BW Technologies BW Clip Real Time Single Gas Detector Owner’s Manual

June 5, 2024
BW Technologies

BW Technologies BW Clip Real Time Single Gas Detector Owner’s Manual
BW Technologies BW Clip Real Time Single Gas Detectorr Owner's
Manual

BW Clip Real Time

BW Clip Real Time is a portable, single-gas detector manufactured by BW Technologies by Honeywell (BW). BW Clip Real Time continuously displays the concentration of a specific gas in the ambient environment, and activates an alarm when concentrations exceed alarm setpoints. It is your responsibility to respond appropriately to the alarms. BW Clip Real Time is a member of the GasAlert family of products. It is compatible with IntelliDoX and MicroDock II automatic test and calibration stations, as well as Fleet Manager II software version 4.2 or higher. For more information, visit the product website at
www.honeywellanalytics.com

Safety Information: Read First

Warning

  1. Substitution of components may impair Intrinsic Safety.
  2. BW recommends performing a bump test prior to each day’s use to confirm sensor response and alarm activation by exposing the detector to a concentration of target gas that exceeds the low alarm set point. BW also recommends performing a bump test if the detector has been subjected to physical impact, liquid immersion, an Over Limit alarm event, custody changes, or anytime the detectors performance is in doubt. Calibrate the detector after a failed bump test. Always bump test and calibrate in a fresh air environment.
  3. Detectors in hibernation mode are not functional.

Caution

  1. Activate the detector before the activation date on the package.
  2. This product is a gas detector, not a measurement device.
  3. Ensure that the sensor grill is free of dirt, debris, and is not obstructed.
  4. Clean the exterior with a soft, damp cloth.
  5. For optimal performance, periodically zero the sensor in a normal atmosphere (20.9% v/v O2 ) that is free of hazardous gas.
  6. Portable safety gas detectors are life safety devices. Accuracy of ambient gas reading(s) is dependent upon factors such as accuracy of the calibration gas standard used for calibration and frequency of calibration. BW recommends performing a calibration at least once every 180 days (6 months).

Getting Started

Product Overview

LCD and Symbols

LCD Display

| Pushbutton prompt. When this symbol is displayed, press and hold the pushbutton until the symbol disappears
---|---

| BW Clip Real Time O2 models only. Oxygen concentration is measured as percent by volume.

| BW Clip Real Time H2 S, CO and SO2 models only. Toxic gas concentration is measured in parts per million

| Low alarm and high alarm. These symbols are displayed when a gas concentration that exceeds alarm setpoints is detected.

| Non-compliance warning. This symbol is displayed when a diagnostic or compliance test fails; when the operating life countdown reaches 24 hours or less; and for up to 30 days after operating life expiry. It is also displayed when certain notifications are enabled through Fleet Manager II via an IntelliDoX station

| Real-time clock. This symbol is displayed when real-time clock is enabled through Fleet Manager II via an IntelliDoX station.

| End of operating life warning. This symbol is displayed when the operating life countdown reaches 24 hours or less.

| Peak reading. This symbol is displayed when a maximum gas exposure event has occurred within the past 24 hours.

| Bump test reminder. If the detector is configured to display a bump test reminder, this symbol is displayed when a bump test is due.

|  Indicates a failed bump test.

| This symbol is displayed when more than 24 hours have elapsed since the last successful zero procedure was performed (O2 models) or indicates a sensor calibration is due (H2 S, CO and SO2 models)

| indicates a failed calibration (H2 S, CO and SO2 models) or a failed zero procedure (O2 models).

| Operating life remaining in months (MM) or days (DD).

Activate a New Detector

  1. Move to a normal atmosphere (20.9% v/v O2 ) that is free of hazardous gas.
  2. Press and hold the pushbutton until a 5 second countdown is displayed, and then continue to hold until the countdown is complete.
  3. When the countdown is complete, the LCD and LEDs turn on and then turn off. The detector performs an internal diagnostic test.
  4. When the diagnostic test is successful, the alarm setpoints are displayed and the LEDs flash.
  5. The sensor stabilization countdown is displayed. During the countdown, the detector vibrates continuously for 20 seconds. The time required to stabilize varies depending on sensor type. When the countdown reaches 0, the LEDs flash and the audible alarm beeps.
  6. The detector is in normal operating mode when the gas type and concentration are displayed.

Normal Operating Mode

When the detector is in normal operating mode, the type of gas detected is permanently displayed. The detected concentration of the gas is displayed until it is disrupted by a pushbutton action, gas alarm, or error event.

Operating Life

When the detector is activated for the first time, the operating life countdown is displayed. Press the pushbutton to display remaining operating life.

90 Days or Less Remaining

When the operating life countdown reaches 90 days or less before expiry, the countdown display changes to days. The countdown displays the remaining operating life in days until the countdown reaches 24 hours or less before expiry.

24 Hours or Less Remaining

When the operating life countdown reaches 24 hours or less before expiry, the countdown display changes to hours, and the EXPIRY! warning and non-compliance symbol are Displayed.

End of Operating Life

When the operating life countdown ends, the detector deactivates and detector safety functions are disabled. It is possible to retrieve event logs for a limited time after expiry. The EXPIRY! warning and non-compliance symbol are displayed for up to 30 days after expiry

Internal Diagnostic Tests

Activated detectors automatically perform one internal diagnostic test every 24 hours. If the internal diagnostic test fails, the diagnostic fail-safe begins.

Diagnostic Fail-Safe

  1. The EXPIRY! warning and non-compliance symbol are displayed.
  2. The detector beeps and vibrates.
  3. The LEDs flash 2 times per second for 15 seconds.
  4.  OFF or an error code is displayed. The detector is deactivated.

It is possible to retrieve the event logs for a limited time after expiry. If you are unable to retrieve the event logs, contact BW.

Alarms and Warnings

An alarm is initiated when the sensor is exposed to a gas concentration that exceeds alarm set points. The alarm persists until the alarm gas concentration returns to an acceptable range. Battery life decreases rapidly when the detector is in alarm condition.

Low Alarm

Audible: 1 beep per second
Visual: 1 flash per second
Vibration: 1 vibration per second

High Alarm

Audible: 2 beeps per second
Visual: 2 flashes per second
Vibration: 2 vibrations per second

Over-Limit and Under-Limit Alarms

If the gas concentration is beyond the sensor range, OL (over-limit) or

-OL (under-limit) is displayed.
Audible: 2 beeps per second
Visual: 2 flashes per second
Vibration: 2 vibrations per second

Automatic Zero Reminder for O2 Models

models only, the Automatic Zero Reminder is displayed when more than 24 hours have elapsed since the last successful zero procedure was performed.

Peak Readings

The peak readings symbol is displayed when the sensor is exposed to a gas concentration that exceeds the alarm setpoints. It is no longer displayed when more than 24 hours have passed since the last alarm, or when a successful bump test is performed via an IntelliDoX or MicroDock II station.

Non-compliance Symbol and LED Indicators

The non-compliance warning symbol is displayed when:

  • an internal diagnostic test fails;
  • a bump test, calibration or zero procedure fails;
  • a bump test or calibration is due, if enabled;
  • the sensor is exposed to a gas concentration that exceeds alarm setpoints, if enabled.

The non-compliance warning symbol is no longer displayed when more than 24 hours have passed since the last alarm, or when a successful bump test or calibration is performed (manual calibration or via an IntelliDoX or MicroDock II station). The non-compliance warning symbol is permanently displayed when the service-life countdown reaches 24 hours or less before expiry, and for up to 30 days after expiry.

When Non-compliance after gas exposure is enabled through Fleet Manager II via an IntelliDoX station, the LEDs flash in an alternating pattern when the non-compliance warning symbol is displayed. When the symbol is first displayed, the top LEDs flash. After 30 seconds, the side LEDs flash. After 30 seconds, the pattern repeats until the noncompliance warning symbol is no longer displayed.

NOTE: When the detector is bump tested via a MicroDock II station, an abnormal test result may occur if non-compliance LEDs are flashing. To suspend the flashing for up to 10 minutes, press and release the pushbutton. When the flashing stops, insert the detector and perform the test. If the bump test fails, the LEDs resume flashing.

Recall Events and Settings

When the detector is in normal operating mode, press the pushbutton at any time to scroll through events and settings that are not empty or disabled through Fleet Manager II via an IntelliDoX station. BW Clip Real Time flashes, vibrates and beeps. While an event or setting is displayed, press the pushbutton to scroll to the next available event or setting. If you do not press the pushbutton, or if you have reached the last available event or setting, the detector returns to normal operation. Events and settings are displayed in this order:

  1. Peak Reading, if available, and when a maximum gas exposure event has occurred within the past 24 hours.
  2. Time of Peak Reading, if Real-time Clock display is enabled.
  3. Operating Life, the remaining operating life of the detector in months, days or hours.
  4. Low Alarm Setpoint
  5. High Alarm Setpoint
  6. Real-time Clock, if enabled. High Alarm
  7. Firmware Version
  8. Next Bump Due, if enabled.
  9. Next Calibration Due, if enabled.
  10. Hibernation, for 2-year H2 S and CO models only.

Zero the Sensor

Over time and through use, the sensor baseline at zero exposure may drift from the manufacturer’s baseline. For optimal performance of O2 models, BW recommends that you zero the O2 sensor once every 24 hours or when the Automatic Zero Reminder is displayed. For all other models, BW recommends that you zero the sensor periodically

Zero Procedure

  1. Move to a normal atmosphere (20.9% v/v O2 ) that is free of hazardous gas.

  2. Press and hold the pushbutton until a 5 second countdown is displayed, and then continue to hold until the countdown is complete.

  3. When the countdown is complete, the Zero procedure begins and ZErO is displayed.

  4. When Zero is successful, PASS and then CAL ? is displayed. Press the pushbutton to calibrate, or wait for the detector to enter normal operating mode after a 5 second countdown.

  5. If Zero is not successful, the non-compliance LED flashes. FAIL and the non-compliance warning symbol are displayed. Press the pushbutton to acknowledge the result and return to normal operation. The non-compliance warning symbol is displayed and the non-compliance LED flashes. Perform the Zero procedure again. If the procedure fails again, contact BW.

Bump Tests

When the detector is configured to display a bump test reminder, the bump test symbol is displayed when a bump test is due. For best results, bump test the detector via an IntelliDoX or MicroDock II station. Bump tests performed via an IntelliDoX or MicroDock II station are logged as bump tests. Manual bump tests are logged as unsafe gas concentrations.

Calibration

By default, BW Clip Real Time is configured to use the following calibration gas mixtures:

H2 S : 20 ppm
balance N2
CO: 100 ppm
balance N2
SO2 : 20 ppm
balance N2
O2: 18.0 % v/v O2

You can use Fleet Manager II software via an IntelliDoX station to change a detector’s default calibration gas mixtures. If the detector’s default calibration mixtures were changed, use the gas mixture values from Fleet Manager II to calibrate the detector.

  1. Move to a normal atmosphere (20.9% v/v O2 ) that is free of hazardous gas.
  2. Connect the calibration hose to the gas cylinder and to the intake inlet on the calibration cap.
  3. Press and hold the pushbutton until a 5 second countdown is displayed, and then continue to hold until the countdown is complete. The zero procedure begins and ZErO is displayed. When Zero is successful, PASS is displayed.
  4. When CAL ? is displayed, press the pushbutton to begin calibration.
  5. Put the calibration cap on the sensor grill.
  6. When GAS ? is displayed, apply calibration gas for two minutes at a recommended flow rate of 500 ml/min. When calibration is successful, PASS is displayed and the detector returns to normal operating mode.

f calibration fails, non-compliance warning symbols are displayed. Press the pushbutton to acknowledge. Repeat calibration. If it fails again, contact BW.

