iGear iG-PO1 Pulse Oximeter Digital Fingertip Monitor User Manual
- June 5, 2024
- iGear
Table of Contents
iGear iG-PO1 Pulse Oximeter Digital Fingertip Monitor
General instructions
The Fingertip Pulse Oximeter is used to detect the blood oxygen saturation and pulse rate of the human body. This product is based on digital blood oxygen technology and uses an advanced DSP algorithm to reduce the effects of motion artifacts and improve the measurement accuracy under weak perfusion.
Technical Parameters
- Display mode: Colourful LED display, 4 display directions.
- Power supply: 1.5 V (AAA size) alkaline batteries x2 Working current: 30mA Measurement performance under weak perfusion Conditions: 0.3%
Measuring Specifications:
Oxygen Saturation (SpO2)
Measuring range| 0-100%
Resolution| 1%
Accuracy| 70~100%, 2%, 0-69%
Pulse Rate|
Measuring range| 25-250 bpm
Resolution| 1 bpm
Accuracy| 2 bpm
Features
- The product is simple and convenient to operate.
- Small in size and light in weigh (including the total weight of the battery is about 50 grams), easy to carry.
- The power consumption is low. Can use Two AAA batteries.
- Show current battery usage. When the voltage is very low, the normal operation of the oximeter may be affected.
- When the finger is not in the measurement position for more than 16 seconds, the product will automatically turn off the power.
Product use range
The Fingertip Pulse Oximeter is used to measure the blood oxygen saturation
and pulse rate of the human body through fingers. The product is suitable for
families, hospitals (including medical and surgical departments,
anaesthesiology, pediatrics, intensive care, etc.), oxygen bars, community
medical, sports and health care.
This device is not intended for continuous monitoring.
Operation guide
- Press the power switch on the front panel to turn on the power pf Oximeter.
- Pinch the clip of the oximeter, and then put one of the fingers completely into cavity of the oximeter.
- Press the power switch again to change the display direction.
- Read the data on the display.
- When the finger is placed in the oximeter, the nail surface must be upward.
- If your fingers are not fully inserted, it may cause measurement errors.
Disclaimer: Before and after each test, please use medical alcohol to clean the oximeter rubber parts that touch your fingers (The rubber in the oximeter is medical rubbber, which is non-toxic and harmless to human skin).
The height of the bar graph can indicate the strength of the pulse.
-
Product accessories
- Lanyard x 1
- User manual x1
-
Battery installation
- Put the AAA batteries x2 into the battery box with correct polarity.
- Close the battery cover.
Note: The battery polarity must be installed correctly. Otherwise, the machine may be damaged. If the oximeter is not used for a long time, the battery must be removed.
-
Lanyard installation
- The thin thread at the end of the lanyard should pass through the perforation used to assemble the lanyard.
- Pass the thick thread at the head of the lanyard through the thin thread at the tail and tighten it.
-
Maintenance and storage
- When the battery is low, please replace the battery in time.
- Please clean the oximeter before diagnosing the patient.
- It is best to place the oximeter in an environment with a temperature of-10 to 400C and a humidity of 10%-95%.
- Avoid direct sunlight and exposure to strong radioactive infrared or ultraviolet rays.
- It is recommended to put the product in a dry environment. A humid environment may affect its life and even damage the product.
- Please dispose of used batteries in accordance with local government laws.
-
Dimensions
- Length: 58mm
- Width: 31mm
- Height: 29mm
- Weight: 50grams (including AAA batteries x2)
-
Environmental requirements
- Working temperature: 5-400C
- Storage temperature: 10-500C
- Ambient humidity: 15%-80% work 10%-95% storage
-
Faults and solutions
- There are no user-serviceable parts in the oximeter, and only qualified personnel can perform repairs.
- If you are not sure whether the measurement is accurate, you can first check the vital signs of the patient through other methods, and then ensure that the patient through other methods, and then ensure that the oximeter is in a normal working state.
- DO not pour, or spill any liquid into the accessories, connectors, switches, or openings in the oximeter because it will damage the oximeter.
Questions| Possible reason| Solution
---|---|---
Blood oxygen saturation and pulse rate cannot be displayed normally| 1. Incorrect finger
insertion.
2. The patient’s perfusion is too low.
| 1. Retry to insert your
finger
2. Try several more measurements, if you are sure that the product does not have any problems, please go to the hospital for an accurate diagnosis in time
The blood oxygen saturation and pulse rate display are unstable| 1. The finger may not
be fully inserted
2. Fingers are shaking or the patient’s body is moving
| 1. Try to insert fingers
again
2. Don’t move
Cannot turn on the power| 1. The battery is low or the battery is not installed
2. The battery is not installed correctly.
3. The oximeter is damaged
| 1. Please replace the battery
2. Reinstall the battery with the correct polarity
3. Please contact your local customer service center
The screen suddenly turns off the display| 1. Shut down without detecting a finger for a long time
2. The battery is exhausted
| 1. Normal
2. Replace the batteries
Sign definition
Manufactured and Marketed By:
iGear Tech Pvt. Ltd. 204, Sai Corporate Park, Laxmi Nagar, Off Link Road,
Goregaon West, Mumbai – 400104.
In case of consumer complaints:
- Toll free number – 1800 22 8495
- Email: sales@igear.asia
(Between 10am to 5pm, Mon- Fri)
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>