Anti-HCV II Reagent Kit
en
Anti-HCV II
04W56
H14970R01
B4W5G0
04W5660
Instructions must be carefully followed. Reliability of assay results cannot be guaranteed if there are any deviations from these instructions.
NAME
Alinity s Anti-HCV II Reagent Kit
Hepatitis C Virus (E coli, Recombinant) NS3 Helicase Antigens and Synthetic
Core Peptide
INTENDED USE
The Alinity s An ti-HCV II assay is a chemiluminescent microparticle
immunoassay (CMIA) used for the qualitative detection of antibodies to
hepatitis C virus (HCV) in human serum and plasma specimens on the Alinity s
System.
The Alinity s An ti-HCV II assay is intended to screen individual human
donors, including volunteer donors of whole blood and blood components, and
other living donors for the presence of anti-HCV. The assay is also intended
for use in testing serum and plasma specimens to screen organ donors when
specimens are obtained while the donor’s heart is still beating, and in
testing serum and EDTA plasma specimens to screen cadaveric (non-heart-
beating) donors. It is not intended for use on cord blood specimens. This test
is not intended for use as an aid in the diagnosis of infection with HCV.
SUMMARY AND EXPLANATION OF THE TEST
Hepatitis C virus (HCV) is the causative agent of acute and chronic hepatitis
infection. Globally, an estimated 58 million individuals are chronically
infected. In 2019, approximately 290 000 people died from HCV-related liver
disease, primarily due to cirrhosis and hepatocellular carcinoma (primary
liver cancer). 1 HCV belongs to the genus Hepacivirus in the family
Flaviviridae and is a linear, single-stranded, positive-sense RNA virus. It is
divided into at least 6 different genotypes (1-6) and several subtypes based
on nucleotide sequence homology. 2 Each HCV genotype can be presen in any
given country, but there are geographical differences in prevalence.
Differences between genotypes are associated with responses to treatment. 3
HCV is spread through contact with blood from an infected person, such as
sharing needles to inject drugs. Less commonly, HCV is transmitted through
blood transfusion, sexual or perinatal routes or contact with contaminated
personal items. Because of effective blood screening using serological and
nucleic acid testing (NAT) methods, the risk of transfusion-transmitted HCV
infections has been reduced.
HCV RNA can be detected within a few days of exposure to HCV, prior to the
development of antibodies. 2 This time period, referred to as the pre-
seroconversion window period, often extends for several weeks after initial
infection with HCV. In general, antibodies to HCV are absent in the early
weeks of infection and are not detected on average until 8-11 weeks after
infection. 5 In general 55%-85% of HCV infected individuals develop chronic
infection, which is characterized by the continued detection of both HCV RNA
and antibodies to HCV, persisting for decades after initial infection. 1, 2
About 30% of infected individuals resolve their infection, which is
characterized by continued detection of antibodies to HCV, but with HCV RNA no
longer being detectable. 1, 2 Anti-HCV assays are used to identify individuals
infected with HCV and to prevent transmission of the virus to recipients of
blood or blood products. The Alinity s Anti-HCV II assay is designed to detect
antibodies to recombinant antigens representing Core and NS3 regions of the
HCV genome.
biological PRINCIPLES OF THE PROCEDURE
i This assay is for the qualitative detection of anti-HCV in human serum and
plasma using chemiluminescent microparticle immunoassay (CMIA) technology.
Sample, streptavidin-coated microparticles complexed with biotinylated HCV
constructs, acridinium-labeled HCV conjugate, and assay diluent are combined
to create a reaction mixture and incubated. The anti-HCV present in the sample
binds to the HCV-coated microparticles and to the acridinium-labeled HCV
conjugate. The mixture is washed. Ancillary wash buffer is added and
incubated. Following a wash cycle, Pre-Trigger and Trigger Solutions are
added.
The resulting chemiluminescent reaction is measured as relative light units
(RLU). There is a direct relationship between the amount of anti-HCV in the
sample and the RLU detected by the system optics. The presence or absence of
anti-HCV in the sample is determined by comparing the chemiluminescent RLU in
the reaction to the cutoff RLU determined from an active calibration. For
additional information on system and assay technology, refer to Alinity s
System Operations Manual, Section 3.
REAGENTS
Kit Contents
Alinity s Anti-HCV II Reagent Kit 04W56
NOTE: This product is composed of 4 components, which are packaged as a
2-cartridge reagent set. Both cartridges are required to perform the assay.
Volumes (mL) listed in the table below indicate the volume per cartridge set.
REF | 04W5660 |
---|---|
Tests per cartridge set | 500 |
Number of cartridge sets per kit | 5 |
Tests per kit | 2500 |
MICROPARTICLES | 27.0 mL |
CONJUGATE | 13.8 mL |
ASSAY DILUENT | 14.0 mL |
ANCILLARY WASH BUFFER | 26.5 mL |
MICROPARTICLES Streptavidin-coated microparticles complexed with
biotinylated HCV antigens (E Coli, recombinant) and biotinylated HCV Core
synthetic peptide in pyrophosphate-buffered saline with surfactants. Minimum
concentration: 0.097% solids. Preservative: sodium azide.
CONJUGATE Acridinium-labeled HCV antigens (E coli, recombinant) and
acridinium-labeled HCV Core synthetic peptide conjugate in pyrophosphate-
buffered saline. Minimum concentration: Peptide 50 ng/mL, antigens 200 ng/mL.
ASSAY DILUENT Preservative: sodium azide. Pyrophosphate-buffered saline
with surfactants. Preservative: sodium azide.
