HIDGEEM SJ-OX2A Homecare In Car Oxygen Concentrator User Manual

June 4, 2024
HIDGEEM

HIDGEEM SJ-OX2A Homecare In Car Oxygen Concentrator

HIDGEEM-SJ-OX2A-Homecare-In-Car-Oxygen-Concentrator-
product

Please read the user manual before using the machine The manufacture date is on the package or Product certificate.

Important Safety Instructions

Warning
To prevent power failure or possible malfunction of the oxygen concentrator, those who urgently need oxygen must configure other backup oxygen supply(eg oxygen cylinder, oxygen bag, etc)! This equipment is suitable for oxygen supplementation, not considered as life support or life extension!

Before Use

The machine is prohibited from being inverted, tilted and placed sideways in any case! Be sure to use a qualified socket! Please install the regulator before using it if the power supply voltage is unstable! Non-professional maintenance personnel should not open the case!

Machine Placement
Keep away from flammable and explosive occasions. The oxygen generator should be placed indoors in a ventilated place and protected from direct sunlight, kept away from walls and furniture for over 10cm.

Do not place the machine in the following environments: temperatures that are too high or too low, near heat and sources of ignition, humidity, obstruction, smoke, and pollution. Do not place debris or any liquid or liquid-filled utensils on the top of the machine. Do not place any objects on the bottom of the machine to prevent the temperature of the inlet and exhaust ports rise and causing the machine to shut down or oxygen concentration to decrease.

Attention
Do not smoke and keep away from light and dark sources of ignition during use to avoid the risk of fire. Do not use oil or grease on any part of the machine to avoid contamination and danger. Do not place the machine in a place where it is easily dropped into water or other liquids. The safe use period of this machine is 5 years from the date of use.

What is SJ-OX2A Oxygen Concentrator?

Overview

The main function of the BE04 Series home care oxygen generator is to produce oxygen. Users can improve their oxygen supply by inhaling oxygen, eliminating fatigue and restoring their body functions. Mainly applicable to the elderly, high-intensity mental or manual workers, game players, the weak, pregnant women, students and other physiological hypoxia populations.

Product Use Range
Produce oxygen for health and beauty care purposes at home.

Functions and Features

  • High-grade environmentally friendly shell, safe, reliable and portable.
  • Infrared remote control: Easy to control the machine remotely.
  • Timing function: Display the total working time on the big screen.
  • Timed shutdown function: The user can customise the oxygen uptake time according to their personal preferences and needs.
  • The compressor has with thermal protector, which can guarantee the safety of the compressor and the whole machine.

Working Principle
Oxygen concentrators are the most reliable, efficient, and convenient source of supplemental oxygen available today, the oxygen concentrator is electrically operated. The unit separates oxygen from room air which allows high-purity supplemental oxygen to be delivered to you through the oxygen outlet, although the concentrator filters the oxygen in a room, it will not affect the normal amount of oxygen in your room.

Specifications

  1. Power:AC220V, 50Hz-60Hz(AC220V 50Hz-60Hz countries or regions)
  2. AC110V, 60Hz (AC110V 60Hz countries or regions)
  3. Input Power:250VA
  4. Oxygen concentration:30%±5%∽93%±3%
  5. Oxygen Flow:2∽10L/Min
  6. Noise:<35±5dB(A)
  7. Net Weight:8.0±0.2kg
  8. Size:40.622.138.9(cm)/168.715.3inch
  9. Compressor high-temperature protection, shutdown
  10. Electrical Classification: ll
  11. Work system: continuous operation
  12. Normal working environment

Environment temperature range:10℃∽40℃;
Relative humidity range:30%∽75%;
Atmospheric pressure range:860hPa∽1060hPa;
Attention: When the storage and transportation temperatures are below 8℃, the equipment should be placed in a normal working temperature environment for more than four hours before use.

Storage and transportation environment
Environment temperature range:-20℃∽+60℃;
Relative humidity range:10%∽93%,no condensation phenomenon;
Atmospheric pressure range:700hPa∽1060hPa.
Attention: Oxygen concentrators should be stored in rooms without strong sunlight, corrosive gases and good ventilation to avoid severe shocks and inversion during transportation.

Opening Inspection
Before installing and operating the product, please check whether the appearance of the product is complete and whether the components are by the quantity and category shown in the packing list. If there is any omission or damage, please contact the distributor or manufacturer in time.

Packing List

  1. Machine: 1 unit
  2. User manual: 1 unit
  3. Nasal Oxygen Tube Device: 2 set
  4. Filters: 2 units (1 unit in the machine)
  5. Glass fuses: 2 units (1 unit in the machine)
  6. Humidifying tank:2 units
  7. Remote Control:1 unit ( No battery included)

Description of buttons and important parts

HIDGEEM-SJ-OX2A-Homecare-In-Car-Oxygen-Concentrator-fig-
\(1\)

  • Front Side
  • Back Side HIDGEEM-SJ-OX2A-Homecare-In-Car-Oxygen-Concentrator-fig- \(2\)
  • Touch Screen PanelHIDGEEM-SJ-OX2A-Homecare-In-Car-Oxygen-Concentrator-fig- \(3\)

Back Side

Cut the belts 1 and belts 2 before use
HIDGEEM-SJ-OX2A-Homecare-In-Car-Oxygen-Concentrator-fig-
\(4\)HIDGEEM-SJ-OX2A-Homecare-In-
Car-Oxygen-Concentrator-fig- \(5\)HIDGEEM-SJ-OX2A-Homecare-In-Car-Oxygen-Concentrator-fig-
\(6\)

Operation of the Machine

  • According to the packing list, after carefully inspecting all parts, wash the “humidifying tank” with a little edible white vinegar and water.
  • Fill in cold distilled water or purified water, but do not exceed the maximum water level, and tighten the humidifying tank cap (do not miss the sealing o-ring).

Note

  1. Do not miss the sealing o-ring
  2. If the “humidifying tank” does not fill with water, the machine can produce oxygen normally, but the user may feel a little dry when inhaling oxygen.
  3. Please refer to Section 3.
    • first, please place the humidifying tank in to correct position, insert the short silicone tube into the oxygen outlet of the machine, and protect the humidifying tank oxygen inlet.
    • The second is the insert pipeline of the nasal oxygen device into the oxygen outlet of the humidifying tank.
    • Third, please correct wearing a nasal oxygen device or headset oxygen device.
    • Please insert t plug of the power cord into the AC220V 50Hz power socket, please check the plug and socket. In the meantime, please ensure that the plug can be kept stable and not loose.
    • Please flick the power switch to“ ”, and the power button will light red, It is ready for the machine self-loading.
    • Next, please press the power button again, will light green, and the machine will enter normal working condition.
  4. The display screen will display the oxygen flows, oxygen concentration, timing and cumulative time. When the oxygen generator works, it emits a “wheeze-” sound every few seconds, which is the normal ventilation sound.

Operation of oxygen inhalation

  • Press the “flow +” button on the touch control panel to adjust the required flow, and the oxygen concentration corresponding to the flow will be displayed on the display screen. At the same time, there should be bubbles in the water tank and there should be oxygen output from the oxygen outlet on the top of the water tank.
  • Connect the hose end of the oxygen inhalation tube to the oxygen outlet on top of the water tank, and wear well the ear-type oxygen inhalation head or nose-suction head, then you can begin the oxygen inhalation.

Timing Setting

  • This machine has an automatic timing function; You can press the “timing +” button on the touch control panel to set up the time in the range of 00 minutes to 3 hours and 00 minutes.
  • When the machine is turned on if the display screen shows “timing time-hour-minute”, means that the automatic timing function is not set, and the machine will be in continuous operation mode until the power switch is turned off.
  • When the “Timing” button is pressed every time, the display number is increased by 15 minutes, indicating that the running time is increased by 15 minutes, if pressed and held for more than 1.5 seconds, the display number will continue to increase until the button is released.

Symbols and Meanings of Safety Requirements in this Machine

HIDGEEM-SJ-OX2A-Homecare-In-Car-Oxygen-Concentrator-fig-
\(7\)

Shutdown

  • If oxygen inhalation needs to be temporarily stopped during use, the power supply (switch) button on the touch control panel can be pressed to stop or restart the oxygen supply.
  • After use, first, pull out the oxygen suction tube or connecting tube from the oxygen outlet, take off the ear-type or nose-type oxygen suction head, turn off the power switch behind the machine, and then pull out the power plug to ensure that the power supply is completely cut off.

Remote control operation
The machine can be operated remotely by a remote controller. The key function of the machine is shown in Fig. F (schematic diagram of the remote controller).
Note: When the remote control is operated, it is better to be in front of the remote control receiving window. But this is not necessary. When the reflection conditions are appropriate, it can be operated in the range of 0-160 degrees. Symbols and Meanings of Safety

Requirements in this Machine
Usage Method of Oxygen Absorber of Ear Mail Type (Optional Accessories)

Function
The ear-type oxygen absorber is designed according to ergonomic design, which can be easily worn on the ear. The excellent personal feeling can make you feel very comfortable to use.

Usage method

  1. Wear a good ear-type oxygen inhaler when inhaling oxygen.
  2. Oxygen inhalation head cleaning: medical alcohol can be used twice.
  3. Use and Cleaning of Adding Liquid Bottle:  Open and pull out the top cap of the filling bottle, add the appropriate amount of aromatic essential oil, and then install the filling bottle in place, you can start using it.
  4. Cleaning: Usually, once a week, it can be cleaned with a lighter detergent (food grade) and then rinsed with clean water. The water in the filling bottle should be dried before use.

