STABILA TECH 1000 DP digital protractor Instruction Manual

June 5, 2024
STABILA

STABILA TECH 1000 DP digital protractor

Intended use

Congratulations on the purchase of your STABILA measuring tool. The STABILA TECH 1000 DP is a digital measuring tool for measuring inclinations.

If you still have questions after reading through the operating instructions, you can obtain advice by telephone:
+49  63 46  3 09  0

Equipment and functions

  • Tough, independent 360° digital protractor for quick and accurate measurements
  • Integrated rare-earth magnet for attachment
  • Integrated V-groove for aligning on round surfaces
  • Integrated T-groove for attachment
  • Integrated Li-ion rechargeable battery
  • Digital protractor with fast, direct transfer of measurement data via a RS485 interface
  • Digital protractor for measuring / monitoring via MODBUS-compatible data traffic
  • Carrying case
  • Mains adapter
  • RS 485 data cable <− > open
  • RS 485 data cable <− > USB (optional)
  • STABILA Analytics evaluation software (optional)

TECH 1000 DP

Components of the unit

STABILA TECH 1000 DP digital protractor FIG
1

  1. Display

  2. Mains adapter connection, M12 socket

  3. Rare-earth magnet

  4. T-groove profile for securing with M4 groove stones, e.g. Bosch Rexroth® or square nut in accordance with DIN 557

  5. V-shape for aligning on round surfaces

  6. Buttons:
    Units of measurement: °, %, mm/m, in/ft

  7. Function selection
    Lighting, acoustic guidance, keylock, unit settings, Auto OFF, baud rate, battery status

  8. Calibration, sensor adjustment

  9. Confirm entry

  10. On/Off

  11. Reference – freely selectable zero position

  12. HOLD – lock measurements

  13. Print mode – manual transmission of measurements
    Display elementsSTABILA TECH 1000 DP digital protractor FIG
2  

  14. Acoustic guidance: activated

  15. See chapter 7.4

  16. Keylock: activated

  17. Data traffic

  18. See chapter 4.1

  19. Units of measurement:  °, %, mm/m, in/ft

  20. See chapter 7.4

  21. Hold:  activated

  22. Reference: activated

  23. Position indicator

Commissioning

STABILA TECH 1000 DP digital protractor FIG
2STABILA TECH 1000 DP digital protractor FIG
3

Power supply

  • Charging the Li-ion rechargeable battery

The Li-ion rechargeable battery is charged using the mains adapter provided. Alternatively, the battery can be charged using the USB connection cable provided, as well as the M12 RS485 connection. The charging time depends on the maximum charging current of the source. Remaining connected to the mains adapter for a long period of time will not damage the Li-ion rechargeable battery.
Other power sources can damage the measuring tool!
Before using the unit for the first time, ensure that the rechargeable battery is fully charged!

Charging time: approx. 3 hours.

  • After 1 hour, the Li-ion rechargeable battery will be charged to approx. 80%.
  • Charging temperature range: 0 °C – 40 °C
  • Do not allow the Li-ion rechargeable battery to become fully discharged.
  • The performance of the Li-ion rechargeable battery is dependant on temperature.

LCD indicator:

  • Symbol not displayed – rechargeable battery is charged
  • Low battery level
  • If connected to the mains – rechargeable battery is being charged
  • If connected to the mains – rechargeable battery is fully charged

4.2 M12 socket pin assignment

When charging via the M12 socket, observe the following:

  • Correct polarity

  • Voltage:
    + 4.75 V (DC) … +5.25 V (DC)

  •  Charging current:
    100 mA … 2000 mA

Connection cable

  • Pin assignment for the enclosed connection cable to the M12 socket

Switching the unit on

  • After switching on with the “ON/OFF” button, an automatic test is carried out. All the display’s segments are shown.
  • After the end of the test, the version number S x.xx of the software is briefly displayed and the automatic switch-off time (Auto OFF) is shown.STABILA TECH 1000 DP digital protractor FIG 4

Functions

STABILA TECH 1000 DP digital protractor FIG
5

Visual guidance

  • Triangles representing the inclination indicate the position of the digital protractor in relation to the horizontal or vertical axis.
  • The 2 “centre display” bars indicate the precise position at which the vertical or horizontal axis is reached.
  • Display showing direction of inclination

Acoustic guidance

  • The acoustic guidance is selected via the “FUNC” button. The tone sequence speeds up as the 0°, 90°, 180° and 270° positions are approached in a range of +/- 2°.
  • A change in the pitch indicates that these positions have been exceeded.
  • A continuous signal tone confirms the precise point at which 0°, 90°,180° and 270° are reached.
  • This function is not active in interface mode.STABILA TECH 1000 DP digital protractor FIG 7

Automatic display inversion

The display is inverted for overhead measurements so that it is always legible.