ANCILLARY WASH BUFFER Pyrophosphate-buffered saline. Preservative: sodium
azide.
Warnings and Precautions
- or In Vitro Diagnostic Use
- Performance characteristics of this product have not been established for laboratory diagnosis of HCV infection.
Safety Precautions
CAUTION: This product requires the handling of human specimens. It is
recommended that all human-sourced materials and all consumables contaminated
with potentially infectious materials be considered potentially infectious and
handled in accordance with the OSHA Standard on Bloodborne Pathogens.
Biosafety Level 2 or other appropriate regional, national, and institutional
biosafety practices should be used for materials that contain, are suspected
of containing, or are contaminated with infectious agents. 6-9
The following warnings and precautions apply to: MICROPARTICLES / CONJUGATE / ASSAY DILUENT/ ANCILLARY WASH BUFFER
Contains sodium azide.
EUH032| Contact with acids liberates very toxic gas.
P501| Dispose of contents/container in accordance with local regulations.
Follow local chemical disposal regulations based on your location along with
recommendations and content in the Safety Data Sheet to determine the safe
disposal of this product.
For the most current hazard information, see the product Safety Data Sheet.
Safety Data Sheets are available at
www.transfusion.abbott or contact your local
representative.
For a detailed discussion of safety precautions during system operation, refer
to Alinity s System Operations Manual, Section 8.
Reagent Handling
- Do not invert reagent cartridges.
- Upon receipt, reagent cartridges can be used immediately or stored in an upright position.
- If a reagent cartridge is dropped, place in an upright position for 1 hour before use to allow bubbles that may have formed to dissipate.
- Reagents are susceptible to the formation of foam and bubbles. Bubbles may interfere with the detection of the reagent level in the cartridge and cause insufficient reagent aspiration that may adversely affect results.
For a detailed discussion of reagent handling precautions during system operation, refer to Alinity s System Operations Manual, Section 7.
Reagent Storage
• Do not freeze.
| Storage Temperature| Maximum Storage Time| Additional Storage Instructions
---|---|---|---
Unopened| 2 to 8°C| Until expiration
date| Store in an upright position.
Opened| 2 to 15°C| 15 days after opening*| Store in an upright position.
Discard after 15 days.
If the cartridge does not remain upright during storage in the system, discard
the cartridge.
Do not reuse original reagent caps or replacement caps due to the risk of
contamination and the potential to compromise reagent performance.
- Includes time on board the system.
Reagents may be stored on or off the system. If removed from the system, store reagents with new replacement caps in an upright position at 2 to 15°C. For reagents stored off the system, it is recommended that they be stored in their original trays or boxes to ensure they remain upright.
For information on unloading reagents, refer to the Alinity s System
Operations Manual, Section 5.
Indications of Reagent Deterioration
Deterioration of the reagents may be indicated when a calibration error occurs or a control value is out of the specified range.
Associated test results are invalid, and samples must be retested.
Assay recalibration may be necessary.
For troubleshooting information, refer to the Alinity s System Operations Manual, Section 10.
INSTRUMENT PROCEDURE
The Alinity s Anti-HCV II Assay File must be installed on the Alinity s System
prior to performing the assay.
For detailed information on assay file installation and viewing and editing
assay parameters, refer to the Alinity s System Operations Manual, Section 2.
For information on printing assay parameters, refer to the Alinity s System
Operations Manual, Section 5.
For a detailed description of system procedures, refer to Alinity s System
Operations Manual.
lSPECIMEN COLLECTION AND PREPARATION FOR ANALYSIS
Specimen Types
The specimen types listed below were verified for use with this assay.
Other specimen types and anticoagulants have not been verified with this
assay.C
Specimen Type | Anticoagulant |
---|
Serum
(including serum separator tubes)| Not Applicable
Plasma| Dipotassium EDTA (including
plasma preparation tubes)
Tripotassium EDTA
Lithium heparin (including plasma
separator tubes)
Sodium citrate
Sodium heparin
ACD-A
ACD-B
CP2D
CPD
CPDA-1
- Liquid anticoagulants may have a dilution effect resulting in lower S/CO values for individual specimens.
- Performance has not been established for the use of umbilical cord blood or bodily fluids such as urine, saliva, semen, amniotic fluid, cerebrospinal fluid, or pleural fluid.
- Performance has been established for the use of cadaveric serum and EDTA plasma specimens (including specimens collected postmortem, non-heart-beating) that have been collected up to 24 hours after death. 10
- Testing of cadaveric serum and EDTA plasma specimens from patients with plasma dilution due to transfusions of > 2000 mL of blood or colloids within 48 hours, or > 2000 mL of crystalloids within 1 hour (or any combination thereof) prior to collection of the specimens has not been verified.
- For cadaveric donors, serum and EDTA plasma may be used; follow general standards and/or regulations for the collection, storage, and handling.
- The system does not provide the capability to verify specimen types. It is the responsibility of the operator to verify that the correct specimen types are used with the assay.
Specimen Conditions
- Do not use:
- heat-inactivated specimens
- pooled specimens
- grossly hemolyzed specimens
- specimens with obvious microbial contamination
- specimens with fungal growth
- For accurate results, serum and plasma specimens should be free of fibrin, red blood cells, and other particulate matter.
- To prevent cross-contamination, the use of disposable pipettes or pipette tips is recommended.
Preparation for Analysis
Failure to follow the specified centrifugation procedure may give erroneous or
inconsistent test results.
-
Clear, nonhemolyzed specimens should be used when possible.
Specimens containing visible particulate matter may give erroneous or inconsistent test results. -
Specimens should be free of bubbles. Remove bubbles with an applicator stick before analysis. Use a new applicator stick for each specimen to prevent cross ontamination.
-
Prior to centrifugation, previously frozen specimens (including previously frozen plasmapheresis specimens) must be mixed gently and thoroughly after thawing.
-
Specimens collected by plasmapheresis, which have not been frozen, do not require centrifugation. All other specimens (including previously frozen plasmapheresis specimens) must be centrifuged between 30 000 – 75 000 g-minutes.
-
All specimens must be tested or retested within 48 hours of initial centrifugation. After 48 hours, these specimens need to be recentrifuged between 30 000 – 75 000 g-minutes.
The acceptable time and force ranges that meet this criterion are listed in
the table below.
Convert rpm to RCF as follows: RCF = 1.12 × rmax (rpm/1000)2
Convert RCF to rpm as follows:
rpm|
---|---
RCF –| The relative centrifugal force is generated during centrifugation.
rpm –| The revolutions per minute of the rotor on which the specimens are
being spun (usually the digital
the readout on the centrifuge will indicate the rpm).
Centrifugation| The time should be measured from the time the
Time –| rotor reaches the required RCF or rpm to the time it begins
decelerating.
max –| The radius of the rotor in millimeters. The radius measured is
dependent on whether the rotor is a fixed angle rotor or a swinging bucket
rotor. This value is typically provided with the rotor by the manufacturer.
For the fixed angle rotor, max is the measure of the distance from the rotor
axis (center) to the bottom of the specimen tube in the rotor or rotor
adapter. For the swinging bucket rotor, max is the measure of the distance
from the rotor axis (center) to the bottom of the specimen tube in the rotor
adapter or bucket at full extension. NOTE: If custom tube adapters (i.e.,
adapters not defined by the centrifuge manufacturer) are used, then the radius
(max) should be manually measured in millimeters and the RCF calculated.
g-minutes –| The unit of measure for the product of RCF (× g) and
centrifugation time (minutes).
Specimen Storage
Specimen Type | Temperature | Maximum Storage Time | Special Instructions |
---|---|---|---|
Living Donor Serum/ Plasma | Room temperature (15 to 30°C) | 7 days | Specimens |
may be stored on or off the clot, red blood cells, or separator gel.
2 to 8°C| 14 days| Specimens may be stored on or off the clot, red blood
cells, or separator gel.
-20°C or colder| 9 months| Remove serum or plasma from the clot, red blood cells, or separator gel.
- Living donor specimens stored at -20°C or colder for greater than 9 months may be used for informational purposes (e.g., lookback testing, discordant sample testing, clinical and validation testing).
- Storage at a combination of 15 to 30°C and 2 to 8°C may not exceed 14 days (inclusive of shipping time) and cannot exceed the maximum durations listed in the table above.
- Performance has not been established for living donor specimens that have undergone more than 6 freeze/thaw cycles.
Specimen Type| Temperature| Maximum
Storage
Time Special Instructions
---|---|---
Cadaveric Serum/ EDTA
Plasma| Room temperature (15 to 30°C)| 3 days If specimens are not processed
directly after initial centrifugation, it is recommended to remove the
supernatant from the clot, red blood cells or separator gel until further
processing.
2 to 8°C| 14 days If specimens are not processed directly after initial
centrifugation, it is recommended to remove the supernatant from the clot, red
blood cells or separator gel until further processing.
-20°C or colder| 9 months If specimens are not processed directly
after initial centrifugation, it is recommended to remove the
supernatant from the clot, red blood cells or separator gel until further
processing.
- Cadaveric specimens stored at -20°C or colder for greater than 9 months may be used for informational purposes (e.g., lookback testing, discordant sample testing, clinical and validation testing).
- Storage at a combination of 15 to 30°C and 2 to 8°C may not exceed 14 days (inclusive of shipping time) and cannot exceed the maximum durations listed in the table above.
- Performance has not been established for cadaveric specimens that have undergone more than 6 freeze/thaw cycles.
Specimen Shipping
Package and label specimens in compliance with applicable state, federal, and
international regulations covering the transport of clinical specimens and
infectious substances.
PROCEDURE
Materials Provided
04W56 Alinity s Anti-HCV II Reagent Kit
Materials Required but not Provided
- Alinity s Anti-HCV II Assay File
- 04W5603 Alinity s Anti-HCV II Calibrator Kit
- 04W5620 Alinity s Anti-HCV II Assay Control Kit
- 04W5624 Alinity s Anti-HCV II Release Control Kit
- Alinity Trigger Solution
- Alinity Pre-Trigger Solution
- Alinity s Concentrated Wash Buffer
For information on materials required for the operation of the system, refer
to the Alinity s System Operations Manual, Section 1.
For information on materials required for maintenance procedures refer to
Alinity s System Operations Manual, Section 9.
Assay Procedure
For a detailed description of how to run an assay, refer to Alinity s System Operations Manual, Section 5.
-
Primary tubes may be on board the system for up to 10 hours.
-
If using primary or aliquot tubes, refer to the Alinity s System Operations Manual, Section 4 to ensure sufficient specimen is present.
-
To minimize the effects of evaporation, verify adequate sample cup volume is present prior to running the test.
-
Maximum number of replicates sampled from the same sample cup: 10
≤ 3 hours on the reagent and sample manager:
– Sample volume for first test: 350 µL
– Sample volume for each additional test from the same sample cup: 150 µL3 hours on the reagent and sample manager:
– Replace with a fresh aliquot of sample. -
Refer to the Alinity s Anti-HCV II Calibrator Kit, Assay Control Kit, and/or Release Control Kit package inserts for preparation and usage.
