Digital Electronic Coating
Thickness Gauge
DPM-816 PRO
USER MANUAL
EN-01.01.21
Device controls
“MODE” button
– Switches the operating mode
– Increases the thickness during calibration
– Switches on two-point calibration
– Switches on six-point calibration of ferrous metals
“POWER” button
– Turns on (off) the device
– Turns on (off) the display backlight
– Switches to the next point of calibration, and saves calibration “UNIT”
button
– Switches the unit of measurement
– Decreases the thickness during calibration
– Resets zero deviation and two-point calibrations
– Switches on the six-point calibration of non-ferrous metals
Introduction
Dear owner of the digital coating thickness gauge (referred to as “the device”
from this point), we congratulate you with the purchase of the device of
Russian origin.
We are confident that the device will be useful to you, and last a long time.
For correct and optimal use of the device, read this manual carefully.
This model has the following features:
- Accurate coating thickness measurement on ferrous, nonferrous and galvanized metals
- Measurement on convex and concave surfaces
- Measurement on galvanized surfaces
- Measurement of deviation from a sample
- Wide range of measurement up to 3500µm (137 Mils)
- Measurement in millimeters, mils, and micrometers
- Wide operating temperature range, -25 to 40ºC (-14 to 104 F) with automatic calibration adjustment
- Informative LCD display with backlight
- Sensor pressure stabilization
- Automatic metal type recognition (Fe/nFe)
- Automatic and manual selection of measurement methods
- Measurement using magnetic induction and eddy current
- Power supplied through one AAA battery at 1.5 volts
- Low battery indicator
- Small and convenient size
- Automatically turns off power and backlight
- 5 modes for various measurement conditions
- Independent calibration using one, two, or six points of reference saved to the non-volatile memory of the device
- Simple, intuitive controls with 3 buttons
- Distinct sound signal for each measurement method and button press
- Complies with industry standards
- 3-year factory warranty
Purpose
The device is designed to measure coating thickness:
Magnetic induction method – thickness of various dielectric (paint,
plastic, primer, etc.) and electrically conductive, nonmagnetic surfaces
(aluminum, zinc, chrome, copper)
on a ferrous metal substrate (steel, iron).
Eddy current method – thickness of various dielectric coatings (paint,
plastic, primer, etc.) on a non-ferrous metal substrate (aluminum, zinc,
copper, etc.)
Application
Process control of the application of paint or other protective coatings in-
service centers and in manufacturing.
Diagnostics of the car body to assess its condition. For example: when buying
a car, before repairing dents without the paint (PDR), or scratch removal.
Device care
Please keep in mind that the device is a high-precision measuring instrument.
Avoid dropping the device and subjecting it to mechanical impact.
The device is protected against moisture but is not waterproof. It must not be
used underwater or in heavy rain. If the surface of the device accidentally
gets wet, wipe off the moisture with a soft, dry piece of cloth.
Do not use the device near sources of strong magnetic fields or in the
presence of strong radio waves. They can impact the performance and
measurement accuracy of the device.
Do not leave the device exposed to high temperatures, such as in a car,
exposed to direct sunlight. Disassembling the device is forbidden.
If the device is brought from a cold to a warm environment, condensation may
form on its surface and its internals. To avoid condensation, first, place the
device in a plastic bag. Before removing the device from the bag, wait for it
to warm up.
Should condensation form, do not use the device. Wait until all the condensate
evaporates.
Cleaning the device
To remove dust from the device case or the sensor, use a soft, dry cloth. Do not use cleaning agents containing organic solvents.
Storage conditions
Store the device in a warm and dry place away from direct sunlight. Do not
leave the battery inside the device during long-term storage.
Do not leave the device near strong sources of magnetic fields, for example,
near magnets, power supplies or electric motors.
Do not store the device in hot, humid, or dusty environments, or in
environments containing corrosive chemical substances.
Display
The device has an informative LCD display with a backlight that automatically
turns off. It shows the coating thickness, mode, method, measurement units and
other parameters.
Although the LCD display is manufactured using FSTN technology with a wide
operating temperature range, it may refresh slower in low and high
temperatures.
