confirmation selection, and after the selection is completed, it will automatically enter the sound switch setting mode.

3. Sound switch setting:
Enter the sound switch setting mode, 0n or DFF is displayed on the screen and flashes, press SET Button to select and press START/STOP Button to confirm selection.

Proper Use Of The Unit

Pre-measurement

About Normal Body Temperature & Fever The temperature in the forehead and temple area differs from the internal temperature, which is taken orally or rectally. Vasoconstriction, an effect which constricts the blood vessels and cools the skin, can occur during the early stages of a fever. In this case, the temperature measured by the Infrared thermometer may be unusually low. If the measurement therefore does not match the patient’s own perception or is unusually low, re peat the measurement every 15 minutes. As a reference, you can also measure the internal body temperature using a conventional oral or rectal thermometer. Body temperature can vary from one individual/person to next. It also varies by location on the body and lime of day. Below shows the statistical normal ranges from different sites. Please keep in mind that temperatures measured from different sites, even at the same time, should not be directly compared. Fever indicates that the body temperature is higher than normal. This symptom may be caused by infection, overdressing or immunization. Some people may not experience fever even when they are ill. These include, but are not limited to, infants younger than 3 months old, persons with compromised immune systems, persons taking antibiotics, steroids or antipyretics (aspirin. ibuprofen, acetaminophen), or persons with certain chronic illnesses. Please consult your physician when you feel ill even if you do not have fever.

*Table 1 Body Site Normal Temperature Range**

Smilecare UFR-106 Infrared Forehead Thermometer Instruction Manual - Table
1

Note: Body Temperature at WebMD;
website: http://firstaid.webmd.com/body-temperature;retrieved at 2010 Jan 7.

As a Body Thermometer

  1. Press the START/STOP Button, all symbols appear on the display, you can hear 1 short beep, then unit °C/°F on the display start to flash.
  2. Press the SET Button to select Body Mode ,the temperature unit flashes.
  3. Move the probe close to the forehead make sure the probe is flat and take measurement with a distance between 1-3cm. Press the START/STOP Button, The meter will take measurement.
  4. The measurement will be finished in 1 second, when it has been completed you can hear 1 short beep, in the meantime, the reading will appears on the display with green LED blinks.

Note:

  • If the readings <37.5°C(99.5°F ), the display will show together with green LED and 1 short beep.
  • If the reading is;;e37.5°C(9g_5°F)and<43°C(109.4°F), the display will show together with red LED and 10 short beeps.
  • Only after the unit °C/°F start lo flash again, ii can continue lo take measurement.
  • As the forehead measurement temperature is likely to be affected by sweat, oil and the surroundings, the reading shall be taken as a reference only.
  • If the probe is placed at an angle close to the forehead measurement, the reading will be affected by surrounding temperature.
  • Babies’ skin reacts very quickly in the ambient temperature. Therefore, do not take their temperature with the non-contact thermometer during/after breastfeeding, because the skin temperature maybe lower than the internal body temperature.

As a Object Thermometer

  1. Press the START/STOP Button, all symbols appear on the display , you can hear 1 short beep, then unit °C/°F on the display start to flash.
  2. Press the SET Button to select Object Mode, the temperature unit flashes.
  3. Move the probe close to the object, make sure the probe is flat and take measurement with a distance between 1-3cm. Press the START/STOP Button, The meter will take measurement.
  4. The measurement will be finished in 1 second, when ii has been completed, you can hear 1 short beep, in the meantime, the reading will appears on the display with green LED blinks.

Note:

  • As the object scan temperature is likely to be affected by the surroundings, the reading shall be taken as a reference only.
  • If the probe is placed at an angle close the object scan, the reading will be affected by surrounding temperature.

Memory-recall of measurements
This Non-contact forehead thermometer automatically stores 20 sets measurements value, the oldest record will be replaced by the latest measurement value when more than 20 sets.

Read memory record
Press the MEM Button to enter the memory mode when power off. Each time you press the MEM Button, a number (from 1 to 20) will be displayed along with the symbol [M], 1 second later, the measurement will be shown, as well as subsequent measurements can be display one after the other by pressing the MEM Button each time.

Note:
The reading number “1” is the latest measurement, “20” is the oldest measurement. When the oldest measurement is read, pressing the MEM Button again will read the reading “1” again.

Smilecare UFR-106 Infrared Forehead Thermometer Instruction Manual - Memory
recall of measurements

Memory-clear Of Measurements

If you are sure that you want to remove all stored memories permanently. Long press the MEM Button for 10 seconds until CLr appears when power off, CLr will flashes for 5 times to clear all the memories along with 5 short beeps.

Exceptional Situation

Smilecare UFR-106 Infrared Forehead Thermometer Instruction Manual -
Exceptional Situation Smilecare UFR-106 Infrared Forehead Thermometer
Instruction Manual - Exceptional Situation

Please contact the distributor if you can not solve the problem, do not disassemble the unit by yourself!

Care And Maintenance

Care for the main unit

  • Keep the unit in the storage case when not in use.
  • Clean the unit with a soft dry cloth. Do not use any abrasive or volatile cleaners.
  • Never immerse the unit or any of its’ component in water.

Maintenance

Smilecare UFR-106 Infrared Forehead Thermometer Instruction Manual -
Maintenance

Note: We will not be responsible for any quality problem if you do not care and maintain the product as instructed.

Specification

Clinical Measurement Accuracy And Safety Verification:

The product has passed clinical trials. The measured results of the infrared forehead thermometer was compared with the measured results of mercury thermometers, the deviation average. lit = 0.011 c not exceeding 0.3°C; the clinical repeatability of the infrared forehead thermometer SR =0.100°C, not exceeding 0.3°C. The measured results up to the laboratory standard and the clinical standard. Therefore, the deviation average and the clinical repeatability of the infrared forehead thermometer are complied with the regulatory requirement ISO 80601-2-56. The conclusions are drawn from the clinical trials, the accuracy and safety are complied with the regulatory requirement.

Warranty Information

▲ The unit is guaranteed to be free of defects in workmanship and materials under normal use for a period of 1 Year from the dale of purchase.
▲ For repair under this warranty. Our authorized service agent must be advised of the fault within the period of the warranty. This warranty only covers parts and labor service under normal operations. Any defect resulting from natural causes, eg. flood, hurricane etc, is not covered in this guarantee. This guarantee also does not cover damage incurred by use of the unit Nolin accordance with the instructions, accidental damage, or being tampered or serviced by unauthorized service agents.
▲ The flowing will be excluded from this warranty-if the thermometer has been misused, abused, or neglect in following the manual’s instructions on purpose and unauthorized repair or modifications.
▲ The device requires no calibration.
▲ The device is not repairable and contains no user serviceable parts.

EMC

IEC 60601-1-2:2014 ME EQUIPMENT and ME SYSTEMS identification, marking and documents for Class B product Instructions for use The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments and so on.
The ME EQUIPMENT or ME SYSTEM is suitable for home healthcare environments and so on.
Warning: Don’t near active HF surgical equipment and the RF shielded room of an ME system for magnetic resonance imaging, where the intensity of EM disturbances is high.
Warning: Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
Warning: Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.”
Warning: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the infrared thermometer, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
If any: a list of all cables and maximum lengths of cables (if applicable), transducers and other ACCESSORIES that are replaceable by the RESPONSIBLE ORGANIZATION and that are likely to affect compliance of the ME EQUIPMENT or ME SYSTEM with the requirements of Clause 7 (EMISSIONS) and Clause 8 (IMMUNITY). ACCESSORIES may be specified either generically (e.g. shielded cable, load impedance) or specifically (e.g. by MANUFACTURER and EQUIPMENT OR TYPE REFERENCE).
If any: the performance of the ME EQUIPMENT or ME SYSTEM that was determined to be ESSENTIAL PERFORMANCE and a description of what the OPERATOR can expect if the ESSENTIAL PERFORMANCE is lost or degraded due to EM DISTURBANCES (the defined term “ESSENTIAL PERFORMANCE” need not be used).

Technical description

  1. All necessary instructions for maintaining BASIC SAFETY and ESSENTIAL PERFORMANCE with regard to electromagnetic disturbances for the excepted service life.
  2. Guidance and manufacturer’s declaration -electromagnetic emissions and Immunity

Smilecare UFR-106 Infrared Forehead Thermometer Instruction Manual - Table
2 Smilecare UFR-106
Infrared Forehead Thermometer Instruction Manual - Table
3

Table 3

Smilecare UFR-106 Infrared Forehead Thermometer Instruction Manual - Table
4

Jiangsu Konsung Bio-Medical Science And Technology Co.,Ltd
No. 8, Shengchang West Rd, Danyang Development Zone,
Jiangsu Province, China
Tel:86-511-86375968
WhatsApp: 0086-18952012596
E-mail: [email protected]
Website: www.konsungmedical.com

Manufacturer
Shenzhen Urion Technology Co.,Ltd.
Modulation
Floor 4-6th of Building D,Jiale Science&Technology Industrial
Zone,No.3,ChuangWei Road,Heshuikou Community,MaTian Street,
GuangMing New District,518106 ShenZhen,PEOPLE’S REPUBLIC OF CHINA
Tel: (86) -755-29231308
MADE IN CHINA

[EC|REP] Shanghai International Holding Corp. GmbH (Europe)
Eiffestrasse 80, 20537 Hamburg,Gennany
Tel:+49-40-251317 E-mail:[email protected]

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Smilecare User Manuals

Related Manuals