Milwaukee MI414 PRO Chloride Photometer User Manual

June 5, 2024
Milwaukee

Milwaukee MI414 PRO Chloride Photometer User Manual

Milwaukee MI414 PRO Chloride Photometer.JPG

MI414

milwaukeeinstruments.com (USA & CAN)
milwaukeeinst.com

Dear Customer,
Thank you for choosing a Milwaukee Instruments product. This manual will provide you with the necessary information for the correct use of the instrument. Please read it carefully before using the meter.

SPECIFICATIONS

FIG 1 SPECIFICATIONS.JPG

MEASUREMENT PROCEDURE

  1. Turn the meter on by pressing ON/OFF.
  2. When the LCD displays “- – -”, the meter is ready.
  3. Fill two cuvets, one with 10 mL of distilled water and one with 10 mL of the unreacted sample, up to the mark.                                                                                     
  4. Using the 1 mL syringe add 0.5 mL of Cl-1 reagent to each cuvet. Replace the caps and swirl gently for 30 seconds.

FIG 2 MEASUREMENT PROCEDURE.JPG

5. Remove the caps and using the second syringe add 0.5 mL of Cl-2 reagent to each cuvet. Replace the caps and swirl gently for 30 seconds.

FIG 3 MEASUREMENT PROCEDURE.JPG

Interferences

  • pH of the sample should be about 2. For alkaline samples, neutralize before adding reagents.
  • Color of the sample will cause interferences, they should be adequately treated before performing the test.
  • Suspended matter in large amounts should be removed by prior filtration.

GUIDE TO DISPLAY CODES

FIG 5 GUIDE TO DISPLAY CODES.JPG

ERROR MESSAGES

On Zero Reading

FIG 6 ERROR MESSAGES.JPG

On Sample Reading

FIG 7 ERROR MESSAGES.JPG

BATTERY REPLACEMENT

Battery replacement must only take place in a non-hazardous environment.

Simply rotate the battery cover on the back of the meter.

Detach the battery from the terminals and attach a fresh 9V battery while paying attention to the correct polarity. Insert the battery and replace the cover.

ACCESSORIES

CERTIFICATION

Milwaukee Instruments conform to the CE European Directives.

Disposal of Electrical & Electronic Equipment. Do not treat this product as household waste. Hand it over to the appropriate collection point for the recycling of electrical and electronic equipment.

Disposal of waste batteries. This product contains batteries. Do not dispose of them with other household waste. Hand them over to the appropriate collection point for recycling.

Please note: proper product and battery disposal prevents potential negative consequences for human health and the environment. For detailed information, contact your local household waste disposal service or go to www.milwaukeeinstruments.com (USA & CAN) or www.milwaukeeinst.com.

RECOMMENDATION
Before using this product, make sure it is entirely suitable for your specific application and for the environment in which it is used. Any modification introduced by the user to the supplied equipment may compromise the meter’s performance. For your and the meter’s safety do not use or store the meter in hazardous environment. To avoid damage or burn, do not perform any measurement in microwave ovens.

WARRANTY

These instruments are warranted against defects in materials and manufacturing for a period of 2 years from the date of purchase. This warranty is limited to repair or free of charge replacement if the instrument cannot be repaired. Damage due to accidents, misuse, tampering or lack of prescribed maintenance is not covered by warranty. If service is required, contact your local Milwaukee Instruments Technical Service. If the repair is not covered by the warranty, you will be notified of the charges incurred. When shipping any meter, make sure it is properly packaged for complete protection.

Read More About This Manual & Download PDF:

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals