Sevenstar CS200-D Controller Mass Flow Meter
Instruction of MFC/MFM
Declaration
The copy right of The user’s manual of mass flow controller and mass flow meter is subjected to Beijing Seven star Flow Co., Ltd (following abbr. Sevenstar), which is not allowed to duplicate, store and distribute any part of this manual in mean of electric, mechanical, photocopy, recording or other way without permission of Sevenstar. This manual is not assured that there is no mistake and missing in which have been corrected strictly, and the publisher is under no obligation to the mistake or missing, meanwhile the publisher is not in charge of any loss that this manual leads to.
Attention
Dear customer, thanks for using CS mass flow controller and mass flow meter.
This manual describes in detail important issues about correct and safe
operations of the product.
User of the product should read and comprehend this manual and pay attention
to the text with labels and notices.
Sevenstar assumes no liability for the customer’s failure to comply with this
manual. This manual is necessary for your installation and maintenance, please
keep it carefully.
The notice of safety
Please pay attention to the below notices when reading this manual. We are not
responsible for any result without abiding by the below notices.
a) Do not replace any components or disassemble instrument.
Do not replace any components, or disassemble the instrument without any
authorized and make sure that the label or/and seal of product is not removed
when returning for reworking, recalibration and maintenance.
b) Please contact professional for technical service.
Do not replace any components. Any technical support is provided by
professional who must be authorized by Sevenstar.
c) Please pay more attention to use dangerous gas.
Instrument should be purged completely and keep safety if dangerous gas is
used. Meanwhile make sure that moist gas must not react with material of seal
and instrument.
d) Please pay attention to purging instrument.
The whole system should be purged by dry gas after and before the instrument
is installed.
e) Please abide by proper steps of purging.
Product should be purged and handled with gloves.
f) Do not use the instrument in explosive environment.
Do not use the instrument in explosive environment, unless the safety
certificate is available.
g) Please use proper fittings and keep the rules.
All fittings of instrument must be matched according to the listing in manual.
Please read manual carefully before screwing tightly.
h) Please do leakage check.
Please check carefully all of vacuum parts and make sure that there is not
leakage in the system.
i) Please make sure that instrument is working under safe pressure.
Please make sure that the pressure of inlet gas is less than maximum working
pressure (referring to maximum working pressure in manual).
j) Please keep the whole system away from pollution.
When system running, do not use polluting gas, such as particle of dust, dirt,
fibre, glass or scrap iron.
k) Please do warm-up instrument before working.
Please do warm-up instrument, especially in using dangerous gas. Please close
valve completely to make sure that there is not error flow.
General
Mass flow controller (MFC) accurately measures and controls mass flow rates,
which is widely applied in the fields as: semiconductor and IC fabrication,
special materials science, chemical industry, petrolic industry,
pharmaceutical industry, environmental protecting and vacuum system
researching, etc.. The typical applications include: microelectronic process
equipment such as diffusion, oxidation, epitaxy, CVD, plasma etching,
sputtering, ion implantation, vacuum deposition equipment, optical fiber
melting, micro-reaction equipment, mixing & matching gas system, capillary
flow control system, gas chromatograph and other analytical instruments.
The CS200-D MFC/MFM are a latest generation MFC for use in semiconductor
applications and demanding industrial uses where its high accuracy and
flexibility in interfacing are required.
The CS200-D MFC/MFM incorporate a dual interface, voltage and current as well
as RS-485,DeviceNet and ProfiBus digital interfaces. It is possible to operate
the instrument completely digitally or it can be operated in analog mode with
digital monitoring. CS200-D series have wide range of power supply (available
for ±8 ~ ±16 VDC or +14 ~ +28 VDC).
In addition, auto-alarm, change of operating gas and range available through
the digital interface. Customer Secondary Development of control and sample
software is available through the open protocol.
CS200-D MFC default setting:
MAC address: 32;
RS485 baud rate: 19200;
Control mold: 0-5V analog signal control.
For more information, please refer to
www.mfcsevenstar.cn.Path: Service >>
Downloads>>Software download>>Communication Protocol .
Specification
Type| CS200-D|
---|---|---
Full scale range (N2)| ( 0~2,3,5,10,20,30,50,100,200,300,500)SCCM (
0~1,2,3,5,10,20,30)SLM
Accuracy| ±1.0% S.P. (≥35% F.S.) ±0.35% F.S. (<35% F.S.)
Linearity| ±0.5% F.S.|
Repeatability| ±0.2% F.S.|
Response Time| ≤ 0.8 sec (SEMI E17-0600)|
Valve Rest Position| Normally Closed or Normally Open| No Valve
** Differential Pressure| (0.05~0.35) MPa (≤10slm) (0.1~0.35) MPa(>10slm)|
<0.02MPa
Maximum Operating Pressure| 0.45MPa|
Temperature Coefficient| Zero: ≤±0.02% F.S./℃;
Span: ≤±0.05% F.S./℃
Proof Pressure| 3MPa (435pisg)|
Leak integrity| 1×10-10 atm·cc / sec He|
Wetted Materials| Metal|
Surface Chemistry| ——|
Surface Finish| 25Ra|
Operation| (0~50) ℃|
Temperature| |
Input Signal| Digital: RS485 or ProfiBus or EtherCAT or DeviceNetTM
Analog** : (0~5)VDC or (4~20)mA or (0~20)mA| N/A
Output Signal| Digital: RS485 or Device Net or ProfiBus or EtherCAT
Analog:(0~5)VDC or (4~20)mA or (0~20)mA
Power Supply| ±8 ~ ±16 VDC or +14 ~ +28 VDC(400mA) or +20.4 ~ +26.4 VDC
Electronic Connector| 9 pin male sub-D, 15 pin male sub-D, DeviceNetTM,
ProfiBus , Analog, EtherCAT
Fittings| VCR1/4” M; Compression FittingΦ6, Compression FittingΦ3, Compression
Fitting1/4”, W-seal, C-seal|
Weight| 1.2kg| 1kg
Notes
MFC/MFM is calibrated by N2 as a standard gas.
Units: SCCM (Standard Cubic Centimeter/Min);
SLM (Standard Liter/Min)
Standard Situation: Tem — 273.15K ( 0 ºC );
Air Pressure — 101325 Pa (760mm Hg)
For Sevenstar MFC/MFM, the unit of SCCM is identical to “mL/min, 0 ºC ,1atm”,
and the unit of SLM is identical to “L/min, 0 ºC ,1atm”.
F.S. : Full Scale
Calibration Features
Normally, the MFC are calibrated close to customer’s requirement ( the requirement should be accorded with the specification). Without customer’s information, the MFC are calibrated under standard conditions.
Standard conditions
Without special conditions specified by the customer, the MFC is calibrated
under the
following standard conditions:
Pressure Outlet : Atmospheric.
Normally gas mass flow rate is transferred to gas volume flow rate at standard
state.
Mass flow rate unit:
SCCM—— standard cubic centimeter per minute.
SLM—— standard liter per minute.
Standard state: Temperature —— 0ºC (273.15K)
Pressure —— 101325Pa (760mmHg)
At standard state, the gas density will be constant. The multiplication of
density and volume flow rate is equal to the mass flow rate. Therefore at the
standard state, the volume flow rate can represent mass flow rate.
The standard mounting position is horizontal, And other positions such as
vertical(inlet up or down),flatwise or customized position are optional. The
mounting position should be specified by the customer to ensure the best
accuracy.
Manufacturing Environment
The MFC are assembled in a class 100 clean room, calibrated, packaged and
controlled in class 1000 environment. The temperature is 22±3℃.
Precision Adjustment
Each MFC is accurately controlled for 24 hours after manufacturing on a
different calibration bench. The accuracy, the dynamic response, the stability
to pressure variations is double checked, only qualified product available for
sale.
INSTALLATION
General
WARNING: Toxic, corrosive or explosive gases must be handled with extreme
care. After installing the MFC, the system should be thoroughly checked to
ensure it is leak free.
Purge the MFC with a dry inert gas for one hour before using corrosive gases.
IMPORTANT: When installing the MFC, ensure that the arrow on the back of
the unit in the same direction as the gas flow.
Unpacking
The CS200-D MFC/MFM are assembled, calibrated and clean packaged under clean
room conditions. These series products are packaged with two separately sealed
plastic
bags. The outside is common plastic bag, the inside is clean bag. The outside
bag should be removed in the entrance to the clean room. In order to minimize
the contamination, the second clean bag should be removed in the clean room
when MFC installed in the system.
Mechanical Installation
General
Most applications will require a positive shutoff valve in line with the MFC.
Pressurized gas trapped between the two devices can cause purge effects, and
consideration must be given to the sitting of the shutoff valve (upstream or
downstream) in relation to the process sequencing. It is recommended that you
install an in-line filter upstream to the
controller in order to prevent MFC from contamination.
CS200-D should be mounted in the position accord with the requirement in the
purchase order. The gas should be clean and dry. The mounting should be free
from shock or vibration. The pictures of the product are shown in
figure2-1,the dimensions of the product are shown in figure2-2,
figure2-3a,figure2-3b,figure2-4 and figure2-5. Different fittings(Compression
Fittingφ6, Compression Fittingφ3,Compression Fitting1/4 , Compression
Fitting1/8,VCR1/4 M, Compression Fitting3/8,1.5〞W-seal,1.5〞C seal) are
optional for customer, The length of the product L are shown in Table 2-1. Do
not remove the protective end caps of the fittings until installation.
Table 2-1 The length of the product with different fittings
Fitting
/
Length| Compression Fittingφ6; Compression Fittingφ3 Compression Fitting1/4;
Compression Fittingφ10 Compression Fitting1/8; Compression Fitting3/8|
Ф4(inner)×1hose| VCR1/4 〞 M
VCO1/4 〞 M
Ф6(inner)×1hose Ф5(inner)×1.5hose| 1.5〞W-seal 1.5〞C-seal A-seal
---|---|---|---|---
L (mm)| 112.8| 117.6| 124| 105.2
Attention :
The height (which is showing in figure 2-2 and figure 2-4 ) of 132mm are height without electric connectors of cable. It should be added around 50mm more after adding the electric connector.
Installation
Place MFC according to the flow direction.
1/4VCR Connection
Refer to figure 2-6 and figure 2-7. Check the gland to gland space, including the gaskets. Remove the plastic gland protector caps. When using loose VCR original style gaskets, inserting the gasket into the female nut. For VCR retainer gaskets, snap the gasket onto the male coupling. Tighten the nuts finger tight. Scribe both nut and body in order to mark the position of the nut. While holding the body with a wrench, tighten the nut: 1/8 turn past finger tight for 316L stainless steel and nickel gaskets.
Two-ferrule (Compression Fitting) Connection
Refer to figure 2-7. Check the gland to gland space. Remove the gland
protector caps.
Insert the tubing to the shoulder inside the fitting, and check that the
ferrules are positioned as shown in figure2-7. Tighten the nuts finger tight.
Use two spanners, one spanner locking the fitting stable, wrench another one
to tighten in 1.25 turns to prove it’s not blow-by after installing the front
ferrule, back ferrule and nut.
Attention :
When installing the fitting, you should manually use spanner to wrench it tighten by 1/2 turn pulling up, (imported Compression Fitting should use spanner to be tighten in 1,25 turns) to prove its not blow-by after your installing the front ferrule 、back ferrule and nut. Please note you should use two spanners to operating, one spanner for locking the fitting stable and another one for revolving the nut. Especially when you dismantle the tube, you should operate by the two spanners otherwise fitting will become flexible which will affect its airproof function.
Electrical Installation
General
With simple switching power supply, CS200-D MFC/MFM are available for ±8 to
±16 VDC (dual-ending) and +14 to +28 VDC (single-ending). Customer can choose
as need.
Customer can choose 9 pin male Sub-D or 15 pin male sub-D connector. 9 pin
male Sub-D connector is the SEMI Standard compatible, only 0-5V analog signal
control and output available. 15 pin male sub-D connector, both 4-20mA or
0-20mA and 0-5V analog signal control and output available. CS200-D MFC/MFM
can communicate with PC via RS485 ,DeviceNet or ProfiBus or EtherCAT.
Connections
The 9 pin male Sub-D connector, 15 pin male sub-D connector, RS485 connector ,DeviceNet connector ,Analog Signal Interface ,RS232-RS485 connector, ProfiBus connector of 9-pin female sub-D and ProfiBus connector of 15-pin male sub-D are shown in figure2-8, figure2-9 , figure 2-10 ,figure 2-11, figure 2-12 , figure2-13, figure2-14 , figure2-15 , figure2-16 and figure2-17.
Attention :
Although sharing with the same appearance of CS200-D MFC, the 0~5V Setpoint Input, the 4~20mA or 0~20mA Setpoint Input and the valve Over-ride are not available. That means pin1 and pin6 of D-sub 9 and pin1, pin7, pin8 and pin12 of D-sub 15 are not available
Table of Selecting Type of Cable and its Illustration of Connection
Table 2-2 Table of Selecting Type of Cable
| D08- 2B/3B/4B 2E/3E/4E| D08- 1/2/4 2F/3F/4F D08 1F/1FM/1FS/8C/8CM/1G/1GM|
Input ±15V| Input +24V| To Serial of PC| To USB of Computer
---|---|---|---|---|---|---
MFC (DB15 Pin)| QCX-19/ QCX-P19 QCX-48|
QCX-17/QCX-P17 QCX-46
| QCX-41| QCX-43| QCX-34| QCX-50
MFC (DB9 pin)| QCX-20/ QCX-P20 QCX-49| QCX-18/QCX-P18 QCX-47| QCX-42| |
QCX-34| QCX-50
MFC (EtherCAT)| | | | | QCX-34| QCX-50
Sevenstar manufactures all standard cables, RS485 adaptor, customized cables and some accessories. For more details, contact Sevenstar or its local agent.
Checking Before Operation
Before operating the MFC the following checks should be completed:
- Check that tubing is leak free.
- Check the process sequence and proper function of all other gas components involved.
- Check the voltage of command signals and power supply to the MFC/MFM.
- Check that the appropriate type of gas is being supplied at the rated pressure.
- Allow the MFC to warm up for 20 minutes, and then check the zero level output.
- Use dry inert gas for test runs.
- Prior to using the MFC for extremely corrosive gases, purge MFC with a dry inert gas for one hour.
FUNCTIONS
General
Based on new sensor driver technology, zero balance technology and VCP technology, CS200-D MFC/MFM presents high performance and reliability. And more digital functions are developed in CS200-D products.
Control Mode
CS200-D MFC/MFM are available for digital,0-5V voltage and 4-20mA or 0-20mA
current control mode and output. When customer choose one of the three control
modes, the others will be screened. The analog output will be available at all
times. The flow output of the CS200-D MFC/MFM will be available through the
RS485 interface, even in the analog control mode. The setting of the CS200-D
MFC/MFM can be provided to the MFC through one of three sources, in digital or
analog mode.
For more information, please refer to
www.mfcsevenstar.cn.Path: Service >>
Downloads>>Software download>>Communication Protocol .
Zero
The Zero function can be commanded through the digital interface or zero- button. Before zero MFC, please make sure that no gas flow through MFC. Then zero MFC by the digital interface or zero-button. The zero-button must be pressed continuously for 0.5 seconds in order to start zero process. The green LED will blink during the zero process. After zero finished, the green LED will be on constantly.
Soft-Start
The CS200-D MFC/MFM support soft-start function. Soft-start allows customer change the setting of MFC with definite rate. For more information, please refer to www.mfcsevenstar.cn.Path:Service>> Downloads>>Software download>>Communication Protocol .3.5 Delay Delay is used to postpone the start of flow from zero flow to the received set-point. It is programmed in millisecond but the MFC internally rounds up any value to 100ms. For example, when the delay value is 200, MFC will delay 200ms then receiving the operate command. When the set point is less than the m
in control rate, the valve will shut off, and when larger than the min control
rate, MFC will start after the set delay time.
Special default: Values from 1 to 49 ms will be programmed as 100ms. Delay
applies to digital and analog set-points.
Valve Command Mode
CS200-D MFC/MFM are available for valve-close or valve-open by input digital signal or analog voltage directly. The Valve Command Mode is used to select one of three ways to interpret the analog signal of Valve Command. For more information, please refer to www.mfcsevenstar.cn.Path:Service>>Downloads>>Softwaredownload>>Communication Protocol .
Valve Type
CS200-D MFC/MFM have two types of valve: Normally open(NO) or Normally closed(NC) . When MFC do not have power supply, the “NO” type valve is open and gas can flow through the MFC; the “NC” type valve is closed and gas can not flow through. Please mention the valve type when you order the MFC.
Multi-Gas and Multi-Flow
The multi-gas and multi-range technology has been developed in the CS200-D
series MFC/MFM. Customer can change gas convert factor, full scale of MFC via
digital interface. Full scale of CS200-D series MFC/MFM can be re-ranged from
30% to 110%F.S.. For example, an MFC with 100SCCM full scale, the new full
scale can be re-ranged between from 30SCCM to 110SCCM.
CS200-D MFC/MFM support customer offset by the target null value. The target
null value is a customer-programmed constant used to offset the flow output
independently of all other sensor offset constants, including the zero
process. For example, set Target Null Value: -20%F.S.,then when no flow
through MFC/MFM,the reading is -20%F.S. with gas flow 20%F.S. through
MFC/MFM,the reading is 0%F.S.。Offset range is from -100%F.S. to100%F.S.。
For more information, please refer to www.mfcsevenstar.cn.Path: Service >> Downloads>>Software download>>Communication Protocol.
Total Flow Accumulator
The Total Flow accumulator will record the amount of gas (in SCC) that has
been delivered by CS200-D MFC/MFM. The amount of purge will not be accumulated
in the total flow amount. For example, the total reading is 3000, means the
amount of gas delivered by MFC is 3000SCC.
For more information, please refer to
www.mfcsevenstar.cn.Path: Service >>
Downloads>>Software download>>Communication Protocol .
Alarm
The CS200-D MFC/MFM will monitor and store certain abnormal conditions as described below. These conditions can be read and reset from the RS485 interface. Provisions will be made for masking (disabling) the alarms or warnings on an individual basis.
Warning and Alarms: Sensor Zero Output Out of Bounds
EEPROM Failure
Valve Coil Failure or Valve Disconnected
Temperature Out of Operating Range
After power up the LED on the top of the CS200-D MFC/MFM is turned GREEN. A warning condition will be announced by a blinking RED and whenever an alarm condition is detected it will be set continuously to RED. For more information, please refer to www.mfcsevenstar.cn.Path:Service>> Downloads>>Software download>>Communication Protocol .
LED
There is a green-red LED located on the top of the MFC. The constant green LED
indicates the power on. Green blinking indicates that the MFC is zeroing. A
blinking red LED indicates warning condition. Constant red indicates error
condition.
In DeviceNet connection, there will be 2 LED on the top of CS200-D, refer to
CS200MFC(CS220)_DnetSpecification_V1.01 for more details.
CS200-D ProfiBus communication product has two LED indicator lights on the
top. When ProfiBus communication is normal, the green indicator light is
always on. Not bright indicates that ProfiBus communication is faulty. The
status of the white indicator light refer to 485 communication products.
Attention :
When the valve of MFC is fully opened , it can be functioned as a MFM. In that case, the maximum flow testing voltage could reach beyond +10V, please be careful, while flow is beyond F.S. +5V(Full Scale), the real flow will have no linearity corresponding with flow testing voltage. While it’s purging, flow display will be inaccurate, even showing “reduce” while the real flow is enhanced, please be sure it’ll be no damaged to device itself.
MAINTENANCE
General
No routine maintenance is required to be performed on the MFM or MFC, Other
than occasional cleaning and re-calibration:
It can be used about 3 or 4 years with an ultra-clean and non corrosive gas.
It can be used about 1 or 2 years with a low purity gas or a corrosive gas.
For any other problems, contact Sevenstar.
Caution
Medium Forbidden
The used gas should be purified without dust, liquid and oil stain. If necessary, the filter should be added to gas system for purification. If the outlet of MFC is connected to liquid sources, a One-way valve should be added to avoid the liquid back to destroy MFC.
Attention :
For the cases that the MFC/MFM used with toxic, pyrophoric, flammable or corrosive gas, you should ensure that the fixing and fitting are airtight. It becomes necessary to remove the controller from the system, purge the controller thoroughly with a dry inert gas such as nitrogen, before disconnecting the gas connections. Failure to purge the controller could cause a fire or explosion resulting in death.
Seal of Valves
The Solenoid Valve of MFC is only for adjustment, can not be used for shut-
off. Generally, shut-off valves should be added in upstream and downstream of
MFC for protection. Normally leakage of MFC valve is not more than 1% F.S..
TROUBLESHOOTING
Initial Check
- Check the gas supply pressure and check the flow-path to the MFC/MFM has been opened.
- Ensure that the power supply and command signals are correctly transmitted to the D-connector pins and RS485.
- Check that the output signal matches the external reading.
Troubleshooting
Use the following table to locate the fault.
| SYMPTOMS| Possible cause| Action
---|---|---|---
1| Output reading, without gas flow, is not zero| Gas flow is actually
present| Check closure of series shutoff valve
Zero drift| Zero MFC
other| Contact Sevenstar
2| MFC will not control| Gas connection incorrect| Check gas connection
Pressure incorrect| Check pressure condition
Wrong control mode| Change the control mode by the software
Power failure| Check Power and pin position
Setpoint incorrect| Check setpoint signal
Contamination| Contact Sevenstar
Defective sensor| Contact Sevenstar
PCB problems| Contact Sevenstar
Defective Mechanics| Contact Sevenstar
3| MFC will not communicate with PC| Power failure| Check Power and pin
position
Cable problems| Check cable and connector
Address conflict| Check address of MFC
Baud rate error| Check baud rate of MFC
PCB problems| Contact Sevenstar
Attention :
- Mark indicates that reparation and adjustment must be dealt under specialist advices. For any other problems, contact Sevenstar.
WARRANTY AND SERVICES
Guarantee of Sevenstar
Sevenstar and its authorized distributors assure that there are not flaw of
the material and quality of product within 12 months since the date of the
product purchased by you.
The compensation for customer is only limited to invalid part for
substitution, installation and processing flaw.
It is guaranteed that all part chose by customers are suitable to relative
manufacturer.
Other relative statements, guarantee and obligation of status and usage of
product, whether direct or indirect, are definite to be excluded. In any
circumstance, Sevenstar and its authorized distributors are not charged of any
obligation of direct or indirect loss for customers or others.
Product Warranty
- Sevenstar products are guaranteed against defects in materials and workmanship if used in accordance with specifications and not subject to physical damage, contamination, alteration or retrofit. Warranty periods: One = year.
- Buyers undertake to check and inspect the goods and to notify Sevenstar of shipment incidents by fax, phone or e-mail as soon as possible after receipting the goods.
- During the warranty period, products must only be repaired by authorized Sevenstar service centers; Otherwise, the Sevenstar product warranty will be invalidated.
- Repairs will be performed free of charge during the one-year warranty period. If MFC are out of warranty, Sevenstar will notify the owner of replacement or repair costs before proceeding. Factory service and repairs are = guaranteed 90 days. The warranty excludes consumable materials and wear parts (in teflon, viton, etc.).
- No MFC will be accepted for repair or warranty without a decontamination and purge certificate.
- Each MFC is individually checked (visual inspection of fittings, helium leak test and flow calibration). Sevenstar shall not be responsible for any damage caused by gas leakage or the use of a dangerous gas. Users are responsible for following the safety rules applicable to each gas they use. Improper use of a Sevenstar MFC will void the warranty, and MFC that have been damaged as a result of improper use will not be replaced by Sevenstar.
- Specific warranty requirements are as follows :
A, Gas must be clean and particle-free, which means a filter must be
fitted in the gas line upstream of the MFC.
B, Gas must comply with the following pressure specifications:
- Gas pressure must never exceed 3MPa.
- Differential pressure must be more than 0.05MPa for full-scale flow through the MFC valve unless another value is specified in the user’s manual.
- Differential pressure must be less than 0.35MPa for the MFC valve to regulate without gas-flow oscillation unless another value is specified in the user’s manual.
- Pressure at the mass-flow inlet must be regulated by an accurate pressure regulator to prevent gas-flow oscillation.
C, Electrical connection requirements are as follows:
The system must be wired carefully: non-observance of the pin-out may irreversibly damage the electronic board inside the MFC, in which case the warranty will be invalidated.
D, Gas connections: The fittings must be handled carefully. Sevenstar guarantees that all fittings have been individually inspected and are scratch- free.
E, Fitting procedure: The fitting procedure set out in the manual must be followed meticulously. Specifically, the purge procedure is very important if corrosive gases or toxic gases are used.
F, The mass-flow must not be dismounted: The MFC warranty will be invalidated if the seal between the MFC block and cover is torn.
Services
Sevenstar can provide services like start-up service, software development,
gas system design, training, etc.
Please visit www.mfcsevenstar.cn for more
information and find your nearest service and calibration centre.
Disclaimer
Beijing Sevenstar Flow Co., Ltd is not responsible to loss as following situation:
- Nature disaster and calamity;
- Unsuitable operation and unreasonable usage;
- Operating and storing in inappropriate or execrable circumstance;
- Usage of instrument beyond user’s manual;
- Unauthorized change or replacement of product.
For example:
It is whether that gas path is not cleared before using corrodible gas or MFC
is contaminated or blocked by particle such as dust.
AppendixⅠ CS200 Selection Guide
Differential Pressure:
(0.05 ~ 0.35) MPa (7.3~50.8 psid) (FLOW≤10SLM)
(0.1 ~ 0.35) MPa (14.5~50.8 psid) (10SLM<FLOW≤30SLM )
(0.2 ~ 0.45) MPa (29.0~65.3 psid) (30SLM<FLOW)
CS200-D XXXN: <0.02MPa (2.9psid)
Proof pressure : 3 MPa (435.1 psig)
Calibration temperature : (22±3)℃
Valve command mode : 2
Power Supply: +24V
Command type: DeviceNet default digital mode; Profibus default mode for
Profibus; other analog voltage
-[ S ] Customized Requirement
For Example: I/O sign:4~20mA;
The proportion of mixture gases should be indicated: N2 (60%) + CO2(40%);
Customer Differential Pressure: (0.05 ~ 0.3) MPa;
Customer Calibration Temperature : 40℃;
The letters on cover and tag: in Chinese;
Valve command mode: 0;
Power Supply: ±15V
Command type: Profibus default mode for analog voltage and other
customized requirements.
Take CS200-A013C500CDGGVHS as the example
APPENDIX II CONVERSION FACTOR
GAS| GAS CODE (SEMIE52-0302)| SPECIFIC HEAT ( Cal/g ℃)|
DENSITY (g/l 0℃)| CONVERSION FACTOR
---|---|---|---|---
Air| 008| 0.2400| 1.2930| 1.001
Ar| 004| 0.1250| 1.7837| 1.420
AsH3| 035| 0.1168| 3.4780| 0.673
BBr3| 079| 0.0647| 11.1800| 0.378
BCl3| 070| 0.1217| 5.2270| 0.450
BF3| 048| 0.1779| 3.0250| 0.508
B2H6| 058| 0.5020| 1.2350| 0.441
CCl4| 101| 0.1297| 6.8600| 0.306
CF4| 063| 0.1659| 3.9636| 0.420
CH4| 028| 0.5318| 0.7150| 0.722
C2H2| 042| 0.4049| 1.1620| 0.596
C2H4| 038| 0.3658| 1.2510| 0.597
C2H6| 054| 0.4241| 1.3420| 0.482
C3H4| 068| 0.3633| 1.7870| 0.421
C3H6| 069| 0.3659| 1.8770| 0.411
C3H8| 089| 0.3990| 1.9670| 0.358
C4H6| 093| 0.3515| 2.4130| 0.322
C4H8| 104| 0.3723| 2.5030| 0.299
C4H10| 117| 0.4040| 2.6500| 0.261
C5H12| 240| 0.3916| 3.2190| 0.217
CH3OH| 176| 0.3277| 1.4300| 0.584
C2H6O| 136| 0.3398| 2.0550| 0.392
C2H3Cl3| 112| 0.1654| 5.9500| 0.278
CO| 009| 0.2488| 1.2500| 1.000
CO2| 025| 0.2017| 1.9640| 0.739
C2N2| 059| 0.2608| 2.3220| 0.451
Cl2| 019| 0.1145| 3.1630| 0.858
D2| 014| 1.7325| 0.1798| 0.997
F2| 018| 0.1970| 1.6950| 0.931
GeCl4| 113| 0.1072| 9.5650| 0.267
GeH4| 043| 0.1405| 3.4180| 0.570
H2| 007| 3.4224| 0.0899| 1.010
HBr| 010| 0.0861| 3.6100| 0.999
HCl| 011| 0.1911| 1.6270| 0.988
HF| 012| 0.3482| 0.8930| 1.001
HI| 017| 0.0545| 5.707| 1.000
H2S| 022| 0.2278| 1.5200| 0.802
He| 001| 1.2418| 0.1786| 1.420
Kr| 005| 0.0593| 3.7390| 1.431
N2| 013| 0.2486| 1.2500| 1.000
Ne| 002| 0.2464| 0.9000| 1.431
NH3| 029| 0.5005| 0.7600| 0.719
NO| 016| 0.2378| 1.3390| 0.978
NO2| 026| 0.1923| 2.0520| 0.737
N2O| 027| 0.2098| 1.9640| 0.710
O2| 015| 0.2196| 1.4270| 0.981
PCl3| 193| 0.1247| 6.1270| 0.358
PH3| 031| 0.2610| 1.5170| 0.690
PF5| 143| 0.1611| 5.6200| 0.302
POCl3| 102| 0.1324| 6.8450| 0.302
SiCl4| 108| 0.1270| 7.5847| 0.284
SiF4| 088| 0.1692| 4.6430| 0.348
SiH4| 039| 0.3189| 1.4330| 0.600
SiH2Cl2| 067| 0.1472| 4.5060| 0.416
SiHCl3| 147| 0.1332| 6.0430| 0.340
SF6| 110| 0.1588| 6.5160| 0.258
SO2| 032| 0.14890| 2.8580| 0.687
TiCl4| 114| 0.1572| 8.4650| 0.206
WF6| 121| 0.0956| 13.2900| 0.217
Xe| 006| 0.0379| 5.8580| 1.431
Conversion Factors Instruction:
MFC and MFM are standard calibrated by N2 while it’s out of factory. Other gas
calibrations can be approximated by converting of conversion factors of our
instruction. While using other gas operating:
One single gas: The conversion factors could find out in the users
specification instruction.
A mixture of two or more gases: Assume there is “n” sorts of gases, could
calculate the conversion factors C by the following formula:
Basic Formula:
C= 0.3106 N /ρ(Cp)
ρ — Density of the gas
Cp — Specific heat of the gas
N — Structure factors of gas-molecule (See the following table)
Table . Gas-Molecule Composing factors
COMPOSITION | EXAMPLE | N VALUE |
---|---|---|
Single atom numerator | Ar He | 1.01 |
Double atom numerator | CO N2 | 1.00 |
Tree atom numerator | CO2 NO2 | 0.94 |
Multi-atom numerator | NH3 C4H8 | 0.88 |
For mixture gases: N = N1 (ω1/ωT )+N2 (ω2/ωT ) + ···+ Nn (ωn/ωT )
Then:
ω1 …ωn — The flow of single gas
ωT — The flow of mixture gas
ρ1…ρn — The density of single gas
CP1…CPn —Specific heat of the single gas
N1…Nn —Structure factors of gas-molecule (See Table 6.)
Attention
- Standard: Temperature 273.15K (0 ºC ); Air Pressure— 101325 Pa (760mm Hg)
- Please feel free to contact us if the request gas conversion factors could not be found in our appendix
Software Download Address:
Please register first and then download software.
Note: Please contact our people if you need disk to download software.
Email address is
weidongxia@sevenstar.com.cn or
mfcoversea@sevenstar.com.cn.
CS Series
Mass Flow Controller
Beijing Sevenstar Flow Co., Ltd.
Address: No.8 Wenchang Avenue Beijing Economic-Technological Development
Area
Post code: 100176
Tel: (+86)-10-56178088
Fax: (+86)-10-56178099
Homepage:www.mfcsevenstar.cn
E-Mail: mfcsales@sevenstar.com.cn
Shanghai Office: Room802-803, Building 3, No. 88 Shengrong Road, Pudong New District, Shanghai Tel: (+86)-21-63532370
Shenzhen Office: Room 202, Section B, No. 1 Chuangjin, No.125 Chuangye Erlu,
28th Section, Baoan\ District, Shenzhen
Tel: (+86)-755-88290258
Fax: (+86)-755-88294770
*Description may be changed following improvements to product. The information contained in this document is subject to change without notice.
*If there is any mistake in this uses manual, please feel free to contact us.