CAPES TD-3128 Blood Pressure Monitoring System Instruction Manual
- June 5, 2024
- CAPES
Table of Contents
TD-3128 Blood Pressure Monitoring System
ri-champion®
Ver 1.0 2020/06 311-3128000-015
smartPRO & smartPRO+ (TD-3128)
BLOOD PRESSURE MONITORING SYSTEM OBERARM BLUTDRUCKMESSGERÄT SYSTÈME DE SURVEILLANCE DE LA PRESSION ARTÉRIELLE SISTEMA DI MONITORAGGIO DELLA PRESSIONE ARTERIOSA SISTEMA DE CONTROL DE PRESIÓN SANGUÍNEA
Operation Instruction / Gebrauchsanleitung / Mode d’emploi / Manuale dell’utente / Instrucciones de Operación /
EN
Dear ri-champion® smartPRO/smartPRO+ (TD-3128) owner:
Thank you for purchasing the smartPRO/smartPRO+ (TD-3128) Blood Pressure
Monitoring System. This manual provides important information to help you to
use this system correctly. Before using this product, please read the
following contents thoroughly and carefully.
With the compact size and easy operation of this smartPRO/smartPRO+ (TD3128)
Blood Pressure Monitoring System, you can easily monitor your blood pressure
by yourself at any time or place. In addition, this system can help you and
your healthcare professionals to monitor and adjust your treatment plans, and
keep your blood pressure under control.
If you have other questions regarding this product, please contact the place
of purchase.
IMPORTANT SAFETY PRECAUTIONS
READ BEFORE USE
1. Use this device ONLY for the intended use described in this manual. 2. Do
NOT use accessories which are not specified by the manufacturer. Other
cables and accessories may negatively affect EMC performance. 3. Do NOT use
the device if it is not working properly or damaged. 4. Do NOT use under any
circumstances on newborns or infants. 5. This device does NOT serve as a cure
for any symptoms or diseases. The data
measured are for reference only. Always consult your doctor to have the
results interpreted. 6. Keep the equipment and its flexible cord away from hot
surfaces. 7. Do NOT apply the cuff to areas other than the place directed. 8.
Use of this instrument in a dry environment, especially if synthetic materials
are present (synthetic clothing, carpets etc.) may cause damaging static
discharges that may cause erroneous results.
1
9. Do not use this instrument in close proximity to sources of strong
electromagnetic radiation, as these may interfere with the accurate operation.
10.Proper maintenance is essential to the longevity of your device. If you are
concerned about your accuracy of measurement, please contact local customer
service for help.
11. Be careful to strangulation due to cables and hoses, particularly due to
excessive length.
12.Do not try to maintain the device while it is in use. 13.Keep the unit out
of reach of small unsupervised children. 14.Do not try to modify the device to
prevent any dangers. 15.Do not place the device in liquid, nor put it where it
could fall into liquid. If the
device becomes wet, unplug the device before touching it. 16.For the reason of
maintain basic safety and essential performance in regards
to EMC, please always contact the manufacturer or the manufacturer’s
representative to report unexpected operation or event. Do not try to fix it
by yourself. 17.When in use, you should stay away from electromagnetic
radiation, such as the mobile in use. 18.Used in close proximity or stacking
to others, EMC must be tested and verified. 19.If abnormal behavior is
observed due to EM disturbances, please relocate the device accordingly.
KEEP THESE INSTRUCTIONS IN A SAFE PLACE
2
TABLE OF CONTENTS
BEFORE YOU BEGIN
04
INTENDED USE
04
TEST PRINCIPLE
04
METER OVERVIEW
04
DISPLAY SCREEN
05
SETTING THE METER
05
USER NUMBER SELECTION
08
TESTING YOUR BLOOD PRESSURE
09
BEFORE MEASUREMENT
09
SELECTING THE CORRECT CUFF
09
FITTING THE CUFF PROPERLY
09
PROPER MEASUREMENT POSITION
10
TAKING MEASUREMENTS
11
AVERAGING MEASUREMENT MODE
12
METER MEMORY
13
REVIEWING TEST RESULTS
13
DOWNLOADING RESULTS
13
MAINTENANCE
15
BATTERY
15
USING AC ADAPTER (optional)
16
CARING FOR YOUR METER
16
DETAILED INFORMATION
17
REFERENCE VALUES
17
SYSTEM TROUBLESHOOTING
18
ERROR MESSAGES
18
TROUBLESHOOTING
19
SYMBOL INFORMATION
20
SPECIFICATIONS
21
SYSTEM PERFORMANCE
21
BLOOD PRESSURE MEASUREMENT PERFORMANCE
21
3
BEFORE YOU BEGIN
INTENDED USE
The smartPRO/smartPRO+ (TD-3128) system is a system designed to measure blood
pressure non-invasively. It is intended for use at home and in clinical
settings by one with a good understanding of the operation instruction, where
the patient may be an operator. The device is not to be used for the diagnosis
or screening of hypertension or for testing on newborns.
TEST PRINCIPLE
Blood pressure is measured non-invasively at the arm based on oscillometric
method.
This device is NOT able to take measurements in the presence of common
arrhythmia, such as atrial or ventricular premature beats or atrial
fibrillation. It may produce reading error.
METER OVERVIEW
1
2
4
3
5
1 DISPLAY SCREEN
2 M BUTTON Enter the meter memory.
3 USER SELECTION BUTTON Assist you with user number selection.
4 ON/OFF BUTTON
5 BATTERY COMPARTMENT
6 AC ADAPTER PORT Connect to a power supply.
6 8
7 9 10
11
7 AIR JACK 8 PRESSURE CUFF
9 AIR TUBE
10 AIR PLUG 11 BT INDICATOR (ri-champion®
smartPRO+ only)
4
DISPLAY SCREEN
1
15
14
2
13
3
4
12
11 5
10 6
7
9
8
1 Error Warning
9 Pulse Rate
2 Systolic Pressure Value
10 Average
3 Diastolic Pressure Value
11 AC Adaptor Symbol
4 Low Battery Symbol
12 Units of Blood Pressure
5 Movement During Measurement 13 Diastolic Pressure Symbol
6 Memory Mode Symbol
14 Systolic Pressure Symbol
7 Pulse Rate Symbol
15 Time & Date
8 User Number
SETTING THE METER
Before using your meter for the first time or if you change the meter battery,
you should check and update these settings. Make sure you complete the steps
below and save the desired settings.
Entering the Setting Mode Start with the meter off. Press and firmly hold
on.
5
for 3 seconds until the meter turns
Setting the Date and the Time
With the year flashing, press Press to set.
until the correct year appears.
With the month flashing, press until the correct month appears.Press to set.
With the day flashing, press Press to set.
until the correct day appears.
With the hour flashing, press Press to set.
until the correct hour appears.
6
With the minute flashing, press appears. Press to set.
until the correct minute
Setting the Time Format
Press to select the desired time format –12h or 24h. Press to set.
Setting the Unit of Measurement
Press to select mmHg or KPa. Press to set.
Delete The Memory
While “dEL” and a flashing ” ” symbol appears on the display, if you do not
wish to delete the saved results, press to skip this step. To delete ALL the
results, press twice. ” ” and ” ” are displayed on the meter, which indicates
that all results have been deleted.
7
Enter the Bluetooth Pairing (ri-champion® smartPRO+ only) NOTE This step is
recommended when the user needs to pair this meter to a Bluetooth receiver for
the first time, or when user needs to pair this meter to another new Bluetooth
receiver. If you wish to enter the pairing mode, with “PAr OFF” shows on the
meter, press once and the meter will display “PAr On” to enable Bluetooth
pairing mode.
Congratulations! You have completed all settings!
NOTE · These parameters can ONLY be changed in the setting mode. · If the
meter is idle for 3 minutes during the setting mode, it will turn off auto-
matically.
USER NUMBER SELECTION
This system stores blood pressure measurements for up to four users. Each
user’s test results are stored separately under each user number.
1. Press to select the desired user number. 2. Press to confirm. The monitor
will repeat
the selected user number and then turn off automatically.
8
M
TESTING YOUR BLOOD PRESSURE
BEFORE MEASUREMENT
· Avoid caffeine, tea, alcohol and tobacco for at least 30 minutes before
measurement.
· Wait 30 minutes after exercising or bathing before measurement. · Sit or lie
down for at least 10 minutes before measurement. · Do not measure when feeling
anxious or tense. · Take a 5-10 minute break between measurements. This break
can be longer if
necessary, depending on your physical condition. · Keep the records for your
doctor as reference. · Blood pressure naturally varies between each arm.
Always measure your blood
pressure on the same arm.
SELECTING THE CORRECT CUFF
Two different cuff sizes are provided with the blood pressure monitoring
system: Small and Wide. Select the cuff size that best matches the
circumference of the patient’s upper arm.
S (Small size) 15-24cm (5.9-9.4 inches)
W (Wide size) 24 43 cm (9.4 16.9 inches)
FITTING THE CUFF PROPERLY
1. Connect the air plug of the tubing to the air jack of the meter.
2. Assemble the cuff as shown. The smooth surface should be inside of the
cuff loop and the metal D-ring should not touch your skin.
9
W
Pile Side Material Sewn Hook Material
D-Ring 2-3 cm Main Arterles
3. Stretch your left (right) arm in front of you with your palm facing up.
Slide and place the cuff onto your arm with the air tube and artery mark
region toward the lower arm.Wrap and tighten the cuff above your elbow. The
bottom edge of the cuff should be approximately 0.8″ to 1.2″ (2 cm to 3 cm)
above your elbow. Align the tube over the main arteries on the inside.
4. Leave a little free space between the arm and the cuff; you should be able
to fit two fingers between them. Clothing must not restrict the arm. Remove
all clothing covering or constricting the measurement arm.
5. Press the hook material firmly against the pile material. The top and bottom edges of the cuff should be tightened evenly around your upper arm.
The range index of cuff should fall within this range
PROPER MEASUREMENT POSITION
1. Sit down for at least 10 minutes before measuring.
2. Place your elbow on a flat surface. Relax your hand with the palm facing
up.
3. Make sure the cuff is about the same height as the location of your heart.
Press . Remain still and do not talk or move during the measurement.
WARNING If the cuff is relatively lower (higher) than the heart, the obtained
blood pressure value could be higher (lower) than the actual value. A 15 cm
difference in height may result in an error around 10 mmHg.
10
4. Measurement is in progress. After the meter is turned on, the cuff will begin to inflate automatically.
TAKING MEASUREMENTS
Always apply the pressure cuff before turning on the meter.
1. Press . All the LCD symbols will appear. Press to select
user ”
” (numbers from 1 to 4) whose test reading
will be stored in its memory. Then the cuff will begin to inflate
automatically.
2. The heart symbol ” the measurement.
” will flash when pulse is detected during
3. After the measurement, the meter displays the systolic pressure, diastolic pressure and pulse rate.
4. Press to switch off or it will switch off automatically after being idle
for 3 minutes.
5. The meter will enter Bluetooth mode automatically. (ri-champion® smartPRO+
only)
11
NOTE
· If you press during measurement, the meter will be turned off.
· If the pulse rate symbol is shown as ”
” instead of ” “, this indicates
that the meter has detected an irregular heart beat.
AVERAGING MEASUREMENT MODE
Always apply the pressure cuff before turning on the meter.
1. Press and hold until the meter displays
“AVERAGE”, then press to select user
”
” (numbers from 1 to 4) whose test
reading will be stored in its memory. Then the
cuff will begin to inflate automatically.
2. After the first measurement is completed, the meter will start counting down before the second measurement begins. The number on the right represents the remaining countdown between each measurement. The meter will take three (3) measurements consecutively with intervals of 20 seconds.
3. After taking three measurements, the results are averaged to produce the
blood pressure measurement with “AVERAGE” symbol display on the meter. Press
to turn off the meter.
12
METER MEMORY
Your meter stores the 400 most recent blood pressure test results along with
respective dates and times in the meter memory. To recall the memory, start
with the meter off.
REVIEWING TEST RESULTS
1. Press and release . The user symbol
”
” (numbers from1 to 4) appears
first, press to select which one you used
to store in the memory.
2. Press again, and the first reading you see is the last blood pressure
result along with date and time.
3. Press Review all test results stored in the meter.
4. Exit the memory Press and the meter will turn off.
DOWNLOADING RESULTS
Data Transmission via Bluetooth (ri-champion® smartPRO+ only) You can transmit
your blood pressure monitoring data from the meter to your device via
Bluetooth. Please contact your local customer service or place of purchase for
assistance.
13
How to Install and Update the Healthy Check App You must connect to the internet to download the app. The App Store or Google Play can be accessed by tapping the App Store or Play Store icon on your iOS or Android devices. It is simple and intuitive to use, for better understanding of your current condition and to achieve better blood pressure control.
System Requirement For the requirement of OS version, please find on App Store or Google Play when you download the app. Please contact your local customer service or place of purchase for assistance. Please note that you must complete the pairing between meter and Bluetooth receiver before transmitting data.
Pairing with your mobile device
1. Turn on the Bluetooth function on your mobile device. 2. Start with the
meter off. Press and firmly hold for 3 seconds until the meter
turns on. “PCL” will appear on the meter. 3. Follow the instruction of the
Healthy Check app to pair the device. (Ex. Search to
find the meter and then add it into app.) 4. After successfully pairing the
app with the device, the Bluetooth function of meter
shall be on before transmitting the data to the Healthy Check app.
Bluetooth indicator on the blood pressure monitor:
BLUETOOTH INDICATOR
STATUS
Flashing Blue
The Bluetooth function is on and waiting for connection.
Solid Blue
The Bluetooth connection is established.
NOTE · While the meter is in transmission mode, it will be unable to perform a
blood
pressure test. · Make sure your device supporting Bluetooth Smart Technology
has turned on
Bluetooth before transmitting the data and the meter is within the receiving
range. For the requirement of OS version, please find on App Store or Google
Play when you download the app. · The Bluetooth functionality is implemented
in different ways by the various mobile device manufacturers, the
compatibility issue between your mobile device and the meter may occur.
14
MAINTENANCE
BATTERY
Your meter comes with four (4)1.5V AA size alkaline batteries.
Low Battery Signal The meter will display either of the two messages below to alert you when the meter power is getting low.
1. The
symbol appears along with display messages: The
meter is functional and the result remains accurate, but it is
time to change the batteries.
2. The
symbol appears with E-b:
The power is not enough to do a test. You must change the
batteries immediately.
Replacing the Battery To replace the batteries, make sure the meter is turned
off.
1. Press the edge of the battery cover and lift it up to remove. 2. Remove
the old batteries and replace with four 1.5V AA size alkaline batteries. 3.
Close the battery cover.
NOTE · Replacing the batteries does not affect the test results stored in
memory. · As with all small batteries, these batteries should be kept away
from small
children. If swallowed, promptly seek medical assistance. · Batteries might
leak chemicals if unused for a long time. Remove the batteries
if you are not going to use the device for an extended period (i.e., 3 months
or more). · Properly dispose of the batteries according to your local
environmental regulations.
15
USING AC ADAPTER (optional)
Connect AC adapter to the meter. 1. Connect AC adapter plug to DC adapter jack
of the meter.
2. Plug AC adapter power plug into an electrical outlet. Press to start the
measurement.
Remove AC adapter from the meter. 1. When the meter is off, remove AC adapter
power plug from the electrical outlet.
2. Disconnect AC adapter plug from DC adapter jack of the meter.
CARING FOR YOUR METER
To avoid the meter attracting dirt, dust or other contaminants, wash and dry
your hands thoroughly before use.
16
Cleaning
· To clean the meter exterior, wipe it with a cloth moistened with tap water
or a mild cleaning agent, then dry the device with a soft dry cloth. Do NOT
flush with water.
· Do NOT use organic solvents to clean the meter. · Do NOT wash the pressure
cuff. · Do NOT iron the pressure cuff.
Meter Storage
· Storage condition: -25°C to 70°C (-13°F to 158°F), 10% to 95% relative
humidity. · Always store or transport the meter in its original storage case.
· Avoid dropping or heavy impact. · Avoid direct sunlight and high humidity.
DETAILED INFORMATION
REFERENCE VALUES
Human blood pressure naturally increases after reaching middle age. This
symptom is a result of continuous ageing of the blood vessels. Further causes
include diabetes, lack of exercise and cholesterol (LDL) adhering to the blood
vessels. Rising blood pressure accelerates hardening of the arteries, and the
body becomes more susceptible to apoplexy and coronary infarction.
Definitions and Classification of blood pressure levels according to 2018
ESC/ESH Guidelines for the management of arterial hypertension:
Category
Systolic (mmHg)
Diastolic (mmHg)
Optimal
< 120
and
< 80
Normal
120129
and/or
8084
High normal Grade 1 hypertension
130139 140159
and/or and/or
8589 9099
Grade 2 hypertension
Grade 3 hypertension
Isolated systolic hypertension
160179 180 140
and/or and/or
and
100109 110 < 90
Isolated systolic hypertension should be graded (1, 2, 3) according to systolic
blood pressure values in the ranges indicated, provided that diastolic values are
<90mmHg.
17
Source: The European Society of Hypertension and European Society of Cardiology Task Force Members. 2018 ESC/ESH Guidelines for the management of arterial hypertension. Journal of Hypertension: October 2018 – Volume 36 – Issue 10 – p 19532041.
SYSTEM TROUBLESHOOTING
If you follow the recommended action but the problem persists, or error messages other than the ones below appear, please call your local customer service. Do not attempt to repair by yourself and never try to disassemble the meter under any circumstances.
ERROR MESSAGES
MESSAGE
CAUSE
Inflation or pressure error.
WHAT TO DO
Please contact local customer service for help.
Blood pressure measurement error.
Refit cuff tightly and correctly. Relax and repeat the measurement. If error still remains, contact local customer service for help.
Appears when the cuff deflates too slow.
Please contact local customer service for help.
Appears when the cuff deflates too fast.
Problems with the meter.
Review the instructions and repeat the test. If the meter still does not work, please contact the local customer service for help.
18
Batteries are too low.
Repeat with new batteries or input AC adapter.
Bluetooth transmission errors. Please contact local customer service for help.
Appears when systolic or diastolic blood pressure value is above the pre-set range.
Review the instructions and repeat the test. If error still remains, contact local customer service for help.
Appears when systolic or diastolic blood pressure value is below the pre-set range.
TROUBLESHOOTING
1. If no display appears after pressing POSSIBLE CAUSE
Batteries exhausted.
Batteries incorrectly installed or absent.
. WHAT TO DO
Replace the batteries.
Check that the batteries are correctly installed.
2. If the heart rate is higher/lower than user’s average:
POSSIBLE CAUSE
WHAT TO DO
Movement during measurement.
Repeat measurement.
Measurement taken just after exercise.
Rest at least 30 minutes before repeating measurement.
19
3. If the result is higher/lower than user’s average measurement:
POSSIBLE CAUSE
WHAT TO DO
May not be in correct position while measuring.
Adjust to the correct position to measure.
Blood pressure naturally varies from Keep in mind for next measurement. time to time.
4. If the cuff inflates again while measuring: POSSIBLE CAUSE
WHAT TO DO
Cuff is not fastened.
Fasten the cuff again.
If user’s blood pressure is higher than the pressure the device has inflated, the device will automatically increase the pressure and start to inflate again. Stay relaxed and wait for the measurement.
SYMBOL INFORMATION
Symbol
Referent
Manufacturer
Serial number
Symbol
Referent
Type BF applied part
Resistant to liquid ingress
Temperature limit
Humidity limitation
0123 CE mark
Caution
Authorized representative in the European Community
Refer to instruction manual / booklet NOTE On ME EQUIPMENT “Follow
instructions for use”
RoHS Compliance
This device does not belong to household waste and must be returned to a
collection point for recycling electric and electronic devices according to
local laws. If it contains batteries, the batteries should be removed and
disposed in accordance with local regulations for separate collection of spent
batteries.
20
SPECIFICATIONS
SYSTEM PERFORMANCE
Model no.: TD-3128 Power Source: Four 1.5V AA alkaline batteries Battery Life:
200 times Size of Meter w/o Cuff: 141 (L) x 121 (W) x 72 (H)mm, 350g without
batteries Cuff Size: S (small) 15 24cm (5.9 9.4 inches) with air tube 80
cm
W (Wide size) 24 43 cm (9.4 16.9 inches) with air tube 80 cm Memory:
Maximum 400 memory records External Output: Bluetooth (ri-champion® smartPRO+
only) (Frequency:2.45GHz, Bandwidth:170MHZ, Modulation:GFSK, ERP:3.54 dBm)
Power Saving: Automatic power off if system idle for 3 minutes Operating
Conditions: 5°C to 40°C (41°F to 104°F), 15% to 93% relative humidity, 700 hPa
to 1060 hPa Storage / Transportation Conditions: -25°C to 70°C (-13°F to
158°F), 10% to 95% relative humidity Power Supply Input: DC + 6V / 1A (max)
via Power Plug IP Classification: IP21 Expected Service Life: 3 years
BLOOD PRESSURE MEASUREMENT PERFORMANCE
Heart Rate Range: 40 199 beat per minute Systolic Measurement Range: 60 mmHg
255 mmHg Diastolic Measurement Range: 30 mmHg 195 mmHg Pulse Rate
Measurement Range: 40 199 beats / minute Maximum Inflation Pressure: 280
mmHg Accuracy of Pressure: ±3 mmHg or ±2% of reading Accuracy of Pulse rate:
±4% of reading Measurement Unit: Either mmHg or KPa This device has been
tested to meet the electrical and safety requirements of: IEC/ EN 60601-1,
IEC/EN 60601-1-2, EN 301 489-17, EN 301 489-1, EN 300 328. Reference to
Standards: · EN 1060-3, NIBP-requirements · IEC60601-1 General requirement for
safety · IEC60601-1-2 Requirements for EMC · EN1060-4, NIBP clinical
investigation · AAMI / ANSI / IEC 80601-2-30, ANSI/AAMI/ISO 81060-2, NIBP
requirements
21
Manufacturer’s declaration-electromagnetic immunity
The TD-3128 is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the TD-3128 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environmentguidance (for home and professional healthcare environment)
Electrostatic discharge (ESD) IEC 61000-4-2
Contact: ±8 kV Air±2 kV, ±4 kV, ±8 kV, ±15 kV
Contact: ±8 kV Air±2 kV, ±4 kV, ±8 kV, ±15 kV
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/burst IEC 61000-4-4
± 2kV for power supply lines ± 1kV for input/ output lines
± 2kV for power supply lines
Not applicable
Mains power quality should be that of a typical home healthcare environment.
Surge IEC 61000-4-5
Voltage Dips, short interruptions and voltage variations on power supply input
lines IEC 61000-4-11
± 0.5kV, ±1kV line(s) to line(s) ± 0.5kV, ±1kV, ±2kV line(s) to earth
Voltage dips: 0 % UT; 0,5 cycle 0 % UT; 1 cycle 70 % UT; 25/30 cycles
Voltage interruptions: 0 % UT; 250/300 cycle
Power frequency (50, 60 Hz) magnetic field IEC 610004-8
30 A/m 50 Hz or 60 Hz
± 0.5kV, ±1kV line(s) to line(s) Not applicable
Mains power quality should be that of a typical home healthcare environment.
Voltage dips: 0 % UT; 0,5 cycle 0 % UT; 1 cycle 70 % UT; 25 cycles
Voltage interruptions: 0 % UT; 250 cycle
30 A/m 50 Hz
Mains power quality should be that of a typical home healthcare environment.
If the user of the TD3128 requires continued operation during power mains
interruptions, it is recommended that the TD-3128 be powered from an
uninterruptible power supply or a battery.
The TD-3128 power frequency magnetic fields should be at levels characteristic
of a typical location in a typical home healthcare environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
22
Manufacturer’s declaration-electromagnetic emissions
The TD-3128 is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the TD-3128 should assure that it is used in such an environment.
Emission test Compliance
Electromagnetic environment-guidance (for home and professional healthcare environment)
RF emissions CISPR 11
Group 1
The TD-3128 uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Class B Class A
The TD-3128 is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Voltage fluctuations / flicker emissions IEC 61000-3-3
Compliance
Manufacturer’s declaration-electromagnetic immunity
The TD-3128 is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the TD-3128 should assure that it is used in such and environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environmentguidance (for home and professional healthcare environment)
23
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
3 Vrms: 0,15 MHz 80 MHz 6 Vrms: in ISM and amateur radio bands between 0,15 MHz and 80 MHz
3 Vrms: 0,15 MHz 80 MHz 6 Vrms: in ISM and amateur radio bands between 0,15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m 80 MHz 2,7 GHz 80 % AM at 1 kHz
80 % AM at 1 kHz
10 V/m 80 MHz 2,7 GHz 80 % AM at 1 kHz
Portable and mobile RF communications equipment should be used no closer to
any part of the TD3128 including cables, than the recommended separation
distance calculated from the equation applicable to the frequency of the
transmitter.
Recommended separation distance: d = 1,2 P d = 1,2 P 80 MHz to 800 MHz d = 2,3
P 800 MHz to 2,7 GHz
Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in metres (m).
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These
guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the TD-3128 is used exceeds
the applicable RF compliance level above, the TD-3128 should be observed to
verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the TD-3128. b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m.
24
Recommended separation distance between portable and mobile RF communications equipment and the TD-3128
The TD-3128 is intended for use in an electromagnetic environment (for home and professional healthcare) in which radiated RF disturbances are controlled. The customer or the user of the TD-3128 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the TD-3128 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter (W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,7 GHz
d =1,2P
d =1,2P
d =2,3P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where p is the
maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies. NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
25
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>