Setting the “MODE” unit of measurementSTABILA TECH 1000 DP digital
protractor FIG 8

  • The unit of measurement is set by pressing the “MODE” button several times.
  • The set unit of measurement is retained after the unit is switched off.

Locking the measurement with “HOLD”
The current measurement can be locked by pressing the “HOLD” button. The relevant inclination triangle and the bars flash. The “Hold” symbol is displayed continuously. The measurement is displayed continuously.
The locked measurement is deleted by pressing the “HOLD” button again or switching the unit off.

Freely selectable zero position “REF”STABILA TECH 1000 DP digital
protractor FIG 10

  • The “REF” button can be used to select any set angle as 0° reference. The angle details now displayed relate to this reference angle. The displayed valueflashes with this setting.
  1. The reference angle value is displayed for 3 seconds by briefly pressing the “REF” button.
  2. The reference angle is deleted by:
    • Pressing and holding (≥ 3 sec) the “REF” button If the keylock is active, this must be disabled first.
    • Switching off
    • The automatic switch-off function

The zero position then refers back to the original setting.

The alignment selected for the digital protractor must not be changed during the reference function, as this could lead to a display error.

“FUNC” button settings

  • The user can switch between the different setting options by repeatedly pressing the “FUNC” button. While the display is flashing, the selected function can be confirmed with the “ENTER” button. If no button is pressed, the “FUNC” menu closes after a short time.

  • The “ON/OFF”, “FUNC” and “ENTER” buttons are always active.

  • With battery operation: switch between 0.2 and 2 hours. The unit is permanently on when connected to an external power supply.

The baud rate can be set from 1200 – 19,200 Bd.

  • The Analytics evaluation software can only be connected with a baud rate of 9600 Bd.STABILA TECH 1000 DP digital protractor FIG 12

Checking the measuring tool

STABILA TECH 1000 DP digital protractor FIG
13

Accuracy check

  • To prevent measuring errors, the accuracy of the measuring tool must be checked at regular intervals; for example, each time before beginning work, or after a heavy impact or extreme fluctuations in temperature.

    1. Step
      Place the unit with the lower measuring sole on as horizon-tal a surface as possible (e.g. a table) with the display side facing the user. Determine the measurement.

    2. Step
      Turn the unit by 180° in the same position.

    3. Step
      The rear of the unit is now facing the user.
      Calculating the error:
      0.00° − display Œ = A
      360.00° − display Ž = B
      If A+B is greater than ± 0.05°, the tool must be recalibrated.

Calibration -– Adjustment

By pressing the “CAL/ADJ” button repeatedly, the user can switch between CAL2P = calibration in relation to the meas-uring sole and ADJ4P = sensor adjustment. The selected function is confirmed by pressing the “ENTER” button.

CalibrationSTABILA TECH 1000 DP digital protractor FIG
14

  1. Step: Once “Calibration” has been selected with the “CAL/ADJ” button, confirm by pressing “Enter”.
    Display: CAL2P

  2. Step:
    Place the unit with the lower measuring sole on as horizontal a surface as possible (e.g. a table) with the display side facing the user. Calibration is started by pressing the
    “CAL/ADJ” button. “CAL” flashes in the display.
    Display: CAL2
    Calibration step 2 successfully completed

  3. Step:
    Turn the unit by 180° in the same position.

  4. Step:
    The rear of the unit is now facing the user. The second calibration is started by pressing the “CAL/ADJ” button.
    “CAL” flashes in the display.
    “rdy” display: Calibration completed successfully!

Adjusting the sensorSTABILA TECH 1000 DP digital protractor FIG
15

The sensor must be adjusted if the “temperature” or “Adj.” symbols are shown in the display.

  1. All 4 planes are adjusted during the sensor adjustment.

  2. The sensor can only be adjusted if the two black bars appear on the display (in the range of 0°, 90°, 180° and 270°).

  3. “ADJ” flashes during sensor adjustment of the respective plane.

  4. STABILA TECH 1000 DP digital protractor FIG 15Planes that have not been adjusted are not displayed. Successfully adjusted planes are permanently indicated in the display.

  5. Step 1:
    Once “Sensor adjustment” has been selected with the
    “CAL/ADJ” button, confirm by pressing “Enter”.
    Display: ADJ4P

  6. Step 2:
    Hold the unit in plane 1.
    Press the “CAL/ADJ” buttons.
    If the plane has been adjusted successfully, it is displayed permanently.

  7. Step 3:
    Turn the unit by 90° to plane 2.
    Press the “CAL/ADJ” button.
    If the plane has been adjusted successfully, it is displayed permanently.

  8. Step 4:STABILA TECH 1000 DP digital protractor FIG 17
    Turn the unit by 90° to plane 3.
    Press the “CAL/ADJ” button.
    If the plane has been adjusted successfully, it is displayed permanently.

  9. Step 5:
    Turn the unit by 90° to plane 4.
    Press the “CAL/ADJ” button.
    “rdy” display: Sensor adjustment completed successfully!

Error messages

The sensor must be adjusted if the “temperature” or “ADJ” symbols are shown in the display.STABILA TECH 1000 DP digital protractor FIG
18

Display: Error
The measuring unit must not be moved or subjected to vi-brations during the calibration/sensor adjustment. This can lead to measurement errors.

Display:
Measuring unit inclination around longitudinal axis > 10°

  • Display: mains connection/battery symbol flashes Temperature too high or too low to operate the rechargea-ble battery
  • Display: mains connection/battery symbol flashes quickly Charging voltage too high or too low
  • Display: only the lightning symbol flashes quickly Rechargeable battery is faulty

 Data transfer

protocol

The transmission protocol is configured in accordance with the MODBUS standard.
Characterformat:| 1 start bit, 8 data bits, 2 stop bits, no parity
Baud rate:| Default setting: 9600 Bd

Possible:              1200 Bd … 19,200 Bd

Idle period:| at least 3.5 characters between two messages
Unit address:| Default setting:        032 d

Possible: 001 d ….. 247 d

Please note: The data connection is interrupted if there is no query for

2 seconds.

To prevent errors when analysing measurements, the angle is only transmitted in degrees. The REF and HOLD settings are deleted.

| MODBUS function|

| Description
---|---|---|---
| | | Query current angle in 1/100°
| Query print angle in 1/100°
| Software version
| Serial number 1
| Serial number 2
| | | Change bus address
| Switch off measuring unit
| | | Analysis of data connection

Multiple-participant mode:
This mode enables multiple participants with different unit addresses but the same baud rate to connect to the MODBUS.

Querying the measurement

Structure of the read command function 03 h

| | | | | | |
| | | | **
Example: querying the current angle ( register ****] )
**
| | | | | | | ****

Structure of the response function 03 h

| | | | | |
| | | |
Example: response at 45.00°
20 h| 03 h| 02 h| 11 h| 94 h| 01 h| 70 h

Changing the bus address

Structure of the write command function 06 h

| | | | | | |
| | | |
**
Example: changing address to 16 d **
20 h| 06 h| 10 h| 04 h| 00 h| 10 h| CB h| 86 h

Structure of the response function 06 h

| | | | | | |
| | | |
**
Example: changing address to 16 d **
20 h| 06 h| 10 h| 04 h| 00 h| 10 h| CB h| 86 h

Error codes

An incorrect query is acknowledged with 8X h in the function code (second byte).

Error codes|
---|---
| | | | | | |
Addr| Function| Start address| Number of registers| CRC16
| 8X h| | | | | |

AUTO MODE:
A measurement is transmitted immediately after each query.
If the inclination of the measuring unit in the longitudinal axis is greater than 10° during the measurement, the TECH 1000 DP supplies the value FFFF h ( 65535 d ).STABILA TECH 1000 DP digital protractor FIG
19

PRINT MODE:

A query is sent from the PC to the measuring unit. If the “PRINT” button has not yet been  pressed, the TECH 1000 DP supplies the value CCCC h ( 52428 d ).

Otherwise, the TECH 1000 DP supplies the angle at the time the button was pressed.

If the inclination of the measuring unit in the longitudinal axis is greater than 10° during  the measurement, the TECH 1000 DP supplies the value FFFF h ( 65535 d ).

STABILA Analytics evaluation software (optional)

STABILA Analytics provides communication between a Windows PC and the TECH 1000 DP digital protractor made by the company STABILA Messgeräte GmbH. The TECH 1000 DP is connected to the computer via the data cable provided.
The Analytics evaluation software can only be connected with a baud rate of 9600 Bd.

Installation requirements:

  • TECH 1000 DP with the data cable provided (RS485 to USB)
  • PC with operating system Microsoft Windows XP SP3, Windows 7, Windows 8 or Windows 10
  • At least Windows installer version 4.5.6001.22159
  • .NetFramework 4

Technical data

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

STABILA User Manuals

Related Manuals