Sound signal
The device has a distinct sound signal for each measurement method. This allows the user to quickly identify the type of metal substrate under the coating when measuring in automatic mode, without having to look at the display.
Sensor
To increase the accuracy of measurements and to enable measurements on convex
and concave surfaces, the device uses a sensor constructed to stabilize the
pressure on its sensitive tip against the surface, as well as a spherical form
for its sensitive tip. The sensitive part of the sensor is made of high-
strength heat-treated steel, and the area of contact
with the measured surface is less than 1 mm².
Do not apply mechanical stress to the sensor. Do not attempt to extract the
sensor from the device, and do not rotate it. This may interfere with the
correct operation of the device.
Included in the set
First of all, make sure that the following items are present:
- Coating thickness gauge
- Aluminum and steel base plates for calibration
- Two sample plastic plates for calibration
- User manual
Getting started
To get started, you need to install one alkaline battery type AAA, 1.5 volts.
Before first use, or in case the device has not been used for an extended
period, as well as to increase the measurement accuracy, you need to calibrate
the device using one or two points of reference.
Installing the battery
Open the battery cover, shifting the cover with your thumb in the direction of the arrow. Insert the battery, taking note of the polarity. Close the battery cover until it clicks.
Low battery indicator
The device consumes very low power and can last an extremely long time from a
single battery without a backlight. When the battery eventually loses its
capacity and does not
supply enough power, the top right corner of the display will show a low
battery indicator.
Turning on the device
To turn on the device, press and hold the button for 2 seconds until the
device emits a sound signal.
After turning it on, the display will show “8888”.
The device will
begin to automatically test and calibrate the sensor.
After 2-3 seconds, the display will show dashes, indicating that the device is
ready to take measurements.
IMPORTANT! When turning on the device, keep it away from metal objects and
strong sources of magnetic fields!
Turning off the device
To turn off the device from any state, press and hold the button for 2
seconds.
Display backlight
The device has an integrated backlight for the display. To turn the backlight
on or off, shortly press the button.
Operating modes
The device has 5 operating modes: A, M, P, F(Fe), N(nFe).
To switch between these modes, shortly press the button, which will switch
modes in a loop. The current model will be shown on the display.
A – Auto : Quick measurement with automatic selection of measurement
method. Dynamic resolution. Great for checking the car body.
M – Manufacturing: Detailed and precise measurements with high resolution
across the entire range. Automatic selection of measurement method. It is
recommended to complete the two-point calibration before measuring in this
mode.
P – Program: Measurement using six points of reference for custom
calibration. Automatic selection of measurement method. To use this mode,
completing the six-point calibration is required. This mode is designed for
cases where the standard factory calibration of the sensor is not suitable.
For example, when measuring coating thickness on a substrate with properties
vastly differing from steel and aluminum.
F(Fe) – Magnetic induction method: Quick, continuous measurement (4
measurements per second) of coating thickness on a ferrous metal substrate,
using magnetic induction. For accurate measurements, it is recommended to
complete the two-point calibration before measuring.
N(nFe) – Eddy current method: Quick, continuous measurement (4
measurements per second) of coating thickness on an aluminum alloy substrate,
using eddy current. For accurate measurements, it is recommended to complete
the two-point calibration before measuring.
Units of measurement
For user convenience, the device can display the results of measurement in several units of measurement: millimeters, mils, and micrometers.
The choice of units of measurement is made with short consecutive presses of the button, which will switch units in a loop. The current unit will be shown on the bottom right corner of the display.
Taking the measurement
Turn on the device (see the “Turning on the device” section above) and wait for automatic calibration to complete. After automatic calibration (which takes 2-3 seconds) the display will show dashes.
If necessary, select the desired mode and unit of measurement in which you
want to get the result.
Taking measurements in A, M, and P modes (automatic modes): Firmly place
the device sensor flat against the surface to be measured. Do not move the
device during measurement (until a sound signal is emitted), otherwise, the
measurement may be inaccurate.
Once the device detects that the sensor is on the surface in a fixed position,
you will hear a sound signal indicating the completion of the measurement.
Move the device away from the measured surface. The display will show the
result for 30 seconds – the measured coating thickness and method of
measurement: Fe (magnetic induction, ferrous metals) or nFe (eddy current,
nonferrous metals).
The accuracy of the measurement depends on how flatly and firmly the device
sensor is pressed against the measured surface.
If you are using the sample plates included in the set to measure or
calibrate, the measurement must be taken from the center of the plates.
Taking measurements in F(Fe), N(none) modes (continuous measurement):
In these modes, the device takes measurements quickly and continuously. To
take measurements, place the device sensor against the surface to be measured.
The device will immediately and continuously (4 times per second) display the
result of the measurement.
These modes are useful when it is difficult to hold the device in a fixed position. For example, when measuring on pipes, convex and concave surfaces.
Calibration
For accurate thickness measurements, the device needs to be calibrated. The device retains the calibration information and in most cases you don’t need to calibrate it again before every use.
For the highest accuracy, it is advisable to calibrate the device using a
substrate that closely resembles the surface to be measured (material, form,
size).
For example, when measuring the coating thickness on carbon steel and chrome
steel substrates, the calibration will vary insignificantly, but aluminum and
zinc substrates will vary greatly. Therefore it is important to calibrate.
The device has separate, independent calibrations:
Zero deviation calibration (one point of reference) and two-point calibration
are performed independently (separately) for A, M, Fe, Fe modes, and for P
mode with each measurement method (magnetic induction and eddy current). This
allows you to set up a one-time calibration for modes A, M, Fe, Fe, for
example on carbon steel and aluminum substrates, and for P mode, chrome steel
and zinc substrates. To switch between calibrations, simply switch between
modes.
Keep a clean environment during calibration, as the slightest bit of dust stuck between plastic and metal base plates may distort the result by 5-10 µm.
Zero deviation calibration (one-point calibration)
When to calibrate:
- When measuring on a metal substrate without a coating, the device will show a result greater or less than zero (depending on the properties of the metal substrate). After calibration, the device will display zero.
- In cases where you need to measure deviation from a sample. In these cases, the zero deviation calibration is done on a sample coating and the device will use that coating as a point of reference – “0”. The device will display the difference from the sample coating in subsequent measurements.
Zero deviation calibration procedure:
For zero deviation calibration, take a measurement of the metal substrate
without a coating, or a measurement of sample coating. When the result is
displayed, press and hold the button for 2 seconds. The display will show
“CAL”, which indicates a successful calibration.
Verify the result
by repeating the measurement.
Removing calibrations
This function will completely remove zero deviation and two-point calibrations
for all modes. This function will restore the device’s functionality in case
of incorrect calibration, but for accurate measurements, it needs to be
calibrated again.
To remove calibrations, when the display shows dashes
, press and hold the button for 3 seconds.
The display will show “-CL-” and a sound signal will be emitted.
The calibrations have been erased.
Two-point calibration
When to calibrate:
– Before first use;
– If the device has not been used for an extended period;
– Before each measurement for maximum accuracy.
Two-point calibration procedure:
To enter calibration mode, press and hold the button for 2 seconds until the
display shows “F-0”. This indicates the zero point calibration.
Take a measurement on a metal substrate without a coating, for which the
device will be calibrated. If necessary, repeat the measurement. To proceed to
the next point for calibration, shortly press the button while the measurement
result is being displayed. The display will show “F-1”, the second point of
calibration.
Take a measurement on the same metal substrate together with a sample coating
of known thickness, for example, 700 or 1000 µm. If necessary, repeat the
measurement. While the measurement result is being displayed, shortly press
the or buttons to adjust the displayed thickness of the measured coating, and
shortly press the button to save the calibration. The display will show
“8888”, indicating that the calibration has been successfully saved.
Verify that the calibration is correct by repeating the measurement of the
metal substrate with and without the sample coating. If unsuccessful, repeat
the calibration.
Six-point calibration
This calibration is available in “P” measurement mode and will be required when measuring coatings on substrates whose properties differ from standard substrates, or for more accurate measurements in a given range.
The new device’s calibration is identical to the factory calibration.
For calibration, a metal base and 5 plates of known thicknesses are required
(not included in the set). All calibration points must be evenly distributed
throughout the measuring range.
F-0 (point 0) – Base without a coating (zero).
F-1 (point 1) – Plate with a thickness of 80-300µm (3-12 mils).
F-2 (point 2) – Plate with a thickness of 400-800µm (16-31 mils).
F-3 (point 3) – Plate with a thickness of 900-1800µm (36-70 mils).
F-4 (point 4) – Plate with a thickness of 1900-2500µm (75-98 mils).
F-5 (point 5) – Plate with a thickness of 2600-3500µm (102-137 mils).
Entering the calibration mode:
Turn off the device, then press and hold the button to enter
calibration mode for a ferrous metal substrate (magnetic induction method) or
the button to enter calibration mode for a non-ferrous metal substrate (eddy
current method), until the display shows “FCAL” or “ACAL”
respectively.
Wait until the display shows “0”, indicating zero point calibration.
From this moment, calibration can begin.
Calibration process:
The first point of calibration is “0”. Take a measurement of the metal base
without a coating, for which the device will be calibrated. If necessary,
repeat the measurement.
Shortly press the button to save the result and proceed to the next point of
calibration.
The display will show “F-1” – the next point of calibration, and the device
will show the recommended thickness of calibration plate for the current point
of calibration.
Take a measurement of the metal base together with the calibration plate. If
necessary, repeat the measurement. Use the buttons to calibrate the displayed
thickness of the measured calibration plate, and shortly press the button to
proceed to the next point.
Repeat the procedure for each point of calibration.
After saving the last point, the display will show “8888”,
and the device will switch to operating mode “P”. Repeat the procedure if
necessary, for the second type of metal base.
Please note, if the calibration is done incorrectly on one of the metal base
types, the device will be unable to automatically detect the type of substrate
material and choose the suitable measurement method.
The new device’s six-point calibration is identical to the factory
calibration.
The device is calibrated at the factory using steel (magnetic induction method
of measurement) and aluminum (eddy current measurement method) substrates.
Checking the car body
You can find repainted spots by taking measurements in places of the car body
that are more susceptible to damage. Every car is painted in the factory in an
automated fashion, and the paint thickness is a relatively constant value.
However, different car manufacturers may have slightly different paint
thicknesses.
Usually, the paint coating thickness is within the range of 60 200µm (2.5-8
mils). In a single car, the coating thickness variance on different parts of
the car should not exceed 20%. If the measured thickness exceeds the average
thickness in one or more spots by twofold, it indicates a second layer of
paint. If the measured paint thickness exceeds
the average thickness by more than twofold, such a spot was treated with
putty. The thicker the paint thickness, the lower the quality of the repair
job on the car body.
Specifications
Measuring range (normalized)
– mm
– µm
– mil| 0-3
0-3000
0-118
---|---
Maximum measurable thickness
– mm
– µm
– mil| 3.5
3500
137
Resolution (A mode)
– mm (dynamic)
– µm (dynamic)
– mil (dynamic)| 0.001-0.01
1-10
0.01-0.5
Maximum resolution (M, P, F, N modes)
– mm (fixed)
– µm (fixed)
– mil (dynamic)| 0.001
1
0.01-0.5
Absolute measurement error
– in the range of 0-700µm (0-27 mils)
mm –
µm –
mils –
– in the range of 700-3000µm (27-118 mils)
mm –
µm –
mils –
– in the range of 3000-3500µm (118-137 mils)
mm –
µm –| ±(1%+0.001)
±(1%+1)
±(1%+0.04)±(1%+0.01)
±(1%+10)
±(1%+0.4)
±(3%+0.01)
±(3%+10)
mils — | ±(3%+0.5) |
---|
Operating conditions
– ambient temperature, °C
-25..+40|
– ambient temperature, °F| -13..+104
– relative humidity, % not exceeding| 80
Storage and transportation conditions
– ambient temperature, °C
-25..+55|
– ambient temperature, °F| -13..+131
– relative humidity, % not exceeding| 80
Supply voltage (battery type AAA), V| 2.
Current consumption, mA, not exceeding – in sleep mode (off state)| 0.5
– in measurement mode| 60
Overall dimensions, – mm| 104x43x29
– in| 4.1×1.7×1.2
Mass (without battery), not exceeding – kg| 0.05
– oz| 2.
Surface base dimensions, not less than – mm| 1.0x30x42
– in| 0.04×1.2×1.6
Average service lifespan, years| 5
The specifications given are general. Specifications of individual devices may
vary. The specifications are subject to change without notice.
Possible malfunctions and suggested remedies
Malfunction | Probable cause | Remedy |
---|---|---|
The device does not turn on, or turns on and immediately turns off. | Incorrect |
installation of the battery, reversed polarity. The battery is discharged, or
its service lifespan has ended.| Check the battery polarity. If necessary,
replace the
battery.
The device turns off when the backlight is turned on. or the low battery
indicator is displayed.| The battery is low, or its service lifespan has
ended.| Replace the battery with a new one.
The device occasionally displays an incorrect result.| The measured surface or
the sensor is dirty.| Check the measured surface and the device sensor for
dust or dirt. The sensor should move smoothly without jerking or jamming. If
necessary.
clean the sensor and get rid of dirt.
The device displays an incorrect result when measuring a calibration plate.|
Knocked device calibration. Dust present on the surface of the calibration
plate.| Complete a two-point calibration. Clean the calibration plate from
dust.
When measuring thickness on a metal surface without a coating, the device
displays a non-zero result.| Knocked or incomplete calibration of the device.|
Complete a zero deviation calibration or a two-point calibration.
The device does not take measurements, or the readings greatly vary from
actual thickness.| Interference from metal objects or strong magnetic fields.|
When turning on the device, it should be located not closer than 0.3 m from
large metal objects and the measured surface. Use the device away from sources
of strong magnetic fields.
During prolonged continuous measurement, the
measurement error starts to increase.| Reduction in battery charge.| Stop the
measurement and leave the device turned on. away from metal objects for 10-20
seconds. The device will automatically recalibrate.
The device selects the wrong method of measurement.| Knocked or incorrectly
completed six-point calibration while in measurement mode P.
Unable to determine the best measurement method.| Complete the six-point
calibration for each method of measurement. if you’re measuring in P mode.
Select the desired method of measurement manually, using F(Fe) and N(nFe)
modes.
When measuring on certain substrates, the thickness reading wrong.| The
current calibration preset is unsuitable for the current substrate.| Complete
a two-point calibration.
---|---|---
When the device is placed against the measured surface. the measurement does
not occur.| Incorrect device positioning relative to the measured surface.|
The device must be firmly pressed against the measured surface and remain
fixed until the sound signal is emitted.
When measuring on convex or concave surfaces, the readings have a large
error.| Unable to firmly press the device against the surface and hold it
perpendicular to the surface.| Use the F(Fe) and N(nFe) modes with continuous
measurement.
Warranty
The manufacturer guarantees gratuitous elimination of device defects, arising
from the fault of the manufacturer during the warranty period, subject to all
warranty conditions and storage and operating conditions. The devices are not
subject to warranty repair if: the warranty stickers are damaged, traces of
corrosion or tampering (self-repair) are present, presence of dirt, insects
and objects that are not part of this product; in the absence of a warranty
card or receipt confirming the date and place of purchase of the device, with
the obligatory indication of the name of the seller and the serial number of
the device (the warranty period is counted from the date of purchase); in case
of mismatch of the serial number in the warranty card (receipt) and on the
device; in the presence of mechanical, electrical (smoke, traces of a short
circuit, flooding) or other damage arising from violations of the operating
and transportation conditions, or natural wear. Warranty repair and
maintenance is carried out through an authorized dealer that sells it.
The warranty period for “CARSYS DPM-816 Pro” is 3 years from the date of
purchase of the device by the buyer. All the details are at https://car-
sys.com/en/ The device is designed and manufactured by: CARSYS, Vladimir
Chuvakin, Russia, 644007, Omsk, 80 Frunze st. CARSYS ELECTRONICS LIMITED, 160
City Road, London, EC1V 2NX United Kingdom
User Manual DPM-816 PRO
car-sys.com