WARNING
The power supply must be disconnected for maintenance of the machine. Do not dismantle the shell in case of electricity.

Case cleaning

  • The outer casing must be cleaned at least once a month.
  • Wipe with a clean and soft wet cotton cloth or sponge, do not infiltrate the liquid into the machine crevice。

Clean or replace the filter element

  • The cleaning and replacement of filter elements are very important for protecting the compressor and adsorption tower and prolonging the service life of the machine. Please clean and replace it in time.

  • Special attention: If the filter element is not installed in the filter, or the filter element is not properly installed in place, or the filter element is not completely dry, the power supply shall not be connected and the machine shall not be operated, which may cause permanent and
    irreversible damage to the machine.

  • The filter element should be cleaned once every 15 natural days. The filter is located at the bottom of the machine. Pull out the filter and take out the filter element in the filter. After removal, please check the filter element the first time. If the filter element is black, please clean or replace it immediately.

  • Cleaning of filter element: Clean with light detergent (food grade) first, rinse with clean water. The filter element must be dry and then installed in the filter.

Replacing dry box and silica gel desiccant

  • Please often check the Dr. Y box and silica gel desiccant, if the bloat is happening for Dr. Y box or silica gel desiccant, please replace bloat units. (Please remove the bottom cover to check the silica gel desiccant)
  • It is every 7 natural days to check the status of Dr.’s box and desiccant.

Maintain your humidifying tank

  • Please dump the water in the water tank.
  • Clean the water tank once a week with clean water, and wash it repeatedly with clean water to ensure neat.
  • If there is a scale or smell in the water tank due to some reasons that have not been cleaned or changed for more than a week, it can be soaked in edible white vinegar for more than one hour, shake it vigorously and wash it with clean water.

Maintain your nasal oxygen tube kit
It should be cleaned every three days. The nose suction head on the oxygen inhalation tube should be cleaned after use every time. It can be soaked in edible white vinegar for about 8 minutes and then cleaned with water or wiped with medical alcohol. The oxygen pipe kit should be kept dry and clean after cleaning. The use time of the oxygen inhalation tube should be 30- 60 natural days according to the use frequency and environment.

Replacement of Glass Safety Pipe
Turn off the ship power switch on the back of the machine and unplug the power cord. The fuse is located on the back of the machine. The fuse seat cover can be removed and replaced by unscrewing counterclockwise with a cross screwdriver.
PS: Glass fuse model: 5.0A 250VAC, Ø5×20)

Troubleshooting Guide

HIDGEEM-SJ-OX2A-Homecare-In-Car-Oxygen-Concentrator-fig-
\(8\)

Attention
Please contact the manufacturer or agent if the fault can not be fixed. ( You can find the manufacturer information in the user manual or invoice)

Diagram of Oxygen Concentrator Operation

HIDGEEM-SJ-OX2A-Homecare-In-Car-Oxygen-Concentrator-fig-
\(9\)

Guidance on the Use of Electromagnetic Environment
To avoid being disturbed by a strong electromagnetic environment or other electrical products, please keep the machine at least 30 cm apart from other household appliances in use. Non-professionals are not allowed to dismantle the machine shell. If professional maintenance personnel need product technical information, please contact the manufacturer or after-sales service vice point.

After-sale service
The company is responsible for the quality problems caused by non-human factors within one week from the date of sale. Under normal use and maintenance conditions, the company will provide free maintenance (the compressor warranty period is three years) if the quality problems occur within one year from the date of production or within one year from the date of purchase (customer’s choice). Machines, users can according to the invoice and warranty card to the company’s after-sales service vice department, offices or dealers to repair, reasonable fees. If the user is unable to provide the invoice, the warranty period shall be confirmed by a one-month extension of the date of our machine’s factor y serial number.

After-sales service will not be provided for the below situation:

  1. Vulnerable and consumable products: ear-type or nose-breathing oxygen inhaler, glass safety pipe, filter element, and humidifying water tank.
  2. Water, drugs and other liquids enter the machine by the user, which causes the machine not to work properly.
  3. The machine was floored or rained.
  4. Unauthorized disassembly, repair or reform of the machine.
  5. Faults caused by accidental falls during use and handling
  6. Failure due to incorrect operation disaccordance with instructions
  7. Damage caused by unforeseen natural disasters such as fires, earthquakes, floods, etc.
  8. Removal scope of free warranty service described in the warranty card

Disposal of Waste and Residue
The disposal of waste and residue should conform to the relevant national laws and regulations. We reserve the right to change the technology and appearance of this product. If there is any change, we will not inform you otherwise.

Remark
please keep your order bill to claim your warranty Please send the warranty card with the machine and provide the invoice for after-service purposes or you can scan the QR code to contact our Tech. service engineer for fast technical support and after-service.
E-mail: service@hidgeemoxygen.com
Web: service@hidgeemoxygen.com

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals