UNI-T UT715 Multifunction Loop Process Calibrator User Manual

June 5, 2024
UNI-T

UNI-T UT715 Multifunction Loop Process Calibrator User Manual
UNI-T UT715 Multifunction Loop Process Calibrator

Preface

Thank you for purchasing this brand new product. In order to use this product safely and correctly, please read this manual thoroughly, especially the safety notes.

After reading this manual, it is recommended to keep the manual at an easily accessible place, preferably close to the device, for future reference.

Limited Warranty and Liability

Uni-Trend guarantees that the product is free from any defect in material and workmanship within one year from the purchase date. This warranty does not apply to damages caused by accident, negligence, misuse, modification, contamination or improper handling. The dealer shall not be entitled to give any other warranty on behalf of Uni-Trend. If you need warranty service within the warranty period, please contact your seller directly.

Uni-Trend will not be responsible for any special, indirect, incidental or subsequent damage or loss caused by using this device.

Overview

UT715 is a high-performance, high-accuracy, handheld, multifunctional loop calibrator, which can be used in loop calibration and repair. It can output and measure direct current and voltage with a high accuracy of 0.02%, it has the functionalities of automatic stepping and automatic sloping output, these functionality help you to rapidly detect the linearity, the storage functionality facilitates the system setup, the data transferring functionality help the customers to rapidly test the communication.

Chart 1 Input and output function

Function Input Output Remark
DC millivolt -10mV – 220mV -10mV – 110mV
DC Voltage 0 — 30V 0 — 10V
DC Current 0 — 24mA 0 — 24mA
0 — 24 mA (LOOP) 0 — 24mA (SIM)
Frequency 1Hz – 100kHz 0.20Hz – 20kHz
Pulse   1-10000Hz The pulse quantity and range can be compiled.
Continuity SOON The buzzer beeps when the resistance is less than 2500.
24V Power   24V

Features

  1. The output accuracy and measurement accuracy reach up to 02%.
  2. It can output “Percentage”, users can easily get different percentage values by pressing
  3. It has the functionality of automatic stepping and automatic sloping output, these functions help you to rapidly detect the linearity.
  4. It can measure mA at the same time of providing the loop power to the
  5. It can save frequently-used setting
  6. The data transferring function helps you to rapidly test the
  7. Adjustable screen
  8. Rechargeable Ni-MH

Accessories

If any of accessories is missing or damaged, please contact your supplier.

  1. UT715: 1 piece
  2. Probes: 1 pair
  3. Alligator clips:1 pair
  4. Use Manual: 1 piece
  5. AA NI-MH battery: 6 pieces
  6. Adaptor: 1 piece
  7. USB cable:1 piece
  8. Cloth bag :1 piece

Operation

Please use the calibrator according to the user manual. “Warning” refers to potential hazard, “Attention” refers to the situation where would damage the calibrator or tested devices.

Warning

To avoid electric shock, damage, explosive gas ignition, please follow bellows:

  • Please use the calibrator according to this
  • Check before use, please do not use a damaged
  • Check the connectivity and insulation of the test leads, replace any exposed test
  • When using the probes, user only hold the protection end of the
  • Do not exert a voltage with more than 0V on any terminals and earth line.
  • If a voltage with more than 0V is applied on any terminals, the factory certificate will be out of effect, moreover, the device will be damaged permanently.
  • Correct terminals, modes, ranges must be used when it is on output
  • To prevent the tested device from being damaged, choose a correct mode before connecting the testing
  • When connecting the leads, first connect the COM test probe and then connect the other While disconnecting the lead, first disconnect the conducted probe and then disconnect the COM probe.
  • Do not open the calibrator
  • Before using the calibrator, please ensure that the battery door is tightly closed. Please refer the “Maintenance and Repair”.
  • When the battery power is insufficient , replace or charge the battery as soon as possible to avoid wrong reading value which may cause electric shock. Before opening the battery door, first remove the calibrator from “Dangerous Zone”. Please refer “Maintenance and Repair”.
  • Disassemble the test leads of the calibrator before opening the battery door.
  • For CAT I, the standard definition of measurement is applicable to the circuit that does not directly connect to a power
  • Specific replacement parts must be used when repairing the
  • The inside of the calibrator must be free from
  • Before using the calibrator, input a voltage value to check if the operation is
  • Do not use the calibrator wherever there is explosive powder
  • For battery, please refer to “Maintenance”.

