Product Usage Instructions

Safety Precautions:

Always handle the O-Ring spares with care to avoid damage.
Ensure proper disposal of used resins as hazardous waste.

Installation Instructions:

  1. Inspect Components: Check all components (labeled A, B, C) in the O-Ring kit for any damage.
  2. Assembly: Follow the provided Fig. 2 for step-by-step installation instructions.
  3. Caution: Pay attention to step 2.3 to ensure proper installation and functionality.

Frequently Asked Questions (FAQ):

Q: How should I dispose of used resins?
A: Take used resins to a waste-disposal center and dispose of them as hazardous waste, following the laws of your country.

Q: What should I do if I encounter any issues during installation?
A: Refer to the provided installation instructions carefully. If problems persist, contact Euronda customer support for assistance.

Thank you for purchasing original Euronda spares.

Please note

  • Ion exchange resin cartridges should be replaced when the red indicator light on the Aquafilter lights up;
  • Once the new cartridges have been fitted, do not remove them until they have been used up, except where the following point applies;
  • Before an extended period of non-use of the Aquafilter, empty the system to prevent the formation of bacteria and biofilm;
  • To avoid compromising the performance of the device, both cartridges must be replaced;
  • It is recommended that you replace the vessels’ O-Ring seal (for guidance in selecting the seal, see Fig. 1);
  • The quality of the inlet water affects the durability of the resins. The higher the water-hardness and electrical-conductivity values, the fewer litres will be produced by the resin cartridges before a value of 15 microsiemens per cm stops the unit functioning. For example, if electrical conductivity is 300 μS/cm at the input, the quantity produced will be approximately 150 litres. For values near or above 600 μS/cm, the number of litres produced may be reduced by 30%. In this case, we recommend using the Aquaosmo reverse osmosis system.

Storage instructions

  • Store at a temperature between 5°C and 35°C;
  • Only remove from original packaging when fitting;
  • Avoid exposure to sources of light and heat;
  • Do not open or tamper with the cartridges.

Remember: take used resins to a waste-disposal centre and dispose of them as hazardous waste, complying with the laws in force in the country of use at all times.

Installation instructions

  • If possible, before replacing the resins, relieve the pressure on the vessels;
  • Unscrew the vessels containing the cartridges using the wrench supplied with the Aquafilter (Fig. 2.1);
  • Insert the spacer ring into the cartridge, ensuring that it is positioned correctly, then insert the cartridge into the vessel with the seal facing upwards (Fig. 2.2);
  • Screw the container back into the Aquafilter (Fig. 2.3);
  • Repeat this step with the second cartridge.

Fig. 1: O-Ring spares

Euronda-816042-Aquafilter-1

Fig. 2: Installation instructions

Euronda-816042-Aquafilter-2

Euronda Spa
Via Chizzalunga 1
36066 Sandrigo
Vicenza, Italy
t +39 0444 656111
f +39 0444 656199
m info@euronda.com
www.euronda.it

51020260 – Ref Code: 816042 – Rev. 11/2021