Event Logs

The detector stores the last 35 events that occurred, including peak readings, bump tests, calibrations and auto zeros. Each record contains:

  • detector serial number, sensor type and life-remaining;
  • total number of events that have occurred;
  • event type and duration;
  • alarm level(s) in ppm or %;
  • time elapsed since the alarm occurred in days/hours/minutes;
  • duration of the alarm (minutes/seconds).

Use Fleet Manager II via an IntelliDoX or MicroDock II station to transfer event logs from the detector to a computer. When logs are transferred via an IntelliDoX station, the most recent 35 events are transferred. When logs are transferred via a MicroDock II station, the most recent 10 events are transferred.

Optional Settings

Use Fleet Manager II via an IntelliDoX station to enable the following optional settings for BW Clip Real Time detectors.

Non-compliance After Gas Exposure

When Non-compliance after gas exposure is enabled, the LEDs flash in an alternating pattern when the sensor is exposed to a gas concentration that exceeds alarm setpoints. When the gas exposure is detected, the top LEDs flash. After 30 seconds, the side LEDs flash. After 30 seconds, the pattern repeats until more than 24 hours have passed since the last alarm, or when a successful bump test is performed via an IntelliDoX or MicroDock II station.

Real-time Clock Display

When real-time Clock Display is enabled, you can display the current time by pressing the pushbutton to Recall Events and Settings, and then continuing to press the pushbutton until the time is displayed. Time may be displayed in 12-hour or 24-hour clock format.

Hibernation

Hibernation is a factory installed feature that is available for 24-month H2 S and CO detectors. When Hibernation is activated, the operating life countdown is suspended. Use Hibernation to deactivate the detector for 7 days or longer, up to a maximum of 12 months. When Hibernation is activated, all detector safety functions are disabled.

Activate Hibernation via BW Clip Real Time Hibernation  Case

  1. Have ready a BW Clip Hibernation Case. To purchase a Case, contact BW or an authorized distributor.
  2. Move to a normal atmosphere (20.9% v/v O2 ) that is free of hazardous gas.
  3. Verify that the detector is in normal operating mode.
  4. Press the pushbutton to Recall Events   and Settings, and then continue to press the pushbutton until Hib is displayed.
  5. While Hib is displayed, press the pushbutton until a 5 second countdown is displayed. When the countdown is complete, the 20-second Hibernation counter is displayed.
  6. While the Hibernation counter is displayed, place the detector into a BW Clip Real Time Hibernation Case and then close the case firmly. The detector enters Hibernation mode. All detector safety functions are disabled.
  7. To reactivate the detector, open the case and remove the detector.

Activate Hibernation through Fleet Manager II via IntelliDoX

  1. then insert the detector in the configured station. The detector event logs are retrieved, and the detector enters Hibernation mode. All safety functions are disabled, and the detector may be removed from the IntelliDoX station.
  2. To reactivate the detector, press and hold the pushbutton until a 5 second countdown is displayed, and then continue to hold until the LCD and LEDs turn on and  off. The detector performs a self-diagnostic test. When the test is successful, the type of gas detected a nd the remaining operating life are displayed. The detector is in normal operating mode.
  3. For more information, refer to the operator manuals for the IntelliDoX station and Fleet Manager II software version 4.2 or higher.

About this Publication

This publication is an operator guide for BW Clip Real Time portable single- gas detectors manufactured by BW Technologies by Honeywell. It is intended for use with the following models:

  • 24- and 36-month H2S and CO Detectors
  • 24-month O2 detector
  • 24-month SO2 detector.

Copyright, Notice, Trademarks

While this information is presented in good faith and believed to be accurate, BW Technologies by Honeywell (BW) disclaims the implied warranties of merchantability and fitness for a particular purpose and makes no express warranties except as may be stated in its written agreement with and for its customers. In no event is BW liable to anyone for any indirect, special or consequential damages. The content in this document is subject to change.

Specifications

Shelf Life

H2S One (1) year before activation
CO One (1) year before activation
O2 Six (6) months before activation
SO2 One (1) year before activation
Instrument Weight alligator clip included
92 grams (3.2 ounces)
H2S One (1) year before activation
CO One (1) year before activation
O2 Six (6) months before activation
SO2 One (1) year before activation
Instrument Weight alligator clip included
92 grams (3.2 ounces)
Operating Humidity
5% to 95% relative humidity (non-condensing)
Audible Alarm
≈ 95 dB at 10 cm (3.9 in)
Visual Alarm
Flashing, wide-angled alarm lens with red
LEDs plus alarm LCD readout
Display
Liquid crystal display (LCD)
Sensor Type
Electrochemical cells
Battery
Lithium, non-replaceable
Ingress Protection
IP 66/67
Alarm Setpoints
Instant low and instant high
Calibration
To ensure accurate calibration, BW Technologies recommends using a premium grade calibration gas approved by the National Institute of Standards and Technology (NIST). Do not use a gas cylinder beyond its expiry date. BW Technologies recommends you calibrate the detector at least once every 180 days depending.

http://www.honeywellanalytics.com/~/media/ honeywell-analytics/products/bw- clip/docu-ments/certificates/ korea-certi ication_bw-clip-15ka-4bo0306 .

on use and sensor exposure to poisons and contaminants. Calibrate only in a safe area that is free of hazardous gas.

Allowable Calibration Gas Mixtures

H2S :5 to 50 ppm
CO2: 25 to 250 ppm
SO2: 5 to 70 ppm
O2: 5.0 % to 19.0 % v/v O2
For O2 detectors, perform the Zero procedure once every 24 hours or when the Automatic Zero Reminder is displayed.
Intended Use
This product is classified for use in hazardous atmospheres that are not more than
21% v/v O2

Recycling

This instrument contains a lithium battery. Do not mix with the solid waste stream. Spent batteries should be disposed of by a qualified recycler or hazardous materials handler.

CAUTION

Products may contain materials that are regulated for transportation under domestic and international dangerous goods regulations. Return product in compliance with appropriate dangerous goods regulations. Contact freight carrier for further instructions.

Limited Warranty and Limitation of Liability

BW Technologies (BW) warrants this product to be free from defects in material and workmanship under normal use and service for a period of two or three years (depending upon detector), beginning on the date of activation. 24-month H2S and CO detectors are covered for up to an additional 12 months when hibernation is used, limited by a total of 24 months of detector operation

This Warranty is valid only if the detector is activated by the date on the package. This warranty extends only to the sale of new and unused products to the original buyer. BW’s warranty obligation is limited, at BW’s option, to refund of the purchase price, repair, or replacement of a defective product that is returned to a BW authorized service center within the warranty period.

In no event shall BW’s liability hereunder exceed the purchase price actually paid by the buyer for the Product.

This warranty does not include:

  • fuses, disposable batteries or the routine replacement of parts due to the normal wear and tear of the product arising from use;
  • any product which in BW’s opinion, has been misused, altered, neglected or damaged by accident or abnormal conditions of operation, handling or use; or
  • any damage or defects attributable to repair of the product by any person other than an authorized dealer, or the installation of unapproved parts on the product.

The obligations set forth in this warranty are conditional on:

  1.   Proper storage, installation, calibration, use, maintenance and compliance with the product manual instructions and any other applicable recommendations of BW;
  2. the buyer promptly notifying BW of any defect and, if required, promptly making the product available for correction. No goods shall be returned to BW until receipt by the buyer of shipping instructions from BW; and
  3. the right of BW to require that the buyer provide proof of purchase such as the original invoice, bill of sale or packing slip to establish that the product is within the warranty period.

THE BUYER AGREES THAT THIS WARRANTY IS THE BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. BW SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT OR RELIANCE OR ANY OTHER THEORY. Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will not affect the validity or enforceability of any other provisions.

China RoHS 2 Card

Part or Assembly
Name| Hazardous Substances
---|---
Lead
(Pb)| Mercury
(Hg)| Cadmium
(Cd)| e H xavalent
Chromium
(Cr (VI))| Polybrominat-
ed bibhenyls
(PBB)| Polybrominated
diphenyl ethers
(PBDE)
Gas Sensors| X| 0| 0| 0| 0| 0
Circuit Board
Assembly| 0| 0| 0| 0| 0| 0
Endosure
Material| 0| 0| 0| 0| 0| 0

All parts and assemblies not listed in this table contain hazardous substances below the limit requirements of GB/T 26572 This table is prepared in accordance with the provisions of SYT 11364 0: Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GBiT 26572 X: Indicates that said hazardous substance contained in at least one of the homogenous materials used for this part is above the limit requirement of GB/T 265721 .1. ,

Contact Us

Corporate Headquarters

BW Technologies by Honeywell
Suite 110 4411-6 Street SE
Calgary, Alberta
Canada, T2G 4E8
Toll-free: 1-800-663-4164

Americas
BW Technologies by Honeywell
405 Barclay Blvd.
Lincolnshire, IL
USA 60069
Toll-free: 1-888-749-8878

Europe
BW Technologies by Honeywell
Javastrasse 2
8604 Hegnau
Switzerland
Toll-free: 00800-333-222-44

Other
BW Technologies by Honeywell
Toll-free: 1-403-248-9226
info@gasmonitors.com
www.honeywellanalytics.com

Warranty Registration

www.honeywellanalytics.com/support/ product-registration

Maximum Operating Life

24-month Detector

2 years after activation, assuming 2 minutes of alarm time per day.
Note: 24-month H2S and CO detectors that use hibernation may extend the service period of the detector up to an additional year. The service life will end once the detector reaches 24 months of actual operation. Hibernation provides a maximum service life of 36 months.

36-month Detector

3 years after activation, assuming 1 minute of alarm time per day.

Detection Range
H2S 0 to 100 ppm
CO2 0 to 300 ppm
O2
0 to 25% v/v O2
SO2
0 to 100 ppm
Factory Alarm Setpoints
Low High Alarm
H 15 ppm
CO 200 ppm
O 23.5% v/v O2
SO 10 ppm
Use Fleet Manager II via an IntelliDoX station to adjust alarm setpoints. For more information, refer to the operator manuals for Fleet Manager II software version 4.2 and the IntelliDoX automatic test and calibration station.

Ratings and Certifications

The BW Clip Real Time is in conformity with the following standards:

UL 913 8th Edition
UL 60079-0:2013,
UL 60079-11:2013
CSA C22.2 No. 157-92:2012,
CSA C22.2 No. 60079-0:2011,
CSA C22.2 No. 60079-11:2014
EN 60079-0:2012 +A11:2013,
EN 60079-11:2012
IEC 60079-0:2011
IEC 60079-11:2011
ABNT NBR IEC 60079-0:2008
ABNT NBR IEC 60079-11:2009

Classified by UL to both US and
UL Canadian Standards as intrinsically safe
for Class I, Division 1, Group A, B, C, D
and Class I, Zone 0, Group IIC.

ATEX
CE 0539 II 1G
Ex ia IIC T4 Ga IP66/67
DEMKO 14 ATEX 1356
CE
European Conformity
EU Declaration of Conformity
www.honeywellanalytics.com
IECEx
Ex ia IIC T4 Ga IP66/67
IECEx UL 14.0063
INMETRO
Ex ia IIC T4 Ga
DNV 15.0024
KTL
15-KA4BO-03061

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>