Attention

To prevent the calibrator or the test device from being damaged:

  • The correct terminals, modes, ranges must be used when it is on output
  • When measuring and outputting current, correct earplug, functionality and ranges must be

Symbol

| Double insulated
---|---

| Warning

Specification

  1. The maximum voltage between the terminal and earth line, or any two terminals is
  2. Range: manually
  3. Operating : -10”C – 55”C
  4. Storage : -20”C – 70“C
  5. Relative Humidity: s95%(0°C – 30”C), 75%(30“C – 40”C), s50%(40“C – 50”C)
  6. Altitude: 0 – 2000m
  7. Battery: AA Ni-MH 2V•6 pieces
  8. Drop test: 1 meter
  9. Dimension: 224• 104 63mm
  10. Weight: About 650g (Including batteries)

Structure

Input terminal and Output terminal

Fig.1 and Fig. 2 Input and output terminal.

Input terminal and Output terminal Overview

No. Name Instruction

(1) (2)

| V, mV, Hz, , PULSE
Measurement/Output Port| (1)Connect red probe, (2)Connect black probe

(2) (3)

| mA, SIM Measurement/Output Port| (3)Connect red probe, (2) Connect black probe.
(3) (4)| LOOP Measurement Port| (4) Connect red probe, (3)Connect black probe.
(5)| Charge/Data Transfer Port| Connect to 12V-1A adaptor for recharging, or computer for data transmission

Button

Fig.3 Calibrator button, Chart 4 Description.

Figure 3

1

| | Power on/off. Long press the button for 2s.
---|---|---

2

| | Backlight adjustment.

3

|

MEAS

| Measurement Mode.
4| SOURŒ| Mode selection.
---|---|---
5| v| Voltage measurement/output.
6| mv| Millivolt measurement/output.
7

8

| mA| Milliampere measurement/output.
Hz| Short press the button to choose frequency measurement/output.
| “Continuity Test”.
10

11

| PULSE| Short press the button to choose pulse output.
100%| Short press to output the 100% value of the currently-set range, Iong press to reset the 100% values.
12| 25%| Short press to increase 25% of the range.
13| 25%| Short press to decrease 25% of the range.
14| 0%| Short press to output 0% value of the currently-set range,

Iong press to reset the 0% value.

15| | Arrow key. Adjust the cursor and parameter.
16| | Cycle selection:

Constantly output 0%-100%-0% at low slope (slow), repeat automatically.
Constantly output 0%-100%-0% at high slope (fast), repeat automatically.
At 25% of the step, step output 0%-100%-0%, repeat automatically.

17| RANGE| Switch range
18| SETUP| Short press to set up the parameter, Iong press to enter Menu.
19| ESC| ESC

LCD Display

Symbol Description Symbol Description
SOURCE Source output mode Battery power
MESUER Measurement mode LOAD Overload
Data adjustment prompt Progress output, slope output, step output
SIM Transmitter output simulation PC Remote control
LOOP Loop measurement AP0 Auto power off

Operation

This part introduces how to operate the UT715 calibrator.

  • Press for more than 2s to power on, LCD will display the model
  • Long press SETUP to enter system setup menu. Press the arrow key to set parameter, short press ESC to exit the setup
    system setup
    Figue 4 system setup
  1. Auto power off:
    Press to AUTO POWER OFF, press to set up the auto power off time. The AUTO POWER OFF time will start when no button is pressed, the counting will restart if any button is pressed. The max. AUTO POWER OFF time is 60 minutes, “0” means auto power off is disabled.

  2. Brightness:
    Press to select the BRIGHTNESS, press to adjust the screen brightness. Press on setup menu to rapidly adjust the brightness.

  3. Remote Control
    Press to select REMOTE CONTROL, press to set up for remote PC control.

  4. Button beep control
    Press to select BEEP CONTROL, press to set up button sound. “Beep” once enables button sound, “Beep” twice disables button sound.

Measurement mode

If the calibrator is on ‘Output’ status, press MEAS to switch to measurement mode

  1. Millivolt
    Press mV to measure the millivolt. Measurement page shown in figure 5. Connection shown in Figure 6.

Measurement mode Voltage
press to measure the voltage .Measurement page shown in figure 7. Connection shown in Figure 8.

Measurement mode

  1. Current
    Constantly press mA until it is switched to measure the milliampere. Measurement page shown in figure 9. Connection shown in Figure 10.
    Note: The buzzer beeps once the resistance is less than 2500

  2. Loop
    Constantly press mA until it is switched to measure the loop. Measurement page shown in figure 11. connection shown in figure 12.

  3. Frequency
    Press to measure the frequency. Measurement page shown in figure 13. Connection shown in Figure 14.
    Measurement mode

  4. Continuity
    Press to measure the continuity. Measurement page shown in figure 15. Connection shown in Figure 16.
    Measurement mode
    Note: The buzzer beeps once the resistance is less than 250 .

Source

Press SOURCE to switch to “Output Mode”.

  1. Millivolt
    Press mV to select millivolt output. Millivolt output page shown in figure 17. Connection shown in figure 18. Press the arrow key (right & left) to choose output digit, press the arrow key (up & down) to set the value.
    Source Induction

  2. Voltage
    Press to select voltage output. Voltage output page shown in figure 19. Connection shown in figure 20. Press the arrow key (right & left) to choose output digit, press the arrow key (up & down) to set the value.

Source Induction

  1. Current
    Press mA to select current output. Current output page shown in figure 21. Connection shown in figure 22.’ Press the arrow key (right & left) to choose output placement, press the arrow key (up & down) to set the value.

Note: If overload, the output value will flicker, the character “LOAD” will display, in this situation, you should check if the connection is correct for safety.

  1. SIM
    Press mA until the calibrator is switched to SIM Output. Passive current output shown in figure 23. Connection shown in 24, press the arrow key (right & left) to choose output placement, press the arrow key (up & down) to set the value.
    Note: The output value will flicker and the character “LOAD” will display when the output is overload, please check if the connection is correct for safety

  2. Source Induction

  3. Frequency
    Press Hz to select frequency output. Frequency output shown in figure 25, connection shown in 26, press the arrow key (right & left) to choose output placement, press the arrow key (up & down) to set the value.

    • Press “RANGE” to choose different ranges (200Hz, 2000Hz, 20kHz).
    • Short Press SETUP to display frequency modification page, as figure 25, in this page, you can modify the frequency by pressing the arrow key. After modification, if you short press SETUP again, the modification will become effective. Short press ESC to give up the modification
      Source Induction
  4. Pulse
    Press PULSE to select frequency output, pulse output page shown in figure 27, connection shown in figure 28, press the arrow key (right & left) to choose output placement, press the arrow key (up & down) to set the value.

    • Press RANGE to choose different ranges (100Hz, 1kHz, 10kHz).
    • Short press SETUP, it will be on the status of editing pulse quantity, then press the arrow key to edit the pulse quantity, short press SETUP again to complete pulse quantity setting, soon after that, it will be on the status of editing pulse range, then you can press the arrow key to edit the pulse range, short press SETUP to complete pulse range modification. The calibrator will output a specific quantity of pulse at a set frequency and range

Source Induction

Remote mode

Based on the instruction, turn on the PC Control Functionality, set the parameter of serial interface on PC and send the protocol command to control UT715. Please refer “UT715 Communication Protocol”.

Advanced application

Percentage

When the calibrator is on output mode, short press Percentage to rapidly output percentage value accordingly, the or  value of each output functionality is as below

Output Functionality 0%vaIue 100%vaIue
Millivolt 100mV 0mV 100mV
Millivolt 1000mV 0mV 1000mV
Voltage 0V 10V
Current 4mA 20mA
Frequency 200Hz 0Hz 200Hz
Frequency 2000Hz 200Hz 2000Hz
Frequency 20kHz 2000Hz 20000kHz

The or  value of each output can be reset by the following methods

  1. Press the arrow key to adjust the value and long press until the buzzer beeps, a new  value will be set as output value.
  2. Long press until the buzzer beeps, a new value will be set as output value

Note: The value must not be less than the   value.
Short press the output value will add % of the range between  value and % value.
Short press , the output value will decrease 25% range between value and value.

No t e: If you short press / or to adjust the value of output functionality, the output value shall not be greater than the  value and not be less than   value

Slope

The automatic output functionality of the slope can constantly provide a dynamic signal to the transmitter. If pressing , the calibrator will produce a constant and repeated slope (0%-100%-0%). There are 3 kinds of slope:

  1. 0%-100%-0% 40 seconds, smooth
  2. 0%-100%-0% 15 seconds, smooth
  3. 0%-100%-0% 25% progress slope, each step keeps for 5

If you want to exit the slope functionality, please press any key except for the slope key.

Indicator

Unless otherwise specified, the calibration period of all indicators is one year, the applicable temperature is +18”C to +28”C, the warm-up time is assumed as 30 minutes

Input Indicator

Indicator Range Resolution Accuracy
DC voltage 200mV 0.01mV +(0.02%+ 5)
30V 1mV ?(0.02%+2)
DC current 24mA 0.001mA ?(0.02%+2)
24mA (LOOP) 0.001mA ?(0.02%+2)
Frequency 100Hz 0.001Hz +(0.01%+1)
1000Hz 0.01Hz +(0.01%+1)
10kHz 0.1Hz +(0.01%+1)
100kHz 1Hz +(0.01%+1)
Continuity detection SOON 10 2500 It beeps

NOTE:

  1. For those temperatures that are not within +18°C-+28°C, the temperature coefficient of -10°C 18°C and +28°C 55°C is +0.005%FS/°C.
  2. The sensitivity of frequency measurement: Vp-p 1V, waveform: Rectangular wave, sine wave, triangular wave, etc

Output indicator

Indicator Range Resolution Accuracy
DC voltage 100mV 0.01mV +(0.02% + 10)
1000mV 0.1mV +(0.02% + 10)
10V 0.001V +(0.02% + 10)
DC current 20mA @ 0 – 24mA 0.001mA +(0.02%+2)
20mA(SIM) @ 0 – 24mA 0.001mA 1(0.02%+2)
Frequency 200Hz 0.01Hz 1(0.01%+1)
2000Hz 0.1Hz 1(0.01%+1)
20kHz 1Hz -+(0.01%+1)
Pulse 1-100Hz 1cyc
1-1000Hz 1cyc
1-10000Hz 1cyc
Loop power supply 24V   +10%

NOTE:

  1. For those temperatures that are not within +18°C *28°C, the temperature coefficient of -10°C 18°C and+28°C 55°C is 0.005%FS/°C
  2. The max load of DC voltage output is 1mA or 10k0, the smaller load shall
  3. The max resistance of DC output: 10000@20mA

Maintenance

Warning: Make sure that the power is off before opening the rear cover of the calibrator or battery cover, and that he probe is away from input terminal and tested circuit.

General maintenance and repair

  • Clean the case by damp cloth and mild detergent, do not use abrasives or solvents. If there is any malfunction, stop using the calibrator and send it for repair.

  • Please ensure that the calibrator is repaired by professionals or designated repair center. Calibrate the meter once a year to ensure its performance.

  • If the meter is not in use, turn off the power. If the meter is not in use for a long time, please take out the batteries.

  • Ensure that the instrumentation is free from moisture, high temperature and strong electromagnetic fields.

Install or replace the battery (Figure 29)

NOTE: When the battery power display , it means the rest of battery power is less than 20%, to ensure that the calibrator can work normally, please replace the battery in time, otherwise the measurement accuracy might be affected. Please replace the old battery by 1.5V alkaline battery or 1.2V NI- MH battery

Install or replace the battery

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals