TOMEY TSL-4000Z Z Series Slit Lamp User Manual
- June 5, 2024
- TOMEY
Table of Contents
SLIT LAMP
OPERATOR MANUAL
Z Series
| Consult instructions for use| | General warning sign
---|---|---|---
| Date of manufacture| ****| Warning: Electricity
| Manufacturer’s name and address| | Warning: Floor level obstacle
| Country of manufacture| | Warning: Non-ionizing radiation
| Waste Electrical and Electronic Equipment (WEEE) recycling| | Warning:
Optical radiation
| This way up| | Warning: Hot surface
| Keep dry| | Conformité Européene
| Fragile| | Type B applied part
| Do not use if the package is damaged| | Class II equipment
| Temperature limit| | Atmospheric pressure limitation
| Authorized Representative in the European Community| | Humidity limitation
| Catalog number| | Serial number
| Translation| | Medical device
The Tomey Slit Lamp is designed and built-in in conformity with Directive
93/42/EEC, Regulation (EU) 2017/745 and ISO 13485 Medical Devices Quality
Management Systems.
Classification: CE: Class I
FDA: Class II
The information contained within this manual must not be reproduced in whole
or part without the manufacturer’s prior written approval. As part of our
policy for continued product development we the manufacturer reserve the right
to make changes to specifications and other information contained in this
document without prior notice. Copyright © Tomey GmbH 2021. Published in the
UK in 2021.
INDICATIONS FOR USE
This device is intended to be used only by suitably trained and authorized
healthcare professionals.
CAUTION: Federal Law restricts this device to sale by or on the order of
a physician or practitioner.
Intended use/purpose of the instrument
This Tomey Slit Lamp is an AC-powered slit lamp biomicroscope and is intended
for use in eye examination of the anterior eye segment, from the cornea
epithelium to the posterior capsule. It is used to aid in the diagnosis of
diseases or trauma which affects the structural properties of the anterior eye
segment.
Brief description of the instrument
This Tomey Slit Lamp can either be mounted onto a custom tabletop supplied by
Tomey or can be mounted on a third parties tabletop (refraction unit) by
suitably trained technicians. The Tomey Slit Lamp consists of 5 assemblies;
Illumination Tower; Observation System; XYZ Translation Base; Chinrest
Assembly and a Table Top with power supply and accessory drawer. The light
intensity is controlled by a variable rheostat located on the XYZ Translation
Base. There are a number of selectable filters allowing the ser to control the
characteristics of the examination light. Many ophthalmic instruments
illuminate the eye with intense light. The light intensity on the Tomey Slit
Lamp is continuously adjustable from maximum to zero.
SAFETY
2.1 PHOTOTOXICITY
CAUTION: The light emitted from this instrument is potentially hazardous. The longer the duration of exposure, the greater the risk of ocular damage. Exposure to light from this instrument when operated at maximum intensity will exceed the safety guideline after 81 seconds While no acute optical radiation hazards have been identified for slit lamps, we recommend keeping the intensity of the light reaching the patient’s retina to the minimum possible for the respective diagnosis. Children, people with aphakia, and people suffering from eye conditions are most at risk. The increased risk may also occur if the retina is exposed to the same or a similar device with a visible light source within 24 hours. This applies, in particular, if the retina has been photographed with a fl ash bulb in advance. Tomey GmbH shall on request, provide the user with a graph showing the relative spectral output of the instrument.
2.2 WARNINGS AND CAUTIONS
Please note that the proper and safe functioning of our instruments is only
guaranteed if both the instruments and their accessories are exclusively from
Tomey GmbH. The use of other accessories may result in increased
electromagnetic emissions or reduced electromagnetic immunity of the device
and may lead to incorrect operation. Observe the following prescriptions in
order to ensure the safe operation of the instrument.
WARNINGS
-
Never use the instrument if visibly damaged and periodically inspect it for signs of damage or misuse.
-
Check your Tomey product for signs of transport/storage damage prior to use.
-
Do not use in the presence of flammable gases/liquids, or in an oxygen-rich environment.
-
US Federal Law restricts this device to sale by or on the order of a physician or practitioner.
-
This device is intended to be used only by suitably trained and authorized healthcare professionals.
-
This product should not be immersed in the fluid.
-
Repairs and modifications to the instrument must be made only by the specialized technicians of the manufacturer’s Technical Service Centre or by personnel trained and
authorized by the manufacturer. The manufacturer declines any and all responsibility for loss and/or damages resulting from unauthorized repairs; furthermore, any such actions will invalidate the warranty. -
The power switch and mains plug are the means of isolating the device from the mains supply – ensure both the power switch and mains plug are accessible at all times.
-
Do not position the equipment so that is difficult to press the power switch or remove the mains plug from the wall socket.
-
Refraction stand variants or adaptors should only be used in combination with EN/IEC 60601-1 and EN/IEC 60601-1-2 compliant power supplies and devices.
-
Route power cords safely to eliminate the risk of tripping or damage to the user.
-
**** Before any cleaning of the instrument or the base unit ensure the power lead is disconnected.
-
LEDs can reach high temperatures in use – allow to cool before handling.
-
Do not exceed the maximum recommended exposure time.
-
Should the instrument suffer shocks (for example, should it accidentally fall), and the optical system or the illumination system are damaged it may be necessary to return the instrument to the manufacturer for repair.
-
After removal of the LED, do not touch the Slit Lamp LED electrical contacts and patient simultaneously.
-
The owner of the instrument is responsible for training personnel in its correct use.
-
Ensure the instrument or instrument table is placed on a level and stable surface.
-
Use only genuine Tomey-approved parts and accessories or device safety and performance may be compromised.
-
Shut down after every use. In case the dust cover is used: risk of overheating.
-
For indoor use only (protect from moisture).
-
Electrical equipment can be affected by electromagnetic interference. If this occurs whilst using this equipment, switch the unit off and reposition it.
-
Do not touch accessible connectors and the patient simultaneously.
-
Before use, the Slit Lamp should be allowed to adjust to the ambient room temperature for several hours. This is especially important when the unit has been stored or transported in a cold environment; this can cause severe condensation to develop on the optical elements.
Before use, the Slit Lamp should be allowed to adjust to the ambient room temperature for several hours. This is especially important when the unit has been stored or transported in a cold environment; this can cause severe condensation to develop on the optical elements.
2.3 CONTRAINDICATION
There is no restriction to the patient population this device can be used with
other than those outlined in the contra-indications stated below. Slit lamps
can produce discomfort in some photophobic patients due to the high
illumination. Additionally, patients must be co-operative and capable of
sitting upright for the duration of the examination, therefore the technique
may not be suitable for patients who are unable to sit upright for long
periods of time or those with limited neck and back mobility.
CLEANING AND DISINFECTION INSTRUCTIONS
**** Before any cleaning of the instrument or the base unit, ensure the power
lead is disconnected.
Only manual non-immersion cleaning as described should be used for this
instrument. Do not autoclave or immerse in cleaning fluids. Always disconnect
the power supply from the source before cleaning.
-
Wipe the external surface with a clean absorbent, non-shedding cloth dampened with de-ionized water/detergent solution (2% detergent by volume) or water/isopropyl
alcohol solution (70% IPA by volume). Avoid optical surfaces. -
Ensure that excess solution does not enter the instrument. Use caution to ensure the cloth is not saturated with the solution.
-
Surfaces must be carefully hand-dried using a clean non-shedding cloth.
-
Safely dispose of used cleaning materials.
NAMES OF CONTROLS AND COMPONENTS
Headrest Assembly
-
Fixation light cable
-
Forehead rest band
-
Patient’s eye height marker
-
Fixation light
-
Chinrest
-
Chinrest height adjuster
-
Patient grab handles
TOMEY SLIT LAMP, Z Series
-
Eyepiece assembly securing knob
-
Yellow filter knob (up = out)
-
Illumination prism
-
Slit offset by rotating prism housing
-
Filter control wheel
-
Slit rotation knob
-
Slit width control knob
-
Aperture control wheel
-
Test bar & tonometer plate hole and cover
-
Illumination arm locking knob
-
Microscope arm locking knob
-
Diffuser
-
Joystick base locking knob
-
Joystick control (X Y Z movement)
-
Illumination control rheostat
-
Axle
-
Runner covers
-
Z type tonometer mounting hole
-
Yellow filter knob (up = out)
-
Lock for securing magnification body
-
Magnifi cation change drum
-
Eyepieces – adjustable for PD and dioptre correction
-
Breath shield securing knob
-
Main power switch
-
Power supply unit
-
Glide plate
ASSEMBLY
Your Tomey Slit Lamp has been designed to fit onto an electric insulated
medical table base or onto an electric insulated and fire-resistant medical
tabletop, e.g. a refraction stand or combi unit. Take care when unpacking your
Slit Lamp that you do not accidentally damage or discard any of the contents.
Leave the Slit Lamp in the packing for several hours after delivery before
unpacking to reduce the risk of condensation forming.
Tomey Slit Lamps can be fitted to most Refraction Stands / Combi units. Tomey
advises that this shall be carried out by a suitably trained technician to
ensure performance and safety are not affected.
The Refraction Stand, Combi Unit, or table leg must be IEC 60601-1 compliant.
If you are fitting or have fitted your Slit Lamp to a medical or Tomey table leg/base ensure it is situated on a fi rm and level floor. If the table leg/base has castors ensure the following before moving it to another location:
- The table is at its lowest position.
- The power cord is removed.
- The Slit Lamp arm and base locking knobs are tightened.
- The runner covers are securely located.
- The system is moved by grasping it at its lowest practical point.
5.1 TABLETOP AND BASE ASSEMBLY PROCEDURE
-
Attach the Slit Lamp tabletop to your table leg using the M6 x 20mm screws and washers. Note that the power supply and accessory drawer should face the operator.
The security of the fitment of the tabletop to the table leg is critical for patient and Slit Lamp safety -
Using the wrench provided, fit the headrest assembly to the tabletop using the hex bolts and washers. The headrest assembly locates on the underside of the tabletop. Take care not to over-tighten the hex bolts.
-
Attach the patient grab handles (7) to the headrest assembly.
-
Place the Slit Lamp base on the runners. Ensure that the wheels are in line with each other. Check that the guide wheels are tight.
-
Fit the runner covers to the runners by gently sliding them inwards, towards each other.
5.2 ILLUMINATION TOWER ASSEMBLY PROCEDURE
-
Remove the hex bolt (a) from the base of the Illumination Tower, and then place the Illumination Tower on the Slit Lamp base with the base notch (b) and pin (c) aligned.
Attach the tower to the base using the hex bolt removed earlier and tighten using the wrench provided.
-
Carefully fit the microscope body to the arm – ensuring it is pushed to the stop. Tighten using the securing knob on the side.
-
Attach the breath shield as indicated in the images below.
5.3 CABLE ATTACHMENT PROCEDURE
-
Connect the power cables.
a) Chinrest fixation light cable to the power supply unit.
b) (3 pins) cable from power supply unit to slit lamp base assembly.
c) (4 pins) main lamp cable from the bottom of the illumination assembly to the slit lamp base assembly.
d) Ensure cables are routed to allow free movement of the XYZ base and to be clear of the patients.
If your Slit Lamp was not supplied with a transformer (Part #3020-P-5040TOM), make sure that the power connection is compatible with the specific cations in this manual and is connected by a qualified ed technician to an available and suitable power supply, see section 9.4 Power Supply page 20.
-
If the Tomey Slit Lamp is used with a power supply or cables other than those supplied, this may result in increased emissions or decreased immunity of the Tomey Slit Lamp in relation to EMC performance.
-
Connect the mains power to the Slit Lamp transformer using the power lead provided.
-
To isolate from the main supply unplug the main plug from the wall socket. Ensure that the product is positioned so that it is easily accessible.
**** Only a hospital-grade 3-conductor electrical power supply cable must be used. For USA and Canada: Detachable power supply cord set, UL listed, type SJE, SJT or SJO, 3-conductor, not smaller than 18 AWG. The plug, cable, and ground lead connection of the socket has to be in perfect condition.
5.4 FITTING APPLANATION TONOMETERS, T TYPE, AND R TYPES
Tomey Applanation Tonometer (T-TYPE)
-
Position the guide plate in the tonometer/test bar support hole on the slit lamp.
-
Lift Tonometer out of the packaging and assemble it by inserting the pin on its base into one of the two possible openings (for the right or left eye) on the horizontal guide
plate above the slit lamp axis. These positions are related to the microscope optics and observation can be made either through the right or the left eye- piece. The tonometer will slip easily onto the support plate; stability is assured by the locking pins. -
To obtain an image as clear and as free of reflexes as possible, the angle between the illumination and the microscope should be about 60° and the slit diaphragm should be
fully opened. -
When not in use the Tonometer should be removed from the Slit Lamp and placed securely back in the packaging or a suitable location.
Applanation Tonometer ‘Tomey Fixed’ (R-TYPE)
This instrument is for those who wish the tonometer to remain permanently on the slit lamp. -
Mount the plate for the tonometer onto the microscope body using the securing screw.
-
Mount the Tonometer onto the mounting post ensuring the grub screws are loose.
-
Swing the Tonometer arm forward in front of the microscope for examination. Rotate the Tonometer body until the prism is in the center of the view through the eyepiece. Carefully tighten the two retaining grub screws in turn until the Tonometer is secure on the mount and the prism is in the center of the view.
-
To obtain an image as clear and as free of reflexes as possible, the angle between the illumination and the microscope should be about 60° and the slit diaphragm should be fully opened.
-
When not in use the tonometer arm should be swung up against the protection plate.
INSTRUCTIONS FOR USE
6.1 SETTING THE BINOCULARS
It is vital that the binoculars are optimized for the user’s optical correction in order to obtain focused binocular images.
-
Remove the Test Bar locating hole cover plate (16) and place the test bar focus in the test bar location hole at the base of the microscope arm. To access the location hole first remove the cover. The test bar should be set with the flat projection face towards the Slit Lamp Microscope. The illumination and the microscope should be in the zero degrees position.
-
Turn on the Slit Lamp and set the slit to full width (14), set the magnification to x16 (27).
-
Adjust the eyepiece pupillary distance by holding both eyepiece bodies and rotating them inwards or outwards until they are correct for your PD.
-
Turn both eyepieces (28) to maximum plus (+) correction.
-
Close one eye, and with the other eye look through the microscope slowly turning the open eye eyepiece towards the minus (-) position until the image of the test bar is in focus. Stop.
-
Repeat the above process for the other eyepiece.
-
Make a note of the positions of the eyepieces so that you can set them quickly if the Slit Lamp has been used by another clinician.
-
Note – younger examiners are recommended to compensate for their ability to accommodate by further adjusting the eyepieces by minus one (-1) or minus two (-2) dioptres.
6.2 PREPARING THE PATIENT AND USING THE SLIT LAMP
Parts of the slit lamp coming into contact with the patient should be cleaned in accordance with these instructions prior to the examination. Tomey recommends the use of disposable hygienic chinrest tissues on the Chinrest before the patients place their chin on it. Never use the instrument if visibly damaged and periodically inspect it for signs of damage or misuse.
-
The patient should be as comfortable as possible and with the patient in the chinrest adjust the chinrest height (6) so that the patient’s eyes are level with the height marking
(3) on the chinrest support. -
Focus the eyepieces using the test bar as described earlier, and if you have not already done so set them to your interpupillary distance by holding both eyepiece bodies and
rotating them inwards or outwards until they are correct for your PD. -
Switch on the illumination, making sure the rheostat (21) is set to a low level to minimize the patient’s exposure to light hazard.
-
Rotate the joystick (20) until the light beam is at eye level.
-
Holding the joystick vertical, move the slit Lamp base towards the patient until the slit beam appears focused on the patient’s cornea.
-
Adjust slit width (14), magnification (27), slit rotation (13) and slit angle, etc. as required to perform the examination.
-
To offset the slit for sclerotic scatter or retro illumination, rotate the illumination prism assembly (11) to the left or right as desired.
-
When using the blue filter (12) the user may wish to insert the yellow barrier filter (9). The yellow barrier filter is out when the knob is up, in when it is down.
-
When the examination is complete, set the rheostat to a low level and switch off the Slit Lamp.
Shut down after every use. In case the dust cover is used: risk of overheating.
6.3 DESCRIPTION OF FILTERS, APERTURES, AND MAGNIFICATIONS
Stereomicroscope
Eyepieces 12.5x
Dioptric adjustment +/– 8D
PD range 8° converging eyepieces 49mm – 77mm
The convergent angle of the optical axis is 13°
Magnifi cation change
3 step drum | Magnifi cation | Field of view |
---|---|---|
3 | 10x | 22mm |
3 | 16x | 14mm |
3 | 25x | 8.5mm |
Filters
a) Clear
b) Neutral Density
c) Red Free
d) Blue
Apertures
Aperture diameters (mm)
Tower illumination
Tower has the facility to tilt towards the user and is positively located at
each step. 0°, 5°, 10°,15°, and 20°.
ROUTINE MAINTENANCE
The maintenance outlined below should only be carried out once the main power cable has been disconnected. If you have any problems that are not covered by the procedures described below, contact Tomey GmbH or your local supplier.
7.1 LED SYSTEMS
- LED typically has a life exceeding 10,000 hours of continuous use and therefore can be considered as a non-consumable item that will not require changing by a user.
- Whilst this is a significant life expectancy, we suggest that the Slit Lamp is always switched off between examinations to conserve energy and LED life.
- In the unlikely event of an LED failure please contact Tomey or your local distributor for guidance on the replacement procedure.
7.2 REGULARLY INSPECT THE DEVICE FOR DAMAGE OR DIRT
- Routinely clean as per section 3 on page 5 cleaning instructions.
7.3 CLEANING THE ILLUMINATION PRISM
- The prism should be cleaned with a soft, clean lens cloth.
- Care must be taken to keep the objective and the eyepiece lenses clean – use only soft, clean lens cloths to clean optical surfaces.
7.4 ELECTRICAL CONNECTIONS
- Routinely check all electrical connections, cables, and connectors.
7.5 OPTICS
- The optics should be wiped clean of any loose dirt or debris with a suitable dust brush and then cleaned with a soft dry lens cloth, washed linen, or other nonabrasive lens cleaning material.
7.6 AXLE AND MECHANICAL PARTS
If the Slit Lamp becomes hard to move on the glide plate, the plate should be
cleaned with a lightly oiled cloth or silicon polish. The axle should be
cleaned only with dry lint-free cloths.
WARRANTY
The Tomey Z-Series Slit Lamps are guaranteed for two years against faulty
workmanship materials or factory assembly. Warranty is on a Return To Base
(RTB) basis at the cost of the customer and may be void if the Slit Lamp has
not been regularly serviced. The manufacturer’s warranty and terms and
conditions are detailed on the Tomey GmbH website.
The mirror, main illumination lamp, and general ‘wear and tear’ are excluded
from our standard warranty.
The manufacturer declines any and all responsibility and warranty coverage
should the instrument be tampered with in any manner or should routine
maintenance be omitted or performed in manners not in accordance with these
manufacturers’ instructions.
There are no user-serviceable parts in this instrument. Any servicing or
repairs should only be carried out by Tomey GmbH or by suitably trained and
authorized distributors. Service manuals will be available to authorized Tomey
service centers and Tomey trained service personnel.
SPECIFICATIONS AND ELECTRICAL RATINGS
The Tomey Slit Lamp is a medical electrical instrument. The instrument requires special care concerning electromagnetic compatibility (EMC). This Section describes its suitability in terms of the electromagnetic compatibility of this instrument. When installing or using this instrument, please read carefully and observe what is described here. Portable or mobile- type radio frequency communication units may have an adverse effect on this instrument, resulting in malfunctioning.
9.1 ELECTROMAGNETIC EMISSIONS
Guidance and manufacturer’s declaration – electromagnetic emissions
The Tomey Slit Lamp is intended for use in the electromagnetic environment
specified below. The customer or user should ensure that it is used in such an
environment.
Emissions test| Compliance| Electromagnetic environment –
guidance
---|---|---
RF emissions CISPR 11| Group 1| The Tomey Slit Lamp uses RF energy only for
its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
RF emissions CISPR 11| Group B| The Tomey Slit lamp is suitable for use in a
professional
healthcare facility environment.
Harmonic emissions
IEC 61000-3-2| Group B| The Tomey Slit lamp is not intended for use in home
environment.
Voltage fl uctuations / fl icker emissions IEC 61000-3-3| Complies|
9.2 ELECTROMAGNETIC IMMUNITY
Guidance and manufacturer’s declaration – electromagnetic immunity
The Tomey Slit Lamp is intended for use in the electromagnetic environment
specified below. The customer or user should ensure that it is used in such an
environment.
Immunity test| IEC 60601 Test level| Compliance level|
Electromagnetic environment — guidance
---|---|---|---
Electrostatic discharge (ESD). IEC 6100-4-2| ± 8 kV contact ± 15 kV air| ± 8
kV contact ± 15 kV air| Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative humidity should be at
least 30%
Electrical fast transient/burst.
IEC 61000-4-4| ± 2 kV for power supply lines
± 1 kV for power supply lines| ± 2 kV for power supply lines
N/A| Mains power quality should be that of a typical professional healthcare
facility
Surge. IEC 61000-4-5| ± 1 kV line(s) to line(s) ± 2 kV line(s) for
input/output line(s)| ± 1 kV line(s) to line(s) ± 2 kV line(s) for
input/output line(s)| Mains power quality should be that of a typical
professional healthcare facility
Voltage dips, short interruptions and voltage variations on power supply input
lines. IEC 61000-4-11| UT = 0% 0.5 cycle 45 90 135 180 (0, 45, 90, 135, 180,
225, 270, 315°) UT = 0%; 1 cycle UT = 70%; 25/30 cycles (@ 0°) UT = 0%;
250/300 cycle| UT = 0% 0.5 cycle (0 45 90 135 180 (0, 45, 90, 135, 180, 225,
270, 315°) UT = 0%; 1 cycle
UT = 70%; 25/30 cycles (@ 0°) UT = 0%; 250/300 cycle| Mains power quality
should be that of a typical professional healthcare facility environment. If
the user of the Tomey Slit Lamp requires continued operations during power
mains interruptions, it is recommended that the instrument be powered from an
uninterruptible power supply.
Power frequency (50/60 Hz)
Magnetic field. IEC 61000-4-8| 30 A/m| 30 A/m| Power frequency magnetic fields
should be at a level characteristic of a typical location in a typical
professional healthcare facility environment.
Note: UT is the a. c. mains voltage prior to application of the test level.
Immunity test| IEC 60601 Test level| Compliance
level| Electromagnetic environment – guidance
---|---|---|---
| | | Portable and mobile RF communications equipment should be used no
closer to any part of the Tomey Slit Lamp, including cables, than the
recommended separation distances calculated from the equation applicable to
the frequency of the transmitter.
| | Recommended separation distance
Conducted RF IEC 61000-4-6| 6 Vrms 1 50kHz to 80MHz| 6 V| d = 1.2 √ p
Radiated RF IEC 61000-4-3| 10 V/m 150kHz to 280MHz| 10 V/m| d = 1.2 √ p 80MHz
to 800 MHz d = 2.3 √ p 800MHz to 2.7GHz
| | | Where p is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and d is the recommended
separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey¹, should be less than the compliance level in each
frequency range.² Interference may occur in the vicinity of equipment marked
with this symbol.
Note: At 80MHz and 800MHz, the higher frequency range applies. These
guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.
-
Field strengths from fixed transmitters, such as base stations (cellular/cordless) telephones and land mobile radios, amateur radio, AM, and FM radio broadcast,s and TV broadcasts cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the Tomey Slit Lamp is used exceeds the applicable RF compliance level above, the Tomey Slit Lamp should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re- orientating or relocating the
Tomey Slit Lamp. -
Over the frequency range 150kHz to 80 MHz, field strengths should be less than 10 V/m.
9.3 RECOMMENDED SAFE DISTANCES
Recommended separation distances between portable and mobile RF
communications equipment and the Tomey Slit Lamp
The Tomey Slit Lamp is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of the
Tomey Slit Lamp can help prevent electromagnetic interference by maintaining a
minimum distance between mobile RF communications equipment (transmitters) and
the Tomey Slit Lamp as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output power of the transmitter (W)| Separation
distance according to the frequency of the transmitter
(m)
---|---
| 150 kHz to 80MHz d = 1.2√ p| 80MHz to 800MHz d = 1.2√ p| 800MHz to 2.7GHz
d = 2.3√ p
0.01| 0.12| 0.12| 0.23
0.1| 0.38| 0.38| 0.73
1| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance in meters (m) can be determined using the
equation applicable to the frequency of the transmitter, where p is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note: At 80MHz and 800MHz, the separation distance for the higher
frequency applies. These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.
9.4 TECHNICAL SPECIFICATIONS
Optical System
Type | Galilean 8° converging |
---|---|
Magnification | Drum change x10, x16, x25 |
Eyepiece | x12.5 |
Field of view | 22, 14, 8.5 and 5.5 mm |
PD range | 8° converging eyepieces, 49-77mm |
Objective lens focal distance | 107mm |
The objective lens convergence angle | 13° |
Slit Projection System & Base
Slit Width | 0-14mm continuously variable |
---|---|
Slit Length | 14mm |
Aperture diameters | 0.2, 1mm square, 2, 3, 5, 9, 14 and reserved |
Filters | Clear; red free; neutral density; blue; |
Slit rotation | 360° continuous, detents at 0° and 180° |
Base travel | 25mm Z-axis, 107mm X-axis, 110mm Y axis |
Horizontal fine adjustment | 12mm |
Tabletop dimensions | 405 x 500mm |
Fixation lamp | LED |
--- | --- |
Light source | LED |
LED output power | 240Klux (+/-20%) |
Weight, packed (approx.)
Slit Lamp complete | 25Kg, 90 x 58 x 45cm W x D x H |
---|
Protection against ingress IPxO
Class II ME equipment
Insulation between the main parts and the functional earth provides at least
two means of protection.
Power Supply
Power supply unit| Switch mode, (100V-240V input) +/- 10% multi-plug
compliant to EN/ IEC 60601-1 EN/IEC 61000-6-2, EN/IEC 61000-6-3
---|---
Fuse| T2.5AH, 250V
Power supply output| 12V DC: 2.5 amps must be EN/IEC 60601 compliant
Complies with| Electrical Safety (Medical) EN/IEC 60601-1
Electromagnetic compatibility EN/IEC 60601-1-2
Ophthalmic instruments – Fundamental requirements and test methods ISO 15004-1
Ophthalmic instruments – Optical radiation hazard ISO 15004-2
When the Slit Lamp is connected to the power supply together they constitute a
Medical Electrical System as defined in EN/IEC 60601-1:2006.
The power supply forms a part of ME equipment.
Fuse Ratings and Quantity
2.5 amp anti-surge
Fuse current 2.5A
Voltage rating V AC 250V
Breaking capacity 1500Amps
Blow characteristic: Time Delay
Environmental Conditions:
USE
|
| Shock (without packing)| 10 g, duration 6 ms
STORAGE CONDITIONS
|
TRANSPORT CONDITIONS
|
| Vibration, sinusoidal| 10 Hz to 500 Hz: 0.5g
| Shock| 30 g, duration 6 ms
| Bump| 10 g, duration 6 ms
*This instrument does not meet the temperature requirements of ISO 15004-1 for storage and transportation. Do not store or transport this instrument in conditions where the temperature may rise above 50 °C.
ACCESSORIES AND SPARES
Item | Part Number |
---|---|
DSL Digital Background Illum.Assy | 3020-P-5039 |
Tomey PSU 12V | 3020-P-5040-TOM |
DSL Accessory Kit Packaging | 3020-P-5221 |
DSL Screw Kit | 3020-P-5240 |
Refraction Cable (XY-Bulb) | 3020-P-7013 |
DSL Refraction Stand Cable Kit | 3020-P-7014 |
DSL Digital Backlight Filters | 3020-P-7054 |
1m DSL to PSU Cable | 3020-P-7068 |
Item | Part Number |
--- | --- |
Tomey user instruction USB | 3020-P-7069 |
Digital Base to Computer Cable | 3020-P-7077 |
Fixation&Illum to 6 Pin Connector | 3020-P-7117 |
Dust Cover – No branding | EP39-19072 |
XY Pad (405) | EP39-80169 |
Linear Gear Rack (405) | EP39-80173 |
Linear Gear Rack Cover/silver | EP39-80174 |
DSL Test/Setting Bar | EP39-80243 |
DSL Illum Projection Mirror-Small | EP39-80250 |
Chin Rest Pins (406) | EP39-80321 |
DSL Table Template | EP39-80324 |
Cable Clip | EP39-80329 |
DSL DIGITAL X-Y PAD | EP39-80364 |
DSL DIGITAL LEFT/RIGHT LABEL | EP39-80365 |
1.27mm A/F ALLEN KEY | EP79-00766 |
5mm L-Shaped Hexagonal Key | EP79-10892 |
6mm L-Shaped Hexagonal Key | EP79-10905 |
Black Patient Handle | EP79-80070 |
PACKAGING AND DISPOSAL INFORMATION
(Applicable in the European Union and other European Countries with separate Collection Systems).
Disposal of old electrical and electronic equipment
This symbol on the product or on its packaging and instructions indicates that
this product shall not be treated as household waste. To reduce the
environmental impact of WEEE (Waste Electrical Electronic Equipment) and
minimize the volume of WEEE entering landfills we encourage at the product’s
end of life that this equipment is recycled and reused.
If you need more information on the collection reuse and recycling then
please contact B2B Compliance on 01691 676124 (+44 1691 676124). (the UK
only). Any serious incident that has occurred in relation to the device
must be reported to the manufacturer and the competent authority of your
Member State.
Contact
Manufacturer
Keeler Limited Clewer Hill Road
Windsor Berkshire
SL4 4AA UK
Freephone 0800 521251
Tel +44 (0) 1753 857177
Fax +44 (0) 1753 827145
SUPPLIER
TOMEY GmbH
Wiesbadener Straße 21
90427 Nürnberg
Germany
Tel +49 911 938 546 2 – 0
Fax +49 911 938 546 2 – 20
Biometrics, S. L., Vinyals, 131
08221 Terrassa, Spain
EP59-41337 Issue 5
Date of Issue 12/05/2021
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>