-
For general operating procedures, refer to the Alinity s System Operations Manual, Section 5.
-
For optimal performance, it is important to perform routine maintenance as described in Alinity s System Operations Manual, Section 9. Perform maintenance more frequently when required by laboratory procedures.
Calibration
For instructions on performing a calibration, refer to the Alinity s System
Operations Manual, Section 5.
Three replicates of Alinity s Anti-HCV II Calibrator 1 are automatically
tested by the system. The calibrator must be priority loaded.
Each assay control must be tested to evaluate the assay calibration.
Once calibration is accepted and stored, it may be used for 14 days. During
this time, all subsequent samples may be tested without further calibration
unless:
- A reagent kit with a new lot number is used.
- Daily quality control results are outside of quality control limits used to monitor and control system performance.
This assay may require recalibration after maintenance to critical parts or subsystems or after service procedures have been performed.
Quality Control Procedures
Assay Controls
The Alinity s Anti-HCV II Assay Controls must be tested once every 24 hours
when the system is being used.
Assay control values must be within the ranges specified in the Alinity s
Anti-HCV II Assay Control Kit package insert. When the assay control values
are within range, sample results are generated, and a valid release control
result is required to release test results. If an assay control value is not
within range, sample results are not generated for in-process or scheduled
samples. For troubleshooting information, refer to the Alinity s System
Operations Manual, Section 10.
Release Controls
The Alinity s Anti-HCV II Release Control must be tested in order to release
test results.
The release control is tested at user-defined intervals. For configuring the
release control, refer to the Alinity s System Operations Manual, Section 2.
For manually ordering the release control, refer to the Alinity s System
Operations Manual, Section 5.
The release control must meet specifications defined in Alinity s Anti-HCV II
Release Control Kit package insert in order to validate the system
functionality and release test results. If the release control does not meet
specifications, refer to the Alinity s System Operations Manual, Section 10,
for additional information.
Other Controls
Additional controls may be tested at the operator’s discretion in accordance
with local, state, and/or federal regulations or accreditation requirements
and your laboratory’s quality control policy. For additional information on
configuring customer controls, refer to Alinity s System Operations Manual,
Section 2.
Invalidate controls: Additional controls may be tested anywhere within a
run as an invalidate control. Specifications may be assigned to invalidating
controls. If an invalidate control fails to meet assigned specifications, no
sample results are calculated or provided by the system. When an invalidate
control meets assigned specifications, sample processing continues, and a
valid release control result is required to release test results.
Non-validating controls: Additional controls may be tested anywhere within a
run as a non-validating control. Specifications may be assigned to non-
validating controls. A valid release control result is required to release
test results. If the user-assigned specifications for the non-validating
control(s) are not met and the release control specifications are met, there
will be no effect on sample processing. In this case, reactive sample results
must not be considered invalid.
Quality Control Guidance
Refer to “Basic QC Practices” by James O Westgard, Ph.D. for guidance on
laboratory quality control practices. 11
RESULTS
Calculation
The Alinity s System calculates results for the Alinity s Anti-HCV II assay
using the ratio of the sample RLU to the cutoff RLU (S/CO) for each specimen
and control.
Cutoff RLU = Calibrator 1 Mean RLU x 0.219
The cutoff RLU is stored for each reagent lot calibration.
S/CO = Sample RLU/Cutoff RLU
Interpretation of Results
The cutoff is 1.00 S/CO.
Initial Results
Initial Result (S/CO) | Interpretation | Retest Procedure |
---|---|---|
< 1.00 | Nonreactive | No retest required. |
Specimen considered negative for antibodies to HCV.
1.00| Reactive| Retest in duplicate.
| Final Interpretation|
Retest Results (S/CO)| Final Results| Final Interpretation
Both results < 1.00| Nonreactive| Specimen considered negative for
antibodies to HCV.
One or both results in 1.00| Repeatedly Reactive| The specimen should be
further tested by supplemental methods.
Supplemental methods should follow appropriate FDA recommendations and
regulations for specimens found to be repeatedly reactive.
Customers outside the US must follow their country’s government
recommendations and regulations for specimens found to be repeatedly reactive.
Flags
Some results may contain information in the Flags field. For a description of
the flags that may appear in this field, refer to Alinity s System Operations
Manual, Section 5.
LIMITATIONS OF THE PROCEDURE
- Potential interference has not been evaluated for substances other than those described in the SPECIFIC PERFORMANCE CHARACTERISTICS – Interference section of this package insert.
- False reactive results can be expected with any test kit. Falsely elevated results have been observed due to non-specific interactions (refer to the SPECIFIC PERFORMANCE CHARACTERISTICS section of this package insert).
- Although the association of infectivity and the presence of antibodies to HCV is strong, it is recognized that presently available methods for antibodies to HCV detection are not sensitive enough to detect all potentially infectious units of blood or possible cases of HCV infection. A non-reactive test result does not exclude infection.
Refer to the SPECIMEN COLLECTION AND PREPARATION FOR ANALYSIS section of this package insert for specimen limitations.
SPECIFIC PERFORMANCE CHARACTERISTICS
Representative performance data are provided in this section. Results obtained
in individual laboratories may vary.
Reproducibility
A study was performed based on guidance from CLSI EP05-A3.
Testing was conducted using 3 lots of the Alinity s Anti-HCV II Reagent Kit,
Calibrator Kit, Assay Control Kit, and Release Control Kit. Panel members and
controls were tested twice a day for 5 days in replicates of 4 at 3 sites.
Sample| N| Mean
S/CO| Within-Run| Between-Run| Between-Day| Within-
Laboratory| Between-Site| Between-Lot| Reproducibility
---|---|---|---|---|---|---|---|---|---
SD| %CV| SD| %CV| SD| %CV| SD| %CV| SD| %CV| SD| %CV| SD| %CV
High| 360| 0.81| 0.022| 3.| 0.007| 0.9| 0.008| 1.0| 0.024| 3.0| 0.004| 0.5|
0.044| 5.| 0.051| 6.
Negative| | | | | | | | | | | | | | | |
Low HCV| 360| 1.| 0.031| 2.| 0.004| 0.3| 0.006| 0.5| 0.031| 3.| 0.010| 0.8|
0.065| 5.| 0.074| 6.
Antibody| | | | | | | | | | | | | | | |
High HCV| 360| 11.| 0.282| 3.| 0.000| 0.0| 0.000| 0.0| 0.282| 3.| 0.000| 0.0|
0.534| 5.| 0.610| 6.
Antibody| | | | | | | | | | | | | | | |
Positive| 360| 3.| 0.053| 2.| 0.000| 0.0| 0.000| 0.0| 0.053| 2.| 0.000| 0.0|
0.072| 3.| 0.093| 3.
Control| | | | | | | | | | | | | | | |
Negative| 360| 0.05| 0.003| NA| 0.001| NA| 0.001| NA| 0.003| NA| 0.001| NA|
0.011| NA| 0.012| NA
Control| | | | | | | | | | | | | | | |
%CV = Coefficient of Variation expressed as a percentage; N = Number of
Replicates; NA = Not Applicable: %CVs are not meaningful when S/CO approaches
zero; SD = Standard Deviation
a Includes within-run, between-run, and between-day variability
b Includes within-run, between-run, between-day, between-site, between-lot,
and site-lot interaction variability
Specificity
A total of 5277 fresh serum specimens and 7082 fresh plasma specimens from
volunteer whole blood donors were collected at 3 distinct blood centers. A
total of 3167 specimens from plasmapheresis donors were collected at one
additional blood center. Both the initial and repeat reactive rates for the
serum specimens were 0.11% (6/5277), both the initial and repeat reactive
rates for the plasma specimens were 0.11% (8/7082), and both the initial and
repeat reactive rates for the plasmapheresis donor specimens were 0.19%
(6/3167). Repeatedly reactive specimens were further tested using an HCV
qualitative RNA assay and an FDA-approved immunoassay for anti-HCV. Based on
supplemental test results for the repeatedly reactive specimens, 14 specimens
were positive, and 6 specimens were negative.
Specificity based on assumed zero prevalence of antibody to HCV in whole blood
and plasmapheresis donors was estimated in this study to be 99.96% (15 506/15
512) with a 95% confidence interval of 99.92% to 99.99%.
Specimen
Category| Number
Tested| IR
(% of Total)
(95% CI)| RR
(% of Total)
(95% CI)| Number
Positive by
Supplemental
Testing
(% of RR)| Specificity
(%)a
(95% CI)
---|---|---|---|---|---
Volunteer Blood
Donors – Serum
Volunteer Blood
Donors
– Plasma
Total Volunteer
Blood Donors
Plasmapheresis
Donors
Total Donors| 5277
7082
12 359
3167
15 526| 6
(0.11)
(0.04 – 0.25)
8
(0.11)
(0.05 – 0.22)
14
(0.11)
(0.06 – 0.19)
6
(0.19)
(0.07 – 0.41)
20
(0.13)
(0.08 – 0.20)| 6
(0.11)
(0.04 – 0.25)
8
(0.11)
(0.05 – 0.22)
14
(0.11)
(0.06 – 0.19)
6
(0.19)
(0.07 – 0.41)
20
(0.13)
(0.08 – 0.20)| 3
(50.00)
6
(75.00)
9
(64.29)
5
(83.33)
14
(70.00)| 99.94
(5271/5274)
(99.83 – 99.99)
99.97
(7074/7076)
(99.90 – 100.00)
99.96
(12 345/12 350)
(99.91 – 99.99)
99.97
(3161/3162)
(99.82 – 100.00)
99.96
(15 506/15 512)
(99.92 – 99.99)
CI = Confidence Interval; IR = Initially Reactive; RR = Repeatedly Reactive a
Based on supplemental test results for the 20 repeatedly reactive specimens,
14 specimens were positive (3 blood donor serum, 6 blood donor plasma, and 5
plasmapheresis donor specimens), and 6 specimens were negative (3 blood donor
serum, 2 blood donor plasma, and 1 plasmapheresis donor specimens); all 14
repeatedly reactive specimens found to be positive by supplemental testing
were excluded from the specificity calculations.
For total donors, the IR rate not reactive on retest was estimated to be 0.00%
(0/15 506) with a 95% confidence interval of 0.00% to 0.08%.
IR Rate Not Reactive on Retest = 100% × (Number of IR – Number of RR) /
(Number Tested – Number of RR)
Sensitivity
A total of 807 specimens from the categories shown in the table below were
tested using Alinity s Anti-HCV II assay at 3 clinical sites. Repeatedly
reactive specimens from individuals at increased risk of HCV infection were
tested using an HCV qualitative RNA assay and an FDA-approved immunoassay for
anti-HCV.
Sensitivity was estimated to be 100.00% (403/403) with a 95% confidence
interval of 99.09% to 100.00% for preselected positive specimens.
Specimen
Category| Number Tested| Number Positive| Number RR (% of Total)| Number RR
Positive by
Supplemental
Testing
(% of RR)| Sensitivity (%)
(95% CI)
---|---|---|---|---|---
Preselected
Anti-HCV
Positive
Individuals at
Increased Risk
of HCV
Infection b
Total| 403c
404
807| 403
70
473| 403
(100.00)
80
(19.80)
483
(59.85)| 403
(100.00)
70
(87.50)
473
(97.93)| 100.00
(403/403)
(99.09 – 100.00)
100.00
(70/70)
(94.87 – 100.00)
100.00
(473/473)
(99.22 – 100.00)
CI = Confidence Interval; RR = Repeatedly Reactive Preselected anti-HCV positive specimens were positive by anti-HCV assays and an FDA-licensed HCV RNA nucleic acid test. b The following risk factors were diagnosed or treated for HIV, sexual contact with HIV-infected individual, hemodialysis patient, high-risk sex behaviors, history of incarceration, illicit drug use (intravenous or intranasal), intranasal cocaine user, intravenous drug user, men who have sex with men, multiple sex partners, occupational exposure, tattoo, body piercing or acupuncture, persons with known exposure to HCV, and transfusion recipient (received transfusion prior to July 1992 or received a transfusion from HCV positive donor). c Three of the 403 preselected anti-HCV positive specimens were from individuals diagnosed with acute infection based on the medical diagnosis and HCV RNA and/or anti-HCV results.
Genotype Detection
A total of 131 preselected HCV positive specimens of known genotype
(genotypes 1-6) obtained from commercial vendors were tested using Alinity s
Anti-HCV II assay. The results were compared to a commercially available anti-
HCV assay. All 131 specimens were repeatedly reactive using the Alinity s
Anti-HCV II assay and the commercially available anti-HCV assay.
Seroconversion Sensitivity
The seroconversion detection of Alinity s Anti-HCV II assay was compared to an
FDA-licensed commercially available anti-HCV assay. Thirty-eight
seroconversion panel sets, consisting of 310 total panel members, obtained
from commercial vendors were tested using the Alinity s Anti-HCV II and
commercially available anti-HCV assays. In 15 panels, the Alinity s Anti-HCV
II and commercially available anti-HCV assays were reactive on the same bleed.
In 10 panels, the 2 assays were reactive within 1 bleed. In 9 panels, the
Alinity s Anti-HCV II assay was reactive 2 to 6 bleeds earlier than the
commercially available anti-HCV assay, and in 4 panels, all bleeds were
nonreactive with the commercially available anti-HCV assay while the Alinity s
Anti-HCV II assay was reactive in the last bleed.
Other Specimen Conditions or the Disease States
A total of 314 specimens from individuals with other disease states or medical
conditions unrelated to HCV infection were evaluated. Of the 314 specimens, 1
was repeatedly reactive using the Alinity s Anti-HCV II assay and was anti-HCV
negative by supplemental testing.
Specimen
Category| Number
Tested| IR (% of Total)| RR (% of Total)| Number Positive
by Supplemental
Testing (% of
Repeatedly
Reactive)
---|---|---|---|---
Other Specimen
Conditions or
Disease States| 314| 1 (0.32)| 1(0.32)| 0 (0.00)
IR = Initially Reactive; RR = Repeatedly Reactive The specimens included the following: Anti-HIV-1/HIV-2 Positive (12), Anti-HTLV I/II Positive (11), HBV Positive (25), Anti-HAV Positive (15), Anti-HDV Positive (12), Anti-CMV (15), Co-infected CMV/EBV/ HSV (14), Anti-T pallidum Positive (15), Non-viral Hepatitis (15), Rheumatoid Factor Positive (15), Anti-ds DNA Positive (11), Pregnant Females (15), Multiparous Females (15), Hyper IgG/IgM (11), Influenza Vaccine Recipient (15), Hemodialysis Patients (15), HAMA Positive (15), E Coli Infection (13), Heterophilic Antibody Positive (14), Fungal (Yeast) Infection (15), ANA Positive (13), and Autoimmune Hepatitis (13). b One HBV positive specimen was repeatedly reactive using the Alinity s Anti-HCV II assay and negative by supplemental testing.
Interference
Potentially Interfering Endogenous Substances
A study was performed testing potential interferents at or above the
concentrations recommended in CLSI EP37. 13
No interference was observed using Alinity s Anti-HCV II assay from
potentially interfering substances (spiked) at the levels shown below.
Potentially Interfering Substance | Interferent Level |
---|---|
Conjugated Bilirubin | ≤ 40 mg/dL |
Unconjugated Bilirubin | ≤ 40 mg/dL |
Hemoglobin | ≤ 1000 mg/dL |
Triglycerides | ≤ 3000 mg/dL |
Total Protein | ≤ 15 g/dL |
In addition, a negative and positive serum specimen was spiked with biotin to
a concentration of 4250 ng/mL. No interference was observed using the Alinity
s Anti-HCV II assay.
The effect of potentially interfering substances has only been evaluated for
those listed in this package insert.
PERFORMANCE CHARACTERISTICS OF CADAVERIC SPECIMEN TESTING
Reproducibility
Twenty-three cadaveric serum specimens, 23 cadaveric EDTA plasma specimens, 23
living donor serum specimens, and 23 living donor plasma specimens were spiked
with human plasma reactive for anti-HCV to create low-level reactive
specimens. Each specimen was tested once per day over 6 days using each of 3
lots of Alinity s Anti-HCV II Reagent Kit. Total %CV values were determined.
Specimen
Category| Specimen
Matrix| Number of
Replicates| Mean S/CO| Iotala
---|---|---|---|---
SD %CV
Cadaveric| Serumb
EDTA
Plasmas| 414
414| 3.41
3.35| 0.125 3.7
0.109 3.2
Living Donor| Serum
EDTA Plasma| 414
414| 3.93
3.90| 0.113 2.9
0.096 2.5
a Total variability contains within-specimen, between-lot, and lot-specimen
interaction variance components.
b Cadaveric serum specimens were collected up to 28.2 hours after death.
c Cadaveric EDTA plasma specimens were collected up to 39.8 hours after death.
Specificity
Specificity was determined by testing 55 cadaveric serum specimens, 55
cadaveric EDTA plasma specimens, 55 living donor serum specimens, and 55
living donor plasma specimens. Each specimen was tested once using each of 3
lots of Alinity s Anti-HCV II Reagent Kit.
Specimen
Category| Specimen
Matrix| Lot| Nonreactive| Repeatedly
Reactive| Specificity (%)
(95% CI)
---|---|---|---|---|---
Cadaveric| Serums| 1| 55| 0| 100.00
(93.51 – 100.00)
2| 55| 0| 100.00
(93.51 – 100.00)
3| 55| 0| 100.00
(93.51 – 100.00)
EDTA
Plasma| 1| 55| 0| 100.00
(93.51 – 100.00)
2| 55| 0| 100.00
(93.51 – 100.00)
3| 55| 0| 100.00
(93.51 – 100.00)
Living Donor| Serum| 1| 55| 0| 100.00
(93.51 – 100.00)
2| 55| 0| 100.00
(93.51 – 100.00)
3| 55| 0| 100.00
(93.51 – 100.00)
Plasma| 1| 55| 0| 100.00
(93.51 – 100.00)
2| 55| 0| 100.00
(93.51 – 100.00)
3| 55| 0| 100.00
(93.51 – 100.00)
CI = Confidence Interval
Cadaveric serum specimens were collected up to 28.2 hours after death.
b Cadaveric EDTA plasma specimens were collected up to 39.8 hours after death.
Analytical Sensitivity
Cadaveric serum and EDTA plasma specimens and living donor serum and plasma
specimens were spiked with human plasma reactive for anti-HCV to create low-
level and high-level reactive specimens. Each specimen was tested once, within
24 hours of spiking, using each of 3 lots of the Alinity s Anti-HCV II Reagent
Kit.
All specimens were reactive on all 3 reagent lots.
Specimen
Category| Analyte
Level| Specimen
Matrix| Lot| Number of
Specimens|
Mean
S/CO
| Sensitivity (%)
(95% CI)
---|---|---|---|---|---|---
Cadaveric| Low
Positive| Serum| 1| 55| 3.69| 100.00
(93.51 – 100.00)
2| 55| 3.63| 100.00
(93.51 – 100.00)
3| 55| 3.69| 100.00
(93.51 – 100.00)
EDTA
Plasma| 1| 55| 3.66| 100.00
(93.51 – 100.00)
2| 55| 3.67| 100.00
(93.51 – 100.00)
3| 55| 3.69| 100.00
(93.51 – 100.00)
High
Positive| Serum| 1| 55| 9.27| 100.00
(93.51 – 100.00)
2| 55| 8.97| 100.00
(93.51 – 100.00)
3| 55| 9.26| 100.00
(93.51 – 100.00)
EDTA
Plasma| 1| 55| 9.14| 100.00
(93.51 – 100.00)
2| 55| 8.97| 100.00
(93.51 – 100.00)
3| 55| 9.09| 100.00
(93.51 – 100.00)
Living
Donor| Low
Positive| Serum| 1| 55| 4.31| 100.00
(93.51 – 100.00)
2| 55| 4.18| 100.00
(93.51 – 100.00)
3| 55| 4.32| 100.00
(93.51 – 100.00)
Serum| 1| 55| 4.09| 100.00
(93.51 – 100.00)
2| 55| 4| 100.00
(93.51 – 100.00)
3| 55| 4.1| 100.00
(93.51 – 100.00)
High
Positive| Serum| 1| 55| 10.89| 100.00
(93.51 – 100.00)
2| 55| 10.43| 100.00
(93.51 – 100.00)
3| 55| 10.86| 100.00
(93.51 – 100.00)
Plasma| 1| 55| 10.6| 100.00
(93.51 – 100.00)
2| 55| 10.08| 100.00
(93.51 – 100.00)
3| 55| 10.54| 100.00
(93.51 – 100.00)
CI = Confidence Interval
Cadaveric serum specimens were collected up to 28.2 hours after death.
b Cadaveric EDTA plasma specimens were collected up to 39.8 hours after death.
Cadaveric Specimen Storage
Cadaveric specimen storage was determined by testing a minimum of 10 low-
level reactive cadaveric serum specimens, 10 low-level reactive cadaveric EDTA
plasma specimens, 10 nonreactive cadaveric serum specimens, and 10 nonreactive
cadaveric EDTA plasma specimens. The low-level reactive specimens were
prepared by spiking 10 nonreactive cadaveric serum specimens and 10
nonreactive cadaveric EDTA plasma specimens to a target S/CO value near the
cutoff with human plasma reactive for anti-HCV. Each specimen was tested at
Day 0, and then subjected to either 2 to 8°C storage for 14 days, room
temperature (15 to 30°C) storage for 3 days, −20°C or colder storage for 12
months, or 6 freeze/thaw cycles. Nonreactive specimens were evaluated by
calculating the differences between the mean S/CO of Day 0 and the mean S/CO
of each storage condition and related timepoint. Reactive specimens were
evaluated by calculating the percent differences between the mean S/CO of Day
0 and the mean S/CO of each storage condition and related timepoint. There
were no changes to the interpretation; the data demonstrate that cadaveric
serum and EDTA plasma specimens can be stored at the following conditions when
tested using the Alinity s Anti-HCV II assay.
Storage
Condition| Timepoint| Specimen Matrix| Nonreactive
Specimens
Upper Limit of
2-sided 95% CI of
Differences| Reactive
Specimens
Lower Limit of
2-sided 95% CI of
% Differences
---|---|---|---|---
Room
Temperature
(15 to 30°C)
2 to 8°C
-20°C or colder
Freeze/Thaw| 3 days
14 days
12 months
6 cycles| Serum
EDTA
Plasma
Serum
EDTA
Plasma
Serum
EDTA
Plasma
Serum
EDTA
Plasma| -0.01 S/CO
-0.01 S/CO
0.00 S/CO
0.00 S/CO
0.00 S/CO
0.02 S/CO
0.01 S/CO
0.01 S/CO| -10.6%
-9.0%
-8.6%
-9.2%
6.2%
5.4%
0.1%
1.1%
CI = Confidence Interval
Hemoglobin levels of cadaveric serum specimens ranged from 37 to 902 mg/dL.
b Cadaveric serum specimens were collected up to 25.5 hours after death.
c Hemoglobin levels of cadaveric EDTA plasma specimens ranged from 10 to 210
mg/dL.
d Cadaveric EDTA plasma specimens were collected up to 24.5 hours after death.
Refer to the SPECIMEN COLLECTION AND PREPARATION FOR
ANALYSIS, Specimen Storage section of this package insert for the maximum
storage times allowed.
BIBLIOGRAPHY
- World Health Organization. Hepatitis C fact sheet. http://www.who.int/news-room/fact-sheets/detail/hepatitis-c. Updated June 24, 2022. Accessed July 8, 2022.
- Dienstag JL. Acute viral hepatitis. In: Longo DL and Fauci AS editors. Harrison’s Gastroenterology and Hepatology. McGraw-Hill; 2010:349– 3 7 7.
- Gower E, Estes C, Blach S, et al. Global epidemiology and genotype distribution of the hepatitis C virus infection. J Hepatol. 2014;61(1 Suppl):S45-57.
- Dwyre DM, Fernando LP, Holland PV. Hepatitis B, hepatitis C and HIV transfusion-transmitted infections in the 21st century. Vox Sang. 2011;100(1):92–98.
- Center for Disease Control. Viral Hepatitis. Hepatitis C Questions and Answers for Health Professionals. https://www.cdc.gov/hepatitis/hcv/ hcvfaq.htm#Ref12. Updated August 7, 2020. Accessed July 8, 2022.
- US Department of Labor, Occupational Safety and Health Administration, 29 CFR Part 1910.1030, Bloodborne pathogens.
- US Department of Health and Human Services. Biosafety in Microbiological and Biomedical Laboratories. 6th ed. Washington, DC: US Government Printing Office; June 2020.
- World Health Organization. Laboratory Biosafety Manual. 4th ed. Geneva: World Health Organization; 2020.
- Clinical and Laboratory Standards Institute (CLSI). Protection of Laboratory Workers From Occupationally Acquired Infections; Approved Guideline—Fourth Edition. CLSI Document M29-A4. Wayne, PA: CLSI; 2014.
- U.S. Department of Health and Human Services, Food and Drug Administration, Center for Biologics Evaluation and Research. Guidance for Industry Recommendations for Obtaining a Labeling Claim for Communicable Disease Donor Screening Tests Using Cadaveric Blood Specimens from Donors of Human Cells, Tissues, and Cellular and Tissue-Based Products (HCT/Ps), November 2004. http://www.fda.gov/CBER/guidelines.htm Accessed July 8, 2022.
- Westgard JO. Basic QC Practices. 3rd ed. Madison, WI: Westgard Quality Corporation; 2010.
- Clinical and Laboratory Standards Institute (CLSI). Evaluation of Precision of Quantitative Measurement Procedures: Approved Guideline—Third Edition. CLSI Document EP05-A3. Wayne, PA: CLSI; 2014.
- Clinical and Laboratory Standards Institute (CLSI). Supplemental Tables for Interference Testing in Clinical Chemistry. 1st ed. CLSI supplement EP37. Wayne, PA: CLSI; 2018.
Note for number formatting:
- A space is used as a thousand separators (example: 10 000 specimens).
- A period is used to separate the integer part from the fractionapart of a number written in decimal form (example: 3.12%).
Key to Symbols
| Consult instructions for use
---|---
| Manufacturer
| Sufficient for
| Temperature limitation
| Use by/Expiration date
ASSAY DILUENT| Assay Diluent
ANCILLARY WASH BUFFER| Ancillary Wash Buffer
CONJUGATE| Conjugate
CONTAINS: AZIDE| Contains Sodium Azide. Contact with acids liberates
very toxic gas.
DISTRIBUTED IN THE USA BY| Distributed in the USA by
INFORMATION FOR USA ONLY| Information needed for United
IVD| In Vitro Diagnostic Medical Device
LOT| Lot Number
MICROPARTICLES| Microparticles
PRODUCT OF GERMANY| Product of Germany
REF| List Number
SN| Serial number
Alinity and related brand marks are trademarks of Abbott. Other trademarks are the property of their respective owners.
Abbott GmbH
Max-Planck-Ring 2
65205 Wiesbaden
Germany
+49-6122-580
DISTRIBUTED IN THE USA BY
Abbott Laboratories
Abbott Park, IL 60064 the USA
Customer Service: Contact your local representative or find
country-specific contact information at
www.transfusion.abbott
US License No. 2095
Created July 2022.
©2022 Abbott Laboratories