Euronda 816042 Aquafilter Instruction Manual
- June 1, 2024
- Euronda
Table of Contents
Euronda 816042 Aquafilter
Product Information
Specifications:
- Product Name: Euronda O-Ring Kit 816042
- Manufacturer: Euronda Spa
- Ref Code: 816042
- Revision Date: 11/2021
Product Usage Instructions
Safety Precautions:
Always handle the O-Ring spares with care to avoid damage.
Ensure proper disposal of used resins as hazardous waste.
Installation Instructions:
- Inspect Components: Check all components (labeled A, B, C) in the O-Ring kit for any damage.
- Assembly: Follow the provided Fig. 2 for step-by-step installation instructions.
- Caution: Pay attention to step 2.3 to ensure proper installation and functionality.
Frequently Asked Questions (FAQ):
Q: How should I dispose of used resins?
A: Take used resins to a waste-disposal center and dispose of them as
hazardous waste, following the laws of your country.
Q: What should I do if I encounter any issues during installation?
A: Refer to the provided installation instructions carefully. If problems
persist, contact Euronda customer support for assistance.
Thank you for purchasing original Euronda spares.
Please note
- Ion exchange resin cartridges should be replaced when the red indicator light on the Aquafilter lights up;
- Once the new cartridges have been fitted, do not remove them until they have been used up, except where the following point applies;
- Before an extended period of non-use of the Aquafilter, empty the system to prevent the formation of bacteria and biofilm;
- To avoid compromising the performance of the device, both cartridges must be replaced;
- It is recommended that you replace the vessels’ O-Ring seal (for guidance in selecting the seal, see Fig. 1);
- The quality of the inlet water affects the durability of the resins. The higher the water-hardness and electrical-conductivity values, the fewer litres will be produced by the resin cartridges before a value of 15 microsiemens per cm stops the unit functioning. For example, if electrical conductivity is 300 μS/cm at the input, the quantity produced will be approximately 150 litres. For values near or above 600 μS/cm, the number of litres produced may be reduced by 30%. In this case, we recommend using the Aquaosmo reverse osmosis system.
Storage instructions
- Store at a temperature between 5°C and 35°C;
- Only remove from original packaging when fitting;
- Avoid exposure to sources of light and heat;
- Do not open or tamper with the cartridges.
Remember: take used resins to a waste-disposal centre and dispose of them as hazardous waste, complying with the laws in force in the country of use at all times.
Installation instructions
- If possible, before replacing the resins, relieve the pressure on the vessels;
- Unscrew the vessels containing the cartridges using the wrench supplied with the Aquafilter (Fig. 2.1);
- Insert the spacer ring into the cartridge, ensuring that it is positioned correctly, then insert the cartridge into the vessel with the seal facing upwards (Fig. 2.2);
- Screw the container back into the Aquafilter (Fig. 2.3);
- Repeat this step with the second cartridge.
Fig. 1: O-Ring spares
Fig. 2: Installation instructions
Euronda Spa
Via Chizzalunga 1
36066 Sandrigo
Vicenza, Italy
t +39 0444 656111
f +39 0444 656199
m info@euronda.com
www.euronda.it
51020260 – Ref Code: 816042 – Rev. 11/2021